SIGNIFICANCE AND USE
5.1 The presence of cell growth medium complicates a direct analysis of cells with SIMS. Attempts to wash out the nutrient medium results in the exposure of cells to unphysiological reagents that may also alter their chemical composition. This obstacle is overcome by using a sandwich freeze-fracture method (1). This cryogenic method has provided a unique way of sampling individual cells in their native state for SIMS analysis.
5.2 The procedure described here has been successfully used for imaging Na+ and K+ ion transport (3), calcium alterations in stimulated cells (4,5), and localization of therapeutic drugs and isotopically labeled molecules in single cells (6). The frozen freeze-dried cells prepared according to this method have been checked for SIMS matrix effects (7). Ion image quantification has also been achieved in this sample type (8).
5.3 The procedure described here is amenable to a wide variety of cell cultures and provides a way for studying the response of individual cells for chemical alterations in the state of health and disease and localization of isotopically-labeled molecules and theraputic drugs in cell culture models.
SCOPE
1.1 This guide provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a cryogenic method for analyzing individual tissue culture cells growing in vitro. This guide is suitable for frozen-hydrated and frozen-freeze-dried sample types. Included are procedures for correlating optical, laser scanning confocal and secondary electron microscopies to complement SIMS analysis.
1.2 This guide is not suitable for cell cultures that do not attach to the substrate.
1.3 This guide is not suitable for any plastic embedded cell culture specimens.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Guide3 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Pressing cryosections flat onto a conducting substrate has been one of the most challenging problems in SIMS analysis of cryogenically prepared tissue specimens. Frozen cryosections often curl or peel off, or both, from the substrate during freeze-drying. The curling of cryosections results in an uneven sample surface for SIMS analysis. Furthermore, if freeze-dried cryosections are not attached tightly to the substrate, the impact of the primary ion beam may result in further curling and even dislodging of the cryosection from the substrate. These problems render SIMS analysis difficult, frustrating and time consuming. The use of indium as a substrate for pressing cryosections flat has provided a practical approach for analyzing cryogenically prepared tissue specimens (1).4
5.2 The procedure described herein has been successfully used for SIMS imaging of calcium and magnesium transport and localization of anticancer drugs in animal models (2, 3, 4, 5).
5.3 The procedure described here is amenable to soft tissues of both animal and plant origin.
SCOPE
1.1 This practice provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a method for analyzing tissue cryosections in the imaging mode of the instrument. This practice is suitable for frozen-freeze-dried and frozen-hydrated cryosection analysis.
1.2 This practice does not describe methods for optimal freezing of the specimen for immobilizing diffusible chemical species in their native intracellular sites.
1.3 This practice does not describe methods for obtaining cryosections from a frozen specimen.
1.4 This practice is not suitable for any plastic embedded tissues.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard3 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Sputter crater depth measurements are performed in order to determine a sputter rate (depth/time) for each matrix sputtered during a sputter depth profile or similar in-depth type analyses. From sputter rate values, a linear depth scale can be calculated and displayed for the sputter depth profile.
4.2 Data obtained from surface profilometry are useful in monitoring instrumental parameters (for example, raster size, shape, and any irregularities in topography of the sputtered crater) used for depth profiles.
SCOPE
1.1 This guide covers the preferred procedure for acquiring and post-processing of sputter crater depth measurements. This guide is limited to stylus-type surface profilometers equipped with a stage, stylus, associated scan and sensing electronics, video system for sample and scan alignment, and computerized system.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Guide3 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Although it would be desirable to measure the extent of profile distortion in any unknown sample by using a standard sample and this guide, measurements of interface width (profile distortion) can be unique to every sample composition (1, 2).4 This guide, describes a method that determines the unique width of a particular interface for the chosen set of operating conditions. It is intended to provide a method for checking on proper or consistent, or both, instrument performance. Periodic analysis of the same sample followed by a measurement of the interface width, in accordance with this guide, will provide these checks.
5.2 The procedure described in this guide is adaptable to any layered sample with an interface between layers in which a nominated element is present in one layer and absent from the other. It has been shown that for SIMS in particular (3, 4) and for surface analysis in general (5, 6), only rigorous calibration methods can determine accurate interface widths. Such procedures are prohibitively time-consuming. Therefore the interface width measurement obtained using the procedure described in this guide may contain significant systematic error (7). Therefore, this measure of interface width may have no relation to similar measures made with other methods. However, this does not diminish its use as a check on proper or consistent instrument performance, or both.
5.3 This guide can be used for both elemental and molecular depth profiles, provided that the materials have constant sputter rates throughout the depth of the overlayer, and minimal interlayer mixing is occurring. For more detailed information regarding measurements of interface widths during organic depth profiling, please see Mahoney (8).
SCOPE
1.1 This guide provides the SIMS analyst with a method for determining the width of interfaces from SIMS sputtering data obtained from analyses of layered specimens (both organic and inorganic). This guide does not apply to data obtained from analyses of specimens with thin markers or specimens without interfaces such as ion-implanted specimens.
1.2 This guide does not describe methods for the optimization of interface width or the optimization of depth resolution.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Guide3 pagesEnglish language
SIGNIFICANCE AND USE
5.1 This practice is intended for use in reporting the experimental and data reduction procedures described in other publications.
SCOPE
1.1 This practice provides the minimum information necessary to describe the instrumental, experimental, and data reduction procedures used in acquiring and reporting secondary ion mass spectrometry (SIMS) mass spectral data.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard2 pagesEnglish language
SIGNIFICANCE AND USE
5.1 This practice is to be used for reporting the experimental and data reduction procedures to be described with the publication of the data.
SCOPE
1.1 This practice lists the minimum information necessary to describe the instrumental, experimental, and data reduction procedures used in acquiring and reporting images generated by secondary ion mass spectrometry (SIMS).
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard3 pagesEnglish language
SIGNIFICANCE AND USE
5.1 This practice is used for reporting the experimental conditions as specified in Section 6 in the “Methods” or “Experimental” sections of other publications (subject to editorial restrictions).
5.2 The report would include specific conditions for each data set, particularly, if any parameters are changed for different sputter depth profile data sets in a publication. For example, footnotes of tables or figure captions would be used to specify differing conditions.
SCOPE
1.1 This practice covers the information needed to describe and report instrumentation, specimen parameters, experimental conditions, and data reduction procedures. SIMS sputter depth profiles can be obtained using a wide variety of primary beam excitation conditions, mass analysis, data acquisition, and processing techniques (1-4).2
1.2 Limitations—This practice is limited to conventional sputter depth profiles in which information is averaged over the analyzed area in the plane of the specimen. Ion microprobe or microscope techniques permitting lateral spatial resolution of secondary ions within the analyzed area, for example, image depth profiling, are excluded.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard3 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Electron multipliers are commonly used in pulse-counting mode to detect ions from magnetic sector mass spectrometers. The electronics used to amplify, detect and count pulses from the electron multipliers always have a characteristic time interval after the detection of a pulse, during which no other pulses can be counted. This characteristic time interval is known as the “dead time.” The dead time has the effect of reducing the measured count rate compared with the “true” count rate.
5.2 In order to measure count rates accurately over the entire dynamic range of a pulse counting detector, such as an electron multiplier, the dead time of the entire pulse counting system must be well known. Accurate count rate measurement forms the basis of isotopic ratio measurements as well as elemental abundance determinations.
5.3 The procedure described herein has been successfully used to determine the dead time of counting systems on SIMS instruments.6 The accurate determination of the dead time by this method has been a key component of precision isotopic ratio measurements made by SIMS.
SCOPE
1.1 This practice provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a method for determining the dead time of the pulse-counting detection systems on the instrument. This practice also allows the analyst to determine whether the apparent dead time is independent of count rate.
1.2 This practice is applicable to most types of mass spectrometers that have pulse-counting detectors.
1.3 This practice does not describe methods for precise or accurate isotopic ratio measurements.
1.4 This practice does not describe methods for the proper operation of pulse counting systems and detectors for mass spectrometry.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard3 pagesEnglish language
SIGNIFICANCE AND USE
5.1 The presence of cell growth medium complicates a direct analysis of cells with SIMS. Attempts to wash out the nutrient medium results in the exposure of cells to unphysiological reagents that may also alter their chemical composition. This obstacle is overcome by using a sandwich freeze-fracture method (1). This cryogenic method has provided a unique way of sampling individual cells in their native state for SIMS analysis.
5.2 The procedure described here has been successfully used for imaging Na+ and K+ ion transport (3), calcium alterations in stimulated cells (4,5), and localization of therapeutic drugs and isotopically labeled molecules in single cells (6). The frozen freeze-dried cells prepared according to this method have been checked for SIMS matrix effects (7). Ion image quantification has also been achieved in this sample type (8).
5.3 The procedure described here is amenable to a wide variety of cell cultures and provides a way for studying the response of individual cells for chemical alterations in the state of health and disease and localization of isotopically-labeled molecules and theraputic drugs in cell culture models.
SCOPE
1.1 This guide provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a cryogenic method for analyzing individual tissue culture cells growing in vitro. This guide is suitable for frozen-hydrated and frozen-freeze-dried sample types. Included are procedures for correlating optical, laser scanning confocal and secondary electron microscopies to complement SIMS analysis.
1.2 This guide is not suitable for cell cultures that do not attach to the substrate.
1.3 This guide is not suitable for any plastic embedded cell culture specimens.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide3 pagesEnglish language
- Guide3 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Pressing cryosections flat onto a conducting substrate has been one of the most challenging problems in SIMS analysis of cryogenically prepared tissue specimens. Frozen cryosections often curl or peel off, or both, from the substrate during freeze-drying. The curling of cryosections results in an uneven sample surface for SIMS analysis. Furthermore, if freeze-dried cryosections are not attached tightly to the substrate, the impact of the primary ion beam may result in further curling and even dislodging of the cryosection from the substrate. These problems render SIMS analysis difficult, frustrating and time consuming. The use of indium as a substrate for pressing cryosections flat has provided a practical approach for analyzing cryogenically prepared tissue specimens (1).4
5.2 The procedure described herein has been successfully used for SIMS imaging of calcium and magnesium transport and localization of anticancer drugs in animal models (2, 3, 4, 5).
5.3 The procedure described here is amenable to soft tissues of both animal and plant origin.
SCOPE
1.1 This practice provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a method for analyzing tissue cryosections in the imaging mode of the instrument. This practice is suitable for frozen-freeze-dried and frozen-hydrated cryosection analysis.
1.2 This practice does not describe methods for optimal freezing of the specimen for immobilizing diffusible chemical species in their native intracellular sites.
1.3 This practice does not describe methods for obtaining cryosections from a frozen specimen.
1.4 This practice is not suitable for any plastic embedded tissues.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard2 pagesEnglish language
- Standard2 pagesEnglish language
SIGNIFICANCE AND USE
Sputter crater depth measurements are performed in order to determine a sputter rate (depth/time) for each matrix sputtered during a sputter depth profile or similar in-depth type analyses. From sputter rate values, a linear depth scale can be calculated and displayed for the sputter depth profile.
Data obtained from surface profilometry are useful in monitoring instrumental parameters (for example, raster size, shape, and any irregularities in topography of the sputtered crater) used for depth profiles.
SCOPE
1.1 This guide covers the preferred procedure for acquiring and post-processing of sputter crater depth measurements. This guide is limited to stylus-type surface profilometers equipped with a stage, stylus, associated scan and sensing electronics, video system for sample and scan alignment, and computerized system.
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide3 pagesEnglish language
- Guide3 pagesEnglish language
SIGNIFICANCE AND USE
This practice is intended for use in reporting the experimental and data reduction procedures described in other publications.
SCOPE
1.1 This practice provides the minimum information necessary to describe the instrumental, experimental, and data reduction procedures used in acquiring and reporting secondary ion mass spectrometry (SIMS) mass spectral data.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard2 pagesEnglish language
- Standard2 pagesEnglish language
SIGNIFICANCE AND USE
This practice is to be used for reporting the experimental and data reduction procedures to be described with the publication of the data.
SCOPE
1.1 This practice lists the minimum information necessary to describe the instrumental, experimental, and data reduction procedures used in acquiring and reporting images generated by secondary ion mass spectrometry (SIMS).
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard3 pagesEnglish language
SIGNIFICANCE AND USE
This practice is used for reporting the experimental conditions as specified in Section 6 in the “Methods” or “Experimental” sections of other publications (subject to editorial restrictions).
The report would include specific conditions for each data set, particularly, if any parameters are changed for different sputter depth profile data sets in a publication. For example, footnotes of tables or figure captions would be used to specify differing conditions.
SCOPE
1.1 This practice covers the information needed to describe and report instrumentation, specimen parameters, experimental conditions, and data reduction procedures. SIMS sputter depth profiles can be obtained using a wide variety of primary beam excitation conditions, mass analysis, data acquisition, and processing techniques (1-4).
1.2 Limitations—This practice is limited to conventional sputter depth profiles in which information is averaged over the analyzed area in the plane of the specimen. Ion microprobe or microscope techniques permitting lateral spatial resolution of secondary ions within the analyzed area, for example, image depth profiling, are excluded.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard3 pagesEnglish language
- Standard3 pagesEnglish language
SIGNIFICANCE AND USE
Although it would be desirable to measure the extent of profile distortion in any unknown sample by using a standard sample and this guide, measurements of interface width (profile distortion) can be unique to every sample composition (1, 2, 3).3 This guide, that describes a method that determines the unique width of a particular interface for the chosen set of operating conditions. It is primarily intended to provide a method for checking on proper or consistent, or both, instrument performance. Periodic analysis of the same sample followed by a measurement of the interface width, in accordance with this guide, will provide these checks.
The procedure described in this guide is adaptable to any layered sample with an interface between layers in which a nominated element is present in one layer and absent from the other. It has been shown that for SIMS in particular (4, 5) and for surface analysis in general (6, 7), only rigorous calibration methods can determine accurate interface widths. Such procedures are prohibitively time-consuming. Therefore the interface width measurement obtained using the procedure described in this guide may contain significant systematic error (8). Therefore, this measure of interface width may have no relation to similar measures made with other methods. However, this does not diminish its use as a check on proper or consistent instrument performance, or both.
SCOPE
1.1 This guide provides the SIMS analyst with a method for determining the width of interfaces from SIMS sputtering data obtained from analyses of layered specimens. This guide does not apply to data obtained from analyses of specimens with thin markers or specimens without interfaces such as ion-implanted specimens.
1.2 This guide does not describe methods for the optimization of interface width or the optimization of depth resolution.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide3 pagesEnglish language
- Guide3 pagesEnglish language
SIGNIFICANCE AND USE
Electron multipliers are commonly used in pulse-counting mode to detect ions from magnetic sector mass spectrometers. The electronics used to amplify, detect and count pulses from the electron multipliers always have a characteristic time interval after the detection of a pulse, during which no other pulses can be counted. This characteristic time interval is known as the “dead time.” The dead time has the effect of reducing the measured count rate compared with the “true” count rate.
In order to measure count rates accurately over the entire dynamic range of a pulse counting detector, such as an electron multiplier, the dead time of the entire pulse counting system must be well known. Accurate count rate measurement forms the basis of isotopic ratio measurements as well as elemental abundance determinations.
The procedure described herein has been successfully used to determine the dead time of counting systems on SIMS instruments. The accurate determination of the dead time by this method has been a key component of precision isotopic ratio measurements made by SIMS.
SCOPE
1.1 This practice provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a method for determining the dead time of the pulse-counting detection systems on the instrument. This practice also allows the analyst to determine whether the apparent dead time is independent of count rate.
1.2 This practice is applicable to most types of mass spectrometers that have pulse-counting detectors.
1.3 This practice does not describe methods for precise or accurate isotopic ratio measurements.
1.4 This practice does not describe methods for the proper operation of pulse counting systems and detectors for mass spectrometry.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard3 pagesEnglish language
- Standard3 pagesEnglish language
SIGNIFICANCE AND USE
Sputter crater depth measurements are performed in order to determine a sputter rate (depth/time) for each matrix sputtered during a sputter depth profile or similar in-depth type analyses. From sputter rate values, a linear depth scale can be calculated and displayed for the sputter depth profile.
Data obtained from surface profilometry are useful in monitoring instrumental parameters (for example, raster size, shape, and any irregularities in topography of the sputtered crater) used for depth profiles.
SCOPE
1.1 This guide covers the preferred procedure for acquiring and post-processing of sputter crater depth measurements. This guide is limited to stylus-type surface profilometers equipped with a stage, stylus, associated scan and sensing electronics, video system for sample and scan alignment, and computerized system.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide3 pagesEnglish language
SIGNIFICANCE AND USE
This practice is to be used for reporting the experimental and data reduction procedures to be described with the publication of the data.
SCOPE
1.1 This practice lists the minimum information necessary to describe the instrumental, experimental, and data reduction procedures used in acquiring and reporting images generated by secondary ion mass spectrometry (SIMS).
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard2 pagesEnglish language
SIGNIFICANCE AND USE
Although it would be desirable to measure the extent of profile distortion in any unknown sample by using a standard sample and this guide, measurements of interface width (profile distortion) can be unique to every sample composition (1, 2, 3).3 This guide, that describes a method that determines the unique width of a particular interface for the chosen set of operating conditions. It is primarily intended to provide a method for checking on proper or consistent, or both, instrument performance. Periodic analysis of the same sample followed by a measurement of the interface width, in accordance with this guide, will provide these checks.
The procedure described in this guide is adaptable to any layered sample with an interface between layers in which a nominated element is present in one layer and absent from the other. It has been shown that for SIMS in particular (4, 5) and for surface analysis in general (6, 7), only rigorous calibration methods can determine accurate interface widths. Such procedures are prohibitively time-consuming. Therefore the interface width measurement obtained using the procedure described in this guide may contain significant systematic error (8). Therefore, this measure of interface width may have no relation to similar measures made with other methods. However, this does not diminish its use as a check on proper or consistent instrument performance, or both.
SCOPE
1.1 This guide provides the SIMS analyst with a method for determining the width of interfaces from SIMS sputtering data obtained from analyses of layered specimens. This guide does not apply to data obtained from analyses of specimens with thin markers or specimens without interfaces such as ion-implanted specimens.
1.2 This guide does not describe methods for the optimization of interface width or the optimization of depth resolution.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide2 pagesEnglish language
SIGNIFICANCE AND USE
This practice is used for reporting the experimental conditions as specified in Section 6 in the “Methods” or “Experimental” sections of other publications (subject to editorial restrictions).
The report would include specific conditions for each data set, particularly, if any parameters are changed for different sputter depth profile data sets in a publication. For example, footnotes of tables or figure captions would be used to specify differing conditions.
SCOPE
1.1 This practice covers the information needed to describe and report instrumentation, specimen parameters, experimental conditions, and data reduction procedures. SIMS sputter depth profiles can be obtained using a wide variety of primary beam excitation conditions, mass analysis, data acquisition, and processing techniques (1-4).
1.2 Limitations This practice is limited to conventional sputter depth profiles in which information is averaged over the analyzed area in the plane of the specimen. Ion microprobe or microscope techniques permitting lateral spatial resolution of secondary ions within the analyzed area, for example, image depth profiling, are excluded.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard3 pagesEnglish language
SIGNIFICANCE AND USE
The presence of cell growth medium complicates a direct analysis of cells with SIMS. Attempts to wash out the nutrient medium results in the exposure of cells to unphysiological reagents that may also alter their chemical composition. This obstacle is overcome by using a sandwich freeze-fracture method (1). This cryogenic method has provided a unique way of sampling individual cells in their native state for SIMS analysis.
The procedure described here has been successfully used for imaging Na+ and K+ ion transport (3), calcium alterations in stimulated cells (4,5), and localization of therapeutic drugs and isotopically labeled molecules in single cells (6). The frozen freeze-dried cells prepared according to this method have been checked for SIMS matrix effects (7). Ion image quantification has also been achieved in this sample type (8).
The procedure described here is amenable to a wide variety of cell cultures and provides a way for studying the response of individual cells for chemical alterations in the state of health and disease and localization of isotopically-labeled molecules and theraputic drugs in cell culture models.
SCOPE
1.1 This guide provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a cryogenic method for analyzing individual tissue culture cells growing in vitro. This guide is suitable for frozen-hydrated and frozen-freeze-dried sample types. Included are procedures for correlating optical, laser scanning confocal and secondary electron microscopies to complement SIMS analysis.
1.2 This guide is not suitable for cell cultures that do not attach to the substrate.
1.3 This guide is not suitable for any plastic embedded cell culture specimens.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide3 pagesEnglish language
SIGNIFICANCE AND USE
This practice is intended for use in reporting the experimental and data reduction procedures described in other publications.
SCOPE
1.1 This practice provides the minimum information necessary to describe the instrumental, experimental, and data reduction procedures used in acquiring and reporting secondary ion mass spectrometry (SIMS) mass spectral data.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard2 pagesEnglish language
SIGNIFICANCE AND USE
Pressing cryosections flat onto a conducting substrate has been one of the most challenging problems in SIMS analysis of cryogenically prepared tissue specimens. Frozen cryosections often curl or peel off, or both, from the substrate during freeze-drying. The curling of cryosections results in an uneven sample surface for SIMS analysis. Furthermore, if freeze-dried cryosections are not attached tightly to the substrate, the impact of the primary ion beam may result in further curling and even dislodging of the cryosection from the substrate. These problems render SIMS analysis difficult, frustrating and time consuming. The use of indium as a substrate for pressing cryosections flat has provided a practical approach for analyzing cryogenically prepared tissue specimens.(1)
The procedure described herein has been successfully used for SIMS imaging of calcium and magnesium transport and localization of anticancer drugs in animal models (2, 3, 4, 5)
The procedure described here is amenable to soft tissues of both animal and plant origin.
SCOPE
1.1 This practice provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a method for analyzing tissue cryosections in the imaging mode of the instrument. This practice is suitable for frozen-freeze-dried and frozen-hydrated cryosection analysis.
1.2 This practice does not describe methods for optimal freezing of the specimen for immobilizing diffusible chemical species in their native intracellular sites.
1.3 This practice does not describe methods for obtaining cryosections from a frozen specimen.
1.4 This practice is not suitable for any plastic embedded tissues.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard2 pagesEnglish language
SIGNIFICANCE AND USE
Electron multipliers are commonly used in pulse-counting mode to detect ions from magnetic sector mass spectrometers. The electronics used to amplify, detect and count pulses from the electron multipliers always have a characteristic time after the detection of a pulse after which no other pulses can be counted. This characteristic time is known as the “dead time.” The dead time has the effect of reducing the measured count rate compared with the “true” count rate.
In order to measure count rates accurately over the entire dynamic range of a pulse counting detector, such as an electron multiplier, the dead time of the entire pulse counting system must be well known. Accurate count rate measurement forms the basis of isotopic ratio measurements as well as elemental abundance determinations.
The procedure described herein has been successfully used to determine the dead time of counting systems on SIMS instruments.5 The accurate determination of the dead time by this method has been a key component of precision isotopic ratio measurements made by SIMS.
SCOPE
1.1 This practice provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a method for determining the dead time of the pulse-counting detection systems on the instrument. This practice also allows the analyst to determine whether the apparent dead time is independent of count rate.
1.2 This practice is applicable to most types of mass spectrometers that have pulse-counting detectors.
1.3 This practice does not describe methods for precise or accurate isotopic ratio measurements, or both.
1.4 This practice does not describe methods for the proper operation of pulse counting systems and detectors for mass spectrometry.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard3 pagesEnglish language
SCOPE
1.1 This practice lists the minimum information necessary to describe the instrumental, experimental, and data reduction procedures used in acquiring and reporting images generated by secondary ion mass spectrometry (SIMS).
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard2 pagesEnglish language
SCOPE
1.1 This guide provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a cryogenic method for analyzing individual tissue culture cells growing in vitro. This guide is suitable for frozen-hydrated and frozen-freeze-dried sample types. Included are procedures for correlating optical, laser scanning confocal and secondary electron microscopies to compliment SIMS analysis.
1.2 This guide is not suitable for cell cultures that do not attach to the substrate.
1.3 This guide is not suitable for any plastic embedded cell culture specimens.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide2 pagesEnglish language
SCOPE
1.1 This guide provides the Secondary Ion Mass Spectometry (SIMS) analyst with a cryogenic method for analyzing individual tissue culture cells growing into vitro. This guide is suitable for frozen-hydrated and frozen-freeze-dried sample types. Included are procedures for correlating optical, laser scanning confocal and secondary electron microscopies to compliment SIMS analysis.
1.2 This guide is not suitable for cell cultures that do not attach to the substrate.
1.3 This guide is not suitable for any plastic embedded cell culture specimens.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide2 pagesEnglish language
SCOPE
1.1 This practice provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a method for analyzing tissue cryosections in the imaging mode of the instrument. This practice is suitable for frozen-freeze-dried and frozen-hydrated cryosection analysis.
1.2 This practice does not describe methods for optimal freezing of the specimen for immobilizing diffusible chemical species in their native intracellular sites.
1.3 This practice does not describe methods for obtaining cryosections from a frozen specimen.
1.4 This practice is not suitable for any plastic embedded tissues.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard2 pagesEnglish language
SCOPE
1.1 This practice provides the Secondary Ion Mass Spectrometry (SIMS) analyst with a method for analyzing tissue cryosections in the imaging mode of the instrument. This practice is suitable for frozen-freeze-dried and frozen-hydrated cryosection analysis.
1.2 This practice does not describe methods for optimal freezing of the specimen for immobilizing diffusible chemical species in their native intracellular sites.
1.3 This practice does not describe methods for obtaining cryosections from a frozen specimen.
1.4 This practice is not suitable for any plastic embedded tissues.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard2 pagesEnglish language
SCOPE
1.1 This practice provides the minimum information necessary to describe the instrumental, experimental, and data reduction procedures used in acquiring and reporting secondary ion mass spectrometry (SIMS) mass spectral data.
1.2 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard2 pagesEnglish language
SCOPE
1.1 The purpose of this guide is to provide the secondary ion mass spectrometry (SIMS) analyst with two procedures for determining relative sensitivity factors (RSFs) from ion implanted external standards. This guide may be used for obtaining the RSFs of trace elements (
1.2 This guide does not describe procedures for obtaining RSFs for major elements (>1 atomic %). In addition, this guide does not describe procedures for obtaining RSFs from implants in heterogeneous (either laterally or in-depth) specimens.
1.3 The values stated in SI units are to be regarded as the standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide3 pagesEnglish language
SCOPE
1.1 This practice provides the minimum information necessary to describe the instrumental, experimental, and data reduction procedures used in acquiring and reporting secondary ion mass spectrometry (SIMS) mass spectral data.
1.2 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard2 pagesEnglish language
SCOPE
1.1 This guide provides the SIMS analyst with a method for determining the width of interfaces from SIMS sputtering data obtained from analyses of layered specimens. This guide does not apply to data obtained from analyses of specimens with thin markers or specimens without interfaces such as ion-implanted specimens.
1.2 This guide does not describe methods for the optimization of interface width or the optimization of depth resolution.
1.3 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide2 pagesEnglish language
SCOPE
1.1 This guide provides the SIMS analyst with a method for determining the width of interfaces from SIMS sputtering data obtained from analyses of layered specimens. This guide does not apply to data obtained from analyses of specimens with thin markers or specimens without interfaces such as ion-implanted specimens.
1.2 This guide does not describe methods for the optimization of interface width or the optimization of depth resolution.
1.3 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Guide2 pagesEnglish language
SCOPE
1.1 This practice covers the information needed to describe and report instrumentation, specimen parameters, experimental conditions, and data reduction procedures. SIMS sputter depth profiles can be obtained using a wide variety of primary beam excitation conditions, mass analysis, data acquisition, and processing techniques (1-4).
1.2 Limitations—This practice is limited to conventional sputter depth profiles in which information is averaged over the analyzed area in the plane of the specimen. Ion microprobe or microscope techniques permitting lateral spatial resolution of secondary ions within the analyzed area, for example, image depth profiling, are excluded.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard3 pagesEnglish language
SCOPE
1.1 This practice covers the information needed to describe and report instrumentation, specimen parameters, experimental conditions, and data reduction procedures. SIMS sputter depth profiles can be obtained using a wide variety of primary beam excitation conditions, mass analysis, data acquisition, and processing techniques (1-4).
1.2 Limitations—This practice is limited to conventional sputter depth profiles in which information is averaged over the analyzed area in the plane of the specimen. Ion microprobe or microscope techniques permitting lateral spatial resolution of secondary ions within the analyzed area, for example, image depth profiling, are excluded.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard3 pagesEnglish language
SIGNIFICANCE AND USE
Interpretation of static SIMS mass spectral data can be complicated due to the complexity and density of data obtained and therefore, variability often occurs when users are not consistent in their methods of data interpretation This guide is intended to help avoid these inconsistencies, by discussing the most commonly observed scenarios in static SIMS analysis and how to approach these scenarios.
This guide can be used as a training guide for employees or students, or both.
SCOPE
1.1 This guide provides time-of-flight secondary ion mass spectrometry (ToF-SIMS) users with a method for forms of interpretation of mass spectral data. This guide is applicable to most ToF-SIMS instruments and may or may not be applicable to other forms of secondary icon mass spectrometry (SIMS).
1.2 This guide does not purport to address methods of sample preparation. It is the responsibility of the user to adhere to strict sample preparation procedures in order to minimize contamination and optimize signals. See Guide E1078 and ISO 18116 for sample preparation guidelines.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
WITHDRAWN RATIONALE
This guide provides time-of-flight secondary ion mass spectrometry (ToF-SIMS) users with a method for forms of interpretation of mass spectral data. This guide is applicable to most ToF-SIMS instruments and may or may not be applicable to other forms of secondary icon mass spectrometry (SIMS).
Formerly under the jurisdiction of Committee E42 on Surface Analysis, this guide was withdrawn in January 2018 in accordance with Section 10.6.3 of the Regulations Governing ASTM Technical Committees, which requires that standards shall be updated by the end of the eighth year since the last approval date.
- Guide12 pagesEnglish language
SCOPE
1.1 The purpose of this guide is to provide the secondary ion mass spectrometry (SIMS) analyst with two procedures for determining relative sensitivity factors (RSFs) from ion implanted external standards. This guide may be used for obtaining the RSFs of trace elements (
1.2 This guide does not describe procedures for obtaining RSFs for major elements (>1 atomic %). In addition, this guide does not describe procedures for obtaining RSFs from implants in heterogeneous (either laterally or in-depth) specimens.
1.3 The values stated in SI units are to be regarded as the standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
WITHDRAWN RATIONALE
The purpose of this guide is to provide the secondary ion mass spectrometry (SIMS) analyst with two procedures for determining relative sensitivity factors (RSFs) from ion implanted external standards.
Formerly under the jurisdiction of Committee E42 on Surface Analysis, this guide was withdrawn in February 2010 in accordance with section 10.5.3.1 of the Regulations Governing ASTM Technical Committees, which requires that standards shall be updated by the end of the eighth year since the last approval date.
- Guide3 pagesEnglish language
Frequently Asked Questions
E42.06 is a Technical Committee within ASTM International. It is named "SIMS". This committee has published 38 standards.
E42.06 develops ASTM standards in the area of Information technology. Currently, there are 38 published standards from this technical committee.
ASTM is a standardization organization that develops and publishes standards to support industry, commerce, and regulatory requirements.
A Technical Committee (TC) in ASTM is a group of experts responsible for developing international standards in a specific technical area. TCs are composed of national member body delegates and work through consensus to create standards that meet global industry needs. Each TC may have subcommittees (SCs) and working groups (WGs) for specialized topics.