ASTM E1324-00(2005)
(Guide)Standard Guide for Measuring Some Electronic Characteristics of Ultrasonic Examination Instruments
Standard Guide for Measuring Some Electronic Characteristics of Ultrasonic Examination Instruments
SIGNIFICANCE AND USE
The recommended measurement procedures described in this guide are intended to provide performance-related measurements that can be reproduced under the specified test conditions using commercially available test instrumentation. These measurements indicate capabilities of sections of the ultrasonic examination instrument independent of specific transducers or examination conditions. Measurements are made from normally available connectors or test points so that no access to internal circuitry is required. Further, this guide is not intended for service, calibration, or maintenance of circuitry for which the manufacturer’instructions are available. It is intended primarily for pulse echo flaw detection instruments operating in the nominal frequency range of 100 kHz to 25 MHz, but the procedures are applicable to measurements on instruments utilizing significantly higher frequency components.
These procedures can be applied to the evaluation of any pulse-echo ultrasonic examination instrument which can be described as a combination of the electronic sections discussed in this guide.
Note 2—These procedures are not intended to preclude the use or application of equipment for which some or all of the measurement techniques of this document are not applicable.
An ultrasonic examination instrument that cannot be completely described as a combination of the electronic sections discussed in this practice can be partially evaluated. Each portion of the ultrasonic examination instrument that is evaluated must fit the description for the corresponding section.
This guide is meant to be used by electronic personnel to evaluate the electronic system components and not the ultrasonic system characteristics.
SCOPE
1.1 This guide describes procedures for electronically measuring the following performance-related characteristics of some sections of ultrasonic instruments:
1.1.1 Power Supply Sectionline regulation,battery discharge time, andbattery charge time.
1.1.2 Pulser Sectionpulse shape,pulse amplitude,pulse rise time, pulse length, andpulse frequency spectrum.
1.1.3 Receiver Sectionvertical linearity, frequency response,noise and sensitivity, and dB controls.
1.1.4 Time Base Sectionhorizontal linearity, andclock (pulse repetition rate).
1.1.5 Gate/Alarm Sectiondelay and width, resolution, alarm level, gain uniformity, analog output, and back echo gate.
1.2 This guide complements Practice E 317, and is not intended for evaluating the performance characteristics of ultrasonic examination instruments on the inspection/production line.Note 1
No access to internal circuity is required.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E1324–00 (Reapproved 2005)
Standard Guide for
Measuring Some Electronic Characteristics of Ultrasonic
1
Examination Instruments
This standard is issued under the fixed designation E1324; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
NOTE 1—No access to internal circuity is required.
1. Scope
1.1 This guide describes procedures for electronically mea-
2. Referenced Documents
suring the following performance-related characteristics of
2
2.1 ASTM Standards:
some sections of ultrasonic instruments:
E317 Practice for Evaluating Performance Characteristics
1.1.1 Power Supply Section:
of Ultrasonic Pulse-Echo Testing Instruments and Systems
line regulation,
2
without the Use of Electronic Measurement Instruments
battery discharge time, and
E1316 Terminology for Nondestructive Examinations
battery charge time.
2.2 Military Standard:
1.1.2 Pulser Section:
3
MIL-STD-45662A Calibration System Requirements
2.3 Other Standard:
pulse shape,
IEEE Std. 100, IEEE Standard Dictionary of Electrical and
pulse amplitude,
4
Electronic Terms
pulse rise time, pulse length, and
pulse frequency spectrum.
3. Summary of Guide
1.1.3 Receiver Section:
3.1 The electronic performance of each section is measured
by identifying that portion of the electrical circuit of the
vertical linearity,
instrument which comprises the section, applying the recom-
frequency response,
mended stimulus or load, or both, and performing the required
noise and sensitivity, and
measurements using commercially available electronic test
dB controls.
equipment. These data are then summarized in tabular or
1.1.4 Time Base Section:
graphical form as performance-related values which can be
comparedwithcorrespondingvaluesofotherultrasonicexami-
horizontal linearity, and
nation instruments or of values for the same instrument
clock (pulse repetition rate).
obtained earlier (see Section 12 for a suggested reporting
1.1.5 Gate/Alarm Section:
format).
3.2 The following describes the sections of the ultrasonic
delay and width,
instrument and their interrelations during measurement:
resolution,
3.2.1 Power Supply Section—The power supply section is
alarm level,
that portion of the total instrument circuitry which supplies the
gain uniformity,
regulated DC voltages required to power all other sections of
analog output, and
the ultrasonic instrument, including the high voltage (that is,
back echo gate.
pulser and CRT voltage) circuitry.
1.2 This guide complements Practice E317, and is not
3.2.2 Pulser Section—The pulser section is that portion of
intended for evaluating the performance characteristics of
the total instrument circuitry which generates the electrical
ultrasonic examination instruments on the inspection/
production line.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
1
This guide is under the jurisdiction of ASTM Committee E07 on Nondestruc- Standards volume information, refer to the standard’s Document Summary page on
tive Testing and is the direct responsibility of Subcommittee E07.06 on Ultrasonic the ASTM website.
3
Method. AvailablefromStandardizationDocumentsOrderDesk,Bldg.4SectionD,700
Current edition approved January 1, 2005. Published January 2005. Originally Robbins Ave., Philadelphia, PA 19111-5094, Attn: NPODS.
4
approved in 1990. Last previous edition approved in 2000 as E1324 - 00. DOI: Available from Institute of Electrical and Electronics Engineers, Inc. (IEEE),
10.1520/E1324-00R05. 445 Hoes Ln., P.O. Box 1331, Piscataway, NJ 08854-1331
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
---------------------- Page: 1 ----------------------
E1324–00 (2005)
pulse used to energize the ultrasonic search unit. The pulser Each portion of the ultrasonic examination instrument that is
section may also include the pulse-shape modification controls evaluated must fit the description for the corresponding sec-
such as pulse length, damping, or tuning controls. tion.
4.4 Thisguideismeanttobeusedbyelectronicpersonnelto
3.2.3 Receiver Section—The receiver section is that portion
evaluate the electronic system components and not the ultra-
ofthetotalinstrumentcircuitrywhichamplifies,ormodifies,or
sonic system characteristics.
both, the radio frequency (RF) pulses received from the
ultrasonic search unit. This includes the RF amplifiers, detec-
5. Apparatus
tors, video amplifiers, suppression and filtering, and the CRT
vertical deflection circuits. Some instruments may not co
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.