Abstract

IEC 61514:2026 specifies tests designed to determine the static and dynamic performance of single-acting or double-acting analogue positioners. The tests apply to positioners which receive standard analogue input signals (as specified in IEC 60381-1, IEC 60381-2 and IEC 60382) and have a pneumatic output. Positioners with pulsed or digital input signals, positioners with digital controllers and positioners with pulsed outputs are outside the scope of this document. Testing is conducted either on a positioner alone, independently of an actuator, or on a positioner mounted and connected to a specific actuator, as a combined unit. The text makes clear where different approaches are required. The methods of evaluation given in this document are intended for use by manufacturers to determine the performance of their products, and by users, or independent testing establishments, to verify manufacturers' performance specifications. The closest liaison between the evaluating body and the manufacturer is indispensable during the tests, including the possibility for the manufacturer to influence the test programme based on the manufacturer's specifications for the instrument and comment on both the test programme and the results. This document is intended to provide definitions of positioner elements, actions, and characteristics, to specify uniform methods of measuring performance errors and effects of influence quantities on those characteristics, and to describe methods of reporting and evaluating the results of the measurement data obtained. The test conditions described in this publication (for example range of ambient temperatures and power supply) relate to conditions which commonly arise in use. Consequently, the values specified are used where no other values are specified by the manufacturer or user. If other values are used, they will be stated. It is recognized that the manufacturer's specifications and instructions for installation and operation apply during all steps. The tests specified in this document are not necessarily sufficient for instruments specifically designed for unusually arduous conditions. Conversely, a reduced series of tests can serve adequately for instruments designed to perform within a more limited range of conditions. When a full evaluation, in accordance with this document, is not required or possible, only the tests which are required are performed and the results reported in accordance with the relevant parts of this document. In such cases, the test report will state that it does not cover the full number of tests specified herein. This second edition cancels and replaces the first edition published in 2000. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) in 6.6.8 and Table 5, the magnetic field has been changed from 100 A/m to of 30 A/m (Mean Root Square);
b) 6.10.4 and Figure 9 have been modified for better understandability;
c) in 7.4, the reference to IEC 61187 has been deleted and replaced with a new Table 4: Document information.

Status
Published
Publication Date
08-Jun-2026
Current Stage
PPUB - Publication issued - 60 Publication
Start Date
09-Jun-2026
Due Date
26-Jun-2026
Completion Date
26-Jun-2026

Buy Documents

Standard

REDLINE IEC 61514:2026 RLV - Industrial-process control systems - Methods of evaluating the performance of valve positioners with pneumatic outputs

ISBN:978-2-8327-1324-2
Release Date:09-Jun-2026
English language (92 pages)
sale 15% off
Preview
sale 15% off
Preview
Standard

iec61514{ed2.0}en - Industrial-process control systems - Methods of evaluating the performance of valve positioners with pneumatic outputs

ISBN:978-2-8327-1323-5
Release Date:09-Jun-2026
English and French language (92 pages)
sale 15% off
Preview
sale 15% off
Preview
Standard

iec61514{ed2.0}fr - Systemes de commande des processus industriels - Méthodes d'évaluation des performances des positionneurs de vannes a sorties pneumatiques

ISBN:978-2-8327-1323-5
Release Date:09-Jun-2026
English and French language (92 pages)
sale 15% off
Preview
sale 15% off
Preview

IEC 61514:2026 is an IEC International Standard from the International Electrotechnical Commission for industrial-process control systems. It specifies methods for evaluating the static and dynamic performance of valve positioners with pneumatic outputs.

IEC 61514:2026 is written for manufacturers, users, and independent testing establishments that need to measure performance, verify stated specifications, and report results in a uniform way. It also defines the key positioner terms, test conditions, and test report content used in practice.

What does IEC 61514:2026 specify?

IEC 61514:2026 specifies test methods for single-acting and double-acting analogue positioners with pneumatic output. The document focuses on positioners that receive standard analogue input signals and it leaves out positioners with pulsed or digital inputs, digital controllers, and pulsed outputs.

IEC 61514:2026 is organized into 8 clauses:

  • Clause 1 - scope and application
  • Clause 2 - normative references
  • Clause 3 - terms and definitions
  • Clause 4 - general test conditions
  • Clause 5 - general test procedures
  • Clause 6 - test procedures
  • Clause 7 - other factors to consider
  • Clause 8 - test report and documentation

The standard covers both bench testing of the positioner alone and combined testing of a positioner mounted on a specific actuator. That matters because the first method isolates the positioner, while the second shows how the assembled unit behaves in service.

What are the key requirements of IEC 61514:2026?

IEC 61514:2026 requires tests to be run under controlled ambient and supply conditions, with manufacturer instructions taking precedence when they give different values. In practice, this means the test setup, stabilization time, and supply quality have to be controlled before results can be compared.

Setting up the test correctly

Clause 5 requires the device to stabilize before readings are taken, and it calls for repeated traverses of the operating range before the measurements are recorded. The measurement points should be spread across the full range, so the result reflects the whole operating band and not just one favorable point.

Clause 5 also separates positioner-only tests from positioner-plus-actuator tests. That distinction matters because the combined unit can be affected by actuator friction, inertia, spring action, and load, while the separate test is the way to compare positioners on their own.

Measuring the main performance characteristics

Clause 6 measures gain, course characteristic, precision terms, dead band, air-flow data, and dynamic response. In practice, this gives a buyer or engineer a full picture of how well the positioner follows its input and how stable and repeatable that behavior is.

Clause 6 defines how to measure gain and the proportional gain factor K, including checks for direct and reverse action and for the effect of supply pressure. That matters because gain changes can affect controllability and tuning in the installed loop.

Clause 6 also requires evaluation of the course characteristic and the main error terms - measured error, conformity or linearity error, hysteresis, and repeatability error. These results show whether the positioner tracks the intended position consistently in both directions.

Checking influence quantities and robustness

Clause 6 requires testing the effect of influence quantities such as supply pressure, ambient temperature, humidity, mounting position, shock, vibration, magnetic field, electromagnetic radiation, electrical fast transients, surge, and series-mode interference. For the user, this shows how much the positioner output or course can shift when the environment changes.

The document also includes an input-overrange test, start-up drift, long-term drift, and accelerated life testing. These checks matter because they reveal whether the positioner stays stable after overload, after power-up, and over extended operation.

Evaluating dynamic response and documenting the result

Clause 6 includes both frequency-response testing and step-response testing. These methods are important when the positioner has to follow fast control signals or when the combined positioner/actuator response is critical to process stability.

Clause 8 requires a full test report in the format of IEC 61298-4, together with the test results listed in the document. In practice, the report has to show what was tested, how it was mounted, what actuator or valve was used, and which tests were omitted if the evaluation was partial.

What terms does IEC 61514:2026 define?

  • Positioner - A control device connected to the moving part of the final control element or its actuator that adjusts output signal Y to maintain the desired course signal X.
  • Single-acting positioner - A positioner with one output signal Y acting on one side of the actuator.
  • Double-acting positioner - A positioner with two output signals Y and Y acting on opposite sides of the actuator.
  • Signal d'entrée W - The reference input signal that represents the desired position of the associated control element.
  • Signal de course X - The signal produced by the movement of the final control element or its actuator.
  • Signal de sortie Y - The air pressure supplied to the actuator of the final control element.
  • Action directe / action inversée - Direct action means Y rises when W rises, while reverse action means Y falls when W rises.
  • Fractionnement de la plage - A special setting that uses only part of the input range to obtain full actuator travel.
  • Facteur de gain proportionnel K - The gain over the full actuator range.
  • Zone d'insensibilité - The range of input or course change that produces no detectable output change.

Who uses IEC 61514:2026?

IEC 61514:2026 is used by positioner manufacturers, industrial users, and independent test laboratories. Manufacturers use it to characterize their products and support performance claims, while users and test houses use it to verify those claims in a consistent way.

It is relevant in industrial process control and valve automation, especially where pneumatic positioners are part of a control loop. Quality teams and buyers can use it to compare devices on the same test basis and to check that test reports include the information needed for acceptance or audit.

What changed in IEC 61514:2026 from the previous edition?

IEC 61514:2026 is the second edition and it cancels and replaces the 2000 edition as a technical revision. The stated technical changes are:

  • the magnetic-field test severity in Clause 6.6.8 and Table 5 was reduced
  • Clause 6.10.4 and Figure 9 were revised for clarity
  • Clause 7.4 now uses a new document-information table instead of the former reference to IEC 61187

Which standards are used with IEC 61514:2026?

IEC 61514:2026 relies on several referenced standards for signals, test conditions, EMC, safety, and reporting.

StandardWhat it contributes
IEC 60050-311Vocabulary for electrical measurements
IEC 60050-351Vocabulary for control and regulation technology
IEC 60068-2-1, IEC 60068-2-2, IEC 60068-2-6, IEC 60068-2-31, IEC 60068-2-78Environmental test methods for cold, dry heat, vibration, shock, and damp heat
IEC 60381-1, IEC 60381-2, IEC 60382Standard analogue input signals used by the positioners in scope
IEC 60529Degree of protection provided by enclosures
IEC 60654 and IEC 60721-3Operating and environmental condition classifications
IEC 61010-1Safety requirements for electrical measuring and control equipment
IEC 61032Test probes for verifying protection by enclosures
IEC 61298-4Required structure of the performance-evaluation report
IEC 61326-1EMC test requirements used for immunity checks

What does the IEC 61514:2026 document contain?

IEC 61514:2026 contains test conditions, general procedures, specific measurement methods, and reporting rules, supported by figures and tables. The tables cover environmental conditions, typical error terms, vibration levels, document information, and the results that must appear in the report.

The figures show typical positioner/actuator arrangements, gain and error curves, the air-flow test setup, a dynamic-response setup, and example frequency- and step-response plots. These elements are used for performance testing, verification, and reporting rather than for installation instructions alone.

Clause 6 provides the core measurement procedures for gain, course characteristic, dead band, air flow, influence quantities, overrange, drift, life testing, and dynamic response. Clause 7 then adds checks on safety, enclosure protection, installation, maintenance, repair, protective treatments, design features, variants, and required tools.

Relations

Effective Date
05-Sep-2023

Buy Documents

Standard

REDLINE IEC 61514:2026 RLV - Industrial-process control systems - Methods of evaluating the performance of valve positioners with pneumatic outputs

ISBN:978-2-8327-1324-2
Release Date:09-Jun-2026
English language (92 pages)
sale 15% off
Preview
sale 15% off
Preview
Standard

iec61514{ed2.0}en - Industrial-process control systems - Methods of evaluating the performance of valve positioners with pneumatic outputs

ISBN:978-2-8327-1323-5
Release Date:09-Jun-2026
English and French language (92 pages)
sale 15% off
Preview
sale 15% off
Preview
Standard

iec61514{ed2.0}fr - Systemes de commande des processus industriels - Méthodes d'évaluation des performances des positionneurs de vannes a sorties pneumatiques

ISBN:978-2-8327-1323-5
Release Date:09-Jun-2026
English and French language (92 pages)
sale 15% off
Preview
sale 15% off
Preview

Get Certified

Connect with accredited certification bodies for this standard

DVS-ZERT GmbH

German welding certification society.

DAKKS Germany Verified

Institut za varilstvo d.o.o. (Welding Institute)

Slovenia's leading welding institute since 1952. ISO 3834, EN 1090, pressure equipment certification, NDT personnel, welder qualification. Only IIW Au

SA Slovenia Verified

CARES (UK Certification Authority for Reinforcing Steels)

UK certification for reinforcing steels and construction.

UKAS United Kingdom Verified

Sponsored listings

Frequently Asked Questions

IEC 61514:2026 is a standard published by the International Electrotechnical Commission (IEC). Its full title is "Industrial-process control systems - Methods of evaluating the performance of valve positioners with pneumatic outputs". This standard covers: IEC 61514:2026 specifies tests designed to determine the static and dynamic performance of single-acting or double-acting analogue positioners. The tests apply to positioners which receive standard analogue input signals (as specified in IEC 60381-1, IEC 60381-2 and IEC 60382) and have a pneumatic output. Positioners with pulsed or digital input signals, positioners with digital controllers and positioners with pulsed outputs are outside the scope of this document. Testing is conducted either on a positioner alone, independently of an actuator, or on a positioner mounted and connected to a specific actuator, as a combined unit. The text makes clear where different approaches are required. The methods of evaluation given in this document are intended for use by manufacturers to determine the performance of their products, and by users, or independent testing establishments, to verify manufacturers' performance specifications. The closest liaison between the evaluating body and the manufacturer is indispensable during the tests, including the possibility for the manufacturer to influence the test programme based on the manufacturer's specifications for the instrument and comment on both the test programme and the results. This document is intended to provide definitions of positioner elements, actions, and characteristics, to specify uniform methods of measuring performance errors and effects of influence quantities on those characteristics, and to describe methods of reporting and evaluating the results of the measurement data obtained. The test conditions described in this publication (for example range of ambient temperatures and power supply) relate to conditions which commonly arise in use. Consequently, the values specified are used where no other values are specified by the manufacturer or user. If other values are used, they will be stated. It is recognized that the manufacturer's specifications and instructions for installation and operation apply during all steps. The tests specified in this document are not necessarily sufficient for instruments specifically designed for unusually arduous conditions. Conversely, a reduced series of tests can serve adequately for instruments designed to perform within a more limited range of conditions. When a full evaluation, in accordance with this document, is not required or possible, only the tests which are required are performed and the results reported in accordance with the relevant parts of this document. In such cases, the test report will state that it does not cover the full number of tests specified herein. This second edition cancels and replaces the first edition published in 2000. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) in 6.6.8 and Table 5, the magnetic field has been changed from 100 A/m to of 30 A/m (Mean Root Square); b) 6.10.4 and Figure 9 have been modified for better understandability; c) in 7.4, the reference to IEC 61187 has been deleted and replaced with a new Table 4: Document information.

IEC 61514:2026 specifies tests designed to determine the static and dynamic performance of single-acting or double-acting analogue positioners. The tests apply to positioners which receive standard analogue input signals (as specified in IEC 60381-1, IEC 60381-2 and IEC 60382) and have a pneumatic output. Positioners with pulsed or digital input signals, positioners with digital controllers and positioners with pulsed outputs are outside the scope of this document. Testing is conducted either on a positioner alone, independently of an actuator, or on a positioner mounted and connected to a specific actuator, as a combined unit. The text makes clear where different approaches are required. The methods of evaluation given in this document are intended for use by manufacturers to determine the performance of their products, and by users, or independent testing establishments, to verify manufacturers' performance specifications. The closest liaison between the evaluating body and the manufacturer is indispensable during the tests, including the possibility for the manufacturer to influence the test programme based on the manufacturer's specifications for the instrument and comment on both the test programme and the results. This document is intended to provide definitions of positioner elements, actions, and characteristics, to specify uniform methods of measuring performance errors and effects of influence quantities on those characteristics, and to describe methods of reporting and evaluating the results of the measurement data obtained. The test conditions described in this publication (for example range of ambient temperatures and power supply) relate to conditions which commonly arise in use. Consequently, the values specified are used where no other values are specified by the manufacturer or user. If other values are used, they will be stated. It is recognized that the manufacturer's specifications and instructions for installation and operation apply during all steps. The tests specified in this document are not necessarily sufficient for instruments specifically designed for unusually arduous conditions. Conversely, a reduced series of tests can serve adequately for instruments designed to perform within a more limited range of conditions. When a full evaluation, in accordance with this document, is not required or possible, only the tests which are required are performed and the results reported in accordance with the relevant parts of this document. In such cases, the test report will state that it does not cover the full number of tests specified herein. This second edition cancels and replaces the first edition published in 2000. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) in 6.6.8 and Table 5, the magnetic field has been changed from 100 A/m to of 30 A/m (Mean Root Square); b) 6.10.4 and Figure 9 have been modified for better understandability; c) in 7.4, the reference to IEC 61187 has been deleted and replaced with a new Table 4: Document information.

IEC 61514:2026 is classified under the following ICS (International Classification for Standards) categories: 23.060.99 - Other valves; 25.040.40 - Industrial process measurement and control. The ICS classification helps identify the subject area and facilitates finding related standards.

IEC 61514:2026 has the following relationships with other standards: It is inter standard links to IEC 61514:2000. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

IEC 61514:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 positioner
  • 3.1.1 single-acting positioner
  • 3.1.2 double-acting positioner
  • 3.2 input signal W
  • 3.3 travel signal X
  • 3.4 output signal Y
  • 3.5 supply pressure
  • 3.6 action
  • 3.6.1 direct action
  • 3.6.2 reverse action
  • 3.7 split ranging
  • 3.8 gain characteristic
  • 3.9 proportional gain factor K
  • 3.10 proportional band X
  • 3.11 travel characteristic
  • 3.12 measured error
  • 4 General conditions for tests
  • 4.1 Environmental test conditions
  • 4.2 Supply conditions
  • 5 General testing procedures
  • 5.1 Test equipment
  • 5.2 Test methods
  • 5.3 Testing precautions
  • 5.4 Basic test arrangement
  • 5.5 Initial setting-up
  • 6 Test procedures
  • 6.1 Gain characteristic
  • 6.2 Travel characteristic
  • 6.3 Accuracy related terms
  • 6.4 Dead band
  • 6.5 Airflow data
  • 6.6 Effects of influence quantities
  • 6.7 Input over-range
  • 6.8 Drift
  • 6.9 Accelerated life test
  • 6.10 Dynamic response
  • 7 Other considerations
  • 7.1 General
  • 7.2 Safety
  • 7.3 Degree of protection provided by enclosures
  • 7.4 Documentary information
  • 7.5 Installation
  • 7.6 Routine maintenance and adjustment
  • 7.7 Repair
  • 7.8 Protective finishes
  • 7.9 Design features
  • 7.10 Variants
  • 7.11 Tools and equipment
  • 8 Test report and documentation
  • 1 GHz
  • Bibliography

1 Scope

This International Standard specifies tests designed to determine the static and dynamic performance of single-acting or double-acting analogue positioners. The tests apply to positioners which receive standard analogue input signals (as specified in IEC 60381-1, IEC 60381-2 and IEC 60382) and have a pneumatic output.

Positioners with pulsed or digital input signals, positioners with digital controllers and positioners with pulsed outputs are outside the scope of this document. Testing is conducted either on a positioner alone, independently of an actuator, or on a positioner mounted and connected to a specific actuator, as a combined unit. The text makes clear where different approaches are required.

The methods of evaluation given in this document are intended for use by manufacturers to determine the performance of their products, and by users, or independent testing establishments, to verify manufacturers' performance specifications. The closest liaison between the evaluating body and the manufacturer is indispensable during the tests, including the possibility for the manufacturer to influence the test programme based on the manufacturer's specifications for the instrument and comment on both the test programme and the results.

This document is intended to provide definitions of positioner elements, actions, and characteristics, to specify uniform methods of measuring performance errors and effects of influence quantities on those characteristics, and to describe methods of reporting and evaluating the results of the measurement data obtained.

The test conditions described in this publication (for example range of ambient temperatures and power supply) relate to conditions which commonly arise in use. Consequently, the values specified are used where no other values are specified by the manufacturer or user. If other values are used, they will be stated. It is recognized that the manufacturer's specifications and instructions for installation and operation apply during all steps. The tests specified in this document are not necessarily sufficient for instruments specifically designed for unusually arduous conditions. Conversely, a reduced series of tests can serve adequately for instruments designed to perform within a more limited range of conditions. When a full evaluation, in accordance with this document, is not required or possible, only the tests which are required are performed and the results reported in accordance with the relevant parts of this document. In such cases, the test report will state that it does not cover the full number of tests specified herein.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

IEC 60050-311, International electrotechnical vocabulary - Electrical and electronic measurements - Part 311: General terms relating to electrical measurements IEC 60050-351, International electrotechnical vocabulary - Part 351: Control technology IEC 60068-2-1, Environmental testing - Part 2-1: Tests - Test A: Cold

IEC 60068-2-2, Environmental testing - Part 2-2: Tests - Test B: Dry heat IEC 60068-2-6, Environmental testing - Part 2-6: Tests - Test Fc: Vibration (sinusoidal) IEC 60068-2-31, Environmental testing - Part 2-31: Tests - Test Ec: Rough handling shocks, primarily for equipment-type specimens IEC 60068-2-78, Environmental testing - Part 2-78: Tests - Test Cab: Damp heat, steady state IEC 60381-1, Analogue signals for process control systems - Part 1: Direct current signals IEC 60381-2, Analogue signals for process control systems - Part 2: Direct voltage signals IEC 60382, Analogue pneumatic signal for process control systems IEC 60529, Degrees of protection provided by enclosures (IP Code) IEC 60654 (all parts), Industrial-process measurement and control equipment - Operating conditions IEC 60721-3 (all parts), Classification of environmental conditions - Part 3 Classification of groups of environmental parameters and their severities IEC 61010-1, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements IEC 61032, Protection of persons and equipment by enclosures - Probes for verification IEC 61298-4, Process measurement and control devices - General methods and procedures for evaluating performance - Part 4: Evaluation report content IEC 61326-1, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements

3 Terms and definitions

For the purposes of this document, the terms and definitions given in IEC 60050-311 and IEC 60050- 351 and the following apply.

ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • IEC Electropedia: available at https://www.electropedia.org/
  • ISO Online browsing platform: available at https://www.iso.org/obp

3.1 positioner

position controller connected to the moving part of a final control element or its actuator; automatically adjusts its output signal Y to the actuator in order to maintain a desired travel signal X that bears a predetermined relationship to the input signal W Note 1 to entry: In this document, only positioners with pneumatic output signals Y are considered. The input signal W may be an air pressure (pneumatic positioner), or an electric current or voltage (electro-pneumatic positioner).

3.1.1 single-acting positioner

positioner having one output signal Y which acts on one side of the actuator SEE: Figure 1 a).

Note 1 to entry: The returning force for the actuator is usually provided by springs.

3.1.2 double-acting positioner

positioner providing two output signals Y and Y connected to opposite sides of the actuator D R diaphragm or piston SEE: Figure 1 b).

a) Single-acting positioner/actuator

b) Double-acting positioner/actuator Figure 1 – Single- and double-acting positioner/actuator

3.2 input signal W

reference input signal which represents the desired position of the associated control element

3.3 travel signal X

signal which results from the linear or angular travel caused by movement of the final control element or its actuator

3.4 output signal Y

air pressure delivered to the actuator of the final control element

3.5 supply pressure

air pressure at the supply connector of the positioner

3.6 action

direction of the output signal Y in relation to the direction of the input signal W

3.6.1 direct action

output signal Y increases as the value of the input signal W increases

3.6.2 reverse action

output signal Y decreases as the value of the input signal W increases

3.7 split ranging

special adjustment in which the full travel of the actuator is achieved from only part of the whole input range (for example 0 % to 50 % or 50 % to 100 %)

3.8 gain characteristic

relationship between input signal W and output signal Y with travel signal X kept constant (i.e. locked stem) SEE: Figure 2.

Note 1 to entry: Incremental gain ΔY/ΔW varies with pressure and the related pressure shall be stated.

3.9 proportional gain factor K

p gain over the full range of the actuator Note 1 to entry: The proportional gain factor for a single-acting positioner can be derived from the gain characteristic (Figure 2 a)):

ΔY max

K = p ΔW max

where ΔW is the change of input signal W as a percentage of span required to change the output signal over the max whole range (ΔY for 100 %). In this case ΔW (%) corresponds to the proportional band X (%). The output max max p signal range ΔY is taken to be the nominal range stated by the manufacturer. max For a double-acting positioner, a proportional gain factor for each output may be derived separately (see Figure 2 b):

ΔY ΔY max max K = K = pD pR ΔW ΔW Dmax Rmax

The proportional gain factor K for a double-acting positioner can then be calculated as the summation of the two p individual gain factors, i.e.:

KK+ K ppDRp

or derived from the differential pressure characteristic (Y – Y ); see Figure 2 b. D R The balance pressure Y is the cross-over point of the direct Y and reverse Y characteristics. Generally, this value B D R depends on the supply pressure applied to the positioner.

a) Single-acting positioner b) Double-acting positioner

Figure 2 – Gain characteristic

3.10 proportional band X

p is defined as 100 % X (%)= p K p

local gain factor K

slope of the gain characteristic at a specific input value

3.11 travel characteristic

closed-loop relationship of a positioner/actuator between the input signal W and the travel signal X Note 1 to entry: The intended relationship between input signal W and travel signal X (for example linear or equal percentage) determines the ideal characteristic.

3.12 measured error

positive or negative value of error of the average up-scale or down-scale value at each point of measurement

4 General conditions for tests

4.1 Environmental test conditions

4.1.1 General

The tests shall be performed under the ambient test conditions recommended below:

Table 1 – Environmental conditions Atmospheric test conditions Temperature Relative humidity Atmospheric pressure °C % kPa Standard reference atmosphere 20 65 101,3 Recommended limits 15 to 25 45 to 75 86 to 106 Referee measurements a: 20 ± 2 65 ± 5 86 to 106 b: 23 ± 2 50 ± 5 86 to 106

The test values shall be corrected back to the standard reference atmosphere conditions listed above. The standard reference atmosphere is equivalent to the normal reference operating conditions commonly identified by the manufacturer.

It is recognized that there cannot be a factor to correct for humidity. When measurements within the recommended range of ambient conditions are unsatisfactory, and the correction factors to adjust parameters to the standard atmosphere are unknown, repeat measurements (referee measurements) may be conducted under the conditions listed in Table 1, a or b, or other reference operating conditions identified by the manufacturer. Special equipment can be required to maintain the basic test conditions within the limits specified.

4.1.2 Limits of ambient conditions for test measurements

Electromagnetic field: value shall be stated, if relevant.

Maximum rate of change of ambient temperature permissible during any test: 1 °C in 10 min, but not more than 3 °C/h.

4.2 Supply conditions

4.2.1 Reference values

Electrical supply: the values specified by the manufacturer.

Pneumatic supply: the values specified by the manufacturer, or a supply pressure of 4,0 bar (400 kPa).

4.2.2 Tolerances

The tolerances given below apply, unless closer tolerances are agreed between user and manufacturer.

a) Electrical supply 1) Rated voltage: ±1 %.

2) Rated frequency: ±1 %.

3) Harmonic distortion (AC supply): less than 5 %.

4) Ripple (DC supply): less than 0,1 %.

b) Pneumatic supply 1) Rated pressure: ±3 %; 2) Supply air temperature: ambient temperature ±2 °C.

3) Supply air humidity: dew-point at least 10 °C below device body temperature.

4) Oil and dust content

i) oil: less than 10 by weight; ii) dust: absence of particles greater than 3 µm in diameter.

5 General testing procedures

5.1 Test equipment

When the accuracy rating of the reference measuring means is one-tenth or less than that of the device under test, the accuracy rating of the reference measuring means may be ignored in calculations, but shall be reported. When the accuracy rating of the reference measuring means is one-third or less, but greater than one-tenth of that of the device under test, the accuracy rating of the reference measuring means shall be stated in the report.

5.2 Test methods

Specific test methods and test configurations are described separately in Clause 6.

5.3 Testing precautions

Unless affecting the influence condition being tested, the following conditions shall apply. An adequate time, as specified by the manufacturer, shall be allowed after switching on the power supply in order to allow stabilization of the positioner and/or associated test equipment. In the absence of a manufacturer specification, a period of at least 15 min shall be allowed (at least 30 min for electrical supplies).

Prior to recording observations, the device under test shall be exercised by three or more full range traverses in each direction.

The measurement points used to determine the relevant performance characteristic should be distributed over the range. They should include points at or near (within 10 %) the lower- and upper-range values. There should be at least six measurement points, and preferably more. The number and location of these measurement points should be consistent with the degree of precision required and the characteristic being evaluated. Each measurement point should be reached avoiding any overshoot of the input signal.

At each point being observed, the recording shall be made after the device becomes stabilized at its apparent steady-state value.

Tapping or vibrating the device under test is not allowed unless the performance characteristic under study requires such action.

All testing should be conducted with positioner covers in place. Any mechanical stops should be adjusted so that they do not interfere with the measurements.

All tests shall be conducted with the device in (an) agreed mounting position(s), which shall be stated in the report.

Characteristics and data which are dependent on the supply pressure value (for example air consumption, flow capacity, etc.) should be measured at minimum and maximum values of the specified supply pressure range.

Positioners fitted with a cam shall be tested with input to travel characteristics which are normally linear.

5.4 Basic test arrangement

5.4.1 General

The basic test arrangements are shown in Figure 3 a) and 3 b).

5.4.2 Positioner tested independently of an actuator

Tests carried out on a positioner, tested independently of an actuator, are conducted with the output Y of the positioner connected only to a dummy load (sometimes referred to as "openloop" operation).

Unless otherwise agreed, a 1 000 cm volume shall be connected to the output. In this way, the performance of the positioner is not affected by the performance of the actuator. This is the only test method which provides positioner data independent of an actuator, and can be useful in comparing the performance of one positioner with another, or in checking a manufacturer's specification for a positioner.

The travel adjustment can be made manually or by using an actuator (manually controlled). The input signal W can be manually adjusted to obtain the required re-balance of output signal Y, or a loop between output signal Y and input signal W can be arranged if desired, i.e. automatic re-balance.

a)

b)

Figure 3 – Basic test arrangement With positioners possessing an additional integral mode (i.e. P+I control action), it is necessary to use an automatic re-balance loop to achieve stable measurements. In this case, the output Y of the positioner is connected to the input (+) of the pneumatic rebalance device (for example summing relay or P+I controller), whose output is fed back to the input W of the positioner. In cases where the input W is an electrical signal, a pressure/electrical convertor must be included in the circuit, after the re-balance device. The arrangement for this additional test is shown in Figure 3 b).

The input signal W can be a pressure or an electrical signal. The output signal Y is a pressure. The normal test procedure will be to set the travel input signal X to some desired position and then to adjust the input signal W sufficiently to balance a change in output signal Y.

5.4.3 Positioner tested in conjunction with an actuator

Tests carried out in conjunction with an actuator are conducted on a positioner when it is mounted on and connected to an actuator, as a complete positioner/actuator assembly. The positioner may be fitted to any commercially available actuator with which it is compatible. The selected actuator is the choice of the user or the manufacturer. Generally, the results of tests can be affected by performance of the actuator used for the test. Therefore, supporting information about the actuator used (such as friction, inertia, type of packing material, etc.) should be included in the presentation of the test report (see Clause 8).

The input signal W can be a pressure or an electrical signal. Some means of accurately measuring the travel X is necessary. The normal test procedure will be to set the input signal W to some desired value and then to read the corresponding value of the travel signal X. Since the output pressure signals Y (or Y and Y for a double-acting positioner) are connected D R to the actuator (sometimes referred to as "closed-loop" operation) and since their actual values are only an internal signal, they shall be measured only if specifically required.

5.5 Initial setting-up

5.5.1 General

The positioner shall be adjusted in accordance with the manufacturer's instructions.

5.5.2 Positioner tested independently of an actuator

Where the gain of the positioner can be altered, it should be set at a value K = 50 or at a p proportional band X = 2 %. If this is not convenient, another value recommended by the p manufacturer may be used.

If required, some characteristics and data may be measured at minimum or maximum K (X ) p p values.

Adjust the travel span and input range as required. When a positioner can be set for a travel span of 25 mm or 90° for a rotary-actuating positioner, this is preferred. The output pressure range for a single-acting positioner will be 0,2 bar to 1 bar. For a doubleacting positioner, it will be ±0,4 bar from the balance pressure Y . Other values may be specified B by the manufacturer.

With the normal characteristics (i.e. without splitting) the lower range point is determined by 0 % input and 0 % travel, the upper range point is determined by 100 % input and 100 % travel.

5.5.3 Positioner tested in conjunction with an actuator

With a single range (i.e. not split) the lower range point is determined by 0 % input and 0 % travel, and the upper range point is determined by 100 % input and 100 % travel. The gain of the positioner shall be set as high as possible in order to minimize the deviations, but "hunting" (permanent movements of the actuator at constant input signal) must be avoided. If the positioner provides additional adjustment means (for example integral control action), this should be adjusted as specified by the manufacturer. The set values of the gain and/or other adjustment means should be reported.

6 Test procedures

6.1 Gain characteristic

6.1.1 General

Adjust the positioner input signal W to bring the output signal Y to 0 % of its range. Then slowly vary the input signal W so that the output signal Y will successively assume values of 20 %, 40 %, 60 %, 80 %, and 100 % (or other sufficiently small increments) and back to 0 %.

At each setting, measure the input signal W and the output signal Y (or outputs Y and Y in the D R case of a double-acting positioner). Plot the values of the output signal Y (or outputs Y and Y D R in the case of a double-acting positioner) against the input signal W. A typical plot is shown in Figure 2 a) for a single acting positioner and in Figure 2 b) for a double acting positioner). From the gain characteristic plot, the proportional band X or proportional gain factor K can be p p calculated as defined in 3.9 and 3.10 and (if required) the local gain factor K may be derived

...

Contents

  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 positioner
  • 3.1.1 single-acting positioner
  • 3.1.2 double-acting positioner
  • 3.2 input signal W
  • 3.3 travel signal X
  • 3.4 output signal Y
  • 3.5 supply pressure
  • 3.6 action
  • 3.6.1 direct action
  • 3.6.2 reverse action
  • 3.7 split ranging
  • 3.8 gain characteristic
  • 3.9 proportional gain factor K
  • 3.10 proportional band X
  • 3.11 travel characteristic
  • 3.12 measured error
  • 4 General conditions for tests
  • 4.1 Environmental test conditions
  • 4.2 Supply conditions
  • 5 General testing procedures
  • 5.1 Test equipment
  • 5.2 Test methods
  • 5.3 Testing precautions
  • 5.4 Basic test arrangement
  • 5.5 Initial setting-up
  • 6 Test procedures
  • 6.1 Gain characteristic
  • 6.2 Travel characteristic
  • 6.3 Accuracy related terms
  • 6.4 Dead band
  • 6.5 Airflow data
  • 6.6 Effects of influence quantities
  • 6.7 Input over-range
  • 6.8 Drift
  • 6.9 Accelerated life test
  • 6.10 Dynamic response
  • 7 Other considerations
  • 7.1 General
  • 7.2 Safety
  • 7.3 Degree of protection provided by enclosures
  • 7.4 Documentary information
  • 7.5 Installation
  • 7.6 Routine maintenance and adjustment
  • 7.7 Repair
  • 7.8 Protective finishes
  • 7.9 Design features
  • 7.10 Variants
  • 7.11 Tools and equipment
  • 8 Test report and documentation
  • 1 GHz
  • Bibliography

1 Scope

This International Standard specifies tests designed to determine the static and dynamic performance of single-acting or double-acting analogue positioners. The tests apply to positioners which receive standard analogue input signals (as specified in IEC 60381-1, IEC 60381-2 and IEC 60382) and have a pneumatic output.

Positioners with pulsed or digital input signals, positioners with digital controllers and positioners with pulsed outputs are outside the scope of this document. Testing is conducted either on a positioner alone, independently of an actuator, or on a positioner mounted and connected to a specific actuator, as a combined unit. The text makes clear where different approaches are required.

The methods of evaluation given in this document are intended for use by manufacturers to determine the performance of their products, and by users, or independent testing establishments, to verify manufacturers' performance specifications. The closest liaison between the evaluating body and the manufacturer is indispensable during the tests, including the possibility for the manufacturer to influence the test programme based on the manufacturer's specifications for the instrument and comment on both the test programme and the results.

This document is intended to provide definitions of positioner elements, actions, and characteristics, to specify uniform methods of measuring performance errors and effects of influence quantities on those characteristics, and to describe methods of reporting and evaluating the results of the measurement data obtained.

The test conditions described in this publication (for example range of ambient temperatures and power supply) relate to conditions which commonly arise in use. Consequently, the values specified are used where no other values are specified by the manufacturer or user. If other values are used, they will be stated. It is recognized that the manufacturer's specifications and instructions for installation and operation apply during all steps. The tests specified in this document are not necessarily sufficient for instruments specifically designed for unusually arduous conditions. Conversely, a reduced series of tests can serve adequately for instruments designed to perform within a more limited range of conditions. When a full evaluation, in accordance with this document, is not required or possible, only the tests which are required are performed and the results reported in accordance with the relevant parts of this document. In such cases, the test report will state that it does not cover the full number of tests specified herein.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

IEC 60050-311, International electrotechnical vocabulary - Electrical and electronic measurements - Part 311: General terms relating to electrical measurements IEC 60050-351, International electrotechnical vocabulary - Part 351: Control technology IEC 60068-2-1, Environmental testing - Part 2-1: Tests - Test A: Cold

IEC 60068-2-2, Environmental testing - Part 2-2: Tests - Test B: Dry heat IEC 60068-2-6, Environmental testing - Part 2-6: Tests - Test Fc: Vibration (sinusoidal) IEC 60068-2-31, Environmental testing - Part 2-31: Tests - Test Ec: Rough handling shocks, primarily for equipment-type specimens IEC 60068-2-78, Environmental testing - Part 2-78: Tests - Test Cab: Damp heat, steady state IEC 60381-1, Analogue signals for process control systems - Part 1: Direct current signals IEC 60381-2, Analogue signals for process control systems - Part 2: Direct voltage signals IEC 60382, Analogue pneumatic signal for process control systems IEC 60529, Degrees of protection provided by enclosures (IP Code) IEC 60654 (all parts), Industrial-process measurement and control equipment - Operating conditions IEC 60721-3 (all parts), Classification of environmental conditions - Part 3 Classification of groups of environmental parameters and their severities IEC 61010-1, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements IEC 61032, Protection of persons and equipment by enclosures - Probes for verification IEC 61298-4, Process measurement and control devices - General methods and procedures for evaluating performance - Part 4: Evaluation report content IEC 61326-1, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements

3 Terms and definitions

For the purposes of this document, the terms and definitions given in IEC 60050-311 and IEC 60050- 351 and the following apply.

ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • IEC Electropedia: available at https://www.electropedia.org/
  • ISO Online browsing platform: available at https://www.iso.org/obp

3.1 positioner

position controller connected to the moving part of a final control element or its actuator; automatically adjusts its output signal Y to the actuator in order to maintain a desired travel signal X that bears a predetermined relationship to the input signal W Note 1 to entry: In this document, only positioners with pneumatic output signals Y are considered. The input signal W may be an air pressure (pneumatic positioner), or an electric current or voltage (electro-pneumatic positioner).

3.1.1 single-acting positioner

positioner having one output signal Y which acts on one side of the actuator SEE: Figure 1 a).

Note 1 to entry: The returning force for the actuator is usually provided by springs.

3.1.2 double-acting positioner

positioner providing two output signals Y and Y connected to opposite sides of the actuator D R diaphragm or piston SEE: Figure 1 b).

a) Single-acting positioner/actuator

b) Double-acting positioner/actuator Figure 1 – Single- and double-acting positioner/actuator

3.2 input signal W

reference input signal which represents the desired position of the associated control element

3.3 travel signal X

signal which results from the linear or angular travel caused by movement of the final control element or its actuator

3.4 output signal Y

air pressure delivered to the actuator of the final control element

3.5 supply pressure

air pressure at the supply connector of the positioner

3.6 action

direction of the output signal Y in relation to the direction of the input signal W

3.6.1 direct action

output signal Y increases as the value of the input signal W increases

3.6.2 reverse action

output signal Y decreases as the value of the input signal W increases

3.7 split ranging

special adjustment in which the full travel of the actuator is achieved from only part of the whole input range (for example 0 % to 50 % or 50 % to 100 %)

3.8 gain characteristic

relationship between input signal W and output signal Y with travel signal X kept constant (i.e. locked stem) SEE: Figure 2.

Note 1 to entry: Incremental gain ΔY/ΔW varies with pressure and the related pressure shall be stated.

3.9 proportional gain factor K

p gain over the full range of the actuator Note 1 to entry: The proportional gain factor for a single-acting positioner can be derived from the gain characteristic (Figure 2 a)):

ΔY max

K = p ΔW max

where ΔW is the change of input signal W as a percentage of span required to change the output signal over the max whole range (ΔY for 100 %). In this case ΔW (%) corresponds to the proportional band X (%). The output max max p signal range ΔY is taken to be the nominal range stated by the manufacturer. max For a double-acting positioner, a proportional gain factor for each output may be derived separately (see Figure 2 b):

ΔY ΔY max max K = K = pD pR ΔW ΔW Dmax Rmax

The proportional gain factor K for a double-acting positioner can then be calculated as the summation of the two p individual gain factors, i.e.:

KK+ K ppDRp

or derived from the differential pressure characteristic (Y – Y ); see Figure 2 b. D R The balance pressure Y is the cross-over point of the direct Y and reverse Y characteristics. Generally, this value B D R depends on the supply pressure applied to the positioner.

a) Single-acting positioner b) Double-acting positioner

Figure 2 – Gain characteristic

3.10 proportional band X

p is defined as 100 % X (%)= p K p

local gain factor K

slope of the gain characteristic at a specific input value

3.11 travel characteristic

closed-loop relationship of a positioner/actuator between the input signal W and the travel signal X Note 1 to entry: The intended relationship between input signal W and travel signal X (for example linear or equal percentage) determines the ideal characteristic.

3.12 measured error

positive or negative value of error of the average up-scale or down-scale value at each point of measurement

4 General conditions for tests

4.1 Environmental test conditions

4.1.1 General

The tests shall be performed under the ambient test conditions recommended below:

Table 1 – Environmental conditions Atmospheric test conditions Temperature Relative humidity Atmospheric pressure °C % kPa Standard reference atmosphere 20 65 101,3 Recommended limits 15 to 25 45 to 75 86 to 106 Referee measurements a: 20 ± 2 65 ± 5 86 to 106 b: 23 ± 2 50 ± 5 86 to 106

The test values shall be corrected back to the standard reference atmosphere conditions listed above. The standard reference atmosphere is equivalent to the normal reference operating conditions commonly identified by the manufacturer.

It is recognized that there cannot be a factor to correct for humidity. When measurements within the recommended range of ambient conditions are unsatisfactory, and the correction factors to adjust parameters to the standard atmosphere are unknown, repeat measurements (referee measurements) may be conducted under the conditions listed in Table 1, a or b, or other reference operating conditions identified by the manufacturer. Special equipment can be required to maintain the basic test conditions within the limits specified.

4.1.2 Limits of ambient conditions for test measurements

Electromagnetic field: value shall be stated, if relevant.

Maximum rate of change of ambient temperature permissible during any test: 1 °C in 10 min, but not more than 3 °C/h.

4.2 Supply conditions

4.2.1 Reference values

Electrical supply: the values specified by the manufacturer.

Pneumatic supply: the values specified by the manufacturer, or a supply pressure of 4,0 bar (400 kPa).

4.2.2 Tolerances

The tolerances given below apply, unless closer tolerances are agreed between user and manufacturer.

a) Electrical supply 1) Rated voltage: ±1 %.

2) Rated frequency: ±1 %.

3) Harmonic distortion (AC supply): less than 5 %.

4) Ripple (DC supply): less than 0,1 %.

b) Pneumatic supply 1) Rated pressure: ±3 %; 2) Supply air temperature: ambient temperature ±2 °C.

3) Supply air humidity: dew-point at least 10 °C below device body temperature.

4) Oil and dust content

i) oil: less than 10 by weight; ii) dust: absence of particles greater than 3 µm in diameter.

5 General testing procedures

5.1 Test equipment

When the accuracy rating of the reference measuring means is one-tenth or less than that of the device under test, the accuracy rating of the reference measuring means may be ignored in calculations, but shall be reported. When the accuracy rating of the reference measuring means is one-third or less, but greater than one-tenth of that of the device under test, the accuracy rating of the reference measuring means shall be stated in the report.

5.2 Test methods

Specific test methods and test configurations are described separately in Clause 6.

5.3 Testing precautions

Unless affecting the influence condition being tested, the following conditions shall apply. An adequate time, as specified by the manufacturer, shall be allowed after switching on the power supply in order to allow stabilization of the positioner and/or associated test equipment. In the absence of a manufacturer specification, a period of at least 15 min shall be allowed (at least 30 min for electrical supplies).

Prior to recording observations, the device under test shall be exercised by three or more full range traverses in each direction.

The measurement points used to determine the relevant performance characteristic should be distributed over the range. They should include points at or near (within 10 %) the lower- and upper-range values. There should be at least six measurement points, and preferably more. The number and location of these measurement points should be consistent with the degree of precision required and the characteristic being evaluated. Each measurement point should be reached avoiding any overshoot of the input signal.

At each point being observed, the recording shall be made after the device becomes stabilized at its apparent steady-state value.

Tapping or vibrating the device under test is not allowed unless the performance characteristic under study requires such action.

All testing should be conducted with positioner covers in place. Any mechanical stops should be adjusted so that they do not interfere with the measurements.

All tests shall be conducted with the device in (an) agreed mounting position(s), which shall be stated in the report.

Characteristics and data which are dependent on the supply pressure value (for example air consumption, flow capacity, etc.) should be measured at minimum and maximum values of the specified supply pressure range.

Positioners fitted with a cam shall be tested with input to travel characteristics which are normally linear.

5.4 Basic test arrangement

5.4.1 General

The basic test arrangements are shown in Figure 3 a) and 3 b).

5.4.2 Positioner tested independently of an actuator

Tests carried out on a positioner, tested independently of an actuator, are conducted with the output Y of the positioner connected only to a dummy load (sometimes referred to as "openloop" operation).

Unless otherwise agreed, a 1 000 cm volume shall be connected to the output. In this way, the performance of the positioner is not affected by the performance of the actuator. This is the only test method which provides positioner data independent of an actuator, and can be useful in comparing the performance of one positioner with another, or in checking a manufacturer's specification for a positioner.

The travel adjustment can be made manually or by using an actuator (manually controlled). The input signal W can be manually adjusted to obtain the required re-balance of output signal Y, or a loop between output signal Y and input signal W can be arranged if desired, i.e. automatic re-balance.

a)

b)

Figure 3 – Basic test arrangement With positioners possessing an additional integral mode (i.e. P+I control action), it is necessary to use an automatic re-balance loop to achieve stable measurements. In this case, the output Y of the positioner is connected to the input (+) of the pneumatic rebalance device (for example summing relay or P+I controller), whose output is fed back to the input W of the positioner. In cases where the input W is an electrical signal, a pressure/electrical convertor must be included in the circuit, after the re-balance device. The arrangement for this additional test is shown in Figure 3 b).

The input signal W can be a pressure or an electrical signal. The output signal Y is a pressure. The normal test procedure will be to set the travel input signal X to some desired position and then to adjust the input signal W sufficiently to balance a change in output signal Y.

5.4.3 Positioner tested in conjunction with an actuator

Tests carried out in conjunction with an actuator are conducted on a positioner when it is mounted on and connected to an actuator, as a complete positioner/actuator assembly. The positioner may be fitted to any commercially available actuator with which it is compatible. The selected actuator is the choice of the user or the manufacturer. Generally, the results of tests can be affected by performance of the actuator used for the test. Therefore, supporting information about the actuator used (such as friction, inertia, type of packing material, etc.) should be included in the presentation of the test report (see Clause 8).

The input signal W can be a pressure or an electrical signal. Some means of accurately measuring the travel X is necessary. The normal test procedure will be to set the input signal W to some desired value and then to read the corresponding value of the travel signal X. Since the output pressure signals Y (or Y and Y for a double-acting positioner) are connected D R to the actuator (sometimes referred to as "closed-loop" operation) and since their actual values are only an internal signal, they shall be measured only if specifically required.

5.5 Initial setting-up

5.5.1 General

The positioner shall be adjusted in accordance with the manufacturer's instructions.

5.5.2 Positioner tested independently of an actuator

Where the gain of the positioner can be altered, it should be set at a value K = 50 or at a p proportional band X = 2 %. If this is not convenient, another value recommended by the p manufacturer may be used.

If required, some characteristics and data may be measured at minimum or maximum K (X ) p p values.

Adjust the travel span and input range as required. When a positioner can be set for a travel span of 25 mm or 90° for a rotary-actuating positioner, this is preferred. The output pressure range for a single-acting positioner will be 0,2 bar to 1 bar. For a doubleacting positioner, it will be ±0,4 bar from the balance pressure Y . Other values may be specified B by the manufacturer.

With the normal characteristics (i.e. without splitting) the lower range point is determined by 0 % input and 0 % travel, the upper range point is determined by 100 % input and 100 % travel.

5.5.3 Positioner tested in conjunction with an actuator

With a single range (i.e. not split) the lower range point is determined by 0 % input and 0 % travel, and the upper range point is determined by 100 % input and 100 % travel. The gain of the positioner shall be set as high as possible in order to minimize the deviations, but "hunting" (permanent movements of the actuator at constant input signal) must be avoided. If the positioner provides additional adjustment means (for example integral control action), this should be adjusted as specified by the manufacturer. The set values of the gain and/or other adjustment means should be reported.

6 Test procedures

6.1 Gain characteristic

6.1.1 General

Adjust the positioner input signal W to bring the output signal Y to 0 % of its range. Then slowly vary the input signal W so that the output signal Y will successively assume values of 20 %, 40 %, 60 %, 80 %, and 100 % (or other sufficiently small increments) and back to 0 %.

At each setting, measure the input signal W and the output signal Y (or outputs Y and Y in the D R case of a double-acting positioner). Plot the values of the output signal Y (or outputs Y and Y D R in the case of a double-acting positioner) against the input signal W. A typical plot is shown in Figure 2 a) for a single acting positioner and in Figure 2 b) for a double acting positioner). From the gain characteristic plot, the proportional band X or proportional gain factor K can be p p calculated as defined in 3.9 and 3.10 and (if required) the local gain factor K may be derived

...

Contents

  • 1 Domaine d’application
  • 2 Références normatives
  • 3 Termes et définitions
  • 3.1 positionneur
  • 3.1.1 positionneur à simple effet
  • 3.1.2 positionneur à double effet
  • 3.2 signal d’entrée W
  • 3.3 signal de course X
  • 3.4 signal de sortie Y
  • 3.5 pression d’alimentation
  • 3.6 action
  • 3.6.1 action directe
  • 3.6.2 action inversée
  • 3.7 fractionnement de la plage
  • 3.8 caractéristique de gain
  • 3.9 facteur de gain proportionnel K
  • 3.10 bande proportionnelle X
  • 3.11 caractéristique de course
  • 3.12 erreur mesurée
  • 4 Conditions générales pour les essais
  • 4.1 Conditions applicables aux essais d’environnement
  • 4.2 Conditions d’alimentation
  • 5 Procédures générales d’essai
  • 5.1 Matériel d’essai
  • 5.2 Méthodes d’essai
  • 5.3 Précautions à prendre lors des essais
  • 5.4 Montage de base pour les essais
  • 5.5 Réglage initial
  • 6 Procédures d’essai
  • 6.1 Caractéristique de gain
  • 6.2 Caractéristique de course
  • 6.3 Termes relatifs à la précision
  • 6.4 Zone d’insensibilité
  • 6.5 Données relatives au débit d’air
  • 6.6 Effets des grandeurs d’influence
  • 6.7 Dépassement d’entrée
  • 6.8 Dérive
  • 6.9 Essai de durée de vie accéléré
  • 6.10 Réponse dynamique
  • 7 Autres facteurs à considérer
  • 7.1 Généralités
  • 7.2 Sécurité
  • 7.3 Degré de protection procuré par les enveloppes
  • 7.4 Information documentaire
  • 7.5 Installation
  • 7.6 Maintenance de routine et réglages
  • 7.7 Réparation
  • 7.8 Traitements de protection
  • 7.9 Caractéristiques de conception
  • 7.10 Variantes
  • 7.11 Outillage et matériel
  • 8 Rapport d’essai et documentation
  • 1 GHz
  • Bibliographie

1 Domaine d’application

La présente Norme internationale spécifie les essais conçus pour déterminer les performances statiques et dynamiques des positionneurs analogiques à simple ou double effet. Ces essais s’appliquent aux positionneurs qui reçoivent des signaux d’entrée analogiques normaux (tels que spécifiés dans l’IEC 60381-1, à l’IEC 60381-2 et à l’IEC 60382) et qui comportent des sorties pneumatiques.

Les positionneurs à signaux d’entrée à impulsions ou à entrées numériques, les positionneurs avec contrôleurs numériques et les positionneurs à sorties à impulsions ne relèvent pas du domaine d’application du présent document.

Les essais sont réalisés sur un positionneur seul, indépendamment de tout actionneur, ou sur un positionneur monté sur et connecté à un actionneur spécifique, comme dans un dispositif combiné. Le texte stipule clairement les cas où une approche différente est exigée. Les méthodes d’évaluation fournies par le présent document sont destinées à être utilisées par les fabricants afin de déterminer les performances de leurs produits, ainsi que par les utilisateurs ou des organismes d’essai indépendants afin de vérifier les spécifications de performances fournies par les fabricants.

Il est indispensable que l’organisme d’évaluation et le fabricant entretiennent un lien étroit pendant les essais, y compris la possibilité pour le fabricant de peser sur le programme d’essai en fonction de ses spécifications relatives à l’instrument et de commenter le programme d’essai et les résultats.

Le présent document a pour objet de fournir les définitions des éléments, des actions et des caractéristiques des positionneurs, de spécifier des méthodes uniformes pour mesurer les erreurs de performance et l’effet des grandeurs d’influence sur ces caractéristiques, ainsi que de décrire les méthodes de compte rendu et d’évaluation des résultats des données de mesure obtenues.

Les conditions d’essai décrites dans la présente publication (par exemple la plage de températures ambiantes ainsi que l’alimentation en énergie) se rapportent aux conditions d’utilisation les plus courantes. Par conséquent, les valeurs spécifiées sont utilisées lorsque ni l’utilisateur ni le fabricant n’en fournissent d’autres. Si d’autres valeurs sont utilisées, elles sont mentionnées. Il est admis d’appliquer les spécifications et les recommandations du fabricant pour l’installation et l’exploitation à toutes les étapes.

Les essais spécifiés dans le présent document ne sont pas nécessairement suffisants pour des instruments spécialement conçus pour des conditions rarement difficiles. À l’inverse, une série d’essais réduite peut suffire pour des instruments conçus pour fonctionner dans une plage de conditions plus limitée.

Lorsqu’une évaluation complète conforme au présent document n’est ni exigée ni possible, seuls les essais nécessaires sont effectués et les résultats font l’objet d’un compte-rendu conformément aux parties pertinentes du présent document. Dans ces cas, le rapport d’essai précise qu’il ne couvre pas la totalité des essais spécifiés ici.

2 Références normatives

Les documents suivants sont cités dans le texte de sorte qu’ils constituent, pour tout ou partie de leur contenu, des exigences du présent document. Pour les références datées, seule l’édition citée s’applique. Pour les références non datées, la dernière édition du document de référence s’applique (y compris les éventuels amendements).

IEC 60050-311, International electrotechnical vocabulary - Electrical and electronic measurements - Part 311: General terms relating to electrical measurements (disponible en anglais seulement) IEC 60050-351, Vocabulaire électrotechnique international - Partie 351: Technologie de commande et de régulation IEC 60068-2-1, Essais d’environnement - Partie 2-1: Essais - Essai A: Froid IEC 60068-2-2, Essais d’environnement - Partie 2-2: Essais - Essai B: Chaleur sèche IEC 60068-2-6, Essais d’environnement - Partie 2-6: Essais - Essai Fc: Vibrations (sinusoïdales) IEC 60068-2-31, Essais d’environnement - Partie 2-31: Essais - Essai Ec: Choc lié à des manutentions brutales, essai destiné en premier lieu aux matériels IEC 60068-2-78, Essais d’environnement - Partie 2-78: Essais - Essai Cab: Chaleur humide, essai continu IEC 60381-1, Signaux analogiques pour systèmes de commande de processus - Partie 1: Signaux à courant continu IEC 60381-2, Signaux analogiques pour systèmes de commande de processus - Partie 2: Signaux en tension continue IEC 60382, Signal analogique pneumatique pour des systèmes de conduite de processus IEC 60529, Degrés de protection procurés par les enveloppes (code IP) IEC 60654 (toutes les parties), Conditions de fonctionnement pour les matériels de mesure et commande dans les processus industriels IEC 60721-3 (toutes les parties), Classification des conditions d’environnement - Partie 3: Classification des groupements des agents d’environnement et de leurs sévérités IEC 61010-1, Règles de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire - Partie 1: Exigences générales IEC 61032:1997, Protection des personnes et des matériels par les enveloppes - Calibres d’essai pour la vérification IEC 61298-4, Dispositifs de mesure et de commande de processus - Méthodes et procédures générales d’évaluation des performances - Partie 4: contenu du rapport d’évaluation IEC 61326-1, Matériel électrique de mesure, de commande et de laboratoire - Exigences relatives à la CEM - Partie 1i Exigences générales

3 Termes et définitions

Pour les besoins du présent document, les termes et les définitions de l’IEC 60050-311 et de l’IEC 60050-351, ainsi que les suivants, s’appliquent.

L'ISO et l'IEC tiennent à jour des bases de données terminologiques destinées à être utilisées en normalisation, consultables aux adresses suivantes:

  • IEC Electropedia: disponible à l’adresse https://www.electropedia.org/
  • ISO Online browsing platform: disponible à l’adresse https://www.iso.org/obp

3.1 positionneur

contrôleur de position connecté à la partie mobile d’un élément de commande final ou à son actionneur et qui ajuste automatiquement son signal de sortie Y par rapport à l’actionneur afin de maintenir un signal de course X souhaité qui a une relation prédéterminée par rapport au signal d’entrée W Note 1 à l’article: Dans le présent document, seuls les positionneurs à signaux de sortie Y pneumatiques sont examinés. Le signal d’entrée W peut être une pression d’air (positionneur pneumatique), une intensité ou une tension électrique (positionneur électro-pneumatique).

3.1.1 positionneur à simple effet

positionneur ayant un seul signal de sortie Y qui agit sur un seul côté de l’actionneur VOIR: Figure 1 a).

Note 1 à l’article: La force de rappel de l’actionneur est généralement fournie par des ressorts.

3.1.2 positionneur à double effet

et Y , connecté aux deux côtés opposés du positionneur fournissant deux signaux de sortie Y D R piston ou diaphragme de l’actionneur VOIR: Figure 1 b).

a) Positionneur/actionneur à simple effet

b) Positionneur/actionneur à double effet Figure 1 – Positionneur/actionneur à simple et double effet

3.2 signal d’entrée W

signal d’entrée de référence qui représente la position souhaitée de l’élément de commande associé

3.3 signal de course X

signal qui résulte de la course linéaire ou angulaire due au mouvement de l’élément de commande final ou de son actionneur

3.4 signal de sortie Y

pression d’air fournie à l’actionneur de l’élément de commande final

3.5 pression d’alimentation

pression d’air au niveau du raccord d’alimentation du positionneur

3.6 action

sens du signal de sortie Y par rapport au sens du signal d’entrée W

3.6.1 action directe

le signal de sortie Y augmente à mesure que la valeur du signal d’entrée W augmente

3.6.2 action inversée

le signal de sortie Y diminue à mesure que la valeur du signal d’entrée W augmente

3.7 fractionnement de la plage

réglage spécial permettant d’obtenir la course complète de l’actionneur en couvrant uniquement une partie de la plage d’entrée (par exemple de 0 % à 50 % ou de 50 % à 100 %)

3.8 caractéristique de gain

relation entre le signal d’entrée W et le signal de sortie Y avec un signal de course X maintenu constant (c’est-à-dire avec la tige bloquée) VOIR: Figure 2.

Note 1 à l’article: Le gain différentiel ΔY/ΔW varie avec la pression et la pression associée doit être mentionnée

3.9 facteur de gain proportionnel K

p gain sur toute la plage de l’actionneur Note 1 à l’article: Le facteur de gain proportionnel d’un positionneur à simple effet peut être déduit de la caractéristique de gain (Figure 2a):

ΔY max K = p ΔW max

où ΔW est la variation du signal d’entrée W en pourcentage de l’intervalle de mesure exigé pour faire varier le max signal de sortie sur la totalité de la plage (ΔY pour 100 %). Dans ce cas, ΔW (%) correspond à la bande max max proportionnelle X (%). La plage ΔY du signal de sortie est considérée comme étant la plage nominale déclarée p max par le fabricant.

Pour un positionneur à double effet, le facteur de gain proportionnel pour chaque sortie peut être déduit séparément (voir Figure 2b):

ΔY ΔY max max K = K = pD pR ΔW ΔW Dmax Rmax

Le facteur de gain proportionnel K pour un positionneur à double effet peut être calculé comme la somme de deux p facteurs de gain individuels, c’est-à-dire:

KK+ K ppDRp

ou peut être déterminé à partir de la caractéristique de pression différentielle (Y – Y ); voir Figure 2b. D R La pression d’équilibre Y , est le point de rencontre des caractéristiques de pression directe Y et inverse Y . Cette B D R valeur dépend en général de la pression d’alimentation appliquée au positionneur.

a) positionneur à simple effet b) positionneur à double effet

Figure 2 – Caractéristique de gain

3.10 bande proportionnelle X

p est définie en tant que 100 % X (%)= p K p

facteur de gain local K

pente de la caractéristique de gain pour une valeur d’entrée spécifique

3.11 caractéristique de course

relation en boucle fermée entre le signal d’entrée W et le signal de course X d’un positionneur/actionneur.

Note 1 à l’article: La relation prévue entre le signal d’entrée W et le signal de course X (par exemple linéaire ou à pourcentage constant) constitue la caractéristique idéale

3.12 erreur mesurée

valeur positive ou négative de l’erreur calculée à partir de la moyenne des valeurs obtenues pour chaque point de mesure dans le sens montant ou descendant

4 Conditions générales pour les essais

4.1 Conditions applicables aux essais d’environnement

4.1.1 Généralités

Les essais doivent être réalisés dans les conditions ambiantes telles que recommandées ciaprès:

Tableau 1 – Conditions pour les essais d’environnement Conditions atmosphériques Température Humidité relative Pression atmosphérique des essais °C % kPa Atmosphère de référence 20 65 101,3 normale Limites recommandées 15 à 25 45 à 75 86 à 106 Mesures de référence a: 20 ± 2 65 ± 5 86 à 106 b: 23 ± 2 50 ± 5 86 à 106

Les valeurs d’essai doivent être ramenées aux conditions atmosphériques normales de référence indiquées ci-dessus. Les conditions atmosphériques normales de référence sont équivalentes aux conditions atmosphériques normales de fonctionnement généralement identifiées par le fabricant.

L’absence de facteur de correction pour l’humidité peut être admise. Lorsque les mesures effectuées dans les limites de la plage recommandée pour les conditions ambiantes ne sont pas satisfaisantes et que les facteurs de correction pour ajuster les paramètres aux conditions atmosphériques normales sont inconnus, les mesures (de référence) peuvent être répétées dans les conditions indiquées dans le Tableau 1, en a ou b, ou dans d’autres conditions de fonctionnement de référence identifiées par le fabricant.

Un équipement spécial peut être exigé pour maintenir les conditions d’essai de base dans les limites spécifiées.

4.1.2 Limites pour les conditions ambiantes des essais

Champ électromagnétique: la valeur doit être indiquée, si cela est pertinent. Vitesse maximale de variation de la température ambiante admissible pendant les essais: 1 °C par 10 min, sans dépasser 3 °C/h.

4.2 Conditions d’alimentation

4.2.1 Valeurs de référence

Alimentation électrique: les valeurs spécifiées par le fabricant. Alimentation pneumatique: les valeurs spécifiées par le fabricant ou une pression d’alimentation de 4,0 bar (400 kPa).

4.2.2 Tolérances

Les tolérances indiquées ci-dessous s’appliquent, sauf accord entre l’utilisateur et le fabricant sur des tolérances plus serrées.

a) Alimentation électrique 1) Tension assignée: ±1 %.

2) Fréquence assignée: ±1 %.

3) Distorsion harmonique (alimentation en courant alternatif): inférieure à 5 %. 4) Ondulation (alimentation en courant continu): inférieure à 0,1 %. b) Alimentation pneumatique 1) Pression assignée: ±3 %.

2) Température de l’air d’alimentation: température ambiante ±2 °C. 3) Humidité de l’air d’alimentation: point de rosée, au moins 10 °C sous la température du corps du dispositif.

4) Teneur en huile et en poussière

i) huile: moins de 10 en poids; ii) poussière: absence de particules de diamètre supérieur à 3 µm.

5 Procédures générales d’essai

5.1 Matériel d’essai

Lorsque la précision assignée des moyens de mesure de référence est inférieure ou égale à un dixième de celle du dispositif en essai, elle peut être ignorée dans les calculs, mais elle doit être mentionnée dans le rapport d’essai. Lorsque la précision assignée des moyens de mesure de référence est comprise entre un dixième et un tiers de celle du dispositif en essai, la précision nominale des moyens de mesure de référence doit être consignée dans le rapport.

5.2 Méthodes d’essai

Les méthodes d’essai spécifiques et les configurations d’essai sont décrites séparément à l’Article 6.

5.3 Précautions à prendre lors des essais

Sauf si elles influent sur la condition d’influence en cours d’essai, les conditions ci-dessous doivent être respectées.

Un délai approprié, tel que celui spécifié par le fabricant, doit être respecté après la mise sous tension, afin de permettre la stabilisation du positionneur et/ou du matériel d’essai associé. En l’absence de spécifications du fabricant, une période d’au moins 15 min (30 min au moins pour les alimentations électriques) doit être prévue.

Avant d’enregistrer les observations, le dispositif en essai doit effectuer au moins trois traversées complètes de la plage dans chaque sens.

Il convient que les points de mesure servant à déterminer les caractéristiques de performance appropriées soient répartis sur la totalité de la plage. Il est recommandé d’y inclure les points situés à ou proches (à 10 % près) des valeurs inférieure et supérieure de la plage. Il convient qu’il y ait au moins six points de mesure, et de préférence plus. Il est recommandé que le nombre et l’emplacement de ces points de mesure soient compatibles avec le degré de précision exigé et la caractéristique évaluée. Il convient que chaque point de mesure soit atteint en évitant tout dépassement du signal d’entrée.

Pour chaque point observé, l’enregistrement ne doit être effectué qu’après stabilisation du dispositif à sa valeur apparente en régime établi.

Il n’est pas permis de soumettre le dispositif en essai à des chocs ou des vibrations, sauf lorsque la caractéristique de performance étudiée exige une telle action. Il convient que les carters du positionneur soient en place lors des essais. Il est recommandé de régler tous les dispositifs de blocage mécaniques de façon à ne pas interférer avec les mesures.

Tous les essais doivent être menés avec le dispositif dans une position de montage convenue qui doit être mentionnée dans le rapport.

Il convient que les caractéristiques et les données qui dépendent de la valeur de la pression d’alimentation (par exemple la consommation d’air, le débit, etc.) soient mesurées aux valeurs minimale et maximale de la plage spécifiée pour la pression d’alimentation. Les positionneurs équipés d’une came doivent être soumis à essai avec des caractéristiques d’entrée/de course qui sont normalement linéaires.

5.4 Montage de base pour les essais

5.4.1 Généralités

Les montages de base pour les essais sont représentés à la Figure 3 a) et à la Figure 3 b).

5.4.2 Essais séparés d’un positionneur et d’un actionneur

Les essais séparés d’un positionneur et d’un actionneur sont réalisés en raccordant uniquement la sortie Y du positionneur à une charge fictive (parfois appelé fonctionnement en "boucle ouverte").

Sauf accord contraire, un volume de 1 000 cm doit être appliqué en sortie. Les performances du positionneur ne sont alors pas affectées par celles de l’actionneur. Cette méthode d’essai est la seule à fournir des données de positionneur indépendantes d’un actionneur et elle peut servir à comparer les performances entre deux positionneurs, ou à vérifier les spécifications fournies par le fabricant du positionneur. Le réglage de la position peut être effectué manuellement ou à l’aide d’un actionneur (commandé manuellement).

Le signal d’entrée W peut être réglé manuellement pour obtenir le rééquilibrage exigé du signal de sortie Y. Une boucle de contre-réaction peut également être mise en place entre le signal de sortie Y et le signal d’entrée W si un rééquilibrage automatique est souhaité.

a)

b) Figure 3 – Montage de base pour les essais Pour les positionneurs à régulation intégrale (c’est-à-dire une commande de type P+I), il est nécessaire d’utiliser une boucle de rééquilibrage automatique pour réaliser des mesures stables. Dans ce cas, la sortie Y du positionneur est reliée à l’entrée (+) du dispositif pneumatique de rééquilibrage (par exemple le relais additionneur ou le contrôleur P+I), dont la sortie est reliée à l’entrée W du positionneur. Si l’entrée W est un signal électrique, un convertisseur de pression/électricité doit être inséré dans le circuit après le dispositif de rééquilibrage. Le montage pour cet essai supplémentaire est représenté sur la Figure 3 b). Le signal d’entrée W peut être une pression ou un signal électrique. Le signal de sortie Y est une pression. La procédure d’essai normale consiste à affecter au signal d’entrée de course X la position souhaitée, puis à régler le signal d’entrée W de façon à équilibrer suffisamment une variation du signal de sortie Y.

5.4.3 Essais conjoints d’un positionneur et d’un actionneur

Les essais conjoints avec un actionneur sont réalisés en montant et connectant le positionneur à un actionneur, pour former un ensemble positionneur/actionneur complet. Le positionneur peut être connecté à tout actionneur du commerce avec lequel il est compatible. Le choix de l’actionneur relève de l’utilisateur ou du fabricant. En général, les résultats des essais peuvent être affectés par les performances de l’actionneur utilisé pour les essais. Il convient donc d’inclure les informations décrivant l’actionneur (par exemple le frottement, l’inertie, le type de matériau d’emballage, etc.) dans le rapport d’essai (voir l’Article 8).

Le signal d’entrée W peut être une pression ou un signal électrique. Il est nécessaire de prévoir des moyens pour mesurer précisément la course X. La procédure d’essai normale consiste à affecter au signal d’entrée W la valeur souhaitée, puis à lire la valeur correspondante du signal de course X.

Comme les signaux de pression de sortie Y (ou Y et Y pour un positionneur à double effet) D R sont appliqués à l’actionneur (fonctionnement parfois dit "en boucle fermée") et qu’ils n’interviennent que comme des signaux internes, ils doivent être mesurés uniquement si cela est spécifiquement exigé.

5.5 Réglage initial

5.5.1 Généralités

Le positionneur doit être réglé conformément aux instructions du fabricant.

...