Determination of explosion limits of gases and vapours

This standard specifies a test method to determine the explosion limits of gases, completely evaporated liquids (vapours) and their mixtures mixed with air. An air/inert gas mixture (volume fraction of the oxygen <21%) can be used as oxidizer in the place of air. This standard applies to gases, vapours and their mixtures at atmospheric pressure and at temperatures from ambient temperature to 200 °C.

Bestimmung der Explosionsgrenzen von Gasen und Dämpfen

Diese Europäische Norm legt zwei Verfahren (Methode T und Methode B) zum Bestimmen der Explosionsgrenzen von Gasen, Dämpfen und deren Gemischen in Mischung mit Luft fest. An Stelle von Luft kann als Oxidator auch ein Luft/Inertgas-Gemisch (Volumenanteil des Sauerstoffs < 21 %) verwendet werden. In dieser Europäischen Norm subsumiert der Ausdruck ?Luft" auch solche Luft/Inertgas-Gemische.
Dieses Verfahren gilt für Gase, Dämpfe und deren Gemische bei Umgebungsdruck und Temperaturen von Raumtemperatur bis 200 °C.

Determination des limites d'explosivité des gaz et des vapeurs

La présente Norme européenne spécifie deux méthodes d'essai (méthode T et méthode B) pour déterminer les limites d'explosivité dans l'air des gaz, des vapeurs et de leurs mélanges. Un mélange d'air et d'un gaz inerte (fraction volumique de l'oxygene inférieure a 21 %) peut etre utilisé comme comburant, au lieu de l'air. Dans cette norme, le terme " air " comprend également les mélanges air/gaz inerte.
La présente norme s'applique aux gaz, aux vapeurs et a leurs mélanges a la pression atmosphérique et a des températures comprises entre la température ambiante et 200 °C.

Ugotavljanje mej eksplozivnosti plinov in hlapov

General Information

Status
Withdrawn
Publication Date
30-Nov-2003
Withdrawal Date
22-Nov-2012
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
22-Nov-2012
Due Date
15-Dec-2012
Completion Date
23-Nov-2012

Relations

Buy Standard

Standard
EN 1839:2003
English language
29 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 1839:2003
01-december-2003
Ugotavljanje mej eksplozivnosti plinov in hlapov
Determination of explosion limits of gases and vapours
Bestimmung der Explosionsgrenzen von Gasen und Dämpfen
Determination des limites d'explosivité des gaz et des vapeurs
Ta slovenski standard je istoveten z: EN 1839:2003
ICS:
13.230 Varstvo pred eksplozijo Explosion protection
SIST EN 1839:2003 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

SIST EN 1839:2003

---------------------- Page: 2 ----------------------

SIST EN 1839:2003
EUROPEAN STANDARD
EN 1839
NORME EUROPÉENNE
EUROPÄISCHE NORM
September 2003
ICS 13.230
English version
Determination of explosion limits of gases and vapours
Determination des limites d'explosivité des gaz et des Bestimmung der Explosionsgrenzen von Gasen und
vapeurs Dämpfen
This European Standard was approved by CEN on 7 August 2003.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CEN member into its own language and notified to the Management Centre has the same status as the official
versions.
CEN members are the national standards bodies of Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece,
Hungary, Iceland, Ireland, Italy, Luxembourg, Malta, Netherlands, Norway, Portugal, Slovakia, Spain, Sweden, Switzerland and United
Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre: rue de Stassart, 36  B-1050 Brussels
© 2003 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN 1839:2003 E
worldwide for CEN national Members.

---------------------- Page: 3 ----------------------

SIST EN 1839:2003
EN 1839:2003 (E)
Contents page
Foreword.3
1 Scope .5
2 Normative references .5
3 Terms and definitions.5
4 Test Methods.6
4.1 Method T (”tube” method) .6
4.1.1 Principle.6
4.1.2 Reagents and Materials.7
4.1.3 Apparatus .7
4.1.4 Preparation of the test mixture.10
4.1.5 Procedure .10
4.2 Method B ("bomb" method).11
4.2.1 Principle.11
4.2.2 Reagents and materials.11
4.2.3 Apparatus .11
4.2.4 Preparation of the test mixture.14
4.2.5 Procedure .15
4.3 Expression of results .16
4.4 Test report .16
Annex A (informative) Examples to describe flame detachment .18
Annex B (informative) Example for a recommended evaporator equipment.19
Annex C (normative) Safety advice .21
Annex D (informative) Example for a form expressing the results .22
Annex E (normative) Verification.23
Annex F (informative) Conversion of the values for the explosion limits .25
Annex ZA (informative) Clauses of this European Standard addressing essential requirements or other
provisions of EU Directives .29
2

---------------------- Page: 4 ----------------------

SIST EN 1839:2003
EN 1839:2003 (E)
Foreword
This document (EN 1839:2003) has been prepared by Technical Committee CEN/TC 305 “Potentially explosive
atmospheres — Explosion prevention and protection”, the secretariat of which is held by DIN.
This European Standard shall be given the status of a national standard, either by publication of an identical text or
by endorsement, at the latest by March 2004, and conflicting national standards shall be withdrawn at the latest by
March 2004.
This document has been prepared under a mandate given to CEN by the European Commission and the European
Free Trade Association, and supports essential requirements of EU Directive(s).
For relationship with EU Directive(s), see informative Annex ZA, B, C or D, which is an integral part of this
document.
In this European Standard the annexes A, B, D and F are informative and the annexes C to E are normative.
According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following
countries are bound to implement this European Standard: Austria, Belgium, Czech Republic, Denmark, Finland,
France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Luxembourg, Malta, Netherlands, Norway, Portugal,
Slovakia, Spain, Sweden, Switzerland and the United Kingdom.
3

---------------------- Page: 5 ----------------------

SIST EN 1839:2003
EN 1839:2003 (E)
Introduction
To avoid the hazard of explosion, an appropriate measure is to prevent explosive mixtures of gases and/or vapours
being formed. To do so, the explosion limits (also known as ”flammability limits”) of the flammable substance have
to be known. These limits depend mainly on:
the properties of the flammable substance,
temperature and pressure,
size and shape of the test vessel,
ignition source (type, energy),
the criterion for self-propagating combustion.
To obtain reliable and comparable results it is therefore necessary to standardize conditions (apparatus and
procedure) under which the explosion limits are to be determined. It is however not possible to propose one
method that is suitable for all substances. The use of an apparatus that can also be used for the determination of
other explosion characteristics would also be a benefit. To these practical reasons the standard offers two
methods, tube method (method T) and bomb method (method B). In general the tube method gives a wider
explosion range. Differences in the explosion limits determined by the two methods can differ by up to 10 %
relative.
4

---------------------- Page: 6 ----------------------

SIST EN 1839:2003
EN 1839:2003 (E)
1 Scope
This European Standard specifies two test methods (method T and method B) to determine the explosion limits of
gases, vapours and their mixtures mixed with air. An air/inert gas mixture (volume fraction of the oxygen < 21 %)
can be used as oxidizer instead of air. In this standard the term “air” includes such air/inert mixtures.
This European Standard applies to gases, vapours and their mixtures at atmospheric pressure and at temperatures
from ambient temperature to 200 °C.
2 Normative references
This European Standard incorporates by dated or undated reference, provisions from other publications. These
normative references are cited at the appropriate places in the text and the publications are listed hereafter. For
dated references, subsequent amendments to or revisions of any of these publications apply to this European
Standard only when incorporated in it by amendment or revision. For undated references the latest edition of the
publication referred to applies (including amendments).
EN 1127-1:1997, Explosive atmospheres — Explosion prevention and protection — Part 1: Basic concepts and
methodology.
3 Terms and definitions
For the purposes of this European Standard, the following terms and definitions apply.
3.1
explosion range
range of the concentration of a flammable substance in air, within which an explosion can occur, respectively range
of the concentration of a flammable substance in mixture with air/inert gas, within which an explosion can occur,
determined under specified test conditions
NOTE Adapted from EN 1127-1.
3.2
explosion limits
limits of the explosion range
[EN 1127-1:1997, 3.7]
3.3
lower explosion limit
LEL
lower limit of the explosion range
3.4
upper explosion limit
UEL
upper limit of the explosion range
[EN 1127-1:1997]
NOTE 1 The values of the explosion limits vary with type and amount of inert gas added.
NOTE 2 In English literature the term ”flammability limits” is also used for describing the limits determined according to this
standard. The term ”explosion limits” is chosen here to designate the same safety characteristics. This is done, because new
regulations now use this term.
5

---------------------- Page: 7 ----------------------

SIST EN 1839:2003
EN 1839:2003 (E)
3.5
explosion criterion — flame detachment
in method T, the criterion of an explosion (self-propagating combustion) is the upward movement of the flame from
the spark gap for at least 100 mm. This process which is referred to as "flame detachment" is visually observed
(examples see annex A). If a halo reaches the top of the tube or has at least a height of 240 mm this shall be also
counted as an ignition
NOTE During the duration of the ignition spark, test mixtures whose test substance content lies slightly outside the
explosion range exhibit a luminous phenomenon (which is referred to as ”halo”) above the spark gap which does not, however,
detach from the latter (see annex A). In the case of certain test substances (e.g. halogenated hydrocarbons), this luminous
phenomenon occupies a large portion of the test vessel. The formation of an halo exclusively is not yet considered to be an
ignition of the test mixture unless it reaches the top of the tube or a minimum height of 240 mm.
3.6
explosion criterion — pressure rise
in method B, the criterion of an explosion (self-propagating combustion) is the amount of the measured explosion
overpressure. The test mixture is considered explosive if the measured explosion overpressure is equal to or
greater than the overpressure created by the ignition source alone in air, plus (5 ± 0,1) % of the initial pressure
3.7
vapour
gaseous phase emanating or being emanated from a liquid. If not otherwise mentioned, the term ”gas” in this
standard also includes such vapours but not mists
NOTE The special features of vapours (possible condensation, difference in mixture composition between liquid and
vapour phase, change in mixture composition when taking out vapour phase from a closed volume) have to be considered
carefully when determining explosion limits for vapours of flammable liquids.
3.8
oxidizer
air or an air/inert gas mixture (volume fraction of the oxygen < 21 %)
3.9
sample
a substance or a mixture of substances for which explosion limits are to be determined
3.10
test substance
the sample in the gaseous state — in the case of liquid samples, after complete evaporation
3.11
test mixture
the mixture of test substance and air
4 Test Methods
4.1 Method T (”tube” method)
4.1.1 Principle
The test mixture is led through the cylindrical test vessel from below towards the top until the mixture being in the
test vessel before has been completely replaced. Then in the quiescent test mixture an ignition is initiated using a
series of induction sparks, and it is observed whether or not a flame detachment occurs. The test substance
content of the test mixture is varied step by step until the LEL or the UEL (explosion criterion — flame detachment)
have been determined or until it is certain that there exists no explosion range.
6

---------------------- Page: 8 ----------------------

SIST EN 1839:2003
EN 1839:2003 (E)
4.1.2 Reagents and Materials
air: the air shall be free of water and oil. If synthetic air is used, it has to be stated in the report.
inert gases: the purity of the inert, or the mixture of inerts, shall be 99,8 % mol. or better. If a mixture of inerts is
used, the composition of the mixture shall be stated in the test report.
flammable substance: the flammable substance may be:
a single substance or a defined mixture of substances,
a process sample (of known or unknown composition).
When a single substance or a mixture of substances is used, the purity of each substance shall be 99,8 % mol.
or better. In the case of a mixture of substances or a process sample of known composition, the precision of
the composition shall be stated in the test report. In the case of a process sample of unknown composition, the
sample shall be defined as well as possible (e.g. process conditions).
If the flammable gas is derived from a liquid containing more than one component, the gas phase composition
can differ from the composition of the liquid phase and when large quantities of the gas are drawn off, the
composition of both the liquid and gas phases can change with time. For these reasons, the test sample shall
be taken from the liquid phase.
heat-resistant, chemically inert material for gaskets and adhesive mountings.
Sample containers are to be kept closed before and after sampling to avoid alterations of the sample (e.g. escape
of volatile constituents in the case of mixtures). If the sample consists of a gas mixture which is removed from a
container containing a liquid phase, take into account that the composition of the gas and the liquid phase is
different. It is recommended to take the test substance from the liquid phase.
4.1.3 Apparatus
4.1.3.1 General
The apparatus (see figure 1) consists of:
test vessel;
ignition source;
equipment for preparing the test mixture;
temperature regulating device;
safety equipment.
4.1.3.2 Test vessel
The test vessel is an upright cylindrical test vessel made of glass or another transparent material (e.g.
polycarbonate) with an inner diameter of (80 ± 2) mm and a minimum length of 300 mm. An inlet pipe for the test
mixture with three-way cock, has to be placed at the bottom, an outlet pipe and pressure venting in the upper part.
Bottom and top may be made of other material. It has to be however free of any catalytic effect and resistant to
corrosion by the test mixture as well as the reaction products.
4.1.3.3 Ignition source
A series of induction sparks between two electrodes is used as the ignition source.
7

---------------------- Page: 9 ----------------------

SIST EN 1839:2003
EN 1839:2003 (E)
The electrodes shall end (60 ± 1) mm above the bottom of the test vessel.
Stainless steel is a suitable material for the electrodes. The electrodes shall be pointed rods with a diameter of
maximum 4 mm. The angle of the tips shall be (60 ± 3)°. The distance between the tips shall be (5 ± 0,1) mm. The
electrodes shall be mounted in the vessel so that they are gas tight at the highest pressures generated during the
test. The mounting shall be resistant to heat and the test mixture, and provide adequate electrical resistance from
the test vessel body.
A high voltage transformer, with a root mean square of 13 kV to 16 kV and a short circuit current of 20 mA to
30 mA, shall be used for producing the ignition spark. The primary winding of the high voltage transformer shall be
connected to the mains via a timer set to the required discharge time.
The spark discharge time shall be adjusted to 0,2 s. If a spark discharge time of 0,2 s does not result in ignition of
the test mixture, the test may be repeated with a spark discharge time up to 0,5 s.
The power of the induction sparks depends on the gas mixture and its pressure. In air at atmospheric conditions
according to calorimetric and electric measurements such a source gives a spark with a power of approximately
10 W.
4.1.3.4 Equipment for preparing the test mixture
The test mixture is prepared by mixing flows of gaseous components. For doing so the following equipment is
necessary:
metering device for air, gaseous samples or additional inert gases (e.g. mass flow controller, volume flow
controller, metering pump for gases);
metering device for liquid samples (e.g. volumetric metering pumps);
evaporator equipment in the case of a liquid sample (for example see annex B);
mixing vessel for homogenizing the test mixture.
The metering devices and the equipment for preparing the test mixture has to be designed in such a way that the
uncertainty of measurement of the test substance content in the test mixture is not higher then the data given in
table 1.
8

---------------------- Page: 10 ----------------------

SIST EN 1839:2003
EN 1839:2003 (E)
Table 1 — Maximum permissible uncertainty of measurement for the amount of test substance in the test
mixture
molar amount of test substance maximum uncertainty of measurement
% %
relative absolute
± 10
2
> 2 ± 0,2
4.1.3.5 Temperature regulating device
For measurements at temperatures higher than ambient temperature, the apparatus has to be equipped with a
temperature regulating device. In such a case it has to be ensured that the temperature difference inside the test
vessel is not more than 10 K. This has to be proven when building up the apparatus and every time when parts are
renewed and with every verification.
Key
1 test vessel 7 timer
2 electrodes 8 facility for keeping the temperature
3 three-way cock 9 flammable substance
4 mixing vessel 10 air
5 metering devices 11 power supply
6 high-voltage transformer
Figure 1 — Scheme of the ‘tube’ apparatus for determining the explosion limits
9

---------------------- Page: 11 ----------------------

SIST EN 1839:2003
EN 1839:2003 (E)
4.1.3.6 Safety equipment
The safety advices according annex C have to be followed.
4.1.4 Preparation of the test mixture
When evaporating liquid samples, it has to be considered that the mixture composition of a gaseous phase in
equilibrium with a liquid phase (”vapour”) generally differs from the mixture composition of the liquid phase. Also the
mixture composition of the liquid and the vapour phase may change, when taking out substance from the vapour
phase. This has to be observed carefully when determining explosion limits for flammable liquids, when handling
liquid samples and when preparing test mixtures by evaporating liquid samples. To avoid mistakes, the method of
dynamic total evaporation is to be used. An example for a recommended evaporator equipment is described in
annex D. When liquids are metered, it has to be ensured that no bubbles form in any part carrying the liquid (e.g.
pipes). To achieve complete homogenization, the test mixture is led through a mixing vessel, preferably made of
glass. For a mixing vessel without built-in elements a volume of at least 600 ml is recommended. It is expedient to
introduce the test mixture tangentially. The mixing vessel is not necessary if the homogenization is already ensured
by the metering device. The temperature of the mixing vessel and of all parts carrying the test mixture is to be kept
constant so that the test substance cannot condense. It is recommended to heat the parts carrying the test mixture
together with the test vessel.
4.1.5 Procedure
If the explosion limits are determined at increased temperature, preheat the test vessel and all parts carrying the
test mixture to the required temperature. For liquid samples the temperature of the test mixture shall be at least
25 K higher than the condensation temperature. Prior to each ignition attempt, it has to be ensured that the
temperature in the test vessel differs not more than 5 K from the required value.
The determination of the explosion limits consists of a series of ignition attempts which are carried out with test
mixtures whose test substance content is varied.
For safety reasons, the ignition attempts are started with a test substance content in the test mixture which lies, if
possible, outside the explosion range to be expected.
NOTE Rough estimate of the amount of substance to start experiments:
For organic substances which consist exclusively of carbon, hydrogen and oxygen (with the exception of peroxides), the LEL
can be roughly estimated. At 20 °C, the LEL is in many cases at approximately half the test substance content of the
stoichiometric composition. The temperature dependence of the LEL has to be taken into account. Up to 200 °C, the LEL
decreases more or less linearly between 30 % and 50 % of the value estimated for 20 °C.
There is at present no method which is easy to apply for estimating the UEL.
Prior to each ignition attempt, the test vessel is purged with the test mixture. The purging volume has to be at least
ten times the volume of the test vessel. Subsequently, the test mixture by-passes the test vessel and is supplied
into the exhaust system; the inlet to the test vessel is sealed. Then an ignition is attempted with the induction spark
when test mixture is quiescent (6 s to 10 s), and it is observed whether a flame detaches from the ignition source
(see annex A).
It is recommended not to interrupt the production of test mixture during the ignition attempts, as after any new start,
it will take a certain time to produce a test mixture of constant composition even if the adjustment has not been
changed.
If an ignition is observed, the test substance content in the test mixture is iteratively varied until no further flame
detachment follows. Close to the explosion limits, the step size for the variation of the test substance content has to
be selected such that it is at the most equal to the relative deviation given in table 1. The test mixture with which a
further flame detachment just failed has to be used to carry out another four tests. The determination is terminated
when all five tests have taken place without flame detachment. If this is not the case, the test substance content
has to be varied again, i.e. for the determination of the LEL the test substance content has to be reduced by one
step size; for the UEL, it has to be increased by one step size. Subsequently, another five tests are to be carried
out. It is recommended to check the quantitative composition of the non-ignited test mixture flowing out of the test
vessel in order to recognise in time maladjustments of the metering devices or leakage.
10

---------------------- Page: 12 ----------------------

SIST EN 1839:2003
EN 1839:2003 (E)
4.2 Method B ("bomb" method)
4.2.1 Principle
The quiescent test mixture in a closed vessel (the bomb) is subjected to an ignition source. The overpressure given
by the ignition is measured and characterises the explosivity of the test mixture. The amount of test substance in
the test mixture is varied incrementally until the LEL or the UEL is determined.
4.2.2 Reagents and materials
air: the air shall be free of water and oil. If synthetic air is used, it has to be stated in the report;
inert gases: the purity of the inert, or the mixture of inerts, shall be 99,8 % mol. or better. If a mixture of inerts is
used, the composition of the mixture shall be stated in the test report;
flammable substance: the flammable substance may be:
a single substance or a defined mixture of substances;
a process sample (of known or unknown composition).
When a single substance or a mixture of substances is used, the purity of each substance shall be 99,8 % mol.
or better. In the case of a mixture of substances or a process sample of known composition, the precision of
the composition shall be stated in the test report. In the case of a process sample of unknown composition, the
sample shall be defined as well as possible (e.g. process conditions).
If the flammable gas is derived from a liquid containing more than one component, the gas phase composition
can differ from the composition of the liquid phase and when large quantities of the gas are drawn off, the
composition of both the liquid and gas phases can change with time. For these reasons, the test sample shall
be taken from the liquid phase.
heat-resistant, chemically inert material for gaskets and adhesive mountings.
Sample containers are to be kept closed before and after sampling to avoid alterations of the sample (e.g. escape
of volatile constituents in the case of mixtures). If the sample consists of a gas mixture which is removed from a
container containing a liquid phase, take into account that the composition of the gas and the liquid phase is
different. It is recommended to take the test substance from the liquid phase.
4.2.3 Apparatus
4.2.3.1 General
The apparatus consists of
a test vessel;
an ignition system;
equipment for preparing the test mixture;
a pressure measuring system for measuring the explosion overpressure;
a temperature measuring system;
safety equipment.
11

---------------------- Page: 13 ----------------------

SIST EN 1839:2003
EN 1839:2003 (E)
4.2.3.2 Test vessel
The test vessel shall be cylindrical or spherical. The internal volume of the test vessel shall be equal to or greater
3
than 0,005 m . If a cylindrical vessel is used, the length to diameter ratio shall be between 1 and 1,5.
The test vessel and any equipment (valves, ignition source, transducer.) fitted on the vessel shall be designed to
withstand a maximum overpressure of at least 15 bar.
The vessel shall be made of stainless steel or any material free of any catalytic effect and resistant to corrosion
from the initial gas mixture and the products of combustion.
The test vessel shall be fitted with sufficient ports to allow filling, evacuating and purging.
If the test mixture is prepared inside the test vessel by partial pressures, it is recommended to disconnect the
pressure measuring system used via a valve to protect it during the ignition trials.
The temperature tracer (thermocouple or PT 100) used for measuring the temperature of the test mixture inside the
test vessel, has to be mounted such that propagation for the flame is not hindered.
4.2.3.3 Ignition system
4.2.3.3.1 General
The ignition system shall be positioned in the centre of the test vessel. Ignition devices are the series of induction
sparks or the fusing wire. In the test report, the type of ignition source used shall be stated.
4.2.3.3.2 Induction spark
A series of induction sparks between two electrodes is used as the ignition source.
Stainless steel is a suitable material for the electrodes. The electrodes shall be positioned in the centre of the
vessel. They shall be pointed rods with a maximum diameter of 4 mm. The angle of the tips shall be (60 ± 3)°. The
distance between the tips shall be (5 ± 0,1) mm. The electrodes shall be mounted in the vessel so that they are gas
tight at the highest pressures generated during the test. The mounting shall be resistant to heat and the test
mixture, and provide adequate electrical resistance from the test vessel body.
A high voltage transformer, with a root mean square of 13 kV to 16 kV and a short circuit current of 20 mA to
30 mA, shall be used for producing the ignition spark. The primary winding of the high voltage transformer shall be
connected to the mains via a timer set to the required discharge time.
The spark discharge time shall be adjusted to 0,2 s. If a spark discharge time of 0,2 s does not result in ignition of
the test mixture, the test may be repeated with a spark discharge time up to 0,5 s.
The power of the spark depends on the gas mixture and its pressure. In air at normal atmospheric conditions, such
a source shall give a
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.