Standard Specification for Color Coding of Reusable Laboratory Pipets

ABSTRACT
This specification covers a system of color coding of reusable laboratory pipets for identification of nominal capacities and units of sub-division. This specification applies to one-mark and graduated pipets of nominal capacities. The color code and marking requirements for one-mark and graduated pipets shall take the form of color bands in accordance to the color coding system indicated in this specification.
SCOPE
1.1 This specification covers a system of color coding for one-mark pipets for identification of nominal capacities, and for graduated pipets for identification of nominal capacities and units of sub-division.
1.2 This specification applies to one-mark and graduated pipets of the nominal capacities listed in Table 1 and Table 2, respectively.
Note 1—Many pipets not covered by other ASTM standards are included in and , in order, firstly, to ensure uniformity of color coding as far as possible for non-standard as well as standard pipets and, secondly, to reserve suitable codes for possible future standards for other types of pipet.

General Information

Status
Historical
Publication Date
30-Oct-1988
Current Stage
Ref Project

Relations

Buy Standard

Technical specification
ASTM E1273-88(2003) - Standard Specification for Color Coding of Reusable Laboratory Pipets
English language
2 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E 1273 – 88 (Reapproved 2003)
Standard Specification for
Color Coding of Reusable Laboratory Pipets
This standard is issued under the fixed designation E 1273; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
TABLE 1 Coding System for One-Mark Pipets
1. Scope
Nominal Capacity, mL Color Code Bands
1.1 This specification covers a system of color coding for
0.001 1 blue
one-mark pipets for identification of nominal capacities, and
0.002 2 red
forgraduatedpipetsforidentificationofnominalcapacitiesand
0.003 1 yellow
units of sub-division.
0.004 2 green
0.005 1 white
1.2 This specification applies to one-mark and graduated
0.01 1 orange
pipets of the nominal capacities listed in Table 1 and Table 2,
0.015 2 blue
respectively.
0.02 1 black
0.025 2 white
NOTE 1—Many pipets not covered by other ASTM standards are
0.03 2 yellow
included in Table 1 and Table 2, in order, firstly, to ensure uniformity of 0.035 2 black
0.04 2 red
color coding as far as possible for non-standard as well as standard pipets
0.05 1 green
and, secondly, to reserve suitable codes for possible future standards for
0.075 2 orange
other types of pipet.
0.1 1 blue
0.15 1 white
2. Referenced Documents
0.2 1 red
0.25 2 green
2.1 ASTM Standards:
0.3 1 yellow
E 1157 Specification for the Sampling and Testing of Reus-
0.4 2 red
able Laboratory Glassware 0.5 2 black
1 1 blue
2 1 orange
3. Significance and Use
3 1 black
3.1 The purpose of this specification is to ensure that if a 4 2 red
5 1 white
color code is used on pipets, all manufacturers will use the
6 2 orange
same code; it is not intended as an encouragement of color
7 2 green
coding if this is not required. 8 1 blue
9 1 black
10 1 red
4. Color Properties
15 1 green
4.1 Coding—The color coding used on one-mark pipets
20 1 yellow
25 1 blue
shall be in accordance with Table 1, and graduated pipets shall
30 1 black
be in accordance with Table 2.
40 1 white
4.2 Variations—Variations in the enamels used in the meth- 50 1 red
75 1 green
ods of application appropriate for pipets made from different
100 1 yellow
types of glass inevitably result in minor variations of color, and
150 2 black
it is therefore not appropriate to specify closely the seven 200 1 blue
colors listed in the tables.
5. Marking Requirements
and situated not more than 70 mm from the top of the pipet and
5.1 The color code shall take the form of color bands
not less than 20 mm above the nearest graduation line.
extending at least 150° around the circumference of the pipet
5.2 For a code consisting of a single band of color, the band
shall be 5 to 10 mm wide. For a code consisting of two bands
Th
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.