Standard Practice for Calculation of Gas Chromatographic Response Factors

SIGNIFICANCE AND USE
5.1 ASTM standard gas chromatographic methods for the analysis of petroleum products require calibration of the gas chromatographic system by preparation and analysis of specified reference mixtures. Frequently, minimal information is given in these methods on the practice of calculating calibration or response factors. Test Methods D2268, D2427, D2804, D2998, D3329, D3362, D3465, D3545, and D3695 are examples. The present practice helps to fill this void by providing a detailed reference procedure for calculating response factors, as exemplified by analysis of a standard blend of C6 to C11 n-paraffins using n-C12 as the diluent.  
5.2 In practice, response factors are used to correct peak areas to a common base prior to final calculation of the sample composition. The response factors calculated in this practice are “multipliers” and prior to final calculation of the results the area obtained for each compound in the sample should be multiplied by the response factor determined for that compound.  
5.3 It has been determined that values for response factors will vary with individual installations. This may be caused by variations in instrument design, columns, and experimental techniques. It is necessary that chromatographs be individually calibrated to obtain the most accurate data.
SCOPE
1.1 This practice covers a procedure for calculating gas chromatographic response factors. It is applicable to chromatographic data obtained from a gaseous mixture or from any mixture of compounds that is normally liquid at room temperature and pressure or solids, or both, that will form a solution with liquids. It is not intended to be applied to those compounds that react in the chromatograph or are not quantitatively eluted. Normal C6 through C11  paraffins have been chosen as model compounds for demonstration purposes.  
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
30-Sep-2023
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM D4626-23 - Standard Practice for Calculation of Gas Chromatographic Response Factors
English language
3 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM D4626-23 - Standard Practice for Calculation of Gas Chromatographic Response Factors
English language
3 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D4626 − 23
Designation: 378/87
Standard Practice for
1
Calculation of Gas Chromatographic Response Factors
This standard is issued under the fixed designation D4626; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope* Hydrocarbons in Gasolines by Gas Chromatography
3
(Withdrawn 2023)
1.1 This practice covers a procedure for calculating gas
D2804 Test Method for Purity of Methyl Ethyl Ketone By
chromatographic response factors. It is applicable to chromato-
Gas Chromatography
graphic data obtained from a gaseous mixture or from any
D2998 Test Method for Polyhydric Alcohols in Alkyd Res-
mixture of compounds that is normally liquid at room tempera-
3
ins (Withdrawn 2004)
ture and pressure or solids, or both, that will form a solution
D3329 Test Method for Purity of Methyl Isobutyl Ketone by
with liquids. It is not intended to be applied to those com-
Gas Chromatography
pounds that react in the chromatograph or are not quantitatively
D3362 Test Method for Purity of Acrylate Esters by Gas
eluted. Normal C through C paraffins have been chosen as
3
6 11
Chromatography (Withdrawn 2011)
model compounds for demonstration purposes.
D3465 Guide for Purity of Monomeric Plasticizers by Gas
1.2 The values stated in SI units are to be regarded as Chromatography
standard. No other units of measurement are included in this D3545 Test Method for Alcohol Content and Purity of
standard. Acetate Esters by Gas Chromatography
D3695 Test Method for Volatile Alcohols in Water by Direct
1.3 This standard does not purport to address all of the
Aqueous-Injection Gas Chromatography
safety concerns, if any, associated with its use. It is the
D4175 Terminology Relating to Petroleum Products, Liquid
responsibility of the user of this standard to establish appro-
Fuels, and Lubricants
priate safety, health, and environmental practices and deter-
D4307 Practice for Preparation of Liquid Blends for Use as
mine the applicability of regulatory limitations prior to use.
Analytical Standards
1.4 This international standard was developed in accor-
E260 Practice for Packed Column Gas Chromatography
dance with internationally recognized principles on standard-
E355 Practice for Gas Chromatography Terms and Relation-
ization established in the Decision on Principles for the
ships
Development of International Standards, Guides and Recom-
mendations issued by the World Trade Organization Technical
3. Terminology
Barriers to Trade (TBT) Committee.
3.1 Definitions:
3.1.1 This test method makes reference to common gas
2. Referenced Documents
chromatographic procedures, terms, and relationships. Detailed
2
2.1 ASTM Standards:
definitions of these can be found in Practices E260, E355
D2268 Test Method for Analysis of High-Purity n-Heptane
and/or Terminology D4175.
and Isooctane by Capillary Gas Chromatography
3.2 Definitions of Terms Specific to This Standard:
D2427 Test Method for Determination of C through C
2 5 3.2.1 response factor (R)—a constant of proportionality
used to convert the observed chromatographic response of
specific compounds to either mass or volume percent compo-
1
sition. The observed response may be measured as peak areas
This practice is under the jurisdiction of ASTM Committee D02 on Petroleum
Products, Liquid Fuels, and Lubricants and is the direct responsibility of Subcom-
or peak heights. Depending on the calculation formula, the
mittee D02.04.0L on Gas Chromatography Methods.
response factor (R) is applied by either multiplying or dividing
Current edition approved Oct. 1, 2023. Published November 2023. Originally
the observed response by the determined factor.
approved in 1986. Last previous edition approved in 2019 as D4626 – 95 (2019).
DOI: 10.1520/D4626-23.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
3
Standards volume information, refer to the standard’s Document Summary page on The last approved version of this historical standard is referenced on
the ASTM website. www.astm.org.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
D4626 − 23
NOTE 1—Rapidly eluting peaks such as those produced by a capillary
3.2.2 In this practice, the response factors determi
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D4626 − 95 (Reapproved 2019) D4626 − 23
Designation: 378/87
Standard Practice for
1
Calculation of Gas Chromatographic Response Factors
This standard is issued under the fixed designation D4626; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope Scope*
1.1 This practice covers a procedure for calculating gas chromatographic response factors. It is applicable to chromatographic data
obtained from a gaseous mixture or from any mixture of compounds that is normally liquid at room temperature and pressure or
solids, or both, that will form a solution with liquids. It is not intended to be applied to those compounds that react in the
chromatograph or are not quantitatively eluted. Normal C through C paraffins have been chosen as model compounds for
6 11
demonstration purposes.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of
regulatory limitations prior to use.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2
2.1 ASTM Standards:
D2268 Test Method for Analysis of High-Purity n-Heptane and Isooctane by Capillary Gas Chromatography
3
D2427 Test Method for Determination of C through C Hydrocarbons in Gasolines by Gas Chromatography (Withdrawn 2023)
2 5
D2804 Test Method for Purity of Methyl Ethyl Ketone By Gas Chromatography
3
D2998 Test Method for Polyhydric Alcohols in Alkyd Resins (Withdrawn 2004)
D3329 Test Method for Purity of Methyl Isobutyl Ketone by Gas Chromatography
3
D3362 Test Method for Purity of Acrylate Esters by Gas Chromatography (Withdrawn 2011)
D3465 Guide for Purity of Monomeric Plasticizers by Gas Chromatography
D3545 Test Method for Alcohol Content and Purity of Acetate Esters by Gas Chromatography
D3695 Test Method for Volatile Alcohols in Water by Direct Aqueous-Injection Gas Chromatography
D4175 Terminology Relating to Petroleum Products, Liquid Fuels, and Lubricants
D4307 Practice for Preparation of Liquid Blends for Use as Analytical Standards
1
This practice is under the jurisdiction of ASTM Committee D02 on Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of Subcommittee
D02.04.0L on Gas Chromatography Methods.
Current edition approved Dec. 1, 2019Oct. 1, 2023. Published December 2019November 2023. Originally approved in 1986. Last previous edition approved in 20152019
as D4626 – 95 (2015).(2019). DOI: 10.1520/D4626-95R19.10.1520/D4626-23.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
3
The last approved version of this historical standard is referenced on www.astm.org.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
D4626 − 23
E260 Practice for Packed Column Gas Chromatography
E355 Practice for Gas Chromatography Terms and Relationships
3. Terminology
3.1 Definitions:
3.1.1 This test method makes reference to common gas chromatographic procedures, terms, and relationships. Detailed definitions
of these can be found in Practices E260, E355 and/or Terminology D4175.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 response factor (R)—a constant of proportionality used to convert the observed chromatographic response of specific
compounds to either mass or volume percent composition. The observed response may be measured as peak areas or peak heights.
Depending on the calculation formula, the response factor (R) is applied by either mu
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.