Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 031: Safety requirements for hand-held and hand-manipulated probe assemblies for electrical test and measurement

IEC 61010-031:2022 specifies safety requirements for hand-held and hand-manipulated probe assemblies for electrical test and measurement, and their related accessories. These probe assemblies are for non-contact or direct electrical connection between a part and electrical test and measurement equipment. They can be fixed to the equipment or be detachable accessories for the equipment. It has the status of a group safety publication in accordance with IEC Guide 104. This third edition cancels and replaces the second edition published in 2015, and Amendment 1:2018. IEC 61010-031 is a stand-alone standard. This edition includes the following significant technical changes with respect to the previous edition:
the scope has been made succinct. General information from the scope of Edition 2 has been moved to a new Clause 4. Consequently, Clause 4 to Clause 8 of Edition 2 have been renumbered. Clause 9 of Edition 2 has been deleted;
in Clause 2, normative references have been dated and new normative references have been added;
in 3.1.4, the definition of probe tip has been modified;
in 4.1, there is no longer any differentiation between high voltage and low voltage probe assemblies. Type C probe assemblies have been merged with Type B probe assemblies;
in 4.1 d) "Kelvin" probes have been added to the list of probe assemblies as a new Type E and a new Figure 5;
in 4.1 e), probes for voltage measurement without electrical connection to conductors have been added to the list of probe assemblies as a new Type F and a new Figure 6;
in 4.2.1, spread of fire is no longer considered as a hazard;
Subclause 4.4.2.5 from Edition 2 has been deleted;
Subclause 4.4.4.3 from Edition 2 has been deleted;
in 5.4.4.1 consideration has been given to spacings and impedance;
in 6.1.1, removable parts of probe tips which bear markings are allowed;
in 6.1.5, the voltage to be marked for measurement categories is the AC line-to-neutral or DC voltage;
in 7.4.2, requirements for unmated connectors have been modified as follows:
Table 2 has been modified and expanded,
a calculation method for clearances of connectors above 20 kV has been defined,
creepage distances have been aligned with clearances;  
in 7.4.3.1 and 7.4.3.5, requirements for IP2X probe tips with retractable sleeve have been added;
in 7.4.3.2, Probe tips are now applicable to non-contact probe assemblies;
in 7.5.2.3.2, the values of Table 5 have been modified;

Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 031: Sicherheitsbestimmungen für handgehaltenes und handgeführtes Messzubehör zum elektrischen Messen und Prüfen

Exigences de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire - Partie 031: Exigences de sécurité pour sondes équipées tenues à la main et manipulées pour mesurage et essais électriques

L’IEC 61010-031:2022 spécifie les exigences de sécurité relatives aux sondes équipées tenues à la main et manipulées prévues pour le mesurage et les essais électriques, ainsi que leurs accessoires connexes. Ces sondes équipées sont prévues pour une connexion sans contact électrique ou une connexion électrique directe entre une partie et un appareil de mesure et d’essai électrique. Elles peuvent être solidaires de l’appareil ou en être des accessoires détachables. Elle a le statut d’une publication groupée de sécurité conformément au Guide IEC 104. Cette troisième édition annule et remplace la deuxième édition parue en 2015 et l’Amendement 1:2018. L’IEC 61010-031 est une norme autonome. Cette édition inclut les modifications techniques majeures suivantes par rapport à l’édition précédente:
le domaine d’application a été raccourci. Les informations générales du domaine d’application de l’Edition 2 ont été déplacées dans un nouvel Article 4. Par conséquent, les Articles 4 à 8 de l’Edition 2 ont été renumérotés. L’Article 9 de l’Edition 2 a été supprimé;
à l’Article 2, les références normatives ont été datées et de nouvelles références normatives ont été ajoutées;
en 3.1.4, la définition de pointe de touche a été modifiée;
en 4.1, il n’y a plus de distinction entre les sondes équipées haute tension et basse tension. Les sondes équipées de types B et C ont été fusionnées;
en 4.1 d), les sondes "Kelvin" ont été ajoutées à la liste des sondes équipées en tant que nouveau type E, ainsi qu’une nouvelle Figure 5;
en 4.1 e), des sondes de mesure de la tension sans connexion électrique aux conducteurs ont été ajoutées à la liste des sondes équipées en tant que nouveau type F, ainsi qu’une nouvelle Figure 6;
en 4.2.1, la propagation du feu n’est plus considérée comme un danger;
le paragraphe 4.4.2.5 de l’Edition 2 a été supprimé;
le paragraphe 4.4.4.3 de l’Edition 2 a été supprimé;
en 5.4.4.1, une attention particulière a été accordée aux espacements et à l’impédance;
en 6.1.1, les parties amovibles des pointes de touche qui portent des marquages sont autorisées;
en 6.1.5, la tension à marquer pour les catégories de mesure est la tension alternative phase-neutre ou la tension continue;
en 7.4.2, les exigences relatives aux connecteurs découplés ont été modifiées comme suit:
le Tableau 2 a été modifié et étendu;
une méthode de calcul des distances d’isolement des connecteurs au-delà de 20 kV a été définie;
les lignes de fuite ont été alignées avec les distances d’isolement;  
en 7.4.3.1 et 7.4.3.5, des exigences ont été ajoutées pour les pointes de touche IP2X avec ma

Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko uporabo - 031. del: Varnostne zahteve za sestave ročnih sond za električne meritve in preskušanja (IEC 61010-031:2022)

Ta del standarda IEC 61010 določa varnostne zahteve za sestave ročnih in ročno vodenih sond za električna preskušanja in meritve ter pripadajočo dodatno opremo. Ti sestavi sond so namenjeni brezkontaktni ali neposredni električni povezavi med delom in električnimi preskušanji ter merilno opremo. Pritrdijo se lahko na opremo ali so odstranljivi dodatki za opremo.
Ta publikacija skupinske varnosti, ki se osredotoča na bistvene varnostne zahteve, se uporablja predvsem kot standard o varnosti izdelka za izdelke, ki so omenjeni v okviru publikacije, vendar je namenjena tudi tehničnim odborom pri pripravi publikacij za izdelke, ki so podobni izdelkom, omenjenim v okviru te publikacije skupinske varnosti, skladno z načeli vodil IEC 104 in ISO/IEC 51.
Ena od dolžnosti tehničnega odbora je, da med pripravo publikacij uporablja osnovne varnostne publikacije in/ali publikacije skupinske varnosti, kadar je to primerno.

General Information

Status
Published
Publication Date
25-May-2023
Current Stage
6060 - Document made available - Publishing
Start Date
26-May-2023
Due Date
02-May-2023
Completion Date
26-May-2023

Relations

Buy Standard

Standard
EN IEC 61010-031:2023 - BARVE
English language
103 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day
Draft
prEN IEC 61010-031:2022
English language
90 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN IEC 61010-031:2023
01-november-2023
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 031. del: Varnostne zahteve za sestave ročnih sond za električne
meritve in preskušanja (IEC 61010-031:2022)
Safety requirements for electrical equipment for measurement, control and laboratory
use - Part 031: Safety requirements for hand-held probe assemblies for electrical
measurement and test (IEC 61010-031:2022)
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil
031: Sicherheitsbestimmungen für handgehaltenes Messzubehör zum elektrischen
Messen und Prüfen (IEC 61010-031:2022)
Règles de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire - Partie 031: Exigences de sécurité pour sondes équipées tenues à la main
pour mesurage et essais électriques (IEC 61010-031:2022)
Ta slovenski standard je istoveten z: EN IEC 61010-031:2023
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
SIST EN IEC 61010-031:2023 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
SIST EN IEC 61010-031:2023

---------------------- Page: 2 ----------------------
SIST EN IEC 61010-031:2023


EUROPEAN STANDARD EN IEC 61010-031

NORME EUROPÉENNE

EUROPÄISCHE NORM May 2023
ICS 19.080 Supersedes EN 61010-031:2015; EN 61010-
031:2015/A1:2021; EN 61010-031:2015/A11:2021
English Version
Safety requirements for electrical equipment for measurement,
control, and laboratory use - Part 031: Safety requirements for
hand-held and hand-manipulated probe assemblies for electrical
test and measurement
(IEC 61010-031:2022)
Exigences de sécurité pour appareils électriques de Sicherheitsbestimmungen für elektrische Mess-, Steuer-,
mesurage, de régulation et de laboratoire - Partie 031: Regel- und Laborgeräte - Teil 031:
Exigences de sécurité pour sondes équipées tenues à la Sicherheitsbestimmungen für handgehaltenes und
main et manipulées pour mesurage et essais électriques handgeführtes Messzubehör zum elektrischen Messen und
(IEC 61010-031:2022) Prüfen
(IEC 61010-031:2022)
This European Standard was approved by CENELEC on 2023-03-06. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Türkiye and the United Kingdom.



European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2023 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
 Ref. No. EN IEC 61010-031:2023 E

---------------------- Page: 3 ----------------------
SIST EN IEC 61010-031:2023
EN IEC 61010-031:2023 (E)
European foreword
The text of document 66/770/FDIS, future edition 3 of IEC 61010-031:2022, prepared by IEC/TC 66
"Safety of measuring control and laboratory equipment" was submitted to the IEC-CENELEC parallel
vote and approved by CEN-CENELEC as EN IEC 61010-031:2023.
The following dates are fixed:
• latest date by which this document has to be (dop) 2023-12-06
implemented at national level by publication of an
identical national standard or by endorsement
• latest date by which the national standards conflicting (dow) 2026-03-06
with this document have to be withdrawn
This document supersedes EN 61010-031:2015 and all of its amendments and corrigenda (if any).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
This document has been prepared under a Standardization Request given to
...

SLOVENSKI STANDARD
oSIST prEN IEC 61010-031:2022
01-junij-2022
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 031. del: Varnostne zahteve za ročne sonde za električne meritve in
preskušanja
Safety requirements for electrical equipment for measurement, control and laboratory
use - Part 031: Safety requirements for hand-held probe assemblies for electrical
measurement and test
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil
031: Sicherheitsbestimmungen für handgehaltenes Messzubehör zum elektrischen
Messen und Prüfen
Règles de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire - Partie 031: Exigences de sécurité pour sondes équipées tenues à la main
pour mesurage et essais électriques
Ta slovenski standard je istoveten z: prEN IEC 61010-031:2022
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
oSIST prEN IEC 61010-031:2022 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
oSIST prEN IEC 61010-031:2022

---------------------- Page: 2 ----------------------
oSIST prEN IEC 61010-031:2022
66/757/CDV

COMMITTEE DRAFT FOR VOTE (CDV)
PROJECT NUMBER:
IEC 61010-031 ED3
DATE OF CIRCULATION: CLOSING DATE FOR VOTING:
2022-04-01 2022-06-24
SUPERSEDES DOCUMENTS:
66/734/CD, 66/753/CC

IEC TC 66 : SAFETY OF MEASURING, CONTROL AND LABORATORY EQUIPMENT
SECRETARIAT: SECRETARY:
United Kingdom Mr David Hyde
OF INTEREST TO THE FOLLOWING COMMITTEES: PROPOSED HORIZONTAL STANDARD:

TC 78,TC 85
Other TC/SCs are requested to indicate their interest, if
any, in this CDV to the secretary.
FUNCTIONS CONCERNED:
EMC ENVIRONMENT QUALITY ASSURANCE SAFETY
SUBMITTED FOR CENELEC PARALLEL VOTING NOT SUBMITTED FOR CENELEC PARALLEL VOTING
Attention IEC-CENELEC parallel voting
The attention of IEC National Committees, members of
CENELEC, is drawn to the fact that this Committee Draft
for Vote (CDV) is submitted for parallel voting.
The CENELEC members are invited to vote through the
CENELEC online voting system.

This document is still under study and subject to change. It should not be used for reference purposes.
Recipients of this document are invited to submit, with their comments, notification of any relevant patent rights of
which they are aware and to provide supporting documentation.

TITLE:
Safety requirements for electrical equipment for measurement, control and laboratory use -
Part 031: Safety requirements for hand-held probe assemblies for electrical measurement and
test

PROPOSED STABILITY DATE: 2025

NOTE FROM TC/SC OFFICERS:


Copyright © 2022 International Electrotechnical Commission, IEC. All rights reserved. It is permitted to
download this electronic file, to make a copy and to print out the content for the sole purpose of preparing National
Committee positions. You may not copy or "mirror" the file or printed version of the document, or any part of it,
for any other purpose without permission in writing from IEC.

---------------------- Page: 3 ----------------------
oSIST prEN IEC 61010-031:2022
IEC CDV 61010-031 © IEC 2022 2 66/757/CDV

1 CONTENTS
2 1 Scope and object . 8
3 1.1 Scope . 8
4 1.1.1 Probe assemblies included in scope . 8
5 1.1.2 Probe assemblies excluded from scope . 11
6 1.2 Object . 12
7 1.2.1 Aspects included in scope . 12
8 1.2.2 Aspects excluded from scope . 12
9 1.3 Verification . 12
10 1.4 Environmental conditions . 12
11 1.4.1 Normal environmental conditions . 12
12 1.4.2 Extended environmental conditions . 12
13 2 Normative references. 13
14 3 Terms and definitions . 13
15 3.1
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.