ISO/DTS 17242
(Main)Quasi-static calibration procedure for belt force transducers
- Abstract
ISO/TS 17242:2014 provides a procedure to calibrate seat belt force transducers with loading capacities up to 25 kN and consistent test specifications and sequences in order to improve comparability of measurement results between testing laboratories.
- Status
- Not Published
- Technical Committee
- ISO/TC 22/SC 36 - Safety and impact testing
- Drafting Committee
- ISO/TC 22/SC 36/WG 3 - Instrumentation
- Current Stage
- 5000 - Final text received or FDIS registered for formal approval - 50 Approval
- Start Date
- 14-Aug-2026
- Due Date
- 14-Jun-2026
- Completion Date
- 14-Jun-2026
Buy Documents
ISO/DTS 17242 - Quasi-static calibration procedure for belt force transducers
REDLINE ISO/DTS 17242 - Quasi-static calibration procedure for belt force transducers
ISO/DTS 17242 is an ISO document on quasi-static calibration of belt force transducers. It specifies the calibration procedure, test conditions and reporting needed to calibrate seat belt force transducers with loading capacities up to 25 kN, so laboratories can compare measurement results more consistently.
What does ISO/DTS 17242 specify?
ISO/DTS 17242 specifies a procedure for calibrating seat belt force transducers, including the test setup, loading conditions, evaluation methods and the information to record in the calibration report. It is written for transducers with or without linearization electronics and is organized to support repeatable use in testing laboratories.
The document is structured as follows:
- Clause 1 - Scope
- Clause 2 - Normative references
- Clause 3 - Terms and definitions
- Clause 4 - General specifications
- Clause 5 - Test conditions
- Clause 6 - Calibration procedure
- Annex A - Test setup
- Annex B - Evaluation method
- Annex C - Belt strap characteristics
- Annex D - Application notes
| Annex | What it covers |
|---|---|
| Annex A | Schematic calibration test setup and mounting arrangement |
| Annex B | Linear regression and third degree polynomial evaluation methods |
| Annex C | Standardized belt strap characteristics and thickness measurement setup |
| Annex D | Application notes on polynomial scaling and unified measurement ranges |
What are the key requirements of ISO/DTS 17242?
ISO/DTS 17242 focuses on getting the same sensor, strap and test conditions from one laboratory to another, because the calibration result depends strongly on how the transducer is mounted and loaded. In practice, the document gives labs a common way to generate calibration data and to document the conditions behind it.
Transducer types and belt strap selection
Clause 4 categorizes belt force transducers as Type 1 baffle, Type 2 loop, Type 3 clamping and Type 4 3-bolt. The document applies to transducers up to 25 kN and covers devices with or without linearization electronics.
Clause 4.2 explains when a standardized calibration belt strap from Annex C should be used and when the seat belt webbing intended for later testing should be used instead. For Type 1 transducers, the standardized strap is mainly for comparability between laboratories, while the actual test webbing is generally used when absolute values are needed. The belt strap type and its average thickness must be recorded in the calibration report, so users can trace results back to the exact test material.
Test conditions that control the measurement
Clause 5 sets the loading method as a continuous load up to full scale. The free clamping length is defined for the setup in Annex A, and the test velocity is controlled so the loading is quasi-static rather than dynamic.
The clause also states that hysteresis is not verified, so the evaluation concentrates on the loading curve. Data acquisition has to follow ISO 6487 filter settings, and the first loading sequence is used for evaluation with a minimum sampling rate of 100 Hz. The evaluation range runs from 3.5 % to 100 % of the calibration range, which matters because the reported sensitivity or polynomial fit is only valid over that span.
Conditioning, mounting and environmental control
Clause 6 requires conditioning of the data acquisition system, test transducer and reference transducer so they reach thermal equilibrium. The calibration is carried out under monitored conditions between 19 °C and 23 °C and relative humidity between 10 % and 70 %, with an acclimatization period of 8 h for the transducers and belt strap specimen.
Before loading, the sensor offset is measured under no load, a new belt strap is inserted, and the transducer is mounted centered and perpendicular. The preload while holding the transducer in place must not exceed 0.2 % of the calibration range. In practice, this is the part that keeps the setup from introducing extra error before the actual calibration starts.
Data treatment and regression methods
Clause 6.4 and Annex B allow two evaluation paths: linear regression or third degree polynomial regression. For linear regression, the zero offset is corrected before the test. For polynomial regression, the zero offset is not corrected before the test, and the calibration coefficient is handled so the original offset is preserved for later use.
Annex B defines the fitting methods used to calculate the calibration coefficients from the recorded data points. This matters because laboratories can use the same recorded curve to produce either a straight-line calibration or a polynomial approximation, depending on the transducer and the intended use.
Reporting and traceability
Clause 6.7 lists the information that has to go into the calibration report. It includes temperature, humidity, belt strap type, strap thickness, sensitivity or polynomial equation, polynomial coefficients, measured offset, interpolation deviation, repeat accuracy and uncertainty.
The report also has to note the sensor excitation voltage, the sensor connection and whether excitation sensing was used. In practice, this is what lets another user judge whether the calibration can be compared with a different sensor, a different strap or a different laboratory setup.
Who uses ISO/DTS 17242?
ISO/DTS 17242 is used by laboratories that calibrate belt force transducers and by crash test laboratories that rely on those sensors in occupant restraint testing. It is also relevant to development engineers, test engineers and quality staff who need consistent calibration data across different sites.
Typical users include:
- calibration laboratories that verify force transducers
- crash test laboratories working with seat belt force measurements
- vehicle safety and impact testing teams
- manufacturers and suppliers of belt force transducers
- quality managers who need traceable calibration records
What changed in ISO/DTS 17242 from the previous edition?
ISO/DTS 17242 is the second edition and it replaces ISO/TS 17242:2014 and ISO/TS 17242:2014/Amd 1:2017. The foreword states that it is technically revised and incorporates the amendment.
The stated changes are:
- the evaluation range for calibration in 5.8 has been updated
- the test report content in 6.7 has been updated for repeat accuracy and uncertainty values
- the derivation of the polynomial regression in Annex B has been corrected
- the belt strap requirements in Annex C have been changed for thickness, breaking load and filling threads
Which standards are used with ISO/DTS 17242?
ISO/DTS 17242 cites three normative references:
- ISO 376 - provides the calibration basis for force-proving instruments used to verify uniaxial testing machines, which supports traceability of the comparison standard
- ISO 6487 - provides measurement techniques and instrumentation requirements for impact tests, including the filter settings used for data acquisition
- SAE-J2517 - provides the calibration procedure referenced for polynomial handling and zero-offset preservation in the calibration process
What does the ISO/DTS 17242 document contain?
ISO/DTS 17242 contains a transducer classification table, a schematic test setup, two evaluation methods and belt strap specification data. It also gives practical application notes that explain when polynomial scaling is useful and how to choose a unified measurement range.
Annex A shows the mounting arrangement for the calibration test, including the reference transducer, test transducer and clamping layout. Annex B gives the linear and third degree polynomial regression methods used to calculate calibration coefficients from measured data. Annex C specifies the characteristics of the standardized calibration belt strap and shows how to measure its average thickness. Annex D explains how electronic linearization affects nonlinearity and why laboratories should align measurement ranges when they want comparable results.
Relations
- Effective Date
- 18-Nov-2023
- Effective Date
- 18-Nov-2023
Buy Documents
ISO/DTS 17242 - Quasi-static calibration procedure for belt force transducers
REDLINE ISO/DTS 17242 - Quasi-static calibration procedure for belt force transducers
Get Certified
Connect with accredited certification bodies for this standard

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DEKRA Certification Inc.
DEKRA US certification services.
Sponsored listings
Frequently Asked Questions
ISO/DTS 17242 is a draft published by the International Organization for Standardization (ISO). Its full title is "Quasi-static calibration procedure for belt force transducers". This standard covers: ISO/TS 17242:2014 provides a procedure to calibrate seat belt force transducers with loading capacities up to 25 kN and consistent test specifications and sequences in order to improve comparability of measurement results between testing laboratories.
ISO/TS 17242:2014 provides a procedure to calibrate seat belt force transducers with loading capacities up to 25 kN and consistent test specifications and sequences in order to improve comparability of measurement results between testing laboratories.
ISO/DTS 17242 is classified under the following ICS (International Classification for Standards) categories: 43.040.80 - Crash protection and restraint systems. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO/DTS 17242 has the following relationships with other standards: It is inter standard links to ISO/TS 17242:2014/Amd 1:2017, ISO/TS 17242:2014. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ISO/DTS 17242 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 General specifications
- 4.1 General
- 1 Baffle
- 2 Loop
- 3 Clamping
- 4.2 Limitations of the application of the test belt strap
- 5 Test conditions
- 5.1 General
- 5.2 Test method
- 5.3 Clamping length
- 5.4 Test velocity
- 5.5 Belt strap
- 5.6 Load relieving
- 5.7 Data acquisition
- 5.8 Data evaluation
- 5.9 Sensor excitation
- 5.10 Environmental conditions
- 6 Calibration procedure
- 6.1 General
- 6.2 Conditioning
- 6.3 Test preparation
- 6.4 Calibration test
- 6.5 Data storage
- 6.6 Data evaluation
- 6.7 Documentation
- Annex A
- Annex B
- Annex C
- Annex D
- Bibliography
1 Scope
This document provides a procedure to calibrate seat belt force transducers with loading capacities up to 25 kN and consistent test specifications and sequences in order to improve comparability of measurement results between testing laboratories.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 6487, Road vehicles — Measurement techniques in impact tests — Instrumentation SAE-J2517, Hybrid III Family Chest Potentiometer Calibration Procedure
3 Terms and definitions
No terms and definitions are listed in this document.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https:// www .iso .org/ obp
- IEC Electropedia: available at https:// www .electropedia .org/
4 General specifications
4.1 General
The described measurement procedure refers to belt force transducers–whether with or without linearization electronics–with the specifications according to Table 1 and a loading capacity of up to 25 kN.
ISO/DTS 17242:2026(en) Table 1 — Design categorization of belt force transducers Type Description Belt orientation
1 Baffle
(test belt strap according to Annex C)
2 Loop
(test belt strap according to customer requirement)
3 Clamping
(test belt strap according to customer requirement) 4 3-bolt (test belt strap according to customer requirement)
4.2 Limitations of the application of the test belt strap
In general, the standardized calibration belt strap (see Clause C.1) is only recommended for the calibration of belt force transducers which are to be utilized in tests that focus on comparability between different laboratories (such as during development efforts). In case absolute values have to be obtained, the seat belt webbing designated for subsequent testing purposes shall be used for calibration. Measurement of absolute values shows significant variation for type 1 transducers as shown in Table 1. Calibration using the standardized calibration belt strap shall be done for highly dynamically loaded transducers type 1 (baffle). Consequently calibration of type 1 transducers for quasi-static loads and measurement ranges up to 500 N (i.e. comfort measurements) as well as types 2, 3 and 4, in general, use the seat belt webbing designated for subsequent testing purposes for calibration. The belt strap type used for calibration as well as the strap’s average thickness (see Annex C) shall be identified in the calibration report.
If a non-standardized belt strap is supplied, it is also recommended to perform a calibration using the standardized belt strap for traceability.
See Annex D for the application notes.
5 Test conditions
5.1 General
The calibration test shall be performed under the conditions identified in 5.2 to 5.10 .
ISO/DTS 17242:2026(en)
5.2 Test method
A continuous loading up to the transducer’s full scale range shall be applied. NOTE See the transducer data sheet for the full scale range.
5.3 Clamping length
As shown in Annex A, Figure A.1, the belt force transducer is tested on a belt strap specimen clamped into the universal tension machine. The free clamping length shall be within 375 mm ± 75 mm.
5.4 Test velocity
The speed of displacement of the tension machine for the continuous loading shall be within 170 mm/ min ± 30 mm/min.
5.5 Belt strap
A standardized belt strap specimen as defined in Annex C should be utilized. Other belt strap types should be identified in the calibration report.
5.6 Load relieving
The hysteresis is not verified.
5.7 Data acquisition
In case of polynomial regression, the zero offset shall not be corrected prior to the test. Therefore, the amplifier offset should be considered for the error of measurement of the whole measurement channel. The filter settings shall comply with the specifications of ISO 6487.
5.8 Data evaluation
For the evaluation, the first loading sequence recorded with a minimum sampling rate of 100 Hz shall be considered in a range from 3,5 % to 100 % of the transducer’s calibration range (e.g. 0,56 kN – 16 kN at 16 kN [1] sensor range). Ensure the comparison standard is traceable (e.g. ISO 376 ). Thenonlinearity is displayed as a percentage [%] of the calibration range.
NOTE 1 The first loading sequence does not include preloading procedures. NOTE 2 A sampling rate of 100 Hz is used to reach appropriate accuracy.
5.9 Sensor excitation
The sensor excitation voltage, as well as the sensor connection, shall be in accordance to the customer requirements and shall be recorded. If excitation sensing is utilized it shall be stated in calibration report.
5.10 Environmental conditions
The calibration test shall be performed under monitored environmental conditions within a temperature range of 19 °C to 23 °C and a relative humidity of 10 % to 70 %. For acclimatization purposes, the transducers as well as the belt strap specimen should be kept in this environment for a period of 8 h prior to the calibration.
ISO/DTS 17242:2026(en)
6 Calibration procedure
6.1 General
Proceed with the following steps to perform the calibration test.
...
Contents
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 General specifications
- 4.1 General
- 1 Baffle
- 2 Loop
- 3 Clamping
- 4.2 Limitations of the application of the test belt strap
- 5 Test conditions
- 5.1 General
- 6 Calibration procedure
- 6.1 General
- Annex A
- Annex B
- Annex C
- Annex D
- Bibliography
1 Scope
This document provides a procedure to calibrate seat belt force transducers with loading capacities up to 25 kN and consistent test specifications and sequences in order to improve comparability of measurement results between testing laboratories.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 376, Metallic materials — Calibration of force-proving instruments used for the verification of uniaxial testing machines ISO 6487, Road vehicles — Measurement techniques in impact tests — Instrumentation SAE-J2517, Hybrid III Family Chest Potentiometer Calibration Procedure
3 Terms and definitions
There are noNo terms and definitions are listed in this document. ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https://www.iso.org/obp
- IEC Electropedia: available at https://www.electropedia.org/
4 General specifications
4.1 General
The described measurement procedure refers to belt force transducers - –whether with or without linearization electronics - –with the specifications according to Table 1 and a loading capacity of up to 25 kN. Table 1 — Design categorization of belt force transducers Type Description Belt Orientationorientation
1 Baffle
(test belt strap according to 0)
ISO/CD TSDTS 17242:2026(en) Type Description Belt Orientationorientation
2 Loop
(test belt strap according to customer requirement)
3 Clamping
(test belt strap according to customer requirement)
4 3-Boltbolt (test belt strap according to customer requirement)
4.2 Limitations of the application of the test belt strap
In general, the standardized calibration belt strap (see C.1) is only recommended for the calibration of belt force transducers which are to be utilized in tests that focus on comparability between different laboratories (such as during development efforts). In case absolute values have to be obtained, the seat belt webbing designated for subsequent testing purposes is toshall be used for calibration. Measurement of absolute values shows significant variation for type 1 transducers as shown in Table 1 . Calibration using the standardized calibration belt strap shall be done for highly dynamically loaded transducers Typetype 1 (baffle). Consequently calibration of Typetype 1 transducers for quasi-static loads and measurement ranges up to 500 N (i.e. comfort measurements) as well as Typestypes 2, 3 and 4, in general, use the seat belt webbing designated for subsequent testing purposes for calibration.
ISO/CD TSDTS 17242:2026(en) The belt strap type used for calibration as well as the strap’s average thickness (see 0) shall be identified in the calibration report.
If a non-standardized belt strap is supplied, it is also recommended to perform a calibration using the standardized belt strap for traceability.
See Annex D for the application notes.
5 Test conditions
5.1 General
The calibration test shall be performed under the conditions identified in 5.2below to 5.10 .
5.15.2 Test method
1) A continuous loading up to the transducer’s full scale range shall be applied. NOTE See the transducer data sheet for the full scale range.
5.25.3 Clamping length
As shown in Annex A, Figure A.1 ,, the belt force transducer is tested on a belt strap specimen clamped into the universal tension machine. The free clamping length shall be within 375 mm ± 75 mm.
5.35.4 Test velocity
The speed of displacement of the tension machine for the continuous loading shall be within 170 mm/min ± 30 mm/min.
5.45.5 Belt strap
A standardized belt strap specimen as defined in 0 should be utilized. Other belt strap types should be identified in the calibration report.
5.55.6 Load relieving
The hysteresis is not verified.
5.65.7 Data acquisition
In case of polynomial regression, the zero offset shall not be corrected prior to the test. SoTherefore, the amplifier offset should be considered for the error of measurement of the whole measurement channel. The filter settings shall comply with the specifications of ISO 6487.
5.75.8 Data evaluation
2) 3) For the evaluation, the first loading sequence recorded with a minimum sampling rate of 100 Hz shall be considered in a range from 3,5 % to 100 % of the transducer’s calibration range (e.g. 0,56 kN – 16 kN @at
1) See transducer data sheet.
2) No preloading procedures.
3) To reach appropriate accuracy.
ISO/CD TSDTS 17242:2026(en) 4) 16 kN sensor range). Ensure the comparison standard is traceable (e.g. [1]) . The nonlinearity). Thenonlinearity is displayed as a percentage in %[%] of the calibration range. NOTE 1 The first loading sequence does not include preloading procedures. NOTE 2 A sampling rate of 100 Hz is used to reach appropriate accuracy.
5.85.9 Sensor excitation
The sensor excitation voltage, as well as the sensor connection, shall be in accordance to the customer requirements and shall be recorded. If excitation sensing wasis utilized it shall be stated in calibration report.
5.95.10 Environmental conditions
The calibration test shall be performed under monitored environmental conditions within a temperature range of 19 °C to 23 °C and a Relative Humidityrelative humidity of 10 % to 70 %. For acclimatization purposes, the transducers as well as the belt strap specimen should be kept in this environment for a period of 8 h prior to the calibration.
6 Calibration procedure
6.1 General
Proceed with the following steps to perform the calibration test.
6.16.2 Conditioning
An adequate warm-up time for the data acquisition system, test and reference transducer should be allowed to reach thermal equilibrium.
6.26.3 Test preparation
The sensor offset should be measured under no load condition. A new belt strap shall be inserted and mounted 5) in the fixations. The fixation can be done by clamp or loop, see [2] . regarding conformance. The test transducer shall be mounted centred and perpendicular in accordance with the sensor user’s manual. The belt strap preload shall not exceed 0,2 % of the calibration range (while holding the transducer in place).
...