IEC 60747-14-12:2026
(Main)Semiconductor devices - Part 14-12: Semiconductor sensors - Performance test methods for CMOS imager-based gas sensors
- Abstract
IEC 60747-14-12:2026 specifies the performance test methods for CMOS imager-based gas sensors. It includes the terms and definitions, the test environmental conditions, the test system, the test methods, and the test report. This document is applicable to the performance test of both CMOS gas sensors based on lens-free imaging and CMOS gas sensors based on lens imaging.
- Status
- Published
- Publication Date
- 03-Nov-2026
- Technical Committee
- SC 47E - Discrete semiconductor devices
- Drafting Committee
- WG 1 - Semiconductor sensors
- Current Stage
- PPUB - Publication issued - 60 Publication
- Start Date
- 04-Nov-2026
- Due Date
- 27-Nov-2026
- Completion Date
- 27-Nov-2026
IEC 60747-14-12:2026 is an International Standard from the International Electrotechnical Commission (IEC) for CMOS imager-based gas sensors. It specifies performance test methods for these sensors and is written for development, production, and delivery acceptance work. The standard covers both CMOS gas sensors based on lens-free imaging and those based on lens imaging.
What does IEC 60747-14-12:2026 specify?
IEC 60747-14-12:2026 specifies how to test the performance of CMOS imager-based gas sensors. Its scope covers the test environment, the test system, the test methods, the required terms and definitions, and the test report.
The document is organized into seven clauses and two informative annexes:
| Annex | What it covers |
|---|---|
| Annex A | Principle and type of CMOS imager-based gas sensor |
| Annex B | Test report template |
Clause 1 states the scope. Clause 2 is reserved for normative references. Clause 3 gives terms and definitions. Clauses 4 and 5 set the environmental conditions and the test system. Clause 6 sets the test methods, and Clause 7 sets the contents of the test report.
What are the key requirements of IEC 60747-14-12:2026?
IEC 60747-14-12:2026 focuses on a practical test setup and a small set of performance measures that describe how a CMOS imager-based gas sensor behaves in use. The main point is to keep the gas delivery, imaging, and data analysis controlled so results can be compared across tests.
Environmental conditions and test setup
Clause 4 says measurements should be made at 15 °C to 35 °C, 25 %RH to 85 %RH, and 86 kPa to 106 kPa unless otherwise specified. In practice, this means the test lab must control the room conditions and record them with the results.
Clause 5 describes a test system made up of a gas flowmeter, gas pipe, gas chamber, image storage device, and image analysis system. The gas chamber should block ambient light, remain airtight except for the inlet and outlet, and have an anti-adsorption inner surface. This matters because the sensor response is optical and gas handling can affect both signal quality and response time.
The gas flow should be stable, and the gas chamber size should be chosen so filling the chamber has little effect on response time. The document notes that a flow rate in the range of 0,5 L/min to 2 L/min is appropriate for a chamber that is usually only several milliliters in volume.
Sensor settings before testing
Clause 6.1 requires the sensor parameters to be specified before testing. These include resolution, frame rate, exposure time, analog gain, and digital gain. For a buyer or test engineer, this means the image sensor must be configured consistently before any performance numbers are compared.
Sensitivity and limit of detection
Clause 6.2 defines sensitivity testing as measuring the sensor at different target gas concentrations and plotting the average sensing unit value against concentration. Sensitivity is the slope of the best linear fit regression line. In practice, the test shows how strongly the sensor signal changes as gas concentration changes.
Clause 6.3 derives the limit of detection (LOD) from the zero noise of the sensor and its sensitivity. Zero noise is taken from repeated images in a stable zero-gas condition. This gives a practical measure of the smallest reliable concentration the sensor can detect.
Selectivity, response time, repeatability, and stability
Clause 6.4 defines selectivity as the ratio of the sensitivity to the target gas and the sensitivity to an interfering gas. This tells the user how well the sensor distinguishes the gas of interest from other gases.
Clause 6.5 defines response time from the change in average sensing unit value between 10 % and 90 % of the step from initial to stable value. The procedure requires continuous image recording with no more than 1 s between frames. This matters because response time is only meaningful if the transition is captured with enough time resolution.
Clause 6.6 defines repeatability from repeated trials at specified concentrations. The result is expressed by standard deviation, so the user can judge how consistent the sensor is from run to run.
Clause 6.7 defines stability as the change in sensor performance over 24 h between two deliveries of the same target gas. The test uses a first and second measurement separated by 24 hours, so the user can see whether the sensor keeps its behavior over time.
Test report
Clause 7 lists the information that should appear in the test report. It includes environmental conditions, zero gas and target gas details, sensor configuration, and the results for sensitivity, LOD, selectivity, response time, repeatability, and stability. Annex B gives a report template that matches these items.
What terms does IEC 60747-14-12:2026 define?
- Zero gas - a gas or gas mixture with very low content of the component of interest, used to produce the zero response of the sensor.
- Target gas - the gas or gas mixture under test, with a known concentration of the component of interest.
- Sensing unit - the sensing area formed by a material that reacts when it contacts a specific gas.
- Average sensing unit value - the average value taken from the sensing unit over specified image frames.
- Stable state - a condition where the fluctuation value of the sensing unit value is less than 3 % within 1 minute, with at least 10 images captured.
- Sensitivity - the ratio of the change of sensing unit value to the concentration of the target gas.
- Limit of detection (LOD) - the smallest reliable concentration of the target gas that the sensor can detect.
- Repeatability (r) and stability (δ) - repeatability is the closeness of agreement between successive measurements, while stability is the ability of the sensor to keep its performance unchanged over a specified time interval.
Who uses IEC 60747-14-12:2026?
IEC 60747-14-12:2026 is used by designers and manufacturers of CMOS imager-based gas sensors who need a common way to measure sensor performance. It is also relevant to test laboratories and quality teams that verify sensitivity, LOD, selectivity, response time, repeatability, and stability before release or acceptance.
The standard is relevant in fields where gas sensing is used for environmental monitoring, healthcare monitoring, leakage and exposure detection, breath analysis, and food detection. It also supports procurement and conformity assessment work when a buyer needs comparable test results from different suppliers or product types.
Which standards are used with IEC 60747-14-12:2026?
This document states that there are no normative references. It does cite the following standards and vocabulary sources in its definitions and bibliography:
- ISO 7504:2015 - source for the definition of zero gas, modified for use with gas sensors.
- IEC 60747-14-10:2019 - source for the definition of LOD, modified for gas sensors.
- IEC 60050-426:2020 - source for the definition of selectivity.
- IEC 60050-300:2001 - source for the definitions of repeatability and stability.
What does the IEC 60747-14-12:2026 document contain?
Clause 5 includes a schematic of the performance test system, showing the gas flowmeter, gas chamber, image storage device, and image analysis system. Clause 6 includes the measurement procedures and data analysis paths for sensitivity, LOD, selectivity, response time, repeatability, and stability.
The document also includes Figure 2 for the concentration-versus-signal curve and Figure 3 for the response-time curve. Annex A explains the operating principle and shows the two sensor types, lensless imaging and lens imaging, with structure diagrams.
Annex B provides a test report template in table form, which is useful for laboratories and manufacturers that need to record test conditions and results in a consistent format.
Frequently Asked Questions
IEC 60747-14-12:2026 is a standard published by the International Electrotechnical Commission (IEC). Its full title is "Semiconductor devices - Part 14-12: Semiconductor sensors - Performance test methods for CMOS imager-based gas sensors". This standard covers: IEC 60747-14-12:2026 specifies the performance test methods for CMOS imager-based gas sensors. It includes the terms and definitions, the test environmental conditions, the test system, the test methods, and the test report. This document is applicable to the performance test of both CMOS gas sensors based on lens-free imaging and CMOS gas sensors based on lens imaging.
IEC 60747-14-12:2026 specifies the performance test methods for CMOS imager-based gas sensors. It includes the terms and definitions, the test environmental conditions, the test system, the test methods, and the test report. This document is applicable to the performance test of both CMOS gas sensors based on lens-free imaging and CMOS gas sensors based on lens imaging.
IEC 60747-14-12:2026 is classified under the following ICS (International Classification for Standards) categories: 31.080.01 - Semiconductor devices in general. The ICS classification helps identify the subject area and facilitates finding related standards.
IEC 60747-14-12:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 General terms
- 3.1.1 zero gas
- 3.1.2 target gas
- 3.1.3 sensing unit
- 3.1.4 pixel value
- 3.1.5 sensing unit value
- 3.1.6 average sensing unit value
- 3.1.7 stable state
- 3.2 Performance parameters
- 3.2.1 sensitivity
- 3.2.2 limit of detection
- 3.2.3 selectivity
- 3.2.4 response time
- 3.2.5 repeatability
- 3.2.6 stability
- 4 Test environmental conditions
- 5 Test system
- 5.1 General
- 4 CMOS imager-based gas sensor
- 5.2 Gas flowmeter
- 5.3 Gas chamber
- 5.4 Image storage device
- 5.5 Image analysis system
- 6 Test methods
- 6.1 Preparation before test
- 6.2 Sensitivity test
- 6.3 Limit of detection test
- 6.4 Selectivity test
- 6.5 Response time test
- 6.6 Repeatability test
- 6.7 Stability test
- 7 Test report
- Annex A
- 1 Lighting source
- 2 Sensing unit on the surface of CMOS image sensor
- 3 CMOS image sensor
- 1 CMOS image sensor
- 2 Lens
- 3 Sensing unit on a substrate
- 4 Lighting source
- Annex B
- Bibliography
INTRODUCTION Gas sensors play a crucial role in various areas, including environmental monitoring, healthcare monitoring, leakage and exposure detection, and so on. In the era of intelligence, gas sensors are evolving towards miniaturization, low power consumption, multiplexing, and smart features. CMOS imager-based gas sensors are at the forefront of this trend, enabling high-density integration and parallel reading of sensing units. This makes them easily integratable and suitable for intelligent applications, aligning with the future development direction. These sensors can be applied to environmental monitoring, breath analysis, food detection, and other fields.
The key parameters of a gas sensor can reflect its performance, such as sensitivity, limit of detection, and so on. Standards are crucial to specify the test methods of sensor performance and regulate the procedure and data analysis during the development, productions and delivery acceptance. IEC 60747-14-12 specifies the performance test methods of CMOS imager-based gas sensors, which includes the test methods of sensitivity, limit of detection, selectivity, response time, repeatability, and stability.
1 Scope
This part of IEC 60747 specifies the performance test methods for CMOS imager-based gas sensors. It includes the terms and definitions, the test environmental conditions, the test system, the test methods, and the test report. This document is applicable to the performance test of both CMOS gas sensors based on lens-free imaging and CMOS gas sensors based on lens imaging.
2 Normative references
There are no normative references in this document.
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- IEC Electropedia: available at https://www.electropedia.org/
- ISO Online browsing platform: available at https://www.iso.org/obp
3.1 General terms
3.1.1 zero gas
gas or gas mixture with sufficiently low content of the component(s) of interest, used to produce the zero response of a gas sensor for a given range of content [SOURCE: ISO 7504:2015 [1], 5.6, modified – The terms "a given instrument" were changed to "gas sensor".]
3.1.2 target gas
test object of the gas sensor, which is a type of gas or gas mixture with a certain concentration of a kind of component of interest, used to test the performance of the gas sensor Note 1 to entry: Gas concentration is expressed as a volume fraction.
3.1.3 sensing unit
sensing area formed by a material that reacts when in contact with a specific gas
3.1.4 pixel value
value of R channel, or G channel, or B channel, or gray value of a pixel
3.1.5 sensing unit value
average of the pixel value covered by a sensing unit in an image
3.1.6 average sensing unit value
average sensing unit value of specified frames of images
3.1.7 stable state
state in which the fluctuation value of the sensing unit value is less than 3 % within 1 minute, with at least 10 images captured
Note 1 to entry: Fluctuation value is the standard deviation of the sensing unit value divided by the average sensing unit value.
3.2 Performance parameters
3.2.1 sensitivity
S ratio of the change of sensing unit value to the concentration of the target gas
3.2.2 limit of detection
LOD smallest reliable measurable concentration of the target gas (3.1.2) that the gas sensor is able to detect [SOURCE: IEC 60747-14-10:2019 [2], 3.2.7, modified – The terms "quantity value" were changed to "concentration", and "glucose sensor" were changed to "gas sensor".]
3.2.3 selectivity
R response of the sensor to the gas of interest compared with the response to other gases [SOURCE: IEC 60050-426:2020 [3], 426-24-54, modified – The term "equipment" was changed to "sensor".]
3.2.4 response time
Δt duration required for the average sensing unit value to change from 10 % to 90 % of the difference between its initial value and stable value following a step change in the measured gas concentration
3.2.5 repeatability
S r closeness of agreement between the results of successive measurements of the same sample, carried out under the same conditions of measurement [SOURCE: IEC 60050-300:2001 [4], 311-06-06, modified – The term "measurand" was changed to "sample", and the description of the same conditions of measurement were deleted. The admitted term S r was added.]
3.2.6 stability
δ ability of a gas sensor to keep its performance characteristics unchanged during a specified time interval, all other conditions being the same [SOURCE: IEC 60050-300:2001 [4], 311-06-12, modified – The term "measuring instrument" was changed to "gas sensor" and the admitted term δ was added.]
4 Test environmental conditions
Measurements shall be carried out under the following environmental conditions, if not otherwise specified. a) temperature: 15 °C to 35 °C; b) relative humidity: 25 %RH to 85 %RH; c) atmospheric pressure: 86 kPa to 106 kPa.
5 Test system
5.1 General
The principle and type of CMOS imager-based gas sensor are shown in Annex A. The test system for CMOS imager-based gas sensor is shown in Figure 1. The gas inlet of a gas chamber is connected to the output port of a gas flowmeter by a gas pipe. The target gas and zero gas are injected in the input port of the gas flowmeter, flow through the gas chamber, and are discharged through the gas outlet. The gas pipe junction shall be sealed.
The gas sensor under test is fixed in the gas chamber, and the output signal is connected to an image storage device. The image storage device is connected to an image analysis system. The image storage device is used to record the output image of CMOS image sensor in real time and transmits the images to the image analysis system. The image analysis system reads and analyzes the image information.
Key 1 gas flowmeter 2 gas pipe 3 gas chamber
4 CMOS imager-based gas sensor
5 image storage device 6 image analysis system
Figure 1 – Schematic diagram of the performance test system for CMOS imager-based gas sensors
5.2 Gas flowmeter
The flow range of the flowmeter should ensure that the gas flow rate remains stable under the operating state. The appropriate flow rate should be set according to the size of the gas chamber, to ensure that the time for the gas to fill the chamber has little effect on the measurement of the response time. NOTE 1 When the flow rate is too large, the air pressure and noise in the gas chamber are increased and the results are affected. Without these effects, the larger the flow, the better.
NOTE 2 The volume of the gas chamber is usually several milliliters, and a flow rate in the range of 0,5 L/min to 2 L/min is appropriate.
...