Low alloyed steel — Determination of C, Si, Mn, P, S, Cr, Ni, Al, Ti and Cu - Glow discharge optical emission spectrometry (routine method)

ISO 19272:2015 specifies a glow discharge optical emission spectrometric (GD-OES) method for determination of the mass fraction Carbon, Silicon, Manganese, Phosphorus, Sulfur, Chromium, Nickel, Aluminium, Titanium and Copper in low alloyed steels. The content ranges to which the method is applicable are shown in Table 1. Table 1 ? Content ranges Element Content ranges mass fraction % C 0,060 to 0,35 Si 0,14 to 1,50 Mn 0,090 to 0,70 P 0,010to 0,070 S 0,005 to 0,050 Cr 0,008 to 0,65 Ni 0,050 to 0,50 Al 0,006 to 0,90 Ti 0,014 to 0,13 Cu 0,005 to 1,00

Aciers faiblement alliés — Détermination de C, Si, Mn, P, S, Cr, Ni, Al, Ti et Cu — Spectrométrie d'émission optique à décharge luminescente (méthode de routine)

General Information

Status
Published
Publication Date
13-Aug-2015
Current Stage
9093 - International Standard confirmed
Completion Date
05-Dec-2020
Ref Project

Buy Standard

Standard
ISO 19272:2015 - Low alloyed steel -- Determination of C, Si, Mn, P, S, Cr, Ni, Al, Ti and Cu - Glow discharge optical emission spectrometry (routine method)
English language
32 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 19272
First edition
2015-08-15
Low alloyed steel — Determination
of C, Si, Mn, P, S, Cr, Ni, Al, Ti and Cu
- Glow discharge optical emission
spectrometry (routine method)
Aciers faiblement alliés — Détermination de C, Si, Mn, P, S, Cr, Ni, Al,
Ti et Cu — Spectrométrie d’émission optique à décharge luminescente
(méthode de routine)
Reference number
ISO 19272:2015(E)
©
ISO 2015

---------------------- Page: 1 ----------------------
ISO 19272:2015(E)

COPYRIGHT PROTECTED DOCUMENT
© ISO 2015, Published in Switzerland
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2015 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 19272:2015(E)

Contents Page
Foreword .iv
1 Scope . 1
2 Normative references . 1
3 Principle . 2
4 Apparatus . 2
4.1 Glow discharge optical emission spectrometer. 2
4.1.1 General. 2
4.1.2 Short term stability . 2
4.1.3 Detection limit . 2
4.1.4 Data acquisition and processing system . 2
4.1.5 Others . 2
5 Sampling . 3
6 Procedure. 3
6.1 Sample preparation . 3
6.2 Selection of spectral lines . 3
6.3 Optimization of the instrument . 4
6.3.1 General. 4
6.3.2 Setting the discharge parameters of a direct current (DC) source . 4
6.3.3 Setting the discharge parameters of radiofrequency (RF) source . 5
6.3.4 Optimization of the detection system . 6
6.3.5 Pre-burning time and integration time . 6
6.3.6 Validation of the instrumental parameters . 6
6.4 Calibration . 6
6.5 Checking of the trueness of the method . . 7
6.6 Drift correction . 7
6.7 Analysis of the samples . 7
7 Expression of results . 7
7.1 Method of calculation . 7
7.2 Precision . 7
8 Test report .11
Annex A (informative) Detection limit.12
Annex B (informative) Additional information on international interlaboratory test .14
Annex C (informative) Graphical representation of precision data .22
Bibliography .32
© ISO 2015 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 19272:2015(E)

Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation on the meaning of ISO specific terms and expressions related to conformity
assessment, as well as information about ISO’s adherence to the WTO principles in the Technical
Barriers to Trade (TBT) see the following URL: Foreword - Supplementary information
The committee responsible for this document is ISO/TC 17, Steel, Subcommittee SC 1, Method of
determination of chemical composition.
iv © ISO 2015 – All rights reserved

---------------------- Page: 4 ----------------------
INTERNATIONAL STANDARD ISO 19272:2015(E)
Low alloyed steel — Determination of C, Si, Mn, P, S,
Cr, Ni, Al, Ti and Cu - Glow discharge optical emission
spectrometry (routine method)
1 Scope
This International Standard specifies a glow discharge optical emission spectrometric (GD-OES)
method for determination of the mass fraction Carbon, Silicon, Manganese, Phosphorus, Sulfur,
Chromium, Nickel, Aluminium, Titanium and Copper in low alloyed steels.
The content ranges to which the method is applicable are shown in Table 1.
Table 1 — Content ranges
Content ranges
Element mass fraction
%
C 0,060 to 0,35
Si 0,14 to 1,50
Mn 0,090 to 0,70
P 0,010to 0,070
S 0,005 to 0,050
Cr 0,008 to 0,65
Ni 0,050 to 0,50
Al 0,006 to 0,90
Ti 0,014 to 0,13
Cu 0,005 to 1,00
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable to its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
ISO 5725-1, Accuracy (trueness and precision) of measurement methods and results — Part 1: General
principles and definitions
ISO 5725-2, Accuracy (trueness and precision) of measurement methods and results — Part 2: Basic method
for the determination of repeatability and reproducibility of a standard measurement method
ISO 5725-3, Accuracy (trueness and precision) of measurement methods and results — Part 3: Intermediate
measures of the precision of a standard measurement method
ISO 14284, Steel and iron — Sampling and preparation of samples for the determination of chemical
composition
ISO 14707, Surface chemical analysis — Glow discharge optical emission spectrometry (GD-OES) —
Introduction to use
© ISO 2015 – All rights reserved 1

---------------------- Page: 5 ----------------------
ISO 19272:2015(E)

3 Principle
The method used involves the following steps:
a) a sample with a flat and smooth surface is used as the cathode of a direct current or radio frequency
glow discharge device;
b) cathodic sputtering of the sample surface, atomization of the sputtered atoms and ions from the
sample surface; excitation of these atoms and ions in the plasma formed in the glow discharge device;
c) spectrometric measurement of the intensity of the emitted light from the ions or the atoms of the
elements to be determined and, optionally, the emitted light from iron at 371,994 nm or 271,441 nm
or another appropriate wavelength (if internal standard is used);
d) conversion of the measured signals to the contents through calibration curves established by
measuring certified reference materials.
4 Apparatus
4.1 Glow discharge optical emission spectrometer
4.1.1 General
An optical emission spectrometer system consists of a Grimm type or similar glow discharge source
(direct current or radio frequency powered) and a simultaneous optical spectrometer as described in
ISO 14707, with suitable spectral lines for the elements to be determined. A sequential optical system
may also be used alone or combined with simultaneous channels.
A 2 mm to 8 mm range of the inner diameter of the anode of the glow discharge is recommended. A
cooling device is also recommended, but not strictly required for implementation of the method.
The anode-cathode gap is normally around 0,1 mm to 0,3 mm; otherwise, follow the instrument
manufacturer’s instructions.
4.1.2 Short term stability
Carry out at least 10 measurements of the emission intensity of an element having a content around
the corresponding upper limit specified in this International Standard on an appropriate sample. Allow
the discharge to stabilize for at least 50 s before each measurement. The data acquisition time should
be in the range of 5 s to 30 s. Each measurement shall be carried out on a newly polished surface of
the sample. Calculate the corresponding standard deviation and mean. The relative standard deviation
(RSD) should not exceed 5 % for contents less than 0,5 %, otherwise, the RSD should not exceed 3 %.
4.1.3 Detection limit
Detection limits may be determined by either the SNR method or SBR-CVB method (see A.1 and A.2).
4.1.4 Data acquisition and processing system
The data acquisition and processing system is conducted by a computer, equipped with software
suitable for controlling the instrument parameters and running the test programs.
4.1.5 Others
For other basic requirements of the instrument, use ISO 14707.
2 © ISO 2015 – All rights reserved

---------------------- Page: 6 ----------------------
ISO 19272:2015(E)

5 Sampling
Carry out sampling in accordance with ISO 14284 or appropriate national standards for steel.
6 Procedure
6.1 Sample preparation
Samples shall be homogeneous, flat and free of porosity. The surface of the sample shall be prepared
by using suitable methods to ensure cleanliness and flatness. Surface preparation may be achieved by
using abrasive-wheel or milling machine. All the reference materials and samples shall be prepared
under the same conditions and their size should be suitable for the glow discharge source used.
6.2 Selection of spectral lines
For each element to be determined, there are a number of spectral lines that can be used. Suitable lines
shall be selected on the basis of several factors, including the spectral range of the spectrometer used,
element concentration, sensitivity of the spectral lines and spectral interference from other elements
present in the samples. Table 2 contains some suggestions concerning suitable spectral lines. With
these lines, no obvious inter-element interferences are observed in low alloyed steels. Spectral lines
other than those listed may be used, if they have favourable characteristics, but special attention needs
to be paid to the influence of co-existing elements and corrections might be required.
Table 2 — Suggested wavelengths for the analysis of low alloy steels
Wavelength
Element
nm
371,994
Fe
271,441
165,700
C
156,143
288,158
Si
251,611
403,449
Mn
257,610
177,497
P
178,287
S 180,731
Cr 425,433
349,296
Ni
341,477
Al 396,152
337,279
Ti
365,350
219,228
Cu
327,396
© ISO 2015 – All rights reserved 3

---------------------- Page: 7 ----------------------
ISO 19272:2015(E)

6.3 Optimization of the instrument
6.3.1 General
Follow the manufacturer’s instructions or other suitable documented procedures to start and set
the GD-OES and let it stabilize. Check, or adjust if necessary, the entrance slit in accordance with the
manufacturer’s instructions to ensure it is correctly positioned so that the intensity is measured at
the peak of the spectral line in order to obtain the best signal to background ratio (make reference to
ISO 14707 for further information). Carry out at least three preliminary discharges on a generic sample
before any measurement procedures.
NOTE Analyte lines are centred by adjusting the entrance slit whilst carrying out discharges on a reference
material. It is advisable to profile the instrument on a daily basis for a newly installed instrument during the first
week of operation and then periodically over time. Profiling is carried out initially on all lines or at least on those
included in each analytical program or by following the recommendations of the instrument manufacturer.
Follow manufacturer’s recommendation or other suitable documented procedures to select or adjust
all instrumental parameters: evacuation time, argon purging time [pumpdown cycles], glow discharge
spectrometer opening mode [startup time], flush time [cleaning time], sputtering time, etc.
6.3.2 Setting the discharge parameters of a direct current (DC) source
DC glow discharge spectrometers usually have three parameters (current, voltage and pressure) to be
controlled. Any two of these three parameters may be fixed to constant values by varying the third one.
The user may adopt the following procedures, or any other mode recommended by the manufacturers.
6.3.2.1 Constant applied voltage and current
The controlled parameters are applied voltage and current. Set the power supply of the glow discharge
source to constant voltage and constant current operation. Firstly, set the voltage and current to typical
values recommended by the manufacturer or reliable values. If the recommended values are not available
or not suitable for the task, following steps could be pursued to determine the optimum parameters.
Set the high voltage of the detectors as described in 6.3.4.
Step one: set an initial value or the value recommended by the manufacturer for current and voltage
(initial discharge condition). For DC glow discharge, electrical current is normally in the range of 5 mA
to 10 mA for a 2 mm to 2,5 mm anode, 15 mA to 45 mA for a 4 mm anode and 40 mA to 100 mA for
a 7 mm to 8 mm anode. Voltage is usually in the range of 600 V to 1250 V. Determine the intensities
corresponding to an appropriate sample under the initial discharge condition. Fix one parameter
(current or voltage), and change the other one gradually. Measure the intensities at least seven times at
each setting. Do this at the entire parameter range allowed by the instrument.
Step two: fix the parameter which changed in step one and gradually change the other parameter.
Repeat the operation as described in step one.
Step three: calculate the CVs of the measurements for each setting. Investigate the influence of the
discharge parameters on the intensities and CVs and fix the optimum instrumental condition.
6.3.2.2 Constant applied current and pressure
The controlled parameters are applied current and pressure. Set the power supply of the glow discharge
source to constant current and constant pressure operation. Firstly, set the current and pressure to
typical values recommended by the manufacturer or values suitably documented. If the recommended
values are not available or not suitable for the task, the following steps could be pursued to determine
the optimum parameters.
Set the high voltage of the detectors as described in 6.3.4.
4 © ISO 2015 – All rights reserved

---------------------- Page: 8 ----------------------
ISO 19272:2015(E)

Step one: set an initial value or the value recommended by the manufacturer for current and pressure
(initial discharge condition). For DC glow discharge, electrical current is normally in the range of 5 mA
to 10 mA for a 2 mm to 2,5 mm anode, 15 mA to 45 mA for a 4 mm anode and 40 mA to 100 mA for a
7 mm to 8 mm anode. Set the initial pressure to a value to obtain a voltage in the range of 600 V to
1250 V or set the initial pressure to a value which is in the middle of the allowed range of the instrument.
Determine the intensities corresponding to an appropriate sample under the initial discharge condition.
Fix one parameter (current or pressure) and change another one gradually. Measure the intensities at
least seven times at each setting. Do this at the entire parameter range allowed by the instrument.
Step two: fix the parameter which changed in step one and gradually change the other parameter.
Repeat the operation as described in step one.
Step three: calculate the CVs of the measurements for each setting. Investigate the influence of the
discharge parameters on the intensities and CVs and fix the optimum instrumental condition.
Use a similar procedure to optimize the analytical parameters for other operation modes.
6.3.3 Setting the discharge parameters of radiofrequency (RF) source
At the time of publication of this International Standard, most RF sources are operated with constant
applied power and constant pressure. Other modes also exist such as:
— constant applied power and constant voltage;
— constant voltage and constant pressure;
— constant power and constant argon flow, etc.
All RF operational modes are allowed in this International Standard, provided they meet the
requirements described in 4.1.2 and 4.1.3 and ensure that constant excitation conditions in calibration
and analysis can be kept for optimum accuracy.
6.3.3.1 Constant applied power and pressure
The controlled parameters are applied power and pressure. Set the power supply of the glow
discharge source to constant power and constant pressure. First set the power and pressure
to typical values recommended by the manufacturer or values suitably documented. If the
recommended values are not available or not suitable for the task, following steps could be pursued
to determine the optimum parameters.
Set the high voltage of the detectors as described in 6.3.4.
Step one: set an initial discharge condition which may be the value recommended by the manufacturer
or a value in the middle of the parameter range allowed by the instrument for power and pressure.
Determine the intensities corresponding to an appropriate sample under the initial discharge condition.
Fix one parameter (power or pressure) and change the other one gradually. Measure the intensities at
least seven times at each setting. Do this at the entire parameter range allowed by the instrument.
Step two: fix the parameter which changed in step one and gradually change the other parameter.
Repeat the operation as described in step one.
Step three: calculate the CVs of the measurements for each setting. Investigate the influence of the
discharge parameters on the intensities and CVs and fix the optimum instrumental condition.
6.3.3.2 Constant power and voltage
The controlled parameters are applied power and voltage. Set the power supply of the glow
discharge source to constant power and constant voltage. First set the power and voltage to typical
values recommended by the manufacturer or values suitably documented. If the recommended
© ISO 2015 – All rights reserved 5

---------------------- Page: 9 ----------------------
ISO 19272:2015(E)

values are not available or not suitable for the task, following steps could be pursued to determine
the optimum parameters.
Set the high voltage of the detectors as described in 6.3.4.
Step one: set an initial discharge condition which may be the value recommended by the manufacturer
or a value in the middle of the parameter range allowed by the instrument for power and voltage.
Determine the intensities corresponding to an appropriate sample under the initial discharge condition.
Fix one parameter (power or voltage) and change the other one gradually. Measure the intensities at
least seven times at each setting. Do this at the entire parameter range allowed by the instrument.
Step two: fix the parameter which changed in step one and gradually change the other parameter.
Repeat the operation as described in step one.
Step three: calculate the CVs of the measurements for each setting. Investigate the influence of the
discharge parameters on the intensities and CVs and fix the optimum instrumental condition.
Use a similar procedure to optimize the analytical parameters for other operation modes.
6.3.4 Optimization of the detection system
Photomultiplier power input shall be selected as a function of the type of detector and the content
range of each element to be determined. Operate the source and observe the output signals from the
detector of the elements concerned, adjust the high voltage of the detectors in such a way that sufficient
sensitivity at the lowest analyte concentrations is ensured, without overflowing the detectors at the
highest analyte concentrations.
6.3.5 Pre-burning time and integration time
After a glow discharge is initiated, it needs some time to reach stabilization. This stabilization time is
dependent on the conditions of the instrument and sample surface. Measurement of the signals can
only be done after stabilization. This requirement is met by setting a suitable pre-burning time. A pre-
burning time reasonably set may also clean the sample surface to eliminate a possible contamination
caused by sample preparation.
Select pre-burning time according to the manufacturer’s recommendation. If the recommended value is
neither suitable nor available, select it as described below:
Excite the sample continuously, record the diagram of intensity vs sputtering time. Select a sputtering
time, at which the intensity became sufficiently stable, as pre-burning time.
Select integration time according to the manufacturer’s recommendation. If the recommended value is
neither suitable nor available, select it as described below:
Set a series of integration times, normally in the range of 5 s to 50 s. Measure the intensities
corresponding to the elements to be determined at least seven times at each integration setting. Check
the influence of the integration times on intensities and CVs. Integration time shall be long enough to
ensure CVs required in 4.1.2.
6.3.6 Validation of the instrumental parameters
The discharge parameters and detector high voltages selected shall be used for the analytical program.
Using a suitable sample, carry out some preliminary tests and ensure that the key instrumental
parameters are appropriate to meet the criteria described in 4.1.2 and 4.1.3.
6.4 Calibration
A series of certified reference materials (CRM) (at least five for each element), which have the same
or at least very similar matrix and metallurgical structure as the samples to be analysed is used to
prepare the calibration curves. The content range of the CRM used shall cover that of all the samples to
6 © ISO 2015 – All rights reserved

---------------------- Page: 10 ----------------------
ISO 19272:2015(E)

be analysed within each specific analytical program. Adjust the source parameters as described in 6.3.
Measure each CRM at least three times and every time on a new area of the sample surface. The mean
intensity is correlated to the corresponding certified content and a regression is calculated by the least
squares method. In order to minimize possible signal fluctuations, it is advisable to use the internal
standard method.
6.5 Checking of the trueness of the method
In order to check the trueness of the method, select a series of certified reference materials, necessarily
independent from that used for the calibration of the spectrometer and analyse it. The composition
of the certified reference materials selected shall cover the full range of the calibration. The results
obtained for each certified reference material shall be situated in the interval “Certified Value ± 2σ”,
where σ is standard deviation of certified value.
If necessary, in order to fulfil this trueness requirement for all the certified reference materials analysed,
carry out adjustments of the calibration curves (inter element corrections and/or modification of the
mathematical function defining each curve). If the trueness criteria cannot be reached, re-prepare the
calibration curve or re-select the parameters.
6.6 Drift correction
Before starting a series of determinations by using an already recorded program, it’s necessary to
check the drift condition of the instrument. The procedure for drift checking is the same as described
in 6.5. Drift correction is not necessary if the results are within specified accuracy; otherwise, the drift
correction shall be carried out.
Whenever the drift correction is performed, an accuracy control needs to be carried out as
described in 6.5.
6.7 Analysis of the samples
Each sample shall be analysed at least three times. Use the mean intensity to derive the content of each
element from the corresponding calibration curve.
In the case where the internal standard method is employed, use the intensity ratio between each
analyte and the internal standard element.
The content of each element to be determined shall be within the range of the calibration curves.
7 Expression of results
7.1 Method of calculation
Using the intensities (or intensity ratios) obtained, the content of each element measured, expressed as
mass fraction, is derived from the calibration curves established in 6.4.
7.2 Precision
A planned precision trial of this method was carried out by 11 laboratorie
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.