Standard Guide for Examination of Hardened Concrete Using Scanning Electron Microscopy

SCOPE
1.1 This guide provides information for the examination of hardened concrete using scanning electron microscopy (SEM) combined with energy-dispersive X-ray spectroscopy (EDX or EDS). Since the 1960s, SEM has been used for the examination of concrete and has proved to be an insightful tool for the microstructural analysis of concrete and its components. There are no standardized procedures for the SEM analysis of concrete. SEM supplements techniques of light microscopy, which are described in Practice C856/C856M, and, when applicable, techniques described in Practice C856/C856M should be consulted for SEM analysis. For further study, see the bibliography at the end of this guide.  
1.2 This guide is intended to provide a general introduction to the application of SEM/EDS analytical techniques for the examination and analysis of concrete. It is meant to be useful to engineers and scientists who want to study concrete and who are familiar with, but not expert in, the operation and application of SEM/EDS technology. The guide is not intended to provide explicit instructions concerning the operation of this technology or interpretation of information obtained through SEM/EDS.  
1.3 It is critical that petrographer or operator or both be familiar with the SEM/EDX (EDS) equipment, specimen preparation procedures, and the use of other appropriate procedures for this purpose. This guide does not discuss data interpretation. Proper data interpretation is best done by individuals knowledgeable about the significance and limitations of SEM/EDX (EDS) and the materials being evaluated.  
1.4 The SEM provides images that can range in scale from a low magnification (for example, 15×) to a high magnification (for example, 50 000× or greater) of concrete specimens such as fragments, polished surfaces, or powders. These images can provide information indicating compositional or topographical variations in the observed specimen. The EDX (EDS) system can be used to qualitatively or quantitatively determine the elemental composition of very small volumes intersecting the surface of the observed specimen (for example, 1-10 cubic microns) and those measured compositional determinations can be correlated with specific features observed in the SEM image. See Note 1.
Note 1: An electronic document consisting of electron micrographs and EDX (EDS) spectra illustrating the materials, reaction products, and phenomena discussed below is available at http://netfiles.uiuc.edu/dlange/www/CML/index.html.  
1.5 Performance of SEM and EDX (EDS) analyses on hardened concrete specimens can, in some cases, present unique challenges not normally encountered with other materials analyzed using the same techniques.  
1.6 This guide can be used to assist a concrete petrographer in performing or interpreting SEM and EDX (EDS) analyses in a manner that maximizes the usefulness of these techniques in conducting petrographic examinations of concrete and other cementitious materials, such as mortar and stucco. For a more in-depth, comprehensive tutorial on scanning electron microscopy or the petrographic examination of concrete and concrete-related materials, the reader is directed to the additional publications referenced in the bibliography section of this guide.  
1.7 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.8 This standard does not purport to address all of the safety concerns, if any, associated with the use of electron microscopes, X-ray spectrometers, chemicals, and equipment used to prepare samples for electron microscopy. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.9 This international standard was developed in accordance with internationally recognized principles on standardization established in ...

General Information

Status
Published
Publication Date
30-Sep-2022
Drafting Committee
Current Stage
Ref Project

Relations

Buy Standard

Guide
ASTM C1723-16(2022) - Standard Guide for Examination of Hardened Concrete Using Scanning Electron Microscopy
English language
9 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: C1723 − 16 (Reapproved 2022)
Standard Guide for
Examination of Hardened Concrete Using Scanning Electron
Microscopy
This standard is issued under the fixed designation C1723; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope can be used to qualitatively or quantitatively determine the
elemental composition of very small volumes intersecting the
1.1 This guide provides information for the examination of
surface of the observed specimen (for example, 1-10 cubic
hardened concrete using scanning electron microscopy (SEM)
microns) and those measured compositional determinations
combined with energy-dispersive X-ray spectroscopy (EDX or
can be correlated with specific features observed in the SEM
EDS).Sincethe1960s,SEMhasbeenusedfortheexamination
image. See Note 1.
of concrete and has proved to be an insightful tool for the
NOTE 1—An electronic document consisting of electron micrographs
microstructural analysis of concrete and its components.There
and EDX (EDS) spectra illustrating the materials, reaction products, and
are no standardized procedures for the SEM analysis of
phenomena discussed below is available at http://netfiles.uiuc.edu/dlange/
concrete. SEM supplements techniques of light microscopy,
www/CML/index.html.
which are described in Practice C856/C856M, and, when
1.5 Performance of SEM and EDX (EDS) analyses on
applicable, techniques described in Practice C856/C856M
hardened concrete specimens can, in some cases, present
should be consulted for SEM analysis. For further study, see
unique challenges not normally encountered with other mate-
the bibliography at the end of this guide.
rials analyzed using the same techniques.
1.2 This guide is intended to provide a general introduction
to the application of SEM/EDS analytical techniques for the
1.6 This guide can be used to assist a concrete petrographer
examination and analysis of concrete. It is meant to be useful
inperformingorinterpretingSEMandEDX(EDS)analysesin
toengineersandscientistswhowanttostudyconcreteandwho
a manner that maximizes the usefulness of these techniques in
are familiar with, but not expert in, the operation and applica-
conducting petrographic examinations of concrete and other
tion of SEM/EDS technology. The guide is not intended to
cementitious materials, such as mortar and stucco. For a more
provide explicit instructions concerning the operation of this
in-depth, comprehensive tutorial on scanning electron micros-
technology or interpretation of information obtained through
copyorthepetrographicexaminationofconcreteandconcrete-
SEM/EDS.
related materials, the reader is directed to the additional
1.3 It is critical that petrographer or operator or both be
publications referenced in the bibliography section of this
familiar with the SEM/EDX (EDS) equipment, specimen
guide.
preparation procedures, and the use of other appropriate
1.7 Units—The values stated in SI units are to be regarded
procedures for this purpose. This guide does not discuss data
asstandard.Nootherunitsofmeasurementareincludedinthis
interpretation. Proper data interpretation is best done by
standard.
individuals knowledgeable about the significance and limita-
tions of SEM/EDX (EDS) and the materials being evaluated.
1.8 This standard does not purport to address all of the
safety concerns, if any, associated with the use of electron
1.4 The SEM provides images that can range in scale from
microscopes, X-ray spectrometers, chemicals, and equipment
alowmagnification(forexample,15×)toahighmagnification
used to prepare samples for electron microscopy. It is the
(for example, 50 000× or greater) of concrete specimens such
as fragments, polished surfaces, or powders.These images can responsibility of the user of this standard to establish appro-
priate safety, health, and environmental practices and deter-
provide information indicating compositional or topographical
variations in the observed specimen. The EDX (EDS) system mine the applicability of regulatory limitations prior to use.
1.9 This international standard was developed in accor-
1 dance with internationally recognized principles on standard-
This guide is under the jurisdiction ofASTM Committee C09 on Concrete and
Concrete Aggregates and is the direct responsibility of Subcommittee C09.65 on
ization established in the Decision on Principles for the
Petrography.
Development of International Standards, Guides and Recom-
Current edition approved Oct. 1, 2022. Published October 2022. Originally
mendations issued by the World Trade Organization Technical
approved in 2010. Last previous edition approved in 2016 as C1723 – 16. DOI:
10.1520/C1723-16R22. Barriers to Trade (TBT) Committee.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
C1723 − 16 (2022)
2. Referenced Documents 3.1.12 stage, n—platform upon which the specimen is
2 placedwithinthevacuumchamberthatcanberemotelymoved
2.1 ASTM Standards:
in various directions.
C125Terminology Relating to Concrete and Concrete Ag-
3.1.13 working distance, n—thedistancebetweenthedetec-
gregates
tor and the sample. Each SEM will have an optimun distance
C294Descriptive Nomenclature for Constituents of Con-
in which X-rays can be collected for EDX (EDS).
crete Aggregates
C295/C295MGuide for Petrographic Examination of Ag-
3.1.14 X-ray detector, n—also known as EDX (EDS) sys-
gregates for Concrete
tem.
C457/C457MTest Method for Microscopical Determination
of Parameters of the Air-Void System in Hardened Con-
4. Description of Equipment
crete
4.1 The principles of the electron system of the scanning
C856/C856MPractice for Petrographic Examination of
electron microscope, the interactions of the electron beam and
Hardened Concrete
the specimen under examination, and the detection systems
C1356 Test Method for Quantitative Determination of
used for the examination are based on concepts that need
Phases in Portland Cement Clinker by Microscopical
understanding if the resulting image and other analytical
Point-Count Procedure
information obtained are to be best resolved and understood.
An abbreviated discussion is provided here. A more compre-
3. Terminology
hensive understanding can be obtained from texts devoted to
3.1 Definitions of Terms Specific to This Standard: 3
this subject (1,2).
3.1.1 BSE, n—backscatter electrons; these are high-energy
4.1.1 SEM Optics:
electronsemittedbackfromthespecimensurface.Elementsof
4.1.1.1 An electron beam is generated in a column consist-
higheratomicnumberwillhavestrongeremissionsandappear
ing of an electron gun and multiple electromagnetic lenses and
brighter.
apertures.Theelectronbeamisgeneratedbyheatingafilament
3.1.2 brightness, n—the amount of energy used to produce
so that it emits electrons. The most common filament for
an X-ray.
general SEM work is tungsten, but other filaments can be used
for increased brightness.The electrons are accelerated towards
3.1.3 charging, n—thebuildupofelectronsonthespecimen
the specimen by an applied potential and then focused by
at the point where the beam impacts the sample. Charging can
lenses and apertures. The energy of the electron beam influ-
alter the normal contrast of the image (usually becomes
ences resolution, image quality, and quantitative and qualita-
brighter)andmaydeflectthebeam.Coatingthespecimenwith
tive X-ray microanalyses.
athinlayerofconductivematerial(suchasgoldorcarbon)can
4.1.1.2 Theelectronbeamisfinelyfocusedthroughelectro-
minimize this effect.
magnetic lenses and apertures. A smaller beam size improves
3.1.4 contrast, n—the difference in intensity of the energy
resolution, but decreases signal intensity.
produced by varying elements when excited.
4.1.1.3 Electron systems operate under vacuum. Specimens
3.1.5 dead-time, n—the time of finite processing during
should be prepared to minimize alteration or damage when
which the circuit is “dead” and unable to accept a new pulse
they are exposed to the vacuum (See 5.1.4). Variable pressure
from the X-rays.
scanning electron microscopes, low vacuum scanning electron
3.1.6 EDX (EDS) (energy-dispersive X-ray spectroscopy),
microscopes (LVSEM), and environmental scanning electron
n—the interaction of the electron beam with atoms in the
microscopes (ESEM) permit the examination of samples con-
sample produces characteristic X-rays having energies and
taining some moisture under low vacuum. The ESEM also
wavelengths unique to atoms.
allows analysis of organic materials. Even in an ESEM,
however, some drying occurs.
3.1.7 live-time, n—howtheacquistionofX-raydataistimed
4.1.2 Signal Generation and Detection:
when the rate of X-ray events between measurements are
4.1.2.1 Theinteractionoftheelectronbeamwiththesample
compared. Opposite of dead-time.
generates several types of signals that can be utilized for
3.1.8 K, L, or M peaks, n—characteristic X-ray intensities
imaging and X-ray microanalysis. The intensities of these
detected for elements.
signals are measured by detectors. The signals allow the
3.1.9 raster, n—to scan as when the beam from the filament
examination and determination of properties such as surface
sweeps back and forth over the sample
topography, elemental composition, and spatial distribution of
3.1.10 SE, n—secondary electrons; these are low-energy
components.Signalintensitiesaregenerallyusedtoprovidean
electrons emitted when the specimen is hit with the beam and
image on a screen.
associated with the topography of the same.
4.1.2.2 The signals that are produced when the electron
beam strikes the specimen surface are secondary electrons
3.1.11 SEM, n—scanning electron microscope.
(SE), backscattered electrons (BSE), and X-rays.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on The boldface numbers in parentheses refer to a list of references at the end of
the ASTM website. this standard.
C1723 − 16 (2022)
4.1.2.3 To generate an image, the electron beam is moved concrete may also contain supplementary cementitious
repeatedly across the specimen to form a raster. The magnifi- materials,organic,inorganic,andmetallicfibers,andentrained
cation is the ratio between the size of the raster and that of the air voids.
screen image.
5.1.3 Cement Paste—The cement paste contains residual
cement,frequentlysupplementarypozzolanicandcementitious
4.1.2.4 Images produced by secondary electrons are most
materials, and various hydration products that together have a
commonlyutilizedfortopographicalimaging.TheSEintensity
complex and porous microstructure. The paste is initially a
depends mainly on the angles between the electron beam and
mixtureofindividualgrainsofcementiousmaterialsandwater,
thespecimensurfaceandbetweenthespecimensurfaceandthe
and may also contain chemical admixtures. Over time, hydra-
detector. The SE intensity is relatively insensitive to the
tion reactions consume the cement and produce various hydra-
specimen composition.
tion products, some of which grow on the surface of cement
4.1.2.5 Images produced by backscattered electrons are
grains, while progressively filling the initial water-filled space.
often used for elemental contrast imaging. The BSE image is
5.1.3.1 Residualportlandcementparticlesappeardenseand
useful for identifying different chemical constituents in con-
angulartosubangular.Aliteusuallyhasatleastonecrystalface
crete. The BSE intensity depends on the average atomic
while belite is usually rounded and sometimes striated. In a
number and density of each phase. The BSE intensity also
BSE image, residual portland cement particles occur as rela-
depends on the angles between the electron beam and the
tively bright objects in a matrix of gray cement hydration
specimen surface and between the specimen surface and the
products.
detector.Therefore,someBSEdetectorscanbemanipulatedto
5.1.3.2 Calcium-silicate-hydrate is the major hydration
observe the sample topography.
product of portland cement and is usually amorphous or very
4.1.2.6 The interaction of the electron beam with atoms in
poorly crystalline. Its morphology varies depending on the
the sample produces characteristic X-rays having energies and
calcium to silica ratio, water to cementitious materials ratio,
wavelengths unique to atoms. Chemical analysis (or micro-
curing conditions, degree of cement hydration, and chemical
analysis) is performed using an X-ray spectrometer that mea-
admixtures. At high magnifications, the morphology of
sures the energies and intensities of the X-rays.The intensities
calcium-silicate-hydrate varies from very fine fibrous growths,
of X-rays depend upon many factors, including electron beam
to sheet-like units, to irregular massive grains.
currents and accelerating voltages, as well as chemical com-
5.1.3.3 Portlandite (calcium hydroxide) is a major phase of
position of the specimen interacting with the electron beam.
cement hydration and occurs in variable sizes and shapes
4.1.2.7 One important parameter for image quality is the
including platy hexagonal crystals and sheet-like masses,
working distance, the distance between specimen surface and
depending on the orientation. Calcium hydroxide is normally
the point where the electron beam exits the electron optics.
observedthroughoutthecementpasteandsometimesdevelops
Small working distances maximize BSE collection efficiency
along paste-aggregate interfaces. It also sometimes occurs as
and improve the image resolution. Long working distances
secondary deposits in voids and cracks.
improve image depth of field for topographical images but
5.1.3.4 Ettringite is a primary product of the reactions
decrease image resolution. The working distance generally
between calcium aluminates and the sulfate phases in cement.
must be within a predetermined range to perform X-ray
It has a characteristic acicular shape. Ettringite often also
microanalysis.
appears as a secondary deposit. Secondary deposits of ettring-
itearecommonlyfoundinvoidsandcracks.X-raymicroanaly-
5. Materials and Features
sis is sometimes required for its identification. A compound
5.1 Important microstructural features include the size and
that has similar morphology is thaumasite (See 5.1.7 on
shape o
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.