ISO 3219-3:2026
(Main)Rheology — Part 3: Test procedure and examples for the evaluation of results when using rotational and oscillatory rheometry
- Abstract
This document provides requirements and guidelines for selecting the measuring device, measuring geometry and temperature control unit for tests involving rotational and oscillatory rheometry. The general principles of test performance are described and example evaluations of rotational and oscillatory rheometry are provided.
- Status
- Published
- Publication Date
- 15-Oct-2026
- Technical Committee
- ISO/TC 35/SC 9 - General test methods for paints and varnishes
- Current Stage
- 6060 - International Standard published - 60 Publication
- Start Date
- 16-Oct-2026
- Due Date
- 16-Oct-2026
- Completion Date
- 16-Oct-2026
Buy Documents
ISO 3219-3:2026 - Rheology — Part 3: Test procedure and examples for the evaluation of results when using rotational and oscillatory rheometry
ISO 3219-3:2026 - Rhéologie — Partie 3: Mode opératoire d'essai et exemples d'évaluation des résultats en cas d'utilisation de la rhéométrie rotative et oscillatoire
ISO 3219-3:2026 is an ISO standard for test procedure and example result evaluation in rotational and oscillatory rheometry. It specifies how to select the measuring device, measuring geometry and temperature control unit, and it describes the main principles for running and evaluating common rheological tests. It is written for people who need reproducible flow and viscoelastic data from samples such as coatings, polymers, pastes and other structured materials.
What does ISO 3219-3:2026 specify?
ISO 3219-3:2026 specifies requirements and guidelines for rotational and oscillatory rheometry, with a focus on choosing suitable equipment and on evaluating test results. The document is organized into clauses on scope, references, terms, selection of the test setup, preconditions, test execution, basic tests, combined tests and the test report, plus an informative annex with calculation examples.
Clause 4 covers selection of the measuring device, geometry, temperature control system and measuring method. Clause 5 covers ambient conditions, sample preparation, filling and cleaning. Clauses 6 to 8 cover how to run measurements and how to evaluate the main rotational and oscillatory tests. Clause 9 defines the minimum report content. Annex A gives calculation examples.
| Annex | What it covers |
|---|---|
| Annex A | Informative calculation examples for shear rate and shear stress in practical cases |
What are the key requirements of ISO 3219-3:2026?
ISO 3219-3:2026 is mainly about matching the test setup to the sample and then reporting the measurement in a way that can be repeated and checked. The practical point is that good rheology results depend as much on the setup and the preconditioning as on the test itself.
Device selection, Clause 4.2 - ISO 3219-3:2026 says a rotational viscometer can be used for shear viscosity, while a rotational rheometer can also run the rotational and oscillatory basic tests. For oscillation, creep and relaxation, the document points to devices with low internal friction, such as air-bearing or magnetic-bearing systems. In practice, this matters because internal friction and instrument limits can distort sensitive viscoelastic measurements.
Geometry selection, Clause 4.3 - ISO 3219-3:2026 uses the geometry types CC, CP, PP, DG and RMG and recommends choosing them according to sample properties, sample volume, cleaning needs and phenomena such as wall slip, sedimentation and phase separation. For absolute geometries, the maximum particle or heterogeneity size should not exceed 10% of the gap width. This helps the user avoid misleading data from a gap that is too small for the sample structure.
Temperature control, Clause 4.4 - ISO 3219-3:2026 stresses that the indicated temperature can differ from the actual sample temperature and that hoods or inert gas may be needed to reduce gradients or air reactions. This is important whenever thermal effects, evaporation or oxidation can change the material during the test.
Method selection, Clause 4.5 and measuring profiles, Clause 6 - ISO 3219-3:2026 distinguishes controlled rate (CR), controlled stress (CS) and controlled deformation (CD) and requires the measuring profile to be defined with constant values, continuous ramps or stepped ramps. For very low rotational shear rates, the point duration should be at least the inverse of the shear rate, and oscillatory measurements should use at least one oscillation period for a data point. In practice, this is what makes the result interpretable and comparable across labs.
Measurement preconditions, Clause 5 - ISO 3219-3:2026 requires ambient conditions to meet the instrument manufacturer’s requirements and recommends a climate-controlled room at 23 ± 2 °C and 50 ± 5% relative humidity. It also requires coordinated sample preparation, complete cleaning of the geometry, careful filling, bubble-free contact and trimming where relevant. These steps matter because sample history and handling can change the measured structure.
Basic tests and example evaluations, Clause 7 - ISO 3219-3:2026 covers time-dependent, temperature-dependent, amplitude sweep and frequency sweep tests in both rotation and oscillation. The examples show how to obtain flow curves, viscosity curves, gel point, gel time, glass transition temperature, crystallization temperature, melting temperature, zero shear viscosity, the LVR and crossover points. This gives the user a practical path from raw curve shape to a usable material property.
Combined tests, Clause 8 - ISO 3219-3:2026 combines basic tests into flow-curve hysteresis, temperature cycle tests, master curves by time/temperature shift, step tests for structural recovery and creep and recovery tests. These combinations are used to assess thixotropy, structure recovery and long-term behavior. They are useful when a single test does not capture the material’s process behavior.
Test report, Clause 9 - ISO 3219-3:2026 requires reporting the sample identity, preparation, instrument and geometry, measurement settings, evaluation method, results, deviations, unusual observations, operator and date. This is what makes the measurement traceable for audits, customer claims and internal quality control.
What terms does ISO 3219-3:2026 define?
- Crystallization temperature, T - The temperature at which crystallization starts and partial order is established as a melt solidifies.
- Gel point - The transition point from sol to gel in oscillation in the linear-viscoelastic range, where the loss angle is independent of frequency.
- Gel time / pot life - The period during which a crosslinking material remains processable while a user-specified criterion is satisfied.
- Glass transition temperature, T_g - The characteristic temperature of the range in which the glass transformation takes place.
- Shear relaxation modulus, G - The ratio of shear stress to shear strain during a relaxation test.
- Sol/gel transition - The transition range from a viscoelastic liquid (sol) to a viscoelastic solid (gel).
- Trimming - Removal of excess sample volume, especially after filling plate-plate or cone-plate geometries.
- Linear-viscoelastic range (LVR) - The amplitude range in which the oscillatory response remains linear and is used as the basis for amplitude and frequency sweep settings.
Who uses ISO 3219-3:2026?
ISO 3219-3:2026 is used by rheology laboratories, product developers, quality managers and technical buyers who need to choose a rheometer setup and interpret rotational or oscillatory test results. It is relevant in paints and varnishes, plastics, coatings, cosmetics, lubricants, adhesives and other materials where flow, recovery and viscoelastic behavior matter.
It is especially useful for work on shelf life, pumping, spraying, brushing, leveling, gelation, crosslinking, thermal behavior and structural recovery. In practice, that means it supports both formulation work and routine quality control.
Which standards are used with ISO 3219-3:2026?
ISO 3219-3:2026 cites ISO 3219-1:2021 as a normative reference. That part provides the vocabulary and symbols used throughout the document.
The bibliography also names related standards and reports that support the methods:
- ISO 3219-2 - general principles of rotational and oscillatory rheometry, including device setup and geometry topics
- ISO 6721-10 - complex shear viscosity using a parallel-plate and cone-and-plate oscillatory rheometer
- ISO/TR 20659-1 - yield point determination
- ISO/TR 20659-2 - time-dependent structural change, including thixotropy
What does the ISO 3219-3:2026 document contain?
ISO 3219-3:2026 contains selection guidance tables, measuring profiles, example curves and evaluation paths for the main rotational and oscillatory tests. It also includes figures for equipment selection, sample filling, ramp profiles, amplitude sweeps, frequency sweeps, creep and recovery, step tests and combined test sequences.
Clause 7 shows example evaluations for time-dependent and temperature-dependent rotational and oscillatory measurements. Clause 8 expands these into combined tests such as hysteresis areas, cycle tests, master curves and structural recovery analyses.
Annex A provides calculation examples for shear rate and shear stress in practical situations such as coating, pipeline flow, compressed-air spraying, sedimentation, pipe flow and sagging. These examples help users relate laboratory settings to processing conditions.
Relations
- Effective Date
- 29-Apr-2026
- Effective Date
- 12-Feb-2026
- Effective Date
- 07-Jan-2025
Buy Documents
ISO 3219-3:2026 - Rheology — Part 3: Test procedure and examples for the evaluation of results when using rotational and oscillatory rheometry
ISO 3219-3:2026 - Rhéologie — Partie 3: Mode opératoire d'essai et exemples d'évaluation des résultats en cas d'utilisation de la rhéométrie rotative et oscillatoire
Get Certified
Connect with accredited certification bodies for this standard
DIN CERTCO
DIN Group product certification.

Smithers Quality Assessments
US management systems and product certification.
Sponsored listings
Frequently Asked Questions
ISO 3219-3:2026 is a standard published by the International Organization for Standardization (ISO). Its full title is "Rheology — Part 3: Test procedure and examples for the evaluation of results when using rotational and oscillatory rheometry". This standard covers: This document provides requirements and guidelines for selecting the measuring device, measuring geometry and temperature control unit for tests involving rotational and oscillatory rheometry. The general principles of test performance are described and example evaluations of rotational and oscillatory rheometry are provided.
This document provides requirements and guidelines for selecting the measuring device, measuring geometry and temperature control unit for tests involving rotational and oscillatory rheometry. The general principles of test performance are described and example evaluations of rotational and oscillatory rheometry are provided.
ISO 3219-3:2026 is classified under the following ICS (International Classification for Standards) categories: 83.080.01 - Plastics in general. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO 3219-3:2026 has the following relationships with other standards: It is inter standard links to EN ISO 11909:2025, EN ISO 3219-3:2026, ISO 22197-4:2021. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ISO 3219-3:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 crystallization temperature
- 3.2 gel point
- 3.3 gel time
- 3.4 glass transition temperature
- 3.5 sample preparation
- 3.6 shear relaxation modulus
- 3.7 sol/gel transition
- 3.8 trimming
- 4 Guidelines and requirements for selection
- 4.1 General
- 4.2 Selection of the measuring device
- 4.3 Selection of the measuring geometry
- 4.4 Selection of the temperature control system
- 4.5 Selection of the measuring method
- 8.4 Master curve by means of time/temperature shift of frequency sweeps (position
- 8.6 Creep and recovery test (analogous to 7.2.1.2)
- 5 Preconditions for a measurement
- 5.1 Ambient conditions
- 5.2 Sample preparation, filling and cleaning
- 6 Performance of the measurement
- 6.1 General
- 6.2 Measuring profile
- 7 Basic tests and example evaluations
- 7.1 General
- 7.2 Rotational tests
- 1 Newtonian flow behaviour (ideal viscous behaviour)
- 7.3 Oscillatory tests
- 8 Combined basic tests
- 8.1 General
- 8.2 Flow curves and hysteresis area
- 8.3 Temperature tests as cycle tests
- 8.4 Master curve by means of time/temperature shift of frequency sweeps
- 8.5 Step tests for determination of the time-dependent structural change
- 8.6 Creep and recovery tests
- 9 Test report
- Annex A
- Bibliography
1 Scope
This document provides requirements and guidelines for selecting the measuring device, measuring geometry and temperature control unit for tests involving rotational and oscillatory rheometry. The general principles of test performance are described and example evaluations of rotational and oscillatory rheometry are provided.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 3219-1:2021, Rheology — Part 1: Vocabulary and symbols for rotational and oscillatory rheometry
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 3219-1 and the following apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https:// www .iso .org/ obp
- IEC Electropedia: available at https:// www .electropedia .org/
3.1 crystallization temperature
T
temperature at which crystallization starts and partial order is established when the melt solidifies Note 1 to entry: The unit of the crystallization temperature, T , is degrees Celsius (°C).
3.2 gel point
transition point from sol to gel in oscillation in the linear-viscoelastic range, where the loss angle, δ, is independent of frequency Note 1 to entry: The gel point can be a time or a temperature. Note 2 to entry: Contrary to this definition, the gel point is usually measured in practice at one frequency and determined on the basis of the loss angle δ = 45° or on the basis of the loss factor tan δ = 1.
ISO 3219-3:2026(en)
3.3 gel time
pot life period of time for which a crosslinking material remains processable while a user-specified criterion is satisfied Note 1 to entry: The gel time can be determined in oscillation in the nonlinear range or in rotation. Note 2 to entry: In practice, relative measuring geometries or alternative methods are also used, such as flow cups. Note 3 to entry: Limiting values for the viscosity or moduli can be used as user-defined criteria.
3.4 glass transition temperature
T g characteristic temperature of the temperature range in which the glass transformation takes place Note 1 to entry: The glass transition temperature depends on the test procedure and the measuring conditions.
3.5 sample preparation
handling, conditioning and load applied to the sample prior to measurement
3.6 shear relaxation modulus
G ratio of shear stress τ to shear strain γ during the relaxation test Note 1 to entry: The unit for the shear relaxation modulus G is Pascal (Pa).
3.7 sol/gel transition
transition range from viscoelastic liquid (sol) to viscoelastic solid (gel)
3.8 trimming
removal of excess sample volume Note 1 to entry: With plate-plate and cone-plate measuring geometries, a trimming position is often selected being larger than the gap width and at which the excess volume is removed so that the sample is flushed to the edge of the measuring geometry. When the measuring position is subsequently approached, this ensures optimum filling of the gap.
4 Guidelines and requirements for selection
4.1 General
The choice of measuring device, measuring geometry, temperature control unit, measuring method and further accessories depends on the properties of the sample and the resulting rheological requirements.
4.2 Selection of the measuring device
In ISO 3219-2, the differences between the different measuring devices for the determination of rheological properties and their setups are described. A rotational viscometer can be used to determine the shear viscosity. With a rotational rheometer, it is in addition possible to carry out all basic rotational and oscillatory tests (see Clause 7). For this reason, in addition to the viscosity function, this can be used to determine the viscoelastic properties, e.g. shear storage modulus and shear loss modulus. For viscometric measurements, all viscometers are principally suitable, regardless of the type of bearing. For oscillatory measurements, creep tests and relaxation tests, devices with low internal friction should be used, e.g. devices with air bearings or magnetic bearings. Some devices with ball bearings also offer the
ISO 3219-3:2026(en) option of performing oscillatory measurements, but only within a restricted measuring range. During the measurement execution, it shall be ensured that the work is carried out within the specifications of the measuring device, e.g. taking into account the torque limits and data acquisition rate. To cover the broadest possible range of applications, the viscometer or rheometer shall be capable of working with different measuring geometries. The ranges of torque and angular displacement, and therefore the accessible measuring range, depend on the measuring device.
4.3 Selection of the measuring geometry
The measuring geometry shall be selected taking into account the available measuring device, the sample properties and the measuring profile. Table 1 contains recommendations on selecting the different measuring geometries. Here, CC stands for coaxial cylinder, CP for cone-plate, PP for plate-plate, DG for double gap and RMG for relative measuring geometries (e.g. vane sensor, for details see ISO 3219-2). In general, when selecting an absolute measuring geometry, the maximum size of heterogeneities, e.g. the particle size in the sample, shall not exceed 10 % of the gap width. If, as a result, it is not possible to use absolute measuring geometries, then it is necessary to select relative measuring geometries. Optional accessories, such as a hood with a solvent trap to avoid solvent loss are described in ISO 3219-2. Table 1 — Recommendations for selection of the measuring geometry Criteria Recommended measuring geometry Coaxial cylin- Cone-plate Plate-plate Double gap Relative ders CC CP PP DG RMG Sample properties that determine the selection of the measuring geometry a Low viscosity x x x High viscosity x x Pasty, gel-like x Low surface tension x x Inhomogeneity, e.g. air bubbles, x foaming Further criteria for the selection of a measuring geometry Availability of a small sample vol- x x ume b Cleaning of the measuring geometry x x x difficult b Cleaning of the measuring geometry x not possible c c c Prevention of sedimentation/sepa- x x x ration/phase separation d d d d Avoidance of wall slip x x x x a For homogeneous samples using small cone angles.
b Use of disposable measuring geometries.
Surface-treated (e.g. helical spiral groove).
d Surface-treated (e.g. sandblasting, profiling). The surface roughness, R , of the sandblasted geometry should be at least 10 a times smaller than the measuring gap.
4.4 Selection of the temperature control system
Figure 1 illustrates the best-suited temperature control system for each measurement. It is important to keep in mind that the temperature display can deviate from the true sample temperature. The temperature gradient within the sample can be minimized by using an active or passive hood. Any chemical reactions that take place with air (e.g. oxygen, humidity) can be avoided under an inert gas atmosphere.
ISO 3219-3:2026(en) Figure 1 — Selection of the temperature control system
4.5 Selection of the measuring method
In Table 2, typical rheological requirements are listed with recommendations for meaningful measuring methods. Here, CR stands for controlled rate, CS for controlled stress and CD for controlled deformation (also called controlled strain, see ISO 3219-1:2021, 3.47 and Table 1). During execution of the measurement, it shall be ensured that the work is carried out within the specifications of the measuring device used. Table 2 — Selection of measuring methods for viscometers and rheometers based on the example of coating materials a a b Shelf life/sedimentation/phase separation x x x x x x Starting pumping (yield point) x x x Sagging/levelling x x Pumping/mixing/stirring x x Painting/brushing/blade application (manual) x x x a Evaluation via zero shear viscosity or yield point (if applicable, adaptation via rheological model function). b For determination of the linear-viscoelastic range (LVR).
Only in oscillation.
Application Flow curve and viscosity curve CR Flow curve and viscosity curve CS Amplitude sweep CD or CS Frequency sweep CD or CS Creep test Step test Rotation - rotation - rotation or Oscillation – oscillation – oscillation or Oscillation - rotation - oscillation Temperature sweep/time-dependent test in rotation or oscillation
ISO 3219-3:2026(en) TTabablele 2 2 ((ccoonnttiinnueuedd)) Spraying x x x High-speed coating x Gel time/open time/processing time x
Gel point x
Temperature behaviour (glass transition tem- x perature, crystallization temperature, melting temperature, crosslinking temperature) Hardening/crosslinking x a Evaluation via zero shear viscosity or yield point (if applicable, adaptation via rheological model function). b For determination of the linear-viscoelastic range (LVR).
Only in oscillation.
The shear rate range plays an important role in rotational measurements. Representative shear rate ranges are shown in Figure 2 for selected applications from Table 2. The shear rates may, for example, be calculated as described in Annex A.
Figure 2 — Typical shear rate ranges for different applications based on the example of coating materials Typical rheological parameters are shown in Table 3 with reference to the corresponding subclause in this document.
Application Flow curve and viscosity curve CR Flow curve and viscosity curve CS Amplitude sweep CD or CS Frequency sweep CD or CS Creep test Step test Rotation - rotation - rotation or Oscillation – oscillation – oscillation or Oscillation - rotation - oscillation Temperature sweep/time-dependent test in rotation or oscillation
ISO 3219-3:2026(en) Table 3 — Rheological parameters and references Rheological parameter Description in subclauses Flow behaviour 7.2.3 Flow curves and viscosity curves
7.2.1.2 Specification of a constant shear stress (evaluation via the creep curve)
7.2.3.1 Specification of a variable shear rate (evaluation via curve fitting; or via the
Yield point viscosity maximum method)
7.2.3.2 Specification of a variable shear stress (analogous to 7.2.3.1; as a plateau value
or by means of regression lines)
7.3.1 Time-dependent tests in oscillation
Gel point
7.3.2 Temperature-dependent tests in oscillation
7.2.1.1 Specification of a constant shear rate
Gel time
7.2.1.2 Specification of a constant shear stress
Processing time
7.2.2 Temperature-dependent tests in rotation
Processing temperature
7.3.1 Time-dependent tests in oscillation
Crosslinking temperature
7.3.2 Temperature-dependent tests in oscillation
Glass transition temperature 7.3.2 Temperature-dependent tests in oscillation Hysteresis area 8.2 Flow curves and hysteresis area Crystallization temperature 7.3.2 Temperature-dependent tests in oscillation Linearity limit 7.3.3 Amplitude sweeps (end of the LVR) LVR (linear-viscoelastic
7.3.3 Amplitude sweeps
range) Master curve 8.4. Master curve by means of time/temperature shift of frequency sweeps
7.2.1.3 Specification of a constant shear strain (value of the initial shear relaxation
modulus)
7.3.4 Frequency sweeps (position of the crossover point of the shear storage modulus
Average molar mass and shear loss modulus)
8.4 Master curve by means of time/temperature shift of frequency sweeps (position
of the crossover point of the shear storage modulus and shear loss modulus)
7.2.1.3 Specification of a constant shear strain (via the time-based curve of the shear
relaxation modulus)
7.3.4 Frequency sweeps (position of the crossover point of the shear storage modulus
Molar mass distribution and shear loss modulus) 8.4. Master curve by means of time/temperature shift of frequency sweeps (position of the crossover point of the shear storage modulus and shear loss modulus)
7.2.1.2 Specification of a constant shear stress (via the creep curve)
7.2.3.1 Specification of a variable shear rate (as a plateau value at low shear rates)
Zero shear viscosity
7.2.3.2 Specification of a variable shear stress (analogous to 7.2.3.1)
7.3.4 Frequency sweeps (plateau of the viscosity at low frequencies)
Recovery 8.6 Creep and recovery test
7.2.1.2 Specification of a constant shear stress (via creep test)
Shear compliance
8.6 Creep and recovery test (analogous to 7.2.1.2)
Melting temperature 7.3.2 Temperature-dependent tests in oscillation
7.3.1 Time-dependent tests in oscillation
Sol/gel transition
7.3.2 Temperature-dependent tests in oscillation
Structure recovery 8.5 Step tests for determination of the time-dependent structural change Thixotropy 8.5 Step tests for determination of the time-dependent structural change
ISO 3219-3:2026(en)
5 Preconditions for a measurement
5.1 Ambient conditions
The ambient conditions under which the rheometer is operated shall meet the requirements of the manufacturer of the apparatus.
It is recommended that the measuring device is placed in a climate-controlled room [temperature:
(23 ± 2) °C, relative humidity: (50 ± 5) %].
5.2 Sample preparation, filling and cleaning
The sample preparation can play a decisive role for the reproducibility of the measurement. In order to obtain comparable results, it is necessary to coordinate all aspects of the sample preparation between the interested parties, to adhere to these and to log any deviations (Clause 9). Before the measurement, the measuring geometry shall be completely cleaned, with no residue left behind. In the ideal scenario, the sample should be filled in such a way that the smallest possible shear load is applied so that any damage to the structure is kept as small as possible. After the approach to the measuring gap, the sample shall fill the measuring gap completely and shall be free of air bubbles. With plate-plate and cone-plate measuring geometries, the sample is trimmed, which means that any excess leaked from the measuring gap is removed using a suitable tool. For specific sample types, such as asphalt binders or hotmelt adhesives, it can be necessary to adjust the temperature of the trimming tool, but to use a certain temperature range for the trimming tool. For the removal of the excess sample, first a trimming position above the measuring position is approached, the sample is trimmed, and then the measuring position is approached, which produces a slightly curved sample edge. Experience shows that a gap is chosen for the trimming position that is around 1 % to 5 % larger than the gap width. During trimming, it is recommended to block the upper measuring geometry so that any structure present in the sample cannot be damaged due to twisting of the geometry. The excess sample can be scraped off e.g. using a spatula (see Figure 3). Use of a lower recess plate makes the trimming process easier. Figure 3 a) through to Figure 3 d) show the complete process of filling a plate-plate measuring geometry, including trimming.
ISO 3219-3:2026(en) a) Sample in the trimming position b) Scraping off the excess sample in the trimming position, e.g. with a spatula c) Trimmed sample in the trimming position d) Optimum sample filling in the measuring position Figure 3 — Filling a plate-plate measuring geometry
6 Performance of the measurement
6.1 General
Depending on the sample and the rheological issue being investigated, a suitable measuring system and a meaningful measuring method shall be selected from the guidelines and requirements (Clause 4) and, where applicable, with the aid of preliminary tests. The measuring system used shall be adjusted, calibrated and verified. Prior to the start of the measurement, it can be necessary to allow a longer waiting time for recovery of the sample structure or to carry out a specified pre-shear, independent of the temperature control of the sample.
All measuring conditions and boundary conditions shall be reported (see Clause 9).
6.2 Measuring profile
Once a measuring method has been chosen, the measuring profile is set. It may contain various measuring segments (see Clause 7), each of which is specified by the preset parameter(s) and by the duration and number of measuring points.
The preset parameters of a measuring segment may be constant (see Figure 4) or may be varied over time. The variation may be continuous [continuous ramp, see Figure 5 a)] or discrete [stepped ramp, see Figure 5 b)].
ISO 3219-3:2026(en) Key X time, t
Y preset parameter: e.g. shear rate, γ , shear stress, τ, shear strain, γ, temperature, T t end time to switch off the preset parameter n Figure 4 — Illustration of a constant preset parameter a) Continuous ramp b) Stepped ramp Key X time, t
Y preset parameter: e.g. shear rate, γ , shear stress, τ, shear strain, γ, temperature, T Figure 5 — Illustration of the continuous or discrete variation of the preset parameter Detailed specifications of the measuring profile shall be provided to ensure reproducible measuring results and comparable evaluations.
In the case of continuous ramps, deviations will increase with increasing rates of change of the specified conditions (e.g. the shear load or temperature).
The adjustment time of each measuring point shall be considered. −1 For rotational tests with shear rates below 1 s , a measuring point duration of at least the inverse shear rate (i.e. 1/shear rate) should be specified. For oscillatory tests, at least one oscillation period should be used for the integration time, i.e. the measuring time to calculate a data point.
ISO 3219-3:2026(en)
7 Basic tests and example evaluations
7.1 General
A basic test consists of a single measuring segment.
The basic tests in 7.2 and 7.3 are described in terms of the following aspects:
a) measuring profiles; b) examples of measuring curves; c) typical evaluations.
7.2 Rotational tests
7.2.1 Time-dependent tests in rotation
7.2.1.1 Specification of a constant shear rate
a) Measuring profiles
A constant, application-relevant value is specified for the shear rate, γ . All other values are also kept constant (see Figure 4).
b) Examples of measuring curves Figure 6 shows examples of measuring curves for time-dependent tests.
Key X time, t Y shear viscosity, η 1 result that is independent of time 2 initial decrease in the shear viscosity as a function of time (e.g. in the case of shear-induced disentangling of polymer molecules or during the orientation of particles or if the sample is in a non-relaxed/non-equilibrium state before the start of the measurement) 3 initial increase in the shear viscosity as a function of time (e.g. in the case of a shear-induced increase in interactions or if the sample is in a non-relaxed/non-equilibrium state before the start of the measurement) 4 increase in the shear viscosity as a function of time (e.g. due to gelation, chemical crosslinking) 5 decrease in the shear viscosity as a function of time (e.g. due to phase separation, shear heating, degradation of polymer molecules) Figure 6 — Example of measuring curves for time-dependent tests c) Typical evaluations 1) preliminary test to determine parameters for following tests (e.g. maximum test duration, limiting values of the shear load);
...
Contents
- 1 Domaine d’application
- 2 Références normatives
- 3 Termes et définitions
- 3.1 température de cristallisation
- 3.2 point de gélification
- 3.3 temps de gel
- 3.4 température de transition vitreuse
- 3.5 préparation des échantillons
- 3.6 module de relaxation du cisaillement
- 3.7 transition sol-gel
- 3.8 ajustage
- 4 Lignes directrices et exigences pour la sélection
- 4.1 Généralités
- 4.2 Sélection du dispositif de mesure
- 4.3 Sélection de la géométrie de mesure
- 4.4 Caractéristiques du système de régulation de la température
- 4.5 Sélection de la méthode de mesurage
- 8.4 Courbe maîtresse au moyen d'un décalage temps/température des balayages de la
- 8.6 Essai de fluage et de récupération (analogue à 7.2.1.2)
- 8.5 Essais avec paliers pour la détermination de la modification structurelle en fonc-
- 8.5 Essais avec paliers pour la détermination de la modification structurelle en fonc-
- 5 Conditions préalables à un mesurage
- 5.1 Conditions ambiantes
- 5.2 Préparation, remplissage et nettoyage des échantillons
- 6 Réalisation de la mesure
- 6.1 Généralités
- 6.2 Profil de mesure
- 7 Essais de base et exemples d'évaluations
- 7.1 Généralités
- 7.2 Essais rotationnels
- 7.3 Essais oscillatoires
- 8 Essais de base combinés
- 8.1 Généralités
- 8.2 Courbes d'écoulement et zone d'hystérésis
- 8.3 Essais de température sous forme d’essais cycliques
- 8.4 Courbe maîtresse au moyen du décalage temps/température des balayages de la
- 8.5 Essais avec paliers pour la détermination de la modification structurelle en fonction du
- 8.6 Essais de fluage et de récupération
- 9 Rapport d’essai
- Annexe A
- Bibliographie
1 Domaine d’application
Le présent document fournit des exigences et des lignes directrices pour la sélection du dispositif de mesure, de la géométrie de mesure et de l'unité de contrôle de la température pour les essais impliquant la rhéométrie rotative et oscillatoire. Les principes généraux de performance des essais sont décrits et des exemples d'évaluations de rhéométrie rotative et oscillatoire sont fournis.
2 Références normatives
Les documents suivants sont cités dans le texte de sorte qu’ils constituent, pour tout ou partie de leur contenu, des exigences du présent document. Pour les références datées, seule l’édition citée s’applique. Pour les références non datées, la dernière édition du document de référence s'applique (y compris les éventuels amendements).
ISO 3219-1:2021, Rhéologie — Partie 1: Vocabulaire et symboles pour la rhéométrie rotative et oscillatoire
3 Termes et définitions
Pour les besoins du présent document, les termes et les définitions de l’ISO 3219-1 ainsi que les suivants s’appliquent.
L’ISO et l’IEC tiennent à jour des bases de données terminologiques destinées à être utilisées en normalisation, consultables aux adresses suivantes:
- ISO Online browsing platform: disponible à l’adresse https:// www .iso .org/ obp
- IEC Electropedia: disponible à l’adresse https:// www .electropedia .org/
3.1 température de cristallisation
T
température à laquelle la cristallisation commence et l'ordre partiel est établi lorsque la matière fondue se solidifie Note 1 à l'article: L’unité de la température de cristallisation, T , est le degré Celsius (°C).
3.2 point de gélification
point de transition du sol au gel en oscillation dans la plage viscoélastique linéaire, où l'angle de perte, δ, est indépendant de la fréquence Note 1 à l'article: Le point de gélification peut être un temps ou une température. Note 2 à l'article: Contrairement à cette définition, le point de gélification est généralement mesuré à une fréquence et déterminé sur la base de l’angle de perte δ = 45° ou sur la base du facteur de perte tan δ = 1.
ISO 3219-3:2026(fr)
3.3 temps de gel
délai d’utilisation après mélange période pendant laquelle un produit de réticulation peut être traité alors qu'un critère spécifié par l'utilisateur est satisfait Note 1 à l'article: Le temps de gel peut être déterminé en oscillation dans la plage non linéaire ou en rotation. Note 2 à l'article: Dans la pratique, des géométries de mesure relatives ou d'autres méthodes sont également utilisées, telles que des coupelles d'écoulement.
Note 3 à l'article: Les valeurs limites pour la viscosité ou les modules peuvent être utilisées comme critères définis par l'utilisateur.
3.4 température de transition vitreuse
T g température caractéristique de la plage de températures auquel se produit la transformation vitreuse Note 1 à l'article: La température de transition vitreuse dépend du mode opératoire d'essai et des conditions de mesure.
3.5 préparation des échantillons
manipulation, conditionnement et charge appliqués à l'échantillon avant le mesurage
3.6 module de relaxation du cisaillement
G rapport de la contrainte de cisaillement τ à la déformation de cisaillement γ pendant l'essai de relaxation Note 1 à l'article: L’unité du module de relaxation de cisaillement G est Pascal (Pa).
3.7 transition sol-gel
plage de transition du liquide viscoélastique (sol) au solide viscoélastique (gel)
3.8 ajustage
élimination du volume d'échantillon excédentaire Note 1 à l'article: Avec les géométries de mesure plateau/plateau et cône/plateau, une position d'ajustage est souvent choisie comme étant plus grande que la largeur de fente et au niveau de laquelle le volume excédentaire est retiré de sorte que l'échantillon soit à ras du bord de la géométrie de mesure. Lorsque la position de mesurage est ensuite approchée, cela garantit un remplissage optimal de l'espace.
4 Lignes directrices et exigences pour la sélection
4.1 Généralités
Le choix du dispositif de mesure, de la géométrie de mesure, de l'unité de contrôle de la température, de la méthode de mesure et d'autres accessoires dépend des propriétés de l'échantillon et des exigences rhéologiques qui en résultent.
4.2 Sélection du dispositif de mesure
L’ISO 3219-2 décrit les différences entre les différents dispositifs de mesure pour la détermination des propriétés rhéologiques et leurs configurations. Un viscosimètre rotatif peut être utilisé pour déterminer la viscosité en cisaillement. Avec un rhéomètre rotatif, il est en outre possible d'effectuer tous les essais rotatifs et oscillatoires de base (voir Article 7). Pour cette raison, outre la fonction de viscosité, il peut être
ISO 3219-3:2026(fr) utilisé pour déterminer les propriétés viscoélastiques, par exemple le module de cisaillement au stockage et le module de perte de cisaillement.
Pour les mesures viscométriques, tous les viscosimètres conviennent principalement, quel que soit le type de roulement utilisé. Pour les mesurages oscillatoires, les essais de fluage et les essais de relaxation, il convient d'utiliser des dispositifs à faible frottement interne, par exemple des dispositifs munis de roulements pneumatiques ou magnétiques. Certains dispositifs équipés de roulements à billes offrent également la possibilité d'effectuer des mesurages oscillatoires, mais uniquement dans une plage de mesure restreinte. Pendant l'exécution du mesurage, il faut s'assurer que le travail est effectué dans les limites des spécifications du dispositif de mesure, par exemple en tenant compte des limites de couple et de la vitesse d'acquisition des données.
Afin de couvrir l’éventail d’applications le plus large possible, le viscosimètre ou le rhéomètre doit pouvoir fonctionner avec différentes géométries de mesure. Les plages de couple et d’angle de rotation, et donc la plage de mesure accessible, dépendent du dispositif de mesure.
4.3 Sélection de la géométrie de mesure
La géométrie de mesure doit être choisie en tenant compte du dispositif de mesure disponible, des propriétés de l'échantillon et du profil de mesure. Le Tableau 1 contient des recommandations sur la sélection des différentes géométries de mesure. Ici, CC signifie cylindre coaxial, CP cône-plateau, PP plateauplateau, DG double fente et RMG géométries de mesure relatives (par exemple, capteur à aube, voir détails dans l’ISO 3219-2). En général, lors du choix d'une géométrie de mesure absolue, la taille maximale des hétérogénéités, par exemple la taille des particules dans l'échantillon, ne doit pas dépasser 10 % de la largeur de fente. Si, par conséquent, il n'est pas possible d'utiliser des géométries de mesure absolues, il est alors nécessaire de sélectionner des géométries de mesure relatives. Les accessoires facultatifs, tels qu’une hotte avec piège à solvant pour éviter la perte de solvant, sont décrits dans l’ISO 3219-2. Tableau 1 — Recommandations pour la sélection de la géométrie de mesure Critères Géométrie de mesure recommandée Cylindres Cône-plateau Plateau Double écarte- Humidité coaxiaux ment CC CP PP B, So RMG Propriétés de l'échantillon qui déterminent le choix de la géométrie de mesure a Faible viscosité x x x Viscosité élevée x x Pâteux, gélifiant x Faible tension superficielle x x Inhomogénéité, par exemple bulles x d'air, mousse Autres critères pour la sélection d'une géométrie de mesure Disponibilité d'un petit volume x x d'échantillon b Nettoyage de la géométrie de x x x mesure difficile a Pour les échantillons homogènes utilisant de petits angles de cône. b Utilisation de géométries de mesure jetables.
Avec traitement de surface (par exemple rainure hélicoïdale).
d Avec traitement de surface (par exemple sablage, profilage). Il convient que la rugosité de surface, R , de la géométrie sablée a soit au moins 10 fois plus petite que la fente de mesure.
ISO 3219-3:2026(fr) TTabableleaauu 1 1 ((ssuuiitte)e) Critères Géométrie de mesure recommandée Cylindres Cône-plateau Plateau Double écarte- Humidité coaxiaux ment CC CP PP B, So RMG b Nettoyage de la géométrie de x mesure impossible c c c Prévention de la sédimentation/ x x x séparation/séparation des phases d d d d Prévention du glissement sur la x x x x paroi a Pour les échantillons homogènes utilisant de petits angles de cône. b Utilisation de géométries de mesure jetables.
Avec traitement de surface (par exemple rainure hélicoïdale).
d Avec traitement de surface (par exemple sablage, profilage). Il convient que la rugosité de surface, R , de la géométrie sablée a soit au moins 10 fois plus petite que la fente de mesure.
4.4 Caractéristiques du système de régulation de la température
La Figure 1 représente le système de régulation de la température le mieux adapté pour chaque mesurage. Il est important de garder à l'esprit que l'affichage de la température peut s'écarter de la température réelle de l'échantillon. Le gradient de température dans l'échantillon peut être réduit au minimum en utilisant une hotte active ou passive. Toute réaction chimique qui se produit avec l'air (par exemple oxygène, humidité) peut être évitée dans une atmosphère de gaz inerte.
Figure 1 — Sélection du système de régulation de la température
4.5 Sélection de la méthode de mesurage
Dans le Tableau 2, les exigences rhéologiques types sont énumérées avec des recommandations pour des méthodes de mesure significatives. Ici, CR signifie vitesse contrôlée, CS contrainte contrôlée et CD déformation contrôlée (également appelée déformation contrôlée, voir l’ISO 3219-1:2021, 3.47 et Tableau 1). Pendant l'exécution du mesurage, il faut s'assurer que le travail est effectué dans les limites des spécifications du dispositif de mesure utilisé.
ISO 3219-3:2026(fr) Tableau 2 — Sélection des méthodes de mesure pour les viscosimètres et les rhéomètres sur la base de l'exemple des produits de peinture a a b Délai de stockage/sédimentation/séparation des x x x x x x phases Démarrage du pompage (seuil d’écoulement) x x x Affaissement/nivellement x x Pompage/mélange/agitation x x Peinture/brossage/application à la lame x x x (manuelle) Pulvérisation x x x Revêtement grande vitesse x Temps de gel/temps d'assemblage ouvert/temps x de traitement
Point de gélification x
Comportement en température (température x de transition vitreuse, température de cristallisation, température de fusion, température de réticulation) Durcissement/réticulation x a Évaluation via la viscosité en cisaillement zéro ou le seuil d’écoulement (le cas échéant, adaptation via la fonction du modèle rhéologique).
b Pour la détermination de la plage viscoélastique linéaire (LVR).
Uniquement en oscillation.
La plage de vitesse de cisaillement joue un rôle important dans les mesurages rotationnels. Des plages de vitesse de cisaillement représentatives sont représentées à la Figure 2 pour des applications sélectionnées à partir du Tableau 2. Les vitesses de cisaillement peuvent, par exemple, être calculées comme décrit à l’Annexe A.
Application Courbe d'écoulement et courbe de viscosité CR Courbe d'écoulement et courbe de viscosité CS Balayage de l'amplitude CD ou CS Balayage de la fréquence CD ou CS Essai de fluage Essai avec paliers Rotation – rotation – rotation ou Oscillation – oscillation – oscillation ou Oscillation – rotation – oscillation Essai de balayage thermique/dépendant du temps en rotation ou oscillation
ISO 3219-3:2026(fr) Figure 2 — Plages types de vitesse de cisaillement pour différentes applications basées sur l'exemple des produits de peinture Les paramètres rhéologiques types sont indiqués dans le Tableau 3 en référence au paragraphe correspondant du présent document.
Tableau 3 — Paramètres rhéologiques et références Paramètre rhéologique Description dans les paragraphes Comportement à l'écoule-
7.2.3 Courbes d'écoulement et courbes de viscosité
ment
7.2.1.2 Spécification d’une contrainte de cisaillement constante (évaluation via la
courbe de fluage)
7.2.3.1 Spécification d’une vitesse de cisaillement variable (évaluation par ajustement
Seuil d’écoulement de courbe; ou par la méthode de viscosité maximale) 7.2.3.2 Spécification d’une contrainte de cisaillement variable (analogue à 7.2.3.1; sous forme de valeur de plateau ou au moyen de droites de régression)
7.3.1 Essais en oscillation en fonction du temps
Point de gélification
7.3.2 Essais dépendant de la température en oscillation
7.2.1.1 Spécification d'une vitesse de cisaillement constante
Temps de gel
7.2.1.2 Spécification d'une contrainte de cisaillement constante
Durée de traitement
7.2.2 Essais en rotation dépendant de la température
Température de traitement
7.3.1 Essais en oscillation en fonction du temps
Température de réticulation
7.3.2 Essais dépendant de la température en oscillation
Température de transition
7.3.2 Essais dépendant de la température en oscillation
vitreuse Zone d'hystérésis 8.2 Courbes d'écoulement et zone d'hystérésis Température de cristallisa-
7.3.2 Essais dépendant de la température en oscillation
tion Limite de linéarité 7.3.3 Balayages de l’amplitude (fin de la LVR) LVR (plage viscoélastique
7.3.3 Balayages de l’amplitude
linéaire) 8.4. Courbe maîtresse au moyen du décalage temps/température des balayages de la Courbe maîtresse fréquence
ISO 3219-3:2026(fr) TTabableleaauu 3 3 ((ssuuiitte)e) Paramètre rhéologique Description dans les paragraphes
7.2.1.3 Spécification d’une déformation de cisaillement constante (valeur du module
initial de relaxation du cisaillement)
7.3.4 Balayages de la fréquence (position du point de croisement du module de cisail-
Masse molaire moyenne lement au stockage et du module de perte de cisaillement)
8.4 Courbe maîtresse au moyen d'un décalage temps/température des balayages de la
fréquence (position du point de croisement du module de cisaillement au stockage et du module de perte de cisaillement)
7.2.1.3 Spécification d’une déformation de cisaillement constante (via la courbe tem-
porelle du module de relaxation du cisaillement)
7.3.4 Balayages de la fréquence (position du point de croisement du module de cisail-
Distribution de masses lement au stockage et du module de perte de cisaillement) molaires 8.4. Courbe maîtresse au moyen d'un décalage temps/température des balayages de la fréquence (position du point de croisement du module de cisaillement au stockage et du module de perte de cisaillement)
7.2.1.2 Spécification d’une contrainte de cisaillement constante (via la courbe de
fluage)
7.2.3.1 Spécification d’une vitesse de cisaillement variable (comme valeur de plateau à
Viscosité en cisaillement zéro faibles vitesses de cisaillement)
7.2.3.2 Spécification d’une contrainte de cisaillement variable (analogue à 7.2.3.1)
7.3.4 Balayages de la fréquence (plateau de la viscosité aux basses fréquences)
Récupération 8.6 Essai de fluage et de récupération
7.2.1.2 Spécification d’une contrainte de cisaillement constante (par essai de fluage)
Conformité de cisaillement
8.6 Essai de fluage et de récupération (analogue à 7.2.1.2)
Température de fusion 7.3.2 Essais dépendant de la température en oscillation
7.3.1 Essais en oscillation en fonction du temps
Transition sol-gel
7.3.2 Essais dépendant de la température en oscillation
8.5 Essais avec paliers pour la détermination de la modification structurelle en fonc-
Récupération de la structure tion du temps
8.5 Essais avec paliers pour la détermination de la modification structurelle en fonc-
Thixotropie tion du temps
5 Conditions préalables à un mesurage
5.1 Conditions ambiantes
Les conditions ambiantes dans lesquelles le rhéomètre est utilisé doivent satisfaire aux exigences du fabricant de l'appareillage.
Il est recommandé de placer le dispositif de mesure dans une pièce climatisée [température: (23 ± 2) %, hygrométrie relative: (50 ± 5) %].
5.2 Préparation, remplissage et nettoyage des échantillons
La préparation de l'échantillon peut jouer un rôle décisif pour la reproductibilité du mesurage. Afin d'obtenir des résultats comparables, il est nécessaire de coordonner tous les aspects de la préparation des échantillons entre les parties intéressées, de respecter ceux-ci et d'enregistrer tout écart (Article 9). Avant le mesurage, la géométrie de mesure doit être entièrement nettoyée, sans laisser de résidus. Dans le scénario idéal, il convient que l'échantillon soit rempli de manière à ce que la charge de cisaillement la plus faible possible soit appliquée de sorte que tout endommagement de la structure soit aussi faible que possible. Après l'approche de la fente de mesure, l'échantillon doit remplir complètement la fente de mesure et être exempt de bulles d'air.
ISO 3219-3:2026(fr) Avec les géométries de mesure plateau/plateau et cône/plateau, l'échantillon est taillé, ce qui signifie que tout excès de fuite de la fente de mesure est éliminé à l'aide d'un outil approprié. Pour des types d'échantillons spécifiques, tels que les liants bitumineux ou les adhésifs thermofusibles, il peut être nécessaire d'ajuster la température de l'outil d’ajustage, mais d'utiliser une certaine plage de températures pour l'outil d’ajustage. Pour l'élimination de l'échantillon en excès, une position d'ajustage au-dessus de la position de mesure est tout d'abord approchée, l'échantillon est taillé, puis la position de mesure est approchée, ce qui produit un bord d'échantillon légèrement incurvé. L'expérience montre qu'un espace est choisi pour la position d'ajustage qui est supérieur d'environ 1 % à 5 % à la largeur de fente. Pendant l’ajustage, il est recommandé de bloquer la géométrie de mesure supérieure de sorte que toute structure présente dans l'échantillon ne puisse pas être endommagée en raison d'une torsion de la géométrie. L'échantillon en excès peut être gratté, par exemple à l'aide d'une spatule (voir Figure 3). L'utilisation d'une plaque d’évidement inférieure facilite le processus d’ajustage. La Figure 3 a) à la Figure 3 d) montrent le processus complet de remplissage d'une géométrie de mesure plateau/plateau, y compris l’ajustage.
a) Échantillon en position d'ajustage b) Raclage de l'échantillon en excès en position d'ajustage, par exemple à l'aide d'une spatule c) Échantillon taillé en position d'ajustage d) Remplissage optimal de l'échantillon à la position de mesurage Figure 3 — Remplissage d’un géométrie de mesure plateau/plateau
6 Réalisation de la mesure
6.1 Généralités
En fonction de l'échantillon et du problème rhéologique étudié, un système de mesure approprié et une méthode de mesure significative doivent être sélectionnés à partir des lignes directrices et des exigences (Article 4) et, le cas échéant, à l'aide d'essais préliminaires. Le système de mesure utilisé doit être réglé, étalonné et vérifié. Avant le début du mesurage, il peut être nécessaire de prévoir un temps d'attente
ISO 3219-3:2026(fr) plus long pour la récupération de la structure de l'échantillon ou d'effectuer un pré-cisaillement spécifié, indépendamment du contrôle de la température de l'échantillon. Toutes les conditions de mesure et les conditions aux limites doivent être consignées dans le rapport (voir Article 9).
6.2 Profil de mesure
Une fois qu'une méthode de mesure a été choisie, le profil de mesure est défini. Il peut contenir différents segments de mesure (voir Article 7) dont chacun est spécifié par le ou les paramètres prédéfinis et par la durée et le nombre de points de mesure.
Les paramètres prédéfinis d'un segment de mesure peuvent être constants (voir Figure 4) ou peuvent varier dans le temps. La variation peut être continue [montée continue, voir Figure 5 a)] ou discrète [montée par paliers, voir Figure 5 b)].
...