Standard Practice for Evaluating Performance Characteristics of Ultrasonic Pulse- Echo Examination Instruments and Systems Without the Use of Electronic Measurement Instruments

SCOPE
1.1 This practice describes procedures for evaluating the following performance characteristics of ultrasonic pulse-echo examination instruments and systems: Horizontal Limit and Linearity; Vertical Limit and Linearity; Resolution - Entry Surface and Far Surface; Sensitivity and Noise; Accuracy of Calibrated Gain Controls. Evaluation of these characteristics is intended to be used for comparing instruments and systems or, by periodic repetition, for detecting long-term changes in the characteristics of a given instrument or system that may be indicative of impending failure, and which, if beyond certain limits, will require corrective maintenance. Instrument characteristics measured in accordance with this practice are expressed in terms that relate to their potential usefulness for ultrasonic examinations. Instrument characteristics expressed in purely electronic terms may be measured as described in E1324.
1.2 Ultrasonic examination systems using pulsed-wave trains and A-scan presentation (rf or video) may be evaluated.
1.3 The procedures are applicable to shop or field conditions; additional electronic measurement instrumentation is not required.
1.4 This practice establishes no performance limits for examination systems; if such acceptance criteria are required, these must be specified by the using parties. Where acceptance criteria are implied herein they are for example only and are subject to more or less restrictive limits imposed by customer's and end user's controlling documents.
1.5 The specific parameters to be evaluated, conditions and frequency of test, and report data required, must also be determined by the user.
1.6 This practice may be used for the evaluation of a complete examination system, including search unit, instrument, interconnections, fixtures and connected alarm and auxiliary devices, primarily in cases where such a system is used repetitively without change or substitution. This practice is not intended to be used as a substitute for calibration or standardization of an instrument or system to inspect any given material. There are limitations to the use of standard reference blocks for that purpose.
1.7 Required test apparatus includes selected test blocks and a precision external attenuator (where specified) in addition to the instrument or system to be evaluated.
1.8 Precautions relating to the applicability of the procedures and interpretation of the results are included.
1.9 Alternate procedures, such as examples described in this document, or others, may only be used with customer approval.
1.10 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
09-Feb-2001
Technical Committee
Drafting Committee
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E317-01 - Standard Practice for Evaluating Performance Characteristics of Ultrasonic Pulse- Echo Examination Instruments and Systems Without the Use of Electronic Measurement Instruments
English language
12 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E 317 – 01
Standard Practice for
Evaluating Performance Characteristics of Ultrasonic Pulse-
Echo Examination Instruments and Systems without the
1
Use of Electronic Measurement Instruments
This standard is issued under the fixed designation E 317; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the Department of Defense.
1. Scope ment, interconnections, fixtures and connected alarm and
auxiliary devices, primarily in cases where such a system is
1.1 This practice describes procedures for evaluating the
used repetitively without change or substitution. This practice
following performance characteristics of ultrasonic pulse-echo
is not intended to be used as a substitute for calibration or
examination instruments and systems: Horizontal Limit and
standardization of an instrument or system to inspect any given
Linearity; Vertical Limit and Linearity; Resolution - Entry
material. There are limitations to the use of standard reference
Surface and Far Surface; Sensitivity and Noise; Accuracy of
2
blocks for that purpose.
Calibrated Gain Controls. Evaluation of these characteristics is
1.7 Requiredtestapparatusincludesselectedtestblocksand
intended to be used for comparing instruments and systems or,
a precision external attenuator (where specified) in addition to
by periodic repetition, for detecting long-term changes in the
the instrument or system to be evaluated.
characteristics of a given instrument or system that may be
1.8 Precautions relating to the applicability of the proce-
indicative of impending failure, and which, if beyond certain
dures and interpretation of the results are included.
limits, will require corrective maintenance. Instrument charac-
1.9 Alternateprocedures,suchasexamplesdescribedinthis
teristics measured in accordance with this practice are ex-
document,orothers,mayonlybeusedwithcustomerapproval.
pressed in terms that relate to their potential usefulness for
1.10 This standard does not purport to address all of the
ultrasonic examinations. Instrument characteristics expressed
safety concerns, if any, associated with its use. It is the
in purely electronic terms may be measured as described in
responsibility of the user of this standard to establish appro-
E 1324.
priate safety and health practices and determine the applica-
1.2 Ultrasonic examination systems using pulsed-wave
bility of regulatory limitations prior to use.
trains and A-scan presentation (rf or video) may be evaluated.
1.3 The procedures are applicable to shop or field condi-
2. Referenced Documents
tions; additional electronic measurement instrumentation is not
2.1 ASTM Standards:
required.
E 114 Practice for Ultrasonic Pulse-Echo Straight-Beam
1.4 This practice establishes no performance limits for
3
Examination by the Contact Method
examination systems; if such acceptance criteria are required,
E 127 Practice for Fabricating and Checking Aluminum
these must be specified by the using parties. Where acceptance
3
Alloy Ultrasonic Standard Reference Blocks
criteria are implied herein they are for example only and are
E 214 Practice for Immersed Ultrasonic Examination by the
subject to more or less restrictive limits imposed by customer’s
3
Reflection Method Using Pulsed Longitudinal Waves
and end user’s controlling documents.
E 428 Practice for Fabrication and Control of Steel Refer-
1.5 The specific parameters to be evaluated, conditions and
3
ence Blocks Used in Ultrasonic Inspection
frequency of test, and report data required, must also be
3
E 1316 Terminology for Nondestructive Examinations
determined by the user.
E 1324 Standard Guide for Measuring Some Electronic
1.6 This practice may be used for the evaluation of a
3
Characteristics of Ultrasonic Examination Instruments
complete examination system, including search unit, instru-
1
This practice is under the jurisdiction of ASTM Committee E07 on Nonde-
2
structive Testing and is the direct responsibility of Subcommittee E07.06 on Beck, K. H., “Limitations to the Use of Reference Blocks for Periodic and
Ultrasonic Testing Procedures. Preinspection Calibration of Ultrasonic Inspection Instruments and Systems,”
Current edition approved Feb. 10, 2001. Published April 2001. Originally Materials Evaluation, Vol 57, No. 3, March 1999.
3
published as E 317 – 67 T. Last previous edition E 317 – 94. Annual Book of ASTM Standards, Vol 03.03.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1

---------------------- Page: 1 --------------
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.