Standard Test Method for Determination of Residual Contamination of Materials and Components by Total Carbon Analysis Using a High-Temperature Combustion Analyzer

SCOPE
1.1 This test method covers the determination of residual contamination in an aqueous sample by the use of a total carbon ( TC) analyzer. When used in conjunction with Practice G 131 and G 136, this procedure may be used to determine the cleanliness of systems, components, and materials requiring a high level of cleanliness, such as oxygen systems. This procedure is applicable for aqueous-based cleaning and sampling methods only.
1.2 This test method is not suitable for the evaluation of particulate contamination, or contaminants that are not soluble in or that do not form an emulsion with water.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
09-Sep-2001
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM G144-01 - Standard Test Method for Determination of Residual Contamination of Materials and Components by Total Carbon Analysis Using a High-Temperature Combustion Analyzer
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:G144–01
Standard Test Method for
Determination of Residual Contamination of Materials and
Components by Total Carbon Analysis Using a
1
HighTemperature Combustion Analyzer
This standard is issued under the fixed designation G 144; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope G 136 Practice for Determination of Soluble Residual Con-
5
taminants in Materials by Ultrasonic Extraction
1.1 This test method covers the determination of residual
contamination in an aqueous sample by the use of a total
3. Terminology
carbon (TC) analyzer. When used in conjunction with Practice
3.1 Definitions of Terms Specific to This Standard:
G 131 and G 136, this procedure may be used to determine the
3.1.1 contaminant (contamination), n—unwantedmolecular
cleanliness of systems, components, and materials requiring a
and particulate matter that could affect or degrade the perfor-
high level of cleanliness, such as oxygen systems. This
mance of the components upon which they reside.
procedure is applicable for aqueous-based cleaning and sam-
3.1.2 nonvolatile residue (NVR), n—molecular and particu-
pling methods only.
late matter remaining following the filtration and controlled
1.2 This test method is not suitable for the evaluation of
evaporation of a liquid containing contaminants.
particulate contamination, or contaminants that are not soluble
3.1.3 Discussion—In this test method, the NVR may be
in or that do not form an emulsion with water.
uniformly distributed as in a solution or an emulsion, or in the
1.3 This standard does not purport to address all of the
form of droplets. Molecular contaminants account for most of
safety concerns, if any, associated with its use. It is the
the NVR.
responsibility of the user of this standard to establish appro-
3.1.4 particle (particulate contaminant), n—a piece of mat-
priate safety and health practices and determine the applica-
terinasolidstatewithobservablelength,width,andthickness.
bility of regulatory limitations prior to use.
3.1.5 Discussion—The size of a particle is usually defined
2. Referenced Documents by its greatest dimension and is specified in micrometres.
3.1.6 molecular contaminant (non-particulate contamina-
2.1 ASTM Standards:
2
tion), n—the molecular contaminant may be in a gaseous,
D 1193 Specification for Reagent Water
liquid, or solid form.
D 2579 Test Methods for Total and Organic Carbon in
3
Water
4. Summary of Test Method
F 331 Test Method for Nonvolatile Residue of Halogenated
4.1 A test method is described for the quantitative analysis
Solvent Extract from Aerospace Components (Using Ro-
4 of aqueous samples and may be used in the determination of
tary Flash Evaporator)
contamination on parts, components, and materials used in
G 121 Practice for Preparation of Contaminated Test Cou-
5 systems requiring a high degree of cleanliness. The residue
pons for the Evaluation of Cleaning Agents
removed during aqueous cleaning or sampling, using cleaning
G 131 Practice for the Cleaning of Materials and Compo-
5
methods such as Practice G 131and Practice G 136, are ana-
nents by Ultrasonic Techniques
lyzed using a high-temperature combustion analyzer with a
sensitivity of 6 0.2 mgC/L(milligrams of carbon per litre).An
1
This test method is under the jurisdiction of ASTM Committee G04 on aqueous sample is injected into the sample port. A stream of
Compatibility and Sensitivity of Materials in Oxygen-EnrichedAtmospheres and is
oxygen or air carries the sample into the catalytic combustion
the direct responsibility of Subcommittee G04.01 on Test Methods.
chamber, which is maintained at a temperature high enough to
Current edition approved Sept. 10, 2001. Published January 2002.Originally
completely pyrolyze the sample. The sample is combusted in
published as G 144–96. Last previous edition G 144–96
2
Annual Book of ASTM Standards, Vol 11.01.
the catalytic combustion chamber and the products are carried
3
Annual Book of ASTM Standards, Vol 11.02.
by the oxygen or air stream into a nondispersive infrared
4
Annual Book of ASTM Standards, Vol 15.03.
5 (NDIR)detectorwheretheamountofcarbondioxideinthegas
Annual Book of ASTM Standards, Vol 14.04.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1

---------------------- Page: 1 ----------------------
G144
stream is determined. Additional information on the use and all reagents conform to the specifications of the Committee on
operation of carbon analyzers is provided in Test Methods Analytical Reagents of the American Chemical Societ
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.