ASTM E1349-06
(Test Method)Standard Test Method for Reflectance Factor and Color by Spectrophotometry Using Bidirectional (45:0 or 0:45) Geometry
Standard Test Method for Reflectance Factor and Color by Spectrophotometry Using Bidirectional (45:0 or 0:45) Geometry
SIGNIFICANCE AND USE
The most direct and accessible methods for obtaining the color coordinates of object colors are by instrumental measurement using spectrophotometers or spectrocolorimeters with either hemispherical or bidirectional optical measuring systems. This test method provides procedures for such measurement by spectrophotometry using a bidirectional (0:45 or 45:0) optical measuring system. The method for color and color difference measurement using filter colorimeters is contained in Test Method E 1347.
This test method is especially suitable for measurement of the following types of specimens for the indicated uses (see also Guide E 179 and Practice E 805):
5.2.1 Object-color specimens of any gloss level for color assessment.
5.2.2 Object-color specimens with physically flat, smooth surfaces from which to obtain data for use in computer colorant formulation.
5.2.3 Retroreflective specimens.
Note 1—To ensure inter-instrument agreement in the measurement of specimens with intermediate gloss, for formulation, or of retroreflective specimens, significantly tighter tolerances than those given in Practice E 1164, Influx and Efflux Conditions, 45°:Normal (45:0) and Normal:45° (0:45) Reflectance Factor section, may be required for the instrument angles of illumination and viewing. Information on the required tolerances is being developed.
This test method is not recommended for measurement of specimens with bare metal surfaces for color assessment, for which the use of hemispherical measurement geometry, as with an integrating-sphere type instrument, is preferable (see Guide E 179).
SCOPE
1.1 This test method covers the instrumental measurement of the reflection properties and color of object-color specimens by use of a spectrophotometer or spectrocolorimeter with a bidirectional optical measuring system, such as annular, circumferential, or uniplanar 45:0 or 0:45 geometry.
1.2 This test method is generally suitable for any non-fluorescent, flat object-color specimen. It is especially recommended for measuring retroreflective specimens and specimens of intermediate gloss.
1.3 Procedures required for the measurement of fluorescent object color are given in Practice E 991 and Practice E 2153.
1.4 Procedures required for the measurement of color using filter colorimeters are contained in Test Method E 1347 and this standard does not address those instruments.
. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E1349 −06
StandardTest Method for
Reflectance Factor and Color by Spectrophotometry Using
1
Bidirectional (45°:0° or 0°:45°) Geometry
This standard is issued under the fixed designation E1349; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope E694 Specification for Laboratory Glass Volumetric Appa-
ratus
1.1 This test method covers the instrumental measurement
E805 Practice for Identification of Instrumental Methods of
of the reflection properties and color of object-color specimens
Color or Color-Difference Measurement of Materials
by use of a spectrophotometer or spectrocolorimeter with a
E991 Practice for Color Measurement of Fluorescent Speci-
bidirectional optical measuring system, such as annular,
mens Using the One-Monochromator Method
circumferential, or uniplanar 45:0 or 0:45 geometry.
E1164 PracticeforObtainingSpectrometricDataforObject-
1.2 This test method is generally suitable for any non-
Color Evaluation
fluorescent, flat object-color specimen. It is especially recom-
E1345 Practice for Reducing the Effect of Variability of
mendedformeasuringretroreflectivespecimensandspecimens
Color Measurement by Use of Multiple Measurements
of intermediate gloss.
E1347 Test Method for Color and Color-Difference Mea-
surement by Tristimulus Colorimetry
1.3 Procedures required for the measurement of fluorescent
object color are given in Practice E991 and Practice E2153. E1767 Practice for Specifying the Geometries of Observa-
tion and Measurement to Characterize the Appearance of
1.4 Procedures required for the measurement of color using
Materials
filter colorimeters are contained inTest Method E1347 and this
E2153 Practice for Obtaining Bispectral Photometric Data
standard does not address those instruments.
for Evaluation of Fluorescent Color
1.5 . This standard does not purport to address all of the
E2214 Practice for Specifying and Verifying the Perfor-
safety concerns, if any, associated with its use. It is the
mance of Color-Measuring Instruments
responsibility of the user of this standard to establish appro-
3. Terminology
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use.
3.1 Definitions:
3.1.1 ThedefinitionscontainedinGuideE179,Terminology
2. Referenced Documents
E284, Practice E1164, Practice E1767, and Practice E2214 are
2
2.1 ASTM Standards:
applicable to this test method.
D2244 Practice for Calculation of Color Tolerances and
4. Summary of Test Method
Color Differences from Instrumentally Measured Color
Coordinates 4.1 Thistestmethodprovidesaprocedureformeasuringthe
E179 Guide for Selection of Geometric Conditions for
reflectance factors of reflecting object-color specimens by
Measurement of Reflection and Transmission Properties usingaspectrophotometerorspectrocolorimeterequippedwith
of Materials a bidirectional optical measuring system.
E284 Terminology of Appearance
4.2 When the specimens exhibit directionality, and an in-
E308 PracticeforComputingtheColorsofObjectsbyUsing
strument with uniplanar geometry is used, information on
the CIE System
directionality may be obtained by measuring the specimens at
more than one rotation angle, typically at two angles 90° apart
1
This test method is under the jurisdiction of ASTM Committee E12 on Color
as described in Practice E1345. When such information is not
and Appearance and is the direct responsibility of Subcommittee E12.02 on
required, these measurements may be averaged, or an instru-
Spectrophotometry and Colorimetry.
ment with annular or circumferential geometry may be used.
Current edition approved March 1, 2006. Published April 2006. Originally
However, even with annular or circumferential influx or efflux
approved in 1990. Last previous edition approved in 1998 as E1349 – 90 (1998).
DOI: 10.1520/E1349-06.
optics, some of the variability induced by specimen — optical
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
system interactions may remain and the application of the
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
methods in Practice E1345 may help to further reduce those
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. measurement errors.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
E1349−06
4.3 This test method includes procedures for standardizing 6.2.3 Verification Standards (recommended)—(See Practice
the instrument or verifying its sta
...