Standard Practice for Evaluating Performance Characteristics of Ultrasonic Pulse-Echo Testing Instruments and Systems without the Use of Electronic Measurement Instruments

SCOPE
1.1 This practice describes procedures for evaluating the following performance characteristics of ultrasonic pulse-echo examination instruments and systems: Horizontal Limit and Linearity; Vertical Limit and Linearity; Resolution - Entry Surface and Far Surface; Sensitivity and Noise; Accuracy of Calibrated Gain Controls. Evaluation of these characteristics is intended to be used for comparing instruments and systems or, by periodic repetition, for detecting long-term changes in the characteristics of a given instrument or system that may be indicative of impending failure, and which, if beyond certain limits, will require corrective maintenance. Instrument characteristics measured in accordance with this practice are expressed in terms that relate to their potential usefulness for ultrasonic examinations. Instrument characteristics expressed in purely electronic terms may be measured as described in E 1324.
1.2 Ultrasonic examination systems using pulsed-wave trains and A-scan presentation (rf or video) may be evaluated.
1.3 The procedures are applicable to shop or field conditions; additional electronic measurement instrumentation is not required.
1.4 This practice establishes no performance limits for examination systems; if such acceptance criteria are required, these must be specified by the using parties. Where acceptance criteria are implied herein they are for example only and are subject to more or less restrictive limits imposed by customer's and end user's controlling documents.
1.5 The specific parameters to be evaluated, conditions and frequency of test, and report data required, must also be determined by the user.
1.6 This practice may be used for the evaluation of a complete examination system, including search unit, instrument, interconnections, fixtures and connected alarm and auxiliary devices, primarily in cases where such a system is used repetitively without change or substitution. This practice is not intended to be used as a substitute for calibration or standardization of an instrument or system to inspect any given material. There are limitations to the use of standard reference blocks for that purpose.
1.7 Required test apparatus includes selected test blocks and a precision external attenuator (where specified) in addition to the instrument or system to be evaluated.
1.8 Precautions relating to the applicability of the procedures and interpretation of the results are included.
1.9 Alternate procedures, such as examples described in this document, or others, may only be used with customer approval.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
30-Nov-2006
Technical Committee
Drafting Committee
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E317-06a - Standard Practice for Evaluating Performance Characteristics of Ultrasonic Pulse-Echo Testing Instruments and Systems without the Use of Electronic Measurement Instruments
English language
13 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E317 – 06a
Standard Practice for
Evaluating Performance Characteristics of Ultrasonic Pulse-
Echo Testing Instruments and Systems without the Use of
1
Electronic Measurement Instruments
This standard is issued under the fixed designation E317; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the Department of Defense.
1. Scope auxiliary devices, primarily in cases where such a system is
used repetitively without change or substitution. This practice
1.1 This practice describes procedures for evaluating the
is not intended to be used as a substitute for calibration or
following performance characteristics of ultrasonic pulse-echo
standardizationofaninstrumentorsystemtoinspectanygiven
examination instruments and systems: Horizontal Limit and
material. There are limitations to the use of standard reference
Linearity; Vertical Limit and Linearity; Resolution - Entry
2
blocks for that purpose.
Surface and Far Surface; Sensitivity and Noise; Accuracy of
1.7 Requiredtestapparatusincludesselectedtestblocksand
Calibrated Gain Controls. Evaluation of these characteristics is
a precision external attenuator (where specified) in addition to
intended to be used for comparing instruments and systems or,
the instrument or system to be evaluated.
by periodic repetition, for detecting long-term changes in the
1.8 Precautions relating to the applicability of the proce-
characteristics of a given instrument or system that may be
dures and interpretation of the results are included.
indicative of impending failure, and which, if beyond certain
1.9 Alternateprocedures,suchasexamplesdescribedinthis
limits, will require corrective maintenance. Instrument charac-
document,orothers,mayonlybeusedwithcustomerapproval.
teristics measured in accordance with this practice are ex-
1.10 This standard does not purport to address all of the
pressed in terms that relate to their potential usefulness for
safety concerns, if any, associated with its use. It is the
ultrasonic examinations. Instrument characteristics expressed
responsibility of the user of this standard to establish appro-
in purely electronic terms may be measured as described in
priate safety and health practices and determine the applica-
E1324.
bility of regulatory limitations prior to use.
1.2 Ultrasonic examination systems using pulsed-wave
trains and A-scan presentation (rf or video) may be evaluated.
2. Referenced Documents
1.3 The procedures are applicable to shop or field condi-
3
2.1 ASTM Standards:
tions; additional electronic measurement instrumentation is not
E114 Practice for Ultrasonic Pulse-Echo Straight-Beam
required.
Contact Testing
1.4 This practice establishes no performance limits for
E127 Practice for Fabricating and Checking Aluminum
examination systems; if such acceptance criteria are required,
Alloy Ultrasonic Standard Reference Blocks
these must be specified by the using parties. Where acceptance
E214 Practice for Immersed Ultrasonic Testing by the
criteria are implied herein they are for example only and are
4
Reflection Method Using Pulsed Longitudinal Waves
subjecttomoreorlessrestrictivelimitsimposedbycustomer’s
E428 Practice for Fabrication and Control of Metal, Other
and end user’s controlling documents.
than Aluminum, Reference Blocks Used in Ultrasonic
1.5 The specific parameters to be evaluated, conditions and
Testing
frequency of test, and report data required, must also be
determined by the user.
1.6 This practice may be used for the evaluation of a
complete examination system, including search unit, instru-
2
ment, interconnections, fixtures and connected alarm and
Beck, K. H., “Limitations to the Use of Reference Blocks for Periodic and
Preinspection Calibration of Ultrasonic Inspection Instruments and Systems,”
Materials Evaluation, Vol 57, No. 3, March 1999.
1 3
This practice is under the jurisdiction of ASTM Committee E07 on Nonde- For referenced ASTM standards, visit the ASTM website, www.astm.org, or
structive Testing and is the direct responsibility of Subcommittee E07.06 on contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Ultrasonic Method. Standards volume information, refer to the standard’s Document Summary page on
Current edition approved Dec. 1, 2006. Published January 2007. Originally the ASTM website.
4
approved in 1967. Last previous edition approved in 2006 as E317 - 06. DOI: Withdrawn. The last approved version of this historical standard is referenced
10.1520/E0317-06A. on www.astm.org.
Copyright © ASTM International, 100 Barr H
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.