SIST EN IEC 61010-2-030:2021/A11:2021
(Amendment)Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-030: Particular requirements for equipment having testing or measuring circuits
Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-030: Particular requirements for equipment having testing or measuring circuits
1 Scope and object
This clause of Part 1 is applicable except as follows:
1.1.1 Equipment included in scope
Replacement:
Replace the text with the following:
This group safety publication is primarily intended to be used as a product safety standard for the products mentioned in the scope, but shall also be used by technical committees in the preparation of their publications for products similar to those mentioned in the scope of this standard, in accordance with the principles laid down in IEC Guide 104 and lSO/lEC Guide 51.
This part of IEC 61010 specifies safety requirements for equipment having testing or measuring circuits which are connected for test or measurement purposes to devices or circuits outside the measurement equipment itself.
These include measuring circuits which are part of electrical test and measurement equipment, laboratory equipment, or process control equipment. The existence of these circuits in equipment requires additional protective means between the circuit and an OPERATOR.
NOTE These testing and measuring circuits can, for example:
– measure voltages in circuits of other equipment,
– measure temperature of a separate device via a thermocouple,
– measure force on a separate device via a strain gauge,
– inject a voltage onto a circuit to analyse a new design.
Equipment having these testing and measuring circuits may be intended for performing tests and measurements on hazardous conductors, including MAINS conductors and telecommunication network conductors. See Annex BB for considerations of HAZARDS involved in various tests and measurements.
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2-030: Besondere Anforderungen für Geräte mit Prüf- oder Messstromkreis
[Not available]
Exigences de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire - Partie 2-030: Exigences particulières pour les appareils équipés de circuits d’essai ou de mesure
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko uporabo - 2-030. del: Posebne zahteve za opremo s preskusnimi ali z merilnimi vezji- Dopolnilo 11
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
SIST EN IEC 61010-2-030:2021/A11:2021
01-julij-2021
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 2-030. del: Posebne zahteve za opremo s preskusnimi ali z merilnimi
vezji- Dopolnilo 11
Safety requirements for electrical equipment for measurement, control, and laboratory
use - Part 2-030: Particular requirements for equipment having testing or measuring
circuits
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2
-030: Besondere Anforderungen für Geräte mit Prüf- oder Messstromkreis
Exigences de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire - Partie 2-030: Exigences particulières pour les appareils équipés de circuits
d’essai ou de mesure
Ta slovenski standard je istoveten z: EN IEC 61010-2-030:2021/A11:2021
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
SIST EN IEC 61010-2-030:2021/A11:2021 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
---------------------- Page: 1 ----------------------
SIST EN IEC 61010-2-030:2021/A11:2021
---------------------- Page: 2 ----------------------
SIST EN IEC 61010-2-030:2021/A11:2021
EUROPEAN STANDARD EN IEC 61010-2-030:2021/A11
NORME EUROPÉENNE
EUROPÄISCHE NORM
April 2021
ICS 19.080; 71.040.10
English Version
Safety requirements for electrical equipment for measurement,
control, and laboratory use - Part 2-030: Particular requirements
for equipment having testing or measuring circuits
Exigences de sécurité pour appareils électriques de Sicherheitsbestimmungen für elektrische Mess-, Steuer-,
mesurage, de régulation et de laboratoire - Partie 2-030: Regel- und Laborgeräte - Teil 2-030: Besondere
Exigences particulières pour les appareils équipés de Anforderungen für Geräte mit Prüf- oder Messstromkreis
circuits d'essai ou de mesure
This amendment A11 modifies the European Standard FprEN 61010-2-030:2016; it was approved by CENELEC on 2021-03-08. CENELEC
members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this amendment the
status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This amendment exists in three official versions (English, French, German). A version in any other language made by translation under the
responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the same status as
the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2021 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN IEC 61010-2-030:2021/A11:2021 E
---------------------- Page: 3 ----------------------
SIST EN IEC 61010-2-030:2021/A11:2021
EN IEC 61010-2-030:2021/A11:2021 (E)
European foreword
This document (EN IEC 61010-2-030:2021/A11:2021) has been prepared by CLC/TC 66X “Safety of
measuring, control, and laboratory equipment”.
The following dates are fixed:
• latest date by which this document has (dop) 2021-10-02
to be implemented at national level by
publication of an identical national
standard or by endorsement
• latest date by which the national (dow) 2024-04-02
standards conflicting with this document
have to be withdrawn
This document amends EN IEC 61010-2-030:2021.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
This document has been prepared under a mandate given to CENELEC by the European Commission
and the European Free Trade Association, and supports essential requirements of EU Directive(s).
For the relationship with EU Directive(s) see informative Annex ZZ, which is an integral part of this
document.
2
---------------------- Page: 4 ----------------------
SIST EN IEC 61010-2-030:2021/A11:2021
EN IEC 61010-2-030:2021/A11:2021 (E)
1 Modification to 6.8.3.1, “The a.c. voltage test”
Replace the first sentence of the first paragraph with the following sentence:
“The voltage tester shall have a regulated output capable of maintaining the test voltage throughout the
test within ± 5% of the specified value.”
2 Modification to 6.8.3.2, “The d.c. voltage test”
Add a new sentence at the beginning of the first paragraph:
“The voltage tester shall have a regulated output capable of maintaining the test voltage throughout the
test within ± 5% of the specified value.”
3 Modifications to K.101.4.1, “General”
Replace the second paragraph with the following note:
"
NOTE 1 The manufacturer is supposed to take the expected life of the equipment into account when selecting
insulating materials.”
As a result of this, renumber the existing note as NOTE 2.
4 Addition of Annex ZA, “Normative references to international publications
...
SLOVENSKI STANDARD
kSIST FprEN 61010-2-030:2016/oprAA:2020
01-maj-2020
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 2-030. del: Posebne zahteve za opremo s preskusnimi ali z merilnimi
vezji
Safety requirements for electrical equipment for measurement, control, and laboratory
use - Part 2-030: Particular requirements for equipment having testing or measuring
circuits
Exigences de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire - Partie 2-030: Exigences particulières pour les appareils équipés de circuits
d’essai ou de mesure
Ta slovenski standard je istoveten z: FprEN 61010-2-030:2016/prAA
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
kSIST FprEN 61010-2- en,fr,de
030:2016/oprAA:2020
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
---------------------- Page: 1 ----------------------
kSIST FprEN 61010-2-030:2016/oprAA:2020
---------------------- Page: 2 ----------------------
kSIST FprEN 61010-2-030:2016/oprAA:2020
EUROPEAN STANDARD DRAFT
FprEN 61010-2-030:2016
NORME EUROPÉENNE
EUROPÄISCHE NORM
prAA
March 2020
ICS
English Version
Safety requirements for electrical equipment for measurement,
control, and laboratory use - Part 2-030: Particular requirements
for equipment having testing or measuring circuits
Exigences de sécurité pour appareils électriques de To be completed
mesurage, de régulation et de laboratoire - Partie 2-030:
Exigences particulières pour les appareils équipés de
circuits d'essai ou de mesure
This draft amendment prAA, if approved, will modify the European Standard FprEN 61010-2-030:2016; it is submitted to CENELEC
members for enquiry.
Deadline for CENELEC: 2020-06-05.
It has been drawn up by CLC/TC 66X.
If this draft becomes an amendment, CENELEC members are bound to comply with the CEN/CENELEC Internal Regulations which
stipulate the conditions for giving this amendment the status of a national standard without any alteration.
This draft amendment was established by CENELEC in three official versions (English, French, German).
A version in any other language made by translation under the responsibility of a CENELEC member into its own language and notified to
the CEN-CENELEC Management Centre has the same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are aware and to
provide supporting documentation.
Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without notice and
shall not be referred to as a European Standard.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2020 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Project: 69654 Ref. No. FprEN 61010-2-030:2016/prAA E
---------------------- Page: 3 ----------------------
kSIST FprEN 61010-2-030:2016/oprAA:2020
FprEN 61010-2-030:2016/prAA:2020 (E)
European foreword
This document (FprEN 61010-2-030:2016/prAA:2020) has been prepared by CLC/TC 66X “Safety of
measuring, control, and laboratory equipment”.
This document is currently submitted to the Enquiry.
The following dates are proposed:
• latest date by which the existence of this (doa) dor + 6 months
document has to be announced at national
level
• latest date by which this document has to be (dop) dor + 12 months
implemented at national level by publication of
an identical national standard or by
endorsement
• latest date by which the national standards (dow) dor + 36 months
conflicting with this document have to be (to be confirmed or
withdrawn modified when voting)
This document amends FprEN 61010-2-030:2016.
This document has been prepared under a mandate given to CENELEC by the European Commission
and the European Free Trade Association, and supports essential requirements of EU Directive(s).
For the relationship with EU Directive(s) see informative Annex ZZ, which is an integral part of this
document.
2
---------------------- Page: 4 ----------------------
kSIST FprEN 61010-2-030:2016/oprAA:2020
FprEN 61010 2-030:2016/prAA:2020 (E)
1 Modification to 6.8.3.1, “The a.c. voltage test”
Replace the first sentence of the first paragraph with the following sentence:
“The voltage tester shall have a regulated output capable of maintaining the test voltage throughout the
test within ± 5% of the specifi
...