ASTM G138-12(2020)e1
(Test Method)Standard Test Method for Calibration of a Spectroradiometer Using a Standard Source of Irradiance
Standard Test Method for Calibration of a Spectroradiometer Using a Standard Source of Irradiance
SIGNIFICANCE AND USE
4.1 This method is intended for use by laboratories performing calibration of a spectroradiometer for spectral irradiance measurements using a spectral irradiance standard with known spectral irradiance values and associated uncertainties traceable to a national metrological laboratory that has participated in intercomparisons of standards of spectral irradiance, known uncertainties and known measurement geometry.
4.2 This method is generalized to allow for the use of different types of input optics provided that those input optics are suitable for the wavelength range and measurement geometry of the calibration.
4.3 This method is generalized to allow for the use of different types of monochromators provided that they can be configured for a bandwidth, wavelength range, and throughput levels suitable for the calibration being performed.
4.4 This method is generalized to allow for the use of different types of optical radiation detectors provided that the spectral response of the detector over the wavelength range of the calibration is appropriate to the signal levels produced by the monochromator.
SCOPE
1.1 This test method covers the calibration of spectroradiometers for the measurement of spectral irradiance using a standard of spectral irradiance that is traceable to a national metrological laboratory that has participated in intercomparisons of standards of spectral irradiance.
1.2 This method is not limited by the input optics of the spectroradiometric system. However, choice of input optics affects the overall uncertainty of the calibration.
1.3 This method is not limited by the type of monochromator or optical detector used in the spectroradiometer system. Parts of the method may not apply to determine which parts apply to the specific spectroradiometer being used. It is important that the choice of monochromator and detector be appropriate for the wavelength range of interest for the calibration. Though the method generally applies to photodiode array detector based systems, the user should note that these types of spectroradiometers often suffer from stray light problems and have limited dynamic range. Diode array spectroradiometers are not recommended for use in the ultraviolet range unless these specific problems are addressed.
1.4 The calibration described in this method employs the use of a standard of spectral irradiance. The standard of spectral irradiance must have known spectral irradiance values at given wavelengths for a specific input current and clearly defined measurement geometry. Uncertainties must also be known for the spectral irradiance values. The values assigned to this standard must be traceable to a national metrological laboratory that has participated in intercomparisons of standards of spectral irradiance. These standards may be obtained from a number of national standards laboratories and commercial laboratories. The spectral irradiance standards consist mainly of tungsten halogen lamps with coiled filaments enclosed in a quartz envelope, though other types of lamps are used. Standards can be obtained with calibration values covering all or part of the wavelength range from 200 to 4500 nm.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.2
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
Relations
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
´1
Designation: G138 − 12 (Reapproved 2020)
Standard Test Method for
Calibration of a Spectroradiometer Using a Standard Source
of Irradiance
This standard is issued under the fixed designation G138; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
ε NOTE—An editorial change was made to 3.1 in July 2020.
INTRODUCTION
Astandardized means of performing and reporting calibration of the spectroradiometer for spectral
irradiance measurements is desirable.
This test method presents specific technical requirements for a laboratory performing calibration of
a spectroradiometer for spectral irradiance measurements. A detailed procedure for performing the
calibration and reporting the results is outlined.
This test method for calibration is applicable to spectroradiometric systems consisting of at least a
monochromator, input optics, and an optical radiation detector, and applies to spectroradiometric
calibrations performed with a standard of spectral irradiance with known irradiance values traceable
to a national metrological laboratory that has participated in intercomparisons of standards of spectral
irradiance. The standard must also have known uncertainties and measurement geometry associated
with its irradiance values.
1. Scope troradiometers are not recommended for use in the ultraviolet
range unless these specific problems are addressed.
1.1 This test method covers the calibration of spectroradi-
ometers for the measurement of spectral irradiance using a 1.4 The calibration described in this method employs the
use of a standard of spectral irradiance. The standard of
standard of spectral irradiance that is traceable to a national
metrological laboratory that has participated in intercompari- spectral irradiance must have known spectral irradiance values
at given wavelengths for a specific input current and clearly
sons of standards of spectral irradiance.
defined measurement geometry. Uncertainties must also be
1.2 This method is not limited by the input optics of the
known for the spectral irradiance values. The values assigned
spectroradiometric system. However, choice of input optics
to this standard must be traceable to a national metrological
affects the overall uncertainty of the calibration.
laboratory that has participated in intercomparisons of stan-
1.3 This method is not limited by the type of monochroma-
dards of spectral irradiance. These standards may be obtained
tor or optical detector used in the spectroradiometer system.
from a number of national standards laboratories and commer-
Parts of the method may not apply to determine which parts
cial laboratories. The spectral irradiance standards consist
apply to the specific spectroradiometer being used. It is
mainly of tungsten halogen lamps with coiled filaments en-
important that the choice of monochromator and detector be
closed in a quartz envelope, though other types of lamps are
appropriate for the wavelength range of interest for the
used. Standards can be obtained with calibration values cov-
calibration. Though the method generally applies to photo-
ering all or part of the wavelength range from 200 to 4500 nm.
diode array detector based systems, the user should note that
1.5 This standard does not purport to address all of the
these types of spectroradiometers often suffer from stray light
safety concerns, if any, associated with its use. It is the
problems and have limited dynamic range. Diode array spec-
responsibility of the user of this standard to establish appro-
priate safety, health, and environmental practices and deter-
mine the applicability of regulatory limitations prior to use.
This test method is under the jurisdiction of ASTM Committee G03 on
Weathering and Durability and is the direct responsibility of Subcommittee G03.09
on Radiometry.
CurrenteditionapprovedJune1,2020.PublishedJuly2020.Originallyapproved Available from the CIE, (International Commission on Illumination), http://
in 1996. Last previous edition approved in 2012 as G138–12. DOI: 10.1520/ www.techstreet.com/ciegate.tmpl CIE Central Bureau, Kegelgasse 27, A-1030
G0138-12R20E01. Vienna, Austria.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
´1
G138 − 12 (2020)
1.6 This international standard was developed in accor- 3.1.11 ultraviolet, adj—optical radiation at wavelengths be-
dance with internationally recognized principles on standard- low 400 nanometres.
ization established in the Decision on Principles for the
3.2 Definitions of Terms Specific to This Standard:
Development of International Standards, Guides and Recom-
3.2.1 calibration subsystems, n—the instruments used to
mendations issued by the World Trade Organization Technical
supply and monitor current to a standard lamp during
Barriers to Trade (TBT) Committee.
calibration, consisting of a DC power supply, a current shunt,
and a digital voltmeter.
2. Referenced Documents
3.2.2 National Metrological Institution (NMI), n—A na-
2.1 ASTM Standards:
tion‘s internationally recognized standardization laboratory.
E772Terminology of Solar Energy Conversion
3.2.2.1 Discussion—The International Bureau of Weights
E1341Practice for Obtaining Spectroradiometric Data from
andMeasurements(abbreviationBIPMfromtheFrenchterms)
Radiant Sources for Colorimetry
establishestherecognitionthroughMutualRecognitionAgree-
2.2 Other Documents: ments. See http://www.bipm.org/en/cipm-mra. The NMI for
the United States of America is the National Institute for
CIEPublicationNo.63 TheSpectrodiometricMeasurement
of Light Sources Standards and Technology (NIST).
NIST Technical Note 1927:Guidelines for Evaluation and
3.2.3 passband,n—theeffectivebandwidth(c.f.),orspectral
Expressing Uncertainty of NIST Measurement Results
interval, over which the spectroradiometer system transmits at
agivenwavelengthsetting.Expressedasfull-widthatone-half
3. Terminology
maximum, as in bandwidth.Afunction of the linear dispersion
(nm/mm) and slit or aperture widths (mm) of the monochro-
3.1 Definitions:
mator system.
3.1.1 General terms pertaining to optical radiation and
opticalmeasurementsystemsaredefinedinTerminologyE772.
3.2.4 primary standard of spectral irradiance, n—a broad
Some of the more important terms from that standard used in
spectrum light source with known spectral irradiance values at
this paper are listed here.
various wavelengths which are traceable to a national metro-
3.1.2 bandwidth, n—the extent of a band of radiation
logical laboratory that has participated in intercomparisons of
reported as the difference between the two wavelengths at
standards of spectral irradiance.
which the amount of radiation is half of its maximum over the
3.2.5 responsivity, n—symbol R = dS/dφ, S is signal from
given band.
spectroradiometer detector, φ is radiant flux at the detector.
3.1.3 diffuser, n—a device used to scatter or disperse light
3.2.6 secondary standard of spectral irradiance, n—a stan-
usually through the process of diffuse transmission or reflec-
dard calibrated by reference to another standard such as a
tion.
primary or reference standard.
3.1.4 integrating sphere, n—a hollow sphere coated inter-
3.2.7 slit scattering function, n—symbol Z(λ ,λ),therespon-
o
nallywithawhitediffusereflectingmaterialandprovidedwith
sivity of the combined detector and monochromator system as
separate openings for incident and exiting radiation.
a function of wavelengths, λ, in the neighborhood of a given
3.1.5 irradiance, n—radiant flux incident per unit area of a
wavelength setting, λ . The slit scattering function is the
o
surface.
spectral responsivity in the neighborhood of specific wave-
length setting, λ .
3.1.6 monochromator, n—aninstrumentforisolatingnarrow
o
portions of the optical spectrum of a light source
3.2.8 spectral scattering (stray light), n— light with wave-
lengthsoutsidethepassbandofaspectroradiometeraparticular
3.1.7 polarization, n—with respect to optical radiation, the
wavelength setting that is received by the detector and contrib-
restriction of the magnetic or electric field vector to a single
utes to the output signal.
plane.
3.1.8 radiant flux, n—thetimerateofflowofradiantenergy
4. Significance and Use
measured in watts.
4.1 Thismethodisintendedforusebylaboratoriesperform-
3.1.9 spectral irradiance, n—irradianceperunitwavelength
ing calibration of a spectroradiometer for spectral irradiance
interval at a given wavelength.
measurements using a spectral irradiance standard with known
3.1.10 spectroradiometer, n—an instrument for measuring
spectral irradiance values and associated uncertainties trace-
the radiant energy of a light source at each wavelength
able to a national metrological laboratory that has participated
throughout the spectrum.
in intercomparisons of standards of spectral irradiance, known
uncertainties and known measurement geometry.
4.2 This method is generalized to allow for the use of
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
different types of input optics provided that those input optics
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
are suitable for the wavelength range and measurement geom-
Standards volume information, refer to the standard’s Document Summary page on
etry of the calibration.
the ASTM website.
G03
4.3 This method is generalized to allow for the use of
Available from American National Standards Institute, 11 West 42nd Street,
13th Floor, New York, NY 10036 different types of monochromators provided that they can be
´1
G138 − 12 (2020)
configured for a bandwidth, wavelength range, and throughput 1). If the monochromator has interchangeable slits, it is
levels suitable for the calibration being performed. important that the manufacturer document the effective band-
width of the monochromator with all possible combinations of
4.4 This method is generalized to allow for the use of
the slits or that these bandwidths be determined experimen-
different types of optical radiation detectors provided that the
tally. Configuration of the slits should be such that the
spectral response of the detector over the wavelength range of
bandpass function of the monochromator is symmetric, pref-
the calibration is appropriate to the signal levels produced by
erably triangular. The bandwidth should be constant across the
the monochromator.
wavelength region of interest and maintained between 85%
5. Apparatus and 100% of the measurement wavelength interval. The
precision of the wavelength positioning of the monochromator
5.1 Laboratory:
should be 0.1 nm with an absolute accuracy of better than 0.5
5.1.1 The room in which the calibrations are performed and
nm (see Practice E1341). For improved performance in the uv,
especially the area surrounding the optical bench should be
itisrecommendedthathighorderrejectionfiltersbeinsertedin
devoidofreflectivesurfaces.Thecalibrationvaluesassignedto
theopticalpathinthemonochromator.Thepurposeofthehigh
the spectral irradiance standard are for direct irradiance from
order rejection filters is to block radiation in the monochroma-
the lamp and any radiation entering the monochromator from
tor of unwanted wavelengths that could otherwise overpower
some other source including ambient reflections will be a
the signals being measured. The effects of variations in
source of error.
temperature and humidity on the performance of the mono-
5.1.2 The temperature and humidity in the laboratory shall
chromator should be addressed in writing by the manufacturer.
be maintained so as to agree with the conditions under which
5.2.1.2 The monochromator shall not be subject to shock or
the calibrations of the spectral irradiance standard and the
mechanical vibration during the calibration. This can be
calibration subsystems were performed (typically 20 °C,
facilitated by the use of a vibration isolated lab table. If any
25°C, 50% relative humidity).
optical parts in the monochromator are configurable by the
5.1.3 Air drafts in the laboratory should be minimized since
they could affect the output of electrical discharge lamps. user, refer to the manufacturer precautions about opening the
monochromator and handling any parts therein.
5.2 Spectroradiometer
5.2.2 Optical Radiation Detector:
5.2.1 Monochromator:
5.2.1.1 This can be a fixed or scanning, single or multiple, 5.2.2.1 The optical radiation detector employed by the
monochromator employing holographic or ruled gratings or spectroradiometer shall be selected for optimal response over
prisms or a combination of these dispersive elements. For the wavelength range of interest. It is also important that the
improved performance in the ultraviolet (UV) portion of the detector is sensitive enough to measure the levels of light that
spectrum, it is recommended that a scanning double mono- will be produced by the monochromator when it is configured
chromator be used to achieve lower stray light levels (see Fig. forthecalibrationprocess.Theactiveareaofthedetectorshall
FIG. 1 Typical Double Grating Monochromator Layout (courtesy Optronic Laboratories, Used with Permission)
´1
G138 − 12 (2020)
be evenly illuminated by the radiation leaving the exit slit of 5.3.2 Compare the signal from the detector with the filter in
the monochromator. A photomultiplier is typically used be- place to the shuttered, or dark signal of the detector. A signal
cause of its high responsivity and good signal-to-noise ratio. between 10% and 90% of the unfiltered signal indicates
For this reason it is recommended for use when measuring significant scattered light is reaching the detector, possibly due
spectral irradiance in the uv portion of the spectrum. to a non-light-tight enclosure.
5.2.2.2 Anyeffectsofvariationintemperatureandhumidity
5.4 Optical Radiation Sources
on the response of the detector documented by the manufac-
5.4.1 Wavelength Calibration Source:
turer shall be reported. Of all components of the
5.4.1.1 A stable wavelength source is required to calibrate
spectroradiometer, the detector is usually the most sensitive to
the wavelength positioning accuracy of the monochromator.
changesintemperature.Somedetectorsmayrequirecoolingin
Thiscanbeagasdischargelamporalaser.Theimportantthing
order to maintain a specific tempe
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.