Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-091: Particular requirements for cabinet X-ray systems

IEC 61010-2-091:2012 specifies particular safety requirements for CABINET X-RAY SYSTEMS. A CABINET X-RAY SYSTEM is a system that contains an X-ray tube installed in a cabinet which, independently of existing architectural structures except the floor on which it may be placed, is intended to contain at least that portion of a material being irradiated, provide radiation attenuation, and exclude personnel from the interior during generation of X-radiation. These CABINET X-RAY SYSTEMS are used in industrial, commercial, and public environments, for example, to inspect materials, to analyze materials, and to screen baggage.

Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2-091: Besondere Anforderungen für Röntgengeräteschränke

Règles de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire - Partie 2-091: Exigences particulières pour les équipements à rayons X montés en armoire

La CEI 61010-2-091:2012 définit les exigences de sécurité particulières pour les EQUIPEMENTS A RAYONS X MONTES EN ARMOIRE. Un EQUIPEMENT A RAYONS X MONTE EN ARMOIRE est un équipement doté d'un tube radiogène installé dans une armoire qui, indépendamment des structures architecturales existantes à l'exception du sol sur lequel il peut être placé, est destiné à contenir au moins une partie d'un matériau en cours d'irradiation, à fournir une atténuation du rayonnement et à maintenir le personnel à l'extérieur de son enceinte durant la production de rayons X. Ces EQUIPEMENTS A RAYONS X MONTES EN ARMOIRE sont utilisés dans des environnements industriel, commercial et public pour l'inspection et l'analyse des matériaux, et pour l'examen des bagages.

Varnostne zahteve za električno opremo za meritve, nadzorovanje in laboratorijsko uporabo - 2-091. del: Posebne zahteve za rentgenske sisteme v ohišjih (IEC 61010-2-091:2012)

Ta del standarda IEC 61010 določa posebne varnostne zahteve za RENTGENSKE SISTEME V OHIŠJIH. RENTGENSKI SISTEM V OHIŠJU je sistem z rentgensko cevjo, nameščeno v ohišju, ki neodvisno od obstoječih arhitekturnih konstrukcij, razen tal, na katerih lahko stoji, predvidoma vsebuje najmanj material, ki ga obseva, zagotavlja zmanjšanje sevanja in izključuje prisotnost osebja v prostoru med rentgenskim sevanjem. Ti RENTGENSKI SISTEMI V OHIŠJU se uporabljajo v industrijskem, trgovinskem in javnem okolju, na primer za pregled in analizo materialov ter pregled prtljage.

General Information

Status
Published
Publication Date
02-Aug-2012
Current Stage
6060 - Document made available
Due Date
03-Aug-2012
Completion Date
03-Aug-2012

RELATIONS

Buy Standard

Standard
EN 61010-2-091:2012
English language
17 pages
sale 10% off
Preview
sale 10% off
Preview

e-Library read for
1 day

Standards Content (sample)

SLOVENSKI STANDARD
SIST EN 61010-2-091:2012
01-oktober-2012

9DUQRVWQH]DKWHYH]DHOHNWULþQRRSUHPR]DPHULWYHQDG]RURYDQMHLQODERUDWRULMVNR

XSRUDERGHO3RVHEQH]DKWHYH]DUHQWJHQVNHVLVWHPHYRKLãMLK ,(&



Safety requirements for electrical equipment for measurement, control and laboratory

use - Part 2-091: Particular requirements for cabinet x-ray systems (IEC 61010-2-

091:2012)

Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2

-091: Besondere Anforderungen für Röntgengeräteschränke (IEC 61010-2-091:2012)
Règles de sécurité pour appareils électriques de mesurage, de régulation et de

laboratoire - Partie 2-091: Exigences particulières pour les équipements à rayons x

montés en armoire (CEI 61010-2-091:2012)
Ta slovenski standard je istoveten z: EN 61010-2-091:2012
ICS:
19.080 (OHNWULþQRLQHOHNWURQVNR Electrical and electronic
SUHVNXãDQMH testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
SIST EN 61010-2-091:2012 en

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
SIST EN 61010-2-091:2012
---------------------- Page: 2 ----------------------
SIST EN 61010-2-091:2012
EUROPEAN STANDARD
EN 61010-2-091
NORME EUROPÉENNE
August 2012
EUROPÄISCHE NORM
ICS 19.080; 71.040.10
English version
Safety requirements for electrical equipment for measurement, control
and laboratory use -
Part 2-091: Particular requirements for cabinet x-ray systems
(IEC 61010-2-091:2012)
Règles de sécurité pour appareils Sicherheitsbestimmungen für elektrische

électriques de mesurage, de régulation et Mess-, Steuer-, Regel- und Laborgeräte -

de laboratoire - Teil 2-091: Besondere Anforderungen für
Partie 2-091: Exigences particulières pour Röntgengeräteschränke
les équipements à rayons x montés en (IEC 61010-2-091:2012)
armoire
(CEI 61010-2-091:2012)

This European Standard was approved by CENELEC on 2012-07-30. CENELEC members are bound to comply

with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard

the status of a national standard without any alteration.

Up-to-date lists and bibliographical references concerning such national standards may be obtained on

application to the CEN-CENELEC Management Centre or to any CENELEC member.

This European Standard exists in three official versions (English, French, German). A version in any other

language made by translation under the responsibility of a CENELEC member into its own language and notified

to the CEN-CENELEC Management Centre has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus,

the Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany,

Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland,

Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom.

CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
Management Centre: Avenue Marnix 17, B - 1000 Brussels

© 2012 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.

Ref. No. EN 61010-2-091:2012 E
---------------------- Page: 3 ----------------------
SIST EN 61010-2-091:2012
EN 61010-2-091:2012 - 2 -
Foreword

The text of document 66/462/FDIS, future edition 1 of IEC 61010-2-091, prepared by IEC/TC 66 "Safety of

measuring, control and laboratory equipment" was submitted to the IEC-CENELEC parallel vote and

approved by CENELEC as EN 61010-2-091:2012.
The following dates are fixed:
(dop) 2013-04-30
• latest date by which the document has
to be implemented at national level by
publication of an identical national
standard or by endorsement
(dow) 2015-07-30
• latest date by which the national
standards conflicting with the
document have to be withdrawn

Attention is drawn to the possibility that some of the elements of this document may be the subject of

patent rights. CENELEC [and/or CEN] shall not be held responsible for identifying any or all such patent

rights.

This standard covers the Principle Elements of the Safety Objectives for Electrical Equipment Designed

for Use within Certain Voltage Limits (LVD - 2006/95/EC).

This Part 2-091 is intended to be used in conjunction with EN 61010-1:2010. Consideration may be given

to future editions of, or amendments to, EN 61010-1.

This Part 2-091 supplements or modifies the corresponding clauses in EN 61010-1 so as to convert that

publication into the EN standard: Particular requirements for CABINET X-RAY SYSTEMS.

Where a particular subclause of Part 1 is not mentioned in this part 2, that subclause applies as far as is

reasonable. Where this part states “addition”, “modification”, “replacement”, or “deletion”, the relevant

requirement, test specification or note in Part 1 should be adapted accordingly.
In this standard:
a) the following print types are used:
– requirements: in roman type;
– NOTES: in small roman type;
– conformity and tests: in italic type;

– terms used throughout this standard which have been defined in Clause 3: SMALL ROMAN CAPITALS.

b) subclauses, figures, and tables which are additional to those in Part 1 are numbered starting from 101;

additional annexes are lettered starting from AA and additional list items are lettered from aa).

Endorsement notice

The text of the International Standard IEC 61010-2-091:2012 was approved by CENELEC as a European

Standard without any modification.
---------------------- Page: 4 ----------------------
SIST EN 61010-2-091:2012
- 3 - EN 61010-2-091:2012
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications

The following documents, in whole or in part, are normatively referenced in this document and are

indispensable for its application. For dated references, only the edition cited applies. For undated

references, the latest edition of the referenced document (including any amendments) applies.

NOTE When an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD

applies.
Addition to Annex ZA of EN 61010-1:
Publication Year Title EN/HD Year
IEC 62061 - Safety of machinery - Functional safety of EN 62061 -
safety-related electrical, electronic and
programmable electronic control systems
ISO 13849-1 - Safety of machinery - Safety-related parts of EN ISO 13849-1 -
control systems -
Part 1: General principles for design
---------------------- Page: 5 ----------------------
SIST EN 61010-2-091:2012
---------------------- Page: 6 ----------------------
SIST EN 61010-2-091:2012
IEC 61010-2-091
Edition 1.0 2012-06
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
GROUP SAFETY PUBLICATION
PUBLICATION GROUPÉE DE SÉCURITÉ
Safety requirements for electrical equipment for measurement, control, and
laboratory use –
Part 2-091: Particular requirements for cabinet X-ray systems
Règles de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire –
Partie 2-091: Exigences particulières pour les équipements à rayons X montés
en armoire
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
CODE PRIX N
ICS 19.080; 71.040.10 ISBN 978-2-83220-145-9

Warning! Make sure that you obtained this publication from an authorized distributor.

Attention! Veuillez vous assurer que vous avez obtenu cette publication via un distributeur agréé.

® Registered trademark of the International Electrotechnical Commission
Marque déposée de la Commission Electrotechnique Internationale
---------------------- Page: 7 ----------------------
SIST EN 61010-2-091:2012
– 2 – 61010-2-091 © IEC:2012
CONTENTS

FOREWORD ........................................................................................................................... 3

1 Scope and object .............................................................................................................. 5

2 Normative references ....................................................................................................... 6

3 Terms and definitions ....................................................................................................... 6

4 Tests ................................................................................................................................ 7

5 Marking and documentation .............................................................................................. 7

6 Protection against electric shock ...................................................................................... 8

7 Protection against mechanical HAZARDS ............................................................................ 8

8 Resistance to mechanical stresses ................................................................................... 9

9 Protection against the spread of fire ................................................................................. 9

10 Equipment temperature limits and resistance to heat ........................................................ 9

11 Protection against HAZARDS from fluids ............................................................................. 9

12 Protection against radiation, including laser sources, and against sonic and

ultrasonic pressure ........................................................................................................... 9

13 Protection against liberated gases and substances, explosion and implosion ................. 10

14 Components and subassemblies .................................................................................... 11

15 Protection by interlocks .................................................................................................. 11

16 HAZARDS resulting from application ................................................................................. 11

17 RISK assessment ............................................................................................................ 12

Annex L (informative) Index of defined terms ....................................................................... 13

Bibliography .......................................................................................................................... 14

---------------------- Page: 8 ----------------------
SIST EN 61010-2-091:2012
61010-2-091 © IEC:2012 – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT
FOR MEASUREMENT, CONTROL, AND LABORATORY USE –
Part 2-091: Particular requirements for CABINET X-RAY SYSTEMS
FOREWORD

1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising

all national electrotechnical committees (IEC National Committees). The object of IEC is to promote

international co-operation on all questions concerning standardization in the electrical and electronic fields. To

this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,

Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC

Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested

in the subject dealt with may participate in this preparatory work. International, governmental and non-

governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely

with the International Organization for Standardization (ISO) in accordance with conditions determined by

agreement between the two organizations.

2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international

consensus of opinion on the relevant subjects since each technical committee has representation from all

interested IEC National Committees.

3) IEC Publications have the form of recommendations for international use and are accepted by IEC National

Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC

Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any

misinterpretation by any end user.

4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications

transparently to the maximum extent possible in their national and regional publications. Any divergence

between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in

the latter.

5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity

assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any

services carried out by independent certification bodies.

6) All users should ensure that they have the latest edition of this publication.

7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and

members of its technical committees and IEC National Committees for any personal injury, property damage or

other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and

expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC

Publications.

8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is

indispensable for the correct application of this publication.

9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of

patent rights. IEC shall not be held responsible for identifying any or all such patent rights.

International Standard IEC 61010-2-91 has been prepared by IEC technical committee 66:

Safety of measuring, control and laboratory equipment.
It has the status of a group safety publication as specified in IEC Guide 104.
The text of this standard is based on the following documents:
FDIS Report on voting
66/462/FDIS 66/470/RVD

Full information on the voting for the approval of this standard can be found in the report on

voting indicated in the above table.
---------------------- Page: 9 ----------------------
SIST EN 61010-2-091:2012
– 4 – 61010-2-091 © IEC:2012

This Part 2-091 is intended to be used in conjunction with IEC 61010-1. It was established on

the basis of the third edition (2010). Consideration may be given to future editions of, or

amendments to, IEC 61010-1.

This Part 2-091 supplements or modifies the corresponding clauses in IEC 61010-1 so as to

convert that publication into the IEC standard: Particular requirements for CABINET X-RAY

SYSTEMS.

Where a particular subclause of Part 1 is not mentioned in this part 2, that subclause applies

as far as is reasonable. Where this part states “addition”, “modification”, “replacement”, or

“deletion”, the relevant requirement, test specification or note in Part 1 should be adapted

accordingly.
In this standard:
a) the following print types are used:
– requirements: in roman type;
– NOTES: in small roman type;
– conformity and tests: in italic type;
– terms used throughout this standard which have been defined in Clause 3: SMALL
ROMAN CAPITALS.

b) subclauses, figures, and tables which are additional to those in Part 1 are numbered

starting from 101; additional annexes are lettered starting from AA and additional list

items are lettered from aa).

This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.

The following differing practices exist in the countries indicated below:
– 7.1: Conveyor systems are required to meet the requirements of
ANSI/ASME B20.1 (USA).

A list of all parts of the IEC 61010 series, published under the general title Safety

requirements for electrical equipment for measurement, control, and laboratory use, can be

found on the IEC website.

The committee has decided that the contents of this publication will remain unchanged until

the stability date indicated o
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.