SIGNIFICANCE AND USE
5.1 In order to be certain that the end user of analytical results obtained from using an ASTM Committee D19 test method can be confident that the values have been obtained through a competent application of the test method, a demonstration of the proficiency of the analytical system shall be performed. Appropriate proficiency is demonstrated by achievement of performance criteria derived from results of the test method collaborative study. The QC measures specified in this practice shall be included in each ASTM test method, as applicable, to ensure the quality of measurements.  
5.2 In order for users of D19 test methods to achieve consistently valid results, a minimum level of QC shall be performed. This minimum level of QC is stipulated in this practice and by the task groups developing D19 test methods. If the specific requirements outlined in this practice are not applicable to the test method, alternative QC shall be defined in the test method.
SCOPE
1.1 This practice provides specific, mandatory requirements for incorporating quality control (QC) procedures into all test methods under the jurisdiction of Committee D19.  
1.2 ASTM International has adopted the following:
Policy on implementation of requirements for a quality control section in standard test methods generated by Committee D19 on Water.  
GENERAL—By July 29, 1998, or at the next reapproval or revision, whichever is later, every D19 Standard Test Method shall contain a QC section that is in full compliance with the requirements of this practice.  
NEW COLLABORATIVE TESTING—As of July 29, 1998, each collaborative study design shall include a QC section as part of the method to be tested. Prior to approval of the study design, the Results Advisor or equivalent shall ascertain the appropriateness of the QC section in meeting the requirements of this practice and Practice D2777, and shall advise the designer of the study of any changes needed to fulfill the requirements of these practices. Before a collaborative study may be conducted, approval of the study design by the Results Advisor or equivalent shall be obtained.  
OLDER VALIDATED METHODS—Standard test methods that were validated using Practices D2777 – 77, D2777 – 86, or D2777 – 94, when balloted for reapproval or revision, shall contain a QC section based upon the best information from the historical record. Where appropriate, information derived from the record of the collaborative study shall be utilized for this purpose. The introduction of the QC section into these standard test methods shall not be construed as a requirement for a new collaborative study, though the Subcommittee may opt for such a study. Any information available regarding QC or precision/bias testing shall be included in the appropriate sections of the published test method.  
1.3 Required QC sections in all applicable test methods are intended to achieve two goals. First, users of Committee D19 test methods will be able to demonstrate a minimum competency in the performance of these test methods by comparison with collaborative study data. Second, all users of test methods will be required to perform a minimum level of QC as part of proper implementation of these test methods to ensure ongoing competency.  
1.4 This practice contains the primary requirements for QC of a specific test method. In many cases, it may be desirable to implement additional QC requirements to assure the desired quality of data.  
1.5 The specific requirements in this practice may not be applicable to all test methods. These requirements may vary depending on the type of test method used as well as the analyte being determined and the sample matrix being analyzed.  
1.5.1 If there are compelling reasons why any of the specific QC requirements listed in this practice are not applicable to a specific test method, these reaso...

  • Standard
    12 pages
    English language
  • Standard
    12 pages
    English language

SIGNIFICANCE AND USE
5.1 Following this practice should result in precision-and-bias statements that can be achieved by any laboratory properly using the test method studied. These precision-and-bias statements provide the basis for generic limits for use in the Quality Control section of the test method. Optionally, the detection and quantitation values provide estimates of the level at which most laboratories should be able to achieve confident detection and meet the minimum precision (expressed as relative standard deviation) expected.  
5.2 The method specifies the matrices for which the test method is appropriate. The collaborative test corroborates the write-up within the limitations of the test design. An extensive test can only use representative matrices so that universal applicability cannot be implied from the results.  
5.3 The fundamental assumption of the collaborative study is that the matrices tested, the concentrations tested, and the participating laboratories are a representative and fair evaluation of the scope and applicability of the test method as written.
SCOPE
1.1 This practice establishes uniform standards for estimating and expressing the precision and bias of applicable test methods for Committee D19 on Water. Statements of precision and bias in test methods are required by the Form and Style for ASTM Standards, “Section A21. Precision and Bias (Mandatory).” In principle, all test methods are covered by this practice. However, the variability equations provided in this standard are applicable only to test methods that yield continuous function values.  
1.2 Except as specified in 1.4, 1.5, and 1.6, this practice requires the task group proposing a new test method to carry out a collaborative study from which statements for precision (overall and single-operator standard-deviation estimates) and bias can be developed. This practice provides general guidance to task groups in planning and conducting such determinations of precision and bias.  
1.3 This practice requires that a task group making a substantive revision to a test method also perform a limited-scale collaborative study (known as a ― comparability study) to evaluate the effect of the revision on the precision and bias statement. This practice provides guidance to task groups for conducting such limited-scale collaborative studies. Examples of substantive modifications may include, but are not limited to, changes in mandatory or allowable instrumentation, reagents, reaction times, etc.  
1.3.1 Changes to applicable water matrices in the Scope of a method may constitute a substantive modification under this provision. Only matrices that have been evaluated in an approved collaborative study may be listed in the Scope of a method. It is recognized that the term “matrix” is generally vague. Terms specifying matrix types can cover significantly different chemical constituents, unless the matrix is synthesized to be of a standardized makeup. Substitute Wastewater (Practice D5905) is one such defined matrix. For purposes of this practice, the importance of this requirement is to assist the user of a D19 standard in determining the applicability of the method to their samples. Evaluated matrices should be described with as much detail as possible to minimize misapplication.  
1.3.2 A method's concentration-range extension that is deemed to merit additional collaborative testing (even without a method modification that would otherwise be considered substantive) shall require a full collaborative study, as described in 7.1 through 7.5, but only at concentrations representative of the extended range. Note that such a collaborative study could involve as little as a single concentration study in a single reproducible matrix.  
1.3.3 Whether a revision to a test method includes substantive modification shall be determined by consensus of the Committee.  
1.4 If a full-scale collaborative study is not technically feasible, because...

  • Standard
    21 pages
    English language
  • Standard
    21 pages
    English language

SIGNIFICANCE AND USE
5.1 Appropriate application of this practice should result in a WQE achievable by the laboratory in applying the tested method/matrix/analyte combination to routine sample analysis. That is, a laboratory should be capable of measuring concentrations greater than WQEZ %, with the associated RSD equal to Z % or less.  
5.2 The WQE values may be used to compare the quantitation capability of different methods for analysis of the same analyte in the same matrix within the same laboratory.  
5.3 The WQE procedure should be used to establish the within-laboratory quantitation capability for any application of a method in the laboratory where quantitation is important to data use. The intent of the WQE is not to impose reporting limits. The intent is to provide a reliable procedure for establishing the quantitative characteristics of the method (as implemented in the laboratory for the matrix and analyte) and thus to provide the laboratory with reliable information characterizing the uncertainty in any data produced. Then the laboratory can make informed decisions about censoring data and has the information necessary for providing reliable estimates of uncertainty with reported data.
SCOPE
1.1 This practice establishes a uniform standard for computing the within-laboratory quantitation estimate associated with Z % relative standard deviation (referred to herein as WQEZ %), and provides guidance concerning the appropriate use and application.  
1.2 WQEZ % is computed to be the lowest concentration for which a single measurement from the laboratory will have an estimated Z % relative standard deviation (Z % RSD, based on within-laboratory standard deviation), where Z is typically an integer multiple of 10, such as 10, 20, or 30. Z can be less than 10 but not more than 30. The WQE10 % is consistent with the quantitation approaches of Currie (1)2 and Oppenheimer, et al. (2).  
1.3 The fundamental assumption of the WQE is that the media tested, the concentrations tested, and the protocol followed in developing the study data provide a representative and fair evaluation of the scope and applicability of the test method, as written. Properly applied, the WQE procedure ensures that the WQE value has the following properties:  
1.3.1 Routinely Achievable WQE Value—The laboratory should be able to attain the WQE in routine analyses, using the laboratory’s standard measurement system(s), at reasonable cost. This property is needed for a quantitation limit to be feasible in practical situations. Representative data must be used in the calculation of the WQE.  
1.3.2 Accounting for Routine Sources of Error—The WQE should realistically include sources of bias and variation that are common to the measurement process and the measured materials. These sources include, but are not limited to intrinsic instrument noise, some typical amount of carryover error, bottling, preservation, sample handling and storage, analysts, sample preparation, instruments, and matrix.  
1.3.3 Avoidable Sources of Error Excluded—The WQE should realistically exclude avoidable sources of bias and variation (that is, those sources that can reasonably be avoided in routine sample measurements). Avoidable sources include, but are not limited to, modifications to the sample, modifications to the measurement procedure, modifications to the measurement equipment of the validated method, and gross and easily discernible transcription errors (provided there is a way to detect and either correct or eliminate these errors in routine processing of samples).  
1.4 The WQE applies to measurement methods for which instrument calibration error is minor relative to other sources, because this practice does not model or account for instrument calibration error, as is true of most quantitation estimates in general. Therefore, the WQE procedure is appropriate when the dominant source of variation is not instrument calibration, but is perhaps one or ...

  • Standard
    19 pages
    English language
  • Standard
    19 pages
    English language

SIGNIFICANCE AND USE
4.1 The user of this guide is not assumed to be a technical practitioner in the water field. This guide is an assembly of the components common to all aspects of water monitoring and fulfills a need in the development of a common framework for a better coordinated and more unified approach to monitoring water.  
4.2 Limitations—This guide does not establish a standard procedure to follow in all cases and it does not cover the details necessary to meet a particular monitoring objective.
SCOPE
1.1 Purpose—This guide is generic in its application to surface or ground water, rivers, lakes, or estuaries (quantity and quality). It proposes a series of options that offer direction without recommending a definite course of action and discusses the major elements that are common to all purposes of water monitoring.  
1.2 The elements described are applicable whether the monitoring is only for one location or integrates multiple measurement sites for the purpose of assessing a whole watershed, estuary, or aquifer system.  
1.3 This guide is intended to outline for planners and administrators the components, process, and procedures which should be considered when proposing, planning, or implementing a monitoring program. The guide is not a substitute for obtaining specific technical advice. The reader is not assumed to be a technical practitioner in the water field; however, practitioners will find it a good summary of practice and a handy checklist. Other standard guides have or will be prepared that address the necessary detail.  
1.4 Monitoring Components—A water monitoring program is composed of a set of activities, practices, and procedures designed to collect reliable information of known accuracy and precision concerning a particular water resource in order to achieve a specific goal or purpose. The purposes may range in scope from tracking status and trends on a regional or national basis to gathering data to determine the effects of a specific management practice or pollution incident such as a spill. This guide suggests and discusses the following process and components:  
1.4.1 Establishment of program goals and objectives and recording of decisions in a written plan (see 6.1),  
1.4.2 Developing background data and a conceptual model (see 6.1.12),  
1.4.3 Establishment of data (quality, quantity, type) objectives (see 6.2),  
1.4.4 Design of field measurement and sampling strategies and specification of laboratory analyses and data acceptance criteria (see 6.3),  
1.4.5 Data storage and transfer (see 6.6),  
1.4.6 Implementation of sampling and analysis strategies (see 6.4),  
1.4.7 Data quality assessment (see 6.5),  
1.4.8 Assessment of data (see 6.7),  
1.4.9 Program evaluation (see 6.8), and  
1.4.10 Reporting (see 6.9). See also Fig. X1.1 in Appendix X1 and the condensed list of headings in Appendix X2.  
1.5 Monitoring Purposes—Establishing goals defines the purpose for monitoring. Each purpose has some monitoring design needs specific to itself. There are six major purposes for water monitoring. They are as follows:  
1.5.1 Determining the Status and Trends of Water Conditions—This can require long term, regular monitoring to determine how parameters change over time.  
1.5.2 Detecting Existing and Emerging Problems—Determining if, how, or where a substance may move through an aquatic system, or if water quantities are changing.  
1.5.3 Developing and Implementing Management and Regulatory Programs—Includes baseline and reconnaissance monitoring to characterize existing conditions such as to identify critical areas or hot spots; implementation monitoring to assess whether activities were carried out as planned; and compliance monitoring to determine if specific water quality or water use criteria were met.  
1.5.4 Responding to an Emergency—Performed to provide information in the near term.  
1.5.5 Evaluating the Effectiveness of Water Monitoring Programs—Is the ...

  • Guide
    11 pages
    English language

SIGNIFICANCE AND USE
5.1 Matrix spiking is commonly used to determine the bias under specific analytical conditions, or the applicability of a test method to a particular sample matrix in that context, by determining the extent to which the spiked analyte or component is recovered from the sample matrix under these conditions. Reactions or interactions of the analyte or component of interest with the sample matrix may cause a significant positive or negative effect on recovery and may render the chosen analytical, or monitoring, process ineffectual for that sample matrix.  
5.2 Matrix spiking can also be used to monitor the performance of a laboratory, individual instrument, or analyst as part of a regular quality assurance program. Changes in spike recoveries or recovery limits from the same or similar matrices over time may indicate variations in the quality of analytical results.  
5.3 Spiking can be used to compare the recoveries of like spikes from reagent water samples and natural matrix samples (measured with and without spike) to distinguish between (1) unusual interference and (2) inherent method recovery and instability effects. This guide does not attempt to deal with the statistical significance of differences in spike recoveries from different matrices.  
5.4 Special precautions shall be observed when nonlaboratory personnel perform spiking in the field. It is recommended that all spike preparation work be performed in a laboratory by experienced analysts so that the field operation consists solely of adding a prepared spiking solution to the sample matrix. Training of field personnel and validation of their spiking techniques are necessary to ensure that spikes are added accurately and reproducibly. Duplicate field spikes can be used to document the reproducibility of the technique. When environmentally labile compounds are used as spikes, the spiking solution shall be protected up to the point of use by appropriate means such as chilling, protection from sunlight and o...
SCOPE
1.1 This guide covers the general technique of “spiking” a broad range of materials into aqueous media. This guide will serve the analyst in preparing spiked samples for quality control purposes. Guidance is also provided to aid the analyst in calculating recoveries and interpreting results. It is the responsibility of the analyst to determine whether the procedures and materials described here are appropriate to the task at hand.  
1.2 The procedures in this guide are focused on “matrix spike” preparation, analysis, and interpretation of results. The applicability of these procedures to the preparation of calibration standards, calibration check standards, laboratory control standards, reference materials, and other quality control materials by spiking is incidental. A sample (the matrix) is fortified (spiked) with the analyte of interest for a variety of analytical and quality control purposes. While the spiking of multiple sample portions is discussed, the method of standard additions is not covered.  
1.3 This guide is intended for use in conjunction with the individual analytical test method that provides procedures for analysis of the analyte or component of interest. The test method is used to determine an analyte or component’s background level and, again after spiking, its now elevated level. Each test method typically provides procedures not only for samples, but also for calibration standards or analytical control solutions, or both. These procedures include preparation, handling, storage, preservation, and analysis techniques. These procedures are applicable by extension, using the analyst’s judgement on a case-by-case basis, to spiking solutions, and are not reiterated in this guide. See also Practice E200 for preparation and storage information.  
1.4 These procedures apply only to analytes that are soluble in water at the concentration of the spike plus any background material, or to analytes soluble in...

  • Guide
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 This substitute ocean water may be used for laboratory testing where a reproducible solution simulating sea water is required. Examples are for tests on oil contamination, detergency evaluation, and corrosion testing.  
Note 2: The lack of organic matter, suspended matter, and marine life in this solution does not permit unqualified acceptance of test results as representing performance in actual ocean water. Where corrosion is involved, the results obtained from laboratory tests may not approximate those secured under natural testing conditions that differ greatly from those of the laboratory, and especially where effects of velocity, salt atmospheres, or organic constituents are involved. Also the rapid depletion of reacting elements present in low concentrations suggests caution in direct application of results.
SCOPE
1.1 This practice covers the preparation of solutions containing inorganic salts in proportions and concentrations representative of ocean water.2  
Note 1: Since the concentrations of ocean water varies with sampling location, the gross concentration employed herein is an average of many reliable individual analyses. Trace elements, occurring naturally in concentrations below 0.005 mg/L, are not included.  
1.2 This practice provides three stock solutions, each relatively concentrated but stable in storage. For preparation of substitute ocean water, aliquots of the first two stock solutions with added salt are combined in larger volume. An added refinement in adjustment of heavy metal concentration is provided by the addition of a small aliquot of the third stock solution to the previous solution.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    3 pages
    English language

SIGNIFICANCE AND USE
5.1 Appropriate application of this practice should result in an estimate of the test-method’s uncertainty (at any concentration within the working range), which can be compared with data-quality objectives to see if the uncertainty is acceptable.  
5.2 With data sets that compare recovered concentration with true concentration, the resulting regression plot allows the correction of the recovery data to true values. Reporting of such corrections is at the discretion of the user.  
5.3 This practice should be used to estimate the measurement uncertainty for any application of a test method where measurement uncertainty is important to data use.
SCOPE
1.1 This practice establishes a standard for computing the measurement uncertainty for applicable test methods in Committee D19 on Water. The practice does not provide a single-point estimate for the entire working range, but rather relates the uncertainty to concentration. The statistical technique of regression is employed during data analysis.  
1.2 Applicable test methods are those whose results come from regression-based methods and whose data are intra-laboratory (not inter-laboratory data, such as result from round-robin studies). For each analysis conducted using such a method, it is assumed that a fixed, reproducible amount of sample is introduced.  
1.3 Calculation of the measurement uncertainty involves the analysis of data collected to help characterize the analytical method over an appropriate concentration range. Example sources of data include: (1) calibration studies (which may or may not be conducted in pure solvent), (2) recovery studies (which typically are conducted in matrix and include all sample-preparation steps), and (3) collections of data obtained as part of the method’s ongoing Quality Control program. Use of multiple instruments, multiple operators, or both, and field-sampling protocols may or may not be reflected in the data.  
1.4 In any designed study whose data are to be used to calculate method uncertainty, the user should think carefully about what the study is trying to accomplish and much variation should be incorporated into the study. General guidance on designing studies (for example, calibration, recovery) is given in Appendix X1. Detailed guidelines on sources of variation are outside the scope of this practice, but general points to consider are included in Appendix X2, which is not intended to be exhaustive. With any study, the user must think carefully about the factors involved with conducting the analysis, and must realize that the computed measurement uncertainty will reflect the quality of the input data.  
1.5 Associated with the measurement uncertainty is a user-chosen level of statistical confidence.  
1.6 At any concentration in the working range, the measurement uncertainty is plus-or-minus the half-width of the prediction interval associated with the regression line.  
1.7 It is assumed that the user has access to a statistical software package for performing regression. A statistician should be consulted if assistance is needed in selecting such a program.  
1.8 A statistician also should be consulted if data transformations are being considered.  
1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.10 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    7 pages
    English language

SIGNIFICANCE AND USE
4.1 The ASTM guidance manual, Form and Style for ASTM Standards,4 Section A21, requires a precision and bias statement in all ASTM test methods. Section A21.2.2 states:    
Precision shall be estimated in accordance with the interlaboratory test program prescribed in Practice E691, Conducting an Interlaboratory Study to Determine the Precision of a Test Method, or by an interlaboratory test program that yields equivalent information, for example, a standard practice developed by an ASTM technical committee.  
4.2 Practice D2777, Section 1.1, states:    
This practice establishes uniform standards for estimating and expressing the precision and bias of applicable test methods for Committee D19 on Water. Statements of precision and bias in test methods are required by the Form and Style for ASTM Standards, “Section A21. Precision and Bias (Mandatory).” In principle, all (ASTM Committee D19) test methods are covered by this practice.  
4.3 Practice D2777, Section 1.2, requires a task group proposing a new test method to carry out a collaborative study from which concentration limits, repeatability and reproducibility precision and bias statements are developed.  
4.3.1 This guide describes options for developing and optimizing chemical test methods for Committee D19, not implementation of a test method by a laboratory. Refer to Guide E2857 for procedures used in validating existing test methods for your laboratory.  
4.3.2 The collaborative study described in Practice D2777 is not the test method validation. The collaborative study verifies the new test method is reproducible among different laboratories, different instruments/apparatus, and different analysts.  
4.3.3 Practice D2777, Section 6.1, assumes the test method has already been optimized prior to conducting the collaborative study.  
4.4 Practice D2777, Section 4 (Summary of Practice), requires, a collaborative study only after the task group has assured itself that preliminary evaluation ...
SCOPE
1.1 This guide identifies procedures for use in developing and optimizing new or modified Subcommitees D19.05 and D19.06 test methods intended for regulatory compliance reporting in EPA drinking water and wastewater programs. This guide may also be useful for developing test methods for emerging contaminants that may not yet have regulatory requirements.  
1.2 This guide also cites statistical procedures that are useful in the single laboratory characterization and optimization and in the inter-laboratory studies (ILSs).  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This guide offers an organized collection of information or a series of options and does not recommend a specific course of action. This document cannot replace education or experience and should be used in conjunction with professional judgment. Not all aspects of this guide may be applicable in all circumstances. This ASTM standard is not intended to represent or replace the standard of care by which the adequacy of a given professional service must be judged, nor should this document be applied without consideration of a project’s many unique aspects. The word “Standard” in the title of this document means only that the document has been approved through the ASTM consensus process.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical...

  • Guide
    12 pages
    English language

SIGNIFICANCE AND USE
4.1 Many analytical laboratories comply with accepted quality system requirements such as NELAC, Chapter 5,5 (see Note 2) and ISO/IEC 17025. When using standard test methods, their test results on the same sample should agree with those from other similar laboratories within the reproducibility estimates (R2) published in the standard. Reproducibility estimates are generated during the standardization process as part of the interlaboratory studies (ILS). Many laboratories participate in proficiency tests to confirm that they perform consistently over time. In both ILS and proficiency testing protocols, it is generally assumed that only one workstation is used to generate the data (see 6.5.1).
Note 2: NELAC, Chapter 5, allows the use of a Work Cell where multiple instruments/operators are treated as one unit: the performance of the Work Cell is tracked rather than each workstation independently. This guide is intended to go beyond the Work Cell to achieve the benefits of monitoring workstations independently.  
4.2 Many laboratories have workloads or logistical requirements, or both, that dictate the use of multiple workstations. Some have multiple stations in the same area (central laboratory format). Others’ stations are scattered throughout a facility (at-line laboratory format). Often, analysis reports do not identify the workstation used for the testing, even if workstations differ in their testing uncertainties. Problems can arise if clients mistakenly attribute variation in report values to process rather then workstation variability. These problems can be minimized if the laboratory organization sets, complies with, and reports a unified set of measurement quality objectives throughout.  
4.3 This guide can be used to harmonize calibration and control protocols for all workstations, thereby providing the same level of measurement traceability and control. It streamlines documentation and training requirements, thereby facilitating flexibility in personnel...
SCOPE
1.1 This guide describes a protocol for optimizing, controlling, and reporting test method uncertainties from multiple workstations in the same laboratory organization. It does not apply when different test methods, dissimilar instruments, or different parts of the same laboratory organization function independently to validate or verify the accuracy of a specific analytical measurement.  
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    7 pages
    English language

SIGNIFICANCE AND USE
4.1 Substitute wastewater may be used for laboratory testing where a reproducible mixture simulating municipal wastewater is required. To provide a more rugged evaluation of a test method's ability to measure analytes and the precision of the test method under more demanding conditions, it is necessary to utilize a matrix more complex than reagent water. Substitute wastewater is intended to be used as a material to be spiked with analytes or interferents of interest to evaluate the performance of test methods.  
4.2 Utilization of substitute wastewater as a matrix in interlaboratory validation studies will allow inclusion of performance statistics for this matrix in the Precision and Bias section of test methods. Users of test methods will be able to evaluate their application of test methods in this matrix against published results.
SCOPE
1.1 This practice covers the preparation of an aqueous mixture containing constituents in concentrations such that it will have physical and chemical matrix characteristics similar to municipal wastewater.  
1.2 Wastewaters are extremely variable, depending on the quantity and nature of the materials being discharged into the collection system. The mixture prepared with this practice is not representative of any particular wastewater. Rather, it allows the user to prepare a mixture that exhibits a similar matrix impact on test method performance as is typical of municipal wastewater and can be prepared from common materials inexpensively and reproducibly. It allows the evaluation of test methods, over time, against the same reference point.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    3 pages
    English language

SIGNIFICANCE AND USE
4.1 The purity of water is relative and is usually characterized by the limits of impurities found in the water as well as by the methods used to prepare and handle the water. Section 7 mentions the suitable methods for water preparation.
SCOPE
1.1 This guide is intended to describe the chemical and biological characteristics of water to be used whenever critical purity is essential to the use intended in laboratory bio-applications, for example, clinical, pharmaceutical, and biomedical. The importance of such a reagent is often underestimated despite the impact that it can have.  
1.2 This guide is not intended to be used as a reference in preparing water for injectables. Generally, the appropriate use of this guide may include experiments involving tissue culture, chromatography, mass spectrometry, polymerase chain reaction (PCR), deoxyribonucleic acid (DNA) sequencing, DNA hybridization, electrophoresis, molecular biology or analyses where molecular concentrations of impurities may be important.  
1.3 For all the other applications linked to an ASTM method and not bio-sensitive that require purified water, it is recommended that Specification D1193 or Guide D5127 be consulted.  
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    4 pages
    English language

SIGNIFICANCE AND USE
5.1 Environmental data are often required for making regulatory and programmatic decisions. These data must be of known quality commensurate with their intended use.  
5.2 Certain minimal criteria must be met by the field organizations in order to meet the objectives of the water monitoring activities.  
5.3 This guide defines the criteria for organizations taking water samples and generating environmental data and identifies other activities that may be required based on the DQOs.  
5.4 This guide emphasizes the importance of communication among those involved in establishing the DQOs, planning, and implementing the sampling and analysis aspects of environmental data generation activities, and assessing data quality.
SCOPE
1.1 This guide covers planning and implementation of the sampling aspects of environmental data generation activities. Environmental data generation efforts are comprised of four parts: (1) establishment of data quality objectives (DQOs); (2) design of field sampling and measurement strategies and specification of laboratory analyses and data acceptance criteria; (3) implementation of sampling and analysis strategies; and (4) data quality assessment.  
1.2 This guide defines the criteria that must be considered to ensure the quality of the field aspects of environmental data and sample generation activities.  
1.3 DQOs should be adopted prior to the application of this guide. The data generated in accordance with this guide are subject to a final assessment to determine whether the DQOs were met. For example, many screening activities do not require all of the quality assurance (QA) and quality control (QC) steps found in this guide to generate data adequate to meet the project needs. The extent to which all of the requirements must be met remains a matter of technical judgement as it relates to the established DQOs.  
1.4 This guide presents extensive management requirements designed to ensure high-quality samples and data. The words “must,” “shall,” “may,” and “should” have been selected carefully to reflect the importance placed on many of the statements made in this guide.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    8 pages
    English language

SIGNIFICANCE AND USE
4.1 This practice is designed to assist suppliers and users of reference materials by identifying the information necessary on the certificate of analysis of materials designated for use in ASTM test methods. This practice is specifically designed to ensure that materials suitable for use as either calibration or quality control standards are available. This practice does not define a specific certification protocol, but rather provides guidance in the development of adequate data to support the use of the material as either a calibration or quality control standard. Suppliers are referred to ISO Guide 35 for guidelines on acceptable certification protocols. End users are referred to ISO Guide 31 for a more complete description of the elements of typical certificates of analysis.
SCOPE
1.1 This practice covers the information that must be provided on certificates of analysis of reference materials designated to support ASTM methods. It provides end users of these materials with a defined set of data that is required to be on a certificate of analysis and provides information to assist the end user in evaluating the independence of the material. Similarly, it provides the suppliers of reference materials with a consistent format for the presentation of certification data.  
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 This guide recommends the water quality required for the electronics and microelectronics industries. High-purity water is required to prevent contamination of products during manufacture, since contamination can lead to an unacceptable, low yield of electronic devices.  
4.2 The range of water purity is defined in accordance with the manufacturing process. The types of ultra-pure water are defined with respect to device line width. In all cases, the water-quality recommendations apply at the point of distribution.  
4.3 The limits on the impurities are related to current contamination specifications and to available analytical methods (either performed in a suitable clean laboratory or by on-line instrumentation). On-line and off-line methods are used in accordance with current industry practice. Concentration of the sample may be required to measure the impurities at the levels indicated in Table 1. (A) The user should be advised that analytical data often are instrument dependent and technique dependent. Thus, the numbers in Table 1 are only guidelines. This table will be revised whenever the semiconductor industry develops new linewidths, thereby keeping the guidelines current.(B) Values shown in Type E-1.3 are a result of aligning ITRS risk factors of known contaminates to the production processes found in current semiconductor processing for the linewidth of interest and may differ in a few cases to those found in Type E-1.2. Users who wish to use the higher numbers for Type E-1.2 water should feel free to do so.
All values are equal to or less than with the exception of Resistivity.(C) Boron is monitored only as an operational parameter for monitoring the ion-exchange beds.
SCOPE
1.1 This guide provides recommendations for water quality related to electronics and semiconductor-industry manufacturing. Seven classifications of water are described, including water for line widths as low as 0.032 μm. In all cases, the recommendations are for water at the point of distribution (POD).  
1.2 Water is used for washing and rinsing of semiconductor components during manufacture. Water is also used for cleaning and etching operations, making steam for oxidation of silicon surfaces, preparing photomasks, and depositing luminescent materials. Other applications are in the development and fabrication of solid-state devices, thin-film devices, communication lasers, light-emitting diodes, photo-detectors, printed circuits, memory devices, vacuum-tube devices, or electrolytic devices.  
1.3 Users needing water qualities different from those described here should consult other water standards, such as Specification D1193 and Guide D5196.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    6 pages
    English language

SIGNIFICANCE AND USE
5.1 In order to obtain meaningful analytical data, sample preservation techniques must be effective from the time of sample collection to the time of analysis. A laboratory must confirm that sample integrity is maintained throughout maximum time periods between sample collection and analysis. In many cases, it is useful to know the maximum holding time. An evaluation of holding time is useful also in judging the efficacy of various preservation techniques.
SCOPE
1.1 This practice covers the means of estimating the period of time during which a water sample can be stored after collection and preservation without significantly affecting the accuracy of analysis.  
1.2 The maximum holding time is dependent upon the matrix used and the specific analyte of interest. Therefore, water samples from a specific source must be tested to determine the period of time that sample integrity is maintained by standard preservation practices.  
1.3 In the event that it is not possible to analyze the sample immediately at the time of collection, this practice does not provide information regarding degradation of the constituent of interest or changes in the matrix that may occur from the time of sample collection to the time of the initial analysis.  
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
5.1 Chain-of-custody procedures are a necessary element in a program to assure one’s ability to support data and conclusions adequately from the time samples are collected until sample disposal. In a legal or regulatory situation custody documentation alone is not sufficient. A complete data defensibility scheme should be followed that fits the given situation.  
5.2 In applying the sample chain-of-custody procedures in this guide, it is assumed that all of the other elements of data defensibility have been applied, if applicable.  
5.3 Because there is no definitive program that guarantees legal defensibility of data integrity in any given situation, this guide provides a description and discussion of a comprehensive list of possible elements of a chain-of-custody program, all of which have been employed in actual programs but are given as options for the development of a specific chain-of-custody program. In addition, within particular chain-of-custody elements, this guide proscribes certain activities to assure that if these options are chosen, they will be implemented properly.
SCOPE
1.1 This guide contains a comprehensive discussion of potential requirements, in the analysis of water, for a sample chain-of-custody program and describes the procedures involved in sample chain-of-custody. The purpose of these procedures is to provide accountability for and documentation of sample integrity from the time samples are collected until sample disposal.  
1.2 These procedures are intended to document sample possession during each stage of a sample’s life cycle, that is, during collection, shipment, storage, and the process of analysis.  
1.3 Sample chain-of-custody is just one aspect of the larger issue of data defensibility (see 3.2.2 and Appendix X1).  
1.4 A sufficient chain-of-custody process, that is, one that provides sufficient evidence of sample integrity in a legal or regulatory setting, is situationally dependent. The procedures presented in this guide are generally considered sufficient to assure legal defensibility of sample integrity. In a given situation, less stringent measures may be adequate. It is the responsibility of the users of this guide to determine their exact needs. Legal counsel may be needed to make this determination.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    9 pages
    English language

SIGNIFICANCE AND USE
4.1 The proper use of analytical data requires adequate documentation of all inputs, that is, the source and history of the sample, laboratory performing the analysis, method of analysis, date of analysis, precision and bias of the measurements, and related quality assurance information.  
4.2 In order to have defensible data, the report must be complete and accurate, providing adequate information to evaluate the quality of the data and contain supporting information that documents sampling and analysis procedures.  
4.3 This guide contains some of the common data qualifiers or “flags” commonly used by laboratories following the good laboratory practices, the government contract program, or found in the commercial laboratories. Examples of these qualifiers are the use of (E) for estimated value, (U) for analyzed for but not detected, and (B) for analyte was found in the blank (see 8.11). The qualifiers included in this guide should help the laboratory and its customers to better understand each other by using standardized qualifiers.  
4.4 Practice D933 is a comprehensive practice for reporting water-formed constituents such as metal oxides, acid anhydrides, and others.
SCOPE
1.1 This guide provides guidelines for reporting inorganic and organic results of analyses of drinking water, waste water, process water, ground water, and surface water, and so forth, to laboratory clients in a complete and systematic fashion.  
1.2 The reporting of bacterial and radiological data are not addressed in this guide.  
1.3 The commonly used data qualifiers for reviewing and reporting information are listed and defined. Client and laboratory specific requirements may make use of other qualifiers. This guide does not preclude the use of other data qualifiers.  
1.4 This guide discusses procedures for and specific problems in the reporting of low level data, potential errors (Type I and Type II), and reporting data that are below the calculated method detection limit and above the analyte.  
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 This guide establishes basic requirements which should be met by water and environmental laboratories that generate and report test chemical analyses which the laboratory client desires to be traceable to SI units (Note 1) or certified reference materials traceable to SI units. Traceability of chemical analyses is important because it provides a uniform basis for the comparison of results from different measurement systems and because it relates those results to our current knowledge of physical laws (Note 2).
Note 1: A certified reference material traceable to SI units is a certified reference material whose value can be related with a stated uncertainty through an unbroken change of comparisons to stated references (usually national or international standards) in SI units, such as a primary measurement made in SI units or a national standard certified in SI units.
Note 2: Not all chemical analysis results can be traceable to SI units or to certified reference material’s traceable to SI units, such as turbidity and or total suspended solids.  
4.2 Many waters-related laboratories comply with ISO Guide 17025 and participate in Proficiency Testing Programs. Laboratories that are connected to the same accreditation bodies and Proficiency Test providers can be expected to report statistically similar results on the same sample. However, some test methods and some certified reference materials are not supported with data traceable to SI units. Therefore, fully compliant laboratories that are not connected to the same providers may report statistically different chemical analysis results if they used the same nontraceable test method on the same sample. This problem could be minimized if they used test methods, measurement devices, and certified reference materials that are traceable to SI units, where available.  
4.3 Although some standard test methods and certified reference materials provide evidence of traceability to SI units, many others do not. Therefor...
SCOPE
1.1 This guide sets a protocol for generating and reporting chemical analyses that are traceable to SI units or to certified reference materials in laboratories that serve the water and environmental industry.  
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.3 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    3 pages
    English language

SIGNIFICANCE AND USE
4.1 This practice is for the use by RM producers in the development and implementation of their quality system and by those concerned with assessing the competence of RM producers. It should be recognized that a RM needs to be characterized mainly to the level of accuracy required for its intended purpose (that is, appropriate measurement uncertainty). The RM producer shall describe the procedure for establishing the quality of materials as a component of the quality system.  
4.2 This practice is for the use of RM users in the establishment if a RM producer has a quality system adequate to produce high quality RMs. It can be used by users to determine if the scientific and technical competence of a RMs producer is adequate to ensure the quality of RMs. This practice is consistent with the requirements for RM producers established in ISO Guide 34.  
4.3 This practice does not specify specific protocols for the contents of RMs certificates of analysis, for calibration in analytical chemistry and use of certified RMs and for certification of RMs. For this information, users are referred to Practice D6362, ISO Guide 32, and ISO Guide 35.
SCOPE
1.1 This practice establishes the general requirements with which a reference materials (RM) producer has to demonstrate that it operates, if it is to be recognized as competent to produce RMs used for water analysis.  
1.2 This practice establishes the quality system requirements in accordance with which waters RMs shall be produced. It is intended to be used as part of a RM producer’s general quality assurance (QA) procedures. RM producers shall define their scope in terms of the application, the measurement methods used in the homogeneity, stability, and characterization studies.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    12 pages
    English language

SIGNIFICANCE AND USE
4.1 This guide is intended to be used in the planning stage or phase of developing a sediment monitoring program. This guide is an assembly of the components common to all aspects of watershed sediment monitoring and fulfills a need in the development of a common framework for a better coordinated and a more unified approach to sediment monitoring in watersheds.  
4.2 The user of this guide is not assumed to be a trained technical practitioner in the water quality, sedimentation, or hydrology fields. The intended users are managers and planners who need information to develop a water quality monitoring program or project with an emphasis in sediment and hydrology. Sediment specialists will also find information on procedures, equipment, methodology, and operations to conduct a monitoring program.  
4.3 This guide is used during the planning process of developing, designing, and re-evaluating a sediment monitoring program.
SCOPE
1.1 Purpose—This guide is intended to provide general guidance on a watershed monitoring program directed toward sediment. The guide offers a series of general steps without setting forth a specific course of action. It gives advice for establishing a monitoring program, not an implementation program.  
1.2 Sedimentation as referred to in this guide is the detachment, entrainment, transportation, and deposition of eroded soil and rock particles. Specific types or parameters of sediment may include: suspended sediment, bedload, bed material, turbidity, wash load, sediment concentration, total load, sediment deposits, particle size distribution, sediment volumes and particle chemistry. Monitoring may include not only sediments suspended in water but sediments deposited in fields, floodplains, and channel bottoms.  
1.3 This guide applies to surface waters as found in streams and rivers; lakes, ponds, reservoirs, estuaries, and wetlands.  
1.4 Limitations—This guide does not establish a standard procedure to follow in all situations and it does not cover the detail necessary to define all of the needs of a particular monitoring objective or project. Other standards and guides included in the reference and standard sections describe in detail the procedures, equipment, operations, and site selection for collecting, measuring, analyzing, and monitoring sediment and related constituants.  
1.5 Additional ASTM and U.S. Geological Survey standards applicable to sediment monitoring are listed in Appendix X1 and Appendix X2. Due to the large number of optional standards and procedures involved in sediment monitoring, most individual standards are not referenced in this document. Standards and procedures have been grouped in the appendices according to the type of analyses or sampling that would be required for a specific type of measurement or monitoring.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    10 pages
    English language

SCOPE
1.1 This practice describes a procedure for developing a graphical model of relative standard deviation versus concentration for analytical methods used in the analysis of water (methods that are subject to non-additive random errors) for the purpose of assigning a statement of noise or randomness to analytical results (commonly referred to as a precision statement), in either a manual or an automated fashion.  
1.2 Data analysis and modeling is done with Committee D19 Adjunct DQCALC2 (a Microsoft Excel3-based tool).  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
4.1 The user of this guide is not assumed to be a trained technical practitioner in the water quality field. The guide is an assembly of the components common to all aspect of watershed nutrient monitoring and fulfills a need in the development of a common framework for a better coordinated and a more unified approach to nutrient monitoring in watersheds.  
4.2 Limitations—This guide does not establish a standard procedure to follow in all situations and it does not cover the detail necessary to meet all of the needs of a particular monitoring objective. Other standards and guides included in the references describe the detail of the procedures.
SCOPE
1.1 Purpose—This guide is intended to provide general guidance on a watershed monitoring program directed toward the plant nutrients nitrogen and phosphorus. The guide offers a series of general steps without setting forth a specific course of action. It gives assistance for developing a monitoring program but not a program for implementing measures to improve water quality.  
1.2 This guide applies to waters found in streams and rivers; lakes, ponds, and reservoirs; estuaries; wetlands; the atmosphere; and the vadose and subsurface saturated zones (including aquifers). This guide does not apply to nutrients found in soils, plants, or animals.  
1.3 Nutrients as used in this guide are intended to include nitrogen and phosphorus in dissolved, gaseous, and particulate forms. Specific species of nitrogen include: nitrate, nitrite, ammonia, organic, total Kjeldahl, and nitrous oxide. The species of phosphorus include total, total dissolved, organic, acid-hydrolyzable, and reactive phosphorus as described in Ref (2).  
1.4 Safety—Health and safety practices developed for a project may need to consider the following:  
1.4.1 During the construction of sampling stations:
1.4.1.1 Drilling practices during monitoring well installations,
1.4.1.2 Overhead and underground utilities during monitoring well drilling,
1.4.1.3 Traffic patterns/concerns during sampling station installation,
1.4.1.4 Traffic patterns/concerns during surveying sampling station locations and elevations,
1.4.1.5 Drilling through materials highly contaminated with fertilizers, and
1.4.1.6 Installing monitoring equipment below the soil surface.  
1.4.2 During the collection of water samples:
1.4.2.1 Using acids for sample preservation,
1.4.2.2 Sampling during flooding events and ice conditions,
1.4.2.3 Traffic on bridges,
1.4.2.4 Condition of sampling stations following flood events,
1.4.2.5 Sampling of water or soils, or both, highly contaminated with fertilizers,
1.4.2.6 Conditions of sampling stations resulting from vandalism,
1.4.2.7 Adverse weather conditions, and
1.4.2.8 Transporting liquid samples.  
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Guide
    7 pages
    English language

ABSTRACT
This specification describes the required characteristics of reagent waters. Four types of waters have been specified, with three additional grades that can be applied to the four types. The grade specifications specifically address contaminants of microbiological origin. Historically, reagent water types I, II, III, and IV have been linked to specific processes for their production. Starting with this revision, these types of waters may be produced with alternate technologies as long as the appropriate constituent specifications are met. The electrical conductivity and resistance, pH, silica, sodium, chlorides, TOC, endotoxins, and microbiological contamination shall be tested to meet the requirements prescribed.
SCOPE
1.1 This specification describes the required characteristics of waters deemed suitable for use with the standards under the jurisdiction of ASTM.  
1.2 The alphanumeric characters ascribed to water types and grades are specified in the manual, Form and Style for ASTM Standards. These have been assigned in order of historical precedence and should not be taken as an indication of a progression in water purity.  
1.3 Four types of waters have been specified, with three additional grades that can be applied to the four types. The grade specifications specifically address contaminants of microbiological origin.  
1.4 All applicable ASTM Standards are expected to reference one or more of these reagent water types where reagent water is needed as a component of an analytical measurement process. Where a different water type or grade is needed for an ASTM Standard, it may be added to this Specification through the ASTM Standard revision process.  
1.5 Although these water types and associated grades have been defined specifically for use with ASTM Standards, they may be appropriate for other applications. It is the responsibility of the users of this specification to ensure that the selected water types or grades are suitable for their intended use. Historically, reagent water Types I, II, III, and IV have been linked to specific processes for their production. Starting with this revision, these types of waters may be produced with alternate technologies as long as the appropriate constituent specifications are met and that water so produced has been shown to be appropriate for the application where the use of such water is specified. Therefore, the selection of an alternate technology in place of the technology specified in Table 1 should be made taking into account the potential impact of other contaminants such as microorganism and pyrogens. Such contaminants were not necessarily considered by the performance characteristics of the technology previously specified. (A) Type I grade of reagent water shall be prepared by distillation or other equal process, followed by polishing with a mixed bed of ion-exchange materials and a 0.2-µm membrane filter. Feed water to the final polishing step must have a maximum conductivity of 20 µS/cm at 298K (25°C). Type I reagent water may be produced with alternate technologies as long as the appropriate constituent specifications are met and that water so produced has been shown to be appropriate for the application where the use of such water is specified.(B) Type II grade of reagent water shall be prepared by distillation using a still designed to produce a distillate having a conductivity of less than 1.0 µS/cm at 298 K (25°C). Ion exchange, distillation, or reverse osmosis and organic adsorption may be required prior to distillation, if the purity cannot be attained by single distillation. Type II reagent water may be produced with alternate technologies as long as the appropriate constituent specifications are met and that water so produced has been shown to be appropriate for the application where the use of such water is specified.(C) Type III grade of reagent water shall be prepared by distillation, ion exchange, continuous electrodeionization, r...

  • Technical specification
    6 pages
    English language

SCOPE
1.1 This software was developed to automate calculations within three ASTM standards: Practices D2777 (outlier removal section), D6091, and D6512.  
1.2 The program calculates detection estimates (DE) and quantitation estimates (QE) for the constant, straight-line, exponential, and hybrid (Rocke-Lorenzato) models of the variation of [inter or intra] laboratory standard deviation (ILSD) with concentration. Calculations are shown in the DE_QE worksheet and results are shown in the DLs & QLs worksheet. Several plots are generated showing how well each model fits the data. The least complex model to fit the data with adequate confidence must be used by the ASTM standards.
Note 1: Modeling techniques automated in this practice and with this software have been shown to work well with most data sets. Users of this software are cautioned that with some, rare data sets, anomalous results may be obtained, and manual forcing of a different model may be required. It has been noted that for some data sets when an exponential model is selected, there may be a lack of convergence on a result or there may be a convergence on two separate results.  
1.3 Users of DQCALC should refer to Practices D2777, D6091, and D6512 for the specifics of the scope and application of the Practices.  
1.4 The IDE Practice (D6091) and the IQE Practice (D6512) are concerned with estimates of limits of detection and limits of quantitation based on inter-laboratory data. DQCALC may also be employed to calculate detection and quantitation estimates based on single laboratory data.  
1.5 The DQCALC Software consists of a Microsoft Excel3 workbook spreadsheet and associated macros and a user manual in Microsoft Word.3

  • Standard
    2 pages
    English language
  • Standard
    75 pages
    English language
  • Standard
    75 pages
    English language

SIGNIFICANCE AND USE
5.1 In order to be certain that the end user of analytical results obtained from using an ASTM Committee D19 test method can be confident that the values have been obtained through a competent application of the test method, a demonstration of the proficiency of the analytical system must be performed. Appropriate proficiency is demonstrated by achievement of performance criteria derived from results of the test method collaborative study. The QC measures specified in this practice must be included in each ASTM test method, as applicable, to ensure the quality of measurements.  
5.2 In order for users of D19 test methods to achieve consistently valid results, a minimum level of QC must be performed. This minimum level of QC is stipulated in this practice and by the task groups developing D19 test methods. If the specific requirements outlined in this practice are not applicable to the test method, alternative QC must be defined in the test method.
SCOPE
1.1 This practice provides specific, mandatory requirements for incorporating quality control (QC) procedures into all test methods under the jurisdiction of Committee D19.  
1.2 ASTM has adopted the following:
Policy on implementation of requirements for a quality control section in standard test methods generated by Committee D19 on Water.  
GENERAL—By July 29, 1998, or at the next reapproval or revision, whichever is later, every D19 Standard Test Method shall contain a QC section that is in full compliance with the requirements of this practice.  
NEW COLLABORATIVE TESTING—As of July 29, 1998, each collaborative study design shall include a QC section as part of the method to be tested. Prior to approval of the study design, the Results Advisor shall ascertain the appropriateness of the QC section in meeting the requirements of this practice and Practice D2777, and shall advise the designer of the study of any changes needed to fulfill the requirements of these practices. Before a collaborative study may be conducted, approval of the study design by the Results Advisor must be obtained.  
OLDER VALIDATED METHODS—Standard test methods that were validated using Practices D2777 – 77, D2777 – 86, or D2777 – 94, when balloted for reapproval or revision, shall contain a QC section based upon the best information from the historical record. Where appropriate, information derived from the record of the collaborative study shall be utilized for this purpose. The introduction of the QC section into these standard test methods shall not be construed as a requirement for a new collaborative study, though the Subcommittee may opt for such a study. Any information available regarding QC or precision/bias testing shall be included in the appropriate sections of the published test method.  
1.3 Required QC sections in all applicable test methods are intended to achieve two goals. First, users of Committee D19 test methods will be able to demonstrate a minimum competency in the performance of these test methods by comparison with collaborative study data. Second, all users of test methods will be required to perform a minimum level of QC as part of proper implementation of these test methods to ensure ongoing competency.  
1.4 This practice contains the primary requirements for QC of a specific test method. In many cases, it may be desirable to implement additional QC requirements to assure the desired quality of data.  
1.5 The specific requirements in this practice may not be applicable to all test methods. These requirements may vary depending on the type of test method used as well as the analyte being determined and the sample matrix being analyzed. See Explanation 1 in Appendix X1.  
1.5.1 If there are compelling reasons why any of the specific QC requirements listed in this practice are not applicable to a specific test method, these reasons must be do...

  • Standard
    12 pages
    English language
  • Standard
    12 pages
    English language

SIGNIFICANCE AND USE
5.1 Matrix spiking is commonly used to determine the bias under specific analytical conditions, or the applicability of a test method to a particular sample matrix in that context, by determining the extent to which the spiked analyte or component is recovered from the sample matrix under these conditions. Reactions or interactions of the analyte or component of interest with the sample matrix may cause a significant positive or negative effect on recovery and may render the chosen analytical, or monitoring, process ineffectual for that sample matrix.  
5.2 Matrix spiking can also be used to monitor the performance of a laboratory, individual instrument, or analyst as part of a regular quality assurance program. Changes in spike recoveries or recovery limits from the same or similar matrices over time may indicate variations in the quality of analytical results.  
5.3 Spiking can be used to compare the recoveries of like spikes from reagent water samples and natural matrix samples (measured with and without spike) to distinguish between (1) unusual interference and (2) inherent method recovery and instability effects. This guide does not attempt to deal with the statistical significance of differences in spike recoveries from different matrices.  
5.4 Special precautions shall be observed when nonlaboratory personnel perform spiking in the field. It is recommended that all spike preparation work be performed in a laboratory by experienced analysts so that the field operation consists solely of adding a prepared spiking solution to the sample matrix. Training of field personnel and validation of their spiking techniques are necessary to ensure that spikes are added accurately and reproducibly. Duplicate field spikes can be used to document the reproducibility of the technique. When environmentally labile compounds are used as spikes, the spiking solution shall be protected up to the point of use by appropriate means such as chilling, protection from sunlight and o...
SCOPE
1.1 This guide covers the general technique of “spiking” a broad range of materials into aqueous media. This guide will serve the analyst in preparing spiked samples for quality control purposes. Guidance is also provided to aid the analyst in calculating recoveries and interpreting results. It is the responsibility of the analyst to determine whether the procedures and materials described here are appropriate to the task at hand.  
1.2 The procedures in this guide are focused on “matrix spike” preparation, analysis, and interpretation of results. The applicability of these procedures to the preparation of calibration standards, calibration check standards, laboratory control standards, reference materials, and other quality control materials by spiking is incidental. A sample (the matrix) is fortified (spiked) with the analyte of interest for a variety of analytical and quality control purposes. While the spiking of multiple sample portions is discussed, the method of standard additions is not covered.  
1.3 This guide is intended for use in conjunction with the individual analytical test method that provides procedures for analysis of the analyte or component of interest. The test method is used to determine an analyte or component's background level and, again after spiking, its now elevated level. Each test method typically provides procedures not only for samples, but also for calibration standards or analytical control solutions, or both. These procedures include preparation, handling, storage, preservation, and analysis techniques. These procedures are applicable by extension, using the analyst's judgement on a case-by-case basis, to spiking solutions, and are not reiterated in this guide. See also Practice E200 for preparation and storage information.  
1.4 These procedures apply only to analytes that are soluble in water at the concentration of the spike plus any background material, or to analytes soluble in...

  • Guide
    6 pages
    English language
  • Guide
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 The user of this guide is not assumed to be a technical practitioner in the water field. This guide is an assembly of the components common to all aspects of water monitoring and fulfills a need in the development of a common framework for a better coordinated and more unified approach to monitoring water.  
4.2 Limitations—This guide does not establish a standard procedure to follow in all cases and it does not cover the details necessary to meet a particular monitoring objective.
SCOPE
1.1 Purpose—This guide is generic in its application to surface or ground water, rivers, lakes, or estuaries (quantity and quality). It proposes a series of options that offer direction without recommending a definite course of action and discusses the major elements that are common to all purposes of water monitoring.  
1.2 The elements described are applicable whether the monitoring is only for one location or integrates multiple measurement sites for the purpose of assessing a whole watershed, estuary, or aquifer system.  
1.3 This guide is intended to outline for planners and administrators the components, process, and procedures which should be considered when proposing, planning, or implementing a monitoring program. The guide is not a substitute for obtaining specific technical advice. The reader is not assumed to be a technical practitioner in the water field; however, practitioners will find it a good summary of practice and a handy checklist. Other standard guides have or will be prepared that address the necessary detail.  
1.4 Monitoring Components—A water monitoring program is composed of a set of activities, practices, and procedures designed to collect reliable information of known accuracy and precision concerning a particular water resource in order to achieve a specific goal or purpose. The purposes may range in scope from tracking status and trends on a regional or national basis to gathering data to determine the effects of a specific management practice or pollution incident such as a spill. This guide suggests and discusses the following process and components:  
1.4.1 Establishment of program goals and objectives and recording of decisions in a written plan (see 6.1),  
1.4.2 Developing background data and a conceptual model (see 6.1.12),  
1.4.3 Establishment of data (quality, quantity, type) objectives (see 6.2),  
1.4.4 Design of field measurement and sampling strategies and specification of laboratory analyses and data acceptance criteria (see 6.3),  
1.4.5 Data storage and transfer (see 6.6),  
1.4.6 Implementation of sampling and analysis strategies (see 6.4),  
1.4.7 Data quality assessment (see 6.5),  
1.4.8 Assessment of data (see 6.7),  
1.4.9 Program evaluation (see 6.8), and  
1.4.10 Reporting (see 6.9). See also Fig. X1.1 in Appendix X1 and the condensed list of headings in Appendix X2.  
1.5 Monitoring Purposes—Establishing goals defines the purpose for monitoring. Each purpose has some monitoring design needs specific to itself. There are six major purposes for water monitoring. They are as follows:  
1.5.1 Determining the Status and Trends of Water Conditions—This can require long term, regular monitoring to determine how parameters change over time.  
1.5.2 Detecting Existing and Emerging Problems—Determining if, how, or where a substance may move through an aquatic system, or if water quantities are changing.  
1.5.3 Developing and Implementing Management and Regulatory Programs—Includes baseline and reconnaissance monitoring to characterize existing conditions such as to identify critical areas or hot spots; implementation monitoring to assess whether activities were carried out as planned; and compliance monitoring to determine if specific water quality or water use criteria were met.  
1.5.4 Responding to an Emergency—Performed to provide information in the near term.  
1.5.5 Evaluating the Effectiveness of Water Monitoring Programs—Is the ...

  • Guide
    11 pages
    English language
  • Guide
    11 pages
    English language

SIGNIFICANCE AND USE
4.1 The purity of water is relative and is usually characterized by the limits of impurities found in the water as well as by the methods used to prepare and handle the water. Section 7 mentions the suitable methods for water preparation.
SCOPE
1.1 This guide is intended to describe the chemical and biological characteristics of water to be used whenever critical purity is essential to the use intended in laboratory Bio-Applications, for example, clinical, pharmaceutical, and biomedical. The importance of such a reagent is often underestimated despite the impact that it can have.  
1.2 This guide is not intended to be used as a reference in preparing water for injectables. Generally, the appropriate use of this guide may include experiments involving tissue culture, chromatography, mass spectrometry, Polymerase Chain Reaction (PCR), DeoxyriboNucleic Acid (DNA) sequencing, DNA hybridization, electrophoresis, molecular biology or analyses where molecular concentrations of impurities may be important.  
1.3 For all the other applications linked to an ASTM method and not bio-sensitive that require purified water, it is recommended that Specification D1193 or Test Method D5127 be consulted.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Guide
    4 pages
    English language

SIGNIFICANCE AND USE
5.1 Appropriate application of this practice should result in a WQE achievable by the laboratory in applying the tested method/matrix/analyte combination to routine sample analysis. That is, a laboratory should be capable of measuring concentrations greater than WQEZ %, with the associated RSD equal to Z % or less.  
5.2 The WQE values may be used to compare the quantitation capability of different methods for analysis of the same analyte in the same matrix within the same laboratory.  
5.3 The WQE procedure should be used to establish the within-laboratory quantitation capability for any application of a method in the laboratory where quantitation is important to data use. The intent of the WQE is not to impose reporting limits. The intent is to provide a reliable procedure for establishing the quantitative characteristics of the method (as implemented in the laboratory for the matrix and analyte) and thus to provide the laboratory with reliable information characterizing the uncertainty in any data produced. Then the laboratory may make informed decisions about censoring data and has the information necessary for providing reliable estimates of uncertainty with reported data.
SCOPE
1.1 This practice establishes a uniform standard for computing the within-laboratory quantitation estimate associated with Z % relative standard deviation (referred to herein as WQEZ %), and provides guidance concerning the appropriate use and application.  
1.2 WQEZ % is computed to be the lowest concentration for which a single measurement from the laboratory will have an estimated Z % relative standard deviation (Z % RSD, based on within-laboratory standard deviation), where Z is typically an integer multiple of 10, such as 10, 20, or 30. Z can be less than 10 but not more than 30. The WQE10 % is consistent with the quantitation approaches of Currie (1)2 and Oppenheimer, et al (2).  
1.3 The fundamental assumption of the WQE is that the media tested, the concentrations tested, and the protocol followed in the developing the study data provide a representative and fair evaluation of the scope and applicability of the test method, as written. Properly applied, the WQE procedure ensures that the WQE value has the following properties:  
1.3.1 Routinely Achievable WQE Value—The laboratory should be able to attain the WQE in routine analyses, using the laboratory‘s standard measurement system(s), at reasonable cost. This property is needed for a quantitation limit to be feasible in practical situations. Representative data must be used in the calculation of the WQE.  
1.3.2 Accounting for Routine Sources of Error—The WQE should realistically include sources of bias and variation that are common to the measurement process and the measured materials. These sources include, but are not limited to intrinsic instrument noise, some typical amount of carryover error, bottling, preservation, sample handling and storage, analysts, sample preparation, instruments, and matrix.  
1.3.3 Avoidable Sources of Error Excluded—The WQE should realistically exclude avoidable sources of bias and variation (that is, those sources that can reasonably be avoided in routine sample measurements). Avoidable sources would include, but are not limited to, modifications to the sample, modifications to the measurement procedure, modifications to the measurement equipment of the validated method, and gross and easily discernible transcription errors (provided there was a way to detect and either correct or eliminate these errors in routine processing of samples).  
1.4 The WQE applies to measurement methods for which instrument calibration error is minor relative to other sources, because this practice does not model or account for instrument calibration error, as is true of quantiation estimates in general. Therefore, the WQE procedure is appropriate when the dominant source of variation is not instrument calibration, but is perhaps one...

  • Standard
    13 pages
    English language

SCOPE
0

  • Standard
    75 pages
    English language
  • Standard
    75 pages
    English language

SIGNIFICANCE AND USE
4.1 This substitute ocean water may be used for laboratory testing where a reproducible solution simulating sea water is required. Examples are for tests on oil contamination, detergency evaluation, and corrosion testing. Note 2—The lack of organic matter, suspended matter, and marine life in this solution does not permit unqualified acceptance of test results as representing performance in actual ocean water. Where corrosion is involved, the results obtained from laboratory tests may not approximate those secured under natural testing conditions that differ greatly from those of the laboratory, and especially where effects of velocity, salt atmospheres, or organic constituents are involved. Also the rapid depletion of reacting elements present in low concentrations suggests caution in direct application of results.
SCOPE
1.1 This practice covers the preparation of solutions containing inorganic salts in proportions and concentrations representative of ocean water.2 Note 1—Since the concentrations of ocean water varies with sampling location, the gross concentration employed herein is an average of many reliable individual analyses. Trace elements, occurring naturally in concentrations below 0.005 mg/L, are not included.  
1.2 This practice provides three stock solutions, each relatively concentrated but stable in storage. For preparation of substitute ocean water, aliquots of the first two stock solutions with added salt are combined in larger volume. An added refinement in adjustment of heavy metal concentration is provided by the addition of a small aliquot of the third stock solution to the previous solution.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    3 pages
    English language

SIGNIFICANCE AND USE
5.1 Following this practice should result in precision-and-bias statements that can be achieved by any laboratory properly using the test method studied. These precision-and-bias statements provide the basis for generic limits for use in the Quality Control section of the test method. Optionally, the detection and quantitation values provide estimates of the level at which most laboratories should be able to achieve confident detection and meet the minimum precision (expressed as relative standard deviation) expected.  
5.2 The method specifies the matrices for which the test method is appropriate. The collaborative test corroborates the write-up within the limitations of the test design. An extensive test can only use representative matrices so that universal applicability cannot be implied from the results.  
5.3 The fundamental assumption of the collaborative study is that the matrices tested, the concentrations tested, and the participating laboratories are a representative and fair evaluation of the scope and applicability of the test method as written.
SCOPE
1.1 This practice establishes uniform standards for estimating and expressing the precision and bias of applicable test methods for Committee D19 on Water. Statements of precision and bias in test methods are required by the Form and Style for ASTM Standards, “Section A21. Precision and Bias (Mandatory).” In principle, all test methods are covered by this practice.  
1.2 Except as specified in 1.4, 1.5, and 1.6, this practice requires the task group proposing a new test method to carry out a collaborative study from which statements for precision (overall and single-operator standard-deviation estimates) and bias can be developed. This practice provides general guidance to task groups in planning and conducting such determinations of precision and bias.  
1.3 This practice also provides guidance to task groups for conducting limited-scale collaborative studies (known as “comparability studies”) for test methods that have been revised, when such revision includes substantive modifications. Examples of substantive modifications may include, but are not limited to, changes in mandatory or allowable instrumentation, reagents, reaction times, etc.  
1.3.1 Changes to applicable water matrices in the Scope of a method may constitute a substantive modification under this provision. However, recognize that even the original collaborative study may not have used all the various matrix types specified in the method’s original Scope.  
1.3.2 A method's concentration-range extension that is deemed to merit additional collaborative testing (even without a method modification that would otherwise be considered substantive) shall require a full collaborative study, as described in Sections 7.1-7.5, but only at Youden-pair concentrations representative of the extended range. Note that such a collaborative study could involve as little as a single-sample Youden-pair study in a single reproducible matrix.  
1.3.3 Whether a revision to a test method includes substantive modification shall be determined by consensus of the Committee.  
1.4 If a full-scale collaborative study is not technically feasible, due to the nature of the test method or instability of samples, the largest feasible scaled-down collaborative study shall be conducted to provide the best possible limited basis for estimating the overall and single-operator standard deviations.  
1.4.1 Examples of acceptable scaled-down studies are the local-area studies conducted by Subcommittee D19.24 on microbiological methods because of inherent sample instability. These studies involve six or more completely independent local-area analysts who can begin analysis of uniform samples at an agreed upon time.  
1.4.2 If uniform samples are not feasible under any circumstances, a statement of single-operator precision will meet the requirements of this practice. Whenever possible, this stat...

  • Standard
    21 pages
    English language
  • Standard
    21 pages
    English language

SIGNIFICANCE AND USE
5.1 Appropriate application of this practice should result in an estimate of the test-method’s uncertainty (at any concentration within the working range), which can be compared with data-quality objectives to see if the uncertainty is acceptable.  
5.2 With data sets that compare recovered concentration with true concentration, the resulting regression plot allows the correction of the recovery data to true values. Reporting of such corrections is at the discretion of the user.  
5.3 This practice should be used to estimate the measurement uncertainty for any application of a test method where measurement uncertainty is important to data use.
SCOPE
1.1 This practice establishes a standard for computing the measurement uncertainty for applicable test methods in Committee D19 on Water. The practice does not provide a single-point estimate for the entire working range, but rather relates the uncertainty to concentration. The statistical technique of regression is employed during data analysis.  
1.2 Applicable test methods are those whose results come from regression-based methods and whose data are intra-laboratory (not inter-laboratory data, such as result from round-robin studies). For each analysis conducted using such a method, it is assumed that a fixed, reproducible amount of sample is introduced.  
1.3 Calculation of the measurement uncertainty involves the analysis of data collected to help characterize the analytical method over an appropriate concentration range. Example sources of data include: 1) calibration studies (which may or may not be conducted in pure solvent), 2) recovery studies (which typically are conducted in matrix and include all sample-preparation steps), and 3) collections of data obtained as part of the method’s ongoing Quality Control program. Use of multiple instruments, multiple operators, or both, and field-sampling protocols may or may not be reflected in the data.  
1.4 In any designed study whose data are to be used to calculate method uncertainty, the user should think carefully about what the study is trying to accomplish and much variation should be incorporated into the study. General guidance on designing studies (for example, calibration, recovery) is given in Appendix A. Detailed guidelines on sources of variation are outside the scope of this practice, but general points to consider are included in Appendix B, which is not intended to be exhaustive. With any study, the user must think carefully about the factors involved with conducting the analysis, and must realize that the computed measurement uncertainty will reflect the quality of the input data.  
1.5 Associated with the measurement uncertainty is a user-chosen level of statistical confidence.  
1.6 At any concentration in the working range, the measurement uncertainty is plus-or-minus the half-width of the prediction interval associated with the regression line.  
1.7 It is assumed that the user has access to a statistical software package for performing regression. A statistician should be consulted if assistance is needed in selecting such a program.  
1.8 A statistician also should be consulted if data transformations are being considered.  
1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    7 pages
    English language

SIGNIFICANCE AND USE
5.1 Environmental data are often required for making regulatory and programmatic decisions. These data must be of known quality commensurate with their intended use.  
5.2 Certain minimal criteria must be met by the field organizations in order to meet the objectives of the water monitoring activities.  
5.3 This guide defines the criteria for organizations taking water samples and generating environmental data and identifies other activities that may be required based on the DQOs.  
5.4 This guide emphasizes the importance of communication among those involved in establishing the DQOs, planning, and implementing the sampling and analysis aspects of environmental data generation activities, and assessing data quality.
SCOPE
1.1 This guide covers planning and implementation of the sampling aspects of environmental data generation activities. Environmental data generation efforts are comprised of four parts: (1) establishment of data quality objectives (DQOs); ( 2) design of field sampling and measurement strategies and specification of laboratory analyses and data acceptance criteria; (3) implementation of sampling and analysis strategies; and ( 4) data quality assessment.  
1.2 This guide defines the criteria that must be considered to ensure the quality of the field aspects of environmental data and sample generation activities.  
1.3 DQOs should be adopted prior to the application of this guide. The data generated in accordance with this guide are subject to a final assessment to determine whether the DQOs were met. For example, many screening activities do not require all of the quality assurance (QA) and quality control (QC) steps found in this guide to generate data adequate to meet the project needs. The extent to which all of the requirements must be met remains a matter of technical judgement as it relates to the established DQOs.  
1.4 This guide presents extensive management requirements designed to ensure high-quality samples and data. The words “must,”“ shall,” “may,” and “should” have been selected carefully to reflect the importance placed on many of the statements made in this guide.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Guide
    8 pages
    English language

SIGNIFICANCE AND USE
4.1 Substitute wastewater may be used for laboratory testing where a reproducible mixture simulating municipal wastewater is required. To provide a more rugged evaluation of a test method's ability to measure analytes and the precision of the test method under more demanding conditions, it is necessary to utilize a matrix more complex than reagent water. Substitute wastewater is intended to be used as a material to be spiked with analytes or interferents of interest to evaluate the performance of test methods.  
4.2 Utilization of substitute wastewater as a matrix in interlaboratory validation studies will allow inclusion of performance statistics for this matrix in the Precision and Bias section of test methods. Users of test methods will be able to evaluate their application of test methods in this matrix against published results.
SCOPE
1.1 This practice covers the preparation of an aqueous mixture containing constituents in concentrations such that it will have physical and chemical matrix characteristics similar to municipal wastewater.  
1.2 Wastewaters are extremely variable, depending on the quantity and nature of the materials being discharged into the collection system. The mixture prepared with this practice is not representative of any particular wastewater. Rather, it allows the user to prepare a mixture that exhibits a similar matrix impact on test method performance as is typical of municipal wastewater and can be prepared from common materials inexpensively and reproducibly. It allows the evaluation of test methods, over time, against the same reference point.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    3 pages
    English language

SIGNIFICANCE AND USE
5.1 In order to obtain meaningful analytical data, sample preservation techniques must be effective from the time of sample collection to the time of analysis. A laboratory must confirm that sample integrity is maintained throughout maximum time periods between sample collection and analysis. In many cases, it is useful to know the maximum holding time. An evaluation of holding time is useful also in judging the efficacy of various preservation techniques.
SCOPE
1.1 This practice covers the means of estimating the period of time during which a water sample can be stored after collection and preservation without significantly affecting the accuracy of analysis.  
1.2 The maximum holding time is dependent upon the matrix used and the specific analyte of interest. Therefore, water samples from a specific source must be tested to determine the period of time that sample integrity is maintained by standard preservation practices.  
1.3 In the event that it is not possible to analyze the sample immediately at the time of collection, this practice does not provide information regarding degradation of the constituent of interest or changes in the matrix that may occur from the time of sample collection to the time of the initial analysis.  
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 This practice is designed to assist suppliers and users of reference materials by identifying the information necessary on the certificate of analysis of materials designated for use in ASTM test methods. This practice is specifically designed to ensure that materials suitable for use as either calibration or quality control standards are available. This practice does not define a specific certification protocol, but rather provides guidance in the development of adequate data to support the use of the material as either a calibration or quality control standard. Suppliers are referred to ISO Guide 35 for guidelines on acceptable certification protocols. End users are referred to ISO Guide 31 for a more complete description of the elements of typical certificates of analysis.
SCOPE
1.1 This practice covers the information that must be provided on certificates of analysis of reference materials designated to support ASTM methods. It provides end users of these materials with a defined set of data that is required to be on a certificate of analysis and provides information to assist the end user in evaluating the independence of the material. Similarly, it provides the suppliers of reference materials with a consistent format for the presentation of certification data.  
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 This guide recommends the water quality required for the electronics and microelectronics industries. High-purity water is required to prevent contamination of products during manufacture, since contamination can lead to an unacceptable, low yield of electronic devices.  
4.2 The range of water purity is defined in accordance with the manufacturing process. The types of ultra-pure water are defined with respect to device line width. In all cases, the water-quality recommendations apply at the point of distribution.  
4.3 The limits on the impurities are related to current contamination specifications and to available analytical methods (either performed in a suitable clean laboratory or by on-line instrumentation). On-line and off-line methods are used in accordance with current industry practice. Concentration of the sample may be required to measure the impurities at the levels indicated in Table 1.TABLE 1 Requirements for Water at the Point of Distribution in the Electronics and Semiconductor IndustriesA    
Parameter  
Type E-1  
Type E-1.1  
Type E-1.2B  
Type E-1.3B  
Type E-2  
Type E-3  
Type E-4  
Linewidth (microns)  
1.0–0.5  
0.35–0.25  
0.18–0.09  
0.065–0.032  
5.0–1.0  
>5.0  
—  
Resistivity, 25°C (On-line)  
18.1  
18.2  
18.2  
18.2  
16.5  
12  
0.5  
TOC (μg/L) (on-line for  
5  
2  
1  
1  
50  
300  
1000  
On-line dissolved oxygen (μg/L)  
25  
10  
3  
10  
—  
—  
—  
On-Line Residue after evaporation (μg/L)  
1  
0.5  
0.1  
—  
—  
—  
On-line particles/L (micron range)  
>0.05 μm  
500  
0.05–0.1  
1000  
200  
N/A  
—  
—  
—  
0.1–0.2  
1000  
350  
N/A  
—  
—  
—  
0.2–0.5  
500  
N/A  
—  
—  
—  
0.5-1.0  
200  
N/A  
—  
—  
—  
1.0  
N/A  
—  
—  
—  
SEM particles/L (micron range)  
...
SCOPE
1.1 This guide provides recommendations for water quality related to electronics and semiconductor-industry manufacturing. Seven classifications of water are described, including water for line widths as low as 0.032 micron. In all cases, the recommendations are for water at the point of distribution (POD).  
1.2 Water is used for washing and rinsing of semiconductor components during manufacture. Water is also used for cleaning and etching operations, making steam for oxidation of silicon surfaces, preparing photomasks, and depositing luminescent materials. Other applications are in the development and fabrication of solid-state devices, thin-film devices, communication lasers, light-emitting diodes, photo-detectors, printed circuits, memory devices, vacuum-tube devices, or electrolytic devices.  
1.3 Users needing water qualities different from those described here should consult other water standards, such as Specification D1193 and Guide D5196.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Guide
    6 pages
    English language
  • Guide
    6 pages
    English language

SIGNIFICANCE AND USE
5.1 In order to obtain meaningful analytical data, sample preservation techniques must be effective from the time of sample collection to the time of analysis. A laboratory must confirm that sample integrity is maintained throughout maximum time periods between sample collection and analysis. In many cases, it is useful to know the maximum holding time. An evaluation of holding time is useful also in judging the efficacy of various preservation techniques.
SCOPE
1.1 This practice covers the means of estimating the period of time during which a water sample can be stored after collection and preservation without significantly affecting the accuracy of analysis.  
1.2 The maximum holding time is dependent upon the matrix used and the specific analyte of interest. Therefore, water samples from a specific source must be tested to determine the period of time that sample integrity is maintained by standard preservation practices.  
1.3 In the event that it is not possible to analyze the sample immediately at the time of collection, this practice does not provide information regarding degradation of the constituent of interest or changes in the matrix that may occur from the time of sample collection to the time of the initial analysis.  
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
The user of this guide is not assumed to be a trained technical practitioner in the water quality field. The guide is an assembly of the components common to all aspect of watershed nutrient monitoring and fulfills a need in the development of a common framework for a better coordinated and a more unified approach to nutrient monitoring in watersheds.
Limitations—This guide does not establish a standard procedure to follow in all situations and it does not cover the detail necessary to meet all of the needs of a particular monitoring objective. Other standards and guides included in the references describe the detail of the procedures.
SCOPE
1.1 Purpose—This guide is intended to provide general guidance on a watershed monitoring program directed toward the plant nutrients nitrogen and phosphorus. The guide offers a series of general steps without setting forth a specific course of action. It gives assistance for developing a monitoring program but not a program for implementing measures to improve water quality.  
1.2 This guide applies to waters found in streams and rivers; lakes, ponds, and reservoirs; estuaries; wetlands; the atmosphere; and the vadose and subsurface saturated zones (including aquifers). This guide does not apply to nutrients found in soils, plants, or animals.
1.3 Nutrients as used in this guide are intended to include nitrogen and phosphorus in dissolved, gaseous, and particulate forms. Specific species of nitrogen include: nitrate, nitrite, ammonia, organic, total Kjeldahl, and nitrous oxide. The species of phosphorus include total, total dissolved, organic, acid-hydrolyzable, and reactive phosphorus as described in (2)
1.4 Safety—Health and safety practices developed for a project may need to consider the following:
1.4.1 During the construction of sampling stations:
1.4.1.1 Drilling practices during monitoring well installations,
1.4.1.2 Overhead and underground utilities during monitoring well drilling,
1.4.1.3 Traffic patterns/concerns during sampling station installation,
1.4.1.4 Traffic patterns/concerns during surveying sampling station locations and elevations,
1.4.1.5 Drilling through materials highly contaminated with fertilizers, and
1.4.1.6 Installing monitoring equipment below the soil surface.
1.4.2 During the collection of water samples:
1.4.2.1 Using acids for sample preservation,
1.4.2.2 Sampling during flooding events and ice conditions,
1.4.2.3 Traffic on bridges,
1.4.2.4 Condition of sampling stations following flood events,
1.4.2.5 Sampling of water or soils, or both, highly contaminated with fertilizers,
1.4.2.6 Conditions of sampling stations resulting from vandalism,
1.4.2.7 Adverse weather conditions, and
1.4.2.8 Transporting liquid samples.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Guide
    7 pages
    English language

SIGNIFICANCE AND USE
This practice is for the use by RM producers in the development and implementation of their quality system and by those concerned with assessing the competence of RM producers. It should be recognized that a RM needs to be characterized mainly to the level of accuracy required for its intended purpose (that is, appropriate measurement uncertainty). The RM producer shall describe the procedure for establishing the quality of materials as a component of the quality system.
This practice is for the use of RM users in the establishment if a RM producer has a quality system adequate to produce high quality RMs. It can be used by users to determine if the scientific and technical competence of a RMs producer is adequate to ensure the quality of RMs. This practice is consistent with the requirements for RM producers established in ISO Guide 34.
This practice does not specify specific protocols for the contents of RMs certificates of analysis, for calibration in analytical chemistry and use of certified RMs and for certification of RMs. For this information, users are referred to Practice D6362, ISO Guide 32, and ISO Guide 35.
SCOPE
1.1 This practice establishes the general requirements with which a reference materials (RM) producer has to demonstrate that it operates, if it is to be recognized as competent to produce RMs used for water analysis.  
1.2 This practice establishes the quality system requirements in accordance with which waters RMs shall be produced. It is intended to be used as part of a RM producer's general QA procedures. RM producers shall define their scope in terms of the application, the measurement methods used in the homogeneity, stability and characterization studies.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    12 pages
    English language

SIGNIFICANCE AND USE
In order to be certain that the end user of analytical results obtained from using an ASTM Committee D-19 test method can be confident that the values have been obtained through a competent application of the test method, a demonstration of the proficiency of the analytical system must be performed. Appropriate proficiency is demonstrated by achievement of performance criteria derived from results of the test method collaborative study. The QC measures specified in this practice must be included in each ASTM test method, as applicable, to ensure the quality of measurements.
In order for users of D-19 test methods to achieve consistently valid results, a minimum level of QC must be performed. This minimum level of QC is stipulated in this practice and by the taskgroups developing D-19 test methods. If the specific requirements outlined in this practice are not applicable to the test method, alternative QC must be defined in the test method.
SCOPE
1.1 This practice provides specific, mandatory requirements for incorporating quality control (QC) procedures into all test methods under the jurisdiction of Committee D-19.
1.2 ASTM has adopted the following:
Policy on implementation of requirements for a quality control section in standard test methods generated by Committee D-19 on Water.      GENERALBy July 29, 1998, or at the next reapproval or revision, whichever is later, every D-19 Standard Test Method shall contain a QC section that is in full compliance with the requirements of this practice.  NEW COLLABORATIVE TESTINGAs of July 29, 1998, each collaborative study design shall include a QC section as part of the method to be tested. Prior to approval of the study design, the Results Advisor shall ascertain the appropriateness of the QC section in meeting the requirements of this Practice and Practice D2777, and shall advise the designer of the study of any changes needed to fulfill the requirements of these practices. Before a collaborative study may be conducted, approval of the study design by the Results Advisor must be obtained.  OLDER VALIDATED METHODSStandard test methods that were validated using D-2777-77, D-2777-86, or D-2777-94, when ballotted for reapproval or revision, shall contain a QC section based upon the best information from the historical record. Where appropriate, information derived from the record of the collaborative study shall be utilized for this purpose. The introduction of the QC section into these standard test methods shall not be construed as a requirement for a new collaborative study, though the Subcommittee may opt for such a study. Any information available regarding QC or precision/bias testing shall be included in the appropriate sections of the published method.  
1.3 Required QC sections in all applicable test methods are intended to achieve two goals. First, users of Committee D-19 test methods will be able to demonstrate a minimum competency in the performance of these test methods by comparison with collaborative study data. Second, all users of test methods will be required to perform a minimum level of QC as part of proper implementation of these test methods to ensure ongoing competency.
1.4 This practice contains the primary requirements for QC of a specific test method. In many cases, it may be desirable to implement additional QC requirements to assure the desired quality of data.
1.5 The specific requirements in this practice may not be applicable to all test methods. These requirements may vary depending on the type of test method used as well as the analyte being determined and the sample matrix being analyzed. See Explanation 1 in Appendix X1.
1.5.1 If there are compelling reasons why any of the specific QC requirements listed in this practice are not applicable to a specific test method, these reasons must be documented in the QC section of the test method.
1.5.2 With the approval of Committee D-19 on the recommendation o...

  • Standard
    11 pages
    English language

SIGNIFICANCE AND USE
This guide recommends the water quality required for the electronics and microelectronics industries. High-purity water is required to prevent contamination of products during manufacture, since contamination can lead to an unacceptable, low yield of electronic devices.
The range of water purity is defined in accordance with the manufacturing process. The types of ultra-pure water are defined with respect to device line width. In all cases, the water-quality recommendations apply at the point of distribution.
The limits on the impurities are related to current contamination specifications and to available analytical methods (either performed in a suitable clean laboratory or by on-line instrumentation). On-line and off-line methods are used in accordance with current industry practice. Concentration of the sample may be required to measure the impurities at the levels indicated in Table 1.
TABLE 1 Requirements for Water at the Point of Distribution in the Electronics and Semiconductor IndustriesA  ParameterType E-1Type E-1.1Type E-1.2BType E-1.3BType E-2Type E-3Type E-4 Linewidth (microns)1.0–0.50.35–0.250.18–0.090.065–0.0325.0–1.0>5.0 Resistivity, 25°C (On-line)18.118.218.218.216.5120.5 TOC (μg/L) (on-line for 10 ppb)5211503001000 On-line dissolved oxygen (μg/L)2510310 On-Line Residue after evaporation (μg/L)10.50.1 On-line particles/L (micron range) >0.05 μm500C  0.05–0.11000200N/AC  0.1–0.21000350100N/A  0.2–0.550010010N/A  0.5-1.0200505N/A  1.0100201N/A SEM particles/L (micron range)  0.1–0.21000700250N/A  0.2–0.5500400100N/A3000  0.5–11005030N/A10 000  10503010N/A100 000 Bacteria in CFU/Volume  100 mL Sample531N/A1050100  1 L Sample101 10 L Sample 1 Silica – total (μg/L)5310.510501000 Silica – dissolved (μg/L)310.50.5 Anions and Ammonium by IC (μg/L)  Ammonium0.10.100.050.050  Bromide0.10.050.020.050  Chloride0.10.050.020.0501101000  Fluoride0.10.050.030.050  Nitrate0.10.050.020.05015500  Nitrite0.10.050.020.050  Phosphate0.10.050.020.05015500  Sulfate0.1...
SCOPE
1.1 This guide provides recommendations for water quality related to electronics and semiconductor-industry manufacturing. Seven classifications of water are described, including water for line widths as low as 0.032 micron. In all cases, the recommendations are for water at the point of distribution (POD).
1.2 Water is used for washing and rinsing of semiconductor components during manufacture. Water is also used for cleaning and etching operations, making steam for oxidation of silicon surfaces, preparing photomasks, and depositing luminescent materials. Other applications are in the development and fabrication of solid-state devices, thin-film devices, communication lasers, light-emitting diodes, photo-detectors, printed circuits, memory devices, vacuum-tube devices, or electrolytic devices.
1.3 Users needing water qualities different from those described here should consult other water standards, such as Specification D1193 and Guide D5196.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Guide
    6 pages
    English language

SIGNIFICANCE AND USE
Following this practice should result in precision-and-bias statements that can be achieved by any laboratory properly using the test method studied. These precision-and-bias statements provide the basis for generic limits for use in the Quality Control section of the test method. Optionally, the detection and quantitation values provide estimates of the level at which most laboratories should be able to achieve confident detection and meet the minimum precision (expressed as relative standard deviation) expected.
The method specifies the matrices for which the test method is appropriate. The collaborative test corroborates the write-up within the limitations of the test design. An extensive test can only use representative matrices so that universal applicability cannot be implied from the results.
The fundamental assumption of the collaborative study is that the matrices tested, the concentrations tested, and the participating laboratories are a representative and fair evaluation of the scope and applicability of the test method as written.
SCOPE
1.1 This practice establishes uniform standards for estimating and expressing the precision and bias of applicable test methods for Committee D19 on Water. In principle, all test methods are covered by this practice.
1.2 Except as specified in 1.4, 1.5, and 1.6, this practice requires the task group proposing a new test method to carry out a collaborative study from which statements for precision (overall and single-operator standard-deviation estimates) and bias can be developed. This practice provides general guidance to task groups in planning and conducting such determinations of precision and bias.  
1.3 This practice also provides guidance to task groups for conducting limited-scale collaborative studies (known as “comparability studies”) for test methods that have been revised, when such revision includes substantive modifications. Examples of substantive modifications may include, but are not limited to, changes in mandatory or allowable instrumentation, reagents, reaction times, etc.
1.3.1 Changes to applicable water matrices in the Scope of a method may constitute a substantive modification under this provision. However, recognize that even the original collaborative study may not have used all the various matrix types specified in the method’s original Scope.
1.3.2 A method's concentration-range extension that is deemed to merit additional collaborative testing (even without a method modification that would otherwise be considered substantive) shall require a full collaborative study, as described in Sections 7.1-7.5, but only at Youden-pair concentrations representative of the extended range. Note that such a collaborative study could involve as little as a single-sample Youden-pair study in a single reproducible matrix.
1.3.3 Whether a revision to a test method includes substantive modification shall be determined by consensus of the Committee.
1.4 If a full-scale collaborative study is not technically feasible, due to the nature of the test method or instability of samples, the largest feasible scaled-down collaborative study shall be conducted to provide the best possible limited basis for estimating the overall and single-operator standard deviations.
1.4.1 Examples of acceptable scaled-down studies are the local-area studies conducted by Subcommittee D19.24 on microbiological methods because of inherent sample instability. These studies involve six or more completely independent local-area analysts who can begin analysis of uniform samples at an agreed upon time.
1.4.2 If uniform samples are not feasible under any circumstances, a statement of single-operator precision will meet the requirements of this practice. Whenever possible, this statement should be developed from data generated by independent multiple operators, each doing replicate analyses on independent samples (of a specific matrix type), which gener...

  • Standard
    21 pages
    English language

SIGNIFICANCE AND USE
This guide is intended to be used in the planning stage or phase of developing a sediment monitoring program. This guide is an assembly of the components common to all aspects of watershed sediment monitoring and fulfills a need in the development of a common framework for a better coordinated and a more unified approach to sediment monitoring in watersheds.
The user of this guide is not assumed to be a trained technical practitioner in the water quality, sedimentation, or hydrology fields. The intended users are managers and planners who need information to develop a water quality monitoring program or project with an emphasis in sediment and hydrology. Sediment specialists will also find information on procedures, equipment, methodology, and operations to conduct a monitoring program.
This guide is used during the planning process of developing, designing, and reevaluating a sediment monitoring program.
SCOPE
1.1 PurposeThis guide is intended to provide general guidance on a watershed monitoring program directed toward sediment. The guide offers a series of general steps without setting forth a specific course of action. It gives advice for establishing a monitoring program, not an implementation program.
1.2 Sedimentation as referred to in this guide is the detachment, entrainment, transportation, and deposition of eroded soil and rock particles. Specific types or parameters of sediment may include: suspended sediment, bedload, bed material, turbidity, wash load, sediment concentration, total load, sediment deposits, particle size distribution, sediment volumes and particle chemistry. Monitoring may include not only sediments suspended in water but sediments deposited in fields, floodplains, and channel bottoms.
1.3 This guide applies to surface waters as found in streams and rivers; lakes, ponds, reservoirs, estuaries, and wetlands.
1.4 LimitationsThis guide does not establish a standard procedure to follow in all situations and it does not cover the detail necessary to define all of the needs of a particular monitoring objective or project. Other standards and guides included in the reference and standard sections describe in detail the procedures, equipment, operations, and site selection for collecting, measuring, analyzing, and monitoring sediment and related constituants.
1.5 Additional ASTM and US Geological Survey standards applicable to sediment monitoring are listed in Appendix X1 and Appendix X2. Due to the large number of optional standards and procedures involved in sediment monitoring, most individual standards are not referenced in this document. Standards and procedures have been grouped in the appendices according to the type of analyses or sampling that would be required for a specific type of measurement or monitoring.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Guide
    10 pages
    English language

SCOPE
1.1 This Practice describes a procedure for developing a graphical model of relative standard deviation vs concentration for a analytical methods used in the analysis of water (methods that are subject to non-additive random errors) for the purpose of assigning a statement of noise or randomness to analytical results (commonly referred to as a precision statement), in either a manual or an automated fashion.
1.2 Data analysis and modeling is done with D19 Adjunct DQCALC (an Excel based tool).
1.3 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
Data on the composition and characteristics of water are frequently used to evaluate the health and safety to humans and the environment.
Moreover, such data are frequently used for process control or to ascertain compliance with regulatory statutes that place limits on acceptable compositions and characteristics of waters.
Laboratories that conduct water sampling and generate analytical data, and those persons who have the responsibility for selecting a laboratory to perform water quality studies, need to use criteria, guidelines, and recommendations that have been developed by consensus and are well accepted in making this selection.
Demonstration and documentation by a laboratory that there was judicious selection and control of organization, facilities, resources, and operations will enhance the credibility of the data produced and promote its acceptance.
SCOPE
1.1 This guide provides information on consensus good laboratory practices for laboratories that provide services in the sampling and analysis of water. As consensus standards, these are the minimum criteria that all laboratories should consider in establishing their good laboratory practices.
1.2 This guide is designed to be used by those responsible for the selection, operation, or control of laboratory organizations engaged in sampling and analysis of water.
1.3 This guide presents features of organization, facilities, resources, and operations which affect the usefulness of the data generated.
1.4 This guide presents criteria for selection and control of the features described in and also makes recommendations for the correction of unacceptable performance.
1.5 This guide describes methodology and practices intended to be completely consistent with the International Organization for Standardization (ISO) 9000 series of standards and Guide 25 - 1990 ().
1.6 The values stated in inch-pound units are to be regarded as the standard. The values given in parentheses are for information only.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Guide
    12 pages
    English language
  • Guide
    12 pages
    English language

SIGNIFICANCE AND USE
This guide establishes basic requirements which should be met by water and environmental laboratories that generate and report test chemical analyses which the laboratory client desires to be traceable to SI units (Note 2) or certified reference materials traceable to SI units. Traceability of chemical analyses is important because it provides a uniform basis for the comparison of results from different measurement systems and because it relates those results to our current knowledge of physical laws. (Note 3)
Note 2—A certified reference material traceable to SI units is a certified reference material whose value can be related with a stated uncertainty through an unbroken change of comparisons to stated references (usually national or international standards) in SI units, such as a primary measurement made in SI units or a national standard certified in SI units.
Note 3—Not all chemical analysis results can be traceable to SI units or to certified reference material's traceable to SI units, such as turbidity and or total suspended solids.
Many waters-related laboratories comply with ISO Guide 17025 and participate in Proficiency Testing Programs. Laboratories that are connected to the same accreditation bodies and Proficiency Test providers can be expected to report statistically similar results on the same sample. However, some test methods and some certified reference materials are not supported with data traceable to SI units. Therefore, fully compliant laboratories that are not connected to the same providers may report statistically different chemical analysis results if they used the same nontraceable test method on the same sample. This problem could be minimized if they used test methods, measurement devices, and certified reference materials that are traceable to SI units, where available.
Although some standard test methods and certified reference materials provide evidence of traceability to SI units, many others do not. Therefore, not all lab...
SCOPE
1.1 This guide sets a protocol for generating and reporting chemical analyses that are traceable to SI units or to certified reference materials in laboratories that serve the water and environmental industry.
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Guide
    3 pages
    English language

SIGNIFICANCE AND USE
The user of this guide is not assumed to be a technical practitioner in the water field. This guide is an assembly of the components common to all aspects of water monitoring and fulfills a need in the development of a common framework for a better coordinated and more unified approach to monitoring water.
Limitations— This guide does not establish a standard procedure to follow in all cases and it does not cover the details necessary to meet a particular monitoring objective.
SCOPE
1.1 Purpose—This guide is generic in its application to surface or ground water, rivers, lakes, or estuaries (quantity and quality). It proposes a series of options that offer direction without recommending a definite course of action and discusses the major elements that are common to all purposes of water monitoring.
1.2 The elements described are applicable whether the monitoring is only for one location or integrates multiple measurement sites for the purpose of assessing a whole watershed, estuary, or aquifer system.  
1.3 This guide is intended to outline for planners and administrators the components, process, and procedures which should be considered when proposing, planning, or implementing a monitoring program. The guide is not a substitute for obtaining specific technical advice. The reader is not assumed to be a technical practitioner in the water field; however, practitioners will find it a good summary of practice and a handy checklist. Other standard guides have or will be prepared that address the necessary detail.
1.4 Monitoring Components—A water monitoring program is composed of a set of activities, practices, and procedures designed to collect reliable information of known accuracy and precision concerning a particular water resource in order to achieve a specific goal or purpose. The purposes may range in scope from tracking status and trends on a regional or national basis to gathering data to determine the effects of a specific management practice or pollution incident such as a spill. This guide suggests and discusses the following process and components:

  • Guide
    9 pages
    English language

Frequently Asked Questions

D19.02 is a Technical Committee within ASTM International. It is named "Quality Systems, Specification, and Statistics". This committee has published 142 standards.

D19.02 develops ASTM standards in the area of Information technology. Currently, there are 142 published standards from this technical committee.

ASTM is a standardization organization that develops and publishes standards to support industry, commerce, and regulatory requirements.

A Technical Committee (TC) in ASTM is a group of experts responsible for developing international standards in a specific technical area. TCs are composed of national member body delegates and work through consensus to create standards that meet global industry needs. Each TC may have subcommittees (SCs) and working groups (WGs) for specialized topics.