Standard Test Method for Color and Color-Difference Measurement by Tristimulus Colorimetry

SIGNIFICANCE AND USE
5.1 The most direct and accessible methods for obtaining the color differences and color coordinates of object colors are by instrumental measurement using colorimeters or spectrophotometers with either hemispherical or bidirectional optical measuring systems. This test method provides procedures for such measurement by use of a colorimeter with either a bidirectional or a hemispherical optical measuring system.  
5.2 This test method is suitable for measurement of color differences of nonmetameric, nonparameric pairs of object-color specimens, or color coordinates of most such specimens. A further limitation to the use of colorimeters having hemispherical geometry is the existence of a chromatic integrating-sphere error that prevents accurate measurement of color coordinates when the colorimeter is standardized by use of a white standard.4  
5.3 For the measurement of retroreflective specimens by this test method, the use of bidirectional geometry is recommended (see Guide E179 and Practice E805).
Note 3: To ensure inter-instrument agreement in the measurement of retroreflective specimens, significantly tighter tolerances than those given in Practice E1164 in the section on Influx and Efflux Conditions for 45°:Normal (45:0) and Normal:45° (0:45) Reflectance Factor are required for the instrument angles of illumination and viewing. Information on the required tolerances is being developed.  
5.4 A requirement for the use of a colorimeter to obtain accurate color coordinates is that the combination of source, filter, and detector characteristics to duplicate accurately the combined characteristics of a CIE standard illuminant and observer. When this requirement is not met, this test method requires the use of local standards for improving accuracy in the measurement of color coordinates (see also 4.2). For the measurement of small color differences between nonmetameric, nonparameric specimens, accuracy in absolute color coordinates is less important and standard...
SCOPE
1.1 This test method covers the instrumental measurement of specimens resulting in color coordinates and color difference values by using a tristimulus colorimeter, also known as a tristimulus filter colorimeter or a color-difference meter.  
1.2 Provision is made in this test method for the measurement of color coordinates and color differences by reflected or transmitted light using either a hemispherical optical measuring system, such as an integrating sphere, or a bidirectional optical measuring system, such as annular, circumferential, or uniplanar 45:0 and 0:45 geometry.  
1.3 Because of the limited absolute accuracy of tristimulus colorimeters, this test method specifies that, when color coordinates are required, the instrument be standardized by use of a standard having similar spectral (color) and geometric characteristics to those of the specimen. This standard is also known as a product standard. The use of a product standard of suitable stability is highly desirable.  
1.4 Because tristimulus colorimeters do not provide any information about the reflectance or transmittance curves of the specimens, they cannot be used to gain any information about metamerism or paramerism.  
1.5 Because of the inability of tristimulus (filter) colorimeters to detect metamerism or paramerism of specimens, this test method specifies that, when color differences are required, the two specimens must have similar spectral (color) and geometric characteristics. In this case, the instrument may be standardized for reflectance measurement by use of a white reflectance standard or, for transmittance measurement, with no specimen or standard at the specimen position.  
1.6 This test method is generally suitable for any non-fluorescent, planar, object-color specimens of all gloss levels. Users must determine whether an instrument complying with this method yields results that are useful to evaluate and characterize retroreflective ...

General Information

Status
Published
Publication Date
30-Apr-2020
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E1347-06(2020) - Standard Test Method for Color and Color-Difference Measurement by Tristimulus Colorimetry
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E1347 − 06 (Reapproved 2020)
Standard Test Method for
Color and Color-Difference Measurement by Tristimulus
Colorimetry
This standard is issued under the fixed designation E1347; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope this method yields results that are useful to evaluate and
characterize retroreflective specimens, or specimens having
1.1 This test method covers the instrumental measurement
optical structures.
ofspecimensresultingincolorcoordinatesandcolordifference
values by using a tristimulus colorimeter, also known as a 1.7 This test method does not apply to the use of a
spectrocolorimeter, which is a spectrometer that provides
tristimulus filter colorimeter or a color-difference meter.
colorimetric data, but not the underlying spectral data. Mea-
1.2 Provision is made in this test method for the measure-
surement by using a spectrocolorimeter is covered in Practice
ment of color coordinates and color differences by reflected or
E1164 and methods on color measurement by spectrophotom-
transmitted light using either a hemispherical optical measur-
etry.
ing system, such as an integrating sphere, or a bidirectional
1.8 This standard does not purport to address all of the
optical measuring system, such as annular, circumferential, or
safety concerns, if any, associated with its use. It is the
uniplanar 45:0 and 0:45 geometry.
responsibility of the user of this standard to establish appro-
1.3 Because of the limited absolute accuracy of tristimulus
priate safety, health, and environmental practices and deter-
colorimeters, this test method specifies that, when color coor-
mine the applicability of regulatory limitations prior to use.
dinates are required, the instrument be standardized by use of
1.9 This international standard was developed in accor-
a standard having similar spectral (color) and geometric
dance with internationally recognized principles on standard-
characteristics to those of the specimen. This standard is also
ization established in the Decision on Principles for the
known as a product standard. The use of a product standard of
Development of International Standards, Guides and Recom-
suitable stability is highly desirable.
mendations issued by the World Trade Organization Technical
1.4 Because tristimulus colorimeters do not provide any Barriers to Trade (TBT) Committee.
informationaboutthereflectanceortransmittancecurvesofthe
specimens, they cannot be used to gain any information about 2. Referenced Documents
metamerism or paramerism.
2.1 ASTM Standards:
1.5 Because of the inability of tristimulus (filter) colorim- D2244 Practice for Calculation of Color Tolerances and
eters to detect metamerism or paramerism of specimens, this Color Differences from Instrumentally Measured Color
test method specifies that, when color differences are required, Coordinates
the two specimens must have similar spectral (color) and E179 Guide for Selection of Geometric Conditions for
geometric characteristics. In this case, the instrument may be
Measurement of Reflection and Transmission Properties
standardized for reflectance measurement by use of a white of Materials
reflectance standard or, for transmittance measurement, with
E284 Terminology of Appearance
no specimen or standard at the specimen position. E805 Practice for Identification of Instrumental Methods of
Color or Color-Difference Measurement of Materials
1.6 This test method is generally suitable for any non-
E1164 PracticeforObtainingSpectrometricDataforObject-
fluorescent, planar, object-color specimens of all gloss levels.
Color Evaluation
Users must determine whether an instrument complying with
E1345 Practice for Reducing the Effect of Variability of
Color Measurement by Use of Multiple Measurements
This test method is under the jurisdiction of ASTM Committee E12 on Color
and Appearance and is the direct responsibility of Subcommittee E12.02 on
Spectrophotometry and Colorimetry. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved May 1, 2020. Published May 2020. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1990. Last previous edition approved in 2015 as E1347 – 06 (2015). Standards volume information, refer to the standard’s Document Summary page on
DOI: 10.1520/E1347-06R20. the ASTM website.
Copyright ©ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA19428-2959. United States
E1347 − 06 (2020)
3. Terminology 4.2 This test method includes two different procedures for
standardizing the colorimeter. The first procedure utilizes a
3.1 Definitions:
white standard of known reflectance factor; the second proce-
3.1.1 The definitions contained in Guide E179 and Termi-
dure utilizes a local standard.
nology E284 are applicable to this test method.
4.2.1 When absolute values of color coordinates are to be
determined, the use of a white standard is recommended only
4. Summary of Test Method
with colorimeters in which there is good conformance of the
4.1 This test method provides procedures for measuring
colorimeter readings to CIE tristimulus values, as determined
object-color specimens in either transmission or reflection with
by measurement of suitable verification standards (see Practice
atristimuluscolorimeter(hereafterreferredtoasacolorimeter)
E1164).Withinstrumentsnotmeetingthisrequirement,theuse
by use of the following geometric conditions and standardiza-
of local standards is recommended, but only when the signal
tion procedures:
level (see Note 2) from the use of each colorimeter filter is
4.1.1 Color differences by reflected light of nonmetameric,
adequately high.
nonparameric pairs of opaque or translucent specimens by use
NOTE 1—Of necessity, the above requirements are in part subjective, as
ofeitherhemisphericalgeometry,withanintegratingsphere,or
the methods for verifying conformance to the requirements may not be
bidirectional geometry, such as annular, circumferential, or
available to the average user. Each user must decide whether the
uniplanar 45:0 or 0:45 geometry. The colorimeter may be
standardization procedure selected results in a loss of accuracy in the
standardized by use of a white reflectance standard.
measurements that is negligibly small for the purpose for which data are
obtained.
4.1.2 Color differences by transmitted light of
NOTE 2—The adequacy of the signal level can be determined by
nonmetameric, nonparameric pairs of transparent or translu-
measuring the short-term repeatability without replacement, and ascer-
cent specimens by use of hemispherical geometry. The colo-
taining that the variation in the answer represents less than 30 % of the
rimeter may be standardized by use of a white standard at the
desired or allowable variation.
reflection port of the integrating sphere with no specimen in
4.2.2 When color differences are to be measured, only
place. When translucent specimens are measured, they should
relative measured values are required for the two members of
be placed flush against the transmission port of the sphere, and
the color-difference pair, and standardization by use of either a
the white standard should, for maximum accuracy, have the
white standard or a local standard is satisfactory. In those cases
samereflectanceandchemicalcompositionasthatofthelining
where a computer program is being used to predict color
of the integrating sphere.
tolerances, accuracy of the absolute values of the product
4.1.3 Color coordinates by reflected light of opaque or
standard color coordinates may become more important (see
translucent specimens by use of either bidirectional or hemi-
4.2.1).
spherical geometry. The colorimeter may be standardized by
4.3 Procedures for selecting specimens suitable for preci-
use of a standard having spectral (color) and geometric
sion measurement are included in this test method.
characteristics similar to those of the specimens. Such
standards, often called hitching-post standards, are hereafter
4.4 Most modern colorimeters compute the color coordi-
referred to as local standards.
natesofthespecimenduringthemeasurement.Whenthisisthe
4.1.4 Color coordinates by transmitted light of transparent
case, the user of this test method must designate the color
or translucent specimens by use of hemispherical geometry.
system to be used in the computation (see Practice D2244).
The colorimeter may be standardized by use of a local
standard.
5. Significance and Use
4.1.5 This test method is not appropriate for fluorescent
5.1 The most direct and accessible methods for obtaining
specimens.
the color differences and color coordinates of object colors are
4.1.6 For the measurement of the daytime color of retrore-
by instrumental measurement using colorimeters or spectro-
flective specimens, the 45:0 or 0:45 conditions are normally
photometers with either hemispherical or bidirectional optical
required. Some modern, high brightness, retroreflective sheet-
measuring systems. This test method provides procedures for
ing has been shown to exhibit geometric artifacts if the cone
such measurement by use of a colorimeter with either a
angles are too narrow. In these cases, it may be more
bidirectional or a hemispherical optical measuring system.
appropriate to use larger cone angles, with appropriate toler-
ances. 5.2 This test method is suitable for measurement of color
differences of nonmetameric, nonparameric pairs of object-
4.1.7 When the specimens exhibit directionality, and a
colorimeter with uniplanar bidirectional geometry is used, color specimens, or color coordinates of most such specimens.
A further limitation to the use of colorimeters having hemi-
information on directionality may be obtained by measuring
the specimens at more than one rotation angle, typically at two spherical geometry is the existence of a chromatic integrating-
sphere error that prevents accurate measurement of color
angles 90° apart. When such information is not required, these
measurements may be averaged, or a colorimeter with annular coordinates when the colorimeter is standardized by use of a
white standard.
or circumferential bidirectional geometry may be used.
3 4
Hunter, R. S., “Photoelectric Tristimulus Colorimetry with Three Filters,” Hoffman, K.,“Chromatic Integrating-Sphere Error in Tristimulus
Journal, Optical Society of America, Vol 32, 1942, pp. 509–558. Colorimeters,” Journal of Color and Appearance, Vol 1, No. 2, 1971, pp. 16–21.
E1347 − 06 (2020)
5.3 For the measurement of retroreflective specimens by 8. Standardization and Verification
this test method, the use of bidirectional geometry is recom-
8.1 Standardization for the Measurement Color Differences
mended (see Guide E179 and Practice E805).
of Specimen Pairs:
NOTE 3—To ensure inter-instrument agreement in the measurement of 8.1.1 Standardize the colorimeter by use of the white
retroreflective specimens, significantly tighter tolerances than those given
standard (mandatory) and the zero-reading standard (if
in Practice E1164 in the section on Influx and Efflux Conditions for
required), following the manufacturer’s instructions.
45°:Normal (45:0) and Normal:45° (0:45) Reflectance Factor are required
8.1.2 Verify the accuracy of the standardization and the
for the instrument angles of illumination and viewing. Information on the
required tolerances is being developed.
instrument performance by measuring a series of verification
standards (recommended).
5.4 A requirement for the use of a colorimeter to obtain
accurate color coordinates is that the combination of source,
8.2 Standardization for Measurement of Color Coordinates:
filter, and detector characteristics to duplicate accurately the
NOTE 4—If the verification tests of 8.2.2 are not to be carried out, omit
combined characteristics of a CIE standard illuminant and
8.2.1 and 8.2.2 and proceed to 8.2.3.
observer. When this requirement is not met, this test method
8.2.1 Standardize the colorimeter by use of the white
requires the use of local standards for improving accuracy in
standard (mandatory) and the zero-reading standard (if
the measurement of color coordinates (see also 4.2). For the
required), following the manufacturer’s instructions.
measurement of small color differences between
8.2.2 Verify the accuracy of the standardization and the
nonmetameric, nonparameric specimens, accuracy in absolute
color coordinates is less important and standardization of the instrument performance by measuring a series of verification
standards (recommended).
colorimeterbyuseofawhitestandardissatisfactory.However,
accurate color-difference measurement requires that specimen
8.2.3 Standardize the colorimeter by use of the appropriate
pairs have similar spectral and geometric characteristics.
local standard for the specimens to be measured (mandatory)
and the zero-reading standard (if required), following the
6. Apparatus
manufacturer’s instructions.
6.1 Colorimeter, designed for the measurement of object-
9. Procedure
color specimens. Use hemispherical geometry for reflection or
transmission measurements or bidirectional geometry for re-
9.1 When required, select the color scales to be used in the
flection measurements.
computation of color coordinates or color differences.
6.2 Standardization Plaques, supplied by the manufacturer.
9.2 Handle the specimen carefully; avoid touching the area
6.2.1 White Reflecting Tile or Standard (Mandatory)—(If to be measured. When necessary, clean the specimen by an
the colorimeter has hemispherical geometry, a standard of
agreed procedure.
hemispherical reflectance factor is required; if bidirectional
9.3 Whenhemisphericalgeometryisused,makethefollow-
geometry, a standard of bidirectional reflectance factor is
ing selections:
required.)
9.3.1 For the measurement of reflecting specimens, select
6.2.2 Local Standardization Plaques (Recommended), hav-
inclusion or exclusion of the specular component of reflection
ing spectral (color) and geometric characteristics similar to
as desired (see Guide E179 and Practice E805).
those of specimens to be measured, as required for the
9.3.1.1 Ifthespecimenistranslucent,backitwithastandard
measurement of color coordinates.
backing material during the measurement.
6.2.3 Light Trap (Hemispherical) or Polished Black Glass
9.3.2 For the measure
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.