SIST EN ISO 22442-3:2008
Medical devices utilizing animal tissues and their derivatives - Part 3: Validation of the elimination and/or inactivation of viruses and transmissible spongiform encephalopathy (TSE) agents (ISO 22442-3:2007)
Medical devices utilizing animal tissues and their derivatives - Part 3: Validation of the elimination and/or inactivation of viruses and transmissible spongiform encephalopathy (TSE) agents (ISO 22442-3:2007)
This part of ISO 22442 specifies requirements for the validation of the elimination and/or inactivation of viruses and TSE agents during the manufacture of medical devices (excluding in vitro diagnostic medical devices) utilizing animal tissue or products derived from animal tissue, which are non-viable or have been rendered non-viable. It applies where required by the risk management process as described in ISO 22442-1. It does not cover other transmissible and non-transmissible agents.
Tierische Gewebe und deren Derivate nutzende Medizinprodukte - Teil 3: Validierung der Abreicherung und/oder Inaktivierung von Viren und Erregern der übertragbaren spongiösen Enzephalopathie (TSE) (ISO 22442-3:2007)
Diese Internationale Norm legt Anforderungen an die Validierung der Abreicherung und/oder Inaktivierung von Viren und TSE Erregern bei der Herstellung von Medizinprodukten (mit Ausnahme von in-vitro-Diagnostika) fest, die unter Verwendung von nicht lebensfähigen oder abgetöteten tierischen Geweben oder deren Derivaten hergestellt wurden. Sie schließt keine anderen übertragbaren oder nichtübertragbaren Krankheits¬erreger ein.
ANMERKUNG 1 Risikoanalyse und –management und konventionelle Sterilisationsverfahren, wenn sie bei der Behandlung tierischer Gewebe für Medizinprodukte angewendet werden, haben sich als nicht vollständig wirksam für die Inaktivierung der Erreger von übertragbarer spongiformer Enzephalopathie erwiesen. Eine selektive Auswahl bei der Gewebegewinnung ist demzufolge von großer Wichtigkeit (siehe ISO 22442 1 und ISO 22442 2).
ANMERKUNG 2 ISO 11135, ISO 11137, ISO 11737 1, ISO 13408, ISO 14160, ISO 14937 und ISO 17665 können für Bakterien, Schimmelpilze und Hefepilze verwendet werden (siehe Literaturhinweise).
Diese Internationale Norm umfasst nicht die Verwendung von menschlichen Geweben in Medizinprodukten.
Diese Internationale Norm legt kein Qualitätsmanagementsystem zur Kontrolle aller Produktionsschritte eines Medizin¬produktes fest.
ANMERKUNG 3 Es ist keine Anforderung dieser Norm, ein vollständiges Qualitätsmanagementsystem für die Herstel¬lung zu haben. Es wird auf die Normen für Qualitätsmanagementsysteme (siehe ISO 13485) zur Kontrolle aller Schritte der Produktion und Wiederaufarbeitung von Medizinprodukten hingewiesen.
Diese Internationale Norm berücksichtigt nicht die Auswirkungen irgendeines Verfahrens der Abreicherung und/oder Inaktivierung auf die Zweckbestimmung des Medizinprodukts.
Tissus animaux et leurs dérivés utilisés dan la fabrication des dispositifs médicaux - Partie 3: Validation de l'élimination et/ou de l'inactivation des virus et autres agents responsables d'encéphalopathie spongiforme transmissible (EST) (ISO 22442-3:2007)
L'ISO 22442-3:2007 spécifie les exigences relatives à la validation de l'élimination et/ou de l'inactivation des virus et agents EST au cours de la fabrication des dispositifs médicaux (à l'exception des dispositifs médicaux de diagnostic in vitro) utilisant des tissus animaux ou des produits dérivés de tissus animaux, non viables ou rendus non viables. Elle s'applique lorsque cela est requis par le processus de gestion des risques décrit dans l'ISO 22442-1. Elle ne couvre pas les autres agents transmissibles et non transmissibles.
Medicinski pripomočki, ki uporabljajo živalska tkiva in njihove derivate - 3. del: Validacija pri izločitvi in/ali inaktivaciji virusov in agensov transmisivnih spongiformnih encefalopatij (TSE) (ISO 22442-3:2007)
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
SIST EN ISO 22442-3:2008
01-april-2008
1DGRPHãþD
SIST EN 12442-3:2001
0HGLFLQVNLSULSRPRþNLNLXSRUDEOMDMRåLYDOVNDWNLYDLQQMLKRYHGHULYDWHGHO
9DOLGDFLMDSULL]ORþLWYLLQDOLLQDNWLYDFLMLYLUXVRYLQDJHQVRYWUDQVPLVLYQLK
VSRQJLIRUPQLKHQFHIDORSDWLM76(,62
Medical devices utilizing animal tissues and their derivatives - Part 3: Validation of the
elimination and/or inactivation of viruses and transmissible spongiform encephalopathy
(TSE) agents (ISO 22442-3:2007)
Tierische Gewebe und deren Derivate nutzende Medizinprodukte - Teil 3: Validierung
der Abreicherung und/oder Inaktivierung von Viren und Erregern der übertragbaren
spongiösen Enzephalopathie (TSE) (ISO 22442-3:2007)
Tissus animaux et leurs dérivés utilisés dan la fabrication des dispositifs médicaux -
Partie 3: Validation de l'élimination et/ou de l'inactivation des virus et autres agents
responsables d'encéphalopathie spongiforme transmissible (EST) (ISO 22442-3:2007)
Ta slovenski standard je istoveten z: EN ISO 22442-3:2007
ICS:
11.120.01
SIST EN ISO 22442-3:2008 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
---------------------- Page: 1 ----------------------
EUROPEAN STANDARD
EN ISO 22442-3
NORME EUROPÉENNE
EUROPÄISCHE NORM
December 2007
ICS 11.100.20 Supersedes EN 12442-3:2000
English Version
Medical devices utilizing animal tissues and their derivatives -
Part 3: Validation of the elimination and/or inactivation of viruses
and transmissible spongiform encephalopathy (TSE) agents
(ISO 22442-3:2007)
Dispositifs médicaux utilisant des tissus animaux et leurs Tierische Gewebe und deren Derivate, die zur Herstellung
dérivés - Partie 3: Validation de l'élimination et/ou de von Medizinprodukten eingesetzt werden - Teil 3:
l'inactivation des virus et autres agents responsables Validierung der Eliminierung und/oder Inaktivierung von
d'encéphalopathie spongiforme transmissible (EST) (ISO Viren und Erregern der übertragbaren spongiösen
22442-3:2007) Enzephalopathie (TSE) (ISO 22442-3:2007)
This European Standard was approved by CEN on 14 December 2007.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the CEN Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CEN member into its own language and notified to the CEN Management Centre has the same status as the
official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland,
France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal,
Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre: rue de Stassart, 36 B-1050 Brussels
© 2007 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 22442-3:2007: E
worldwide for CEN national Members.
---------------------- Page: 2 ----------------------
EN ISO 22442-3:2007 (E)
Contents Page
Foreword.3
2
---------------------- Page: 3 ----------------------
EN ISO 22442-3:2007 (E)
Foreword
This document (EN ISO 22442-3:2007) has been prepared by Technical Committee ISO/TC 194 "Biological
evaluation of medical devices" in collaboration with Technical Committee CEN/TC 316 “Medical devices
utilizing tissues” the secretariat of which is held by NBN.
This European Standard shall be given the status of a national standard, either by publication of an identical
text or by endorsement, at the latest by June 2008, and conflicting national standards shall be withdrawn at
the latest by June 2008.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent rights.
This document supersedes EN 12442-3:2000.
This document has been prepared under a mandate given to CEN by the European Commission and the
European Free Trade Association, and supports essential requirements of EC Directive(s).
This European Standard has been developed for medical devices regulated by the Medical Device Directive
93/42/EC as amended by 2003/32/EC (see Annex ZA). By analogy, it could be applied for active implantable
medical devices regulated by the Active Implantable Medical Device Directive 90/385/EC.
For relationship with EC Directive(s), see informative Annex ZA, which is an integral part of this document.
According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following
countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Cyprus, Czech
Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain,
Sweden, Switzerland and the United Kingdom.
Endorsement notice
The text of ISO 22442-3:2007 has been approved by CEN as a EN ISO 22442-3:2007 without any
modification.
3
---------------------- Page: 4 ----------------------
EN ISO 22442-3:2007 (E)
Annex ZA
(informative)
Relationship between this European Standard and the Essential Requirements
of EU Directive 93/42/EEC as amended by Commission Directive 2003/32/EC
This European Standard has been prepared under a mandate given to CEN by the European Commission
and the European Free Trade Association to provide a means of conforming to Essential Requirements of the
New Approach Directive 93/42/EEC, concerning medical devices, as amended by Commission Directive
2003/32/EC in relation to detailed specifications regarding requirements for medical devices utilizing tissues of
animal origin.
Once this standard is cited in the Official Journal of the European Communities under that Directive and has
been implemented as a national standard in at least one Member State, compliance with the clauses of this
standard given in table ZA confers, within the limits of the scope of this International Standard, a presumption
of conformity with the corresponding Essential Requirements of that Directive and associated EFTA
regulations.
Table ZA — Correspondence between this European Standard and Directive 93/42/EEC as amended by
Commission Directive 2003/32/EC
Clause(s)/subclause(s) of this Essential requirements (ERs) of Qualifying remarks/Notes
International Standard Directive 93/42/EEC as amended
by Commission Directive
2003/32/EC
4, 5, 6, 7, 8, 9, Annex A Annex I, 7.1, 7.2, 8.1, 8.2
4, 5, 6, 7, 8, 9, Annex A Annex of Commission Directive
2003/32/EC
WARNING — Other requirements and other EU Directives may be applicable to the product(s) falling
within the scope of this standard.
4
---------------------- Page: 5 ----------------------
INTERNATIONAL ISO
STANDARD 22442-3
First edition
2007-12-15
Medical devices utilizing animal tissues
and their derivatives —
Part 3:
Validation of the elimination and/or
inactivation of viruses and transmissible
spongiform encephalopathy (TSE) agents
Dispositifs médicaux utilisant des tissus animaux et leurs dérivés —
Partie 3: Validation de l'élimination et/ou de l'inactivation des virus et
autres agents responsables d'encéphalopathie spongiforme
transmissible (EST)
Reference number
ISO 22442-3:2007(E)
©
ISO 2007
---------------------- Page: 6 ----------------------
ISO 22442-3:2007(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
COPYRIGHT PROTECTED DOCUMENT
© ISO 2007
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2007 – All rights reserved
---------------------- Page: 7 ----------------------
ISO 22442-3:2007(E)
Contents Page
Foreword. iv
Introduction . v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions. 2
4 General requirements. 3
4.1 Risk management . 3
4.2 Sourcing and manufacturing process. 3
4.3 General requirements related to validation . 3
5 Literature review . 4
5.1 Conduct of the literature review. 4
5.2 Application of literature review output. 4
5.3 Viruses . 4
5.4 TSE agents . 4
6 Elimination and/or inactivation study of viruses and TSE agents . 5
6.1 General. 5
6.2 Protocol . 5
6.3 Conduct of the study. 6
6.4 Interpretation of data. 6
7 Final report . 6
8 Review of final report . 6
9 Routine monitoring and control of critical process parameters . 6
Annex A (normative) Requirements related to literature review. 7
Annex B (informative) Guidance on the elimination and/or inactivation study for viruses . 11
Annex C (informative) Guidance on the elimination and/or inactivation study for TSE agents . 16
Annex D (informative) Guidance on scaling down . 17
Annex E (informative) Statistical evaluation of virus titres and reduction factors and assessment
of their validity . 18
Annex F (informative) Calculation of reduction factors . 19
Annex G (informative) Probability of detection of agents at low concentrations. 20
Bibliography . 21
© ISO 2007 – All rights reserved iii
---------------------- Page: 8 ----------------------
ISO 22442-3:2007(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 22442-3 was prepared by Technical Committee ISO/TC 194, Biological evaluation of medical devices,
Subcommittee SC 1, Tissue product safety.
ISO 22442 consists of the following parts, under the general title Medical devices utilizing animal tissues and
their derivatives:
⎯ Part 1: Application of risk management
⎯ Part 2: Controls on sourcing, collection and handling
⎯ Part 3: Validation of the elimination and/or inactivation of viruses and transmissible spongiform
encephalopathy (TSE) agents
iv © ISO 2007 – All rights reserved
---------------------- Page: 9 ----------------------
ISO 22442-3:2007(E)
Introduction
Certain medical devices utilize materials of animal origin.
Animal tissues and their derivatives are used in the design and manufacture of medical devices to provide
performance characteristics that were chosen for advantages over non-animal based materials. The range
and quantities of materials of animal origin in medical devices vary. These materials can comprise a major
part of the device (e.g. bovine/porcine heart valves, bone substitutes for use in dental or orthopaedic
applications, haemostatic devices), can be a product coating or impregnation (e.g. collagen, gelatine, heparin),
or can be used in the device manufacturing process (e.g. tallow derivatives such as oleates and stearates,
foetal calf serum, enzymes, culture media).
It is important to be aware that the exposure to a properly validated and accurately controlled method of viral
and TSE inactivation/elimination is not the only factor associated with demonstrating product safety. Attention
has also to be given to a number of factors including sourcing, collecting, handling, storage, processing,
testing of tissues and/or cells of animal origin, and to the control of the environment in which the product is
manufactured, assembled and packaged. The manufacturer should consider the fact that each manufacturing
phase can contribute to contamination as well as elimination and/or inactivation of viruses and TSE agents.
For the safety of medical devices there are two complementary approaches (see ISO 22442-1) that can be
adopted to control the potential contamination of tissues. These typically are:
a) selecting source material for minimal contamination with viruses and/or TSE agents (see ISO 22442-1
and ISO 22442-2);
b) providing valid scientific evidence to demonstrate the ability of the production processes to eliminate or
inactivate viruses and/or TSE agents (this part of ISO 22442).
Requirements for a quality system for medical devices for regulatory use are specified in ISO 13485. The
standards for quality management systems recognize that, for certain processes used in manufacturing, the
effectiveness of that process cannot be fully verified by subsequent inspection and testing of the product. The
elimination and/or inactivation of viruses and TSE agents is an example of a special process because process
efficacy cannot be verified by inspection and testing of the product. For this reason, the following need to be
considered in particular:
⎯ definition of the process(es) and materials to be used;
⎯ adequate inactivation validation before routine use;
⎯ performance monitoring of the process during manufacture;
⎯ appropriate equipment maintenance;
⎯ staff training, etc.
Historically there have been many instances of unknown or unsuspected viral contamination during
manufacture. For this reason, evaluation of the manufacturing process can provide a measure of confidence
that a wide number of viruses, including unknown pathogenic viruses are eliminated. Similar principles may
apply to TSE agents.
NOTE To show compliance with this part of ISO 22442, its specified requirements should be fulfilled. The guidance
given in the Notes and informative annexes is not normative and is not provided as a checklist for auditors.
© ISO 2007 – All rights reserved v
---------------------- Page: 10 ----------------------
INTERNATIONAL STANDARD ISO 22442-3:2007(E)
Medical devices utilizing animal tissues and their derivatives —
Part 3:
Validation of the elimination and/or inactivation of viruses and
transmissible spongiform encephalopathy (TSE) agents
1 Scope
This part of ISO 22442 specifies requirements for the validation of the elimination and/or inactivation of viruses
and TSE agents during the manufacture of medical devices (excluding in vitro diagnostic medical devices)
utilizing animal tissue or products derived from animal tissue, which are non-viable or have been rendered
non-viable. It applies where required by the risk management process as described in ISO 22442-1. It does
not cover other transmissible and non-transmissible agents.
NOTE 1 Analysis and management of risk is described in ISO 22442-1. Conventional processes used for sterilization,
when used for the treatment of animal tissues for medical devices, have not been shown to be completely effective in
inactivating the causative agents of transmissible spongiform encephalopathy. Selective sourcing is extremely important
(see ISO 22442-1 and ISO 22442-2).
NOTE 2 ISO 11135, ISO 11137, ISO 11737-1, ISO 13408, ISO 14160, ISO 14937 and ISO 17665 may be relevant for
bacteria, moulds and yeast (see Bibliography).
This part of ISO 22442 does not cover the utilization of human tissues in medical devices.
This part of ISO 22442 does not specify a quality management system for the control of all stages of
production of medical devices.
NOTE 3 It is not a requirement of this part of ISO 22442 to have a full quality management system during manufacture,
but it does specify requirements for some of the elements of a quality management system. Attention is drawn to the
standards for quality management systems (see ISO 13485) that control all stages of production or reprocessing of
medical devices. The quality management system elements that are required by this part of ISO 22442 can form part of a
quality management system conforming to ISO 13485.
This part of ISO 22442 does not consider the effect of any method of elimination and/or inactivation on the
suitability of the medical device for its intended use.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 22442-1:2007, Medical devices utilizing animal tissues and their derivatives — Part 1: Application of risk
management
ISO 22442-2, Medical devices utilizing animal tissues and their derivatives — Part 2: Controls on sourcing,
collection and handling
© ISO 2007 – All rights reserved 1
---------------------- Page: 11 ----------------------
ISO 22442-3:2007(E)
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 22442-1 and the following apply.
3.1
model TSE agent
TSE agent that displays a known resistance to physical and/or chemical processing used as reference by
analogy for the inactivation of relevant TSE agents, and thereby demonstrating the effectiveness of the
process used for inactivation
3.2
model virus
virus that displays a known resistance to physical and/or chemical processing used as reference by analogy
for the inactivation of relevant viruses, and thereby demonstrating the effectiveness of the process used for
inactivation
NOTE This includes viral models (RNA, DNA, enveloped, non-enveloped) and bacteriophage models.
3.3
overall reduction factor
sum of the reduction factors of the individual process steps
3.4
permissive cell
cell that can become infected with the virus under study and in which that virus replicates
3.5
reduction factor
ratio of the virus or TSE agent load in the relevant material used or the device prior to the inactivation or
elimination step and the virus or TSE agent load after the inactivation or elimination step when it is ready for
the next step in the manufacturing process, expressed as the number of ten fold reduction (log )
10
3.6
relevant TSE agent
TSE agent known to, or likely to, contaminate the source material or other materials used in the manufacturing
process
3.7
relevant virus
virus known to, or likely to, contaminate the source material or other materials used in the manufacturing
process
3.8
revalidation
set of documented procedures to confirm an established validation
3.9
scaled down process
scaling down
process at a specified reduced scale which simulates the performance parameters as used in the full scale
production process
3.10
sterilization
validated process used to render a product free of all forms of viable microorganisms
2 © ISO 2007 – All rights reserved
---------------------- Page: 12 ----------------------
ISO 22442-3:2007(E)
3.11
validation
documented procedure for obtaining, recording and interpreting the results required to establish that a process
will consistently yield product complying with predetermined specifications
[ISO/TS 11139:2006, definition 2.55]
4 General requirements
4.1 Risk management
Analysis and management of risk shall be carried out in accordance with ISO 22442-1.
Due account shall be taken of manufacturing processes that are considered to be effective for certain animal
materials as discussed in Annex C of ISO 22442-1:2007.
4.2 Sourcing and manufacturing process
A documented system shall be established and maintained to control the source of raw materials of animal
origin. ISO 22442-2 shall be used to meet this requirement as far as applicable.
The manufacturing process shall be established to minimize the load of viruses and TSE agents in starting
materials, intermediate products and finished products.
Appropriate documented protocols and procedures shall be established to ensure that the validated
processing parameters will be applied during the routine manufacturing processes.
NOTE Employing a quality management system complying with ISO 13485 could be used to meet the requirements
of this subclause.
4.3 General requirements related to validation
4.3.1 Documented procedures
The documented procedures and requirements of this part of ISO 22442 shall be implemented.
Documentation and records shall be reviewed and approved by designated personnel (see 4.3.2).
Procedures for any literature review and/or any inactivation study shall be documented and records shall be
retained for a period defined by the manufacturer.
4.3.2 Personnel
Responsibility for the implementation of this part of ISO 22442 shall be assigned to qualified personnel.
The requirements for the qualification, training or experience of personnel shall be documented and
appropriate to the individual’s work, responsibility and authority.
NOTE The level of qualification, training and experience required by personnel at various levels depends upon the
activities being performed.
4.3.3 Calibration
An effective system shall be established, documented and maintained for the calibration of all controlling,
indicating and recording instruments used for validation.
© ISO 2007 – All rights reserved 3
---------------------- Page: 13 ----------------------
ISO 22442-3:2007(E)
4.3.4 Equipment
Appropriate equipment as specified in a protocol shall be used. All equipment requiring planned maintenance
shall be maintained in accordance with documented procedures. Records of maintenance shall be retained.
In particular, any equipment shall be capable of delivering its intended process within defined limits. In
addition, if the equipment used during validation is not identical to that used in normal production cycles,
adequate documentation shall be available to demonstrate that the performance parameters are equivalent to
those used in the production cycle.
4.3.5 Experimental systems
Additional parts of the experimental systems used for validation studies such as chemicals, cell systems and
laboratory animals shall be adequately identified, justified, controlled and documented.
5 Literature review
5.1 Conduct of the literature review
A literature review shall be performed as specified in Annex A, in order to identify and analyse data on the
elimination and/or inactivation of viruses and TSE agents.
5.2 Application of literature review output
Technical information from the literature review shall be used in optimizing the design of an inactivation and/or
elimination study.
Any extrapolation based on the inactivation of viruses and TSE agents shall be justified and documented.
Intrinsic variability of materials of animal origin utilized in medical devices and of manufacturing processes can
lead to misinterpretation of the validity of published data and shall be taken into account.
5.3 Viruses
The manufacturer shall demonstrate whether the literature review provides an indication of which inactivation
and/or elimination steps are likely to be effective. A literature review is a prerequisite to performing a viral
inactivation study. In exceptional cases, if a manufacturer chooses not to perform a study, this shall be
justified and documented.
If the available information does not support the elimination and/or inactivation of viruses, then an alternative
risk management strategy shall be implemented (see ISO 22442-1).
5.4 TSE agents
The literature review shall consider which of the published methods for elimination and/or inactivation of TSE
agents are likely to be suitable for the medical device under consideration. In particular, the materials of
animal origin and manufacturing processes referred to in the literature shall be comparable to those used for
the medical device under consideration (see Annex A). A validated inactivation study shall be performed when
the comparability of materials and processes cannot be demonstrated or specific claims are made for
inactivation of TSE agents by the manufacturer (see Clause 6).
If the available information does not support the elimination and/or inactivation of TSE agents, then an
alternative risk management strategy shall be implemented (see ISO 22442-1).
Special considerations for the manufacture of some animal ma
...
Questions, Comments and Discussion