SIGNIFICANCE AND USE
5.1 This test method can be used to evaluate unused mixed bed ion exchange cartridges for conformance to specifications.
5.2 This test method provides for the calculation of capacity in terms of the volume of water treated to a conductivity end point.
5.3 The test method as written assumes that the ion exchange resins in the cartridge are either partially or fully converted to the H+ or OH– form. Regeneration of the resins is not part of this method.
5.4 This test method provides for the calculation of capacity on a cartridge basis.
5.5 This test method may be used to test different size mixed bed resin cartridges. The flow rate of test water and the frequency of sampling are varied to compensate for the approximate volume of resin in the test cartridge.
SCOPE
1.1 This test method covers the determination of the performance of mixed bed ion exchange resin cartridges in the active form when used for deionization. The test can be used to determine the initial capacity of unused cartridges or the remaining capacity of used cartridges. In this case performance is defined as ion exchange capacity (or throughput) to two defined endpoints. The method does not measure organics and does not attempt to determine the ultimate water quality attainable by the cartridge.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Relations
Buy Standard
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D7513 − 17
Standard Test Method for
1
Capacity of Mixed Bed Ion Exchange Cartridges
This standard is issued under the fixed designation D7513; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope D5391 Test Method for Electrical Conductivity and Resis-
tivity of a Flowing High Purity Water Sample
1.1 This test method covers the determination of the perfor-
mance of mixed bed ion exchange resin cartridges in the active
3. Terminology
form when used for deionization. The test can be used to
3.1 Definitions:
determine the initial capacity of unused cartridges or the
3.1.1 For definitions of terms used in this standard, refer to
remainingcapacityof used cartridges. In this case performance
Terminology D1129.
is defined as ion exchange capacity (or throughput) to two
defined endpoints. The method does not measure organics and
4. Summary of Test Method
does not attempt to determine the ultimate water quality
4.1 This test method consists of exhausting a cartridge of
attainable by the cartridge.
mixedbedionexchangematerialtoaspecificendpointwithan
1.2 The values stated in SI units are to be regarded as
influent solution of known composition and volume.
standard. No other units of measurement are included in this
5. Significance and Use
standard.
1.3 This standard does not purport to address all of the 5.1 This test method can be used to evaluate unused mixed
safety concerns, if any, associated with its use. It is the bed ion exchange cartridges for conformance to specifications.
responsibility of the user of this standard to establish appro-
5.2 This test method provides for the calculation of capacity
priate safety and health practices and determine the applica-
in terms of the volume of water treated to a conductivity end
bility of regulatory limitations prior to use
point.
1.4 This international standard was developed in accor-
5.3 The test method as written assumes that the ion ex-
dance with internationally recognized principles on standard-
change resins in the cartridge are either partially or fully
ization established in the Decision on Principles for the
+ –
converted to the H or OH form. Regeneration of the resins is
Development of International Standards, Guides and Recom-
not part of this method.
mendations issued by the World Trade Organization Technical
Barriers to Trade (TBT) Committee. 5.4 This test method provides for the calculation of capacity
on a cartridge basis.
2. Referenced Documents
5.5 Thistestmethodmaybeusedtotestdifferentsizemixed
2
2.1 ASTM Standards:
bed resin cartridges. The flow rate of test water and the
D1125 Test Methods for Electrical Conductivity and Resis-
frequency of sampling are varied to compensate for the
tivity of Water
approximate volume of resin in the test cartridge.
D1129 Terminology Relating to Water
6. Apparatus
D1193 Specification for Reagent Water
D3375 Test Method for Column Capacity of Particulate
6.1 Test Assembly (Fig. 1), consisting of the following:
Mixed Bed IonExchange Materials
6.1.1 Cartridge, and pressure vessel, if required. Connec-
tions shall be provided at the top and the bottom for the
admission and removal of the ion exchange test water as
1
This test method is under the jurisdiction of ASTM Committee D19 on Water
described in 7.4.2.
and is the direct responsibility of Subcommittee D19.08 on Membranes and Ion
6.1.2 Test solution tank. Adequate means of regulating and
Exchange Materials.
measuring flow through the cartridge shall be provided. If the
Current edition approved June 1, 2017. Published June 2017. Originally
test flow rate (8.1) can not be obtained by gravity feed, then a
approved in 2009. Last previous edition approved in 2009 as D7513 – 09. DOI:
10.1520/D7513-17.
pump should be used.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
6.1.3 Measuring circuit and in-line conductivity cells shall
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
be as described in Test Methods D1125 or D5391. A continu-
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. ous recorder is recommended.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
D7513 − 17
FIG. 1 Typical Apparatus for Performance Testing of Mixed-Bed Ion Exchange Cartridges
7. Reagents and Materials ye
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D7513 − 09 D7513 − 17
Standard Test Method for
1
Capacity of Mixed Bed Ion Exchange Cartridges
This standard is issued under the fixed designation D7513; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method covers the determination of the performance of mixed bed ion exchange resin cartridges in the active form
when used for deionization. The test can be used to determine the initial capacity of unused cartridges or the remaining capacity
of used cartridges. In this case performance is defined as ion exchange capacity (or throughput) to two defined endpoints. The
method does not measure organics and does not attempt to determine the ultimate water quality attainable by the cartridge.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use
1.4 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2
2.1 ASTM Standards:
D1125 Test Methods for Electrical Conductivity and Resistivity of Water
D1129 Terminology Relating to Water
D1193 Specification for Reagent Water
D2777 Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water
D3375 Test Method for Column Capacity of Particulate Mixed Bed IonExchange Materials
D5391 Test Method for Electrical Conductivity and Resistivity of a Flowing High Purity Water Sample
3. Terminology
3.1 Definitions—Definitions: For definitions of terms related to water, refer to Terminology D1129.
3.1.1 For definitions of terms used in this standard, refer to Terminology D1129.
4. Summary of Test Method
4.1 This test method consists of exhausting a cartridge of mixed bed ion exchange material to a specific end point with an
influent solution of known composition and volume.
5. Significance and Use
5.1 This test method can be used to evaluate unused mixed bed ion exchange cartridges for conformance to specifications.
5.2 This test method provides for the calculation of capacity in terms of the volume of water treated to a conductivity end point.
5.3 The test method as written assumes that the ion exchange resins in the cartridge are either partially or fully converted to
+ -–
the H or OH form. Regeneration of the resins is not part of this method.
5.4 This test method provides for the calculation of capacity on a cartridge basis.
1
This test method is under the jurisdiction of ASTM Committee D19 on Water and is the direct responsibility of Subcommittee D19.08 on Membranes and Ion Exchange
Materials.
Current edition approved March 15, 2009June 1, 2017. Published April 2009June 2017. Originally approved in 2009. Last previous edition approved in 2009 as D7513
– 09. DOI: 10.1520/D7513-09.10.1520/D7513-17.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
D7513 − 17
5.5 This test method may be used to test different size mixed bed resin cartridges. The flow rate of test water and the frequency
of sampling are varied to compensate for the approximate volume of resin in the test cartridge.
6. Apparatus
6.1 Test Assembly (Fig. 1) , ), consisting of the following:
6.1.1 Cartridge, and pressure vessel, if required. Connections shall be provided at the top and the bottom for the admission and
removal of the ion exchange test water as described in 7.37.4.2.
6.1.2 Test solution tank. Adequate means of regulating and measuring flow through the cartridge
...
Questions, Comments and Discussion