EN ISO 11810-2:2007
(Main)Lasers and laser-related equipment - Test method and classification for the laser-resistance of surgical drapes and/or patient-protective covers - Part 2: Secondary ignition (ISO 11810-2:2007)
Lasers and laser-related equipment - Test method and classification for the laser-resistance of surgical drapes and/or patient-protective covers - Part 2: Secondary ignition (ISO 11810-2:2007)
This International Standard applies to disposable and reusable, as well as woven and non-woven materials used as surgical drapes and other patient protective covers which claim to be laser-resistant.
The purpose of this standard is to provide a standardized method for testing and classifying surgical drapes and other patient protective covers with respect to laser-induced hazards. An appropriate classification system is given. It is not the purpose of this standard to serve as a general fire safety specification. This standard is limited to testing the secondary ignition of materials that are rated I1 or I2 from ISO 11810-1.
All materials reflect portions of the beam and it is necessary for the user to decide whether specular reflection may be a hazard. This measurement, however, is not covered in this standard.
Laser und Laseranlagen - Prüfverfahren und Einstufung zur Laserresistenz von Operationstüchern und/oder anderen Abdeckungen zum Schutz des Patienten - Teil 2: Sekundäre Entzündung (ISO 11810-2:2007)
Dieser Teil von ISO 11810 gilt für Einwegprodukte und wieder verwendbare gewebte Materialien und Vliesstoffe, die als Operationstücher und andere Abdeckungen zum Schutz des Patienten mit dem Anspruch auf Laserstrahlungsbeständigkeit verwendet werden.
Der Zweck dieses Teils von ISO 11810 ist, ein genormtes Verfahren zur Prüfung und Einstufung von Operationsabdecktüchern und anderen Abdeckungen zum Schutz des Patienten hinsichtlich der durch Laser hervorgerufenen Gefährdungen zur Verfügung zu stellen. Ein geeignetes Einstufungssystem wird angegeben. Der Zweck dieses Teils von ISO 11810 besteht nicht darin, als allgemeine Brandsicherheitsvorschrift zu dienen. Dieser Teil von ISO 11810 beschränkt sich darauf, ein Verfahren zur Prüfung der sekundären Entzündung für Materialien festzulegen, die nach ISO 11810 1 mit I1 oder I2 bewertet wurden.
Alle Materialien reflektieren Anteile des Strahls, und es ist für den Anwender erforderlich, entscheiden zu können, ob die spiegelnde Reflexion eine Gefährdung darstellen kann. Diese Messungen werden jedoch in diesem Teil von ISO 11810 nicht erfasst.
Prüfergebnisse nach diesem Teil von ISO 11810 sollten nicht auf andere Wellenlängen und Zeitverhaltensweisen angewendet werden.
Für diesen Teil von ISO 11810 wurde der 20 W CO2 Laser (Dauerstrich) ausgewählt.
ANMERKUNG Für diesen Teil von ISO 11810 wurde der 20 W CO2 Laser (Dauerstrich) ausgewählt. Die Anwender der nach diesem Verfahren geprüften Produkte werden warnend darauf hingewiesen, dass die Laserstrahlungsbeständigkeit eines Operationstuches und/oder einer anderen Abdeckung zum Schutz des Patienten von der Wellenlänge abhängt und dass das Tuch und/oder die Abdeckung bei der Wellenlänge geprüft werden sollte, für die ihr Einsatz vorgesehen ist. Bei Prüfung unter Anwendung anderer Wellenlängen müssen die Leistungswerte und Anregungsarten explizit angegeben werden.
Lasers et équipements associés aux lasers - Méthode d'essai et classification de la résistance au laser pour des draps chirurgicaux et/ou des couvertures de protection des patients - Partie 2: Inflammation secondaire (ISO 11810-2:2007)
Laserji in laserska oprema - Preskusna metoda in razvrstitev za ugotavljanje odpornosti proti laserju za kirurške zastirke in/ali za varovalna pokrivala za paciente - 2. del: Sekundarno zgorevanje (ISO 11810-2:2007)
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
01-oktober-2007
Laserji in laserska oprema - Preskusna metoda in razvrstitev za ugotavljanje
odpornosti proti laserju za kirurške zastirke in/ali za varovalna pokrivala za
paciente - 2. del: Sekundarno zgorevanje (ISO 11810-2:2007)
Lasers and laser-related equipment - Test method and classification for the laser-
resistance of surgical drapes and/or patient-protective covers - Part 2: Secondary ignition
(ISO 11810-2:2007)
Laser und Laseranlagen - Prüfverfahren und Einstufung zur Laserresistenz von
Operationstüchern und/oder anderen Abdeckungen zum Schutz des Patienten - Teil 2:
Sekundäre Entzündung (ISO 11810-2:2007)
Lasers et équipements associés aux lasers - Méthode d'essai et classification de la
résistance au laser des draps chirurgicaux et/ou des couvertures de protection des
patients - Partie 2: Inflammation secondaire (ISO 11810-2:2007)
Ta slovenski standard je istoveten z: EN ISO 11810-2:2007
ICS:
11.040.30 Operacijski instrumenti in Surgical instruments and
materiali materials
13.340.99 Druga varovalna oprema Other protective equipment
31.260 Optoelektronika, laserska Optoelectronics. Laser
oprema equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD
EN ISO 11810-2
NORME EUROPÉENNE
EUROPÄISCHE NORM
May 2007
ICS 11.040.30; 31.260; 13.340.99
English Version
Lasers and laser-related equipment - Test method and
classification for the laser-resistance of surgical drapes and/or
patient-protective covers - Part 2: Secondary ignition (ISO
11810-2:2007)
Lasers et équipements associés aux lasers - Méthode Laser und Laseranlagen - Prüfverfahren und Einstufung zur
d'essai et classification de la résistance au laser pour des Laserresistenz von Operationstüchern und/oder anderen
draps chirurgicaux et/ou des couvertures de protection des Abdeckungen zum Schutz des Patienten - Teil 2:
patients - Partie 2: Inflammation secondaire (ISO 11810- Sekundäre Entzündung (ISO 11810-2:2007)
2:2007)
This European Standard was approved by CEN on 26 April 2007.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the CEN Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CEN member into its own language and notified to the CEN Management Centre has the same status as the
official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland,
France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal,
Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre: rue de Stassart, 36 B-1050 Brussels
© 2007 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 11810-2:2007: E
worldwide for CEN national Members.
Foreword
This document (EN ISO 11810-2:2007) has been prepared by Technical Committee ISO/TC 172
"Optics and optical instruments" in collaboration with Technical Committee CEN/TC 123 "Lasers
and photonics", the secretariat of which is held by DIN.
This European Standard shall be given the status of a national standard, either by publication of
an identical text or by endorsement, at the latest by November 2007, and conflicting national
standards shall be withdrawn at the latest by November 2007.
This document has been prepared under a mandate given to CEN by the European Commission
and the European Free Trade Association, and supports essential requirements of EU
Directive(s).
For relationship with EU Directive(s), see informative Annex ZA, which is an integral part of this
document.
According to the CEN/CENELEC Internal Regulations, the national standards organizations of
the following countries are bound to implement this European Standard: Austria, Belgium,
Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece,
Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United
Kingdom.
Endorsement notice
The text of ISO 11810-2:2007 has been approved by CEN as EN ISO 11810-2:2007 without any
modifications.
Annex ZA
(informative)
Relationship between this European Standard and the Essential
Requirements of EU Directive 2006/42/EC
This International Standard has been prepared under a mandate given to CEN by the European
Commission and the European Free Trade Association to provide one means of conforming to
Essential Requirements of the New Approach Directive 2006/42/EC.
Once this standard is cited in the Official Journal of the European Communities under that
Directive and has been implemented as a national standard in at least one Member State,
compliance with the normative clauses of this standard given in Table ZA.1 confers, within the
limits of the scope of this standard, a presumption of conformity with the corresponding Essential
Requirements of that Directive and associated EFTA regulations.
Table ZA.1 — Correspondence between this International Standard and EU Directive
2006/42/EC
Clause(s)/sub-clause(s) of this Essential requirements (ERs) of Qualifying remarks/Notes
International Standard EU Directive 2006/42/EC
1.5.10 Radiation
1.5.12 Laser radiation
WARNING: Other requirements and other EU Directives may be applicable to the product(s)
falling within the scope of this standard.
INTERNATIONAL ISO
STANDARD 11810-2
First edition
2007-05-01
Lasers and laser-related equipment —
Test method and classification for the
laser-resistance of surgical drapes and/or
patient-protective covers —
Part 2:
Secondary ignition
Lasers et équipements associés aux lasers — Méthode d'essai et
classification de la résistance au laser pour des draps chirurgicaux et/ou
des couvertures de protection des patients —
Partie 2: Inflammation secondaire
Reference number
ISO 11810-2:2007(E)
©
ISO 2007
ISO 11810-2:2007(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
© ISO 2007
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2007 – All rights reserved
ISO 11810-2:2007(E)
Contents Page
Foreword. iv
Introduction . v
1 Scope. 1
2 Normative references. 1
3 Terms and definitions. 2
4 Test methods. 2
4.1 General conditions . 2
4.1.1 Sampling. 2
4.1.2 Test equipment. 2
4.2 Testing procedure. 8
4.2.1 Sequence of testing. 8
4.2.2 Specimen preparation . 8
4.2.3 Laser-induced secondary ignition . 8
5 Classification. 9
5.1 Laser-induced secondary ignition of test material .9
5.2 Classification definition . 9
6 Test Report. 10
Bibliography . 11
ISO 11810-2:2007(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 11810-2 was prepared by Technical Committee ISO/TC 172, Optics and photonics, Subcommittee SC 9,
Electro-optical systems.
ISO 11810 consists of the following parts, under the general title Lasers and laser-related equipment — Test
method and classification for the laser-resistance of surgical drapes and/or patient-protective covers:
⎯ Part 1: Primary ignition and penetration
⎯ Part 2: Secondary ignition
For the purposes of this part of ISO 11810 the CEN annex regarding fulfilment of European Council Directives
will be removed at publication stage.
iv © ISO 2007 – All rights reserved
ISO 11810-2:2007(E)
Introduction
Some laser applications in medicine may require laser-resistant surgical drapes or other patient protective
covers. Surgical drapes or other patient protective covers are necessary when a sterile procedure is
performed and the surrounding area needs to be protected from liquids, secretions and inadvertent laser
radiation. While conventional surgical drapes or other patient protective covers are not necessarily laser-
resistant, specifically designed drapes offer the possibility of laser resistance.
Laser-induced risks include ignition, inflammability, melting, penetration, thermal transfer and reflectivity.
Textile and non-woven drape materials may have other risks but they may provide a laser barrier. While there
are many potential ignition devices present in the operating room, e.g. fibre optic illumination systems, electro-
surgical units, hot wire cauteries, etc., this test method addresses only the laser ignition source. A surgical
drape or other patient protective cover that claims to be laser-resistant must be tested according to this part of
ISO 11810.
CO lasers may provide the most challenging conditions of all medical lasers. Ignition/inflammability tests and
penetration tests may disclose more challenging laser wavelengths as well as modes of laser delivery, for
example Q-switching in the nanosecond range. The 20 W CO laser (continuous wave) has been selected as
the laser to be used for this part of ISO 11810. For laser-induced secondary ignition of drapes and
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.