CISPR 16-4-2:2011/AMD2:2018/COR1:2019
(Amendment)Corrigendum 1 - Amendment 2 - Specification for radio disturbance and immunity measuring apparatus and methods - Part 4-2: Uncertainties, statistics and limit modelling - Measurement instrumentation uncertainty
Corrigendum 1 - Amendment 2 - Specification for radio disturbance and immunity measuring apparatus and methods - Part 4-2: Uncertainties, statistics and limit modelling - Measurement instrumentation uncertainty
Corrigendum 1 - Amendement 2 - Spécifications des méthodes et des appareils de mesure des perturbations radioélectriques et de l'immunité aux perturbations radioélectriques - Partie 4-2: Incertitudes, statistiques et modélisation des limites - Incertitudes de mesure de l'instrumentation
General Information
Relations
Buy Standard
Standards Content (Sample)
CISPR 16-4-2:2011/AMD2:2018/
COR1:2019 IEC 2019 – 1 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
COMMISSION ÉLECTROTECHNIQUE INTERNATIONALE
____________
CISPR 16-4-2 CISPR 16-4-2
Edition 2.0 2011-06 Édition 2.0 2011-06
Amendment 2: 2018-08 Amendement 2:2018-08
Specification for radio disturbance and immunity Spécifications des méthodes et des appareils de
measuring apparatus mesure des perturbations
and methods – radioélectriques et de l'immunité aux
perturbations radioélectriques –
Part 4-2: Uncertainties, statistics and limit
modelling – Measurement Partie 4-2: Incertitudes, statistiques et
instrumentation uncertainty modélisation des limites – Incertitudes
de mesure de l’instrumentation
CO RRI G E NDUM 1
Corrections to the French version appear after the English text.
Les corrections à la version française sont données après le texte anglais.
Table B.8 − Conducted disturbances measurements from 150 kHz to 30 MHz using a 150
Ω Δ-AN
Replace the existing values for the input quantity AN Impedance (CM) tolerancesδΖ as
AN-CM
follows:
B26)
δΖ + 5,37/- 3,67 Triangular 1,84
AN Impedance (CM) tolerances
AN-CM
Replace the existing values for the combined standard uncertainty u and the expanded
c
uncertainty (U ) 2 u as follows:
CISPR c
u
Combined standard uncertainty 2,93
c
Expanded uncertainty (U ) 2 u
5,86
CISPR c
B.10 Rationale for the estimates of input quantities specific to the measurement method
using a Δ-AN
Replace the existing second sentence in Superscript B2
...