kSIST FprEN ISO 13169:2026
(Main)Water quality - Uranium - Test method using alpha liquid scintillation counting (ISO/FDIS 13169:2026)
- Abstract
This document specifies the measurement method for the determination of total activity concentration of uranium isotopes in non-saline waters by extraction and liquid scintillation counting.
This method covers the measurement of soluble uranium isotopes in water in activity concentrations between approximately 2·10−3 Bq/kg and 10 Bq/kg when analysing a 1 l test sample volume with a 60 000 s counting time with a typical alpha LSC instrument.
The ratio 234U/238U can also be determined. This method has not been tested for the measurement of other uranium isotopes.
- Status
- Not Published
- Public Enquiry End Date
- 02-Feb-2026
- Technical Committee
- KAV - Water quality
- Current Stage
- 5020 - Formal vote (FV) (Adopted Project)
- Start Date
- 13-Oct-2026
- Due Date
- 01-Dec-2026
kSIST FprEN ISO 13169:2026 is a SIST draft of an ISO water-quality standard for measuring uranium isotopes in non-saline waters by extraction and alpha liquid scintillation counting. It tells laboratories how to prepare the sample, calibrate the instrument, calculate activity concentration, and report the result; it also allows determination of the 234U/238U ratio.
What does kSIST FprEN ISO 13169:2026 specify?
kSIST FprEN ISO 13169:2026 specifies a method for determining the total activity concentration of uranium isotopes in non-saline waters. The method is based on extraction followed by liquid scintillation counting, and the scope states that it is for soluble uranium isotopes in a defined activity range with a 1 l test sample and a typical 60 000 s counting time.
The document also states that the 234U/238U ratio can be determined from the measurements. It has not been tested for other uranium isotopes, so users should not assume coverage beyond the isotopes named in the standard.
| Annex | What it covers |
|---|---|
| Annex A | Set-up parameters and validation data, including example spectra, spillover behaviour, calibration parameters, characteristic limits and performance checks. |
The document is organized into Clauses 1 to 11: Scope, Normative references, Terms and definitions, Principle, Sampling, Reagents and equipment, Instrument set-up and calibration, Procedure, Quality assurance and quality control, Expression of results, and Test report.
What are the key requirements of kSIST FprEN ISO 13169:2026?
kSIST FprEN ISO 13169:2026 is built around a practical laboratory workflow: take a representative non-saline water sample, extract uranium into a scintillation cocktail, count it with a liquid scintillation counter, and calculate activity with uncertainty and decision limits. The most important points are the sampling controls, the discriminator setting, efficiency verification, interference control, and the way results are reported.
Sampling and sample preparation
Clause 5 requires sampling, handling, and storage to follow the cited water-sampling guidance, and the sample has to remain representative from collection to analysis. The water is stored in plastic bottles, filtered to remove suspended matter, acidified with nitric acid, and kept at 4 °C if it is not analysed immediately.
Clause 5 also describes two preparation paths: direct measurement and pre-concentration by evaporation. In practice, this lets the laboratory choose between a simpler routine procedure and a more sensitive one when lower detection limits are needed, especially for soft waters.
Reagents and equipment
Clause 6 specifies analytical-grade reagents and the main laboratory items: a balance, a hot plate with magnetic stirring, a liquid scintillation counter with α/β discrimination, and polyethylene scintillation vials with PTFE coating. The standard also names the acids, scintillation cocktail components, and radioactive standard solutions used for calibration.
This matters because the counting performance depends strongly on the vial material, the cocktail composition, and the background of the counter. The text explicitly prefers low-background counters and PTFE-coated vials because they improve discrimination and reduce unwanted radon uptake.
Instrument set-up and calibration
Clause 7 requires the extractant scintillation cocktail to be prepared first, then the alpha emitter calibration source and beta emitter calibration source to be made with the named standards. The working point is chosen by testing several discriminator settings and selecting the one that minimises alpha and beta spillover.
The standard treats extraction efficiency and counting efficiency together as alpha efficiency. In practice, the laboratory must define acceptance limits, verify efficiency periodically, and repeat verification after changes in materials, staff, or maintenance.
Measurement procedure and interference control
Clause 8 gives two procedures: one without pre-concentration and one with evaporation pre-concentration. After acidification and heating, or after concentration, the sample is extracted twice into the scintillation cocktail, the vial is purged with argon or nitrogen, cleaned with ethanol, and then counted under the selected conditions.
Clause 9 warns that thorium isotopes can be co-extracted and can interfere with uranium measurement. If thorium peaks are visible, the standard says an alternative method shall be used for uranium quantification, so laboratories need a check for this interference before accepting the result.
Quality assurance, uncertainty and decision limits
Clause 9 requires quality control consistent with ISO/IEC 17025. Clause 10 then sets out how to calculate the activity per unit mass, the total activity concentration, the standard uncertainty, the decision threshold, the limit of detection, and the coverage intervals using the ISO 11929 framework.
In practice, this means the result is not just a count rate. The laboratory must also show the uncertainty and decide whether the measured activity is above the decision threshold or detection limit, using the stated statistical treatment.
Test report
Clause 11 requires the report to identify the sample, state the units, give the result in the correct decision form, and record the date used for the mass-activity calculation and the report issue date. The report may also include sampling and receipt dates, uncertainty intervals, decision probabilities, and the decision threshold and detection limit.
What terms does kSIST FprEN ISO 13169:2026 define?
- Activity concentration - the activity of uranium isotopes expressed per unit volume, used when the result is reported for water.
- Total mass activity - the total activity of uranium isotopes per unit mass of sample, before conversion to concentration by density.
- Decision threshold - the level used to decide whether the measured signal is distinguishable from zero with the chosen false-positive probability.
- Limit of detection - the smallest activity level that the method is intended to detect with the chosen false-negative probability.
- Alpha efficiency - the combined extraction and counting efficiency for the uranium measurement.
- Spillover parameter - the amount of alpha counts seen in the beta window, or beta counts seen in the alpha window, used to select the working point.
- Coverage interval - the interval around the result that expresses uncertainty at a stated probability, either as a probabilistically symmetric interval or as the shortest interval.
Who uses kSIST FprEN ISO 13169:2026?
kSIST FprEN ISO 13169:2026 is used by water-testing laboratories, radiological analysis laboratories, and environmental monitoring teams that need uranium results in non-saline waters. It is also relevant to quality managers, technical assessors, and buyers who need a defined laboratory method for routine monitoring, compliance checks, or emergency response measurements.
The introduction links the method to drinking-water monitoring and to situations where national limits, guidance levels, or emergency conditions require low detection limits. The standard therefore fits laboratories that need defensible uranium results with stated uncertainty, not just a count rate.
What changed in kSIST FprEN ISO 13169:2026 from the previous edition?
kSIST FprEN ISO 13169:2026 replaces ISO 13169:2018. The foreword states that this second edition has a technically revised introduction, an updated Clause 8 on expression of results, and an updated bibliography.
Which standards are used with kSIST FprEN ISO 13169:2026?
- ISO 5667-1 - sampling programme design and sampling techniques.
- ISO 5667-3 - preservation and handling of water samples.
- ISO 5667-14 - quality assurance and quality control of environmental water sampling and handling.
- ISO 80000-10 - quantities and units for atomic and nuclear physics.
- ISO 11929 (all parts) - decision threshold, detection limit, and coverage interval calculations for ionizing-radiation measurements.
- ISO/IEC 17025 - laboratory competence and the quality system expected for testing and calibration work.
- ISO/IEC Guide 98-3 - uncertainty of measurement and the GUM approach used in the result calculations.
- ISO 5667-10 - cited in Clause 5 for sampling guidance.
- ISO 13166:2014 - cited when thorium interference is visible and an alternative uranium method is needed.
What does the kSIST FprEN ISO 13169:2026 document contain?
The document contains the full laboratory method: principle, sampling guidance, reagent and apparatus requirements, cocktail preparation, calibration source preparation, discriminator optimisation, efficiency determination, blank preparation, sample extraction, and sample counting. It also contains the calculation framework for activity, uncertainty, decision threshold, detection limit, and coverage intervals.
Annex A provides worked validation material rather than routine requirements. It includes example spectra, a spillover diagram, calibration parameters, characteristic limits, linearity checks, precision data, accuracy observations, detection-limit behaviour, and an uncertainty example, which helps laboratories judge whether their own setup performs similarly.
Get Certified
Connect with accredited certification bodies for this standard
BSMI (Bureau of Standards, Metrology and Inspection)
Taiwan's standards and inspection authority.
CIS Institut d.o.o.
Personal Protective Equipment (PPE) certification body. Notified Body NB-2890 for EU Regulation 2016/425 PPE.

Kiwa BDA Testing
Building and construction product certification.
Sponsored listings
Frequently Asked Questions
kSIST FprEN ISO 13169:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Water quality - Uranium - Test method using alpha liquid scintillation counting (ISO/FDIS 13169:2026)". This standard covers: This document specifies the measurement method for the determination of total activity concentration of uranium isotopes in non-saline waters by extraction and liquid scintillation counting. This method covers the measurement of soluble uranium isotopes in water in activity concentrations between approximately 2·10−3 Bq/kg and 10 Bq/kg when analysing a 1 l test sample volume with a 60 000 s counting time with a typical alpha LSC instrument. The ratio 234U/238U can also be determined. This method has not been tested for the measurement of other uranium isotopes.
This document specifies the measurement method for the determination of total activity concentration of uranium isotopes in non-saline waters by extraction and liquid scintillation counting. This method covers the measurement of soluble uranium isotopes in water in activity concentrations between approximately 2·10−3 Bq/kg and 10 Bq/kg when analysing a 1 l test sample volume with a 60 000 s counting time with a typical alpha LSC instrument. The ratio 234U/238U can also be determined. This method has not been tested for the measurement of other uranium isotopes.
kSIST FprEN ISO 13169:2026 is classified under the following ICS (International Classification for Standards) categories: 13.060.60 - Examination of physical properties of water; 17.240 - Radiation measurements. The ICS classification helps identify the subject area and facilitates finding related standards.
kSIST FprEN ISO 13169:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI STANDARD
oSIST prEN ISO 13169:2026
01-januar-2026
Kakovost vode - Uran - Preskusna metoda s štetjem alfa delcev s tekočinskim
scintilatorjem (ISO/DIS 13169:2025)
Water quality - Uranium - Test method using alpha liquid scintillation counting (ISO/DIS
13169:2025)
Wasserbeschaffenheit - Uran - Verfahren mittels AlphaFlüssigszintillationszählung
(ISO/DIS 13169:2025)
Qualité de l'eau - Uranium - Méthode d'essai par comptage des scintillations alpha en
milieu liquide (ISO/DIS 13169:2025)
Ta slovenski standard je istoveten z: prEN ISO 13169
ICS:
13.060.60 Preiskava fizikalnih lastnosti Examination of physical
vode properties of water
17.240 Merjenje sevanja Radiation measurements
oSIST prEN ISO 13169:2026 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
oSIST prEN ISO 13169:2026
oSIST prEN ISO 13169:2026
DRAFT
International
Standard
ISO/DIS 13169
ISO/TC 147/SC 3
Water quality — Uranium —
Secretariat: AFNOR
Test method using alpha liquid
Voting begins on:
scintillation counting
2025-12-01
Qualité de l'eau — Uranium — Méthode d'essai par comptage des
Voting terminates on:
scintillations alpha en milieu liquide
2026-02-23
ICS: 17.240; 13.060.60
THIS DOCUMENT IS A DRAFT CIRCULATED
FOR COMMENTS AND APPROVAL. IT
IS THEREFORE SUBJECT TO CHANGE
AND MAY NOT BE REFERRED TO AS AN
INTERNATIONAL STANDARD UNTIL
PUBLISHED AS SUCH.
This document is circulated as received from the committee secretariat.
IN ADDITION TO THEIR EVALUATION AS
BEING ACCEPTABLE FOR INDUSTRIAL,
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
ISO/CEN PARALLEL PROCESSING
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
NATIONAL REGULATIONS.
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION.
Reference number
ISO/DIS 13169:2025(en)
oSIST prEN ISO 13169:2026
DRAFT
ISO/DIS 13169:2025(en)
International
Standard
ISO/DIS 13169
ISO/TC 147/SC 3
Water quality — Uranium —
Secretariat: AFNOR
Test method using alpha liquid
Voting begins on:
scintillation counting
Qualité de l'eau — Uranium — Méthode d'essai par comptage des
Voting terminates on:
scintillations alpha en milieu liquide
ICS: 17.240; 13.060.60
THIS DOCUMENT IS A DRAFT CIRCULATED
FOR COMMENTS AND APPROVAL. IT
IS THEREFORE SUBJECT TO CHANGE
AND MAY NOT BE REFERRED TO AS AN
INTERNATIONAL STANDARD UNTIL
PUBLISHED AS SUCH.
This document is circulated as received from the committee secretariat.
IN ADDITION TO THEIR EVALUATION AS
BEING ACCEPTABLE FOR INDUSTRIAL,
© ISO 2025
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
STANDARDS MAY ON OCCASION HAVE TO
ISO/CEN PARALLEL PROCESSING
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
BE CONSIDERED IN THE LIGHT OF THEIR
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
or ISO’s member body in the country of the requester.
NATIONAL REGULATIONS.
ISO copyright office
RECIPIENTS OF THIS DRAFT ARE INVITED
CP 401 • Ch. de Blandonnet 8
TO SUBMIT, WITH THEIR COMMENTS,
CH-1214 Vernier, Geneva
NOTIFICATION OF ANY RELEVANT PATENT
Phone: +41 22 749 01 11
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION.
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland Reference number
ISO/DIS 13169:2025(en)
ii
oSIST prEN ISO 13169:2026
ISO/DIS 13169:2025(en)
Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms, definitions, symbols and units . 2
4 Principle . 3
5 Sampling . 4
6 Chemical reagents and equipment . 4
6.1 Chemical reagents .4
6.2 Equipment and Apparatus .5
7 Instrument set up and calibration . 5
7.1 Preparation of extractant scintillation cocktail .5
7.2 Preparation of alpha emitter calibration source .5
7.3 Preparation of beta emitter calibration source .6
7.4 Optimization of counting conditions .6
7.5 Detection efficiency .6
7.6 Blank sample preparation and measurement.7
8 Procedure . 7
8.1 Without pre-concentration .7
8.2 Pre-concentration by evaporation .7
8.3 Sample preparation .7
8.4 Sample measurement .8
9 Quality assurance and quality control program . 8
9.1 Interference control .8
9.2 Quality control .8
10 Expression of results . 8
10.1 Calculation of activity per unit of mass .8
10.2 Standard uncertainty .8
10.3 Decision threshold .9
10.4 Limit of detection .9
10.5 Limits of the coverage intervals . .9
10.5.1 Limits of the probabilistically symmetric coverage interval .9
10.5.2 The shortest coverage interval .10
10.6 Calculations using the activity concentration .10
11 Test report . 10
Annex A (informative) Set-up parameters and validation data .12
Bibliography .16
iii
oSIST prEN ISO 13169:2026
ISO/DIS 13169:2025(en)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 147, Water quality, Subcommittee SC 3,
Radioactivity measurements.
This second edition cancels and replaces the first edition (ISO 13169:2018), which has been technically
revised.
The main changes are as follows:
— Modification of the introduction
— Clause 8: updating of the expression of results
— Updating of the Bibliography
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
iv
oSIST prEN ISO 13169:2026
ISO/DIS 13169:2025(en)
Introduction
Radionuclides are present throughout the environment, thus, water bodies (e.g. surface waters, ground
waters, sea waters) contain radionuclides, which can be of either natural, or anthropogenic origin.
3 14 40
— Naturally-occurring radionuclides, including H, C, K, and those originating from the thorium and
210 210 222 226 228 227 232 231 234 238
uranium decay series, in particular Pb, Po, Rn, Ra, Ra, Ac, Th, Pa, U and U,
can be found in water bodies due to either natural processes (e.g. desorption from the soil and runoff by
rain water) or released from technological processes involving naturally occurring radioactive materials
(e.g. mining, mineral processing, oil, gas, and coal production, water treatment and the production and
use of phosphate fertilisers).
55 59 63 90 99
— Anthropogenic radionuclides such as Fe, Ni, Ni, Sr, Tc, transuranic elements (e.g., Np, Pu, Am,
60 137
and Cm) and some gamma emitting radionuclides such as Co and Cs can also be found in natural
waters. Small quantities of anthropogenic radionuclides can be discharged from nuclear facilities into
the environment as a result of authorized routine releases. The radionuclides present in liquid effluents
[1]
are usually controlled before being discharged to the environment and water bodies. Anthropogenic
radionuclides used for medical and industrial applications can be released into the environment after
use. Anthropogenic radionuclides are also found in waters due to contamination fallout resulting from
above-ground nuclear detonations and accidents such as those that have occurred at the Chornobyl and
Fukushima nuclear facilities.
Radionuclide activity concentrations in water bodies can vary according to local geological characteristics
and climatic conditions and can be locally and temporally enhanced by releases from nuclear facilities
[2][3]
during planned, existing, and emergency exposure situations. Some drinking-water sources can thus
contain radionuclides at activity concentrations that can present a human health risk. The World Health
[4]
Organization (WHO) recommends to routinely monitor radioactivity in drinking waters and to take
proper actions when needed to minimize the health risk.
National regulations usually specify the activity concentration limits that are authorized in drinking waters,
water bodies, and liquid effluents to be discharged to the environment. These limits can vary for planned,
existing, and emergency exposure situations. As an example, during either a planned or existing situation,
-1 -1 [4]
the WHO guidance level in drinking water is 10 Bq·l for uranium 238 and 1 Bq·l for uranium 234 see
NOTES 1 and 2. Compliance with these limits is assessed by measuring radioactivity in water samples and
by comparing the results obtained, with their associated uncertainties, as specified by ISO/IEC Guide 98-3
[5]
and ISO 5667-20 .
-1
The provisional guideline value for total content of uranium in drinking water is 30 µg·l based on its
chemical toxicity, which is predominant compared with its radiological toxicity.
NOTE 1 If the value is not specified in Annex 6 of Reference [4], the value has been calculated using the formula
provided in Reference [4] and the dose coefficient data from References [6] and [7].
NOTE 2 The guidance level calculated in Reference [4] is the activity concentration that results in an effective dose
-1 −1
of 0,1 mSv·a to members of the public for an intake of 2 l·d of drinking water for one year. This is an effective
dose that represents a very low level of risk to human health and which is not expected to give rise to any detectable
[4]
adverse health effects .
This document contains method(s) to support laboratories, which need to determine uranium isotopes
activity in water samples. The method described in this document can be used for various types of waters
(see Clause 1). Minor modifications such as sample volume and counting time can be made if needed to
ensure that the decision threshold, detection limit and uncertainties are below the required limits. This can
be done for several reasons such as emergency situations, lower national guidance limits, and operational
requirements.
v
oSIST prEN ISO 13169:2026
oSIST prEN ISO 13169:2026
DRAFT International Standard ISO/DIS 13169:2025(en)
Water quality — Uranium — Test method using alpha liquid
scintillation counting
WARNING — Persons using this document should be familiar with normal laboratory practice. This
document does not purport to address all of the safety problems, if any, associated with its use. It is
the responsibility of the user to establish appropriate safety and health practices.
IMPORTANT — It is absolutely essential that tests conducted according to this document be carried
out by suitably qualified staff.
1 Scope
This document specifies a method to measure uranium isotopes in non-saline waters by extraction and
liquid scintillation counting.
This method covers the measurement of soluble uranium isotopes in water in activity concentrations
−3 -1 -1
between approximately 2·10 Bq·kg and 10 Bq·kg when analysing a 1 l test sample volume with a 60 000 s
counting time with a LSC instrument with α/β discrimination function.
234 238
The ratio U/ U can also be determined. This method has not been tested for the measurement of other
uranium isotopes.
The method is applicable to test samples of non-saline waters after proper sampling, handling, and
preparation.
The detection limit depends on the sample volume, the instrument used, the background count rate, the
detection efficiency, the counting time, and the chemical yield. The limit of detection of the method described
-1
in this document, using currently available liquid scintillation apparatus, is approximately 0,03 Bq·l , which
-1 238 -1
is lower than the WHO criteria for safe consumption of drinking water (10 Bq·l for U and 1 Bq·l for
234 [4]
U). This value can typically be achieved with a counting time of 30 000 s for a test sample volume of 1 l.
The method described in this document is also applicable in the event of an emergency situation.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO 5667-1, Water quality — Sampling — Part 1: Guidance on the design of sampling programmes and sampling
techniques
ISO 5667-3, Water quality — Sampling — Part 3: Guidance on the preservation and handling of water samples
ISO 5667-14, Water quality — Sampling — Part 14: Guidance on quality assurance and quality control of
environmental water sampling and handling
ISO 80000-10, Quantities and units — Part 10: Atomic and nuclear physics
ISO 11929 (all parts), Determination of the characteristic limits (decision threshold, detection limit and limits of
the confidence interval) for measurements of ionizing radiation — Fundamentals and application
ISO/IEC 17025, General requirements for the competence of testing and calibration laboratories
oSIST prEN ISO 13169:2026
ISO/DIS 13169:2025(en)
ISO/IEC Guide 98-3, Uncertainty of measurement — Part 3: Guide to the expression of uncertainty in
measurement (GUM:1995)
3 Terms, definitions, symbols and units
For the purposes of this document, the terms and definitions given in ISO 80000-10, ISO 11929 series and
ISO/IEC Guide 98-3 apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
236 -1
A activity concentration of U, the standard solution Bq·g
S
-1
a total mass activity of uranium isotopes Bq·g
−1
a possible or assumed true quantity values of the activity per mass Bq g
-1
a* decision threshold of the measurand Bq·g
# -1
a limit of detection of the measurand Bq·g
-1
Bq·g
lower and upper limits of the probabilistically symmetric coverage interval
aa,
<> -1
lower and upper limits of the shortest coverage interval Bq·g
aa,
-1
c total activity concentration of the uranium isotopes Bq·l
A
ε alpha efficiency
distribution function of the standardized normal distribution, Φ(k ) = p ap-
p
φ
plies
α, β false positive and false negative probability respectively
m sample mass g
m
standard solution mass g
S
mass of U standard solution used for the preparation of the calibration g
m
S(U)
sample
mass of Sr standard solution used for the preparation of the calibration g
m
S(Sr)
sample
r sample count rate in the alpha window
g
-1
r count rate of the calibration sample in the alpha window s
S
-1
r blank sample count rate in the alpha window s
-1
r
background count rate, from the alpha window s
0,α
-1
r background count rate, of the beta window s
0,β
-1
r
background count rate, from the total window s
0,T
-1
r alpha emitter count rate in beta window when measuring alpha standard s
Sα,β
oSIST prEN ISO 13169:2026
ISO/DIS 13169:2025(en)
-1
alpha emitter count rate in total (alpha+beta) window when measuring alpha s
r
Sα,T
standard
-1
r beta emitter count rate in alpha window when measuring beta standard s
Sβ,α
-1
beta emitter count rate in total (alpha+beta) window when measuring beta s
r
Sβ,T
standard
t blank sample counting time s
t sample counting time s
g
t calibration or sample counting time s
s
-1
u(a) standard uncertainty associated with the measurand Bq·g
u
relative standard uncertainty -
rel
−1
ua() standard uncertainty of a as a function of its true value Bq·g
# −1
standard uncertainty of a when the true value is the detection limit Bq·g
ua
()
-1
expanded uncertainty of the measurand, calculated by U = ku (c ) with k= 2, Bq·g
A
U(a)
…,
τ alpha spillover parameter
α
τ beta spillover parameter
β
quantiles of the standardized normal distribution for the probabilities p (for
k
p
instance p = 1 − α, 1 − β or 1 − γ/2)
k quantiles of the standardized normal distribution for the probabilities q (for
q
instance q = 1−α, 1− β or 1−γ/2)
ρ -1
density of water samples g·l
s(ε) standard deviation of the alpha efficiency
s standard deviation of net alpha cps
αcps
ω auxiliary quantity
V volume of test sample L
-1
w Calibration factor L
4 Principle
234 235 238
Natural uranium is composed of a mixture of three isotopes, all radioactive: U, U and U.
Human activity can modify the natural ratio of uranium isotopes. As well, dissolved uranium in natural
waters can show isotope ratios different from natural ones due to leaching mechanisms.
oSIST prEN ISO 13169:2026
ISO/DIS 13169:2025(en)
Table 1 — Uranium isotopes features
Isotope Mass abundance (%) Half-life (a) Specific activity of natural
-1
uranium (Bq·g )
238 9 4
U 99,3 4,468·10 1,235·10
235 8 2
U 0,7 7,04·10 5,579·10
234 5 4
U <0,1 2,455 ·10 1,243·10
The water sample is first acidified with a fixed amount of nitric acid and then, if a lower detection limit is
required, concentrated by evaporation. Uranium isotopes are then separated (liquid-liquid extraction) by
using a scintillation cocktail containing a complexing agent such as bis(2-ethylhexyl) hydrogen phosphate,
HDEHP and measured by liquid scintillation. The total activity of uranium isotopes can be thus determined.
U standard solution is used for calibration (external standard).
234 238
Spectral deconvolution can be performed using a suitable software to assess the U/ U ratio.
5 Sampling
Sampling, handling, and storage of the water shall be done as specified
...