This document specifies requirements for non-sterile, single-use graduated flow regulators used as subcomponents in sterilized infusion sets for single use to control the flow of intravenous infusion solutions with fluid contact under gravity feed conditions.

  • Standard
    13 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and gives the test method for medical gloves for single use in order to determine freedom from holes.

  • Standard
    10 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for sterile, single-use, extracorporeal blood-gas exchangers (oxygenators) intended for the supply of oxygen to, and the removal of carbon dioxide from, human blood, during cardiopulmonary bypass (CPB) for up to 6 h, extracorporeal lung assist [ECLA with veno-venous (VV), veno-arterial (VA) or veno-arterial-venous (VAV) cannulation strategies], cardiopulmonary support (CPS), extracorporeal life support (ECLS with VA cannulation strategy), extracorporeal carbon dioxide removal (ECCO2R), and other extracorporeal circulation techniques requiring blood-gas exchange.
This document also applies to heat exchangers and arterial filters that are integral parts of the oxygenator.
This document also applies to external equipment unique to the use of the oxygenator.
This document does not apply to
—     implanted oxygenators,
—     liquid oxygenators,
—     extracorporeal circuits (blood tubing),
—     separate heat exchangers,
—     separate ancillary devices, and
—     separate arterial line filters.

  • Standard
    30 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the general requirements for the preparation of fluids for haemodialysis and related therapies and substitution fluid for use in online therapies, such as haemodiafiltration and haemofiltration, for dialysis practitioners. This document gives guidance on the user's responsibility for fluids used in haemodialysis and related therapies once the equipment used in its preparation has been delivered and installed. As dialysis water used to prepare dialysis fluid can also be used to reprocess dialysers not marked intended for single use, this aspect of water use is also covered by this document.
This document is applicable to
—     the quality management of equipment used to treat and distribute water used for the preparation of dialysis fluid and substitution fluid, from the point at which municipal water enters the dialysis facility to the point at which the final dialysis fluid enters the dialyser or the point at which substitution fluid is infused.
—     the quality management of the equipment used to prepare acid and bicarbonate concentrate from powdered or other highly concentrated media at a dialysis facility, and
—     the preparation of the final dialysis fluid or substitution fluid from dialysis water and concentrates.
This document does not apply to
—     sorbent-based dialysis fluid regeneration systems that regenerate and recirculate small volumes of dialysis fluid,
—     systems for continuous renal replacement therapy that use pre-packaged solutions, and
—     systems and solutions for peritoneal dialysis.
This document does not address clinical issues associated with inappropriate usage of such fluids.

  • Standard
    114 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and recommendations for individual water treatment devices and water treatment systems assembled from one or more of such devices. This document is directed at the individual or company that specifies the complete water treatment system and, the supplier who assembles and installs the system. Since systems can be assembled from a number of individual water treatment devices, the provisions of this document are also directed at the manufacturers of these devices, provided that the manufacturer indicates that the device is intended to be used to supply water for haemodialysis and related therapies.
This document is applicable to all devices, piping and fittings between the point at which water is delivered to the water purification system and the point of use of the purified water. Such components include but are not necessarily limited to water purification devices, online water quality monitors (such as conductivity monitors) and piping systems for the distribution of purified water.
This document does not apply to
—     equipment used in the preparation of concentrates from powder or other highly concentrated media at a dialysis facility either for a single patient or multiple patients,
—     dialysis fluid supply systems that proportion water and concentrates to produce dialysis fluid,
—     sorbent dialysis fluid regeneration systems that regenerate and recirculate small volumes of the dialysis fluid,
—     dialysis concentrates,
—     haemodiafiltration or haemofiltration systems,
—     systems that process dialysers for multiple uses, and
—     peritoneal dialysis systems.
Requirements for the ongoing monitoring of water purity in terms of chemical and microbiological quality are given in ISO 23500-3.

  • Standard
    42 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and gives test methods for physical properties of single-use medical gloves (i.e. surgical gloves and examination/procedure gloves) in order to ensure that they provide and maintain in use an adequate level of protection from cross contamination for both patient and user.
This document does not specify the size of a lot. Attention is drawn to the difficulties that can be associated with the distribution and control of very large lots. The recommended maximum individual lot size for production is 500 000.

  • Standard
    14 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the chemical and microbiological requirements for concentrates used for haemodialysis and related therapies and applies to the manufacturer of such concentrates.
This document is applicable to:
—     concentrates in both liquid and powder forms;
—     additives, also called spikes, which are chemicals that can be added to the concentrate to supplement or increase the concentration of one or more of the existing ions in the concentrate and thus in the final dialysis fluid;
—     equipment used to mix acid and bicarbonate powders into concentrate at the user's facility.
This document does not apply to:
—     concentrates prepared from pre-packaged salts and water at a dialysis facility for use in that facility;
—     pre-packaged and sterile dialysis fluid;
—     sorbent dialysis fluid regeneration systems that regenerate and recirculate small volumes of the dialysis fluid;
—     equipment to perform patient treatment; this is addressed IEC 60601-2-16.
This document does not cover the dialysis fluid that is used to clinically dialyse patients. Dialysis fluid is covered in ISO 23500-5. The making of dialysis fluid involves the proportioning of concentrate and water at the bedside or in a central dialysis fluid delivery system. Although the label requirements for dialysis fluid are placed on the labelling of the concentrate, it is the user's responsibility to ensure proper use.

  • Standard
    31 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the minimum chemical and microbiological quality requirements, for water used for preparation of dialysis fluids, concentrates, and for the reprocessing of haemodialysers, together with the necessary steps to ensure conformity with the requirements. The document also provides guidance for the ongoing monitoring of the purity of such water in terms of chemical and microbiological quality.
This document is applicable to
—     water used in the preparation of dialysis fluids for haemodialysis, haemodiafiltration and haemofiltration and the reprocessing of haemodialysers, and
—     water used in the preparation of concentrates.
This document does not apply to dialysis fluid regenerating systems.
The operation of water treatment equipment and the final mixing of treated water with concentrates to produce dialysis fluid are the sole responsibility of dialysis professionals.

  • Standard
    26 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the minimum chemical and microbiological quality requirements for dialysis fluids used in haemodialysis and related therapies.
This document applies to
—     dialysis fluids used for haemodialysis and haemodiafiltration,
—     substitution fluid produced online for haemodiafiltration and haemofiltration based on dialysis fluid
This document does not apply to
—     the water and concentrates used to prepare dialysis fluid or the equipment to produce dialysis fluid
—     sorbent-based dialysis fluid regeneration systems that regenerate and recirculate small volumes of dialysis fluid,
—     systems for continuous renal replacement therapy that use pre-packaged solutions, and
—     systems and solutions for peritoneal dialysis.
The delivery and monitoring of the dialysis fluid composition and its permitted deviation from set points is governed by protective systems defined in IEC 60601-2-16.

  • Standard
    20 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for disposable extracorporeal blood and fluid circuits and accessories used in combination with haemodialysis equipment intended for extracorporeal blood treatment therapies such as, but not limited to, haemodialysis, haemodiafiltration, haemofiltration.
This document does not apply to:
—     haemodialysers, haemodiafilters or haemofilters;
—     plasmafilters;
—     haemoperfusion devices;
—     vascular access devices.
NOTE 1        Requirements for haemodialysers, haemodiafilters, haemofilters and haemoconcentrators are specified in ISO 8637-1.
NOTE 2        Requirements for plasmafilters are specified in ISO 8637-3.

  • Standard
    36 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies minimum requirements and a test method for the antimicrobial (microbicidal or microbistatic) activity of wound dressing products. It applies to all wound dressing products that specifically claim antimicrobial activity according to this document.

  • Standard
    47 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for balloon dilatation catheters supplied sterile and intended for single use.
This document does not specify requirements for vascular stents (see ISO 25539-2).
NOTE            Guidance on the selection of balloon materials is given in Annex G.

  • Standard
    25 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies general requirements for intravascular catheters, supplied sterile and intended for single use, for any application.
This document does not apply to intravascular catheter accessories, e.g. those covered by ISO 11070.

  • Standard
    52 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This part of EN 455 specifies requirements for the evaluation of biological safety for medical gloves for single use. It gives requirements for labelling and the disclosure of information relevant to the test methods used.

  • Standard
    43 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies test methods for the evaluation of aspects of absorption of wound dressings, test methods for the evaluation of moisture vapour transmission rate of permeable film wound and fixation dressings, and test methods to assess waterproofness and extensibility.

  • Standard
    70 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies a test method to assess bacterial retention ability of finished stand-alone and integrated air-inlet filters on administration devices for infusion and transfusion applications.

  • Standard
    14 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This part of ISO 11608 specifies requirements and test methods for needle-based injection systems with automated functions (referred to in the standard as NIS-AUTO), for the administration of medicinal products in humans. This document does not cover remote communication from the NIS-AUTO.

  • Standard
    30 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the requirements for infusion sets for single use that use light-protective agents in the fluid path materials (henceforth abbreviated as "light-protective infusion sets”).
This document also provides guidelines for performance and quality specifications of materials used in light-protective infusion sets.

  • Amendment
    7 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the requirements and methods for the clinical investigation of continuous automated non-invasive sphygmomanometers used for the measurement of the blood pressure of a patient.
This document does not cover usability aspects such as the form and manner of the data display or output. This document does not specify a numerical threshold on the minimum output period. A continuous automated non-invasive sphygmomanometer providing blood pressure parameters (e.g., systolic blood pressure, diastolic blood pressure or mean arterial pressure) with an output period considerably larger than 30 s is not typically considered a continuous automated non-invasive sphygmomanometer.
This document covers both trending continuous automated non-invasive sphygmomanometers and absolute accuracy continuous automated non-invasive sphygmomanometers and focuses solely on requirements for the clinical investigation. Representation of output is not covered by this document.
NOTE 1    IEC 62366-1 provides requirements on the application of usability engineering to medical devices. The usability engineering process can be used to clarify for the intended user whether the displayed data concerns absolute accurate values or trending values.
The requirements and methods for the clinical investigation of continuous automated non-invasive sphygmomanometers provided in this document are applicable to any subject population, and any condition of use of the continuous automated non-invasive sphygmomanometers.
NOTE 2    Subject populations can, for example, be represented by age or weight ranges.
NOTE 3    This document does not provide a method to assess the effect of artefacts during the clinical investigation (e.g. motion artefacts induced by the movement of the subject or the movement of the platform supporting the subject).
This document specifies additional disclosure requirements for the accompanying documents of continuous automated non-invasive sphygmomanometers that have undergone clinical investigation according to this document.
This document is not applicable to:
—    the clinical investigation of a non-automated sphygmomanometer as given in ISO 81060-1,
—    the clinical investigation of an intermittent automated non-invasive sphygmomanometer as given in ISO 81060-2,
—    an automated non-invasive sphygmomanometer as given in IEC 80601-2-30, or
—    invasive blood pressure monitoring equipment as given in IEC 60601‑2‑34.

  • Standard
    44 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and test methods for needle-based injection systems (NISs) containing electronics with or without software (NIS-Es).
The needle-based injection system containing electronics can be single use or reusable and can be operated with or without electrical/conductive connections to other devices. The system is intended to deliver medication to a patient by self-administration or by administration by one other operator (e.g. caregiver or health care provider).
This document applies to electronic accessories that are intended to be physically connected to a NIS or NIS-E according to the NIS/NIS-E intended use.
This document also applies to electronic accessories that are intended to have electrical/conductive connections to a NIS or NIS-E according to the NIS/NIS-E intended use.
This document does not specify requirements for software in programmable NIS-E.
NOTE       IEC 60601-1:2005+AMD1:2012+AMD2:2020, Clause 14 addresses software life cycle processes.
This document does not specify requirements for cybersecurity.

  • Standard
    72 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and test methods for single-use, double-ended, sterile needles intended to be used with some needle-based injection systems (NISs) that use a non-integrated double-ended needle according to ISO 11608-1.
This document is not applicable to the following:
—    needles for dental use;
—    pre-attached syringe needles;
—    hypodermic needles;
—    needles intended for different routes of administration (e.g. intravenous, intrathecal, intraocular);
—    materials that form the medicinal product contact surfaces of the primary container closure.
However, while this document is not intended to directly apply to these needle products, it does contain requirements and tests methods that can be used to help design and evaluate them.
NOTE       Needles provided by the manufacturer integrated into the fluid path or container are covered in ISO 11608‑3, and hypodermic needles provided separately are covered in ISO 7864.

  • Standard
    40 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and test methods for design verification of containers and integrated fluid paths used with Needle-Based Injection Systems (NISs) according to ISO 11608-1.
It is applicable to single and multi-dose containers either filled by the manufacturer (primary container closure) or by the end-user (reservoir) (e.g. cartridges, syringes) and fluid paths that are integrated with the NIS at the point of manufacture.
This document is also applicable to prefilled syringes (see ISO 11040-8) when used with a NIS (see also scope of ISO 11608-1:2022).
This document is not applicable to the following products:
—    sterile hypodermic needles;
—    sterile hypodermic syringes;
—    sterile single-use syringes, with or without needle, for insulin;
—    containers that can be refilled multiple times;
—    containers intended for dental use;
—    catheters or infusion sets that are attached or assembled separately by the user.

  • Standard
    35 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the requirements and methods for the clinical investigation of me equipment used for the intermittent non-invasive automated estimation of the arterial blood pressure by utilizing a cuff.
This document is applicable to all sphygmomanometers that sense or display pulsations, flow or sounds for the estimation, display or recording of blood pressure. These sphygmomanometers need not have automatic cuff inflation.
This document covers sphygmomanometers intended for use in all patient populations (e.g. all age and weight ranges), and all conditions of use (e.g. ambulatory blood pressure monitoring, stress testing blood pressure monitoring and blood pressure monitors for the home healthcare environment for self-measurement as well as use in a professional healthcare facility).
EXAMPLE Automated sphygmomanometer as given in IEC 80601-2-30 undergoing clinical investigation according to this document.
This document specifies additional disclosure requirements for the accompanying documents of sphygmomanometers that have passed a clinical investigation according to this document.
This document is not applicable to clinical investigations of non-automated sphygmomanometers as given in ISO 81060-1 or invasive blood pressure monitoring equipment as given in IEC 60601-2-34.
This document is not applicable to clinical investigations of a set of cuffs that are not of same materials and construction. Each type of cuff set is required to be evaluated separately according to this document.

  • Standard
    47 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the requirements for infusion sets for single use that use light-protective agents in the fluid path materials (henceforth abbreviated as "light-protective infusion sets”).
This document also provides guidelines for performance and quality specifications of materials used in light-protective infusion sets.

  • Standard
    16 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and test methods for Needle-Based Injection Systems (NISs) for single-patient use intended to deliver discrete volumes (bolus) of medicinal product, which can be delivered through needles or soft cannulas for intradermal, subcutaneous and/or intramuscular delivery, incorporating pre-filled or user-filled, replaceable or non-replaceable containers.
This document applies in cases where the NIS incorporates a prefilled syringe. However, stand-alone prefilled syringes defined by ISO 11040-8 are not covered by this document (see exclusions below).
It is important to note that other functions and characteristics of the prefilled syringe, such as dose accuracy, are subject to the requirements (delivered volume) in ISO 11040-8 and not this document, unless the addition impacts the delivery function (e.g. a mechanism that intends to restrict or stop the plunger movement, which would limit the dose delivered). In that case, the system is completely covered by this document and applicable requirements of the ISO 11608 series.
Excluded from the scope are:
—    stand-alone prefilled syringes defined by ISO 11040-8 (with noted exceptions above);
—    NISs that provide continuous delivery and require a delivery rate clinically specified in the medicinal product labelling or determined by a physician based on clinical relevance (i.e. medication efficacy) as would be the case with insulin patch pumps or traditional infusion pumps (e.g. IEC 60601-2-24, ISO 28620) associated with continuous delivery of medicinal products (e.g. insulin);
—    NISs with containers that can be refilled multiple times;
—    requirements relating to methods or equipment associated with user filling of containers unless they are dedicated accessories (a component necessary for primary function, whether included in the original kitted product or not);
—    NISs intended for dental use;
—    NISs intended for different routes of administration (e.g. intravenous, intrathecal, intraocular).
NOTE       These products that are excluded might benefit from elements in this document but might not completely fulfil the basic safety and effectiveness of such products.

  • Standard
    81 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and test methods for sterilized single-use transfer sets that are used for pharmaceutical preparations.

  • Standard
    23 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for check valves intended for single use and used with infusion equipment both with gravity-feed infusion and with pressure infusion apparatus.
The functional requirements in this document also apply to inline check valves.

  • Standard
    18 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 8536-14:2016 specifies requirements for non-sterile clamps and flow regulators used as a subcomponent to control the flow of intravenous solutions and/or blood components through sterilized infusion and blood transfusion sets and blood bag assemblies without fluid contact.
In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over ISO 8536-14:2016.

  • Standard
    16 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies identification and characterization procedures applicable to elastomeric parts including coated stoppers used for drug containers and medical devices.
The physical and chemical test procedures specified in this document permit the determination of the typical characteristics of elastomeric parts including coatings and surface treatments and can serve as a basis for agreements between manufacturer and user regarding the product consistency in subsequent supplies. Depending upon the type of elastomer and its application, an appropriate set of tests is selected.

  • Standard
    31 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies the properties and performance of sterile single-use hypodermic syringes with an auto-disable syringe feature intended to deliver a fixed dose of vaccine immediately after filling. The syringes can be made of plastic, rubber or other materials and can be with or without needle and needle protection feature.
This document does not specify the design of the auto-disable syringe feature.
This document is not applicable to syringes for use with insulin (covered by ISO 8537), syringes for use with power-driven syringe pumps (covered by ISO 7886-2), reuse prevention syringes (covered by ISO 7886-4) or syringes designed to be prefilled (covered by the ISO 11040 series). It does not address compatibility with injection fluids/vaccines.

  • Standard
    24 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for sterile single-use hypodermic syringes of nominal capacity 1 ml and above, made of plastic materials and intended for use with power-driven syringe pumps.
This document does not apply to syringes with auto-disable syringe features (ISO 7886-3[2]), syringes for use with insulin (ISO 8537[3]), single-use syringes made of glass, syringes prefilled with the injection by the manufacturer and syringes supplied with the injection as a kit for filling by a pharmacist. It does not address compatibility with injection fluids.

  • Standard
    27 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for single use, gravity feed infusion sets for medical use in order to ensure their compatibility with containers for infusion solutions and intravenous equipment.
Secondary aims of this document are to provide guidance on specifications relating to the quality and performance of materials used in infusion sets and to present designations for infusion set components.

  • Standard
    25 pages
    English language
    sale 10% off
    e-Library read for
    1 day
  • Standard
    25 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 80601-2-56:2017 applies to the basic safety and essential performance of a clinical thermometer in combination with its accessories, hereafter referred to as me equipment. This document specifies the general and technical requirements for electrical clinical thermometers. This document applies to all electrical clinical thermometers that are used for measuring the body temperature of patients.
Clinical thermometers can be equipped with interfaces to accommodate secondary indicators, printing equipment, and other auxiliary equipment to create me systems. This document does not apply to auxiliary equipment.
Me equipment that measures a body temperature is inside the scope of this document.
ISO 80601-2-56:2017 does not specify the requirements for screening thermographs intended to be used for the individual non-invasive human febrile temperature screening of groups of individual humans under indoor environmental conditions, which are given in IEC 80601‑2‑59[4].
If a clause or subclause is specifically intended to be applicable to me equipment only, or to me systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to me equipment and to me systems, as relevant.
Hazards inherent in the intended physiological function of me equipment or me systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+A1:2012, 7.2.13 and 8.4.1.
NOTE          Additional information can be found in IEC 60601?1:2005+A1:2012, 4.2.

  • Standard
    65 pages
    English language
    sale 10% off
    e-Library read for
    1 day

2020-06-05-JO- BT N 11989 to delink EN ISO 80601-2-56:2017/prA1 from MDD and M/023
2019-03-07-JO - Awaiting for TC to resolve the negative assessment of the HAS consultant at ENQ stages. FV dispactch was due  on the 18th of January 2019.
DOW = DAV + 36 months

  • Amendment
    8 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 10555-6:2015 specifies requirements, performance, and user safety issues related to subcutaneous implanted ports and catheters for intravascular long-term use supplied in sterile condition and intended for single use.
ISO 10555-6:2015 does not specify requirements, performance, and user safety issues related to non-coring needles.

  • Amendment
    7 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements, including performance requirements, for plastics collapsible, non-vented, sterile containers (known as plastics containers) complete with collecting tube outlet port(s), integral needle, and with optional transfer tube(s), for the collection, storage, processing, transport, separation, and administration of blood and blood components. The plastics containers can contain anticoagulant and/or preservative solutions, depending on the application envisaged.
This document is also applicable to multiple units of plastics containers, e.g. to double, triple, quadruple, or multiple units.
Unless otherwise specified, all tests specified in this document apply to the plastics container as prepared ready for use.
This document is not applicable to plastics containers with an integrated filter.

  • Standard
    31 pages
    English language
    sale 10% off
    e-Library read for
    1 day

ISO 8637-1:2017 specifies requirements for haemodialysers, haemodiafilters, haemofilters and haemoconcentrators, hereinafter collectively referred to as "the device", for use in humans.
ISO 8637-1:2017 does not apply to:
-      extracorporeal blood circuits;
-      plasmafilters;
-      haemoperfusion devices;
-      vascular access devices;
-      blood pumps;
-      pressure monitors for the extracorporeal blood circuit;
-      air detection devices;
-      systems to prepare, maintain or monitor dialysis fluid;
-      systems or equipment intended to perform haemodialysis, haemodiafiltration, haemofiltration or haemoconcentration;
-      reprocessing procedures and equipment.
NOTE       Requirements for the extracorporeal blood circuit for haemodialysers, haemodiafilters and haemofilters are specified in ISO 8637‑2.

  • Standard
    30 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for enteral feeding systems comprising enteral giving sets, enteral extension sets, enteral syringes, enteral feeding catheters, and enteral accessories.
This document is not applicable to oral syringes.

  • Standard
    52 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies construction, design, performance requirements and test methods for medical face masks intended to limit the transmission of infective agents from staff to patients during surgical procedures and other medical settings with similar requirements. A medical face mask with an appropriate microbial barrier can also be effective in reducing the emission of infective agents from the nose and mouth of an asymptomatic carrier or a patient with clinical symptoms.
This European Standard is not applicable to masks intended exclusively for the personal protection of staff.
NOTE 1   Standards for masks for use as respiratory personal protective equipment are available.
NOTE 2   Annex A provides information for the users of medical face masks.

  • Standard
    23 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for single-use sharps containers intended to hold potentially hazardous sharps medical waste with or without sharps protection features, e.g. scalpel blades, trocars, hypodermic needles and syringes.
It is applicable to single-use sharps containers that are supplied complete by the manufacturer and to those that are supplied as components intended to be assembled by the user.
It is not applicable to reusable sharps containers or to the outer containers used in the transportation of filled single-use sharps containers.

  • Standard
    19 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This European Standard specifies information to be supplied to users and third party verifiers in addition to the usual labelling of medical devices (see EN 1041 and EN ISO 15223-1), concerning manufacturing and processing requirements. This European Standard gives information on the characteristics of single-use and reusable surgical gowns and surgical drapes used as medical devices for patients, clinical staff and equipment, intended to prevent the transmission of infective agents between clinical staff and patients during surgical and other invasive procedures. This European Standard specifies test methods for evaluating the identified characteristics of surgical drapes and gowns and sets performance requirements for these products.
EN 13795-1 does not cover requirements for resistance to penetration by laser radiation of products. Suitable test methods for resistance to penetration by laser radiation, together with an appropriate classification system, are given in EN ISO 11810.
EN 13795-1 does not cover requirements for incise drapes or films.
EN 13795-1 does not cover requirements for antimicrobial treatments for surgical gowns and drapes. Antimicrobial treatment may cause environmental risks such as resistance and pollution. However, antimicrobial treated surgical gowns and drapes fall under the scope of this standard with respect to their use as surgical gowns and drapes.

  • Standard
    33 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This European Standard specifies information to be supplied to users and third party verifiers in addition to the usual labelling of medical devices (see EN 1041 and EN ISO 15223-1), concerning manufacturing and processing requirements. This European Standard gives information on the characteristics of single-use and reusable clean air suits used as medical devices for clinical staff, intended to prevent the transmission of infective agents between clinical staff and patients during surgical and other invasive procedures. This European Standard specifies test methods for evaluating the identified characteristics of clean air suits and sets performance requirements for these products.

  • Standard
    29 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements to the safe handling and the physical, chemical and biological testing of plastic containers for parenterals.
This document is applicable to plastic containers for parenterals having one or more chambers and having a total nominal capacity in the range of 50 ml to 5 000 ml such as film bags or blow-moulded plastic bottles for direct administration of infusion (injection) solutions.
NOTE       In some countries, national or regional pharmacopoeias or other government regulations are legally binding and these requirements take precedence over this document.

  • Standard
    26 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements for sterile single-use hypodermic syringes made of plastic and rubber materials with or without needle, and intended for the aspiration of fluids or for the injection of fluids immediately after filling and of design such that the syringe can be rendered unusable after use.
This document is not applicable to syringes made of glass [specified in ISO 595 (withdrawn)], auto-disable syringes for fixed dose immunization (ISO 7886‑3) and syringes designed to be pre-filled. It does not address compatibility with injection fluids. Other standards can be applicable when syringes are used for any other intended purpose than those specified in this document.
NOTE       Syringes designed to reduce the risk of needle-stick injuries can also comply with this document with regard to their re-use prevention properties, but it is stressed that anti-needle-stick properties of syringes are not in themselves addressed in this document.

  • Standard
    19 pages
    English language
    sale 10% off
    e-Library read for
    1 day

This document specifies requirements and test methods for sterile urethral catheters for single use, with or without a balloon.
This document does not include drainage catheters covered by ISO 20697, e.g. ureteral catheters, nephrostomy catheters, and suprapubic catheters. This document also excludes ureteral stents.
NOTE       Ureteral stents are covered in ASTM F1828‑97.

  • Standard
    43 pages
    English language
    sale 10% off
    e-Library read for
    1 day