Surface chemical analysis — Vocabulary — Part 1: General terms and terms used in spectroscopy

ISO 18115-1:2013 defines terms for surface chemical analysis. It covers general terms and those used in spectroscopy.

Analyse chimique des surfaces — Vocabulaire — Partie 1: Termes généraux et termes utilisés en spectroscopie

General Information

Status
Withdrawn
Publication Date
03-Nov-2013
Current Stage
9599 - Withdrawal of International Standard
Completion Date
19-Jun-2023
Ref Project

Relations

Buy Standard

Standard
ISO 18115-1:2013 - Surface chemical analysis -- Vocabulary
English language
104 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 18115-1
Second edition
2013-11-15
Surface chemical analysis —
Vocabulary —
Part 1:
General terms and terms used in
spectroscopy
Analyse chimique des surfaces — Vocabulaire —
Partie 1: Termes généraux et termes utilisés en spectroscopie
Reference number
ISO 18115-1:2013(E)
©
ISO 2013

---------------------- Page: 1 ----------------------
ISO 18115-1:2013(E)

COPYRIGHT PROTECTED DOCUMENT
© ISO 2013
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2013 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 18115-1:2013(E)

Contents Page
Foreword .iv
Introduction .v
0 Scope . 1
1 Abbreviated terms . 1
2 Format . 3
2.1 Use of terms printed italic in definitions . 3
2.2 Non-preferred and deprecated terms . 3
2.3 Subject fields . 4
3 Definitions of the surface analysis methods . 4
4 Definitions of terms for surface analysis . 8
5 Definitions of terms for multivariate analysis .83
6 Definitions of supplementary terms for surface analysis methods .90
7 Definitions of supplementary terms for surface analysis .95
8 Definitions of supplementary terms for multivariate analysis .101
[11]
Annex A (informative) Extract from IEC 60050-111 .102
Bibliography .104
© ISO 2013 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 18115-1:2013(E)

Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation on the meaning of ISO specific terms and expressions related to conformity
assessment, as well as information about ISO’s adherence to the WTO principles in the Technical Barriers
to Trade (TBT) see the following URL: Foreword - Supplementary information
The committee responsible for this document is ISO/TC 201, Surface chemical analysis, Subcommittee
SC 1, Terminology.
This second edition cancels and replaces the first edition (ISO 18115-1:2010), which has been
technically revised.
ISO 18115 consists of the following parts, under the general title Surface chemical analysis — Vocabulary:
— Part 1: General terms and terms used in spectroscopy
— Part 2: Terms used in scanning-probe microscopy
iv © ISO 2013 – All rights reserved

---------------------- Page: 4 ----------------------
ISO 18115-1:2013(E)

Introduction
Surface chemical analysis is an important area which involves interactions between people with
different backgrounds and from different fields. Those conducting surface chemical analysis might be
materials scientists, chemists, or physicists and might have a background that is primarily experimental
or primarily theoretical. Those making use of the surface chemical data extend beyond this group into
other disciplines.
With the present techniques of surface chemical analysis, compositional information is obtained for
regions close to a surface (generally within 20 nm) and composition-versus-depth information is
obtained with surface analytical techniques as surface layers are removed. The surface analytical
terms covered in this part of ISO 18115 extend from the techniques of electron spectroscopy and mass
spectrometry to optical spectrometry and X-ray analysis. The terms covered in ISO 18115-2 relate to
scanning-probe microscopy. Concepts for these techniques derive from disciplines as widely ranging as
nuclear physics and radiation science to physical chemistry and optics.
The wide range of disciplines and the individualities of national usages have led to different meanings
being attributed to particular terms and, again, different terms being used to describe the same concept.
To avoid the consequent misunderstandings and to facilitate the exchange of information, it is essential
to clarify the concepts, to establish the correct terms for use, and to establish their definitions.
The terms and definitions in this International Standard have been prepared in conformance with the
principles and style defined in ISO 1087-1:2000 and ISO 10241:1992. Essential aspects of these standards
appear in 2.1 to 2.3. This part of ISO 18115 comprises the 78 abbreviations and 590 definitions of the
combined ISO 18115-1:2010 and Amendment 1 to ISO 18115-1:2010. Corrections have been made to
terms 4.61, backscattering factor, and 4.480, unified atomic mass unit that appeared in ISO 18115-1:2010.
The terms are given in alphabetical order, classified under Clauses 3, 4, and 5 from the former
International Standard with corrections and Clauses 6, 7, and 8 from Amendment 1:
— Clause 3: Definitions of the surface analysis methods;
— Clause 4: Definitions of terms for surface analysis;
— Clause 5: Definitions of terms for multivariate analysis;
— Clause 6: Definitions of supplementary terms for the surface analysis methods;
— Clause 7: Definitions of supplementary terms for surface analysis;
— Clause 8: Definitions of supplementary terms for multivariate analysis.
Additional terms, important for surface analysis, are given in an extract from IEC 60050-111 in Annex A.
© ISO 2013 – All rights reserved v

---------------------- Page: 5 ----------------------
INTERNATIONAL STANDARD ISO 18115-1:2013(E)
Surface chemical analysis — Vocabulary —
Part 1:
General terms and terms used in spectroscopy
0 Scope
This part of ISO 18115 defines terms for surface chemical analysis. It covers general terms and those
used in spectroscopy while ISO 18115-2 covers terms used in scanning-probe microscopy.
1 Abbreviated terms
AC alternating current
AES Auger electron spectroscopy
AMRSF average matrix relative sensitivity factor
ANN artificial neural network
APECS Auger photoelectron coincidence spectroscopy
ARAES angle-resolved Auger electron spectroscopy
AREPES angle-resolved elastic peak electron spectroscopy
ARXPS angle-resolved X-ray photoelectron spectroscopy
CDP compositional depth profile
CRM certified reference material
DA/DFA discriminant analysis/discriminant function analysis
DAPCI desorption atmospheric pressure chemical ionization
DAPPI desorption atmospheric pressure photoionization
DART direct analysis in real time
DC direct current
DESI desorption electrospray ionization
DRS direct recoil spectroscopy
eV electron volts
EELS electron energy loss spectroscopy
EESI extractive electrospray ionization
EIA energetic-ion analysis
ELDI electrospray enhanced laser desorption mass spectrometry
© ISO 2013 – All rights reserved 1

---------------------- Page: 6 ----------------------
ISO 18115-1:2013(E)

EPES elastic peak electron spectroscopy
EPMA electron probe microanalysis
ERD elastic recoil detection
ERDA elastic recoil detection analysis
ESCA electron spectroscopy for chemical analysis
EXAFS extended X-ray absorption fine structure spectroscopy
FABMS fast atom bombardment mass spectrometry
FIB focused ion beam system
FWHM full width at half maximum
GDMS glow discharge mass spectrometry
GDOES glow discharge optical emission spectrometry
GDS glow discharge spectrometry
GISAXS grazing-incidence small-angle X-ray scattering
HSA hemispherical sector analyser
IBA ion beam analysis
ISS ion-scattering spectrometry
LAESI laser ablation electrospray ionization
LB Langmuir-Blodgett
LDI laser desorption ionization
LEIS(S) low-energy ion scattering spectrometry
LMIG liquid-metal ion gun
LMIS liquid-metal ion source
MAF analysis maximum autocorrelation factor analysis
MALDI matrix-assisted laser desorption/ionization mass spectrometry
MALDESI matrix-assisted laser desorption electrospray ionization
MCR multivariate curve resolution
MEIS(S) medium-energy ion scattering spectrometry
MVA multivariate analysis
NEXAFS near-edge extended X-ray absorption fine structure spectroscopy
PADI plasma-assisted desorption ionization
PCA principal-component analysis
PERSF pure-element relative sensitivity factor
2 © ISO 2013 – All rights reserved

---------------------- Page: 7 ----------------------
ISO 18115-1:2013(E)

PIXE particle-induced X-ray emission
PLS partial least squares
RBS Rutherford backscattering spectrometry
REELS reflection electron energy loss spectroscopy
RISR relative instrument spectral response function
rf radio-frequency
RM reference material
RSF relative sensitivity factor
SALDI surface-assisted laser desorption/ionization
SAM self-assembled monolayer
SAXS small-angle X-ray scattering
SDP sputter depth profile
SEM scanning electron microscope
SEP surface excitation parameter
SEXAFS surface extended X-ray absorption fine structure spectroscopy
SIMS secondary-ion mass spectrometry
SNMS sputtered neutral mass spectrometry
SSA spherical sector analyser
TOF or ToF time of flight
TXRF total-reflection X-ray fluorescence spectroscopy
UPS ultraviolet photoelectron spectroscopy
XAFS X-ray absorption fine structure spectroscopy
XANES X-ray absorption near-edge spectroscopy
XPS X-ray photoelectron spectroscopy
XRR X-ray reflectometry
XSW X-ray standing waves
2 Format
2.1 Use of terms printed italic in definitions
A term printed in italics in a definition or a note is defined in another entry in this part of ISO 18115.
However, the term is printed in italics only the first time it occurs in each entry.
2.2 Non-preferred and deprecated terms
A term listed lightface is non-preferred or deprecated. The preferred term is listed boldface.
© ISO 2013 – All rights reserved 3

---------------------- Page: 8 ----------------------
ISO 18115-1:2013(E)

2.3 Subject fields
Where a term designates several concepts, it is necessary to indicate the subject field to which each concept
belongs. The field is shown lightface, between angle brackets, preceding the definition, on the same line.
3 Definitions of the surface analysis methods
3.1
Auger electron spectroscopy
AES
method in which an electron spectrometer (4.190) is used to measure the energy distribution of Auger
electrons (4.37) emitted from a surface (4.458)
Note 1 to entry: An electron beam in the energy range 2 keV to 30 keV is often used for excitation of the Auger
electrons. Auger electrons can also be excited with X-rays, ions, and other sources but the term Auger electron
spectroscopy, without additional qualifiers, is usually reserved for electron-beam-induced excitation. Where an
X-ray source is used, the Auger electron energies are referenced to the Fermi level (4.211) but, where an electron
beam is used, the reference can either be the Fermi level or the vacuum level (4.483). Spectra, conventionally, can
be presented in the direct (4.173) or differential (4.171) forms.
3.2
desorption electrospray ionization
DESI
method in which a mass spectrometer is used to measure the mass-to-charge quotient and abundance of
ionized entities emitted from a sample in air as a result of the bombardment by ionized solvent droplets
generated by pneumatically assisted electrospray ionization
Note 1 to entry: Water and methanol are often used as the solvents to create the droplets. Acids and alkalis are
added to control the solution pH.
Note 2 to entry: DESI is one of the few surface analysis methods designed to analyse materials without exposure
to vacuum. It is used for complex molecules, organic molecules, and biomolecules. In vivo analysis is claimed
to be possible.
3.3
dynamic SIMS
SIMS (3.17) in which the material surface (4.458) is sputtered at a sufficiently rapid rate that the original
surface cannot be regarded as undamaged during the analysis
Note 1 to entry: Dynamic SIMS is often simply termed SIMS.
16 2
Note 2 to entry: The ion areic dose (4.175) during measurement is usually more than 10 ions/m .
3.4
elastic peak electron spectroscopy
EPES
method in which an electron spectrometer (4.190) is used to measure the energy, intensity, and/or energy
broadening distribution of quasi-elastically scattered electrons from a solid or liquid surface (4.458)
Note 1 to entry: See recoil effect (4.366) and reflection electron energy loss spectroscopy (REELS) (3.16).
Note 2 to entry: An electron beam in the energy range 100 eV to 3 keV is often used for this kind of spectroscopy.
Note 3 to entry: In general, electron sources with energy spreads that are less than 1 eV are required to provide
adequate information.
Note 4 to entry: EPES is often an auxiliary method of AES (3.1) and REELS (3.16), providing information on the
composition of the surface layer. EPES is suitable for the experimental determination of the inelastic mean free path
(4.243), the electron differential elastic scattering cross section (4.127), and the surface excitation parameter (4.461).
4 © ISO 2013 – All rights reserved

---------------------- Page: 9 ----------------------
ISO 18115-1:2013(E)

3.5
DEPRECATED: electron spectroscopy for chemical analysis
DEPRECATED: ESCA
method encompassing both AES (3.1) and XPS (3.23)
Note 1 to entry: The term ESCA has fallen out of use as, in practice, it was only used to describe situations more
clearly defined by the term X-ray photoelectron spectroscopy (XPS). Since 1980, the latter term has been preferred.
3.6
fast atom bombardment mass spectrometry
FABMS
DEPRECATED: FAB
method in which a mass spectrometer is used to measure the mass-to-charge quotient and abundance
of secondary ions (4.406) emitted from a sample as a result of the bombardment by fast neutral atoms
3.7
G-SIMS
variant of static SIMS (3.20) in which the intensities for each mass in two spectra from the same area,
recorded with different beam energies or different bombarding ions, are ratioed to each other and the
result is used to scale one of the spectra to generate a new spectrum
Note 1 to entry: As with static SIMS, the ion areic dose (4.175) during measurement is restricted to less than
16 2
10 ions/m to an extent that depends on both the material of the sample and the size of the molecular fragments
(4.302) being analysed.
Note 2 to entry: The G-SIMS spectrum enables the mass of whole molecules on the surface (4.458) to be determined
more readily than in static SIMS.
Note 3 to entry: The “G” in G-SIMS originally indicated the gentleness of the process generated.
3.8
glow discharge mass spectrometry
GDMS
method in which a mass spectrometer is used to measure the mass-to-charge quotient and abundance
of ions from a glow discharge (4.228) generated at a surface (4.458)
3.9
glow discharge optical emission spectrometry
GDOES
method in which an optical emission spectrometer is used to measure the wavelength and intensity of
light emitted from a glow discharge (4.228) generated at a surface (4.458)
3.10
glow discharge spectrometry
GDS
method in which a spectrometer is used to measure relevant intensities emitted from a glow discharge
(4.228) generated at a surface (4.458)
Note 1 to entry: This is a general term that encompasses GDOES (3.9) and GDMS (3.8).
3.11
ion beam analysis
IBA
method designed to elucidate composition and structure of the near-surface atomic layers of a solid
material, in which principally monoenergetic, singly charged probe ions (4.349) scattered from the surface
(4.458) are detected and recorded as a function of their energy or angle of scattering (4.18), or both
Note 1 to entry: LEIS(S) (3.12), MEIS(S) (3.13), and RBS (3.15) are all forms of IBA in which the probe ion energies
are typically in the ranges 0,1 keV to 10 keV, 100 keV to 200 keV, and 1 MeV to 2 MeV, respectively. These
classifications represent three ranges in which fundamentally different physics is involved.
© ISO 2013 – All rights reserved 5

---------------------- Page: 10 ----------------------
ISO 18115-1:2013(E)

3.12
low-energy ion scattering spectrometry
LEIS(S)
method designed to elucidate composition and structure of the very outermost atomic layers of a solid
material, in which principally monoenergetic, singly charged probe ions (4.349) scattered from the surface
(4.458) are detected and recorded as a function of their energy or angle of scattering (4.18), or both
Note 1 to entry: LEIS(S) is a form of IBA (3.11) in which the probe ions, typically He or Ne, have energies in the
range 0,1 keV to 10 keV.
Note 2 to entry: The acronym usually has only one “S”.
3.13
medium-energy ion scattering spectrometry
MEIS(S)
method designed to elucidate composition and structure of the outermost atomic layers of a solid material,
in which principally monoenergetic, singly charged probe ions (4.349) scattered from the surface (4.458)
are detected and recorded as a function of their energy or angle of scattering (4.18), or both
Note 1 to entry: MEIS is a form of IBA in which the probe ions, typically protons, have energies in the range
100 keV to 200 keV.
Note 2 to entry: By using channelling (4.94) and aligning the incident-ion beam along a crystal axis, the scattering
from the substrate can be suppressed so that enhanced signal quality and visibility are obtained for amorphous
overlayers. By further aligning the detector along a second crystal axis, the double-alignment mode, the scattering
from the substrate can be further suppressed, improving the signal quality and visibility for amorphous overlayers
to a high level.
Note 3 to entry: In some cases, an angle-sensitive detector is used that allows extensive structure and depth
profile (4.350) information to be obtained.
Note 4 to entry: The acronym usually has only one “S”.
3.14
matrix-assisted laser desorption/ionization mass spectrometry
MALDI
method in which a time of flight (4.473) mass spectrometer is used to measure the mass-to-charge ratio
(4.296) and abundance of ions emitted, as a result of a short pulse of laser illumination, from a sample
whose analyte is contained in an ion-assisting matrix
Note 1 to entry: The matrix used for assisting the ion emission needs a strong absorbance at the laser
wavelength and a low enough mass to be sublimable. Examples of matrices for 337 nm wavelength laser light
are 2,5-dihydroxybenzoic acid (DHB), 3,5-dimethoxy-4-hydroxycinnamic acid (sinapinic acid), and α-cyano-4-
hydroxycinnamic acid (CHCA).
Note 2 to entry: MALDI is used to analyse non-volatile polar biological and organic macromolecules as well as
polymers to masses of over 3 000 kDa.
3.15
Rutherford backscattering spectrometry
RBS
method designed to elucidate composition and structure of layers at the surface (4.458) of a solid material,
in which principally monoenergetic, singly charged probe ions (4.349) scattered from the surface with
a Rutherford cross section (4.133) are detected and recorded as a function of their energy or angle of
scattering (4.18), or both
Note 1 to entry: RBS is a form of IBA (3.11) in which the probe ions, typically He but sometimes H, have energies in
the range 1 MeV to 2 MeV. In its traditional form, a solid-state energy-dispersive detector is used. In the form of high-
resolution RBS, the energy can be reduced to 300 keV and a high-resolution (ion optical) spectrometer can be used.
Note 2 to entry: By using channelling (4.94)and aligning the incident-ion beam along a crystal axis, the scattering from
the substrate can be suppressed so that enhanced signal quality and visibility are obtained for amorphous overlayers.
6 © ISO 2013 – All rights reserved

---------------------- Page: 11 ----------------------
ISO 18115-1:2013(E)

3.16
reflection electron energy loss spectroscopy
REELS
method in which an electron spectrometer (4.190) is used to measure the energy distribution of electrons
quasi-elastically scattered by atoms at or in a surface layer and the associated electron energy loss
spectrum (4.197)
Note 1 to entry: See elastic peak electron spectroscopy (3.4) (EPES)
3.17
secondary-ion mass spectrometry
SIMS
method in which a mass spectrometer is used to measure the mass-to-charge quotient and abundance of
secondary ions (4.406) emitted from a sample as a result of bombardment by energetic ions
Note 1 to entry: See dynamic SIMS (3.3), static SIMS (3.20), and G-SIMS (3.7).
Note 2 to entry: SIMS is, by convention, generally classified as dynamic, in which the material surface layers are
continually removed as they are being measured, and static, in which the ion areic dose (4.175) during measurement
16 2
is restricted to less than 10 ions/m in order to retain the surface (4.458) in an essentially undamaged state.
3.18
small-angle X-ray scattering
SAXS
method in which the elastically scattered intensity of X-rays is measured for small-angle deflections
Note 1 to entry: The angular scattering is usually measured within the range 0,1° to 10°. This provides structural
information on macromolecules as well as periodicity on length scales typically larger than 5 nm and less than
200 nm for ordered or partially ordered systems.
Note 2 to entry: Wide-angle X-ray scattering (WAXS) is an analogous technique, similar to X-ray crystallography,
in which scattering at larger angles, which is sensitive to periodicity on smaller length scales, is measured.
Note 3 to entry: The X-ray source can be a synchrotron, in which case the term synchrotron radiation (4.465)
small-angle X-ray scattering (SRXAS) is occasionally encountered.
3.19
sputtered neutral mass spectrometry
SNMS
method in which a mass spectrometer is used to measure the mass-to-charge quotient and abundance of
secondary ionized neutral species emitted from a sample as a result of particle bombardment
Note 1 to entry: The neutral species can be detected by using plasma (4.337), electron, or photon-ionization methods.
3.20
static SIMS
SIMS (3.17) in which the material surface (4.458) is sputtered at a sufficiently low rate that the original
surface is insignificantly damaged during the analysis
Note 1 to entry: See dynamic SIMS (3.3).
16 2
Note 2 to entry: The ion areic dose (4.175)during measurement is restricted to less than 10 ions/m to an extent
that depends on both the material of the sample and the size of the molecular fragments (4.302) being analysed.
3.21
total reflection X-ray fluorescence spectroscopy
TXRF
method in which an X-ray spectrometer is used to measure the energy distribution of fluorescence
(4.219) X-rays emitted from a surface (4.458) irradiated by primary X-rays under the condition of total
reflection (4.475)
© ISO 2013 – All rights reserved 7

---------------------- Page: 12 ----------------------
ISO 18115-1:2013(E)

3.22
ultraviolet photoelectron spectroscopy
UPS
method in which an electron spectrometer (4.190) is used to measure the energy distribution of
photoelectrons emitted from a surface (4.458) irradiated by ultraviolet photons
Note 1 to entry: Ultraviolet sources in common use include various types of discharges that can generate
the resonance lines of various gases (e.g. the He I and He II emission lines at energies of 21,2 eV and 40,8 eV,
respectively). For variable energies, synchrotron radiation (4.465) is used.
3.23
X-ray photoelectron spectroscopy
XPS
method in which an electron spectrometer (4.190) is used to measure the energy distribution of
photoelectrons and Auger electrons (4.37) emitted from a surface (4.458) irradiated by X-ray photons
Note 1 to entry: X-ray sources in common use are unmonochromated Al Kα and Mg Kα X-rays at 1 486,6 eV, and
1 253,6 eV, respectively. Modern instruments also use monochromated Al Kα X-rays. Some instruments make use
of various X-ray sources with other anodes (4.27) or of synchrotron radiation (4.465).
4 Definitions of terms for surface analysis
4.1
absorption coefficient, linear
linear attenuation coefficient
4.2
absorption coefficient, mass
attenuation coefficient, mass
quantity μ/ρ in the expression, (μ/ρ)Δ(ρx), for the fraction of a parallel beam of specified
particles or radiation removed in passing through a thin layer of mass thickness Δ(ρx) of a substance in
the limit as Δ(ρx) approaches zero, where Δ(ρx) is measured in the direction of the beam
Note 1 to entry: See attenuation length (4.34).
Note 2 to entry: The mass density of the substance is ρ and x is the distance in the direction of the beam.
Note 3 to entry: The intensity or number of particles in the beam decays as exp(−μx) with the distance x.
Note 4 to entry: The mass attenuation (absorption) coefficient is the quotient of the linear attenuation (absorption)
coefficient by the mass density of the substance.
4.3
abundance sensitivity
ratio of the maximum ion current recorded at a mass m to the ion current arising from the same
species recorded at an adjacent mass (m ± 1)
[SOURCE: IUPAC]
4.4
adventitious carbon referencing
determining the charging potential (4.103) of a particular sample from a comparison of the
experimentally determined C 1s binding energy (4.82), arising from adsorbed hydrocarbons on the
sample, with a standard binding energy value
Note 1 to entry: See Fermi level referencing (4.212) and internal carbon referencing (4.257).
Note 2 to entry: A nominal value of 285,0 eV is often used for the binding energy of the relevant C 1s peak, although
some analysts prefer specific values in the range 284,6 eV to 285,2 eV, depending on the nature of the substrate.
This method does not determine the true charging potential (4.103) since the true binding energy of the adsorbed
hydrocarbons is not known.
8 © ISO 2013 – All rights reserved

---------------------- Page: 13 ----------------------
ISO 18115-1:2013(E)

Note 3 to entry: Different sample charging (4.392) potentials can occur on different areas on the surface (4.458),
or at different depths, arising, for example, from sample inhomogeneities or non-uniform intensity of the incident-
radiation flux (4.221).
4.5
afterglow
luminescence of the decaying plasma (4.337) present in a glow discharge (4
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.