SIGNIFICANCE AND USE
4.1 Harmful biological or particulate contaminants may enter the medical package through leaks. These leaks are frequently found at seals between package components of the same or dissimilar materials. Leaks may also result from a pinhole in the packaging material.  
4.2 It is the objective of this test method to visually observe the presence of channel defects by the leakage of dye through them.  
4.3 This dye penetrant procedure is applicable only to individual leaks in a package seal. The presence of a number of small leaks, as found in porous packaging material, which could be detected by other techniques, will not be indicated.  
4.4 There is no general agreement concerning the level of leakage that is likely to be deleterious to a particular package. However, since these tests are designed to detect leaks, components that exhibit any indication of leakage are normally rejected.  
4.5 These procedures are suitable to verify and locate leakage sites. They are not quantitative. No indication of leak size can be inferred from these tests. The methods are usually employed as a pass/fail test.  
4.6 The dye solution will wick through any porous material over time, but usually not within the maximum time suggested. If wicking does occur, it may be verified by observing the porous side of the subject seal area. The dye will have discolored the surface of the material. Refer to Appendix X1 for details on wicking and guidance on the observance of false positives.
SCOPE
1.1 This test method defines materials and procedures that will detect and locate a leak equal to or greater than a channel formed by a 50 µm (0.002 in.) wire in package edge seals formed between a transparent material and a porous sheet material. A dye penetrant solution is applied locally to the seal edge to be tested for leaks. After contact with the dye penetrant for a specified time, the package is visually inspected for dye penetration.  
1.2 Three dye application methods are covered in this test method: injection, edge dip, and eyedropper.  
1.3 These test methods are intended for use on packages with edge seals formed between a transparent material and a porous sheet material. The test methods are limited to porous materials which can retain the dye penetrant solution and prevent it from discoloring the seal area for a minimum of 5 seconds. Uncoated papers are especially susceptible to leakage and must be evaluated carefully for use with each test method.  
1.4 These test methods require that the dye penetrant solution have good contrast to the opaque packaging material.  
1.5 The values are stated in International System of Units (SI units) and English units. Either is to be regarded as standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    6 pages
    English language
  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 Contaminants may enter the package through leaks. Alternatively, product may be lost from the package through leaks. These leaks are frequently found at seals between package components of the same or dissimilar materials.  
4.2 Ingress or egress of gas or moisture through leaks in a package can also degrade sensitive contents.  
4.3 There is no general agreement concerning the level of leakage that is likely to be deleterious to a particular package. However, since these tests are designed to detect leakage, components that exhibit any indication of leakage may be rejected.  
4.4 These procedures are suitable for use to verify and locate leakage sites. They are not quantitative. No indication of leak size can be inferred from the test. Therefore, this method is employed as a go/no-go test.  
4.5 These tests are destructive. No package or component test samples exposed to dye penetration testing may be used for final product packaging.
SCOPE
1.1 Method A of this test method defines a procedure that will detect and locate a leak equal to or greater than a channel formed by a 50 µm [0.002 in.] wire in the edge seals of a nonporous package. A dye penetrant solution is applied locally to the seal edge to be tested for leaks. After contact with the dye penetrant for a minimum specified time, the package is visually inspected for dye penetration or, preferably, the seal edge is placed against an absorbent surface and the surface inspected for staining from the dye.  
1.2 Method B for this test method also defines a procedure that will detect and locate a leak equal to or greater than 10 µm [0.00039 in] diameter in a nonporous flat sheet. The flat sheet is placed on an absorbent surface and then a dye penetrant is spread across the surface of the sheet, preferably using a small roller to apply pressure on the sheet to ensure adequate contact between the absorbent surface and the bottom surface of the sample being tested. The flat sheet is carefully removed and the absorbent surface is inspected for staining from the dye.  
1.3 These test methods are used for both transparent and opaque nonporous surfaces.  
1.4 These test methods require that the dye penetrant have good contrast to the materials being tested and/or the absorbent surface.  
1.5 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    5 pages
    English language
  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
5.1 The vacuum, bubble test method, as described in Test Method D3078, and various other leak detection methods described elsewhere (Test Method D4991, Guide E432, Test Method E493, Test Method E498, Test Method E499, and Test Method E1603) have been successfully used widely in various industries and applications to determine that a given package is or is not a “leaker.” The sensitivity of any selected leak test method has to be considered to determine its applicability to a specific situation.  
5.2 The procedures presented in this test method allow the user to carry out package and seal integrity testing with sufficient sensitivity to quantify seals in the previously defined moderate to fine seal ranges.  
5.3 By employing seal-isolating leak testing fixtures, packages constructed of various materials can be tested in the full range of seal performance requirements. Design of these fixtures is beyond the scope of this method.  
5.4 These seal/package integrity test procedures can be utilized as:  
5.4.1 A design tool,  
5.4.2 For tooling qualification,  
5.4.3 Process setup,  
5.4.4 Process validation tool,  
5.4.5 Quality assurance monitoring, or  
5.4.6 Research and development.
SCOPE
1.1 This test method includes several procedures that can be used for the measurement of overall package and seal barrier performance of a variety of package types and package forms, as well as seal/closure types. The basic elements of this method include:  
1.1.1 Helium (employed as tracer gas),  
1.1.2 Helium leak detector (mass spectrometer), and  
1.1.3 Package/product-specific test fixtures.  
1.1.4 Most applications of helium leak detection are destructive, in that helium needs to be injected into the package after the package has been sealed. The injection site then needs to be sealed/patched externally, which often destroys its saleability. Alternatively, if helium can be incorporated into the headspace before sealing, the method can be non-destructive because all that needs to be accomplished is to simply detect for helium escaping the sealed package.  
1.2 Two procedures are described; however the supporting data in Section 14 only reflects Procedure B (Vacuum Mode). The alternative, Sniffer Mode, has proven to be a valuable procedure for many applications, but may have more variability due to exactly the manner that the operator conducts the test such as whether the package is squeezed, effect of multiple small leaks compared to fewer large leaks, background helium concentration, package permeability and speed at which the scan is conducted. Further testing to quantify this procedure’s variability is anticipated, but not included in this version.  
1.2.1 Procedure A: Sniffer Mode—the package is scanned externally for helium escaping into the atmosphere or fixture.  
1.2.2 Procedure B: Vacuum Mode—the helium containing package is placed in a closed fixture. After drawing a vacuum, helium escaping into the closed fixture (capture volume) is detected. Typically, the fixtures are custom made for the specific package under test.  
1.3 The sensitivity of the method can range from the detection of:  
1.3.1 Large leaks—10-2 Pa·m 3/s to 10-5 Pa·m3/s (10–1 cc/sec/atm to 10-4 cc/sec/atm).  
1.3.2 Moderate leaks—10-5 Pa·m 3/s to 10-7 Pa·m3/s (10-4 cc/sec/atm to 10-6 cc/sec/atm).  
1.3.3 Fine leaks—10-7 Pa·m 3/s to 10-9 Pa·m3/s (10-6 cc/sec/atm to 10-8 cc/sec/atm).  
1.3.4 Ultra-Fine leak—10-9 Pa·m 3/s to 10-11 Pa·m3/s (10-8 cc/sec/atm to 10-10 cc/sec/atm).
Note 1: Conversion from cc/sec/atm to Pa·m3/s is achieved by multiplying by 0.1.  
1.4 The terms large, moderate, fine and ultra-fine are relative terms only and do not imply the acceptability of any leak rate. The individual application dictates the level of integrity needed. For many packaging applications, only “large leaks” are considered unacceptable and the ability to detect smaller leaks is immaterial. All leak rates referred...

  • Standard
    7 pages
    English language
  • Standard
    7 pages
    English language

SIGNIFICANCE AND USE
5.1 Harmful biological or particulate contaminants may enter the package through imperfections such as pinholes or cracks in trays.  
5.2 After initial instrument set-up and calibration, the operations of individual tests and test results do not need operator interpretation.  
5.3 Leak test results that exceed the permissible threshold setting are indicated by audible or visual signal responses, or both, or by other means.  
5.4 This non-destructive test method may be performed in either laboratory or production environments and may be undertaken on either a 100 % or a statistical sampling basis. This test method, in single instrument use and current implementation, may not be fast enough to work on a production packaging line, but is well suited for statistical testing as well as package developmental design work.
SCOPE
1.1 This non-destructive test method detects pinhole leaks in trays, as small as 50 μm (0.002 in.) in diameter, or equivalently sized cracks, subject to trace gas concentration in the tray, tray design and manufacturing tolerances.  
1.2 The values stated in SI units are to be regarded as standard. The values given in parentheses after SI units are provided for information only and are not considered standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    4 pages
    English language

SIGNIFICANCE AND USE
5.1 Harmful biological or particulate contaminants may enter the package through incomplete seals or imperfections such as pinholes or cracks in the trays.  
5.2 After initial instrument set-up and calibration, the operations of individual tests and test results do not need operator interpretation. The non-destructive nature of the test may be important when testing high value added products.  
5.3 Leak test results that exceed the permissible threshold setting are indicated by audible or visual signal responses, or both, or by other means.  
5.4 This non-destructive test method may be performed in either laboratory or production environments. This testing may be undertaken on either a 100 % or a statistical sampling basis. This test method, in single instrument use and current implementation, may not be fast enough to work on a production packaging line, but is well suited for statistical testing as well as package developmental design work.
SCOPE
1.1 This non-destructive test method detects leaks in non-porous rigid thermoformed trays, as well as the seal between the porous lid and the tray. The test method detects channel leaks in packages as small as 100 μm (0.004 in.) diameter in the seal as well as 50 μm (0.002 in.) diameter pinholes, or equivalently sized cracks in the tray, subject to trace gas concentration in the package, package design and manufacturing tolerances.
Note 1: This test method does not claim to challenge the porous (breathable) lidding material. Any defects that may exist in the porous portion of the package will not be detected by this test method.  
1.2 The values stated in SI units are to be regarded as standard. The values given in parentheses after SI units are provided for information only and are not considered standard.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    5 pages
    English language

ABSTRACT
This test method covers the determination of gross leaks in flexible packaging containing a headspace gas by bubble emission. A vacuum chamber shall be any transparent container capable of withstanding approximately one atmosphere pressure differential, fitted with a vacuum-tight cover. A vacuum gage, an inlet tube from a source of vacuum, and an outlet tube to the atmosphere shall be connected to the chamber cover. Use an immersion fluid which does not degrade the package being tested. The test sample and test fluid shall be at equilibrium with normal room temperature. The procedures for testing are presented in details.
SCOPE
1.1 This test method covers the determination of gross leaks in flexible packaging containing a headspace gas. Test sensitivity is limited to 1 × 10−5 atm cm3/s (1 × 10−6 Pa m3/s) or even less sensitive as indicated in a recent interlaboratory test (reported in Section 12).  
1.2 Small leaks may not be detected by this procedure. Viscoelastic effects on the products, or entrapped air, become significant and prevent passage through small openings. Positive pressure inside the pouch after the vacuum is drawn may force the product to plug small leaks. The size of the leak that can be detected is dependent upon the products contained, the nature of the packaging material, and the test parameters selected.  
1.3 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    3 pages
    English language

SIGNIFICANCE AND USE
3.1 The oxygen content of a package’s headspace is an important determinant of the packaging protection afforded by barrier materials. The package under test is typically MAP (modified atmosphere packaging) packaged.  
3.2 Oxygen content is a key contributor to off-flavors and spoilage of various products, such as chemicals, food and pharmaceuticals.  
3.3 The method determines the oxygen in a closed package headspace. This ability has application in:  
3.3.1 Package Permeability Studies—The change of headspace composition over a known length of time allows the calculation of permeation. Since the headspace oxygen is measured as a percentage, the volume of the container’s headspace must be known to allow conversion into a quantity such as millilitres (ml) of oxygen. The use of this approach to measure permeation generally applies to empty package systems only as oxygen uptake or outgassing of contained products could affect results.  
3.3.2 Leak Detection—If the headspace contains more oxygen than expected or is increasing faster than expected, a leak can be suspected. A wide variety of techniques can be employed to verify that a leak is present and to identify its location. If necessary or of interest, a leak rate may be calculated with known headspace volume and measured oxygen concentration change over time.  
3.3.3 Efficacy of the MAP Packaging Process—If the headspace oxygen concentration is found to be higher than expected soon after packaging, the gas flushing process may not be working as well as expected. Various techniques can evaluate whether the MAP system is functioning properly.  
3.3.4 Storage Studies—As the method is non-destructive, the headspace can be monitored over time on individual samples to insure that results of storage studies such as shelf life testing are correctly interpreted.
SCOPE
1.1 This test method covers a procedure for determination of the oxygen concentration in the headspace within a sealed package without opening or compromising the integrity of the package.  
1.2 This test method requires that chemically coated components be placed on the inside surface of the package before closing.  
1.3 The package must be either transparent, translucent, or a transparent window must be affixed to the package surface without affecting the package’s integrity.  
1.4 As this test method determines the oxygen headspace over time, the oxygen permeability can easily be calculated as ingress per unit time as long as the volume of the container is known.  
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    4 pages
    English language

SIGNIFICANCE AND USE
5.1 These test methods provide a rapid, simple to apply method to detect small leaks in flexible package seals or walls at the leak rate level of greater than 1 × 10−4 sccs, thus providing a measure of package integrity. Porous barrier film packages made non-porous with an impermeable film forming coating may demonstrate lateral leakage through the barrier material. Verification of leakage differences from background leakage must be included in validation methods. The use of calibrated hole sizes or orifices may be appropriate to determine leakage sensitivity or barrier integrity for these materials.  
5.2 While theoretical leak rate sensitivity can be established by calculation, the test measurement is in pressure units and the measuring instrument must be calibrated, certified, and verified with these units.  
5.3 The pressure decay method of leak testing is a physical measure of package integrity. When testing medical packaging which must conform to ISO 11607–1: 2006 standards, it may necessary to verify the results of the pressure decay test method with other sterile package integrity test methods.  
5.4 Test Method A allows packages to be pressurized without restraint. In Test Method A the pouch, tray, or other type package will contain a volume of air defined by its mechanical configuration and its ability to resist internal pressure applied. This test method requires that the package reach a stable volume configuration (stop stretching) to make a measurement.  
5.5 Test Method B allows the use of rigid restraining plates against the walls of the package to limit its volume and stabilize the package volume.
SCOPE
1.1 These test methods cover the measurement of leaks in nonporous film, foil, or laminate flexible pouches and foil-sealed trays, which may be empty or enclose solid product. If product is enclosed, seals or surfaces cannot be in contact with water, oils, or other liquid.  
1.2 These test methods will detect leaks at a rate of 1 × 10−4 sccs (standard cubic centimetres per second) or greater, in flexible packages. The limitation of leak rate is dependent on package volume as tested.  
1.3 The following test methods are included:  
1.3.1 Test Method A—Pressure Decay Leak Test for Flexible Packages Without Restraining Plates  
1.3.2 Test Method B—Pressure Decay Leak Test for Flexible Packages With Restraining Plates  
1.4 These test methods are destructive in that they require entry into the package to supply an internal pressure of gas, typically air or nitrogen, although other gases may be used. The entry connection into the flexible package must be leak-tight.  
1.5 For porous packages, see 9.3.  
1.6 The values stated in SI units are to be regarded as standard. The values given in parentheses after SI units are provided for information only and are not considered standard.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    7 pages
    English language

SIGNIFICANCE AND USE
5.1 This method allows for the evaluation of seal quality by passing an ultrasound signal through the sealed area of a package or item. Poorly sealed areas will not transmit as much ultrasonic energy as properly sealed areas.  
5.2 This method relies on quantitative analysis of ultrasound signal strength, providing a non-subjective approach to assessing package seal quality and detecting defects.  
5.3 This technique has been used for inspecting a variety of materials including flexible pouch seals, rigid tray seals and other packaging components such as affixed valves. The precision and bias for any specific package and seal configuration needs to be individually determined and validated.  
5.4 The C-Scan approach is useful for laboratory applications or off-line seal inspection. The L-Scan approach can be used for on-line, real time inspection of seal quality. The sensitivity of either approach to detect a given defect size and level of severity needs to be individually determined.  
5.5 Sound waves propagate at different speeds through different materials generally moving faster through more dense materials. The acoustic impedance (expressed as g/cm2·μs) is the product of density (g/cm3) and velocity (cm/μs). Of particular importance is the extreme difference between the impedance of air and that of any solid material. Any gap or poorly bonded area can be readily detected.    
Material  
Velocity
(cm/μsec)  
Density
(g/cm3)  
Acoustic
Impedance
(g/cm2-μsec)  
Air (20°C, 1 bar)  
0.0344  
0.00119  
0.000041  
Water (20°C)  
0.148  
1.0  
0.148  
Polyethylene  
0.267  
1.1  
0.294  
Aluminum  
0.632  
2.7  
1.710
SCOPE
1.1 This standard method describes the technology and testing procedures that can be used to detect seal defects in the size range of 1 mm and characterize seal quality in a variety of packaging styles using airborne ultrasound technology.  
1.2 This test method does not purport to be the only method for measurement of seal quality.  
1.3 Heat seals and other package components can be tested in flexible, semi-rigid and rigid packages. Only the precision and bias for flexible package seals were evaluated in a recent ILS included in the method. The precision and bias for any specific package needs to be individually determined.  
1.4 On-line, real time inspection of seals can be considered particularly in the L-Scan mode.  
1.5 This method provides a non-destructive, quantitative, non-subjective approach to flexible package seal inspection.  
1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
5.1 Leaks in medical device, pharmaceutical, and food packages may result in the ingress of unwanted gases (most commonly oxygen), harmful microbiological, or particulate contaminants. Package leaks may appear as imperfections in the package components themselves or at the seal juncture between mated components. The ability to detect leaks is necessary to ensure consistency and integrity of packages.  
5.2 After initial set-up and calibration, individual test operation may be semi-automatic, automatic, or manual. The test method permits non-destructive detection of leaks not visibly detectable. The test method does not require the introduction of any extraneous materials or substances, such as dyes or gases. However, it is important to physically mask or block off any package porous barrier surface during the test to prevent rapid loss of chamber vacuum resulting primarily from gas migration through the porous surface. Leak detection is based solely on the ability to detect the change in pressure inside the test chamber resulting from gas or vapor egress from a package challenged with vacuum.  
5.3 This test is a useful research tool for optimizing package sealing parameters and for comparatively evaluating various packages and materials. This test method is also applicable to production settings as it is rapid, non-invasive, and non-destructive, making it useful for either 100 % on-line testing or to perform tests on a statistical sampling from the production operation.  
5.4 Leak test results that exceed the permissible limits for the vacuum decay test are indicated by audible or visual signal responses, or both.
SCOPE
1.1 Test Packages—Packages that can be nondestructively evaluated by this test method include:  
1.1.1 Rigid and semi-rigid non-lidded trays.  
1.1.2 Trays or cups sealed with porous barrier lidding material.  
1.1.3 Rigid, nonporous packages.  
1.1.4 Flexible, nonporous packages.  
1.2 Leaks Detected—This test method detects package leaks by measuring the rise in pressure (vacuum loss) in an enclosed evacuated test chamber containing the test package. Vacuum loss results from leakage of test package headspace gases and/or volatilization of test package liquid contents located in or near the leak. When testing for leaks that may be partially or completely plugged with the package’s liquid contents, the test chamber is evacuated to a pressure below the liquid’s vaporization pressure. All methods require a test chamber to contain the test package and a leak detection system designed with one or more pressure transducers. Test method sensitivities cited below were determined using specific product-package systems selected for the precision and bias studies summarized in Table 1. Table 1 also lists other examples of relevant product-package systems that can be tested for leakage by vacuum decay.    
1.2.1 Trays or Cups (Non-lidded) (Air Leakage)—Hole or crack defects in the wall of the tray/cup of at least 50 μm in diameter can be detected. Nonlidded trays were tested at a Target Vacuum of –4·E4 Pa (–400 mbar).  
1.2.2 Trays Sealed with Porous Barrier Lidding Material (Headspace Gas Leakage)—Hole or crack defects in the wall of the tray/cup of at least 100 μm in diameter can be detected. Channel defects in the seal area (made using wires of 125 μm in diameter) can be detected. Severe seal bonding defects in both continuous adhesive and dot matrix adhesive package systems can be detected. Slightly incomplete dot matrix adhesive bonding defects can also be detected. All porous barrier lidding material packages were tested at a Target Vacuum of –4·E4 Pa (–400 mbar). The sensitivity of the test for porous lidded packages is approximately E-2 Pa·m3·s-1 using a calibrated volumetric airflow meter.  
1.2.3 Rigid, Nonporous Packages (Headspace Gas Leakage)—Hole defects of at least 5 μm in diameter can be detected. Plastic bottles with screw caps were tested at a target vacuum...

  • Standard
    13 pages
    English language

SIGNIFICANCE AND USE
5.1 The internal pressurization test method provides a practical way to examine packages for gross leaks.  
5.2 This test method is extremely useful in a test laboratory environment where no common package material/size exists.  
5.3 This test method may apply to large or long packages that do not fit into any other package integrity test method apparatus.  
5.4 This test method may be used as a means to evaluate package integrity. Package integrity is crucial to consumer safety since heat sealed packages are designed to provide a contamination free or sterile environment, or both, to the product.  
5.5 This test method may be used to detect substrate holes and channels.
SCOPE
1.1 This test method covers the detection of gross leaks in packaging. Method sensitivity is down to 250 μm (0.010 in.) with an 81 % probability (see Section 11). This test method may be used for tray and pouch packages.  
1.2 The sensitivity of this test method has not been evaluated for use with porous materials other than spunbonded polyolefin or with nonporous packaging.  
1.3 This test method is destructive in that it requires entry into the package to supply an internal air pressure  
1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
5.1 Leaks in medical, pharmaceutical, or food product packages can affect product quality and consumer safety. Such leaks can arise from imperfections in package material or between mated components designed to seal the package. Defects can allow unwanted gas (e.g. oxygen or water vapor), particulates, liquids, or microbiological contaminants into or out of the package. Package defect detection can be a critical part of ensuring product quality and consumer safety. Use of a physical CCI test method for sterile products can be used to assure the stability of the package sterility property during transportation and product shelf life.  
5.2 Mass extraction is a useful non-destructive test method for testing a wide variety of packages. Package shape and dimensions that can be tested using mass extraction are essentially unlimited, as long as a vacuum test chamber can be designed and manufactured to accommodate the package.  
5.3 This method produces quantitative flow measurement results that are useful in comparing package sealing properties, different batches of product, material properties, and combinations of process parameters.  
5.4 Applications for mass extraction range from manually loaded and operated machines to automatic unattended work cells. This method can be applied for audit testing or 100% in-line testing.
Note 2: Leak test methods that rely on gas or vapor transport, such as mass extraction, are not able to detect defects if they become plugged by solid or nonvolatile matter. Plugging is possible by exposure to environmental contaminants. In some cases, the packaged product itself can clog defects. For example, leak paths may become blocked by suspended solids, gelatinous matter or dried-out solutions. Product clogging propensity is a function of the product formulation, defect size and geometry, and may be linked to product storage and handling conditions as well as the time allotted to defect exposure. An investigation into the impact of repeate...
SCOPE
1.1 This method provides a nondestructive means to detect holes (leaks) in a variety of non-porous rigid and semi-rigid packages.  
1.2 This test method detects package leaks by measuring the mass flow extracted from a package while the package is enclosed inside an evacuated test chamber. The test system is a closed system during the leakage measurement portion of the test cycle. The closed system includes a vacuum reservoir, Intelligent Molecular Flow Sensor (IMFS), and vacuum test chamber. Mass extracted from the test package into the vacuum test chamber flows to the vacuum reservoir through the IMFS to equalize the system. Mass flow rate from the vacuum chamber to the vacuum reservoir is measured by the IMFS. Based on the conservation of mass law, mass flow into the closed system is equal to the mass loss from the test package. The test system is capable of producing quantitative (variable data) or qualitative (pass/fail) results depending on the requirements.  
1.2.1 Headspace gas leakage defects equivalent to a 1µm diameter glass micropipette (sharp edge defect) can be detected at a 95% confidence level.  
1.2.2 Liquid leakage defects equivalent to a 1µm diameter glass micropipette can be detected at a 95% confidence level for glass vials and LDPE bottles. Liquid leakage defects equivalent to a 2 µm diameter glass micropipette can be detected for glass syringes.  
1.3 Units—The values stated in SI units are to be regarded as standard. Pressure units are expressed as Pa, mbar, or Torr.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization esta...

  • Standard
    12 pages
    English language

SIGNIFICANCE AND USE
5.1 Leaks in blister packs may affect product quality and such defects can arise from imperfections in the packaging material or bond between the sealed surfaces.  
5.2 This method of leak testing is a useful tool as it allows non-destructive and non-subjective leak testing of blister packs. It allows the operator to evaluate how different packaging materials and packaging machine conditions affect the integrity of the packaging. It can also provide indication of unwanted changes in the packaging conditions.  
5.3 This type of testing is typically used in pharmaceutical packaging production, during stability trials and for package research and development operations because of its non-destructive nature, cleanliness, and speed.
SCOPE
1.1 Test Packages—This test method can be applied to non-porous blister packs sealed with flexible films such as those used in pharmaceutical packaging. Such blister packs typically consist of thermoformed polymer or cold formed aluminum trays that contain a number of individual blister pockets into which tablets or capsules are placed. The trays are then sealed with a polymer, paper-backed or foil-based flexible laminate lidding material.  
1.2 Leaks Detected—This test method detects leaks in blister packs by measuring the deflection of the blister pack surface in response to an applied vacuum. This deflection of the blister pack surface results from the difference in pressure between the gas inside the blister pack and the applied vacuum. Air loss from within a blister pocket as a result of a leak will alter this pressure differential causing a measureable variation in blister pocket deflection. This test method requires that the blister packs are held in appropriate tooling inside a suitable test chamber.  
1.3 Test Results—Test results are reported qualitatively (pass/fail). Appropriate acceptance criteria for deflection, height, and collapse values are established by comparing non-leaking packs with those containing defects of a known size. Suitably sized defects in the laminate, tray material, and seal can be detected using this test method. The sensitivity of this test method depends upon a range of factors including blister pocket headspace, blister pocket size, lidding material type, lidding material thickness, lidding material tension, printing, surface texture, test conditions, and the values selected for the pass/fail acceptance criteria. The ability of the test to detect 15 µm, 50 µm, and catastrophic sized holes in four blister pack designs was demonstrated in a study.  
1.4 The values stated in SI units are to be regarded as standard and no other units of measurement are included in this test method.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
5.1 Seal attributes can be linked directly to a number of variables in process parameters, equipment, or material, as well as environmental (room temperature and relative humidity). Visual seal characteristics and defects can provide evidence of package integrity and production sealing problems.  
5.2 Visual seal defects often will be the first indication of heat sealing process variation. They also will indicate a lack of, or potential compromise to, package integrity after physical package performance testing.
SCOPE
1.1 This test method covers the determination of channels in the package seal down to a width of 75 μm [0.003 in.] with a 60–100 % probability (see Section 8).  
1.1.1 The ability to visually detect channel defects in package seals is highly dependent on the size of channel, the degree of contrast from sealed and unsealed areas, the amount and type of adhesive between the two package layers, reflecting light angle, types of material used, the use of magnification, and the inspector's level of training and experience.  
1.2 This test method is applicable to packages with at least one transparent side so that the seal area may be clearly viewed.  
1.3 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    4 pages
    English language
  • Standard
    4 pages
    English language

SIGNIFICANCE AND USE
5.1 Leaks in medical, pharmaceutical, or food product packages can affect product quality and consumer safety. Such leaks can arise from imperfections in package material or between mated components designed to seal the package. Defects can allow unwanted gas (e.g. oxygen or water vapor), particulates, liquids, or microbiological contaminants into or out of the package. Package defect detection can be a critical part of ensuring product quality and consumer safety. Use of a physical CCI test method for sterile products can be used to assure the stability of the package sterility property during transportation and product shelf life.  
5.2 Mass extraction is a useful non-destructive test method for testing a wide variety of packages. Package shape and dimensions that can be tested using mass extraction are essentially unlimited, as long as a vacuum test chamber can be designed and manufactured to accommodate the package.  
5.3 This method produces quantitative flow measurement results that are useful in comparing package sealing properties, different batches of product, material properties, and combinations of process parameters.  
5.4 Applications for mass extraction range from manually loaded and operated machines to automatic unattended work cells. This method can be applied for audit testing or 100% in-line testing.
Note 2: Leak test methods that rely on gas or vapor transport, such as mass extraction, are not able to detect defects if they become plugged by solid or nonvolatile matter. Plugging is possible by exposure to environmental contaminants. In some cases, the packaged product itself can clog defects. For example, leak paths may become blocked by suspended solids, gelatinous matter or dried-out solutions. Product clogging propensity is a function of the product formulation, defect size and geometry, and may be linked to product storage and handling conditions as well as the time allotted to defect exposure. An investigation into the impact of repeate...
SCOPE
1.1 This method provides a nondestructive means to detect holes (leaks) in a variety of non-porous rigid and semi-rigid packages.  
1.2 This test method detects package leaks by measuring the mass flow extracted from a package while the package is enclosed inside an evacuated test chamber. The test system is a closed system during the leakage measurement portion of the test cycle. The closed system includes a vacuum reservoir, Intelligent Molecular Flow Sensor (IMFS), and vacuum test chamber. Mass extracted from the test package into the vacuum test chamber flows to the vacuum reservoir through the IMFS to equalize the system. Mass flow rate from the vacuum chamber to the vacuum reservoir is measured by the IMFS. Based on the conservation of mass law, mass flow into the closed system is equal to the mass loss from the test package. The test system is capable of producing quantitative (variable data) or quantitative (pass/fail) results depending on the requirements.  
1.2.1 Headspace gas leakage defects equivalent to a 1µm diameter glass micropipette (sharp edge defect) can be detected at a 95% confidence level.  
1.2.2 Liquid leakage defects equivalent to a 1µm diameter glass micropipette can be detected at a 95% confidence level for glass vials and LDPE bottles. Liquid leakage defects equivalent to a 2µm diameter glass micropipette can be detected for glass syringes.  
1.3 Units—The values stated in SI units are to be regarded as standard. Pressure units are expressed as Pa, mbar, or Torr.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization esta...

  • Standard
    12 pages
    English language

SIGNIFICANCE AND USE
5.1 The vacuum, bubble test method, as described in Test Method D3078, and various other leak detection methods described elsewhere (Test Method D4991, Guide E432, Guide E479, Test Method E493, Test Method E498, Test Method E499, and Test Method E1603) have been successfully used widely in various industries and applications to determine that a given package is or is not a “leaker.” The sensitivity of any selected leak test method has to be considered to determine its applicability to a specific situation.  
5.2 The procedures presented in this test method allow the user to carry out package and seal integrity testing with sufficient sensitivity to quantify seals in the previously defined moderate to very fine seal ranges.  
5.3 By employing seal-isolating leak testing fixtures, packages constructed of various materials can be tested in the full range of seal performance requirements. Design of these fixtures is beyond the scope of this method.  
5.4 These seal/package integrity test procedures can be utilized as:  
5.4.1 A design tool,  
5.4.2 For tooling qualification,  
5.4.3 Process setup,  
5.4.4 Process validation tool,  
5.4.5 Quality assurance monitoring, or  
5.4.6 Research and development.
SCOPE
1.1 This test method includes several procedures that can be used for the measurement of overall package and seal barrier performance of a variety of package types and package forms, as well as seal/closure types. The basic elements of this method include:  
1.1.1 Helium (employed as tracer gas),  
1.1.2 Helium leak detector (mass spectrometer), and  
1.1.3 Package/product-specific test fixtures.  
1.1.4 Most applications of helium leak detection are destructive, in that helium needs to be injected into the package after the package has been sealed. The injection site then needs to be sealed/patched externally, which often destroys its saleability. Alternatively, if helium can be incorporated into the headspace before sealing, the method can be non-destructive because all that needs to be accomplished is to simply detect for helium escaping the sealed package.  
1.2 Two procedures are described; however the supporting data in Section 14 only reflects Procedure B (Vacuum Mode). The alternative, Sniffer Mode, has proven to be a valuable procedure for many applications, but may have more variability due to exactly the manner that the operator conducts the test such as whether the package is squeezed, effect of multiple small leaks compared to fewer large leaks, background helium concentration, package permeability and speed at which the scan is conducted. Further testing to quantify this procedure’s variability is anticipated, but not included in this version.  
1.2.1 Procedure A: Sniffer Mode—the package is scanned externally for helium escaping into the atmosphere or fixture.  
1.2.2 Procedure B: Vacuum Mode—the helium containing package is placed in a closed fixture. After drawing a vacuum, helium escaping into the closed fixture (capture volume) is detected. Typically, the fixtures are custom made for the specific package under test.  
1.3 The sensitivity of the method can range from the detection of:  
1.3.1 Large leaks—10-2 Pa·m 3/s to 10-5 Pa·m3/s (10–1 cc/sec/atm to 10-4 cc/sec/atm).  
1.3.2 Moderate leaks—10-5 Pa·m 3/s to 10-7 Pa·m3/s (10-4 cc/sec/atm to 10-6 cc/sec/atm).  
1.3.3 Fine leaks—10-7 Pa·m 3/s to 10-9 Pa·m3/s (10-6 cc/sec/atm to 10-8 cc/sec/atm).  
1.3.4 Ultra-Fine leak—10-9 Pa·m 3/s to 10-11 Pa·m3/s (10-8 cc/sec/atm to 10-10 cc/sec/atm).
Note 1: Conversion from cc/sec/atm to Pa·m3/s is achieved by multiplying by 0.1.  
1.4 The terms large, moderate, fine and ultra-fine are relative terms only and do not imply the acceptability of any leak rate. The individual application dictates the level of integrity needed. For many packaging applications, only “large leaks” are considered unacceptable and the ability to detect smaller leaks is immaterial. All le...

  • Standard
    7 pages
    English language
  • Standard
    7 pages
    English language

SIGNIFICANCE AND USE
4.1 Contaminants may enter the package through leaks. Alternatively, product may be lost from the package through leaks. These leaks are frequently found at seals between package components of the same or dissimilar materials.  
4.2 Ingress or egress of gas or moisture through leaks in a package can also degrade sensitive contents.  
4.3 There is no general agreement concerning the level of leakage that is likely to be deleterious to a particular package. However, since these tests are designed to detect leakage, components that exhibit any indication of leakage may be rejected.  
4.4 These procedures are suitable for use to verify and locate leakage sites. They are not quantitative. No indication of leak size can be inferred from the test. Therefore, this method is employed as a go/no-go test.  
4.5 These tests are destructive. No package or component test samples exposed to dye penetration testing may be used for final product packaging.
SCOPE
1.1 Method A of this test method defines a procedure that will detect and locate a leak equal to or greater than a channel formed by a 50 µm [0.002 in.] wire in the edge seals of a nonporous package. A dye penetrant solution is applied locally to the seal edge to be tested for leaks. After contact with the dye penetrant for a minimum specified time, the package is visually inspected for dye penetration or, preferably, the seal edge is placed against an absorbent surface and the surface inspected for staining from the dye.  
1.2 Method B for this test method also defines a procedure that will detect and locate a leak equal to or greater than 10 µm [0.00039 in] diameter in a nonporous flat sheet. The flat sheet is placed on an absorbent surface and then a dye penetrant is spread across the surface of the sheet, preferably using a small roller to apply pressure on the sheet to ensure adequate contact between the absorbent surface and the bottom surface of the sample being tested. The flat sheet is carefully removed and the absorbent surface is inspected for staining from the dye.  
1.3 These test methods are used for both transparent and opaque nonporous surfaces.  
1.4 These test methods require that the dye penetrant have good contrast to the materials being tested and/or the absorbent surface.  
1.5 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    5 pages
    English language
  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
4.1 Harmful biological or particulate contaminants may enter the medical package through leaks. These leaks are frequently found at seals between package components of the same or dissimilar materials. Leaks may also result from a pinhole in the packaging material.  
4.2 It is the objective of this test method to visually observe the presence of channel defects by the leakage of dye through them.  
4.3 This dye penetrant procedure is applicable only to individual leaks in a package seal. The presence of a number of small leaks, as found in porous packaging material, which could be detected by other techniques, will not be indicated.  
4.4 There is no general agreement concerning the level of leakage that is likely to be deleterious to a particular package. However, since these tests are designed to detect leaks, components that exhibit any indication of leakage are normally rejected.  
4.5 These procedures are suitable to verify and locate leakage sites. They are not quantitative. No indication of leak size can be inferred from these tests. The methods are usually employed as a pass/fail test.  
4.6 The dye solution will wick through any porous material over time, but usually not within the maximum time suggested. If wicking does occur, it may be verified by observing the porous side of the subject seal area. The dye will have discolored the surface of the material. Refer to Appendix X1 for details on wicking and guidance on the observance of false positives.
SCOPE
1.1 This test method defines materials and procedures that will detect and locate a leak equal to or greater than a channel formed by a 50 µm (0.002 in.) wire in package edge seals formed between a transparent material and a porous sheet material. A dye penetrant solution is applied locally to the seal edge to be tested for leaks. After contact with the dye penetrant for a specified time, the package is visually inspected for dye penetration.  
1.2 Three dye application methods are covered in this test method: injection, edge dip, and eyedropper.  
1.3 These test methods are intended for use on packages with edge seals formed between a transparent material and a porous sheet material. The test methods are limited to porous materials which can retain the dye penetrant solution and prevent it from discoloring the seal area for a minimum of 5 seconds. Uncoated papers are especially susceptible to leakage and must be evaluated carefully for use with each test method.  
1.4 These test methods require that the dye penetrant solution have good contrast to the opaque packaging material.  
1.5 The values are stated in International System of Units (SI units) and English units. Either is to be regarded as standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    6 pages
    English language
  • Standard
    6 pages
    English language
  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 Contaminants may enter the package through leaks. Alternatively, product may be lost from the package through leaks. These leaks are frequently found at seals between package components of the same or dissimilar materials.  
4.2 Ingress or egress of gas or moisture through leaks in a package can also degrade sensitive contents.  
4.3 There is no general agreement concerning the level of leakage that is likely to be deleterious to a particular package. However, since these tests are designed to detect leakage, components that exhibit any indication of leakage may be rejected.  
4.4 These procedures are suitable for use to verify and locate leakage sites. They are not quantitative. No indication of leak size can be inferred from the test. Therefore, this method is employed as a go/no-go test.  
4.5 This test is destructive. All package or component test samples exposed to dye penetration testing may not be used for final product packaging.
SCOPE
1.1 This test method defines a procedure that will detect and locate a leak equal to or greater than a channel formed by a 50 µm [0.002 in.] wire in the edge seals of a nonporous package. A dye penetrant solution is applied locally to the seal edge to be tested for leaks. After contact with the dye penetrant for a minimum specified time, the package is visually inspected for dye penetration or, preferably, the seal edge is placed against an absorbent surface and the surface inspected for staining from the dye.  
1.2 This test method is used for both transparent and opaque nonporous surfaces.  
1.3 This test method requires that the dye penetrant have good contrast to the materials being tested and/or the absorbent surface.  
1.4 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
5.1 Harmful biological or particulate contaminants may enter the package through incomplete seals or imperfections such as pinholes or cracks in the trays.  
5.2 After initial instrument set-up and calibration, the operations of individual tests and test results do not need operator interpretation. The non-destructive nature of the test may be important when testing high value added products.  
5.3 Leak test results that exceed the permissible threshold setting are indicated by audible or visual signal responses, or both, or by other means.  
5.4 This non-destructive test method may be performed in either laboratory or production environments. This testing may be undertaken on either a 100 % or a statistical sampling basis. This test method, in single instrument use and current implementation, may not be fast enough to work on a production packaging line, but is well suited for statistical testing as well as package developmental design work.
SCOPE
1.1 This non-destructive test method detects leaks in non-porous rigid thermoformed trays, as well as the seal between the porous lid and the tray. The test method detects channel leaks in packages as small as 100 μm (0.004 in.) diameter in the seal as well as 50 μm (0.002 in.) diameter pinholes, or equivalently sized cracks in the tray, subject to trace gas concentration in the package, package design and manufacturing tolerances.Note 1—This test method does not claim to challenge the porous (breathable) lidding material. Any defects that may exist in the porous portion of the package will not be detected by this test method.  
1.2 The values stated in SI units are to be regarded as standard units. Values in parentheses are for information only.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    5 pages
    English language
  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
3.1 The oxygen content of a package’s headspace is an important determinant of the packaging protection afforded by barrier materials. The package under test is typically MAP (modified atmosphere packaging) packaged.  
3.2 Oxygen content is a key contributor to off-flavors and spoilage of various products, such as chemicals, food and pharmaceuticals.  
3.3 The method determines the oxygen in a closed package headspace. This ability has application in:  
3.3.1 Package Permeability Studies—The change of headspace composition over a known length of time allows the calculation of permeation. Since the headspace oxygen is measured as a percentage, the volume of the container’s headspace must be known to allow conversion into a quantity such as millilitres (ml) of oxygen. The use of this approach to measure permeation generally applies to empty package systems only as oxygen uptake or outgassing of contained products could affect results.  
3.3.2 Leak Detection—If the headspace contains more oxygen than expected or is increasing faster than expected, a leak can be suspected. A wide variety of techniques can be employed to verify that a leak is present and to identify its location. If necessary or of interest, a leak rate may be calculated with known headspace volume and measured oxygen concentration change over time.  
3.3.3 Efficacy of the MAP Packaging Process—If the headspace oxygen concentration is found to be higher than expected soon after packaging, the gas flushing process may not be working as well as expected. Various techniques can evaluate whether the MAP system is functioning properly.  
3.3.4 Storage Studies—As the method is non-destructive, the headspace can be monitored over time on individual samples to insure that results of storage studies such as shelf life testing are correctly interpreted.
SCOPE
1.1 This test method covers a procedure for determination of the oxygen concentration in the headspace within a sealed package without opening or compromising the integrity of the package.  
1.2 This test method requires that chemically coated components be placed on the inside surface of the package before closing.  
1.3 The package must be either transparent, translucent, or a transparent window must be affixed to the package surface without affecting the package’s integrity.  
1.4 As this test method determines the oxygen headspace over time, the oxygen permeability can easily be calculated as ingress per unit time as long as the volume of the container is known.  
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    4 pages
    English language

SIGNIFICANCE AND USE
5.1 Seal attributes can be linked directly to a number of variables in process parameters, equipment, or material, as well as environmental (room temperature and relative humidity). Visual seal characteristics and defects can provide evidence of package integrity and production sealing problems.  
5.2 Visual seal defects often will be the first indication of heat sealing process variation. They also will indicate a lack of, or potential compromise to, package integrity after physical package performance testing.
SCOPE
1.1 This test method covers the determination of channels in the package seal down to a width of 75 μm [0.003 in.] with a 60–100 % probability (see Section 8).  
1.1.1 The ability to visually detect channel defects in package seals is highly dependent on the size of channel, the degree of contrast from sealed and unsealed areas, the amount and type of adhesive between the two package layers, reflecting light angle, types of material used, the use of magnification, and the inspector's level of training and experience.  
1.2 This test method is applicable to packages with at least one transparent side so that the seal area may be clearly viewed.  
1.3 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    4 pages
    English language

SIGNIFICANCE AND USE
5.1 Leaks in medical device, pharmaceutical, and food packages may result in the ingress of unwanted gases (most commonly oxygen), harmful microbiological, or particulate contaminants. Package leaks may appear as imperfections in the package components themselves or at the seal juncture between mated components. The ability to detect leaks is necessary to ensure consistency and integrity of packages.  
5.2 After initial set-up and calibration, individual test operation may be semi-automatic, automatic, or manual. The test method permits non-destructive detection of leaks not visibly detectable. The test method does not require the introduction of any extraneous materials or substances, such as dyes or gases. However, it is important to physically mask or block off any package porous barrier surface during the test to prevent rapid loss of chamber vacuum resulting primarily from gas migration through the porous surface. Leak detection is based solely on the ability to detect the change in pressure inside the test chamber resulting from gas or vapor egress from a package challenged with vacuum.  
5.3 This test is a useful research tool for optimizing package sealing parameters and for comparatively evaluating various packages and materials. This test method is also applicable to production settings as it is rapid, non-invasive, and non-destructive, making it useful for either 100 % on-line testing or to perform tests on a statistical sampling from the production operation.  
5.4 Leak test results that exceed the permissible limits for the vacuum decay test are indicated by audible or visual signal responses, or both.
SCOPE
1.1 Test Packages—Packages that can be nondestructively evaluated by this test method include:  
1.1.1 Rigid and semi-rigid non-lidded trays.  
1.1.2 Trays or cups sealed with porous barrier lidding material.  
1.1.3 Rigid, nonporous packages.  
1.1.4 Flexible, nonporous packages.  
1.2 Leaks Detected—This test method detects package leaks by measuring the rise in pressure (vacuum loss) in an enclosed evacuated test chamber containing the test package. Vacuum loss results from leakage of test package headspace gases and/or volatilization of test package liquid contents located in or near the leak. When testing for leaks that may be partially or completely plugged with the package’s liquid contents, the test chamber is evacuated to a pressure below the liquid’s vaporization pressure. All methods require a test chamber to contain the test package and a leak detection system designed with one or more pressure transducers. Test method sensitivities cited below were determined using specific product-package systems selected for the precision and bias studies summarized in Table 1. Table 1 also lists other examples of relevant product-package systems that can be tested for leakage by vacuum decay. TABLE 1 Summary of Vacuum Decay Leak Tests Applications for Various Product-Packages Systems    
Package ExamplesA  
Package Content Examples  
ASTM P&B Data Tables  
Target VacuumB  
GAS LEAK TEST
PACKAGE APPLICATIONS AND PRECISION & BIAS STUDIES  
Porous barrier lidded traysC  
Empty
Solids (tablets, capsules, powders, devices)  
3, 4, 5  
–400 mbar  
Nonlidded traysC or cups  
Empty  
2  
–400 mbar  
Plastic screw capped bottlesC  
Solids (tablets, capsules, powders)
Liquids (with significant gas headspace volume)  
6  
–500 mbar  
Glass syringesC  
Solids (lyophilized powders)  
7, 8  
+250 mbar  
ADDITIONAL GAS LEAK TEST PACKAGE APPLICATIONSA  
Lidded (nonporous) trays or cups containing solid materials (for example, powders, tablets, capsules, devices)  
Glass or plastic vials closed with elastomeric closures containing solid materials (for example, powders)  
Glass or plastic vials closed with elastomeric closures, containing liquid materials, but with significant gas headspace volume  
Flexible packages (for ...

  • Standard
    13 pages
    English language

SIGNIFICANCE AND USE
5.1 Harmful biological or particulate contaminants may enter the package through imperfections such as pinholes or cracks in trays.  
5.2 After initial instrument set-up and calibration, the operations of individual tests and test results do not need operator interpretation.  
5.3 Leak test results that exceed the permissible threshold setting are indicated by audible or visual signal responses, or both, or by other means.  
5.4 This non-destructive test method may be performed in either laboratory or production environments and may be undertaken on either a 100 % or a statistical sampling basis. This test method, in single instrument use and current implementation, may not be fast enough to work on a production packaging line, but is well suited for statistical testing as well as package developmental design work.
SCOPE
1.1 This non-destructive test method detects pinhole leaks in trays, as small as 50 μm (0.002 in.) in diameter, or equivalently sized cracks, subject to trace gas concentration in the tray, tray design and manufacturing tolerances.  
1.2 The values stated in SI units are to be regarded as standard units. Values in parentheses are for information only.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    4 pages
    English language
  • Standard
    4 pages
    English language

SIGNIFICANCE AND USE
5.1 This method allows for the evaluation of seal quality by passing an ultrasound signal through the sealed area of a package or item. Poorly sealed areas will not transmit as much ultrasonic energy as properly sealed areas.  
5.2 This method relies on quantitative analysis of ultrasound signal strength, providing a non-subjective approach to assessing package seal quality and detecting defects.  
5.3 This technique has been used for inspecting a variety of materials including flexible pouch seals, rigid tray seals and other packaging components such as affixed valves. The precision and bias for any specific package and seal configuration needs to be individually determined and validated.  
5.4 The C-Scan approach is useful for laboratory applications or off-line seal inspection. The L-Scan approach can be used for on-line, real time inspection of seal quality. The sensitivity of either approach to detect a given defect size and level of severity needs to be individually determined.  
5.5 Sound waves propagate at different speeds through different materials generally moving faster through more dense materials. The acoustic impedance (expressed as g/cm2·μs) is the product of density (g/cm3) and velocity (cm/μs). Of particular importance is the extreme difference between the impedance of air and that of any solid material. Any gap or poorly bonded area can be readily detected.    
Material  
Velocity
(cm/μsec)  
Density
(g/cm3)  
Acoustic
Impedance
(g/cm2-μsec)  
Air (20°C, 1 bar)  
0.0344  
0.00119  
0.000041  
Water (20°C)  
0.148  
1.0  
0.148  
Polyethylene  
0.267  
1.1  
0.294  
Aluminum  
0.632  
2.7  
1.710
SCOPE
1.1 This standard method describes the technology and testing procedures that can be used to detect seal defects in the size range of 1 mm and characterize seal quality in a variety of packaging styles using airborne ultrasound technology.  
1.2 This test method does not purport to be the only method for measurement of seal quality.  
1.3 Heat seals and other package components can be tested in flexible, semi-rigid and rigid packages. Only the precision and bias for flexible package seals were evaluated in a recent ILS included in the method. The precision and bias for any specific package needs to be individually determined.  
1.4 On-line, real time inspection of seals can be considered particularly in the L-Scan mode.  
1.5 This method provides a non-destructive, quantitative, non-subjective approach to flexible package seal inspection.  
1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
5.1 This method allows for the evaluation of seal quality by passing an ultrasound signal through the sealed area of a package or item. Poorly sealed areas will not transmit as much ultrasonic energy as properly sealed areas.  
5.2 This method relies on quantitative analysis of ultrasound signal strength, providing a non-subjective approach to assessing package seal quality and detecting defects.  
5.3 This technique has been used for inspecting a variety of materials including flexible pouch seals, rigid tray seals and other packaging components such as affixed valves. The precision and bias for any specific package and seal configuration needs to be individually determined and validated.  
5.4 The C-Scan approach is useful for laboratory applications or off-line seal inspection. The L-Scan approach can be used for on-line, real time inspection of seal quality. The sensitivity of either approach to detect a given defect size and level of severity needs to be individually determined.  
5.5 Sound waves propagate at different speeds through different materials generally moving faster through more dense materials. The acoustic impedance (expressed as g/cm2·μs) is the product of density (g/cm3) and velocity (cm/μs). Of particular importance is the extreme difference between the impedance of air and that of any solid material. Any gap or poorly bonded area can be readily detected.    
Material  
Velocity
(cm/μsec)  
Density
(g/cm3)  
Acoustic
Impedance
(g/cm2-μsec)  
Air (20°C, 1 bar)  
0.0344  
0.00119  
0.000041  
Water (20°C)  
0.148  
1.0  
0.148  
Polyethylene  
0.267  
1.1  
0.294  
Aluminum  
0.632  
2.7  
1.710
SCOPE
1.1 This standard method describes the technology and testing procedures that can be used to detect seal defects in the size range of 1 mm and characterize seal quality in a variety of packaging styles using airborne ultrasound technology.  
1.2 This test method does not purport to be the only method for measurement of seal quality.  
1.3 Heat seals and other package components can be tested in flexible, semi-rigid and rigid packages. Only the precision and bias for flexible package seals were evaluated in a recent ILS included in the method. The precision and bias for any specific package needs to be individually determined.  
1.4 On-line, real time inspection of seals can be considered particularly in the L-Scan mode.  
1.5 This method provides a non-destructive, quantitative, non-subjective approach to flexible package seal inspection.  
1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    5 pages
    English language

ABSTRACT
This test method covers the determination of gross leaks in flexible packaging containing a headspace gas by bubble emission. A vacuum chamber shall be any transparent container capable of withstanding approximately one atmosphere pressure differential, fitted with a vacuum-tight cover. A vacuum gage, an inlet tube from a source of vacuum, and an outlet tube to the atmosphere shall be connected to the chamber cover. Use an immersion fluid which does not degrade the package being tested. The test sample and test fluid shall be at equilibrium with normal room temperature. The procedures for testing are presented in details.
SCOPE
1.1 This test method covers the determination of gross leaks in flexible packaging containing a headspace gas. Test sensitivity is limited to 1 × 10−5 atm cm3/s (1 × 10 −6 Pa m3/s) or even less sensitive as indicated in a recent interlaboratory test (reported in Section 12).  
1.2 Small leaks may not be detected by this procedure. Viscoelastic effects on the products, or entrapped air, become significant and prevent passage through small openings. Positive pressure inside the pouch after the vacuum is drawn may force the product to plug small leaks. The size of the leak that can be detected is dependent upon the products contained, the nature of the packaging material, and the test parameters selected.  
1.3 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    3 pages
    English language

SIGNIFICANCE AND USE
5.1 These test methods provide a rapid, simple to apply method to detect small leaks in flexible package seals or walls at the leak rate level of greater than 1 × 10−4 sccs, thus providing a measure of package integrity. Porous barrier film packages made non-porous with an impermeable film forming coating may demonstrate lateral leakage through the barrier material. Verification of leakage differences from background leakage must be included in validation methods. The use of calibrated hole sizes or orifices may be appropriate to determine leakage sensitivity or barrier integrity for these materials.  
5.2 While theoretical leak rate sensitivity can be established by calculation, the test measurement is in pressure units and the measuring instrument must be calibrated, certified, and verified with these units.  
5.3 The pressure decay method of leak testing is a physical measure of package integrity. When testing medical packaging which must conform to ISO 11607–1: 2006 standards, it may necessary to verify the results of the pressure decay test method with other sterile package integrity test methods.  
5.4 Test Method A allows packages to be pressurized without restraint. In Test Method A the pouch, tray, or other type package will contain a volume of air defined by its mechanical configuration and its ability to resist internal pressure applied. This test method requires that the package reach a stable volume configuration (stop stretching) to make a measurement.  
5.5 Test Method B allows the use of rigid restraining plates against the walls of the package to limit its volume and stabilize the package volume.
SCOPE
1.1 These test methods cover the measurement of leaks in nonporous film, foil, or laminate flexible pouches and foil-sealed trays, which may be empty or enclose solid product. If product is enclosed, seals or surfaces cannot be in contact with water, oils, or other liquid.  
1.2 These test methods will detect leaks at a rate of 1 × 10−4 sccs (standard cubic centimetres per second) or greater, in flexible packages. The limitation of leak rate is dependent on package volume as tested.  
1.3 The following test methods are included:  
1.3.1 Test Method A—Pressure Decay Leak Test for Flexible Packages Without Restraining Plates  
1.3.2 Test Method B—Pressure Decay Leak Test for Flexible Packages With Restraining Plates  
1.4 These test methods are destructive in that they require entry into the package to supply an internal pressure of gas, typically air or nitrogen, although other gases may be used. The entry connection into the flexible package must be leak-tight.  
1.5 For porous packages, see 9.3.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
4.1 Harmful biological or particulate contaminants may enter the medical package through leaks. These leaks are frequently found at seals between package components of the same or dissimilar materials. Leaks may also result from a pinhole in the packaging material.  
4.2 It is the objective of this test method to visually observe the presence of channel defects by the leakage of dye through them.  
4.3 This dye penetrant procedure is applicable only to individual leaks in a package seal. The presence of a number of small leaks, as found in porous packaging material, which could be detected by other techniques, will not be indicated.  
4.4 There is no general agreement concerning the level of leakage that is likely to be deleterious to a particular package. However, since these tests are designed to detect leaks, components that exhibit any indication of leakage are normally rejected.  
4.5 These procedures are suitable to verify and locate leakage sites. They are not quantitative. No indication of leak size can be inferred from these tests. The methods are usually employed as a pass/fail test.  
4.6 The dye solution will wick through any porous material over time, but usually not within the maximum time suggested. If wicking does occur, it may be verified by observing the porous side of the subject seal area. The dye will have discolored the surface of the material. Refer to Appendix X1 for details on wicking and guidance on the observance of false positives.
SCOPE
1.1 This test method defines materials and procedures that will detect and locate a leak equal to or greater than a channel formed by a 50 µm (0.002 in.) wire in package edge seals formed between a transparent material and a porous sheet material. A dye penetrant solution is applied locally to the seal edge to be tested for leaks. After contact with the dye penetrant for a specified time, the package is visually inspected for dye penetration.  
1.2 Three dye application methods are covered in this test method: injection, edge dip, and eyedropper.  
1.3 These test methods are intended for use on packages with edge seals formed between a transparent material and a porous sheet material. The test methods are limited to porous materials which can retain the dye penetrant solution and prevent it from discoloring the seal area for a minimum of 5 seconds. Uncoated papers are especially susceptible to leakage and must be evaluated carefully for use with each test method.  
1.4 These test methods require that the dye penetrant solution have good contrast to the opaque packaging material.  
1.5 The values are stated in International System of Units (SI units) and English units. Either is to be regarded as standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    6 pages
    English language
  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
The internal pressurization test method provides a practical way to examine packages for gross leaks.
This test method is extremely useful in a test laboratory environment where no common package material/size exists.
This test method may apply to large or long packages that do not fit into any other package integrity test method apparatus.
This test method may be used as a means to evaluate package integrity. Package integrity is crucial to consumer safety since heat sealed packages are designed to provide a contamination free or sterile environment, or both, to the product.
This test method may be used to detect substrate holes and channels.
SCOPE
1.1 This test method covers the detection of gross leaks in packaging. Method sensitivity is down to 250 μm (0.010 in.) with an 81 % probability (see Section 11). This test method may be used for tray and pouch packages.
1.2 The sensitivity of this test method has not been evaluated for use with porous materials other than spunbonded polyolefin or with nonporous packaging.
1.3 This test method is destructive in that it requires entry into the package to supply an internal air pressure
1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    5 pages
    English language
  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
The vacuum, bubble test method, as described in Test Method D3078, and various other leak detection methods described elsewhere (Test Method D4991, Guide E432, Guide E479, Test Method E493, Test Method E498, Test Method E499, and Test Method E1603) have been successfully used widely in various industries and applications to determine that a given package is or is not a “leaker.” The sensitivity of any selected leak test method has to be considered to determine its applicability to a specific situation.
The procedures presented in this test method allow the user to carry out package and seal integrity testing with sufficient sensitivity to quantify seals in the previously defined moderate to very fine seal ranges.
By employing seal-isolating leak testing fixtures, packages constructed of various materials can be tested in the full range of seal performance requirements. Design of these fixtures is beyond the scope of this method.
These seal/package integrity test procedures can be utilized as:
A design tool,  
For tooling qualification,  
Process setup,  
Process validation tool,  
Quality assurance monitoring, or
Research and development.
SCOPE
1.1 This test method includes several procedures that can be used for the measurement of overall package and seal barrier performance of a variety of package types and package forms, as well as seal/closure types. The basic elements of this method include:

  • Standard
    8 pages
    English language

SIGNIFICANCE AND USE
Seal attributes can be linked directly to a number of variables in process parameters, equipment, or material, as well as environmental (room temperature and relative humidity). Visual seal characteristics and defects can provide evidence of package integrity and production sealing problems.
Visual seal defects often will be the first indication of heat sealing process variation. They also will indicate a lack of, or potential compromise to, package integrity after physical package performance testing.
SCOPE
1.1 This test method covers the determination of channels in the package seal down to a width of 75 μm [0.003 in.] with a 60–100 % probability (see Section 8).
1.1.1 The ability to visually detect channel defects in package seals is highly dependent on the size of channel, the degree of contrast from sealed and unsealed areas, the amount and type of adhesive between the two package layers, reflecting light angle, types of material used, the use of magnification, and the inspector's level of training and experience.  
1.2 This test method is applicable to packages with at least one transparent side so that the seal area may be clearly viewed.
1.3 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    4 pages
    English language
  • Standard
    4 pages
    English language

SIGNIFICANCE AND USE
Leaks in medical device, pharmaceutical, and food packages may result in the ingress of unwanted gases (most commonly oxygen), harmful microbiological, or particulate contaminants. Package leaks may appear as imperfections in the package components themselves or at the seal juncture between mated components. The ability to detect leaks is necessary to ensure consistency and integrity of packages.
After initial set-up and calibration, individual test operation may be semi-automatic, automatic, or manual. The test method permits non-destructive detection of leaks not visibly detectable. The test method does not require the introduction of any extraneous materials or substances, such as dyes or gases. However, it is important to physically mask or block off any package porous barrier surface during the test to prevent rapid loss of chamber vacuum resulting primarily from gas migration through the porous surface. Leak detection is based solely on the ability to detect the change in pressure inside the test chamber resulting from gas or vapor egress from a package challenged with vacuum.
This test is a useful research tool for optimizing package sealing parameters and for comparatively evaluating various packages and materials. This test method is also applicable to production settings as it is rapid, non-invasive, and non-destructive, making it useful for either 100 % on-line testing or to perform tests on a statistical sampling from the production operation.
Leak test results that exceed the permissible limits for the vacuum decay test are indicated by audible or visual signal responses, or both.
SCOPE
1.1 Test Packages—Packages that can be nondestructively evaluated by this test method include:
1.1.1 Rigid and semi-rigid non-lidded trays.
1.1.2 Trays or cups sealed with porous barrier lidding material.
1.1.3 Rigid, nonporous packages.
1.1.4 Flexible, nonporous packages.
1.2 Leaks Detected—This test method detects package leaks by measuring the rise in pressure (vacuum loss) in an enclosed evacuated test chamber containing the test package. Vacuum loss results from leakage of test package headspace gases and/or volatilization of test package liquid contents located in or near the leak. When testing for leaks that may be partially or completely plugged with the package’s liquid contents, the test chamber is evacuated to a pressure below the liquid’s vaporization pressure. All methods require a test chamber to contain the test package and a leak detection system designed with one or more pressure transducers. Test method sensitivities cited below were determined using specific product-package systems selected for the precision and bias studies summarized in Table 1. Table 1 also lists other examples of relevant product-package systems that can be tested for leakage by vacuum decay.  
1.2.1 Trays or Cups (Non-lidded) (Air Leakage)—Hole or crack defects in the wall of the tray/cup of at least 50 μm in diameter can be detected. Nonlidded trays were tested at a Target Vacuum of –4·E4 Pa (–400 mbar).
1.2.2 Trays Sealed with Porous Barrier Lidding Material (Headspace Gas Leakage)—Hole or crack defects in the wall of the tray/cup of at least 100 μm in diameter can be detected. Channel defects in the seal area (made using wires of 125 μm in diameter) can be detected. Severe seal bonding defects in both continuous adhesive and dot matrix adhesive package systems can be detected. Slightly incomplete dot matrix adhesive bonding defects can also be detected. All porous barrier lidding material packages were tested at a Target Vacuum of –4·E4 Pa (–400 mbar). The sensitivity of the test for porous lidded packages is approximately E-2 Pa·m3·s-1 using a calibrated volumetric airflow meter.  
1.2.3 Rigid, Nonporous Packages (Headspace Gas Leakage)—Hole defects of at least 5 μm in diameter can be detected. Plastic bottles with screw caps were tested at a target vacuum of –5·E4 Pa (–500 mbar). U...

  • Standard
    13 pages
    English language
  • Standard
    13 pages
    English language

SIGNIFICANCE AND USE
The oxygen content of a package’s headspace is an important determinant of the packaging protection afforded by barrier materials. The package under test is typically MAP (modified atmosphere packaging) packaged.
Oxygen content is a key contributor to off-flavors and spoilage of various products, such as chemicals, food and pharmaceuticals.  
The method determines the oxygen in a closed package headspace. This ability has application in:
Package Permeability Studies—The change of headspace composition over a known length of time allows the calculation of permeation. Since the headspace oxygen is measured as a percentage, the volume of the container’s headspace must be known to allow conversion into a quantity such as millilitres (ml) of oxygen. The use of this approach to measure permeation generally applies to empty package systems only as oxygen uptake or outgassing of contained products could affect results.
Leak Detection—If the headspace contains more oxygen than expected or is increasing faster than expected, a leak can be suspected. A wide variety of techniques can be employed to verify that a leak is present and to identify its location. If necessary or of interest, a leak rate may be calculated with known headspace volume and measured oxygen concentration change over time.
Efficacy of the MAP Packaging Process— If the headspace oxygen concentration is found to be higher than expected soon after packaging, the gas flushing process may not be working as well as expected. Various techniques can evaluate whether the MAP system is functioning properly.
Storage Studies—As the method is non-destructive, the headspace can be monitored over time on individual samples to insure that results of storage studies such as shelf life testing are correctly interpreted.
SCOPE
1.1 This test method covers a procedure for determination of the oxygen concentration in the headspace within a sealed package without opening or compromising the integrity of the package.
1.2 This test method requires that chemically coated components be placed on the inside surface of the package before closing.
1.3 The package must be either transparent, translucent, or a transparent window must be affixed to the package surface without affecting the package’s integrity.
1.4 As this test method determines the oxygen headspace over time, the oxygen permeability can easily be calculated as ingress per unit time as long as the volume of the container is known.
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    4 pages
    English language

ABSTRACT
This test method covers the determination of gross leaks in flexible packaging containing a headspace gas by bubble emission. A vacuum chamber shall be any transparent container capable of withstanding approximately one atmosphere pressure differential, fitted with a vacuum-tight cover. A vacuum gage, an inlet tube from a source of vacuum, and an outlet tube to the atmosphere shall be connected to the chamber cover. Use an immersion fluid which does not degrade the package being tested. The test sample and test fluid shall be at equilibrium with normal room temperature. The procedures for testing are presented in details.
SCOPE
1.1 This test method covers the determination of gross leaks in flexible packaging containing a headspace gas. Test sensitivity is limited to 1 × 10−5 atm cm3/s (1 × 10−6 Pa m3/s) or even less sensitive as indicated in a recent interlaboratory test (reported in Section 12).
1.2 Small leaks may not be detected by this procedure. Viscoelastic effects on the products, or entrapped air, become significant and prevent passage through small openings. Positive pressure inside the pouch after the vacuum is drawn may force the product to plug small leaks. The size of the leak that can be detected is dependent upon the products contained, the nature of the packaging material, and the test parameters selected.
1.3 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    3 pages
    English language
  • Standard
    3 pages
    English language

ABSTRACT
This test method covers the determination of gross leaks in flexible packaging containing a headspace gas by bubble emission. A vacuum chamber shall be any transparent container capable of withstanding approximately one atmosphere pressure differential, fitted with a vacuum-tight cover. A vacuum gage, an inlet tube from a source of vacuum, and an outlet tube to the atmosphere shall be connected to the chamber cover. Use an immersion fluid which does not degrade the package being tested. The test sample and test fluid shall be at equilibrium with normal room temperature. The procedures for testing are presented in details.
SCOPE
1.1 This test method covers the determination of gross leaks in flexible packaging containing a headspace gas. Test sensitivity is limited to 1 × 10−5 atm cm3/s (1 × 10−6 Pa m3/s) or even less sensitive as indicated in a recent interlaboratory test (reported in Section 12).
1.2 Small leaks may not be detected by this procedure. Viscoelastic effects on the products, or entrapped air, become significant and prevent passage through small openings. Positive pressure inside the pouch after the vacuum is drawn may force the product to plug small leaks. The size of the leak that can be detected is dependent upon the products contained, the nature of the packaging material, and the test parameters selected.
1.3 The values stated in inch-pound units are to be regarded as standard. The values given in parentheses are mathematical conversions to SI units that are provided for information only and are not considered standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    3 pages
    English language

SIGNIFICANCE AND USE
Leaks in medical device, pharmaceutical and food packages may result in the ingress of unwanted gases (most commonly oxygen), harmful microbiological or particulate contaminants. Package leaks may appear as imperfections in the package components themselves or at the seal juncture between mated components. The ability to detect leaks is necessary to ensure consistency and integrity of packages.
After initial set-up and calibration, the operations of individual tests may be semi-automatic, automatic or manual. The test method permits the non-destructive detection of leaks not visibly detectable. The test method does not require the introduction of any extraneous materials or substances, such as dyes or gases. However, it is important to physically mask or block off any porous barrier surface of the package during the test to prevent a rapid loss of chamber vacuum resulting primarily from gas migration through the porous surface. Leak detection is based solely on the ability to detect the change in pressure inside the test chamber as a result of air egress from the properly masked package when challenged with vacuum conditions.
This test is a useful research tool for optimization of package sealing parameters and for comparative evaluation of various packages and materials. This test method is also applicable to production settings as it is rapid, non-invasive and non-destructive, making it useful for either 100 % on-line testing or to perform tests on a statistical sampling from the production operation.
Leak test results that exceed the permissible limits for the vacuum decay test are indicated by audible or visual signal responses, or both.
SCOPE
1.1 Test PackagesPackages that can be nondestructively evaluated by this test method include:
1.1.1 Rigid and semi-rigid non-lidded trays.
1.1.2 Trays or cups sealed with porous barrier lidding material.
1.1.3 Rigid, nonporous packages.
1.1.4 Flexible, nonporous packages (see 1.2.4).
1.2 Leaks Detected - This test method is capable of detecting package leaks using an absolute or differential pressure transducer leak detector. The sensitivity of a test is a function of the sensitivity of the transducer, the package design, the design of the package test fixture, and critical test parameters of time and pressure. Types and sizes of leaks that may be detected for various package systems, as well as test sensitivities are described below. These data are based on precision and bias confirmation studies.
1.2.1 Trays or Cups (Non-lidded) - Hole or crack defects in the wall of the tray/cup of at least 50 m in diameter can be detected at a Target Vacuum of 4104 Pa (400 mbar) using an absolute pressure transducer test instrument.
1.2.2 Trays Sealed with Porous Barrier Lidding Material - Hole or crack defects in the wall of the tray/cup of at least 100 m in diameter can be detected. Channel defects in the seal area (made using wires of 125 m in diameter) can be detected. Severe seal bonding defects in both continuous adhesive and dot matrix adhesive package systems can be detected. Slightly incomplete dot matrix adhesive bonding defects can also be detected. All porous barrier lidding material packages were tested at a Target Vacuum of 4104 Pa (400 mbar) using an absolute pressure transducer test instrument. Using a calibrated volumetric airflow meter, the sensitivity of the test for porous lidded packages is shown to be approximately 10-2 Pam3s-1.
1.2.3 Rigid, Nonporous Packages - Hole defects of at least 5 m in diameter can be detected. All rigid, nonporous packages were tested at a target vacuum of 5104 Pa (500 mbar) using a differential pressure transducer test instrument. Using a calibrated volumetric airflow meter, the sensitivity of the test for rigid, nonporous packages is shown to be approximately 10-4 Pam3s -1.
1.2.4 Flexible, Nonporous Packages - Such packages may also be tested by the vacuum decay method using either an absolute or ...

  • Standard
    9 pages
    English language

SIGNIFICANCE AND USE
These test methods provide a rapid, simple to apply method to detect small leaks in flexible package seals or walls at the leak rate level of greater than 1 × 10−4 sccs, thus providing a measure of package integrity. Porous barrier film packages made non-porous with an impermeable film forming coating may demonstrate lateral leakage through the barrier material. Verification of leakage differences from background leakage must be included in validation methods. The use of calibrated hole sizes or orifices may be appropriate to determine leakage sensitivity or barrier integrity for these materials.
While theoretical leak rate sensitivity can be established by calculation, the test measurement is in pressure units and the measuring instrument must be calibrated, certified, and verified with these units.
The pressure decay method of leak testing is a physical measure of package integrity. When testing medical packaging which must conform to ISO 11607–1: 2006 standards, it may necessary to verify the results of the pressure decay test method with other sterile package integrity test methods.
Test Method A allows packages to be pressurized without restraint. In Test Method A the pouch, tray, or other type package will contain a volume of air defined by its mechanical configuration and its ability to resist internal pressure applied. This test method requires that the package reach a stable volume configuration (stop stretching) to make a measurement.
Test Method B allows the use of rigid restraining plates against the walls of the package to limit its volume and stabilize the package volume.
SCOPE
1.1 These test methods cover the measurement of leaks in nonporous film, foil, or laminate flexible pouches and foil-sealed trays, which may be empty or enclose solid product. If product is enclosed, seals or surfaces cannot be in contact with water, oils, or other liquid.
1.2 These test methods will detect leaks at a rate of 1 104 sccs (standard cubic centimetres per second) or greater, in flexible packages. The limitation of leak rate is dependent on package volume as tested.
1.3 The following test methods are included:
1.3.1 Test Method A Pressure Decay Leak Test for Flexible Packages Without Restraining Plates
1.3.2 Test Method B Pressure Decay Leak Test for Flexible Packages With Restraining Plates
1.4 These test methods are destructive in that they require entry into the package to supply an internal pressure of gas, typically air or nitrogen, although other gases may be used. The entry connection into the flexible package must be leak-tight.
1.5 For porous packages, see .
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    6 pages
    English language
  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
These test methods provide a rapid, simple to apply method to detect small leaks in flexible package seals or walls at the leak rate level of greater than 1 × 10−4 sccs, thus providing a measure of package integrity. Porous barrier film packages made non-porous with an impermeable film forming coating may demonstrate lateral leakage through the barrier material. Verification of leakage differences from background leakage must be included in validation methods. The use of calibrated hole sizes or orifices may be appropriate to determine leakage sensitivity or barrier integrity for these materials.
While theoretical leak rate sensitivity can be established by calculation, the test measurement is in pressure units and the measuring instrument must be calibrated, certified, and verified with these units.
The pressure decay method of leak testing is a physical measure of package integrity. When testing medical packaging which must conform to ISO 11607–1: 2006 standards, it may necessary to verify the results of the pressure decay test method with other sterile package integrity test methods.
Test Method A allows packages to be pressurized without restraint. In Test Method A the pouch, tray, or other type package will contain a volume of air defined by its mechanical configuration and its ability to resist internal pressure applied. This test method requires that the package reach a stable volume configuration (stop stretching) to make a measurement.
Test Method B allows the use of rigid restraining plates against the walls of the package to limit its volume and stabilize the package volume.
SCOPE
1.1 These test methods cover the measurement of leaks in nonporous film, foil, or laminate flexible pouches and foil-sealed trays, which may be empty or enclose solid product. If product is enclosed, seals or surfaces cannot be in contact with water, oils, or other liquid.
1.2 These test methods will detect leaks at a rate of 1 × 10−4 sccs (standard cubic centimetres per second) or greater, in flexible packages. The limitation of leak rate is dependent on package volume as tested.
1.3 The following test methods are included:
1.3.1 Test Method A—Pressure Decay Leak Test for Flexible Packages Without Restraining Plates
1.3.2 Test Method B—Pressure Decay Leak Test for Flexible Packages With Restraining Plates
1.4 These test methods are destructive in that they require entry into the package to supply an internal pressure of gas, typically air or nitrogen, although other gases may be used. The entry connection into the flexible package must be leak-tight.
1.5 For porous packages, see 9.3.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    6 pages
    English language
  • Standard
    6 pages
    English language

SIGNIFICANCE AND USE
Harmful biological or particulate contaminants may enter the package through imperfections such as pinholes or cracks in trays.
After initial instrument set-up and calibration, the operations of individual tests and test results do not need operator interpretation.  
Leak test results that exceed the permissible threshold setting are indicated by audible or visual signal responses, or both, or by other means.
This non-destructive test method may be performed in either laboratory or production environments and may be undertaken on either a 100 % or a statistical sampling basis. This test method, in single instrument use and current implementation, may not be fast enough to work on a production packaging line, but is well suited for statistical testing as well as package developmental design work.
SCOPE
1.1 This non-destructive test method detects pinhole leaks in trays, as small as 50 m (0.002 in.) in diameter, or equivalently sized cracks, subject to trace gas concentration in the tray, tray design and manufacturing tolerances.
1.2 The values stated in SI units are to be regarded as standard units. Values in parentheses are for information only.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    4 pages
    English language

SIGNIFICANCE AND USE
Harmful biological or particulate contaminants may enter the package through incomplete seals or imperfections such as pinholes or cracks in the trays.  
After initial instrument set-up and calibration, the operations of individual tests and test results do not need operator interpretation. The non-destructive nature of the test may be important when testing high value added products.
Leak test results that exceed the permissible threshold setting are indicated by audible or visual signal responses, or both, or by other means.
This non-destructive test method may be performed in either laboratory or production environments. This testing may be undertaken on either a 100 % or a statistical sampling basis. This test method, in single instrument use and current implementation, may not be fast enough to work on a production packaging line, but is well suited for statistical testing as well as package developmental design work.
SCOPE
1.1 This non-destructive test method detects leaks in non-porous rigid thermoformed trays, as well as the seal between the porous lid and the tray. The test method detects channel leaks in packages as small as 100 m (0.004 in.) diameter in the seal as well as 50 m (0.002 in.) diameter pinholes, or equivalently sized cracks in the tray, subject to trace gas concentration in the package, package design and manufacturing tolerances.
Note 1—This test method does not claim to challenge the porous (breathable) lidding material. Any defects that may exist in the porous portion of the package will not be detected by this test method.
1.2 The values stated in SI units are to be regarded as standard units. Values in parentheses are for information only.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    5 pages
    English language

SIGNIFICANCE AND USE
The vacuum, bubble test method, as described in Test Method D 3078, and various other leak detection methods described elsewhere (Test Method D 4991, Guide E 432, Guide E 479, Test Method E 493, Test Method E 498, Test Method E 499, and Test Method E 1603) have been successfully used widely in various industries and applications to determine that a given package is or is not a “leaker.” The sensitivity of any selected leak test method has to be considered to determine its applicability to a specific situation.
The procedures presented in this test method allow the user to carry out package and seal integrity testing with sufficient sensitivity to quantify seals in the previously defined moderate to very fine seal ranges.
By employing seal-isolating leak testing fixtures, packages constructed of various materials can be tested in the full range of seal performance requirements. Design of these fixtures is beyond the scope of this method.
These seal/package integrity test procedures can be utilized as:
5.4.1 A design tool,  
5.4.2 For tooling qualification,  
5.4.3 Process setup,  
5.4.4 Process validation tool,  
5.4.5 Quality assurance monitoring, or
5.4.6 Research and development.
SCOPE
1.1 This test method includes several procedures that can be used for the measurement of overall package and seal barrier performance of a variety of package types and package forms, as well as seal/closure types. The basic elements of this method include:
1.1.1 Helium (employed as tracer gas),
1.1.2 Helium leak detector (mass spectrometer), and
1.1.3 Package/product-specific test fixtures.
1.1.4 Most applications of helium leak detection are destructive, in that helium needs to be injected into the package after the package has been sealed. The injection site then needs to be sealed/patched externally, which often destroys its saleability. Alternatively, if helium can be incorporated into the headspace before sealing, the method can be non-destructive because all that needs to be accomplished is to simply detect for helium escaping the sealed package.
1.2 Two procedures are described; however the supporting data in Section 14 only reflects Procedure B (Vacuum Mode). The alternative, Sniffer Mode, has proven to be a valuable procedure for many applications, but may have more variability due to exactly the manner that the operator conducts the test such as whether the package is squeezed, effect of multiple small leaks compared to fewer large leaks, background helium concentration, package permeability and speed at which the scan is conducted. Further testing to quantify this procedures variability is anticipated, but not included in this version.
1.2.1 Procedure A: Sniffer Mode—the package is scanned externally for helium escaping into the atmosphere or fixture.
1.2.2 Procedure B: Vacuum Mode— the helium containing package is placed in a closed fixture. After drawing a vacuum, helium escaping into the closed fixture (capture volume) is detected. Typically, the fixtures are custom made for the specific package under test.
1.3 The sensitivity of the method can range from the detection of:
1.3.1 Large leaks-10-2 Pam 3/s to 10-5 Pam3/s (10-1 cc/sec/atm to 10-4 cc/sec/atm).
1.3.2 Moderate leaks-10-5 Pam 3/s to 10-7 Pam3/s (10-4 cc/sec/atm to 10-6 cc/sec/atm).
1.3.3 Fine leaks-10-7 Pam 3/s to 10-9 Pam3/s (10-6 cc/sec/atm to 10-8 cc/sec/atm).
1.3.4 Ultra-Fine leak-10-9 Pam 3/s to 10-11 Pam3/s (10-8 cc/sec/atm to 10-10 cc/sec/atm).
Note 1—Conversion from cc/sec/atm to Pam3/s is achieved by multiplying by 0.1.
1.4 The terms large, moderate, fine and ultra-fine are relative terms only and do not imply the acceptability of any leak rate. The individual application dictates the level of integrity needed. For many packaging applications, only "large leaks" are considered unacceptable and the ability to detect smaller leaks is immaterial. All leak rates referred to in this method are based on c...

  • Standard
    7 pages
    English language

SCOPE
1.1 Test Packages—Packages that can be nondestructively evaluated by this test method include:
1.1.1 Rigid and semi-rigid non-lidded trays.
1.1.2 Trays or cups sealed with porous barrier lidding material.
1.1.3 Rigid, nonporous packages.
1.1.4 Flexible, nonporous packages (see 1.2.4).
1.2 Leaks Detected—This test method is capable of detecting package leaks using an absolute or differential pressure transducer leak detector. The sensitivity of a test is a function of the sensitivity of the transducer, the package design, the design of the package test fixture, and critical test parameters of time and pressure. Types and sizes of leaks that may be detected for various package systems, as well as test sensitivities are described below. These data are based on precision and bias confirmation studies.
1.2.1 Trays or Cups (Non-lidded)—Hole or crack defects in the wall of the tray/cup of at least 50 μm in diameter can be detected at a Target Vacuum of 4·104 Pa (400 mbar) using an absolute pressure transducer test instrument.
1.2.2 Trays Sealed with Porous Barrier Lidding Material—Hole or crack defects in the wall of the tray/cup of at least 100 μm in diameter can be detected. Channel defects in the seal area (made using wires of 125 μm in diameter) can be detected. Severe seal bonding defects in both continuous adhesive and dot matrix adhesive package systems can be detected. Slightly incomplete dot matrix adhesive bonding defects can also be detected. All porous barrier lidding material packages were tested at a Target Vacuum of 4104 Pa (400 mbar) using an absolute pressure transducer test instrument. Using a calibrated volumetric airflow meter, the sensitivity of the test for porous lidded packages is shown to be approximately 10-2 Pam3s-1.
1.2.3 Rigid, Nonporous Packages—Hole defects of at least 5 μm in diameter can be detected. All rigid, nonporous packages were tested at a target vacuum of 5·104 Pa (500 mbar) using a differential pressure transducer test instrument. Using a calibrated volumetric airflow meter, the sensitivity of the test for rigid, nonporous packages is shown to be approximately 10-4 P·am3s -1.
1.2.4 Flexible, Nonporous Packages—Such packages may also be tested by the vacuum decay method using either an absolute or differential pressure tranducer test instrument. The instrument should be selected based on the leak test sensitivity desired. Sensitivity data for flexible packages were not included in the precision and bias studies, although the use of vacuum decay for testing such packages is well known.
1.3 Test Results—The test results are qualitative (Accept/Reject). Acceptance criteria for test results are established from quantitative baseline vacuum decay measurements obtained from control, non-leaking packages.
1.4 Standard Value Units—The values used in this test method are stated in SI units and are to be regarded as standard units. Values in parentheses are for information only.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    9 pages
    English language

SIGNIFICANCE AND USE
Seal attributes can be linked directly to a number of variables in process parameters, equipment, or material, as well as environmental (room temperature and relative humidity). Visual seal characteristics and defects can provide evidence of sterile package integrity and production sealing problems.
Visual seal defects often will be the first indication of heat sealing process variation. They also will indicate a lack of, or potential compromise to, package integrity after physical package performance testing.
SCOPE
1.1 This test method covers the determination of channels in the package seal down to a width of 75 μm (0.003 in.) with a 60-100 % probability (see Section 8).
1.1.1 The ability to visually detect channel defects in package seals is highly dependent on the size of channel, the degree of contrast from sealed and unsealed areas, the amount and type of adhesive between the two package layers, reflecting light angle, types of material used, the use of magnification, and the inspector's level of training and experience.
1.2 This test method is applicable to flexible and rigid packages with at least one transparent side so that the seal area may be clearly viewed.
1.3 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    4 pages
    English language

SIGNIFICANCE AND USE
Harmful biological or particulate contaminants may enter the device through leaks. These leaks are frequently found at seals between package components of the same or dissimilar materials. Leaks may also result from a pinhole in the packaging material.
This dye penetrant procedure is applicable only to individual leaks in a package seal. The presence of a number of small leaks, as found in porous packaging material, which could be detected by other techniques, will not be indicated.
There is no general agreement concerning the level of leakage that is likely to be deleterious to a particular package. However, since these tests are designed to detect leakage, components that exhibit any indication of leakage are normally rejected.
Since leaks may change in size with different ambient conditions, comparisons between test stations are not conclusive. Therefore this method is usually employed as a go, no-go test.
The dye solution will wick through any porous material over time, but usually not within the maximum time suggested. If wicking does occur, it may be verified by observing the porous side of the subject seal area. The dye will have discolored the surface of the material.
When puncturing the packaging to allow injection of the dye penetrant solution, care should be taken not to puncture other package surfaces. Puncturing of the package is facilitated if it is done adjacent to a dummy device inside the package. The device will provide a tenting effect that will separate the two sides of the package, reducing the chance of accidental puncture of both sides.
SCOPE
1.1 This test method defines materials and a procedure that will detect and locate a leak equal or greater than a channel formed by a 50 μm (0.002 in.) wire in package edge seals formed between a transparent film and a porous sheet material. A dye penetrant solution is applied locally to the seal edge to be tested for leaks. After contact with the dye penetrant for a specified time, the package is visually inspected for dye penetration.
1.2 This test method is intended for use on packages with edge seals formed between a transparent film and a porous sheet material. This test method is limited to porous materials which can retain the dye penetrant solution and prevent it from discoloring the entire seal area for a minimum of 20 s. Uncoated papers are especially susceptible to leakage and must be evaluated carefully fro use with this test method.
1.3 This test method requires that the dye penetrant have good contrast to the opaque packaging material.
1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    3 pages
    English language

SIGNIFICANCE AND USE
The internal pressurization test method provides a practical way to examine packages for gross leaks, which may render the product non-sterile.
This test method is extremely useful in a test laboratory environment where no common package material/size exists.
This test method may apply to very large or long packages, which do not fit into any other package integrity test method apparatus.
This test method may be used as a means to evaluate package integrity. Package integrity is crucial to consumer safety since heat sealed packages are designed to provide a contamination free and sterile environment to the product.
SCOPE
1.1 This test method covers the detection of gross leaks in medical packaging. Method sensitivity is down to 250 μm with an 81 % probability (see Section 11). This test method may be used for tray and pouch packages.
1.2 The sensitivity of this test method has not been evaluated for use with porous materials other than spunbonded polyolefin or with nonporous packaging.
1.3 This test method is destructive in that it requires entry into the package to supply an internal air pressure
1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    5 pages
    English language

SCOPE
1.1 This non-destructive test method detects pinhole leaks in trays, as small as 50 m (0.002 in.) in diameter, or equivalently sized cracks, subject to trace gas concentration in the tray, tray design and manufacturing tolerances.
1.2 The values stated in SI units are to be regarded as standard units. Values in parentheses are for information only.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    4 pages
    English language

SCOPE
1.1 This test method covers the determination of gross leaks in flexible packaging containing a headspace gas. Test sensitivity is limited to 1 x 10-5 atm cm3/s (1 x 10-6 Pa m3/s) or even less sensitive as indicated in a recent interlaboratory test (reported in Section 12).
1.2 Small leaks may not be detected by this procedure. Viscoelastic effects on the products, or entrapped air, become significant and prevent passage through small openings. Positive pressure inside the pouch after the vacuum is drawn may force the product to plug small leaks. The size of the leak that can be detected is dependent upon the products contained, the nature of the packaging material, and the test parameters selected.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    3 pages
    English language

SCOPE
DESIG: F2228 02 ^TITLE: Standard Test Method for Non-Destructive Detection of Leaks in Medical Packaging Which Incorporates Porous Barrier Material by CO2 Tracer Gas Method ^SCOPE:1. Scope
1.1 This non-destructive test method detects leaks in non-porous rigid thermoformed trays, as well as the seal between the porous lid and the tray. The test method detects channel leaks in packages as small as 100 m (0.004 in.) diameter in the seal as well as 50 m (0.002 in.) diameter pinholes, or equivalently sized cracks in the tray, subject to trace gas concentration in the package, package design and manufacturing tolerances.Note 1
This test method does not claim to challenge the porous (breathable) lidding material. Any defects that may exist in the porous portion of the package will not be detected by this test method.
1.2 The values stated in SI units are to be regarded as standard units. Values in parentheses are for information only.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    5 pages
    English language

SCOPE
1.1 These test methods cover the measurement of leaks in nonporous film, foil, or laminate flexible pouches and foil-sealed trays, which may be empty or enclose solid product. If product is enclosed, seals or surfaces cannot be in contact with water, oils, or other liquid.
1.2 These test methods will detect leaks at a rate of 1 104 sccs (standard cubic centimetres per second) or greater, in flexible packages. The limitation of leak rate is dependent on package volume as tested.
1.3 The following test methods are included:
1.3.1 Test Method A-- Pressure Decay Leak Test for Nonporous Flexible Packages Without Restraining Plates
1.3.2 Test Method B-- Pressure Decay Leak Test for Nonporous Flexible Packages With Restraining Plates
1.4 These test methods are destructive in that they require entry into the package to supply an internal pressure of gas, typically air or nitrogen, although other gases may be used. The entry connection into the flexible package must be leak-tight.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  • Standard
    6 pages
    English language

Frequently Asked Questions

F02.40 is a Technical Committee within ASTM International. It is named "Package Integrity". This committee has published 53 standards.

F02.40 develops ASTM standards in the area of Information technology. Currently, there are 53 published standards from this technical committee.

ASTM is a standardization organization that develops and publishes standards to support industry, commerce, and regulatory requirements.

A Technical Committee (TC) in ASTM is a group of experts responsible for developing international standards in a specific technical area. TCs are composed of national member body delegates and work through consensus to create standards that meet global industry needs. Each TC may have subcommittees (SCs) and working groups (WGs) for specialized topics.

Loading...