Standard Method for Force Calibration Of Thermomechnical Analyzers

SCOPE
1.1 This method describes the calibration or performance confirmation of the electronically applied force signal for thermomechanical analyzers over the range of 0 to 1 N.
1.2 SI units are the standard.
1.3 There is no ISO method equivalent to this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
09-May-2002
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E2206-02 - Standard Method for Force Calibration Of Thermomechnical Analyzers
English language
3 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E2206–02
Standard Method for
1
Force Calibration Of Thermomechnical Analyzers
This standard is issued under the fixed designation E2206; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope be in conformance if the performance is within established
limits, typically 1%. Alternatively, the force signal may be
1.1 This method describes the calibration or performance
calibrated using a two-point calibration method.
confirmation of the electronically applied force signal for
thermomechanical analyzers over the range of 0 to 1 N.
5. Significance and Use
1.2 SI units are the standard.
5.1 Most thermomechanical analysis experiments are car-
1.3 There is no ISO method equivalent to this standard.
ried out with some force applied to the test specimen. This
1.4 This standard does not purport to address all of the
force is often created electronically. It may be constant or
safety concerns, if any, associated with its use. It is the
changed during the experiment.
responsibility of the user of this standard to establish appro-
5.2 Thismethoddemonstratesconformanceorcalibratesthe
priate safety and health practices and determine the applica-
electronically applied force signal.
bility of regulatory limitations prior to use.
5.3 Thismethodmaybeusedforresearchanddevelopment,
2. Referenced Documents quality control, manufacturing or regulatory applications.
5.4 Other thermomechanical analyzer calibration functions
2.1 ASTM Standards:
2
include temperature byTest Method E1363 and length change
E4 Practice for Load Verification of Testing Machines
3
by Test Method E2113.
E473 Terminology Relating to Thermal Analysis
E617 Specification for Laboratory Weights and Precision
6. Apparatus
4
Mass Standards
6.1 Thermomechanical Analyzer—The essential instrumen-
E831 Test Method for Linear Thermal Expansion of Solid
3 tationrequiredtoprovideaminimumthermomechanicalanaly-
Materials by Thermomechanical Analysis
sis or thermodilatometric capability for this method includes:
E1142 Terminology Relating to Thermophysical Proper-
3 6.1.1 Rigid Specimen Holder, inert, low expansivity mate-
ties
rial [typically < 0.6 µm/(m · K)] to center the specimen in the
E1363 Test Method for Temperature Calibration of Ther-
3 furnace and to fix the specimen to mechanical ground.
momechanical Analyzers
E2113 Test Method for Length Change Calibration of
NOTE 1—Materials of construction with greater expansivity may be
3
Thermomechanical Analyzers used but shall be reported.
6.1.2 Rigid (Expansion or Compression) Probe, inert, low
3. Terminology
expansivity material [typically < 0.6 µm/(m·K)] which con-
3.1 The technical terms used in this standard are defined in
tactsthespecimenwithanappliedcompressiveforce(seeNote
Terminologies E473 and E1142.
1).
6.1.3 Sensing Element, linear over a minimum range of
4. Summary of Test Method
2mmtomeasurethedisplacementoftherigidprobeto 61µm
4.1 The electronic force signal generated by a thermome-
resulting from changes in length of the specimen.
chanical analyzer is compared to that exerted by gravity on a
6.1.4 Programmable Force Transducer, to generate a con-
known mass. The thermomechanical analyzer may be said to
stant force (6 1.0%) of up to 1.0 N that is applied through the
rigid probe to the specimen.
1
This method is under the jurisdiction of ASTM Committee E37 on Thermal
NOTE 2—Other force ranges may be used but shall be reported.
Measurements and is the direct responsibility of Subcommittee E37.01 on Thermal
6.1.5 Furnace, capable of providing uniform controlled
Analysis Test Methods.
Current edition approved May 10, 2002. Published August 2002.
heating (cooling) of the specimen to a constant temperature or
2
Annual Book of ASTM Standards, Vol 03.01.
at a constant rate within the temperature range of −100 to
3
Annual Book of ASTM Standards, Vol 14.02.
600°C.
4
Annual Book of ASTM Standards, Vol 14.04.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1

---------------------- Page: 1 ----------------------
E2206–02
NOTE 3—Other temperature ranges may be used but shall be reported.
force transducer so that no force is applied by the probe to the
specimen holder. Record this force as F in mN.
6.1.6 Temperature Controller, capable of executing a spe- 2
cific temperature program by operating the furnace between
NOTE 7—Other masses may be used but shall be reported.
selected temperature limits at a rate of change of up to
8.3 Calculate the force calibration constant (S) and confor-
10°C/min constant to 0.1°C/min or at an isothermal t
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.