Sterilization of medical devices - Low temperature steam and formaldehyde - Requirements for development, validation and routine control of a sterilization process for medical devices

1.1   Inclusions
1.1.1   This European Standard specifies requirements for the development, validation and routine control of a Low Temperature Steam and Formaldehyde (LTSF) sterilization process for medical devices.
NOTE   Although the scope of this standard is limited to medical devices, it specifies requirements and provides guidance that may be applicable to other products and equipment.
1.1.2   This European Standard is intended to be applied by process developers, manufacturers of sterilization equipment, manufacturers of medical devices to be sterilized and the organizations with responsibility for sterilizing medical devices. (See EN ISO 14937:2000, Table E.1)
1.1.3   This European Standard covers sterilization processes which use a mixture of low temperature steam and formaldehyde as sterilant, and which are working below ambient pressure only.
1.2   Exclusions
1.2.1   Sterilization processes validated and controlled in accordance with the requirements of this standard should not be assumed to be effective in inactivating the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeld-Jakob disease. Specific recommendations have been produced in particular countries for the processing of materials potentially contaminated with these agents.
1.2.2   This standard does not specify requirements for designating a medical device as "STERILE". Such requirements are given in EN 556-1.
1.2.3   This standard does not specify a quality management system for the control of all stages of production of medical devices.
NOTE   Attention is drawn to the standards for quality management systems (see EN ISO 13485) that control all stages of production or reprocessing of medical devices including the sterilization process. It is not a requirement of this standard to have a complete quality management system during manufacture or reprocessing, but those elements of such a system that are required are normatively referenced at appro

Sterilisation von Medizinprodukten - Niedertemperatur-Dampf-Formaldehyd - Anforderungen an die Entwicklung, Validierung und Routineüberwachung von Sterilisationsverfahren für Medizinprodukte

Stérilisation des dispositifs médicaux - Vapeur d'eau a basse température et au formaldéhyde - Exigences pour le développement, la validation et le contrôle de routine des procédés de stérilisation

1.1   Inclusions
1.1.1   La présente Norme européenne spécifie les exigences relatives a la mise au point, a la validation et au contrôle de routine d'un procédé de stérilisation a la vapeur et au formaldéhyde a température basse (LTSF) pour les dispositifs médicaux.
NOTE   Bien que le domaine d'application de la présente norme se limite aux dispositifs médicaux, elle spécifie les exigences et fournit un guide qui peut etre applicable a d'autres produits et équipements.
1.1.2   La présente Norme européenne est destinée a etre appliquée par les personnes chargées de la mise au point des procédés, les fabricants d’équipements de stérilisation, les fabricants de dispositifs médicaux destinés a etre stérilisés et les organisations responsables de la stérilisation des dispositifs médicaux (voir EN ISO 14937:2000, Tableau E.1).
1.1.3   La présente Norme européenne couvre les procédés de stérilisation employant un mélange de vapeur et de formaldéhyde a température basse en tant que stérilisant et fonctionnant uniquement en dessous de la pression ambiante.
1.2   Exclusions
1.2.1   Il convient que les procédés de stérilisation validés et contrôlés conformément aux exigences de la présente norme ne soient pas supposés efficaces pour l'inactivation des agents responsables des encéphalopathies spongiformes telles que la tremblante du mouton, l'encéphalopathie spongiforme bovine et la maladie de Creutzfeld-Jacob. Des recommandations spécifiques ont été formulées dans certains pays pour le traitement des matériaux potentiellement contaminés par ces agents.
1.2.2   La présente norme ne spécifie pas d’exigence pour qualifier un dispositif médical de « STERILE ». De telles exigences sont fournies dans l’EN 556-1.
1.2.3   La présente norme ne spécifie pas de systeme de management de la qualité pour le contrôle de toutes les étapes de production des dispositifs médicaux.

Sterilizacija medicinskih pripomočkov - Para nizke temperature in formaldehid - Zahteve za razvoj, validacijo in rutinsko kontrolo sterilizacijskih postopkov za medicinske pripomočke

General Information

Status
Withdrawn
Publication Date
15-Aug-2007
Withdrawal Date
30-Aug-2011
Technical Committee
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
31-Aug-2011
Due Date
23-Sep-2011
Completion Date
31-Aug-2011

Relations

Buy Standard

Standard
EN 15424:2007
English language
46 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Sterilization of medical devices - Low temperature steam and formaldehyde - Requirements for development, validation and routine control of a sterilization process for medical devicesStérilisation des dispositifs médicaux - Vapeur d'eau a basse température et au formaldéhyde - Exigences pour le développement, la validation et le contrôle de routine des procédés de stérilisationSterilisation von Medizinprodukten - Niedertemperatur-Dampf-Formaldehyd - Anforderungen an die Entwicklung, Validierung und Routineüberwachung von Sterilisationsverfahren für MedizinprodukteTa slovenski standard je istoveten z:EN 15424:2007SIST EN 15424:2007en,fr,de11.080.01Sterilizacija in dezinfekcija na splošnoSterilization and disinfection in generalICS:SLOVENSKI
STANDARDSIST EN 15424:200701-september-2007







EUROPEAN STANDARDNORME EUROPÉENNEEUROPÄISCHE NORMEN 15424April 2007ICS 11.080.01 English VersionSterilization of medical devices - Low temperature steam andformaldehyde - Requirements for development, validation androutine control of a sterilization process for medical devicesStérilisation des dispositifs médicaux - Développement,validation et contrôle de routine des procédés destérilisation - Vapeur d'eau à basse température et auformaldéhydeSterilisation von Medizinprodukten - Niedertemperatur-Dampf-Formaldehyd - Anforderungen an die Entwicklung,Validierung und Routineüberwachung vonSterilisationsverfahren für MedizinprodukteThis European Standard was approved by CEN on 17 February 2007.CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this EuropeanStandard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such nationalstandards may be obtained on application to the CEN Management Centre or to any CEN member.This European Standard exists in three official versions (English, French, German). A version in any other language made by translationunder the responsibility of a CEN member into its own language and notified to the CEN Management Centre has the same status as theofficial versions.CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland,France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal,Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.EUROPEAN COMMITTEE FOR STANDARDIZATIONCOMITÉ EUROPÉEN DE NORMALISATIONEUROPÄISCHES KOMITEE FÜR NORMUNGManagement Centre: rue de Stassart, 36
B-1050 Brussels© 2007 CENAll rights of exploitation in any form and by any means reservedworldwide for CEN national Members.Ref. No. EN 15424:2007: E



EN 15424:2007 (E) 2 Contents Page Foreword.4 Introduction.5 1 Scope.7 1.1 Inclusions.7 1.2 Exclusions.7 2 Normative references.8 3 Terms and definitions.8 4 Quality Management System Elements.14 4.1 Documentation.14 4.2 Management responsibility.14 4.3 Product realization.14 4.4 Control of non-conforming product.15 5 Sterilizing agent characterization.15 5.1 General.15 5.2 Sterilizing agent.15 5.3 Microbicidal effectiveness.15 5.4 Material effects.15 5.5 Environmental considerations.15 6 Process and equipment characterization.16 6.1 General.16 6.2 Process.16 6.3 Equipment.16 7 Product definition.17 8 Process definition.18 9 Validation.19 9.1 General.19 9.2 Installation qualification.19 9.2.1 General.19 9.2.2 Installation.19 9.2.3 Equipment.20 9.3 Operational qualification.20 9.4 Performance qualification.20 9.4.1 General.20 9.4.2 Performance qualification – physical.21 9.4.3 Performance qualification – microbiological.21 9.4.4 Performance qualification – desorption and drying.22 9.5 Review and approval of validation.22 10 Routine monitoring and control.23 10.1 General.23 10.2 Biological indicators.23 10.3 Chemical indicators.23 10.4 Records.23 11 Product release from sterilization.24 12 Maintaining process effectiveness.24 12.1 General.24 12.2 Maintenance of equipment.24 12.3 Requalification.24 12.4 Assessment of change.25 Annex A (normative)
Process definition based on inactivation of reference microorganisms and knowledge of bioburden on product items to be sterilized.26



EN 15424:2007 (E) 3 A.1 General.26 A.2 Procedure.26 Annex B (normative)
Process definition based on inactivation of reference microorganisms.27 B.1 General.27 B.1.1 Overkill approach.27 B.1.2 Penetration characteristics into medical devices.27 B.2 Test procedure.28 B.2.1 General.28 B.2.2 Biological indicators.28 B.2.3 Test systems.28 B.2.4 Load configuration.28 B.2.5 Testing.28 Annex C (informative)
Guidance on application of this European Standard.30 C.1 Scope.30 C.2 Normative references.30 C.3 Terms and definitions.30 C.4 Quality management system elements.30 C.5 Sterilizing agent characterization.30 C.5.1 Neutralization.30 C.5.2 Studies of microbial inactivation.31 C.6 Process and equipment characterization.31 C.7 Product definition.31 C.7.1 General.31 C.7.2 Design considerations for medical devices intended for sterilization.32 C.7.3 Packaging considerations.32 C.8 Process definition.32 C.8.1 General.32 C.8.2 Influence on product and packaging.33 C.8.3 Determination of process effectiveness.33 C.9 Validation.34 C.9.1 General.34 C.9.2 Installation qualification (IQ).34 C.9.3 Operational qualification (OQ).35 C.9.4 Performance qualification (PQ).36 C.9.5 Documentation and approval of validation.37 C.10 Routine monitoring and control.37 C.11 Product release from sterilization.38 C.11.1 General.38 C.11.2 Product release using biological indicators.38 C.11.3 Parametric release.38 C.12 Maintaining process effectiveness.38 C.12.1 Change control.38 C.12.2 Requalification.39 Annex D (informative)
Environmental aspects regarding development, validation and routine control of Low Temperature Steam and Formaldehyde processes.40 D.1 General.40 D.2 Formaldehyde (brief description).40 D.3 Environmental impact of formaldehyde.40 D.4 Other environmental burdens.41 Annex ZA (informative)
Relationship between this European Standard and the Essential Requirements of EU Directive 93/42/EEC Medical devices.44 Bibliography.45



EN 15424:2007 (E) 4 Foreword This document (EN 15424:2007) has been prepared by Technical Committee CEN/TC 204 “Sterilization of medical devices”, the secretariat of which is held by BSI. This European Standard shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest by October 2007, and conflicting national standards shall be withdrawn at the latest by October 2007. This document has been prepared under a mandate given to CEN by the European Commission and the European Free Trade Association, and supports essential requirements of EU Directive(s). For relationship with EU Directive(s), see informative Annex ZA, which is an integral part of this document. According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.



EN 15424:2007 (E) 5
Introduction A sterile medical device is one which is free of viable microorganisms. European Standards, which specify requirements for validation and routine control of a sterilization process require, when it is necessary to supply a sterile medical device, that adventitious microbiological contamination of a medical device prior to sterilization be minimized. Even so, medical devices produced under standard manufacturing conditions in accordance with the requirements for quality management systems (see EN ISO 13485) or which have been subjected to a cleaning process as part of their reprocessing in a health care establishment may, prior to sterilization, have microorganisms on them, albeit in low numbers. Such products are non-sterile. The purpose of sterilization is to inactivate the microbiological contaminants and thereby transform the non-sterile products into sterile ones. The kinetics of inactivation of a pure culture of microorganisms by physical and/or chemical agents used to sterilize medical devices can generally best be described by an exponential relationship between the numbers of microorganisms surviving and the extent of treatment with the sterilizing agent; inevitably this means that there is always a finite probability that a microorganism may survive regardless of the extent of treatment applied. For a given treatment, the probability of survival is determined by the number and resistance of microorganisms and by the environment in which the organisms exist during treatment. It follows that the sterility of any one product in a population subjected to sterilization cannot be guaranteed and the sterility of a processed population is defined in terms of the probability of there being a viable microorganism present on a product item. This standard describes requirements which will enable the demonstration that a low temperature steam and formaldehyde sterilization process intended to sterilize medical devices has appropriate microbicidal activity, and that this activity is both reliable and reproducible, such that the relationship for the inactivation of microorganisms can be extrapolated with reasonable confidence to low levels of probability of there being a viable microorganism present on a product after sterilization. This standard does not specify the maximal value to be taken by this probability; specification of this probability is given in EN 556-1.
Requirements of the quality management system for medical device design/development, production, installation and servicing are given in EN ISO 13485. The standard for quality management systems recognizes that, for certain processes used in manufacturing or reprocessing, the effectiveness cannot be fully verified by subsequent inspection and testing of the product. Sterilization is an example of such a process. For this reason, sterilization processes are validated for use, the performance of the sterilization process monitored routinely and the equipment maintained. Exposure to a properly validated, accurately controlled sterilization process is not the only factor associated with the provision of reliable assurance that the product is sterile and, in this regard, suitable for its intended use. Attention is therefore given to a number of factors including: a) the microbiological status of incoming raw materials and/or components, b) the validation and routine control of any cleaning and disinfection procedures used on the product,
c) the control of the environment in which the product is manufactured, assembled and packaged, d) the control of equipment and processes, e) the control of personnel and their hygiene, f) the manner and materials in which the product is packaged, and, g) the conditions under which the product is transported and stored.



EN 15424:2007 (E) 6 The type of contamination on a product to be sterilized varies and this impacts upon the effectiveness of a sterilization process. Products that have been used in a health care setting and are being presented for re-sterilization in accordance with the manufacturer's instructions (see EN ISO 17664) should be regarded as a special case. There is the potential for such products to possess a wide range of contaminating microorganisms and residual inorganic and/or organic contamination in spite of the application of a cleaning process. Hence, particular attention has to be given to the validation and control of the cleaning and disinfection processes used during reprocessing. The requirements are the normative parts of this standard with which compliance is claimed. The guidance given in the informative annexes is not normative and is not provided as a checklist for auditors. The guidance provides explanations as well as methods that are accepted as being suitable means for complying with the requirements. Approaches other than those given in the guidance may be used, if they are effective in achieving compliance with the requirements of this European Standard. The development, validation and routine control of a sterilization process comprise a number of discrete but interrelated activities, for example calibration, maintenance, product definition, process definition, installation qualification, operational qualification, and performance qualification. While the activities required by this standard have been grouped together and are presented in a particular order, this European Standard does not require that the activities be performed in the order that they are presented. The activities required are not necessarily sequential, as the programs of development and validation may be iterative. It is possible that performing these different activities will involve a number of separate individuals and/or organizations, each of whom undertake one or more of these activities. This European Standard does not specify the particular individuals or organizations to carry out the activities. Activities required by this standard might also give rise to an environmental burden that should be considered and minimized, e.g. by utilizing flexibility in planning. Environmental aspects are addressed in Annex D of this standard.



EN 15424:2007 (E) 7 1 Scope 1.1 Inclusions 1.1.1 This European Standard specifies requirements for the development, validation and routine control of a Low Temperature Steam and Formaldehyde (LTSF) sterilization process for medical devices. NOTE Although the scope of this standard is limited to medical devices, it specifies requirements and provides guidance that may be applicable to other products and equipment.
1.1.2 This European Standard is intended to be applied by process developers, manufacturers of sterilization equipment, manufacturers of medical devices to be sterilized and the organizations with responsibility for sterilizing medical devices. (See EN ISO 14937:2000, Table E.1) 1.1.3 This European Standard covers sterilization processes which use a mixture of low temperature steam and formaldehyde as sterilant, and which are working below ambient pressure only. 1.2 Exclusions 1.2.1 Sterilization processes validated and controlled in accordance with the requirements of this standard should not be assumed to be effective in inactivating the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeld-Jakob disease. Specific recommendations have been produced in particular countries for the processing of materials potentially contaminated with these agents. 1.2.2 This standard does not specify requirements for designating a medical device as "STERILE". Such requirements are given in EN 556-1.
1.2.3 This standard does not specify a quality management system for the control of all stages of production of medical devices.
NOTE It is not a requirement of this standard to have a complete quality management system during manufacture or reprocessing, but those elements of such a system that are required are normatively referenced at appropriate places in the text. Attention is drawn to the standards for quality management systems (see EN ISO 13485) that control all stages of production or reprocessing of medical devices including the sterilization process. Further guidance is given in E.2 of EN ISO 14937:2000. 1.2.4 This standard does not specify requirements for occupational safety associated with the design and operation of LTSF sterilization facilities. NOTE 1 Safety requirements for sterilizers are specified in EN 61010-2-040. NOTE 2 Attention is also drawn to the existence in some countries of regulations stipulating safety requirements. 1.2.5 This European Standard does not cover analytical methods for determining levels or residues of formaldehyde and/or its reaction products. NOTE 1 Attention is drawn to EN 14180. NOTE 2 Attention is drawn to the possible existence in some countries of statutory regulation specifying limits for the level of formaldehyde residues on medical devices and products. 1.2.6 This European Standard does not cover preparatory measures that may be necessary before sterilization such as cleaning, disinfection and packing. NOTE For re-sterilizable medical devices, the manufacturer(s) of these devices should supply information on the preparatory measures (see EN ISO 17664).



EN 15424:2007 (E) 8 2 Normative references The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. EN 14180:2003, Sterilizers for medical purposes – Low temperature steam and formaldehyde sterilizers – Requirements and testing
EN ISO 11138-1, Sterilization of health care products - Biological indicators - Part 1: General requirements (ISO 11138-1:2006) EN ISO 11138-5:2006, Sterilization of health care products - Biological indicators - Part 5: Biological indicators for low-temperature steam and formaldehyde sterilization processes (ISO 11138-5:2006) EN ISO 11140, Sterilization of health care products - Chemical indicators (Parts as appropriate) EN ISO 11737-1, Sterilization of medical devices - Microbiological methods - Part 1: Determination of a population of microorganisms on products (ISO 11737-1:2006) EN ISO 11737-2, Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the validation of a sterilization process (ISO 11737-2:1998) EN ISO 13485, Medical devices - Quality management systems - Requirements for regulatory purposes (ISO 13485:2003) 3 Terms and definitions For the purposes of this document, the following terms and definitions apply. 3.1 adjustment
correction of a measurement device or system to indicate the value as established by calibration 3.2 aeration part or parts of the sterilization process in which defined conditions are u
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.