Abstract

This document specifies general requirements and test methods for portable dental equipment for use in non-permanent healthcare environments.
This document applies to portable dental units, portable patient chairs, portable operator’s stools, portable operating lights, portable suction source equipment, portable air compressors and other portable dental equipment in instances where these devices are designed and constructed to be transported for use in non-permanent healthcare environments.
NOTE            Particular requirements for specific types of portable dental equipment for use in non-permanent healthcare environments are specified in subsequent parts of this document.
This document does not apply to stationary dental equipment, wearable equipment (e.g. headlamps and loupes), mobile dental equipment or portable dental equipment that is not intended to be used in non-permanent healthcare environments or not designed to be disassembled, folded or packed for human transport between non-permanent healthcare environments. Requirements for stationary dental equipment that can be installed in a dental mobile medical facility (e.g. vehicular or containerized mobile dental clinic) are not considered in this document.

Status
Published
Publication Date
21-Jul-2026
Technical Committee
CEN/TC 55 - Dentistry
Drafting Committee
CEN/TC 55 - Dentistry
Current Stage
6060 - Definitive text made available (DAV) - Publishing
Start Date
22-Jul-2026
Due Date
22-Jul-2026
Completion Date
22-Jul-2026

Buy Documents

Standard

EN ISO 23402-1:2026

English language (18 pages)
Preview
Preview
e-Library read for
×1 day

EN ISO 23402-1:2026 is a European Standard published by CEN and identical to ISO 23402-1:2026. It specifies general requirements and test methods for portable dental equipment used in non-permanent healthcare environments. The standard is for manufacturers, test laboratories, and buyers who need equipment designed to be carried, transported, unpacked, used, and repacked at temporary care sites.

What does EN ISO 23402-1:2026 specify?

EN ISO 23402-1:2026 specifies the general requirements and test methods for portable dental equipment intended for non-permanent healthcare environments. It covers portable dental units, chairs, stools, lights, suction source equipment, air compressors, and related portable dental equipment when they are designed to be transported for use in these settings.

The scope also sets clear boundaries. EN ISO 23402-1:2026 does not cover stationary dental equipment, wearable equipment such as headlamps and loupes, mobile dental equipment, or portable dental equipment that is not meant for non-permanent healthcare environments or not designed to be disassembled, folded, or packed for human transport.

The document is organized into clauses 1 to 10, plus Annex A and a bibliography:

  • Clause 1 - Scope
  • Clause 2 - Normative references
  • Clause 3 - Terms and definitions
  • Clause 4 - Classification
  • Clause 5 - Requirements
  • Clause 6 - Sampling
  • Clause 7 - Measurement and test methods
  • Clause 8 - Manufacturer's instructions
  • Clause 9 - Marking
  • Clause 10 - Packaging
  • Annex A - Applicable testing methodology from IEC 60601-1

What are the key requirements of EN ISO 23402-1:2026?

EN ISO 23402-1:2026 focuses on portability, environmental robustness, usability, and clear manufacturer information. In practice, it tells manufacturers how to classify the equipment, what transport limits to declare, what tests to use, and what instructions and markings users need.

Classification and transport conditions

Clause 4 classifies portable dental equipment by electrical characteristics, intended use environment, supply source, and exposure during transport. The intended use environment is either indoor use or outdoor use, and transport exposure is either limited outdoor exposure and limited carrying distance or rugged outdoor exposure and prolonged carrying distance.

This matters because the classification drives the required transport tests and enclosure protection. A manufacturer has to declare the relevant class so the equipment can be selected and tested for the real conditions it will face.

Transport performance

Clause 5.2 requires the equipment, including any storage case or transport case, to survive transport in the condition described by the manufacturer. Handling features such as grips or straps have to support transport by the intended number of people, and the packed mass must not exceed 20 kg per person for a one-person carry or per person in a multi-person carry.

For indoor use, packed equipment modules must fit through a 0.8 m by 1.9 m door. That is a practical check on whether the equipment can move through typical building openings without special handling.

Clause 5.2 also covers environmental exposure, impact, drop, vibration, and foreign object and liquid ingress during transport. The impact, drop, and vibration tests use ISO 4180:2019, with more severe classes required for rugged transport conditions. Limited outdoor exposure maps to more severe test levels than indoor transport, and rugged outdoor exposure requires the most demanding classification.

Ingress protection

Clause 5.2.9 sets ingress protection for transport conditions, and clause 5.4.4 does the same for the unpacked, ready-for-use condition. For limited outdoor exposure, the enclosure has to meet IP21; for rugged outdoor exposure, it has to meet IP55. For indoor use, the unpacked equipment also has to meet IP21, while outdoor use requires IP55.

This matters for dust and moisture control. In practice, manufacturers need to choose enclosures and transport cases that match the setting and can protect components from the expected environment.

Utilities, operation, and surface processing

Clause 5.3 requires the manufacturer to specify the utility needs, including electrical mains voltage, frequency, and power, and any safety requirements such as fuses, circuit breakers, residual-current device, or protective earth connection. If the equipment uses batteries, the battery type has to be specified.

Clause 5.4 requires the equipment to operate within the ambient temperature and atmospheric pressure range stated by the manufacturer. It also brings in the usability requirements of IEC 62366-1, so manufacturers have to address risks from poor identification, marking, and documents.

Clause 5.4.3 requires instructions for processing external surfaces and connection interfaces, and the materials used on external and touchable surfaces and on the transport case have to withstand the recommended cleaning and disinfecting agents under ISO 21530. For users, this means the surfaces must be practical to clean and disinfect without damage or discoloration.

Sampling, test methods, instructions, marking, and packaging

Clause 6 requires type tests to be performed on one representative sample. Clause 7 limits conformity checking to visual inspection of the equipment and the documentation. Clauses 8 to 10 then define what the manufacturer must provide: durable instructions, technical description data, markings on the equipment, markings on the packaging, and packaging that protects the equipment during anticipated transport.

What terms does EN ISO 23402-1:2026 define?

  • dental equipment - Furniture, machines, apparatus, and accessories made for dentistry or related procedures.
  • portable dental equipment - Transportable dental equipment intended to be moved from one location to another while being carried by one or more persons.
  • non-permanent healthcare environment - A setting used temporarily to provide dental or medical treatment.
  • transport case - A packaging container intended both for transport and for protecting the portable dental equipment.
  • externally supplied source - A supply source, such as electrical power, water, compressed air, or suction, that must be connected during use.
  • self-contained source - An internal supply source, such as a battery, air compressor, suction source, or water reservoir, that does not need an external connection during use.
  • limited outdoor exposure and limited carrying distance - A transport condition with short outdoor exposure and short carrying distance.
  • rugged outdoor exposure and prolonged carrying distance - A transport condition with severe weather exposure and extended carrying distance, such as over rugged terrain.

Who uses EN ISO 23402-1:2026?

EN ISO 23402-1:2026 is used by manufacturers of portable dental equipment, test laboratories, and compliance teams that need to verify transport durability, enclosure protection, usability, labeling, and packaging. It is also relevant to procurement staff and quality managers selecting equipment for temporary clinics, field operations, outreach programs, and similar settings.

The equipment is described for use by dental professionals in military field environments, humanitarian aid clinics, public health outreach clinics, patient residences, long-term care facilities, prisons, schools, workplaces, and temporary field clinics. Those users need equipment that can be carried, assembled, disinfected, and put into service reliably in variable conditions.

What changed in EN ISO 23402-1:2026 from the previous edition?

EN ISO 23402-1:2026 supersedes EN ISO 23402-1:2020 and is the second edition.

The main changes stated for this edition are:

  • Clause 5.2.6, 5.2.7, and 5.2.8 were updated so impact, drop, and vibration requirements depend on the transport exposure classification.
  • Clause 5.4.4 was added.
  • Clause 8.3 was expanded to include the classification level for drop height and the classification level for rough road transport distance.
  • General editorial updates were made.

Which standards are used with EN ISO 23402-1:2026?

StandardWhat it contributes
ISO 1942Vocabulary for dentistry and the definition source for dental equipment
ISO 4180:2019Transport-package test schedules and the impact, drop, and vibration classifications used in clause 5.2
ISO 9687Graphical symbols for dental equipment used for marking
ISO 15223-1Symbols used with manufacturer information, including marking symbols
ISO 21530Resistance of dental equipment surfaces to chemical disinfectants
IEC 60601-1Basic safety and essential performance requirements and several applicable test methods
IEC 62366-1Usability engineering requirements
IEC 80601-2-60Particular requirements for the basic safety and essential performance of dental equipment
IEC 60529Ingress protection code used for IP21 and IP55

Clause 4 also depends on the relevant IEC classification when the equipment is electrically operated, and the ISO 23402 series is intended to continue with later parts for specific types of portable dental equipment.

What does the EN ISO 23402-1:2026 document contain?

EN ISO 23402-1:2026 contains classification rules, transport-performance requirements, utility declarations, operating requirements, and the content that manufacturers must include in instructions, technical descriptions, marking, and packaging. It also defines the visual inspection approach used to check equipment and documentation.

Annex A is informative and lists testing methodology from IEC 60601-1 that often applies to portable dental equipment, including tests for markings, leakage current, dielectric withstand, mechanical strength, vibration, X-radiation, and microwave radiation. In practice, that annex helps test teams map this standard to relevant electrical-safety checks.

The document is written for design verification, type testing, and documentation review. It gives manufacturers and buyers a common basis for deciding whether a portable dental product is suitable for use in non-permanent healthcare environments.

Relations

Effective Date
08-Apr-2024
Effective Date
21-Oct-2026
Effective Date
21-Oct-2026
Effective Date
21-Oct-2026
Effective Date
21-Oct-2026
Effective Date
21-Oct-2026
Effective Date
21-Oct-2026
Effective Date
21-Oct-2026
Effective Date
21-Oct-2026
Effective Date
21-Oct-2026
Effective Date
12-Feb-2026

Buy Documents

Standard

EN ISO 23402-1:2026

English language (18 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

EN ISO 23402-1:2026 is a standard published by the European Committee for Standardization (CEN). Its full title is "Dentistry - Portable dental equipment for use in non‐permanent healthcare environments - Part 1: General requirements (ISO 23402-1:2026)". This standard covers: This document specifies general requirements and test methods for portable dental equipment for use in non-permanent healthcare environments. This document applies to portable dental units, portable patient chairs, portable operator’s stools, portable operating lights, portable suction source equipment, portable air compressors and other portable dental equipment in instances where these devices are designed and constructed to be transported for use in non-permanent healthcare environments. NOTE            Particular requirements for specific types of portable dental equipment for use in non-permanent healthcare environments are specified in subsequent parts of this document. This document does not apply to stationary dental equipment, wearable equipment (e.g. headlamps and loupes), mobile dental equipment or portable dental equipment that is not intended to be used in non-permanent healthcare environments or not designed to be disassembled, folded or packed for human transport between non-permanent healthcare environments. Requirements for stationary dental equipment that can be installed in a dental mobile medical facility (e.g. vehicular or containerized mobile dental clinic) are not considered in this document.

This document specifies general requirements and test methods for portable dental equipment for use in non-permanent healthcare environments. This document applies to portable dental units, portable patient chairs, portable operator’s stools, portable operating lights, portable suction source equipment, portable air compressors and other portable dental equipment in instances where these devices are designed and constructed to be transported for use in non-permanent healthcare environments. NOTE            Particular requirements for specific types of portable dental equipment for use in non-permanent healthcare environments are specified in subsequent parts of this document. This document does not apply to stationary dental equipment, wearable equipment (e.g. headlamps and loupes), mobile dental equipment or portable dental equipment that is not intended to be used in non-permanent healthcare environments or not designed to be disassembled, folded or packed for human transport between non-permanent healthcare environments. Requirements for stationary dental equipment that can be installed in a dental mobile medical facility (e.g. vehicular or containerized mobile dental clinic) are not considered in this document.

EN ISO 23402-1:2026 is classified under the following ICS (International Classification for Standards) categories: 11.060.20 - Dental equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

EN ISO 23402-1:2026 has the following relationships with other standards: It is inter standard links to EN ISO 23402-1:2020, ISO 1942:2020, ISO 21530:2004, IEC 60529:1989, ISO 4180:2019, ISO 9687:2015, IEC 62366-1:2015, IEC 80601-2-60:2019, ISO 15223-1:2021, IEC 60601-1:2005, ISO 23402-1:2026. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

EN ISO 23402-1:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 dental equipment
  • 3.2 portable dental equipment
  • 3.3 non‐permanent healthcare environment
  • 3.4 transport case
  • 4 Classification
  • 4.1 General
  • 4.2 For electrically operated devices
  • 4.3 According to intended use environment
  • 4.4 According to supply sources
  • 4.5 According to exposure during transport
  • 5 Requirements
  • 5.1 General
  • 5.2 Transport requirements
  • 5.3 Utility requirements
  • 5.4 Operational requirements
  • 6 Sampling
  • 7 Measurement and test methods
  • 7.1 Visual inspection of the equipment
  • 7.2 Visual inspection of the documentation
  • 8 Manufacturer's instructions
  • 8.1 General
  • 8.2 Instructions for use
  • 8.3 Technical description
  • 9 Marking
  • 9.1 Marking on the equipment
  • 9.2 Marking of packaging
  • 10 Packaging
  • Annex A
  • 4.11 Power input
  • 5.7 Humidity conditioning
  • 8.7 Leakage current tests
  • 9.7 Hydrostatic pressure test
  • 10.3 Microwave radiation
  • Bibliography

Introduction

Transportable dental equipment is used by dental professionals to provide care to patients in a variety of settings. Because the intended use applications and intended means for transporting such equipment vary considerably, a wide variety of transportable dental equipment is commercially available. For example, certain transportable equipment is designed and constructed to be carried or rolled on its own wheels between rooms within a healthcare facility, while other transportable dental equipment is made to be folded and packed to carry over terrain which can be rugged and used in transient dental care settings which can have only limited shelter and utility services.

Transportable equipment that can be moved from one location to another while being carried by one or more persons is referred to as portable equipment. The term "portable equipment" applies to equipment that can be carried from room to room in a given facility or to remote parts of the world. This document focuses on portable dental equipment which is specifically designed and constructed to be transported between non-clinical environments and used by dental professionals to provide dental care in such settings, including temporary field clinics.

Such portable dental equipment for use in non-permanent healthcare environments enables dental professionals to provide a high standard of care to patients who do not have access to, or are not able to, travel to traditional health care facilities. Settings in which this equipment is commonly used include military field environments, humanitarian aid field clinics, public health outreach clinics, patient residences, long-term care facilities, prisons, schools and workplaces.

A number of trends in health care have driven increased utilization of portable dental equipment in nonpermanent healthcare environments. Military forces use portable dental equipment in support of mobilized forces or for humanitarian outreach. A variety of government and non-government organizations are increasingly providing humanitarian dental care to underserved populations and populations affected by disasters. Civilian health care workers are also increasingly providing dental services to a growing population who are simply unable to visit traditional dental clinics due to age, disability or income. Academic and research bodies regularly conduct dental education programmes (e.g. dentistry, dental hygiene and dental assisting), particularly at external/off-site locations. The transport and end-use conditions for portable dental equipment used in non-permanent healthcare environments drive certain unique requirements which generally do not apply to portable, mobile or stationary dental equipment used in traditional dental clinics or hospitals. Because portable equipment used in non-permanent healthcare environments is intended to be moved between venues, and in some cases carried over rugged terrain or in inclement conditions, it needs to be designed and constructed to be safely transported by humans without damage, be efficiently assembled and disassembled, and deliver reliable service at the point of use. Special consideration is given to the austerity of the environment in which the equipment can be used and the availability and quality of utility supplies (e.g. electrical power, water, compressed air). In order for the equipment to be sufficiently portable and capable of operating in extreme conditions, certain requirements for dental equipment intended for use in traditional clinical settings can be impractical and should be reconsidered for portable dental equipment for use in non-permanent healthcare environments. There can also be unique safety and infection control concerns to consider. The ISO 23402 series aims to standardize requirements for portable dental equipment for use in nonpermanent healthcare environments.

SIST EN ISO 23402-1:2026

SIST EN ISO 23402-1:2026 International Standard ISO 23402‐1:2026(en) Dentistry — Portable dental equipment for use in non‐ permanent healthcare environments — Part 1:

General requirements

1 Scope

This document specifies general requirements and test methods for portable dental equipment for use in non-permanent healthcare environments.

This document applies to portable dental units, portable patient chairs, portable operator’s stools, portable operating lights, portable suction source equipment, portable air compressors and other portable dental equipment in instances where these devices are designed and constructed to be transported for use in nonpermanent healthcare environments. NOTE Particular requirements for specific types of portable dental equipment for use in non-permanent healthcare environments are specified in subsequent parts of this document. This document does not apply to stationary dental equipment, wearable equipment (e.g. headlamps and loupes), mobile dental equipment or portable dental equipment that is not intended to be used in nonpermanent healthcare environments or not designed to be disassembled, folded or packed for human transport between non-permanent healthcare environments. Requirements for stationary dental equipment that can be installed in a dental mobile medical facility (e.g. vehicular or containerized mobile dental clinic) are not considered in this document.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 1942, Dentistry — Vocabulary ISO 4180:2019, Packaging — Complete, filled transport packages — General rules for the compilation of performance test schedules ISO 4180, Packaging — Complete, filled transport packages — General rules for the compilation of performance test schedules ISO 9687, Dentistry — Graphical symbols for dental equipment ISO 15223-1, Medical devices — Symbols to be used with information to be supplied by the manufacturer — Part 1: General requirements ISO 21530, Dentistry — Materials used for dental equipment surfaces — Determination of resistance to chemical disinfectants IEC 60601-1, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance IEC 62366-1, Medical devices — Part 1: Application of usability engineering to medical devices

SIST EN ISO 23402-1:2026 ISO 23402‐1:2026(en) IEC 80601-2-60, Medical electrical equipment — Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment IEC 60529:1989+AMD 1:1999+AMD2: 2013, Degrees of protection provided by enclosures (IP Code)

3 Terms and definitions

For the purposes of this document, the terms and definitions given in IEC 60601-1 and ISO 1942 as well as the following apply.

ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • IEC Electropedia: available at http:// www .electropedia .org/
  • ISO Online browsing platform: available at http:// www .iso .org/ obp

3.1 dental equipment

furniture, machines, apparatus and accessories thereto, specially made for use in the practice of either dentistry or its associated procedures or both [1] [SOURCE: ISO 1942:2020 , 3.5.1]

3.2 portable dental equipment

transportable dental equipment (3.1) intended to be moved from one location to another while being carried by one or more persons

3.3 non‐permanent healthcare environment

setting for temporarily providing dental or medical treatment

3.4 transport case

packaging container intended to be used for transport and for protection of the portable dental equipment (3.2)

4 Classification

4.1 General

The classifications outlined in this clause pertain to electrical characteristics, end-use facility provisions, supply provisions and portability considerations applicable to portable dental equipment for use in nonpermanent healthcare environments.

4.2 For electrically operated devices

If the equipment is electrically operated, classification shall be in accordance with IEC 60601-1 and IEC 80601-2-60.

4.3 According to intended use environment

Portable dental equipment for use in non-permanent healthcare environments is classified according to the intended use environment as follows:

a) For indoor use: Equipment of this class is intended to be used in indoor environments, such as residential settings, long-term care facilities (e.g. nursing homes, assisted living communities), offices, schools, convention centres, corrections facilities (e.g. prisons) and other permanent building facilities providing protection against weather elements.

SIST EN ISO 23402-1:2026 ISO 23402‐1:2026(en) b) For outdoor use: Equipment of this class is intended to be used in outdoor environments where only rudimentary shelter, such as a tent or tarpaulin, can be available for limiting exposure to weather elements, e.g. military or humanitarian field operations.

4.4 According to supply sources

Portable dental equipment for use in non-permanent healthcare environments is classified according to the supply sources (e.g. electrical power, water, compressed air, suction) required to operate the equipment as follows:

a) Externally supplied source: Equipment of this class requires connection to a specified supply source during use.

...