Amendment 1 - Human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices - Human models, instrumentation, and procedures - Part 2: Procedure to determine the specific absorption rate (SAR) for wireless communication devices used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)

Amendement 1 - Exposition humaine aux champs radiofréquence produits par les dispositifs de communications sans fils tenus à la main ou portés près du corps - Modèles de corps humain, instrumentation et procédures - Partie 2: Procédure de détermination du débit d'absorption spécifique produit par les appareils de communications sans fil utilisés très près du corps humain (gamme de fréquences de 30 MHz à 6 GHz)

General Information

Status
Replaced
Publication Date
16-May-2019
Current Stage
DELPUB - Deleted Publication
Completion Date
29-Oct-2020
Ref Project

Relations

Buy Standard

Standard
IEC 62209-2:2010/AMD1:2019 - Amendment 1 - Human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices - Human models, instrumentation, and procedures - Part 2: Procedure to determine the specific absorption rate (SAR) for wireless communication devices used in close proximity to the human body (frequency range of 30 MHz to 6 GHz)
English and French language
8 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


IEC 62209-2 ®
Edition 1.0 2019-05
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
A MENDMENT 1
AM ENDEMENT 1
Human exposure to radio frequency fields from hand-held and body-mounted
wireless communication devices –
Human models, instrumentation, and procedures –
Part 2: Procedure to determine the specific absorption rate (SAR) for wireless
communication devices used in close proximity to the human body (frequency
range of 30 MHz to 6 GHz)
Exposition humaine aux champs radiofréquence produits par les dispositifs
de communications sans fils tenus à la main ou portés près du corps –
Modèles de corps humain, instrumentation et procédures –
Partie 2: Procédure de détermination du débit d'absorption spécifique produit
par les appareils de communications sans fil utilisés très près du corps humain
(gamme de fréquences de 30 MHz à 6 GHz)
IEC 62209-2:2010-03/AMD1:2019-05(en-fr)

All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.

Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni
utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et
les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.

IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé info@iec.ch
CH-1211 Geneva 20 www.iec.ch
Switzerland
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.

About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigendum or an amendment might have been published.

IEC publications search - webstore.iec.ch/advsearchform Electropedia - www.electropedia.org
The advanced search enables to find IEC publications by a The world's leading online dictionary on electrotechnology,
variety of criteria (reference number, text, technical containing more than 22 000 terminological entries in English
committee,…). It also gives information on projects, replaced and French, with equivalent terms in 16 additional languages.
and withdrawn publications. Also known as the International Electrotechnical Vocabulary

(IEV) online.
IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published IEC Glossary - std.iec.ch/glossary
details all new publications released. Available online and once 67 000 electrotechnical terminology entries in English and
a month by email. French extracted from the Terms and Definitions clause of IEC
publications issued since 2002. Some entries have been
IEC Customer Service Centre - webstore.iec.ch/csc collected from earlier publications of IEC TC 37, 77, 86 and
If you wish to give us your feedback on this publication or need CISPR.

further assistance, please contact the Customer Service

Centre: sales@iec.ch.
A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.

A propos des publications IEC
Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.

Recherche de publications IEC - Le premier dictionnaire d'électrotechnologie en ligne au monde,
webstore.iec.ch/advsearchform avec plus de 22 000 articles terminologiques en anglais et en
La recherche avancée permet de trouver des publications IEC français, ainsi que les termes équivalents dans 16 langues
en utilisant différents critères (numéro de référence, texte, additionnelles. Egalement appelé Vocabulaire
comité d’études,…). Elle donne aussi des informations sur les Electrotechnique International (IEV) en ligne.

projets et les publications remplacées ou retirées.
Glossaire IEC - std.iec.ch/glossary
IEC Just Published - webstore.iec.ch/justpublished 67 000 entrées terminologiques électrotechniques, en anglais
Restez informé sur les nouvelles publications IEC. Just et en français, extraites des articles Termes et Définitions des
Published détaille les nouvelles publications parues. publications IEC parues depuis 2002. Plus certaines entrées
Disponible en ligne et une fois par mois par email. antérieures extraites des publications des CE 37, 77, 86 et
CISPR de l'IEC.
Service Clients - webstore.iec.ch/csc
Si vous désirez nous donner des commentaires sur cette
publication ou si vous avez des questions contactez-nous:
sales@iec.ch.
Electropedia - www.electropedia.org

IEC 62209-2 ®
Edition 1.0 2019-05
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
A MENDMENT 1
AM ENDEMENT 1
Human exposure to radio frequency fields from hand-held and body-mounted

wireless communication devices –

Human models, instrumentation, and procedures –

Part 2: Procedure to determine the specific absorption rate (SAR) for wireless

communication devices used in close proximity to the human body (frequency

range of 30 MHz to 6 GHz)
Exposition humaine aux champs radiofréquence produits par les dispositifs

de communications sans fils tenus à la main ou portés près du corps –

Modèles de corps humain, instrumentation et procédures –

Partie 2: Procédure de détermination du débit d'absorption spécifique produit

par les appareils de communications sans fil utilisés très près du corps humain

(gamme de fréquences de 30 MHz à 6 GHz)

INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 33.050.10 ISBN 978-2-8322-6923-7

– 2 – IEC 62209-2:2010/AMD1:2019
© IEC 2019
FOREWORD
This amendment has been prepared by IEC technical committee 106: Methods for the
assessment of electric, magnetic and electromagnetic fields associated with human exposure.
The text of this amendment is based on the following documents:
FDIS Report on voting
106/484/FDIS 106/492/RVD
Full information on the voting for the approval of this amendment can be found in the report
on voting indicated in the above table.
The committee has decided that the contents of this amendment and the base publication will
remain unchanged until the stability date indicated on the IEC website under
"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
_____________
6.3 Measurement procedure
6.3.1 General procedure
Replace paragraph d) with the following:
d) Measure the three-dimensional SAR distribution at each of the local maxima locations
identified in step c) (zoom scan procedure).
For frequencies at or below 3 GHz, the following procedure shall be applied (see Table 8):
The horizontal grid step shall be 8 mm or less. The grid step in the vertical direction shall
be 5 mm or less if uniform spacing is used. If variable spacing is used in the vertical
direction, the maximum spacing between the two closest measured points to the phantom
shell (M1 and M2, see Figure 14) shall be 4 mm or less and the spacing between farther
points shall increase by a factor of 1,5 or less. The minimum size of the zoom scan
volume shall be 30 mm by 30 mm by 30 mm. For other parameters, see Table 8 and
Figure 14.
For frequencies above 3 GHz, the minimum size of the zoom scan volume may be
reduced to 22 mm by 22 mm by 22 mm. The horizontal grid step shall be (24/f [GHz]) mm
or less. If uniform spacing in the vertical direction is used, the grid step in the vertical
direction shall be (10/(f [GHz] − 1)) mm or less. If variable spacing is used in the vertical
direction, the maximum spacing between the two measured points closest to the phantom
shell shall be (12/f [GHz]) mm or less and the spacing between further points shall
increase by a factor of 1,5 or less. For other parameters, see Table 8 and Figure 14.
When the highest 1 g or 10 g cube is touching the boundary of a zoom-scan volume, the
entire zoom scan shall be repeated with the new centre located at the maximum psSAR
location indicated by the preceding zoom scan measurement.

© IEC 2019
If the zoom scan measured as defined above complies with both of the following criteria,
or if the peak spatial-average SAR is below 0,1 W/kg, no additional measurements are
needed:
1) the smallest horizontal distance from the local SAR peaks to all points 3 dB below the
SAR peak shall be larger than the horizontal grid steps in both x and y directions
(Δx, Δy). This shall be checked for the measured zoom scan plane conformal to the
phantom at the distance z . The minimum distance shall be recorded in the SAR test
M1
report;
2) the ratio of the SAR at the second measured point (M2) to the SAR at the closest
measured point (M1) at the x-y location of the measured maximum SAR value shall be
at least 30 % (see Figure 14). This ratio (in %) shall be recorded in the SAR test report.

NOTE M1 to M8 are example measurement points used for extrapolation to the surface. The maximum of the
angle α between the evaluation axis and the surface normal line is given in Table 2 and Table 8. The distance z
M1
is from the phantom shell to the first measurement point M1, and its maximum value is given in Table 2 and
Table 8. The distances Δz (i = 1, 2, 3, .) are the distances from measurement points M to M . For uniform grids,
i i i-1
Δz are equal. For graded grids, Δz > Δz . R = Δz /Δz is a ratio with a maximum value given in Table 8. The
i i+1 i z i+1 i
z direction corresponds to the vertical direction, the x direction is horizontal and the y direction is horizontal into
the page.
Figure 14 – Orientation of the probe with respect to the line normal
to the phantom surface, shown at two different locations
NOTE 1 The evaluation of the zoom scan is typically done by the post-processor by interpolation and
extrapolation and without reconstruction of the field. More focused induced SAR distributions (e.g., for
more localized sources such as capacitively coupled sources) require a more dense grid such that the
same integration and extrapolation algorithms can be used for the same assessment uncertainty.
NOTE 2 The minimum ratio of 30 % is derived from the plane wave penetration depth at 6 GHz.
If one or both of the above criteria are not met, the zoom scan measurement shall be
repeated using a finer resolution while keeping the other zoom scan parameters
compatible with Table 8. New horizontal and vertical grid steps shall be determined from
the measured SAR distribution so that the above criteria are met. Compliance with the
above two criteria shall be demonstrated for the new measured zoom scan. The size of the
higher resolution zoom scan and other parameters of Table 8 shall apply. The closest
point to the phantom shell shall be 2 mm or less for graded grids and the grading factor
shall be 1,5 or less.
Uncertainties due to field distortion between the media boundary and the dielectric
enclosure of the probe should also be minimized, which is achieved if the distance
between the phantom surface and physical tip of the probe is larger than the probe tip
diameter. Other methods may utilize correction procedures to compensate for boundary
effects that enable high precision measurements closer than half the probe tip
diameter [2],[61]. For all measurement points, the angle of the probe normal to the flat
phantom surface shall be less than 5°. If this cannot be achieved, an additional uncertainty
evaluation according to 7.2.2.6 is required.

– 4 – IEC 62209-2:2010/AMD1:2019
© IEC 2019
Table 8 – Zoom scan parameters
Parameter DUT transmit frequency being tested
f ≤ 3 GHz 3 GHz < f ≤ 6 GHz
a
Maximum distance between the closest 5 δ ln(2)/2
measured points and the phantom surface
(z in Figure 14 and Table 2, in mm)
M1
Maximum angle between the probe axis and the
5° 5°
flat phantom surface normal (α in Figure 14)
b,c
Maximum spacing between measured points in 8 24/f
the x- and y-directions (Δx and Δy, in mm)
For uniform grids: 5 10/(f − 1)
Maximum spacing between measured points in
the direction normal to the phantom shell
(Δz in Figure 14, in mm)
For graded grids: 4 12/f
Maximum spacing between the two closest
measured points in the direction normal to the
phantom shell (Δz in Figure 14, in mm)
For graded grids: 1,5 1,5
Maximum incremental increase in the spacing
between measured points in the direction normal
to the phantom shell (R = Δz /Δz in Figure 14)
z 2 1
Minimum edge length of the zoom scan volume 30 22
in the x- and y-directions (L in 7.2.5.3, in mm)
z
Minimum edge length of the zoom scan volume 30 22
in the direction normal to the phantom shell
(L in 7.2.5.3, in mm)
h
Tolerance in the probe angle 1° 1°
a
δ is the penetration depth for a plane-wave incident normally on a planar half-space.
b
This is the maximum spacing allowed, which may not work for all circumstances.
c
f is the frequency in GHz.
_____________
– 6
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.