IEC 80601-2-35
(First edition – 2009)
Medical electrical equipment –
Part 2-35: Particular requirements for the basic safety
and essential performance of heating devices using blankets,
pads or mattresses and intended for heating in medical use
CO RRI G E NDUM 1
201.4.3.101 Additional requirements for ESSENTIAL PERFORMANCE
In Table 201.101, replace the reference to "201.12.3.103" by "201.12.3.102" as follows:
Table 201.101 – *Additional ESSENTIAL PERFORMANCE requirements
Requirement Subclause
ESSENTIAL PERFORMANCE requirement 1 201.12.4.104 or generation of a TECHNICAL
ALARM CONDITION in compliance with
201.12.3.102
201.15.4.2.1.101 *Temperature and overload control devices – Additional requirements
for application
In the fifth paragraph, replace the reference to "42 °C" by "41 °C".
201.15.4.101.1 Securing the fixation of heating elements and internal wiring
In the fifth paragraph, replace the reference to "201.15.4.1.102.2" by "201.15.4.101.2".