IEC 60601-2-43:2010/AMD1:2017
(Main)Amendment 1 - Medical electrical equipment - Part 2-43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures
Amendment 1 - Medical electrical equipment - Part 2-43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures
Amendement 1 - Appareils électromédicaux - Partie 2-43: Exigences particulières pour la sécurité de base et les performances essentielles des appareils à rayonnement X lors d'interventions
General Information
Buy Standard
Standards Content (Sample)
IEC 60601-2-43
Edition 2.0 2017-05
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
A MENDMENT 1
AM ENDEMENT 1
Medical electrical equipment –
Part 2-43: Particular requirements for the basic safety and essential performance
of X-ray equipment for interventional proceduresAppareils électromédicaux –
Partie 2-43: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X lors d’interventionsIEC 60601-2-43:2010-03/AMD1:2017-05(en-fr)
---------------------- Page: 1 ----------------------
THIS PUBLICATION IS COPYRIGHT PROTECTED
Copyright © 2017 IEC, Geneva, Switzerland
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.
IEC Central Office Tel.: +41 22 919 02 113, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.IEC Catalogue - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
The stand-alone application for consulting the entire The world's leading online dictionary of electronic and
bibliographical information on IEC International Standards, electrical terms containing 20 000 terms and definitions in
Technical Specifications, Technical Reports and other English and French, with equivalent terms in 16 additional
documents. Available for PC, Mac OS, Android Tablets and languages. Also known as the International Electrotechnical
iPad. Vocabulary (IEV) online.IEC publications search - www.iec.ch/searchpub IEC Glossary - std.iec.ch/glossary
The advanced search enables to find IEC publications by a 65 000 electrotechnical terminology entries in English and
variety of criteria (reference number, text, technical French extracted from the Terms and Definitions clause of
committee,…). It also gives information on projects, replaced IEC publications issued since 2002. Some entries have been
and withdrawn publications. collected from earlier publications of IEC TC 37, 77, 86 and
CISPR.IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published IEC Customer Service Centre - webstore.iec.ch/csc
details all new publications released. Available online and If you wish to give us your feedback on this publication or
also once a month by email. need further assistance, please contact the Customer Service
Centre: csc@iec.ch.A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications IECLe contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.Catalogue IEC - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
Application autonome pour consulter tous les renseignements Le premier dictionnaire en ligne de termes électroniques et
bibliographiques sur les Normes internationales, électriques. Il contient 20 000 termes et définitions en anglais
Spécifications techniques, Rapports techniques et autres et en français, ainsi que les termes équivalents dans 16
documents de l'IEC. Disponible pour PC, Mac OS, tablettes langues additionnelles. Egalement appelé Vocabulaire
Android et iPad. Electrotechnique International (IEV) en ligne.Recherche de publications IEC - www.iec.ch/searchpub Glossaire IEC - std.iec.ch/glossary
La recherche avancée permet de trouver des publications IEC 65 000 entrées terminologiques électrotechniques, en anglais
en utilisant différents critères (numéro de référence, texte, et en français, extraites des articles Termes et Définitions des
comité d’études,…). Elle donne aussi des informations sur les publications IEC parues depuis 2002. Plus certaines entrées
projets et les publications remplacées ou retirées. antérieures extraites des publications des CE 37, 77, 86 et
CISPR de l'IEC.IEC Just Published - webstore.iec.ch/justpublished
Restez informé sur les nouvelles publications IEC. Just Service Clients - webstore.iec.ch/csc
Published détaille les nouvelles publications parues. Si vous désirez nous donner des commentaires sur cette
Disponible en ligne et aussi une fois par mois par email. publication ou si vous avez des questions contactez-nous:
csc@iec.ch.---------------------- Page: 2 ----------------------
IEC 60601-2-43
Edition 2.0 2017-05
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
A MENDMENT 1
AM ENDEMENT 1
Medical electrical equipment –
Part 2-43: Particular requirements for the basic safety and essential performance
of X-ray equipment for interventional proceduresAppareils électromédicaux –
Partie 2-43: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils à rayonnement X lors d’interventionsINTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.50; 37.040.25 ISBN 978-2-8322-4318-3
Warning! Make sure that you obtained this publication from an authorized distributor.
Attention! Veuillez vous assurer que vous avez obtenu cette publication via un distributeur agréé.
® Registered trademark of the International Electrotechnical CommissionMarque déposée de la Commission Electrotechnique Internationale
---------------------- Page: 3 ----------------------
– 2 – IEC 60601-2-43:2010/AMD1:2017
IEC 2017
FOREWORD
This amendment has been prepared by subcommittee 62B: Diagnostic imaging equipment, of
IEC technical committee 62: Electrical equipment in medical practice.The text of this amendment is based on the following documents:
CDV Report on voting
62B/1012/CDV 62B/1037/RVC
Full information on the voting for the approval of this amendment can be found in the report
on voting indicated in the above table.The committee has decided that the contents of this amendment and the base publication will
remain unchanged until the stability date indicated on the IEC website under"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
_____________
INTRODUCTION to the Amendment
The purpose of this first amendment to IEC 60601-2-43:2010 is to introduce changes as
follows:– refer to IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012 and its applicable collateral
standards;– refer to IEC 60601-2-54:2009 and IEC 60601-2-54:2009/AMD1:2015 and consequent
subclause adaptations;
– include a requirement to have a maximum time of 10 min to recover all functions after a
recoverable failure in 201.4.101;– include several aspects from IEC 61910-1:2014 and remove the reference to
IEC PAS 61910-1:2007 in 201.4.102;
– include an alternative way of testing in 201.11.6.5.103;
– include a clarification for tableside controls in 201.12.4.106.
In addition, a number of technical errors have been corrected.
_____________
201.1 Scope, object and related standards
Replace the text of the existing footnote by the following:
---------------------- Page: 4 ----------------------
IEC 60601-2-43:2010/AMD1:2017 – 3 –
IEC 2017
1) The general standard is IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, Medical electrical equipment –
Part 1: General requirements for basic safety and essential performance.201.1.3 Collateral standards
Replace the existing second sentence of the second paragraph by the following:
1) 2) 3)
IEC 60601-1-8, IEC 60601-1-10 , IEC 60601-1-11 and IEC 60601-1-12 do not apply.
201.2 Normative references
Replace, under "Amendment", the existing references to IEC 60601-1-2 and to IEC 60601-1-3
as follows:IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral standard: Electromagnetic disturbances –
Requirements and testsIEC 60601-1-3:2008, Medical electrical equipment – Part 1-3: General requirements for basic
safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-
ray equipmentIEC 60601-1-3:2008/AMD1:2013
Add, under "Addition", the following new references:
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performanceIEC 60601-1:2005/AMD1:2012
IEC 61910-1:2014, Medical electrical equipment – Radiation dose documentation – Part 1:
Radiation dose structured reports for radiography and radioscopyIEC 62220-1-1:2015, Medical electrical equipment – Characteristics of digital X-ray imaging
devices – Part 1-1: Determination of the detective quantum efficiency – Detectors used in
radiographic imagingReplace, under "Addition", the existing references to IEC 60601-2-54 and to IEC 60788 as
follows:IEC 60601-2-54:2009, Medical electrical equipment – Part 2-54: Particular requirements for
the basic safety and essential performance of X-ray equipment for radiography and
radioscopyIEC 60601-2-54:2009/AMD1:2015
IEC TR 60788:2004, Medical electrical equipment – Glossary of defined terms
___________
1) IEC 60601-1-10, Medical electrical equipment – Part 1-10: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for the development of physiologic closed-loop controllers
IEC 60601-1-11, Medical electrical equipment – Part 1-11: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical
systems used in the home healthcare environment3) IEC 60601-1-12, Medical electrical equipment – Part 1-12: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical
systems intended for use in the emergency medical services environment---------------------- Page: 5 ----------------------
– 4 – IEC 60601-2-43:2010/AMD1:2017
IEC 2017
201.3 Terms and definitions
Replace the introductory paragraph by the following new paragraph:
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012, IEC 60601-1-3:2008, IEC 60601-2-54:2009,IEC TR 60788:2004, IEC 61910-1:2014, IEC 62220-1-1:2015 and the following apply.
Add, after 201.3.203, the following new term and definition:
201.3.204
EMERGENCY RADIOSCOPY
RADIOSCOPY with availability of a limited set of functions (emergency functions), for use during
recovery from a recoverable failure of the INTERVENTIONAL X-RAY EQUIPMENT201.4.3 ESSENTIAL PERFORMANCE
Replace, in the first sentence, “IEC 60601-2-54:2009” by “IEC 60601-2-54:2009 and
IEC 60601-2-54:2009/AMD1:2015”.Delete the first sentence of the addition (sentence starting with ‘The list in Table…’).
201.4.101 * Recovery managementReplace the existing text of this subclause by the following new text:
The time to recover all of the functions necessary for performing EMERGENCY RADIOSCOPY,
after a failure recoverable automatically or by the OPERATOR shall be as short as reasonably
practicable. The RISK MANAGEMENT shall take into account the availability of emergency power
supply in the determination of the recovery time.When the recovery is complete, a reinitiation of IRRADIATION shall be required to produce
further IRRADIATION.The time to recover all functions, after a failure recoverable automatically or by the operator,
shall be as short as reasonably practicable.In case of a manually recoverable failure, the time to recover all functions shall not exceed
10 min from the time the OPERATOR has initiated the recovery to the time the INTERVENTIONAL
X-RAY EQUIPMENT has all functions available.In case of an automatically detected and automatically recoverable failure, the time to recover
all functions shall not exceed 10 min from the time of the failure of the INTERVENTIONAL X-RAY
to the time the INTERVENTIONAL X-RAY EQUIPMENT has all functions available.EQUIPMENT
INTERVENTIONAL X-RAY EQUIPMENT may have both recovery modes.
NOTE Less than 1 min is a desirable value for the time to recover all functions for performing EMERGENCY
RADIOSCOPY. Less than 3 min is a desirable value to recover all functions.The instructions for use shall indicate:
– the time necessary to get all functions for EMERGENCY RADIOSCOPY operable;
– the time to restore all functions of the INTERVENTIONAL X-RAY EQUIPMENT;
– for failures recoverable by the OPERATOR, the required procedure which the OPERATOR
must follow to perform this recovery.---------------------- Page: 6 ----------------------
IEC 60601-2-43:2010/AMD1:2017 – 5 –
IEC 2017
When the system is in the EMERGENCY RADIOSCOPY mode, this mode shall be indicated at the
working position of the OPERATOR.The functions necessary for performing EMERGENCY RADIOSCOPY shall include, at minimum:
– RADIOSCOPY MODE OF OPERATION, in priority order:• RADIOSCOPY in the MODE OF OPERATION that was used at the time of the recoverable
equipment failure;• or, if this is not possible, RADIOSCOPY in the MODE OF OPERATION as close as possible to
the one which was used at the time of the recoverable equipment failure;– normal operation of the PATIENT SUPPORT;
– normal operation of the gantry;
– normal operation of tableside controls for all functions described above;
– normal operation of the irradiation disabling switch (see 203.6.103);
– normal operation of the motion disabling switch (see 201.9.2.3.1 in IEC 60601-2-54:2009
and IEC 60601-2-54:2009/AMD1:2015);– normal operation of anti-collision functions (see 201.9.2.4).
Compliance is checked by inspection of the risk management file and by functional tests.
201.4.102 * RADIATION dose documentationReplace the existing text of this subclause by the following new text:
The INTERVENTIONAL X-RAY EQUIPMENT shall create RADIATION DOSE STRUCTURED REPORTS
(RDSR) and shall have the ability to perform RDSR END OF PROCEDURE TRANSMISSION.The RDSR shall contain the data elements that are required (‘shall’) in 5.1.2 and 5.1.3 of
IEC 61910-1:2014.The RDSR should contain the data elements that are recommended (‘should’) in 5.1.2 and
5.1.3 of IEC 61910-1:2014.NOTE The conditional statements associated with the data elements in IEC 61910-1:2014 are considered to be
part of these data elements.If the INTERVENTIONAL X-RAY EQUIPMENT does not have means to determine GANTRY
angulations, the RDSR need not contain the data elements related to positioner angles.
Compliance is checked by functional test.201.7.2.101 Beam limiting device
Replace the reference to “IEC 60601-2-54:2009” by “IEC 60601-2-54:2009 and IEC 60601-2-
54:2009/AMD1:2015”.201.7.2.105 * Protection against ingress of liquids
Add, after the existing Note 2, the following new note:
NOTE 3 Parts that are IPX0 need not be marked.
201.7.9.2.12 * Cleaning, disinfection and sterilization
Add, after the existing note, the following new subclause:
---------------------- Page: 7 ----------------------
– 6 – IEC 60601-2-43:2010/AMD1:2017
IEC 2017
201.7.9.2.17 ME EQUIPMENT emitting radiation
Subclause 201.7.9.2.17 of IEC 60601-2-54:2009 and IEC 60601-2-54:2009/AMD1:2015
applies.
201.9.2 HAZARDS associated with moving parts
Replace the existing title of this subclause by the following:
201.9.2 MECHANICAL HAZARDS associated with moving parts
Add, before the existing 201.9.2.2.4, the following new title:
201.9.2.2 TRAPPING ZONE
201.9.2.2.4 GUARDS and protective measures
Replace the existing title and text of this subclause by the following:
201.9.2.2.4 GUARDS and other RISK CONTROL measures
Subclause 201.9.2.2.4 of IEC 60601-2-54:2009 and IEC 60601-2-54:2009/AMD1:2015 applies.
201.9.2.3 Other HAZARDS associated with moving partsReplace the existing title and text of this subclause by the following:
201.9.2.3 Other MECHANICAL HAZARDS associated with moving parts
Subclause 201.9.2.3 of IEC 60601-2-54:2009 and IEC 60601-2-54:2009/AMD1:2015 applies.
201.11.6.5.101 FootswitchesReplace, in the second paragraph, “a saline solution of at least 0.9 %” by “a saline solution of
at least 0,9 % weight to volume of sodium chloride in water”.201.11.6.5.103 * ENCLOSURES
Replace the third dash by the following new dash:
– PATIENT SUPPORT should have a minimum degree of protection of IPX2 or should be
protected against spraying water at any angle up to 15° from the vertical.Renumber the existing Note as NOTE 1, and add, after this note, the following new note:
NOTE 2 For the PATIENT SUPPORT test, the PATIENT SUPPORT can be placed on a 15° slope.
201.11.101 Protection against excessive temperatures of X-RAY TUBE ASSEMBLIESReplace the reference to “IEC 60601-2-54:2009” by “IEC 60601-2-54:2009 and IEC 60601-2-
54:2009/AMD1:2015”.201.12 Accuracy of controls and instruments and protection against
hazardous output
201.12.4 * Protection against hazardous output
Add, after the title of this subclause, the following new subclause:
---------------------- Page: 8 ----------------------
IEC 60601-2-43:2010/AMD1:2017 – 7 –
IEC 2017
201.12.4.5.2 Diagnostic X-ray equipment
Replacement:
INTERVENTIONAL X-RAY EQUIPMENT shall comply with IEC 60601-1-3:2008 and
IEC 60601-1-3:2008/AMD1:2013 as modified by this particular standard.
Compliance is checked as specified in IEC 60601-1-3:2008 and
IEC 60601-1-3:2008/AMD1:2013 as modified by this particular standard.
201.12.4.106 * Tableside controls
Replace the existing text of this subclause by the following new text:
For tableside controls, as a minimum, the following user interface controls, requiring operation
by touch, shall be individually and unambiguously identifiable both by touch alone and also by
sight alone:– GANTRY and PATIENT SUPPORT motions controls (not including motion controls for
preselecting INTERVENTIONAL X-RAY EQUIPMENT positions);
– IRRADIATION SWITCHES (other than footswitches);
– collimation blade control (not including WEDGE FILTER control).
Collimation blade control may additionally be operated by a duplicated tableside control, such
as a touchscreen user interface.All tableside controls shall be identifiable under the lighting conditions for the INTENDED USE,
and if applicable, when covered by transparent protective drapes.Compliance is checked by inspection and by functional tests.
NOTE A tableside control is a control that can be operated adjacent to the PATIENT during a procedure regardless
of whether or not it is physically attached to the PATIENT SUPPORT. A footswitch is not a tableside control for the
purposes of this subclause.202 Electromagnetic compatibility – Requirements and tests
Replace the existing clause title by the following:
202 Electromagnetic disturbances – Requirements and tests
Replace “IEC 60601-1-2:2007” by “IEC 60601-1-2:2014”.
202.101 Immunity testing of ESSENTIAL PERFORMANCE
Replace the existing text of this subclause by the following sentence:
Subclause 202.101 of IEC 60601-2-54:2009 and IEC 60601-2-54:2009/AMD1:2015 applies.
203 Radiation protection in diagnostic X-ray equipmentReplace, in the first paragraph, “IEC 60601-1-3:2008 applies” by “IEC 60601-1-3:2008 and
IEC 60601-1-3:2008/AMD1:2013 apply”.---------------------- Page: 9 ----------------------
– 8 – IEC 60601-2-43:2010/AMD1:2017
IEC 2017
203.4.1 Statement of compliance
Replace the existing text after "Replacement" and before "Additional subclause" by the
following:If, for INTERVENTIONAL X-RAY EQUIPMENT, compliance with this standard is to be stated, the
statement shall be made with the following content:– model or type reference;
– IEC 60601-2-43:2010.
203.5.2.4.5.101 * Dosimetric information for X-RAY EQUIPMENT specified for RADIOSCOPY
and/or SERIAL RADIOGRAPHYIn item d), last dash, replace “shall by specified” by “shall be specified”.
203.6.4.5 * Dosimetric indications
Replace, in the note, “IEC 60601-2-54:2009” by “IEC 60601-2-54:2009 and
IEC 60601-2-54:2009/AMD1:2015”.
Replace, in the third paragraph, the second, third and fourth dashes by the following:
– The value of the cumulative REFERENCE AIR KERMA resulting from RADIOSCOPY andRADIOGRAPHY since the last reset operation shall be
• continuously displayed at the working position of the OPERATOR in mGy together with
this unit and updated at least once every 5 s; or• displayed not later than 5 s after the interruption or termination of LOADING.
– The values for the REFERENCE AIR KERMA RATE and the cumulative REFERENCE AIR KERMA
shall be clearly distinguishable from each other.The correction to the fifth paragraph starting with "The instructions…" applies to the French
text only.203.7 RADIATION QUALITY
Replace the reference to “IEC 60601-2-54:2009” by “IEC 60601-2-54:2009 and IEC 60601-2-
54:2009/AMD1:2015”.Annex AA
Subclause 201.4.101 – Recovery management
Delete the third and fourth paragraphs (both starting with “Less than …”).
Bibliography
Delete the reference [23] to DICOM and the reference [24] to IEC/PAS 61910-1:2007.
Index of defined terms used in this particular standardReplace, in the existing list, all references to "IEC 60788:2004" by references to
"IEC TR 60788:2004".Replace the following references as follows:
---------------------- Page: 10 ----------------------
IEC 60601-2-43:2010/AMD1:2017 – 9 –
IEC 2017
CLEARLY LEGIBLE ........................... IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.15
ESSENTIAL PERFORMANCE ............... IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.27
HARM ............................................ IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.38
HAZARD ......................................... IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.39
HAZARDOUS SITUATION ................... IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.40
INTENDED USE ................................ IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.44
MANUFACTURER ............................. IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.55
NORMAL USE .................................. IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.71
ORIGINAL DATA .................................................................................. IEC 62220-1-1:2015, 3.13
PATIENT ......................................... IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.76
PROCEDURE ................................... IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.88
PROCESS ....................................... IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.89
RISK ............................................ IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.102
RISK MANAGEMENT ....................... IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.107
RISK MANAGEMENT FILE ................. IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.108
SEVERITY ..................................... IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.114
SINGLE FAULT CONDITION .............. IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, 3.116
Add, to the existing list, the following new references:EMERGENCY RADIOSCOPY ........................................................................................... 201.3.204
GUARD ................................................................................................. IEC 60601-1:2005, 3.36
MECHANICAL HAZARD ............................................................................. IEC 60601-1:2005, 3.61
RADIATION DOSE STRUCTURED REPORT (RDSR)........................................... IEC 61910-1:2014, 3.3
RDSR END OF PROCEDURE TRANSMISSION .................................................. IEC 61910-1:2014, 3.5
RISK CONTROL ................................. IEC 60601-1:2005 et IEC 60601-1:2005/AMD1:2012, 3.105
TRAPPING ZONE ................................................................................... IEC 60601-1:2005, 3.131
___________---------------------- Page: 11 ----------------------
– 10 – IEC 60601-2-43:2010/AMD1:2017
IEC 2017
AVANT-PROPOS
Le présent amendement a été établi par le sous-comité 62B: Appareils d'imagerie de
diagnostic, du comité d’études 62 de l'IEC: Équipements électriques dans la pratique
médicale.Le texte de cet amendement est issu des documents suivants:
CDV Rapport de vote
62B/1012/CDV 62B/1037/RVC
Le rapport de vote indiqué dans le tableau ci-dessus donne toute information sur le vote ayant
abouti à l'approbation de cet amendement.Le comité a décidé que le contenu de cet amendement et de la publication de base ne sera
pas modifié avant la date de stabilité indiquée sur le site web de l'IEC sous"http://webstore.iec.ch" dans les données relatives à la publication recherchée. A cette date,
la publication sera• reconduite,
• supprimée,
• remplacée par une édition révisée, ou
• amendée.
_____________
INTRODUCTION à l'amendement
Ce premier amendement à l’IEC 60601-2-43:2010 a pour but d’introduire les modifications
suivantes:– référence à l’IEC 60601-1:2005 et à à l’IEC 60601-1:2005/AMD1:2012 et ses normes
collatérales applicables;– référence à l’IEC 60601-2-54:2009 et à l’IEC 60601-2-54:2009/AMD1:2015 et adaptations
des articles en conséquence;– intégration d’une exigence relative à un temps maximal de 10 min nécessaire à la reprise
de toutes les fonctions après une défaillance qui peut être traitée, au 201.4.101;
– ajout de plusieurs aspects tirés de l’IEC 61910-1:2014 et suppression de la référence à
l’IEC PAS 61910-1:2007 au 201.4.102;– intégration d’une méthode alternative d’essai au 201.11.6.5.103;
– ajout d’une clarification relative aux commandes côté table au 201.12.4.106.
De plus, un certain nombre d’erreurs techniques ont été corrigées.
_____________
201.1 Domaine d’application, objet et normes connexes
Remplacer le texte de la note de bas de page existante par la note suivante:
---------------------- Page: 12 ----------------------
IEC 60601-2-43:2010/AMD1:2017 – 11 –
IEC 2017
1) La norme générale est l’IEC 60601-1:2005 et l'IEC 60601-1:2005/AMD1:2012, Appareils électromédicaux –
Partie 1: Exigences générales pour la sécurité de base et les performances essentielles
201.1.3 Normes collatéralesRemplacer la deuxième phrase existante du deuxième alinéa par le texte suivant:
1) ) )
2 3
L'IEC 60601-1-8, l'IEC 60601-1-10 , l'IEC 60601-1-11 et l'IEC 60601-1-12 ne s’appliquent
pas.201.2 Références normatives
Remplacer, sous "Amendement", les références existantes à l'IEC 60601-1-2 et à
l'IEC 60601-1-3 par les références suivantes:
IEC 60601-1-2:2014, Appareils électromédicaux – Partie 1-2: Exigences générales pour la
sécurité de base et les performances essentielles – Norme collatérale: Perturbations
électromagnétiques – Exigences et essaisIEC 60601-1-3:2008, Appareils électromédicaux – Partie 1-3: Exigences générales pour la
sécurité de base et les performances essentielles – Norme collatérale: Radioprotection dans
les appareils à rayonnement X de diagnosticIEC 60601-1-3:2008/AMD1:2013
Ajouter, sous "Addition", les nouvelles références suivantes:
IEC 60601-1:2005, Appareils électromédicaux – Partie 1: Exigences générales pour la
sécurité de base et les performances essentiellesIEC 60601-1:2005/AMD1:2012
IEC 61910-1:2014, Appareils électromédicaux – Documentation sur la dose de rayonnement –
Partie 1: Rapports structurés sur la dose de rayonnement pour la radiographie et la
radioscopieIEC 62220-1-1:2015, Appareils électromédicaux – Caractéristiques des dispositifs d'imagerie
à rayonnement X – Partie 1-1: Détermination de l'efficacité quantique de détection –
Détecteurs utilisés en imagerie radiographiqueRemplacer, sous "Addition", les références existantes à l'IEC 60601-2-54 et à l'IEC 60788 par
les références suivantes:IEC 60601-2-54:2009, Appareils électromédicaux – Partie 2-54: Exigences particulières pour
la sécurité de base et les performances essentielles des appareils à rayonnement X utilisés
pour la radiographie et la radioscopieIEC 60601-2-54:2009/AMD1:2015
IEC T
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.