Remote alcohol monitoring devices - Test methods and performance requirements - Part 1: Instruments for assessment programmes

The purpose of this new standard is to specify test methods and performance requirements for remotely monitored breath alcohol testing devices. It covers remote alcohol monitoring devices intended to be used by participants in programmes designed to monitor abstinence or restricted alcohol consumption.

Geräte zur Fernüberwachung von Alkoholkonsum - Prüfverfahren und Anforderungen an das Betriebsverhalten - Teil 1: Geräte in Beurteilungsprogrammen

Dispositifs de contrôle à distance de l'alcoolémie - Méthodes d'essai et exigences de performance - Partie 1: Instruments pour les programmes d'évaluation

Le présent document spécifie les exigences de performance, ainsi que les méthodes d'essai relatives aux dispositifs de contrôle à distance de l'alcoolémie qui utilisent des technologies d'essai du taux d'alcoolémie. Il couvre les dispositifs de contrôle à distance de l'alcoolémie équipés d'un embout et destinés à être utilisés par les participants à des programmes de contrôle de la consommation d'alcool.
Le présent document s'adresse aux laboratoires d'essais et aux fabricants de dispositifs de contrôle à distance de l'alcoolémie. Il définit les exigences et les procédures d'essai relatives aux essais de type.
Plusieurs réglages de paramètres (tels que la concentration d'alcool, le volume d'air expiré et les unités de mesure) sont spécifiés dans le présent document pour les besoins des essais de type dans le cadre de la présente norme seulement. Toutefois, en raison de réglementations nationales ou selon les demandes des utilisateurs, il peut être nécessaire de définir les réglages de paramètres prescrits sur des valeurs différentes lorsque les dispositifs de contrôle à distance de l'alcoolémie sont utilisés.

Naprave za daljinsko nadzorovanje alkohola - Preskusne metode in zahtevane lastnosti - 1. del: Instrumenti za ocenjevalne programe

Namen tega novega standarda je določanje preskusnih metod in zahtevanih lastnosti za naprave za daljinsko nadzorovanje alkohola v izdihanem zraku. Zajema naprave za daljinsko nadzorovanje alkohola, ki so namenjene udeležencem programov, zasnovanim za nadzorovanje abstinence ali omejenega uživanja alkohola.

General Information

Status
Published
Public Enquiry End Date
31-Jan-2019
Publication Date
24-Nov-2019
Technical Committee
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
20-Nov-2019
Due Date
25-Jan-2020
Completion Date
25-Nov-2019

Buy Standard

Standard
EN 50980-1:2020
English language
39 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day
Draft
prEN 50980-1:2019
English language
38 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 50980-1:2020
01-januar-2020
Naprave za daljinsko nadzorovanje alkohola - Preskusne metode in zahtevane
lastnosti - 1. del: Instrumenti za ocenjevalne programe
Remote alcohol monitoring devices - Test methods and performance requirements - Part
1: Instruments for assessment programmes
Geräte zur Fernüberwachung von Alkoholkonsum - Prüfverfahren und Anforderungen an
das Betriebsverhalten - Teil 1: Geräte in Beurteilungsprogrammen
Dispositifs de contrôle à distance de l'alcoolémie - Méthodes d'essai et exigences de
performance - Partie 1: Instruments pour les programmes d'évaluation
Ta slovenski standard je istoveten z: EN 50980-1:2019
ICS:
13.200 Preprečevanje nesreč in Accident and disaster control
katastrof
SIST EN 50980-1:2020 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
SIST EN 50980-1:2020

---------------------- Page: 2 ----------------------
SIST EN 50980-1:2020

EUROPEAN STANDARD EN 50980-1

NORME EUROPÉENNE

EUROPÄISCHE NORM
November 2019
ICS 71.040.40
English Version
Remote alcohol monitoring devices - Test methods and
performance requirements - Part 1: Instruments for assessment
programmes
Dispositifs de contrôle à distance de l'alcoolémie - Geräte zur Fernüberwachung von Alkoholkonsum -
Méthodes d'essai et exigences de performance - Partie 1: Prüfverfahren und Anforderungen an das Betriebsverhalten
Instruments pour les programmes d'évaluation - Teil 1: Geräte in Beurteilungsprogrammen
This European Standard was approved by CENELEC on 2019-09-09. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.



European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2019 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
 Ref. No. EN 50980-1:2019 E

---------------------- Page: 3 ----------------------
SIST EN 50980-1:2020
EN 50980-1:2019 (E)
Contents
European foreword . 5
Introduction . 6
1 Scope . 7
2 Normative references. 7
3 Terms and definitions . 8
4 General requirements .11
4.1 Objectives .11
4.2 Mouthpiece .11
4.3 Breath alcohol concentration limit .11
4.4 Breath test request .12
4.5 User interface .12
4.6 Retry breath sample .12
4.7 Repeat test request .12
4.8 Data memory records .12
4.9 Communication integrity .13
4.10 Tampering and circumvention .13
4.11 Service notification .13
4.12 Programme notification .13
4.13 Battery status .13
4.14 Calibration interval .13
4.15 Combination with other systems .14
5 General test methods.14
5.1 Samples .14
5.2 Preparation of remote alcohol monitoring devices before testing .14
5.3 Functional status classification .14
5.3.1 General .14
5.3.2 Class A .14
5.3.3 Class B .14
5.3.4 Class C.15
5.3.5 Class D.15
5.3.6 Class E .15
5.4 Sequence of tests .15
5.4.1 Remote Alcohol Monitoring Device .15
5.4.2 Optional elements .15
5.4.3 Communication testing .15
5.5 Normal conditions for tests .15
5.6 Functional test .16
6 Electrical tests .17
6.1 Excess supply voltage .17
6.2 Short-circuit .17
6.3 Reversed polarity .17
6.4 Electrical disturbances .17
6.4.1 Supply lines .17
6.4.2 Lines other than supply lines .18
6.5 Electrostatic discharge .18
6.6 Electromagnetic compatibility .18
2

---------------------- Page: 4 ----------------------
SIST EN 50980-1:2020
EN 50980-1:2019 (E)
7 Calibration curve .18
8 Durability tests .19
8.1 General .19
8.2 Temperature cycles .19
8.3 Condensed water .20
8.4 Vibrations .20
8.4.1 General .20
8.4.2 Random vibrations .20
8.4.3 Vibration at fixed frequency .20
8.5 Drop test.21
8.6 Mechanical shock .21
9 Environmental tests .21
9.1 General .21
9.2 Temperature .22
9.2.1 General .22
9.2.2 Operation within the specified temperature range .22
9.2.3 Operation outside the specified temperature range .22
9.3 Temperature and humidity .22
9.4 Start-up time .22
9.5 Battery life .23
9.6 Pressure .23
9.7 Protection by enclosure .23
10 Breath sample .23
10.1 Volume .23
10.2 Flow .24
10.3 Exhalation time .24
10.4 Response time .24
11 Analytical specificity .24
11.1 Test gases .24
11.2 Cigarette smoke .25
12 Circumvention and tampering .25
12.1 General .25
12.2 Pressurized air .25
12.3 Provision of the sample with a mouthpiece attached .25
12.4 Provision of the sample without a mouthpiece attached .26
12.5 Obstruction of the mouthpiece .26
12.6 Filter .26
12.7 Condensation .27
12.8 Water .27
12.9 Low power behaviour .28
12.10 Disconnection of components .29
13 Documentation and identity verification.29
13.1 Documentation .29
13.1.1 Image capture .29
13.1.2 Characteristics of the image .29
13.1.3 Environmental lighting conditions .31
13.2 Image processing .31
13.2.1 General .31
13.2.2 Face detection .31
3

---------------------- Page: 5 ----------------------
SIST EN 50980-1:2020
EN 50980-1:2019 (E)
13.2.3 Missing face on image .31
13.3 Identification .31
14 Timer and intervals .31
14.1 Test schedule .31
14.1.1 User initiated test .31
14.1.2 Scheduled test .31
14.1.3 Test on demand .31
14.2 Service reminder .31
14.3 Calibration interval .32
15 Long term behaviour.32
16 Instructions .32
16.1 Instructions for use .32
16.2 Instructions for servicing the remote alcohol monitoring device .34
17 Test report .34
18 Labelling and marking .34
Annex A (normative)  Description of events .35
Annex B (informative)  Performance testing .38
Bibliography .39

4

---------------------- Page: 6 ----------------------
SIST EN 50980-1:2020
EN 50980-1:2019 (E)
European foreword
This document (EN 50980-1:2019) has been prepared by CLC/BTTF 116-2 “Alcohol Interlocks”.
The following dates are fixed:
• latest date by which this document has (dop) 2020-09-09
to be implemented at national level by
publication of an identical national
standard or by endorsement
• latest date by which the national (dow) 2022-09-09
standards conflicting with this document
have to be withdrawn
5

---------------------- Page: 7 ----------------------
SIST EN 50980-1:2020
EN 50980-1:2019 (E)
Introduction
The purpose of remote alcohol monitoring devices is to monitor the compliance of persons subject to
mandatory or voluntary programmes focused on monitoring alcohol consumption.
The purpose of EN 50980-1 is to specify essential performance requirements and to provide the
respective test methods for available technologies. The technology of remote alcohol monitoring
devices continues to evolve, and further innovations can be expected. These could be considered in
new parts or revisions of this European Standard.
6

---------------------- Page: 8 ----------------------
SIST EN 50980-1:2020
EN 50980-1:2019 (E)
1 Scope
This document specifies performance requirements and test methods for remote alcohol monitoring
devices that use breath alcohol testing technology. It covers remote alcohol monitoring devices having
a mouthpiece and which are intended to be used by participants in programmes designed to monitor
alcohol consumption.
This document is directed at test laboratories and manufacturers of remote alcohol monitoring devices.
It defines requirements and test procedures for type testing.
Several parameter settings (such as alcohol concentration, breath volume or units of measurement)
are specified in this document for the purpose of type testing according to this standard only. However,
it may be necessary due to national regulations or depending on user requests to set the values of the
prescribed parameter settings differently when the remote alcohol monitoring devices are in use.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
EN 50436-6, Alcohol interlocks — Test methods and performance requirements — Part 6: Data
security
EN 60068-2-1:2007, Environmental testing — Part 2-1: Tests — Test A: Cold (IEC 60068-2-1:2007)
EN 60068-2-2:2007, Environmental testing — Part 2-2: Tests — Test B: Dry heat (IEC 60068-2-
2:2007)
EN 60068-2-6, Environmental testing — Part 2-6: Tests — Test Fc: Vibration (sinusoidal) (IEC 60068-
2-6)
EN 60068-2-27, Environmental testing — Part 2-27: Tests — Test Ea and guidance: Shock (IEC
60068-2-27)
EN 60068-2-30:2005, Environmental testing — Part 2-30: Tests — Test Db: Damp heat, cyclic (12 h +
12 h cycle) (IEC 60068-2-30:2005)
EN 60068-2-31, Environmental testing — Part 2-31: Tests — Test Ec: Rough handling shocks,
primarily for equipment-type specimens (IEC 60068-2-31)
EN 60068-2-64, Environmental testing — Part 2-64: Tests — Test Fh: Vibration, broadband random
and guidance (IEC 60068-2-64)
EN 60068-2-78, Environmental testing — Part 2-78: Tests — Test Cab: Damp heat, steady state (IEC
60068-2-78)
EN 60529, Degrees of protection provided by enclosures (IP Code) (IEC 60529)
ISO 7637-2, Road vehicles — Electrical disturbances from conduction and coupling — Part 2:
Electrical transient conduction along supply lines only
ISO 7637-3, Road vehicles — Electrical disturbances from conduction and coupling — Part 3:
Electrical transient transmission by capacitive and inductive coupling via lines other than supply lines
ISO 10605, Road vehicles — Test methods for electrical disturbances from electrostatic discharge
7

---------------------- Page: 9 ----------------------
SIST EN 50980-1:2020
EN 50980-1:2019 (E)
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— IEC Electropedia: available at http://www.electropedia.org/
— ISO Online browsing platform: available at http://www.iso.org/obp
3.1
remote alcohol monitoring device
device or system that measures the breath alcohol concentration of a person in an unsupervised
environment and which provides means for identification of the person providing the breath sample
and means of data communication in a secure format
Note 1 to entry: Devices can be fixed, mobile or portable and may be a system consisting of more than one
element.
3.2
breath alcohol concentration
mass concentration of ethanol, expressed in mg/l (milligram ethanol per litre breath), in a breath
sample delivered into a remote alcohol monitoring device
3.3
breath alcohol concentration limit
set value of the breath alcohol concentration at or above which the remote alcohol monitoring device
will register a failed breath test
3.4
mouthpiece
part that is fitted to the remote alcohol monitoring device through which the subject under test provides
the breath specimen
Note 1 to entry: A mouthpiece is used to prevent the breath sample from being mixed with ambient air and
thus diluting the alcohol concentration.
3.5
breath sample
breath sample taken under forced expiration
3.6
accepted breath sample
breath sample fulfilling set requirements for volume, flow, exhalation time and other human breath
sample characteristics
Note 1 to entry: The acceptance of a breath sample is independent from its alcohol concentration.
3.7
breath test
provision of an accepted breath sample to a remote alcohol monitoring device
3.8
passed breath test
breath test whereby the subject under test has provided an accepted breath sample with a breath
alcohol concentration below the breath alcohol concentration limit
8

---------------------- Page: 10 ----------------------
SIST EN 50980-1:2020
EN 50980-1:2019 (E)
3.9
failed breath test
breath test whereby the subject under test has provided an accepted breath sample with a breath
alcohol concentration equal to or above the breath alcohol concentration limit
3.10
test time interval
period of time following a test request within which the subject under test shall provide an accepted
breath sample
3.11
missed breath test
absence of the provision of an accepted breath sample within the test time interval
3.12
ready for test
condition whereby the remote alcohol monitoring device is able to accept a breath test
3.13
error state
condition whereby the remote alcohol monitoring device detects an internal system non conformity
that may cause the device not to function according to specification and normal function is suspended
3.14
breath test request
signal to alert the subject that a breath test is required
3.14.1
repeat test request
additional breath test request following a failed breath test
3.14.2
subject initiated test request
breath test request initiated by the subject
3.14.3
test on demand request
breath test request initiated by an external authority through communication with the remote alcohol
monitoring device
3.14.4
scheduled test request
breath test request initiated according to a fixed or random time schedule
3.15
retry breath sample
additional breath sample request following a not accepted breath sample
3.16
programme authority
responsible authority for monitoring the compliance of the subject using a remote alcohol monitoring
device
3.17
compliance
act of conformance of the subject to the requirements of the programme
9

---------------------- Page: 11 ----------------------
SIST EN 50980-1:2020
EN 50980-1:2019 (E)
3.18
non-compliance
act of non-conformance of the subject to the requirements of the programme
3.19
circumvention
manipulation of a breath test, subject identification and/or data communication for the purpose of
achieving programme compliance
3.20
tampering
unauthorised change to or interference with the remote alcohol monitoring device or its function
3.21
supply voltage
voltage as obtained from the electrical power source (mains or battery) for the operation of the remote
alcohol monitoring device
3.22
nominal operation voltage
supply voltage as specified by the manufacturer for the normal operation of the device
3.23
calibration interval
set time period after calibration, verification or adjustment following which the remote alcohol
monitoring device will request the next calibration, verification or adjustment
3.24
calibration adjustment
set of operations carried out on a remote alcohol monitoring device so that it provides correct breath
alcohol concentration results within the accepted tolerances
3.25
service
service is the sum of actions that are necessary to maintain the proper function of the remote alcohol
monitoring device
3.26
verification test
test of the remote alcohol monitoring device with a gas sample of known alcohol concentration to
determine the proper function of the device and the accuracy of the alcohol sensor response within the
stability requirements
3.27
service notification
notification to alert the subject that a service or inspection of the remote alcohol monitoring device is
required
3.28
programme notification
notification to alert a subject of a special request from the programme authorities
3.29
data memory record
record of breath test results and other events with date, time and other parameters stored in the
memory of the remote alcohol monitoring device
10

---------------------- Page: 12 ----------------------
SIST EN 50980-1:2020
EN 50980-1:2019 (E)
3.30
data communication
transmission of data to or from the remote alcohol monitoring device
3.31
manufacturer
person or organisation responsible for the design, construction and production of the remote alcohol
monitoring device
3.32
g
n
standard acceleration due to the earth gravitational field, which varies with altitude and geographical
latitude
Note 1 to entry: For the purposes of this standard, the value of gn is rounded to the nearest whole number, that
2
is 10 m/s .
3.33
operational state
state in which the remote alcohol device is powered and ready to operate and to communicate. It may
be in a low power consumption mode
4 General requirements
4.1 Objectives
The remote alcohol monitoring device shall monitor the compliance of persons participating in
mandatory or voluntary programmes focused on monitoring alcohol consumption. This shall be
achieved through the recurrent measurement of breath alcohol concentration as well as means for
subject identification and data communication.
4.2 Mouthpiece
The remote alcohol monitoring device shall have an exchangeable mouthpiece for the delivery of the
breath sample that shall prevent
...

SLOVENSKI STANDARD
oSIST prEN 50980-1:2019
01-januar-2019
Naprave za daljinsko nadzorovanje alkohola - Preskusne metode in zahtevane
lastnosti - 1. del: Instrumenti za ocenjevalne programe
Remote alcohol monitoring devices - Test methods and performance requirements - Part
1: Instruments for assessment programmes
Geräte zur Fernüberwachung von Alkoholkonsum - Prüfverfahren und Anforderungen an
das Betriebsverhalten - Teil 1: Geräte in Beurteilungsprogrammen
Ta slovenski standard je istoveten z: prEN 50980-1:2018
ICS:
13.200 3UHSUHþHYDQMHQHVUHþLQ Accident and disaster control
NDWDVWURI
oSIST prEN 50980-1:2019 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
oSIST prEN 50980-1:2019

---------------------- Page: 2 ----------------------
oSIST prEN 50980-1:2019

EUROPEAN STANDARD DRAFT
prEN 50980-1
NORME EUROPÉENNE

EUROPÄISCHE NORM

November 2018
ICS
English Version
Remote alcohol monitoring devices - Test methods and
performance requirements - Part 1: Instruments for assessment
programmes
To be completed Geräte zur Fernüberwachung von Alkoholkonsum -
Prüfverfahren und Anforderungen an das Betriebsverhalten
- Teil 1: Geräte zur Bewertung von Programmen
This draft European Standard is submitted to CENELEC members for enquiry.
Deadline for CENELEC: 2019-02-08.

It has been drawn up by CLC/BTTF 116-2.

If this draft becomes a European Standard, CENELEC members are bound to comply with the CEN/CENELEC Internal Regulations which
stipulate the conditions for giving this European Standard the status of a national standard without any alteration.

This draft European Standard was established by CENELEC in three official versions (English, French, German).
A version in any other language made by translation under the responsibility of a CENELEC member into its own language and notified to
the CEN-CENELEC Management Centre has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden,
Switzerland, Turkey and the United Kingdom.

Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are aware and to
provide supporting documentation.

Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without notice and
shall not be referred to as a European Standard.



European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2018 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Project: 65289 Ref. No. prEN 50980-1:2018 E

---------------------- Page: 3 ----------------------
oSIST prEN 50980-1:2019
prEN 50980-1:2018 (E)
1 Contents
2 European foreword . 5
3 Introduction . 6
4 1 Scope . 7
5 2 Normative references. 7
6 3 Terms and definitions . 8
7 General requirements .11
4
8 4.1 Objectives .11
9 4.2 Mouthpiece .11
10 4.3 Breath alcohol concentration limit .11
11 4.4 Breath test request .12
12 4.5 User interface .12
13 4.6 Test request .12
14 4.7 Retry breath sample .12
15 4.8 Repeat test request .12
16 4.9 Data memory records .12
17 4.10 Communication integrity .13
18 4.11 Tampering and circumvention .13
19 4.12 Service notification .13
20 4.13 Programme notification .13
21 4.14 Battery status .13
22 4.15 Calibration interval .13
23 4.16 Combination with other systems .13
24 5 General test methods.14
25 5.1 Samples .14
26 5.2 Preparation of remote alcohol monitoring devices before testing .14
27 5.3 Functional status classification .14
28 5.3.1 General .14
29 5.3.2 Class A .14
30 5.3.3 Class B .14
31 5.3.4 Class C.14
32 5.3.5 Class D.14
33 5.3.6 Class E .15
34 5.4 Sequence of tests .15
35 5.4.1 Remote Alcohol Monitoring Device .15
36 5.4.2 Optional elements .15
37 5.4.3 Communication Testing .15
38 5.5 Normal conditions for tests .15
39 5.6 Functional test .16
40 6 Electrical tests .16
41 6.1 Excess supply voltage .16
42 6.2 Short-circuit .17
43 6.3 Reversed polarity .17
44 6.4 Electrical disturbances .17
45 6.4.1 Supply lines .17
46 6.4.2 Lines other than supply lines .17
47 6.5 Electrostatic discharge .18
2

---------------------- Page: 4 ----------------------
oSIST prEN 50980-1:2019
prEN 50980-1:2018 (E)
48 6.6 Electromagnetic compatibility .18
49 7 Calibration curve .18
50 8 Durability tests .18
51 8.1 General .18
52 8.2 Temperature cycles .18
53 8.3 Condensed water .20
54 8.4 Vibrations .20
55 8.4.1 General .20
56 8.4.2 Random vibrations .20
57 8.4.3 Vibration at fixed frequency .20
58 8.5 Drop test.21
59 8.6 Mechanical Shock .21
60 9 Environmental tests .21
61 9.1 General .21
62 9.2 Temperature .21
63 9.2.1 General .21
64 9.2.2 Operation within the specified temperature range .22
65 9.2.3 Operation outside the specified temperature range .22
66 9.3 Temperature and humidity .22
67 9.4 Start-up time .22
68 9.5 Battery life .22
69 9.6 Pressure .23
70 9.7 Protection by enclosure .23
71 10 Breath sample .23
72 10.1 Volume .23
73 10.2 Flow .23
74 10.3 Exhalation time .24
75 10.4 Response time .24
76 11 Analytical specificity .24
77 11.1 Test gases .24
78 11.2 Cigarette smoke .24
79 12 Circumvention and tampering .25
80 12.1 General .25
81 12.2 Pressurized air .25
82 12.3 Provision of the sample with a mouthpiece attached .25
83 12.4 Provision of the sample without a mouthpiece attached .26
84 12.5 Obstruction of the mouthpiece .26
85 12.6 Filter .26
86 12.7 Condensation .26
87 12.8 Water .27
88 12.9 Low power behaviour .28
89 12.10 Disconnection of components .28
90 13 Documentation and identity verification.28
91 13.1 Documentation .28
92 13.1.1 Image capture .28
93 13.1.2 Characteristics of the image .28
94 13.1.3 Environmental lighting conditions .30
95 13.2 Image processing .30
96 13.2.1 General .30
3

---------------------- Page: 5 ----------------------
oSIST prEN 50980-1:2019
prEN 50980-1:2018 (E)
97 13.2.2 Face detection .30
98 13.2.3 Missing Face on image .30
99 13.3 Identification .30
100 14 Timer and Intervals .30
101 14.1 Test schedule .30
102 14.1.1 User initiated test .30
103 14.1.2 Scheduled test .30
104 14.1.3 Test on demand .30
105 14.2 Service reminder .31
106 14.3 Calibration interval .31
107 15 Long term behaviour.31
108 16 Instructions .31
109 16.1 Instructions for use .31
110 16.2 Instructions for servicing the remote alcohol monitoring device .33
111 17 Test report .33
112 18 Labelling and marking .33
113 Annex A (normative) Description of events.34
114 Annex B (informative) Performance testing .37
115 Bibliography .38
116
117
4

---------------------- Page: 6 ----------------------
oSIST prEN 50980-1:2019
prEN 50980-1:2018 (E)
118 European foreword
119 This document (prEN 50980-1:2018) has been prepared by CLC/BTTF 116-2 “Alcohol Interlocks”.
120 This document is currently submitted to the enquiry.
121 The following dates are proposed:
- Latest date by which the existence of (doa) dor + 6 months
this document has to be announced at
national level.
- latest date by which this document has (dop) dor + 12 months
to be implemented at national level by
publication of an identical national
standard or by endorsement
- latest date by which the national (dow) dor + 36 months
standards conflicting with this (to be confirmed or
document have to be withdrawn modified when voting)
5

---------------------- Page: 7 ----------------------
oSIST prEN 50980-1:2019
prEN 50980-1:2018 (E)
122 Introduction
123 The purpose of remote alcohol monitoring devices is to monitor the compliance of persons subject to
124 mandatory or voluntary programmes focused on monitoring alcohol consumption.
125 The purpose of prEN 50980-1 is to specify essential performance requirements and to provide the
126 respective test methods for available technologies. The technology of remote alcohol monitoring
127 devices continues to evolve, and further innovations can be expected. These could be considered in
128 new parts or revisions of this European Standard.
6

---------------------- Page: 8 ----------------------
oSIST prEN 50980-1:2019
prEN 50980-1:2018 (E)
129 1 Scope
130 This document specifies performance requirements and test methods for remote alcohol monitoring
131 devices that use breath alcohol testing technology. It covers remote alcohol monitoring devices having
132 a mouthpiece and which are intended to be used by participants in programmes designed to monitor
133 alcohol consumption.
134 This document is directed at test laboratories and manufacturers of remote alcohol monitoring devices.
135 It defines requirements and test procedures for type testing.
136 Several parameter settings (such as alcohol concentration, breath volume or units of measurement)
137 are specified in this document for the purpose of type testing according to this standard only. However,
138 it may be necessary due to national regulations or depending on user requests to set the values of the
139 prescribed parameter settings differently when the remote alcohol monitoring devices are in use.
140 2 Normative references
141 The following documents are referred to in the text in such a way that some or all of their content
142 constitutes requirements of this document. For dated references, only the edition cited applies. For
143 undated references, the latest edition of the referenced document (including any amendments) applies.
144 EN 50436-6, Alcohol interlocks - Test methods and performance requirements - Part 6: Data security
145 EN 60068-2-1, Environmental testing - Part 2-1: Tests - Test A: Cold (IEC 60068-2-1)
146 EN 60068-2-2, Environmental testing - Part 2-2: Tests - Test B: Dry heat (IEC 60068-2-2)
147 EN 60068-2-6, Environmental testing - Part 2-6: Tests - Test Fc: Vibration (sinusoidal) (IEC 60068-2-
148 6)
149 EN 60068-2-27, Environmental testing - Part 2-27: Tests - Test Ea and guidance: Shock (IEC 60068-
150 2-27)
151 EN 60068-2-31, Environmental testing - Part 2-31: Tests - Test Ec: Rough handling shocks, primarily
152 for equipment-type specimens (IEC 60068-2-31)
153 EN 60068-2-64, Environmental testing - Part 2-64: Tests - Test Fh: Vibration, broadband random and
154 guidance (IEC 60068-2-64)
155 EN 60068-2-78, Environmental testing - Part 2-78: Tests - Test Cab: Damp heat, steady state (IEC
156 60068-2-78)
157 EN 60529, Degrees of protection provided by enclosures (IP Code) (IEC 60529)
158 ISO 7637-2, Road vehicles — Electrical disturbances from conduction and coupling — Part 2:
159 Electrical transient conduction along supply lines only
160 ISO 7637-3, Road vehicles — Electrical disturbances from conduction and coupling — Part 3:
161 Electrical transient transmission by capacitive and inductive coupling via lines other than supply lines
162 ISO 10605, Road vehicles — Test methods for electrical disturbances from electrostatic discharge
7

---------------------- Page: 9 ----------------------
oSIST prEN 50980-1:2019
prEN 50980-1:2018 (E)
163 3 Terms and definitions
164 For the purposes of this document, the following terms and definitions apply.
165 ISO and IEC maintain terminological databases for use in standardization at the following addresses:
166 • IEC Electropedia: available at http://www.electropedia.org/
167 • ISO Online browsing platform: available at http://www.iso.org/obp
168 3.1
169 remote alcohol monitoring device
170 device or system that measures the breath alcohol concentration of a person in an unsupervised
171 environment and which provides means for identification of the person providing the breath sample
172 and means of data communication in a secure format
173 Note 1 to entry: Devices can be fixed, mobile or portable and may be a system consisting of more than one
174 element.
175 3.2
176 breath alcohol concentration
177 mass concentration of ethanol, expressed in mg/l (milligram ethanol per litre breath), in a breath
178 sample delivered into a remote alcohol monitoring device
179 3.3
180 breath alcohol concentration limit
181 set value of the breath alcohol concentration at or above which the remote alcohol monitoring device
182 will register a failed breath test
183 3.4
184 mouthpiece
185 part that is fitted to the remote alcohol monitoring device through which the subject under test provides
186 the breath specimen
187 Note 1 to entry: A mouthpiece is used to prevent the breath sample from being mixed with ambient air and
188 thus diluting the alcohol concentration.
189 3.5
190 breath sample
191 breath sample taken under forced expiration
192 3.6
193 accepted breath sample
194 breath sample fulfilling set requirements for volume, flow, exhalation time and other human breath
195 sample characteristics
196 Note 1 to entry: The acceptance of a breath sample is independent from its alcohol concentration.
197 3.7
198 breath test
199 provision of an accepted breath sample to a remote alcohol monitoring device
200 3.8
201 passed breath test
202 breath test whereby the subject under test has provided an accepted breath sample with a breath
203 alcohol concentration below the breath alcohol concentration limit
204 3.9
205 failed breath test
206 breath test whereby the subject under test has provided an accepted breath sample with a breath
207 alcohol concentration equal to or above the breath alcohol concentration limit
8

---------------------- Page: 10 ----------------------
oSIST prEN 50980-1:2019
prEN 50980-1:2018 (E)
208 3.10
209 test time interval
210 period of time following a test request within which the subject under test shall provide an accepted
211 breath sample
212 3.11
213 missed breath test
214 absence of the provision of an accepted breath sample within the test time interval
215 3.12
216 ready for test
217 condition whereby the remote alcohol monitoring device is able to accept a breath test
218 3.13
219 error state
220 condition whereby the remote alcohol monitoring device detects an internal system non conformity
221 that may cause the device not to function according to specification and normal function is suspended
222 3.14
223 breath test request
224 signal to alert the subject that a breath test is required
225 3.14.1
226 repeat test request
227 additional breath test request following a failed breath test
228 3.14.2
229 subject initiated test request
230 breath test request initiated by the subject
231 3.14.3
232 test on demand request
233 breath test request initiated by an external authority through communication with the remote alcohol
234 monitoring device
235 3.14.4
236 scheduled test request
237 breath test request initiated according to a fixed or random time schedule
238 3.15
239 retry breath sample
240 additional breath sample request following a not accepted breath sample
241 3.16
242 programme authority
243 responsible authority for monitoring the compliance of the subject using a remote alcohol monitoring
244 device
245 3.17
246 compliance
247 act of conformance of the subject to the requirements of the programme
248 3.18
249 non-compliance
250 act of non-conformance of the subject to the requirements of the programme
9

---------------------- Page: 11 ----------------------
oSIST prEN 50980-1:2019
prEN 50980-1:2018 (E)
251 3.19
252 circumvention
253 manipulation of a breath test, subject identification and/or data communication for the purpose of
254 achieving programme compliance
255 3.20
256 tampering
257 unauthorised change to or interference with the remote alcohol monitoring device or its function
258 3.21
259 supply voltage
260 voltage as obtained from the electrical power source (mains or battery) for the operation of the remote
261 alcohol monitoring device
262 3.22
263 nominal operation voltage U
n
264 supply voltage as specified by the manufacturer for the normal operation of the device
265 3.23
266 stability period
267 period of time over which the remote alcohol monitoring device fulfils the stability requirements for the
268 measurement of breath alcohol concentration
269 3.24
270 calibration interval
271 set time period after calibration, verification or adjustment following which the remote alcohol
272 monitoring device will request the next calibration, verification or adjustment
273 3.25
274 working day
275 working day means Monday to Friday inclusive, and does not include public holidays and weekends
276 Note 1 to entry: For the purposes of this standard, assume a 5 day working week.
277 3.26
278 calibration adjustment
279 set of operations carried out on a remote alcohol monitoring device so that it provides correct breath
280 alcohol concentration results within the accepted tolerances
281 3.27
282 service
283 service is the sum of actions that are necessary to maintain the proper function of the remote alcohol
284 monitoring device
285 3.28
286 verification test
287 test of the remote alcohol monitoring device with a gas sample of known alcohol concentration to
288 determine the proper function of the device and the
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.