IEC/IEEE 62704-1:2026
(Main)Determining the peak spatial-average specific absorption rate (SAR) in the human body from wireless communications devices, 100 kHz to 10 GHz - Part 1: General requirements for using the finite-difference time-domain (FDTD) method for SAR calculations
- Abstract
IEC/IEEE 62704-1:2026 defines the methodology for the application of the finite-difference time domain (FDTD) technique when used for determining the peak spatial-average specific absorption rate (psSAR) in the human body exposed to wireless communication devices with known uncertainty. It defines methods to validate the computational model of the device under test (DUT) and to assess its uncertainty when used in SAR simulations. Moreover, it defines procedures to determine the psSAR in a cubical volume and to verify the correct implementation of the FDTD simulation software. The applicable frequency range is 100 kHz to 10 GHz.
This document does not recommend specific SAR limits; these are found elsewhere, for example, in the guidelines published by the International Commission on Non-Ionizing Radiation Protection (ICNIRP) or in IEEE Std C95.1.
This second edition cancels and replaces the first edition published in 2017. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) specification of a SAR averaging algorithm for the calculation of the psSAR from computational results for arbitrarily shaped homogeneous phantoms, which avoids inconsistencies with respect to the algorithm for SAR averaging algorithm specified in IEC/IEEE 62209-1528 for SAR measurement methods;
b) provision of SHA-256 checksums for all supplemental files;
c) addition of brief informative discussion on developmental stages of IEC and IEEE computational modelling standards for SAR evaluations.
- Status
- Published
- Publication Date
- 11-Oct-2026
- Current Stage
- PPUB - Publication issued - 60 Publication
- Start Date
- 12-Oct-2026
- Due Date
- 17-Apr-2026
- Completion Date
- 17-Apr-2026
Buy Documents
IEC/IEEE 62704-1:2026 - Determining the peak spatial-average specific absorption rate (SAR) in the human body from wireless communications devices, 100 kHz to 10 GHz - Part 1: General requirements for using the finite-difference time-domain (FDTD) method for SAR calculations
IEC/IEEE 62704-1:2026 is a dual-logo International Standard published by IEC and IEEE. It defines how to apply the finite-difference time-domain (FDTD) method to determine peak spatial-average specific absorption rate (psSAR) in the human body from wireless communications devices over 100 kHz to 10 GHz. It also covers model validation, uncertainty evaluation, whole-body average SAR, and code verification for people who build or review SAR simulation tools.
What does IEC/IEEE 62704-1:2026 specify?
IEC/IEEE 62704-1:2026 specifies the computational method, reporting, uncertainty, and verification framework for SAR simulations based on FDTD. It is aimed at wireless communications devices and human exposure assessment, not at setting SAR limits.
The scope says the document defines methods to:
- validate the computational model of the device under test (DUT)
- assess simulation uncertainty
- determine psSAR in a cubical volume
- verify that FDTD simulation software has been implemented correctly
The document does not set exposure limits. The scope points readers to other sources for limits, such as ICNIRP guidelines and IEEE Std C95.1.
The document is organized into clauses 1 to 8, with supporting annexes and supplemental files.
| Annex | What it covers |
|---|---|
| Annex A | Basic FDTD requirements, update coefficients, stability, and numerical-dispersion background |
| Annex B | SAR Star benchmark, including CAD files, mesh lines, and benchmark output formats |
| Annex C | Practical FDTD considerations, sources and loads, S-parameters, power and efficiency, inhomogeneous meshes, and standards background |
| Annex D | Background on tissue modelling, anatomical models, and recommended phantom models |
| Annex E | Download links and SHA-256 checksums for supplemental files |
What are the key requirements of IEC/IEEE 62704-1:2026?
IEC/IEEE 62704-1:2026 requires a specific three-dimensional FDTD implementation, a defined SAR averaging method, uncertainty evaluation for each test configuration, and code verification against built-in benchmarks. In practice, this means a simulation result only counts if the numerical method, the post-processing, and the software verification all follow the document.
FDTD method and implementation
Clause 5 requires the SAR calculation to use the FDTD method in Cartesian coordinates with a three-dimensional mesh. The electric field and magnetic field are placed on a staggered grid, the field update is explicit and leapfrog-based, and the discretization uses second-order central differences.
That matters because the standard is not accepting any time-domain solver as equivalent. If a tool uses FDTD-like features or an extension such as finite integration technique (FIT) in time domain, the extra features must still be fully verified and at least as accurate as the FDTD method defined here.
SAR calculation and averaging
Clause 6 requires SAR to be calculated from voxel data and then averaged according to the document’s rules. For a single voxel, the standard uses the 12-field-components approach to interpolate the electric field to the voxel centre before calculating SAR with the voxel’s conductivity and mass density.
Clause 6 also defines two psSAR averaging algorithms:
- aligned psSAR for heterogeneous anatomical models, where the averaging cube follows the mesh axes
- homogeneous psSAR for homogeneous phantoms, where the averaging cube is aligned to the phantom surface
This distinction matters in practice because anatomical models and homogeneous phantoms are treated differently. The document says the chosen algorithm and the reference coordinate system must be reported, along with the cube corner coordinates and centre point.
Clause 6 also covers whole-body average SAR. It is calculated from the power dissipated in all body voxels divided by total body mass, so the whole-body result is not a separate measurement but a body-wide sum from the simulation mesh.
Power scaling and reporting
Clause 6 requires SAR results to be scaled to the desired input power or, for plane-wave excitation, to incident power density. The computed input power comes from the complex voltage and current at the feed-point, and the document also allows frequency-domain extraction from DFT or FFT records.
This matters because SAR values must be tied to a stated excitation condition. If device performance changes with temperature, electrical tolerance, or similar effects, the document says that variation has to be handled in the DUT validation or the uncertainty budget.
Clause 6.6 also requires reporting the peak value of the spatial-average SAR distribution and, where applicable, separate values for body and extremity tissues. The psSAR and whole-body average SAR must be reported with their computational uncertainty.
Uncertainty evaluation
Clause 7 requires uncertainty to be evaluated for each exposure configuration. The document separates uncertainty from the test setup, such as positioning, mesh resolution, open boundary conditions, power budget, convergence, and phantom or body dielectric properties, from uncertainty in the developed computational model of the DUT.
That matters because computational SAR is only useful if the reader can judge how much confidence to place in the result. The standard requires the uncertainty budget to combine the identified contributions, and it ties the model validation to representative measurements of the DUT.
Code verification
Clause 8 requires code verification at two levels: code accuracy and canonical benchmarks. The code accuracy tests examine free-space characteristics, planar dielectric boundaries, open boundary conditions, and SAR averaging. The canonical benchmarks include a generic dipole, a microstrip with open boundary conditions, SAR calculation for a SAM phantom and generic phone, and a setup for a system performance check.
In practice, this is the gate that separates a working implementation from an assumed one. Clause 8 says a code that passes the required tests can be labelled as compliant with IEC/IEEE 62704-1 Ed. 2.
What terms does IEC/IEEE 62704-1:2026 define?
IEC/IEEE 62704-1:2026 defines the main terms used in the calculations, averaging, and uncertainty work. These definitions are important because the standard uses them with a precise technical meaning.
- device under test (DUT) - the device or radiating structure being modelled and evaluated in the SAR simulation.
- finite-difference time-domain (FDTD) - a time-domain numerical method for solving Maxwell’s equations on a mesh.
- specific absorption rate (SAR) - the rate at which electromagnetic energy is absorbed per unit mass.
- peak spatial-average specific absorption rate (psSAR) - the maximum average SAR within a local region based on a defined averaging volume or mass.
- spatial-average specific absorption rate (sSAR) - the total power dissipated in the averaging volume divided by the target mass in that volume.
- whole-body average specific absorption rate - the total power dissipated in the whole body divided by body mass.
- voxel - a volume element in the mesh used to discretize the geometry.
- coverage factor (k) - the factor used to convert combined uncertainty into expanded uncertainty with a stated probability.
Who uses IEC/IEEE 62704-1:2026?
IEC/IEEE 62704-1:2026 is used by engineers and organizations that evaluate human exposure from wireless communications devices with computational methods. Typical users are simulation engineers, antenna designers, SAR specialists, compliance laboratories, and manufacturers that need validated SAR results.
It is also relevant to quality managers and technical buyers who need to check whether a SAR tool or study is traceable, repeatable, and backed by the required verification and uncertainty analysis. The standard is especially useful when SAR measurement is difficult or impractical and a computational approach is used instead.
What changed in IEC/IEEE 62704-1:2026 from the previous edition?
IEC/IEEE 62704-1:2026 cancels and replaces the first edition published in 2017. The foreword says this second edition is a technical revision.
The stated technical changes are:
- a SAR averaging algorithm was added for psSAR calculations on arbitrarily shaped homogeneous phantoms
- SHA-256 checksums were added for all supplemental files
- a brief informative discussion was added on the development stages of IEC and IEEE computational modelling standards for SAR evaluations
Which standards are used with IEC/IEEE 62704-1:2026?
IEC/IEEE 62704-1:2026 directly cites one normative reference and also names related documents used alongside it.
| Document | What it contributes |
|---|---|
| IEC/IEEE 62209-1528:2020 | Measurement procedure for SAR assessment, including human models, instrumentation, procedures, and the SAM phantom and testing positions |
| IEC/IEEE 62704-2 | A companion computational SAR standard cited as a way to add device-specific or application-specific definitions |
| IEC/IEEE 62704-3 | A companion computational SAR standard cited as a way to add device-specific or application-specific definitions |
| IEC/IEEE 62704-4 | Identified in the standards history as the FEM part of the IEC/IEEE 62704 series |
| ICNIRP guidelines | Example source for SAR limits outside this document |
| IEEE Std C95.1 | Example source for SAR limits outside this document |
What does the IEC/IEEE 62704-1:2026 document contain?
IEC/IEEE 62704-1:2026 contains the FDTD method definition, the required implementation features, and the SAR post-processing rules used to turn voxel data into psSAR and whole-body average SAR results. It also contains the uncertainty framework needed to support simulations with known uncertainty.
The document includes:
- formulas and procedures for voxel SAR, power scaling, and whole-body averaging
- two psSAR averaging algorithms, one aligned to the mesh and one aligned to a homogeneous phantom surface
- uncertainty tables and budgets for test setup and DUT model contributions
- verification procedures for free-space propagation, dielectric boundaries, open boundaries, and SAR averaging
- benchmark setups such as the generic dipole, microstrip, SAR Star, and SAM phantom/generic phone cases
- practical notes on sources, loads, S-parameters, power, efficiency, and inhomogeneous meshes
- background on tissue properties, anatomical modelling, and recommended phantoms
Annex B supplies benchmark geometry files, mesh lines, and output file formats for verification work. Annex C explains how the FDTD method is used in practice for SAR and related electromagnetic calculations, and Annex D gives the modelling background that supports realistic human-body simulations.
Relations
- Effective Date
- 05-Sep-2023
Buy Documents
IEC/IEEE 62704-1:2026 - Determining the peak spatial-average specific absorption rate (SAR) in the human body from wireless communications devices, 100 kHz to 10 GHz - Part 1: General requirements for using the finite-difference time-domain (FDTD) method for SAR calculations
Get Certified
Connect with accredited certification bodies for this standard

ANCE
Mexican certification and testing association.
BSMI (Bureau of Standards, Metrology and Inspection)
Taiwan's standards and inspection authority.

Intertek Slovenia
Intertek testing, inspection, and certification services in Slovenia.
Sponsored listings
Frequently Asked Questions
IEC/IEEE 62704-1:2026 is a standard published by the International Electrotechnical Commission (IEC). Its full title is "Determining the peak spatial-average specific absorption rate (SAR) in the human body from wireless communications devices, 100 kHz to 10 GHz - Part 1: General requirements for using the finite-difference time-domain (FDTD) method for SAR calculations". This standard covers: IEC/IEEE 62704-1:2026 defines the methodology for the application of the finite-difference time domain (FDTD) technique when used for determining the peak spatial-average specific absorption rate (psSAR) in the human body exposed to wireless communication devices with known uncertainty. It defines methods to validate the computational model of the device under test (DUT) and to assess its uncertainty when used in SAR simulations. Moreover, it defines procedures to determine the psSAR in a cubical volume and to verify the correct implementation of the FDTD simulation software. The applicable frequency range is 100 kHz to 10 GHz. This document does not recommend specific SAR limits; these are found elsewhere, for example, in the guidelines published by the International Commission on Non-Ionizing Radiation Protection (ICNIRP) or in IEEE Std C95.1. This second edition cancels and replaces the first edition published in 2017. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) specification of a SAR averaging algorithm for the calculation of the psSAR from computational results for arbitrarily shaped homogeneous phantoms, which avoids inconsistencies with respect to the algorithm for SAR averaging algorithm specified in IEC/IEEE 62209-1528 for SAR measurement methods; b) provision of SHA-256 checksums for all supplemental files; c) addition of brief informative discussion on developmental stages of IEC and IEEE computational modelling standards for SAR evaluations.
IEC/IEEE 62704-1:2026 defines the methodology for the application of the finite-difference time domain (FDTD) technique when used for determining the peak spatial-average specific absorption rate (psSAR) in the human body exposed to wireless communication devices with known uncertainty. It defines methods to validate the computational model of the device under test (DUT) and to assess its uncertainty when used in SAR simulations. Moreover, it defines procedures to determine the psSAR in a cubical volume and to verify the correct implementation of the FDTD simulation software. The applicable frequency range is 100 kHz to 10 GHz. This document does not recommend specific SAR limits; these are found elsewhere, for example, in the guidelines published by the International Commission on Non-Ionizing Radiation Protection (ICNIRP) or in IEEE Std C95.1. This second edition cancels and replaces the first edition published in 2017. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) specification of a SAR averaging algorithm for the calculation of the psSAR from computational results for arbitrarily shaped homogeneous phantoms, which avoids inconsistencies with respect to the algorithm for SAR averaging algorithm specified in IEC/IEEE 62209-1528 for SAR measurement methods; b) provision of SHA-256 checksums for all supplemental files; c) addition of brief informative discussion on developmental stages of IEC and IEEE computational modelling standards for SAR evaluations.
IEC/IEEE 62704-1:2026 is classified under the following ICS (International Classification for Standards) categories: 17.220.20 - Measurement of electrical and magnetic quantities; 33.060.20 - Receiving and transmitting equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
IEC/IEEE 62704-1:2026 has the following relationships with other standards: It is inter standard links to IEC/IEEE 62704-1:2017. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
IEC/IEEE 62704-1:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 1 Scope
- 2 Normative references
- 4 MHz to 10 GHz)
- 3 Terms and definitions
- 3.1 source
- 3.2 feed-point
- 3.3 antenna feed-point impedance
- 3.4 background material
- 3.5 benchmark
- 3.6 body
- 3.7 conductivity
- 3.8 conservative exposure
- 3.9 coverage factor
- 3.10 far-field
- 3.11 mesh
- 3.12 near-field
- 3.13 perfect electric conductor
- 3.14 relative permittivity
- 3.15 permeability
- 3.16 reactive field
- 3.17 root-mean-square
- 3.18 root-sum-square
- 3.19 scattering
- 3.20 specific absorption rate
- 3.20.1 peak spatial-average specific absorption rate
- 3.20.2 spatial-average specific absorption rate
- 3.20.3 whole-body average specific absorption rate
- 3.21 update coefficient
- 3.22 voxel
- 4 Abbreviated terms
- 5 Finite-difference time-domain method
- 5.1 Definition
- 5.2 Requirements on the implementation of the FDTD method
- 6 SAR calculation and averaging
- 6.1 General
- 6.2 Calculation of SAR in FDTD voxels
- 6.3 Calculation of the psSAR
- 6.4 Calculation of the whole-body average SAR
- 6.5 Power scaling
- 6.6 Reporting peak spatial-average SAR and whole-body average SAR
- 6.7 Referencing psSAR and whole-body average SAR
- 7 SAR simulation uncertainty
- 7.1 Considerations for the uncertainty evaluation
- 7.2 Uncertainty of the test setup with respect to simulation parameters
- 7.3 Uncertainty and validation of the developed computational model of the DUT
- 7.4 Uncertainty budget
- 8 Code verification
- 8.1 General
- 8.2 Code accuracy
- 2 TE TM
- 8.3 Canonical benchmarks
- Annex A
- Annex B
- Annex C
- Annex D
- Annex E
- Bibliography
INTRODUCTION Computational techniques have reached a level of maturity which allows their use in specific absorption rate (SAR) assessment of wireless communication devices. Some wireless communication devices are used in situations where SAR measurement is extremely complex or not possible at all. The benefits of consensus standards to users and regulators include standardized and accepted protocols, anatomically correct body models, verification and validation techniques, benchmark data, reporting format and means for estimating the computational uncertainty in order to produce valid, accurate, repeatable, and reproducible data.
1 Scope
This part of IEC/IEEE 62704 defines the methodology for the application of the finite-difference time domain (FDTD) technique when used for determining the peak spatial-average specific absorption rate (psSAR) in the human body exposed to wireless communication devices with known uncertainty. It defines methods to validate the computational model of the device under test (DUT) and to assess its uncertainty when used in SAR simulations. Moreover, it defines procedures to determine the psSAR in a cubical volume and to verify the correct implementation of the FDTD simulation software. The applicable frequency range is 100 kHz to 10 GHz. This document does not recommend specific SAR limits; these are found elsewhere, for example, in the guidelines published by the International Commission on Non-Ionizing
Radiation Protection (ICNIRP) [1] or in IEEE Std C95.1 [2].
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
IEC/IEEE 62209-1528:2020, Measurement procedure for the assessment of specific absorption rate of human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices - Human models, instrumentation, and procedures (Frequency range of
4 MHz to 10 GHz)
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply. IEC, ISO and IEEE maintain terminological databases for use in standardization at the following addresses:
- IEC Electropedia: available at https://www.electropedia.org/
- ISO Online browsing platform: available at https://www.iso.org/obp
- IEEE Dictionary Online: available at https://dictionary.ieee.org
3.1 source
one or more edges of the mesh to which an associated signal feeds electric or magnetic energy Note 1 to entry: The amplitude of the signal is proportional to an arbitrary function of time.
3.2 feed-point
<antenna> part of the radiating structure where the radio-frequency currents start to support the electromagnetic fields that carry energy away from the antenna Note 1 to entry: Often the feed-point of the antenna is not accessible because of mechanical support requirements; in this case a connection point is available to inject radio-frequency energy into the antenna. Normally, the connection point is a simple connector or a waveguide flange. If not collocated, the connection and the feed-point of an antenna are interconnected by one or more sections of transmission line. By measuring the antenna impedance at the connection point, if the electrical characteristics of the transmission lines between the connection and the feed-point are known, it is possible to calculate the antenna feed-point impedance.
Numbers in square brackets refer to the bibliography.
3.3 antenna feed-point impedance
terminal or driving-point impedance ratio of complex voltage to complex current at the terminals of a transmitting antenna, or the ratio of the open-circuit voltage to the short-circuit current at the terminals of a receiving antenna
3.4 background material
material or tissue which is not considered for the averaging volume Note 1 to entry: Most typically, a background material will be any lossless material, such as the free-space or air surrounding the anatomical model. It also includes air enclosures or other lumina inside the body and tissues that have been excluded from averaging, for example, by user selection.
3.5 benchmark
simulation test specifically defined to verify simulation results based on comparison with a reference
3.6 body
geometrical distribution of the dielectric properties and the mass densities of all live body tissues including body fluids Note 1 to entry: The contents of body lumina or foreign matter, such as medical implants or jewellery, are not considered as part of the body.
3.7 conductivity
σ ratio of the magnitude of the conduction-current density in a medium to the electric field strength Note 1 to entry: Conductivity is expressed in units of siemens per metre (S/m).
3.8 conservative exposure
estimate of the peak spatial-average SAR and whole-body average SAR as defined in this document that is representative of what is expected to occur in the body of a significant majority of the population during normal operating conditions of wireless communication devices [SOURCE: IEC/IEEE 62209-1528:2020, modified – In the definition, whole-body average SAR has been added and explicit mention of uncertainties has been deleted.]
3.9 coverage factor
k factor that is used to obtain the expanded uncertainty from the combined uncertainty with a known probability of containing the true value of the measurand
3.10 far-field
region of the field of an antenna where the angular field distribution is essentially independent of the distance from the antenna Note 1 to entry: In this region, the field has predominantly plane wave characteristics, i.e. the electric field strength and magnetic field strength distributions are locally uniform in planes transverse to the direction of propagation. Note 2 to entry: For larger antennas especially, the far-field is also referred to as the Fraunhofer region.
3.11 mesh
discrete representation of the simulation model as a set of voxels in a regular three-dimensional Cartesian arrangement Note 1 to entry: In the scientific literature, the mesh is often referred to as "grid".
3.12 near-field
region in the field of an antenna, located near the antenna, in which the electric and magnetic fields do not have substantial plane-wave characteristics, but vary considerably from point to point Note 1 to entry: The term near-field is often vaguely defined and has different meanings for large and small antennas. The near-field is further subdivided into the reactive near-field, which is closest to the antenna and contains most or nearly all of the stored energy associated with the field of the antenna, and the radiating near-field. If the antenna has a maximum overall dimension that is not large compared to the wavelength, the radiating near-field possibly does not exist. For antennas large in terms of wavelength, the radiating near-field is sometimes referred to as the Fresnel region.
Note 2 to entry: For most antennas, the outer boundary of the reactive near-field is commonly taken to exist at a distance of λ/2π from the antenna surface.
3.13 perfect electric conductor
PEC material with infinite electrical conductivity which does not dissipate any energy
3.14 relative permittivity
ε r ratio of the complex permittivity to the permittivity of free space Note 1 to entry: The complex relative permittivity, ε = ε /ε , of an isotropic, linear, lossy dielectric medium is from r 0 Formula (1) ε ′′ σ r
ε = ε − jε = ε − j = ε 1− j = ε (1− j tanδ ) (1) r r r r r r
ωε ε ′ 0 r
where −12 ε is the free space permittivity (8,854 × 10 F/m);
ε is the relative permittivity or dielectric constant; r σ is the conductivity in siemens per metre (S/m); tanδ is the loss tangent.
jωt Note 2 to entry: For the purposes of this document, the convention e is used to describe time-varying electric fields.
3.15 permeability
µ ratio of the magnetic flux density to the magnetic field strength at a point Note 1 to entry: The permeability is expressed in units of henry per metre (H/m).
3.16 reactive field
electric and magnetic fields surrounding an antenna or other electromagnetic devices that result in storage rather than propagation of electromagnetic energy
3.17 root-mean-square
RMS <of a periodic function> positive square root of the mean value of the square of the function taken over a given period Note 1 to entry: For a periodic function y of t, the positive square root of y is from Formula (2):
(2) Y = y (t) dt RMS ∫
T where Y is the RMS value of y; RMS t is any value of time; T is the period.
3.18 root-sum-square
RSS positive square root of the sum of the squares of the elements of a set of numbers
3.19 scattering
process that causes waves incident on discontinuities or boundaries of media to be changed in direction, phase, or polarization
3.20 specific absorption rate
SAR time derivative (rate) of the incremental energy (dW) absorbed by (dissipated in) an incremental mass (dm) contained in a volume element (dV) of a given density (ρ) d dW d dW
SAR =
dt dm dt ρ dV
Note 1 to entry: SAR is expressed in units of watts per kilogram (W/kg) or equivalently milliwatts per gram (mW/g). Note 2 to entry: SAR can be related to the electric field at a point by
σ | E | SAR = 2ρ
where σ is the conductivity of the tissue in siemens per metre (S/m);
ρ is the mass density of the tissue in kilograms per cubic metre (kg/m ); E is the peak electric field vector in volts per metre (V/m).
Note 3 to entry: SAR can be related to the increase in temperature at a point by c∆T SAR ≈ ∆t t=0
where ∆T is the change in temperature in degree Celsius (°C); ∆t is the duration of exposure in seconds (s); c is the specific heat capacity in joules per kilogram and degree Celsius (J/kg °C).
This assumes that measurements are made under "ideal" circumstances, i.e. no heat loss by thermal diffusion, heat radiation, or thermoregulation (blood flow, sweating, etc.). Moreover, this expression is valid only in the initial, linear regime of the temperature-time curve. If this linear regime does not exist, the provided expression is not used.
3.20.1 peak spatial-average specific absorption rate
psSAR maximum average SAR within a local region based on a specific averaging volume or mass Note 1 to entry: SAR is expressed in watts per kilogram (W/kg) or equivalently milliwatts per gram (mW/g). Note 2 to entry: The specific averaging volume or mass can be, for example, any 1 g or 10 g of tissue in the shape of a cube.
Note 3 to entry: For the purpose of this document, when the averaging mass is not specified the peak spatialaverage SAR refers to both 1 g and 10 g quantities. In several tables in this document, the peak spatial-average SAR for 1 g and 10 g is simply denoted as 1 g and 10 g SAR.
3.20.2 spatial-average specific absorption rate
sSAR ratio of the total power dissipated in the averaging volume to the target mass of tissue within that volume Note 1 to entry: SAR is expressed in watts per kilogram (W/kg) or equivalently milliwatts per gram (mW/g). Note 2 to entry: The specific averaging volume or mass can be, for example, any 1 g or 10 g of tissue in the shape of a cube.
3.20.3 whole-body average specific absorption rate
quantity calculated as the power dissipated in the entire body divided by its mass Note 1 to entry: SAR is expressed in watts per kilogram (W/kg) or equivalently milliwatts per gram (mW/g).
3.21 update coefficient
numerical value assigned to the voxel edges and faces calculated from the dielectric and magnetic properties of the voxels that surround it Note 1 to entry: The update coefficient is used to advance the numerical values of the electric and magnetic fields in the mesh in time based on their present and past values.
3.22 voxel
volume element of the geometry to be discretized in the mesh Note 1 to entry: In FDTD simulations for SAR calculation, the voxel is assigned the permittivity, permeability, the electric and magnetic conductivities and the mass density of the geometry. The electric-field components of the mesh are located in a tangential direction on the centres of the voxel edges, and the magnetic-field components of the mesh are located in a normal direction on the centres of the voxel faces. Each edge and face is assigned an update coefficient.
4 Abbreviated terms
ASCII American Standard Code for Information Interchange CFL Courant–Friedrichs–Lewy CT computed tomography DFT discrete Fourier transformation DUT device under test FDTD finite-difference time-domain FFT fast Fourier transformation FIT finite integration technique MIMO multiple input multiple output MRI magnetic resonance imaging PEC perfect electric conductor PMC perfect magnetic conductor PML perfectly matched layers psSAR peak spatial-average specific absorption rate RMS root mean square RSS root sum square SAM specific anthropomorphic mannequin SAR specific absorption rate sSAR spatial-average specific absorption rate
5 Finite-difference time-domain method
5.1 Definition
This document applies the finite-difference time-domain (FDTD) method to calculate SAR. FDTD is extensively used for bio-electromagnetic calculations, dating back to 1975 for the analysis of microwave heating of a human eye (Taflove and Brodwin [3]). Initially, the user shall determine the applicability of this document by determining whether the specific computational solution of Maxwell’s equations under consideration qualifies as an implementation of FDTD. This is not entirely straightforward since there have been many variations of the FDTD method developed over the time since Kane Yee’s original paper in 1966 [4], which may be taken as defining the fundamental FDTD method. Since then, the method has received significant attention in the scientific community. Numerous articles on the method and its extensions and applications have been published. Several textbooks on the FDTD method recapitulate its fundamentals, its further developments and its use ([5], [6]). Details on the application of the FDTD method are given in Annex C.
For the application of this document, the implementation of FDTD in the region where SAR is being calculated shall include the following characteristics.
- The electric field components are located on the edges of a Cartesian mesh composed of rectangular parallelepipeds.
- The magnetic field components are located on the edges of a Cartesian mesh composed of rectangular parallelepipeds which is offset from the electric field mesh by a half-voxel in each direction.
- The computational method is a finite-difference approximation of the Maxwell curl equations using central differences which are (at least globally) second-order accurate and may therefore include non-uniform meshes, which are also second-order accurate (Monk and Süli [7]).
- The computational method solves for both electric and magnetic fields by a fully explicit leapfrog time-stepping process.
- Gauss’s laws are implicitly enforced, the fields are divergence-free, and charge is conserved.
- The time-stepping algorithm is non-dissipative. Thus, there is no spurious decay of energy due to non-physical artefacts of the algorithm, and artificial dissipation is not required for stability.
NOTE Complex numerical dispersion in non-uniform meshes leads to spurious rise or decay of field amplitudes [7]. For the applications of the scope of this document, these effects are generally negligible. The impact of the numerical dispersion is assessed within the uncertainty evaluation (Clause 7).
5.2 Requirements on the implementation of the FDTD method
The basic requirements on the FDTD method are given in Annex A. For compliance with this document, the FDTD method shall be implemented according to these requirements. The method shall be implemented in Cartesian coordinates in three spatial dimensions. Moreover, SAR calculation and averaging shall be implemented according to Clause 6. The implementation of the FDTD method shall be verified according to Clause 8. Other methods or extensions, or both, of FDTD that satisfy the above definitions but provide additional features may be applied provided that these features have been fully verified according to the procedures in this document; code verification and benchmarks, etc. These alternative methods shall provide at least the same accuracy as the FDTD method defined above. The finite integration technique (FIT) in time domain can be regarded as an extension of FDTD in this context.
6 SAR calculation and averaging
6.1 General
Present RF exposure standards and guidelines specify time averaged whole-body average SAR limits and psSAR limits, neither of which shall be exceeded. The psSAR is usually averaged over a specified mass of tissue in a cubical volume, such as the trunk, the head, an extremity (e.g. limbs) or the pinnae. Averaging masses are specified in [1], [2] or other applicable guidelines.
The measurement standard that defines the SAM phantom and the testing positions is IEC/IEEE 62209-1528. The frequency range of IEC/IEEE 62209-1528 is from 4 MHz to 10 GHz, whereas the lower end of the frequency range of this document is 100 kHz. The assumptions on which the SAM phantom and the testing positions are based and which are outlined in IEC/IEEE 62209-1528 apply for SAR assessment regardless of the frequency limit. As the FDTD method does not have any intrinsic frequency limitations, the SAM phantom and the testing positions of IEC/IEEE 62209-1528 can be applied over the entire frequency range of this document.
Most computational studies involving the calculation of the psSAR in anatomical models of the human body with complex tissue distribution apply Cartesian meshes and align the cubical averaging volume with the mesh axes [8]. SAR measurement methods using homogeneous body phantoms filled with tissue simulant align the cubical averaging volume to the surface normal of the phantom at the location of the maximum SAR as described in IEC/IEEE 62209-1528:2020. As computational methods are applied to both anatomical models and homogeneous body phantoms, this document specifies two algorithms to calculate the psSAR.
- The algorithm specified in 6.3.1 applies to anatomical models with heterogeneous tissue distribution and calculates the psSAR in a cubical volume that is aligned with the mesh axes. It works on the original computational mesh and is referred to as “aligned”. Its principal application range is the computation of the psSAR in heterogeneous phantoms, such as anatomical models of the human body.
- The algorithm in 6.3.2 calculates the psSAR in a cubical volume that is aligned with the phantom surface. It is specified for phantoms with a constant dielectric over their entire volume disregarding the material of the phantom shell. It is based on the approach of the averaging algorithm specified in IEC/IEEE 62209-1528:2020 for measurement of the psSAR in homogeneous, liquid-filled phantoms. It is intended for the computational evaluation of the psSAR in computational models of such phantoms and removes the uncertainty associated with the differences between the SAR averaging volumes that are aligned with the mesh like the one specified in 6.3.1. This averaging algorithm is referred to as "homogeneous".
NOTE Deviations of the psSAR of more than 1,3 dB have been observed at 900 MHz, due to differences in the alignment with respect to the phantom surface of the averaging cubes calculated by the averaging algorithms of 6.3.1 and IEC/IEEE 62209-1528:2020 [9].
The remainder of Clause 6 is structured as follows.
...