Standard Practice for Color Measurement of Fluorescent Specimens Using the One-Monochromator Method

SCOPE
1.1 This practice applies to the instrumental color measurement of fluorescent specimens excited by near ultraviolet and visible radiation that results in fluorescent emission within the visible range. It is not intended for other types of photoluminescent materials such as phosphorescent, chemiluminescent, or electroluminescent, nor is this practice intended for the measurement of the fluorescent properties for chemical analysis.
1.2 This practice describes the instrumental measurement requirements, calibration procedures, and material standards needed for the color measurement of fluorescent specimens when illuminated by simulated daylight approximating CIE Standard Illuminant D65 (CIE D65).
1.3 This practice is limited in scope to colorimetric spectrometers providing continuous broadband polychromatic illumination of the specimen and employing only a viewing monochromator for analyzing the radiation leaving the specimen.
1.4 This practice can be used for calculating total tristimulus values and total chromaticity coordinates for fluorescent colors in the CIE Color System for either the CIE 1931 Standard Colorimetric Observer or the CIE 1964 Supplementary Standard Colorimetric Observer.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
30-Jun-2006
Technical Committee
Drafting Committee
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E991-06 - Standard Practice for Color Measurement of Fluorescent Specimens Using the One-Monochromator Method
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E991–06
Standard Practice for
Color Measurement of Fluorescent Specimens Using the
1
One-Monochromator Method
This standard is issued under the fixed designation E991; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
Thefundamentalprocedureforevaluatingthecolorofafluorescentobjectistoobtainspectrometric
data for specified illuminating and viewing conditions, and then use this data to compute tristimulus
values based on an International Commission on Illumination (CIE) standard observer and a CIE
standard illuminant. For a fluorescent object-color specimen, the spectral radiance factors used to
calculate tristimulus values are made up of two components — an ordinary reflectance factor and a
fluorescencefactor(b= b + b ).Themagnitudeofthefluorescentradiancefactors,andconsequently
S F
the measured total radiance factors and derived color values, vary directly with the spectral
distribution of the instrument source illuminating the specimen. Consequently, the colorimetry of
fluorescent object-color specimens requires greater control of the measurement parameters in order to
obtainprecisespectrometricandcolorimetricdata.Inordertoobtainrepeatableandreproduciblecolor
values for fluorescent objects it is necessary that the illumination at the specimen surface closely
duplicate the standard illuminant used in the color calculations. The considerations involved and the
procedures used to obtain spectrometric data and compute colorimetric values for fluorescent
specimens using a one-monochromator spectrometer are contained in this practice.
1. Scope mination of the specimen and employing only a viewing
monochromator for analyzing the radiation leaving the speci-
1.1 This practice applies to the instrumental color measure-
men.
ment of fluorescent specimens excited by near ultraviolet and
1.4 Thispracticecanbeusedforcalculatingtotaltristimulus
visible radiation that results in fluorescent emission within the
values and total chromaticity coordinates for fluorescent colors
visible range. It is not intended for other types of photolumi-
in the CIE Color System for either the CIE 1931 Standard
nescent materials such as phosphorescent, chemiluminescent,
Colorimetric Observer or the CIE 1964 Supplementary Stan-
or electroluminescent, nor is this practice intended for the
dard Colorimetric Observer.
measurement of the fluorescent properties for chemical analy-
1.5 This standard does not purport to address all of the
sis.
safety concerns, if any, associated with its use. It is the
1.2 This practice describes the instrumental measurement
responsibility of the user of this standard to establish appro-
requirements, calibration procedures, and material standards
priate safety and health practices and determine the applica-
needed for the color measurement of fluorescent specimens
bility of regulatory limitations prior to use.
when illuminated by simulated daylight approximating CIE
Standard Illuminant D65 (CIE D65).
2. Referenced Documents
1.3 This practice is limited in scope to colorimetric spec-
2
2.1 ASTM Standards:
trometers providing continuous broadband polychromatic illu-
D985 Test Method for Brightness of Pulp, Paper, and
1
This practice is under the jurisdiction of ASTM Committee E12 on Color and
Appearance and is the direct responsibility of Subcommittee E12.05 on Fluores-
2
cence. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved July 1, 2006. Published September 2006. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1984. Last previous edition approved in 1998 as E991 – 98. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/E0991-06. the ASTM website.
Copyright ©ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA19428-2959, United States.
1

---------------------- Page: 1 ----------------------
E991–06
3
Paperboard (Directional Reflectance at 457 nm) 3. Terminology
D2244 Practice for Calculation of Color Tolerances and
3.1 Definitions: The definitions contained in Guide E179,
Color Differences from Instrumentally Measured Color
Terminology E284, Practice E1164, Practice E1767, and Prac-
Coordinates
tice E2153 are applicable to this test method.
E179 Guide for Selection of Geometric Conditions for
3.2 Definitions of Terms Specific to This Standard:
Measurement of Reflection andTransmission Properties of
3.2.1 fluorescence, n—this standard uses the term “fluores-
Materials
cence” as a general term, including both true fluorescence
-8
E284 Termin
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.