ASTM D7066-04(2017)
(Test Method)Standard Test Method for dimer/trimer of chlorotrifluoroethylene (S-316) Recoverable Oil and Grease and Nonpolar Material by Infrared Determination
Standard Test Method for dimer/trimer of chlorotrifluoroethylene (S-316) Recoverable Oil and Grease and Nonpolar Material by Infrared Determination
SIGNIFICANCE AND USE
5.1 The presence and concentration of oil and grease in domestic and industrial wastewater is of concern to the public because of its deleterious aesthetic effect and its impact on aquatic life.
5.2 Regulations and standards have been established that require monitoring of oil and grease in water and wastewater.
SCOPE
1.1 This test method covers the determination of oil and grease and nonpolar material in water and wastewater by an infrared (IR) determination of dimer/trimer of chlorotrifluoroethylene (S-316)2 extractable substances from an acidified sample. Included in this estimation of oil and grease are any other compounds soluble in the solvent.
1.2 This test method is applicable to measurement of the light fuel although loss of some light ends during extraction can be expected.
1.3 This test method defines oil and grease in water and wastewater as that which is extractable in the test method and measured by IR absorption at 2930 cm-1 or 3.4 microns. Similarly, this test method defines nonpolar material in water and wastewater as that oil and grease which is not adsorbed by silica gel in the test method and measured by IR absorption at 2930 cm-1.
1.4 This test method covers the range of 5 to 100 mg/L and may be extended to a lower or higher level by extraction of a larger or smaller sample volume collected separately.
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine (Guide D3856) the applicability of regulatory limitations prior to use.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Relations
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D7066 −04 (Reapproved 2017)
Standard Test Method for
dimer/trimer of chlorotrifluoroethylene (S-316) Recoverable
Oil and Grease and Nonpolar Material by Infrared
Determination
This standard is issued under the fixed designation D7066; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.7 This international standard was developed in accor-
dance with internationally recognized principles on standard-
1.1 This test method covers the determination of oil and
ization established in the Decision on Principles for the
grease and nonpolar material in water and wastewater by an
Development of International Standards, Guides and Recom-
infrared (IR) determination of dimer/trimer of chlorotrifluoro-
2 mendations issued by the World Trade Organization Technical
ethylene (S-316) extractable substances from an acidified
Barriers to Trade (TBT) Committee.
sample. Included in this estimation of oil and grease are any
other compounds soluble in the solvent.
2. Referenced Documents
1.2 This test method is applicable to measurement of the
2.1 ASTM Standards:
light fuel although loss of some light ends during extraction
D1129 Terminology Relating to Water
can be expected.
D1193 Specification for Reagent Water
1.3 This test method defines oil and grease in water and
D3370 Practices for Sampling Water from Closed Conduits
wastewater as that which is extractable in the test method and
D3856 Guide for Management Systems in Laboratories
-1
measured by IR absorption at 2930 cm or 3.4 microns.
Engaged in Analysis of Water
Similarly, this test method defines nonpolar material in water
D2777 Practice for Determination of Precision and Bias of
and wastewater as that oil and grease which is not adsorbed by
Applicable Test Methods of Committee D19 on Water
silica gel in the test method and measured by IR absorption at
D5847 Practice for Writing Quality Control Specifications
-1
2930 cm .
for Standard Test Methods for Water Analysis
1.4 This test method covers the range of 5 to 100 mg/L and E168 Practices for General Techniques of Infrared Quanti-
may be extended to a lower or higher level by extraction of a
tative Analysis
larger or smaller sample volume collected separately. E178 Practice for Dealing With Outlying Observations
1.5 The values stated in SI units are to be regarded as
3. Terminology
standard. No other units of measurement are included in this
standard.
3.1 Definitions:
3.1.1 For definitions of terms used in this standard, refer to
1.6 This standard does not purport to address all of the
Terminology D1129 and Practices E168.
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
3.2 Definitions of Terms Specific to This Standard:
priate safety, health, and environmental practices and deter-
3.2.1 nonpolar material, n—the oil and grease remaining in
mine (Guide D3856) the applicability of regulatory limitations
solution after contact with silica gel and measured by this test
prior to use.
method.
3.2.2 oil and grease, n—the organic matter extracted from
1 water or wastewater and measured by this test method.
This test method is under the jurisdiction of ASTM Committee D19 on Water
andisthedirectresponsibilityofSubcommitteeD19.06onMethodsforAnalysisfor
3.2.3 solvent, n—dimer/trimerofchlorotrifluoroethylene(S-
Organic Substances in Water.
316).
Current edition approved Sept. 1, 2017. Published September 2017. Originally
approved in 2004. Last previous edition approved in 2011 as D7066 – 04 (2011).
DOI: 10.1520/D7066-04R17.
The sole source of supply of the material S-316 known to the committee at this
time is Horiba Instruments, Irvine, CA. If you are aware of alternative suppliers, For referenced ASTM standards, visit the ASTM website, www.astm.org, or
please provide this information to ASTM International Headquarters. Your com- contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
ments will receive careful consideration at a meeting of the responsible technical Standards volume information, refer to the standard’s Document Summary page on
committee, which you may attend. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7066 − 04 (2017)
4. Summary of Test Method 7.10 Volumetric flasks, glass, various (10, 25, 50, 100, and
200-mL).
4.1 An acidified 250-mL sample of water or wastewater is
extracted serially with three 15-mLvolumes of dimer/trimer of 7.11 TFE-fluorocarbon spritz bottle, one-piece wash bottle
chlorotrifluoroethylene(S-316).Theextractisdilutedto50mL for rinsing.
and a portion is examined by infrared spectroscopy (IR) for an
7.12 Repeating pipetter, glass, 15-mL, (optional).
oil and grease measurement. A portion of the extract is
7.13 Volumetric pipettes, glass, various (0.50, 1.00, 5.00,
contacted with silica gel to remove polar substances, thereby
10.0 and 25.0-mL, including a 1.00 serological pipet graduated
producing a solution containing nonpolar material. The non-
in 0.01-mL increments and a 5.00-mL serological pipet gradu-
polar material is measured by infrared spectroscopy.
ated in 0.1-mL increments, or equivalent).
5. Significance and Use
7.14 Benchtop shaker, (optional).
5.1 The presence and concentration of oil and grease in
7.15 Glass stirring rod, (optional).
domestic and industrial wastewater is of concern to the public
7.16 Analytical balance.
because of its deleterious aesthetic effect and its impact on
7.17 Syringes, 50 and 500 mL.
aquatic life.
5.2 Regulations and standards have been established that
8. Reagents
require monitoring of oil and grease in water and wastewater.
8.1 Purity of Reagents—Reagent grade chemicals shall be
used in all tests. Unless otherwise indicated, it is intended that
6. Interferences
allreagentsshallconformtothespecificationoftheCommittee
6.1 Soaps, detergents, surfactants, and other materials may
on Analytical Reagents of the American Chemical Society,
form emulsions that may reduce the amount of oil and grease
where such specifications are available. Other grades may be
extracted from a sample. This test method contains procedures
used, provided it is first ascertained that the reagent is of
that can assist the analyst in breaking such emulsions.
sufficiently high purity to permit its use without lessening the
6.2 Organic compounds and other materials not considered accuracy of the determination.
as oil and grease on the basis of chemical structure may be
8.2 Purity of Water—Unless otherwise indicated, references
extracted and measured as oil and grease. Of those measured,
to laboratory or reagent water shall be understood to mean
certain ones may be adsorbed by silica gel while others may
reagent water conforming to Specification D1193, Type II.
not. Those not adsorbed are measured as nonpolar material.
8.3 Isooctane (2,2,4-trimethylpentane) 98 % minimum
purity, for use in calibration.
7. Apparatus
8.4 Octanoic Acid 98 % minimum purity, for use in calibra-
7.1 All glassware that will come in contact with the sample
tion.
must be rinsed with dimer/trimer of chlorotrifluoroethylene
(S-316) prior to beginning this procedure.
8.5 Silica Gel, Anhydrous, 75–150 micrometres, Davisil
Grade 923 (Supelco 21447-7A, or equivalent). Dry at
7.2 Cell(s), quartz, 10-mm path length (lower concentra-
200–250°C for 24 hour minimum and store in a desiccator or
tions may require a longer pathlength), two required for
tightly sealed container. Determine the dimer/trimer of chloro-
double-beam operation, one required for single-beam
trifluoroethylene (S-316) soluble material content of the silica
operation, or built-in or drop-in cell for infrared filtometer
gel by extracting 10 g of silica gel with 25 mLof dimer/trimer
analyzer operation.
of chlorotrifluoroethylene (S-316) and collect the elute in a
7.3 Filter paper, ashless, quantitative, general-purpose, 11-
flask. Filter and fill a quartz cell for analysis by IR. The
cm, Whatman #40 or equivalent.
dimer/trimer of chlorotrifluoroethylene (S-316) soluble mate-
7.4 Glass funnel.
rial must be less than 5 mg/L.
7.5 Glass wide mouth sample bottle, minimum 250-mL,
8.6 Sodium Sulfate (Na SO ), ACS, granular anhydrous.
2 4
with screw cap having a fluoropolymer liner.
Dry at 200–250°C for 24 hours minimum and store in a tightly
sealed container until use.
7.6 Glass graduated cylinder, 100-mL
NOTE 1—Powdered sodium sulfate should not be used because water
7.7 Infrared spectrometer, double-beam dispersive, single-
may cause it to solidify.
beam dispersive, Fourier transform, filtometers or other ca-
-1
pable of making measurements at 2930 cm . 8.7 Solvent—dimer/trimer of chlorotrifluoroethylene,IR
spectroscopy grade.
7.8 Magnetic stirrer, with small TFE-fluorocarbon stirring
bar. 8.8 Sulfuric Acid (1 + 1)—Slowly and carefully add 1
volume of sulfuric acid (H SO , sp gr 1.84) to 1 volume of
2 4
7.9 Glass separatory-funnel, 500mL, with fluoropolymer
water, stirring and cooling the solution during the addition
stopcock and stopper.
(optional HCl replacement).
4 5
Consult the manufacturer’s operation manual for the specific instructions The material S-316, available from Horiba Instruments, Irvine, CA, or
related to the infrared spectrometer or analyzer to be used. equivalent, has been found suitable for use.
D7066 − 04 (2017)
8.9 Hydrochloric acid,ACS,1+1.Mix equal volumes of 10. Preparation of Calibration and Spiking Solutions
NOTE 2—The calibration standard specified in this procedure reflects
concentrated HCl and water
the objective of the test to detect recoverable oil and grease and nonpolar
8.10 Sodium Chloride (NaCl), crystalline, ACS, or use in
material in wastewater with an unknown composition of oil and grease. In
breaking emulsions, if needed. Wet thoroughly with solvent
a few cases, the composition of the oil and grease in a sample will be
known.However,inordertoobtainconsistentresultsbetweensamplesets
before using.
and between laboratories with different wastewater matrices, calibration
with the known oil and grease in a sample should not be used in this test
9. Sampling
method.
9.1 Collect the sample in accordance with the principles
10.1 Calibration and Solvent Mixtures:
described in Practices D3370, using a glass bottle equipped
NOTE 3—The calibration procedure below calls for transferring, by
with a screw cap having a fluoropolymer liner. Prerinse the
pipette or syringe, a volume of standard into a volumetric flask to obtain
sample bottle and cap with the solvent prior to sample
a desired concentration. Transfer volumes have been rounded for ease of
collection. Do not rinse the sample bottle with the sample to be
measurement and calculation. It is highly recommended that calibration
analyzed. Fill bottle with minimal headspace to prevent loss of
standards be prepared on a weight basis (that is, pipette a volume into a
tared flask and weigh the amount pipetted), then converted to mg/mL by
volatile constituants. Do not allow the sample to overflow the
using the densities of octanoic acid (0.9100 g/mL) and isooctane (0.6920
bottle during collection. Preventing overflow may not be
g/mL). A solution containing equal volumes of isooctane and octanoic
possible in all sampling situations, however, measures should
acid will have a density of 0.801 g/mL.To assure the most accurate
be taken to minimize overflow at all times.
concentrations, use the smallest serological pipet or syringe for
9.2 A sample of about 250 mL is required for this test. Use
measurements. The volume should always be greater than ⁄2
the entire sample because removing a portion would not
the volume of the pipet or syringe.
apportion the oil and grease that adheres to the bottle surfaces.
Ideally,alinearcalibrationcurvewillbeobtainedfromthese
The high probability that extractable matter may adhere to
standards. As discussed in Section 11, the concentrations of
samplingequipmentandresultinmeasurementsthatarebiased
these standards can be adjusted to stay within the linear range
low precludes the collection of composite samples for deter-
of the IR instrument.
mination of oil and grease. Therefore, samples must be
10.1.1 Calibration Stock Solution—Place 0.55 mL of oc-
collected as grab samples. If a composite measurement is
tanoic acid and 0.72 mL of isooctane in a 10-mL volumetric
required, individual grab samples collected at prescribed time
flask and fill to the mark with solvent. Mix well. The resulting
intervals may be analyzed separately and the concentrations
concentration is 50 mg/mL each octanoic acid and isooctane
averaged. Alternatively, samples can be collected in the field
(100 mg/mLtotal oil and grease). This solution will be termed
and composited in the laboratory. For example, collect four
“Stock Solution”.
individual 63-mL samples over the course of a day. In the
10.1.2 Diluted Stock Solution—Place 2.5 mL of the Stock
laboratory, pour each 63-mLsample into the separatory funnel,
Solution to a 50-mL volumetric flask and fill to mark with
rinse each of the four bottles (and caps) sequentially with 10
solvent. Diluted Stock Solution = 5.0 mg/mL (5000 µg/mL).
mL of solvent, and use the solvent for the extraction (12.2.2).
10.1.3 Calibration Solution A—Place 1.0 mL of Diluted
Do not exceed 50 mLof total solvent during the extraction and
Stock Solution in a 10-mLvolumetric flask and fill to the mark
rinse procedure.
withsolvent.CalibrationSolutionA=0.5mg/mL(500µg/mL),
9.3 Preserve the sample with a sufficient quantity of either
equivalent to 100 mg/L oil and grease in a 250-mL water
sulfuric (see 8.8) or hydrochloric acid (see 8.9)toapHof2or
sample extracted into a 50-mL volume of solvent.
lower and refrigerate at 0–4°C from the time of collection until
10.1.4 Calibration Solution B—Place 0.50 mL of Diluted
extraction. The amount of acid required will be dependent
Stock Solution in a 10-mLvolumetric flask and fill to the mark
upon the pH and buffer capacity of the sample at the time of
with solvent. Calibration Solution B = 0.25 mg/mL (250
collection. If the amount of acid required is not known, make
µg/mL), equivalent to 50 mg/L oil and grease in a 250-mL
the pH measurement on a separate sample that will not be
water sample extracted into a 50-mL volume of solvent.
analyzed. Introduction of pH paper to an actual sample or
10.1.5 Ca
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.