EN ISO 14596:1998/AC:1999
(Corrigendum)Petroleum products - Determination of sulfur content - Wavelength-dispersive X-ray fluorescence spectrometry (ISO 14596:1998)
Petroleum products - Determination of sulfur content - Wavelength-dispersive X-ray fluorescence spectrometry (ISO 14596:1998)
ISO - Take over of ISO Technical Corrigendum
Mineralölerzeugnisse - Bestimmung des Schwefelgehaltes - Wellenlängendispersive Röntgenfluoreszenz-Analyse (ISO 14596:1998)
Produits pétroliers - Dosage du soufre - Spectrométrie de fluorescence X dispersive en longueur d'onde (ISO 14596:1998)
Naftni proizvodi - Določevanje žvepla - Valovno disperzivna rentgenska fluorescenčna spektrometrija (ISO 14596:1998)
General Information
Relations
Standards Content (Sample)
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.HQVNDMineralölerzeugnisse - Bestimmung des Schwefelgehaltes - Wellenlängendispersive Röntgenfluoreszenz-Analyse (ISO 14596:1998)Produits pétroliers - Dosage du soufre - Spectrométrie de fluorescence X dispersive en longueur d'onde (ISO 14596:1998)Petroleum products - Determination of sulfur content - Wavelength-dispersive X-ray fluorescence spectrometry (ISO 14596:1998)75.080Naftni proizvodi na splošnoPetroleum products in generalICS:Ta slovenski standard je istoveten z:EN ISO 14596:1998/AC:1999SIST EN ISO 14596:1999/AC:2002en01-oktober-2002SIST EN ISO 14596:1999/AC:2002SLOVENSKI
STANDARD
SIST EN ISO 14596:1999/AC:2002
SIST EN ISO 14596:1999/AC:2002
SIST EN ISO 14596:1999/AC:2002
ICS
75.080Ref. No. ISO 14596:1998/Cor.1:1999(E)©
ISO 1999Printed in SwitzerlandAINTERNATIONAL STANDARD ISO 14596:1998TECHNICAL CORRIGENDUM 1Published 1999-07-15INTERNATIONAL ORGANIZATION FOR STANDARD
...
This May Also Interest You
This document provides guidance on taking and handling samples related to oil spill identification in legal proceedings. Guidance is given on obtaining samples from both the spill and its potential source.
Preservation of evidence is an essential part of legal procedures and this document presents appropriate oil sampling procedures.
WARNING - The use of this document can involve hazardous materials, operations and equipment.
This document does not purport to address all of the safety problems associated with its use. It is the responsibility of users of this document to take appropriate measures to ensure the safety and health of personnel prior to the application of the standard, and to determine the applicability of any other restrictions for this purpose.
IMPORTANT - Most countries have teams with specialists trained in sampling on board of ships. Do not take unnecessary risks, seek assistance from such teams where available.
NOTE For the sake of clarity, the word ‘oil’ is used throughout this document. It can equally refer to crude oil, a petroleum product or mixtures of such.
- Standard31 pagesEnglish languagesale 10% offe-Library read for×1 day
- Draft29 pagesEnglish languagesale 10% offe-Library read for×1 day
This document specifies a method to identify and compare the compositional characteristics of oil samples. Specifically, it describes the detailed analytical and data processing methods for identifying the characteristics of spill samples and establishing their correlation to suspected source oils. Even when samples or data from suspected sources are not available for comparison, establishing the specific nature (e.g. refined petroleum, crude oil, waste oil, etc.) of the spilled oil still helps to constrain the possible source(s).
This methodology is restricted to petroleum related products containing a significant proportion of hydrocarbon-components with a boiling point above 150 °C. Examples are: crude oils, higher boiling condensates, diesel oils, residual bunker or heavy fuel oils, lubricants, and mixtures of bilge and sludge samples, as well as distillate fuels and blends. While the specific analytical methods are perhaps not appropriate for lower boiling oils (e.g. kerosene, jet fuel, or gasoline), the general concepts described in this methodology, i.e. statistical comparison of weathering-resistant diagnostic ratios, are applicable in spills involving these kinds of oils.
Paraffin based products (e.g. waxes, etc.) are outside the scope of this method because too many compounds are removed during the production process [37]. However, the method can be used to identify the type of product involved.
Although not directly intended for identifying oil recovered from groundwater, vegetation, wildlife/tissues, soil, or sediment matrices, they are not precluded. However, caution is needed as extractable compounds can be present in these matrices that alter and/or contribute additional compounds compared to the source sample. If unrecognized, the contribution from the matrix can lead to false “non-matches”. It is therefore advisable to analyse background sample(s) of the matrix that appear unoiled.
When analysing “non-oil” matrices additional sample preparation (e.g. clean-up) is often required prior to analysis and the extent to which the matrix affects the correlation achieved is to be considered. Whether the method is applicable for a specific matrix depends upon the oil concentration compared to the “matrix concentration”. In matrices containing high concentrations of oil, a positive match can still be concluded. In matrices containing lower concentrations of oil, a false “non-match” or an “inconclusive match” can result from matrix effects. Evaluation of possible matrix effects is beyond the scope of this document.
- Standard219 pagesEnglish languagesale 10% offe-Library read for×1 day
- Draft194 pagesEnglish languagesale 10% offe-Library read for×1 day
This document describes three procedures (A, B and C) covering determinations of flash no-flash and flash point.
Rapid equilibrium procedures A and B are applicable to flash no-flash and flash point tests of paints, including water-borne paints, varnishes, binders for paints and varnishes, adhesives, solvents, petroleum products including aviation turbine, diesel and kerosene fuels, fatty acid methyl esters and related products over the temperature range –30 °C to 300 °C. The rapid equilibrium procedures are used to determine whether a product will or will not flash at a specified temperature (flash no-flash procedure A) or the flash point of a sample (procedure B). When used in conjunction with the flash detector (A.1.6), this document is also suitable to determine the flash point of fatty acid methyl esters (FAME). The validity of the precision is given in Table 2.
Non-equilibrium procedure C is applicable to petroleum products including aviation turbine, diesel and kerosine fuels, and related petroleum products, over the temperature range –20 °C to 300 °C. The non-equilibrium procedure is automated to determine the flash point. Precision has been determined over the range 40 °C to 135 °C.
For specifications and regulations, procedures A or B are routinely used (see 10.1.1).
- Standard35 pagesEnglish languagesale 10% offe-Library read for×1 day
- Draft33 pagesEnglish languagesale 10% offe-Library read for×1 day
This document specifies the process and methodology for the construction, operation, and maintenance of statistical control charts to assess if a laboratory's execution of a standard test method is in-statistical-control and how to establish and validate the 'in-statistical-control' status.
It specifies control charts that are most appropriate for ISO/TC 28 test methods where the dominant common cause variation is associated with the long term, multiple operator conditions. The control charts specified for determination of in-statistical-control are: individual (I), moving range of 2 (MR2), and either the exponentially weighted moving average (EWMA) or zone-based run rules [similar to Western Electric (WE) run rules[3]] as sensitivity enhancement strategy to support the I-chart.
The procedures in this document have been primarily designed for numerical results obtained from testing of control samples prepared from a homogenous source of petroleum and related products in a manner that preserves the homogeneity of properties of interest between control samples. If the test method permits, a certified reference material (CRM) sample is used as a control sample provided the sample composition is representative of the material being tested and is not a pure compound; if this is done then the laboratory best establishes its own mean for the CRM sample.
This document is applicable to properties of interest that are (known to be) stable over time, and for data sets with sufficient resolution to support validation of the assumption that the data distribution can be approximately represented by the normal (Gaussian) model. Mitigating strategies are suggested for situations where the assumption cannot be validated.
- Standard44 pagesEnglish languagesale 10% offe-Library read for×1 day
- Draft37 pagesEnglish languagesale 10% offe-Library read for×1 day
- Amendment7 pagesEnglish languagesale 10% offe-Library read for×1 day
- Draft5 pagesEnglish languagesale 10% offe-Library read for×1 day
This document specifies a test method for the determination of the content of mono-aromatic, di-aromatic and tri+-aromatic hydrocarbons in diesel fuels, paraffinic diesel fuels and petroleum distillates.
This document defines two procedures, A and B.
Procedure A is applicable to diesel fuels that may contain fatty acid methyl esters (FAME) up to 30 % (V/V) (as in [1], [2] or [3]) and petroleum distillates in the boiling range from 150 °C to 400 °C (as in [4].
Procedure B is applicable to paraffinic diesel fuels with up to 7 % (V/V) FAME. This procedure does not contain a dilution of the sample in order to determine the low levels of aromatic components in these fuels.
The polycyclic aromatic hydrocarbons content is calculated from the sum of di-aromatic and tri+-aromatic hydrocarbons and the total content of aromatic compounds is calculated from the sum of the individual aromatic hydrocarbon types.
Compounds containing sulfur, nitrogen and oxygen can interfere in the determination; mono-alkenes do not interfere, but conjugated di-alkenes and poly-alkenes, if present, can do so.
NOTE 1 For the purpose of this European Standard, the terms "% (m/m)" and "% (V/V)" are used to represent the mass fraction, µ, and the volume fraction, φ, of a material respectively.
NOTE 2 By convention, the aromatic hydrocarbon types are defined on the basis of their elution characteristics from the specified liquid chromatography column relative to model aromatic compounds. Their quantification is performed using an external calibration with a single aromatic compound for each of them, which may or may not be representative of the aromatics present in the sample. Alternative techniques and test methods may classify and quantify individual aromatic hydrocarbon types differently.
NOTE 3 Backflush is part of laboratory-internal maintenance.
WARNING - The use of this standard can involve hazardous materials, operations and equipment. This standard does not purport to address all of the safety problems associated with its use. It is the responsibility of users of this standard to take appropriate measures to ensure the safety and health of personnel prior to application of the standard, and fulfil statutory and regulatory requirements for this purpose.
- Standard21 pagesEnglish languagesale 10% offe-Library read for×1 day
This document specifies a fluorescent indicator adsorption method for the determination of hydrocarbon types over the concentration ranges from 5 % (V/V) to 99 % (V/V) aromatic hydrocarbons, 0,3 % (V/V) to 55 % (V/V) olefins, and 1 % (V/V) to 95 % (V/V) saturated hydrocarbons in petroleum fractions that distil below 315 °C. This method can apply to concentrations outside these ranges, but the precision has not been determined.
When samples containing oxygenated blending components are analysed, the hydrocarbon type results can be reported on an oxygenate-free basis or, when the oxygenate content is known, the results can be corrected to a total-sample basis.
This test method is applicable to full boiling range products. Cooperative data have established that the precision statement does not apply to petroleum fractions with narrow boiling ranges near the 315 °C limit. Such samples are not eluted properly, and results are erratic.
It does not apply to samples containing dark-coloured components that interfere with reading the chromatographic bands that cannot be analysed.
NOTE 1 The oxygenated blending components methanol, ethanol, tert-butyl methyl ether (MTBE), methyl tert-pentyl ether (TAME) and tert-butyl ethyl ether (ETBE) do not interfere with the determination of hydrocarbon types at concentrations normally found in commercial petroleum blends. These oxygenated compounds are not detected since they elute with the alcohol desorbent. The effects of other oxygenated compounds are individually verified.
NOTE 2 For the purposes of this document, the terms “% (m/m)” and “% (V/V)” are used to represent respectively the mass fraction and the volume fraction.
WARNING — The use of this document can involve hazardous materials, operations and equipment. This document does not purport to address all of the safety problems associated with its use. It is the responsibility of the user of this document to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard19 pagesEnglish languagesale 10% offe-Library read for×1 day
- Draft19 pagesEnglish languagesale 10% offe-Library read for×1 day
This European Standard describes a method for the determination of the boiling range distribution of petroleum products by capillary gas chromatography using flame ionization detection. The standard is applicable to stabilized crude oils and for the boiling range distribution and the recovery up to and including n-nonane. A stabilized crude oil is defined as having a Reid Vapour Pressure equivalent to or less than 82,7 kPa as determined by IP 481 [3].
NOTE For the purposes of this European Standard, the terms “% (m/m)” and “% (V/V)” are used to represent respectively the mass fraction, ω, and the volume fraction, φ.
WARNING —The use of this European Standard may involve hazardous materials, operations and equipment. This European Standard does not purport to address all of the safety problems associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and to determine the applicability of regulatory limitations prior to use.
- Standard31 pagesEnglish languagesale 10% offe-Library read for×1 day
- Draft30 pagesEnglish languagesale 10% offe-Library read for×1 day
- Amendment7 pagesEnglish languagesale 10% offe-Library read for×1 day
- Draft4 pagesEnglish languagesale 10% offe-Library read for×1 day
This document specifies the gas chromatographic (GC) method for the determination of saturated, olefinic and aromatic hydrocarbons in automotive motor gasoline and ethanol (E85) automotive fuel. Additionally, the benzene and toluene content, oxygenated compounds and the total oxygen content can be determined.
NOTE 1 For the purposes of this document, the terms % (m/m) and % (V/V) are used to represent respectively the mass fraction, w, and the volume fraction, φ.
This document defines two procedures, A and B.
Procedure A is applicable to automotive motor gasoline with total aromatics of 19,32 % (V/V) up to 46,29 % (V/V); total olefins from 0,40 % (V/V) up to 26,85 % (V/V); oxygenates from 0,61 % (V/V) up to 9,85 % (V/V); oxygen content from 1,50 % (m/m) to 12,32 % (m/m); benzene content from 0,38 % (V/V) up to 1,98 % (V/V) and toluene content from 5,85 % (V/V) up to 31,65 % (V/V).
The method has also been tested for individual oxygenates. A precision has been determined for a total volume of methanol from 1,05 % (V/V) up to 16,96 % (V/V); a total volume of ethanol from 0,50 % (V/V) up to 17,86 % (V/V); a total volume of MTBE from 0,99 % (V/V) up to 15,70 % (V/V), a total volume of ETBE from 0,99 % (V/V) up to 15,49 % (V/V), a total volume of TAME from 0,99 % (V/V) up to 5,92 % (V/V), and a total volume of TAEE from 0,98 % (V/V) up to 15,59 % (V/V).
Although this test method can be used to determine higher-olefin contents of up to 50 % (V/V), the precision for olefins was tested only in the range from 0,40 % (V/V) to 26,85 % (V/V).
Although specifically developed for the analysis of automotive motor gasoline that contains oxygenates, this test method can also be applied to other hydrocarbon streams having similar boiling ranges, such as naphthas and reformates.
NOTE 2 For Procedure A, applicability of this document has also been verified for the determination of n-propanol, acetone, and di-isopropyl ether (DIPE). However, no precision data have been determined for these compounds.
Procedure B describes the analysis of oxygenated groups (ethanol, methanol, ethers, C3 – C5 alcohols) in ethanol (E85) automotive fuel containing ethanol between 50 % (V/V) and 85 % (V/V). The gasoline is diluted with an oxygenate-free component to lower the ethanol content to a value below 20 % (V/V) before the analysis by GC.
The sample can be fully analysed including hydrocarbons. Precision data for the diluted sample are only available for the oxygenated groups.
NOTE 3 For Procedure B, the precision can be used for an ethanol fraction from about 50 % up to 85 % (V/V). For the ether fraction, the precision as specified in Table 6 can be used for samples containing at least 11 % (V/V) of ethers. For the higher alcohol fraction, too few data were obtained to derive a full precision statement and the data presented in Table 6 are therefore only indicative.
NOTE 4 An overlap between C9 and C10 aromatics can occur. However, the total is accurate. Isopropyl benzene is resolved from the C8 aromatics and is included with the other C9 aromatics.