Abstract

IEC 80601-2-52:2026 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of MEDICAL BEDS as defined in 201.3.214, intended for ADULTS as defined in 201.3.222. Included in the scope are both electrical and non-electrical (manual) MEDICAL BEDS with or without adjustable functions. This document is applicable to either a BED-LIFT or a detachable MATTRESS SUPPORT PLATFORM or both. The combination of BED-LIFT or a detachable MATTRESS SUPPORT PLATFORM with a compatible non-MEDICAL BED as specified by the MANUFACTURER is also considered a MEDICAL BED. This document does not apply to: - MEDICAL BEDS for CHILDREN and ADULTS with atypical anatomies (ADULTS ranging outside the definition for ADULTS in 202.3.222) covered by IEC 80601-2-89; - SPECIALITY MATTRESS covered by ISO 20342 series; - devices for which the INTENDED USE is mainly for examination or transportation under medical supervision (e.g. stretcher, examination table); - all requirements for MEDICAL BEDS with special functionality. If a clause or subclause is specifically intended to be applicable to a MEDICAL BED only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to MEDICAL BED and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of MEDICAL BED or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020.

Status
Published
Publication Date
09-Jul-2026
Current Stage
6060 - Document made available - Publishing
Start Date
10-Jul-2026
Due Date
10-Jul-2026
Completion Date
10-Jul-2026

Buy Documents

Standard

EN IEC 80601-2-52:2026 - BARVE

English language (103 pages)
Preview
Preview
e-Library read for
×1 day

EN IEC 80601-2-52:2026 is a European Standard published by CENELEC for the basic safety and essential performance of medical beds for adults. It covers both electrical and non-electrical medical beds, including beds with a bed-lift, a detachable mattress support platform, or both, and it is meant for manufacturers, test labs, and risk managers who need to design, mark, test, or assess these products.

What does EN IEC 80601-2-52:2026 specify?

EN IEC 80601-2-52:2026 specifies the particular requirements for medical beds intended for adults and explains how the document is organized around the IEC 60601-1 structure. It adds medical-bed-specific rules to the general safety standard and uses annexes for guidance, additional design requirements, and inspection advice.

The scope in Clause 201.1 includes:

  • electrical and manual medical beds
  • beds with or without adjustable functions
  • a bed-lift, a detachable mattress support platform, or both
  • a compatible non-medical bed combined with a bed-lift or mattress support platform, when the manufacturer specifies that combination

The document excludes:

  • medical beds for children and adults with atypical anatomies covered by IEC 80601-2-89
  • speciality mattresses covered by the ISO 20342 series
  • devices mainly intended for examination or transportation under medical supervision, such as stretchers and examination tables
  • medical beds with special functionality

Document structure

AnnexWhat it covers
Annex AAParticular guidance and rationale
Annex BBAdditional design requirements and recommendations for medical beds
Annex CCGuidance for assessing risk of entrapment in V-shaped openings
Annex DDGuidance and recommendations for periodic inspection

The main body follows the IEC 60601-1 clause structure from 201.1 to 201.17, so a reader can find scope, definitions, marking, mechanical hazards, temperature, cleaning, usability, fault conditions, construction, and EMC in familiar places.

What are the key requirements of EN IEC 80601-2-52:2026?

EN IEC 80601-2-52:2026 focuses on risk management, entrapment prevention, stability, load-bearing strength, usability, and cleaning-related hazards. In practice, it asks the manufacturer to show that the bed remains safe for its intended use, environment, and accessories.

Risk management and application environments

Clause 201.4.2.2 requires the manufacturer to include in the risk management process hazards for patients taller than 185 cm, patients outside the ADULT definition, and bariatric patients. For medical beds intended for APPLICATION ENVIRONMENT 6, the risk management also has to consider strangulation, snagging, disassembly that could lead to choking or ingestion, and material flammability.

Clause 201.4.6 extends patient-contact considerations to all parts within the patient’s reach, including parts under the mattress support platform. For APPLICATION ENVIRONMENT 4 and APPLICATION ENVIRONMENT 6, all accessible parts are treated as APPLIED PARTs, which means the whole bed has to be assessed with that level of attention.

Marking and traceability

Clause 201.7.2.2 requires the medical bed and detachable components to carry the manufacturer identification, model or type reference, and traceability information. Clause 201.7.2.2.101 adds marking of the maximum patient weight and SAFE WORKING LOAD (SWL), which matters because users must match the bed, mattress, and accessories to the load they are actually using.

Entrapment and fall protection

Table 201.101 sets the entrapment approach for openings around side rails, head boards, foot boards, and the mattress support platform. The tests use defined tools and opening limits, so the reader can verify whether a gap is small enough to block access or large enough to avoid creating a trapping point.

Table 201.103 covers inadvertent patient falls by requiring the side rail top edge to reach a defined height over enough of the mattress support platform. Clause 201.9.2.2.3 also sets a 200 mm safe distance for hand reach when the patient is lying in the bed.

Annex CC gives a proposed method for checking V-shaped openings in areas B and C. It is informative, but it is useful for design reviews because it explains how geometry can create entrapment risk even when a simple gap check is not enough.

Strength, stability, and durability

Clause 201.9.8.3.1 defines minimum SWLs for different application environments and for the bed-lift. Clause 201.9.8.3.2 then requires static loading at two times the SWL or 4 000 N, whichever is greater, plus separate requirements for accessories and lifting poles. In practice, the manufacturer must prove that the bed, accessories, and lifting pole do not deform or fail in a way that creates a hazard.

Clause 201.9.8.3.3 adds endurance tests for loading from persons, height adjustment, and side rail latch reliability. The document also includes stability tests for lateral and longitudinal tipping, movement over a threshold, and rough handling, so the bed has to survive both daily use and movement through real facilities.

Control of movement and usability

Clause 201.9.2.2.5 requires continuous activation for movement controls, meaning motion stops when the control is released. Clause 201.9.2.3.1 addresses unintended movement, and for APPLICATION ENVIRONMENT 6 it requires a MOTION LOCKOUT CONTROL when the patient can access the controls.

Clause 201.12.2 asks the USABILITY ENGINEERING PROCESS to consider reach, ergonomic operation, button and switch placement, and prompt access to the patient for treatment. For buyers and designers, that means the bed has to be safe to operate without slowing care.

Temperature, flammability, ingress, and cleaning

Clause 201.11.1.1 replaces the temperature limits for patient-contact APPLIED PARTS, so skin-contact surfaces have to stay within specified limits for different materials and contact durations. Clause 201.11.2.101 requires the manufacturer to consider flammability of textiles, coated textiles, and plastics coverings, and to warn users about safe material combinations.

Clause 201.11.6.5 sets a minimum IPX4 protection degree, with IPX9 required for beds marked for jet stream washing. Clause 201.11.6.6 requires the manufacturer to specify the cleaning or disinfection process when the bed is intended for use with such equipment.

What terms does EN IEC 80601-2-52:2026 define?

  • MEDICAL BED - A device intended for sleeping, resting, or both, with a mattress support platform, and intended to assist diagnosis, monitoring, prevention, treatment, alleviation of disease, or compensation for injury or disability.
  • ADULT - A patient with a physical size of at least 146 cm, a mass of at least 40 kg, and a body mass index of at least 17.
  • APPLICATION ENVIRONMENT 4 - Care in a domestic area where the medical bed is used to alleviate or compensate for injury, disability, or disease.
  • APPLICATION ENVIRONMENT 6 - Psychiatric or mental health care, including prisons, jails, and correctional facilities.
  • SAFE WORKING LOAD (SWL) - The sum of the patient, the mattress, supporting accessories, and the loads supported by those accessories, excluding patient weight.
  • TRAPPING ZONE - A location where a patient or another person can become entrapped, entangled, wedged, or stuck between parts of the medical bed.
  • PENDANT CONTROL - A hand-held control connected to the bed that operates articulations or movements, and may be wired or wireless.

Who uses EN IEC 80601-2-52:2026?

EN IEC 80601-2-52:2026 is used by medical bed manufacturers, accessory makers, and test laboratories when designing and verifying beds for adult care. It is also used by quality managers, risk managers, and regulatory teams when preparing the risk management file, instructions for use, marking, and compliance test evidence.

Hospitals, long-term care providers, home-care suppliers, and procurement teams use it to compare beds for a specific application environment and to check that mattresses, side rails, lifting poles, and controls are compatible. Maintenance and inspection personnel can also use Annex DD to plan periodic inspections and record the results.

What changed in EN IEC 80601-2-52:2026 from the previous edition?

EN IEC 80601-2-52:2026 supersedes EN 60601-2-52:2010, EN 60601-2-52:2010/A1:2015, and EN 60601-2-52:2010/AC:2011. The foreword also says this first edition cancels and replaces IEC 60601-2-52:2009 and Amendment 1:2015.

The stated technical changes include:

  • scope updated to cover both electrical and non-electrical medical beds
  • scope clarified to show overlap with IEC 80601-2-89
  • introduction of APPLICATION ENVIRONMENT 6
  • alignment with IEC 60601-1:2005/AMD2:2020
  • a new snagging test
  • durability tests moved into one subclause for most application environments
  • a new jet stream washing rating and updated machine-washing requirements
  • updated test tools with tolerances and drawings

Which standards are used with EN IEC 80601-2-52:2026?

StandardWhat it contributes
IEC 60601-1:2005 and its amendmentsGeneral requirements for basic safety and essential performance
IEC 60529Degrees of protection provided by enclosures, including IPX4 and IPX9 checks
ISO 3746Sound power and sound pressure measurement method for acoustic output
EN 597-1 and EN 597-2Ignitability tests for mattress and upholstered bed-base claims
IEC 80601-2-89:2025Particular requirements for medical beds for children, used to define the adult-bed scope boundary
ISO 20342 seriesSpeciality mattresses for tissue integrity when lying down
ISO 9999:2022Assistive products classification and terminology, used in definitions
IEC 60068-2-31Rough handling test method referenced for PENDANT CONTROL testing
ISO/IEC Guide 71:2014Accessibility guidance referenced in the bibliography
IEC 60601-1-10Named in the bibliography, while Clause 201.1.3 states it does not apply
IEC 60601-1-3Named in Clause 201.1.3 as not applying

What does the EN IEC 80601-2-52:2026 document contain?

EN IEC 80601-2-52:2026 contains defined test tools, loading devices, and measurement set-ups for entrapment, stability, impacts, and load-bearing checks. It also contains tables for entrapment protection, minimum SWLs, fall protection, maximum skin-contact temperatures, and the machine-washable compliance procedure.

The figures show the bed layout, tilt positions, side rail examples, clearance areas, dynamic loading positions, and the test tools used in the procedures. Annex BB adds design guidance on human factors and geometry, Annex CC gives an additional V-shaped-opening assessment method, and Annex DD gives a practical periodic inspection checklist and record content.

Relations

Effective Date
23-Jun-2026
Effective Date
23-Jun-2026
Effective Date
23-Jun-2026
Effective Date
30-Jun-2026
Effective Date
30-Jun-2026
Effective Date
30-Jun-2026
Effective Date
30-Jun-2026
Effective Date
30-Jun-2026
Effective Date
30-Jun-2026
Effective Date
30-Jun-2026
Effective Date
14-Jul-2026
Effective Date
14-Jul-2026
Effective Date
14-Jul-2026

Buy Documents

Standard

EN IEC 80601-2-52:2026 - BARVE

English language (103 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

EN IEC 80601-2-52:2026 is a standard published by CLC. Its full title is "Medical electrical equipment - Part 2-52: Particular requirements for the basic safety and essential performance of medical beds for adults". This standard covers: IEC 80601-2-52:2026 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of MEDICAL BEDS as defined in 201.3.214, intended for ADULTS as defined in 201.3.222. Included in the scope are both electrical and non-electrical (manual) MEDICAL BEDS with or without adjustable functions. This document is applicable to either a BED-LIFT or a detachable MATTRESS SUPPORT PLATFORM or both. The combination of BED-LIFT or a detachable MATTRESS SUPPORT PLATFORM with a compatible non-MEDICAL BED as specified by the MANUFACTURER is also considered a MEDICAL BED. This document does not apply to: - MEDICAL BEDS for CHILDREN and ADULTS with atypical anatomies (ADULTS ranging outside the definition for ADULTS in 202.3.222) covered by IEC 80601-2-89; - SPECIALITY MATTRESS covered by ISO 20342 series; - devices for which the INTENDED USE is mainly for examination or transportation under medical supervision (e.g. stretcher, examination table); - all requirements for MEDICAL BEDS with special functionality. If a clause or subclause is specifically intended to be applicable to a MEDICAL BED only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to MEDICAL BED and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of MEDICAL BED or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020.

IEC 80601-2-52:2026 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of MEDICAL BEDS as defined in 201.3.214, intended for ADULTS as defined in 201.3.222. Included in the scope are both electrical and non-electrical (manual) MEDICAL BEDS with or without adjustable functions. This document is applicable to either a BED-LIFT or a detachable MATTRESS SUPPORT PLATFORM or both. The combination of BED-LIFT or a detachable MATTRESS SUPPORT PLATFORM with a compatible non-MEDICAL BED as specified by the MANUFACTURER is also considered a MEDICAL BED. This document does not apply to: - MEDICAL BEDS for CHILDREN and ADULTS with atypical anatomies (ADULTS ranging outside the definition for ADULTS in 202.3.222) covered by IEC 80601-2-89; - SPECIALITY MATTRESS covered by ISO 20342 series; - devices for which the INTENDED USE is mainly for examination or transportation under medical supervision (e.g. stretcher, examination table); - all requirements for MEDICAL BEDS with special functionality. If a clause or subclause is specifically intended to be applicable to a MEDICAL BED only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to MEDICAL BED and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of MEDICAL BED or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020.

EN IEC 80601-2-52:2026 is classified under the following ICS (International Classification for Standards) categories: 11.140 - Hospital equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

EN IEC 80601-2-52:2026 has the following relationships with other standards: It is inter standard links to EN 60601-2-52:2010/A1:2015, EN 60601-2-52:2010/AC:2011, EN 60601-2-52:2010, EN 60529:1991/A1:2000, EN 60601-1:2006/A2:2021, EN 60529:1991, EN 60601-1:2006/A1:2013, EN 60529:1991/A2:2013, EN ISO 3746:2010, EN 60601-1:2006, EN 13718-1:2008, EN 1789:2007+A2:2014, EN 1789:2007+A1:2010. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

EN IEC 80601-2-52:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • 1 HEAD BOARD
  • 2 Back section (part of MATTRESS SUPPORT PLATFORM)
  • 3 Seat section (part of MATTRESS SUPPORT PLATFORM)
  • 4 Upper leg section (part of MATTRESS SUPPORT PLATFORM)
  • 5 Lower leg section (part of MATTRESS SUPPORT PLATFORM)
  • 6 FOOT BOARD
  • 1 HEAD BOARD
  • 2 FOOT BOARD
  • 3 MATTRESS SUPPORT PLATFORM
  • 1 HEAD BOARD
  • 2 FOOT BOARD
  • 3 MATTRESS SUPPORT PLATFORM
  • 1 Marked centre line
  • 2 Drilling hole for weight adjustment, if necessary
  • 3 The surface roughness of the TOOL is less than or equal to 1,6 µm
  • 1 Drilling hole for weight adjustment, if necessary
  • 2 The surface roughness of the TOOL is less than or equal to 1,6 µm
  • 1 Connecting part of the lifting device which shall not restrain the free fall
  • 2 Striking surface
  • 1 Ball chain loop with a length of 200 ± 2 mm
  • 2 Spherical mass, weight 2,5 ± 0,05 kg and diameter 115 ± 1 mm
  • 3 Fixing point
  • 1 MATTRESS SUPPORT PLATFORM
  • 2 HEAD BOARD
  • 3 FOOT BOARD
  • 4 Mattress
  • 5 SIDE RAIL
  • 1 MATTRESS SUPPORT PLATFORM
  • 2 HEAD BOARD
  • 3 FOOT BOARD
  • 4 Mattress
  • 5 SIDE RAIL
  • 1 Pinch point (TRAPPING ZONE)
  • 2 MATTRESS SUPPORT PLATFORM
  • 2 HEAD BOARD
  • 3 FOOT BOARD
  • 2 HEAD BOARD
  • 3 FOOT BOARD removed
  • 1 Two loads of 1 100 N, two loads of 925 N for APPLICATION ENVIRONMENT 4, evenly distributed
  • 2 HEAD BOARD
  • 3 FOOT BOARD
  • 1 Back section
  • 2 Seat section
  • 3 Upper leg section
  • 4 Lower leg section
  • 5 SWL = Safe Working Load
  • 1 Head end of MEDICAL BED
  • 2 Foot end of MEDICAL BED
  • 1 MATTRESS SUPPORT PLATFORM
  • 2 FOOT BOARD
  • 3 HEAD BOARD
  • 4 Mattress
  • 4 Operate each articulation control within its entire range, from end stop to end
  • 6 Repeat test 4. (Operate each articulation)
  • 12 Perform a visual inspection for ingress of water that can result in an
  • 1 Back section
  • 2 Leg section
  • 3 Upper leg section
  • 4 Seat section
  • 5 Lower leg section
  • Annex AA
  • 1 Normal force from SIDE RAIL
  • 2 Normal force from MATTRESS SUPPORT PLATFORM
  • 3 Neck
  • 4 Resultant (ejecting) force
  • 1 Normal force from SIDE RAIL
  • 2 Normal force to neck from MATTRESS
  • 3 Neck
  • 4 Resultant force from MATTRESS
  • 5 Mattress
  • 1 SIDE RAIL
  • 2 MATTRESS SUPPORT PLATFORM
  • 3 Gap measurement (< 60 mm)
  • 1 Mattress
  • 2 SIDE RAIL
  • 3 Mattress height(s) (specified by the MANUFACTURER)
  • 4 Angle measurement zone
  • 1 SIDE RAILS
  • 2 Mattress
  • 3 Cone TOOL

INTRODUCTION In 1996, the IEC published the first edition of the particular standard for electrically operated hospital beds, IEC 60601-2-38. The publication was in response to demand in the field for a universal standard addressing HAZARDS specific to the safety of the hospital bed. Used in conjunction with a MANUFACTURER'S RISK ASSESSMENT, the standard was felt to be the current thinking on establishing a BASIC SAFETY benchmark for industry. An amendment of IEC 60601-2-38 issued in 1999 recognized the need to mitigate against a RISK of PATIENT entrapment in the SIDE RAILS, again combined with the use of the MANUFACTURER'S RISK ASSESSMENT. Although this improved the particular standard, it still was centered upon electrically operated hospital beds, and failed to take into account manually operated hospital beds and products in other medical environments. In 2000, the EN 1970 standard (Adjustable beds for disabled persons – Requirements and test methods) was published, which addressed beds used by PERSONS WITH DISABILITY to alleviate or compensate for a disability or handicap. This standard offered a broadened scope in conjunction with IEC 60601-2-38, but after the edition of Amendment 1 to IEC 60601-2-38, the opportunity presented itself to combine the two standards to a common, international standard. As work began on the integration, the IEC adjusted its stance on BASIC SAFETY and ESSENTIAL PERFORMANCE, integrating them into the third edition of IEC 60601-1 published in 2005. It therefore became necessary to align the new standard with the third edition. The particular standard was given a new number, IEC 60601-2-52, and work began on alignment to third edition.

This particular standard, therefore, is the realization of much work in alignment, and scope adjustment between IEC 60601-2-38, EN 1970, and IEC 60601-1:2005. It represents the current thinking in BASIC SAFETY and ESSENTIAL PERFORMANCE of the MEDICAL BED as used to alleviate illness of PATIENTS and disability of PERSONS WITH DISABILITY. This is the effort of a joint working group of the IEC and the ISO.

IEC 80601-2-52 is a continuation of the IEC 60601-2-52 standard. It will align with the CHILDREN version of the standard, IEC 80601-2-89. The standard number was changed from IEC 60601-2-52 to IEC 80601-2-52 to indicate joint work between the ISO and the IEC.

SIST EN IEC 80601-2-52:2026

201.1 Scope, object and related standards IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, Clause 1 applies, except as follows:

201.1.1 * Scope Replacement:

This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of MEDICAL BEDS as defined in 201.3.214, intended for ADULTS as defined in 201.3.222. Included in the scope are both electrical and non-electrical (manual) MEDICAL BEDS with or without adjustable functions.

This document is applicable to either a BED-LIFT or a detachable MATTRESS SUPPORT PLATFORM or both. The combination of BED-LIFT or a detachable MATTRESS SUPPORT PLATFORM with a compatible non-MEDICAL BED as specified by the MANUFACTURER is also considered a MEDICAL BED.

This document does not apply to:

  • MEDICAL BEDS for CHILDREN and ADULTS with atypical anatomies (ADULTS ranging outside

the definition for ADULTS in 201.3.222) covered by IEC 80601-2-89[1] ; - SPECIALITY MATTRESS covered by ISO 20342 series[2]; - devices for which the INTENDED USE is mainly for examination or transportation under medical supervision (e.g. stretcher, examination table); - all requirements for MEDICAL BEDS with special functionality. If a clause or subclause is specifically intended to be applicable to a MEDICAL BED only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to MEDICAL BED and to ME SYSTEMS, as relevant. AZARDS inherent in the intended physiological function of MEDICAL BED or ME SYSTEMS within H the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020.

NOTE 1 See also 4.2 of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020. NOTE 2 Whenever the term MEDICAL ELECTRICAL EQUIPMENT (MEE, ME EQUIPMENT) is used within the series of IEC 60601 standards, it refers to MEDICAL BEDS (both electrical and non-electrical). 201.1.2 Object Replacement:

The object of this document is to establish particular BASIC SAFETY and ESSENTIAL PERFORMANCE requirements for MEDICAL BEDS as defined in 201.3.214, intended for ADULTS as defined in 201.3.222.

Numbers in square brackets refer to the Bibliography.

SIST EN IEC 80601-2-52:2026

201.1.3 Collateral standards Addition:

This document refers to those applicable collateral standards that are listed in Clause 2 of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 and Clause 2 of this document.

IEC 60601-1-3 and IEC 60601-1-10 do not apply. All other published collateral standards in the IEC 60601-1 series apply as published.

NOTE Some IEC 60601-1-8 requirements can be excluded if they do not affect PATIENT safety, could lead to user confusion, or are inappropriate to MEDICAL BED usage.

201.1.4 Particular standards Addition:

In the IEC 60601 series, particular standards may modify, replace or delete requirements contained in IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, including the collateral standards as appropriate for the particular ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL PERFORMANCE requirements.

A requirement of a particular standard takes priority over IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 The numbering of clauses and subclauses of this particular standard corresponds to that of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 with the prefix "201" (e.g. 201.1 in this document addresses the content of Clause 1 of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020) or applicable collateral standard with the prefix "20x" where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this document addresses the content of Clause 4 of the IEC 60601-1-2 collateral standard, 203.4 in this document addresses the content of Clause 4 of the IEC 60601-1-3 collateral standard, etc.). The changes to the text of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 are specified by the use of the following words:

"Replacement" means that the clause or subclause of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or applicable collateral standard is replaced completely by the text of this document.

"Addition" means that the text of this document is additional to the requirements of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or applicable collateral standard.

"Amendment" means that the clause or subclause of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or applicable collateral standard is amended as indicated by the text of this document. Subclauses, figures or tables which are additional to those of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 are numbered starting from 201.101. However, due to the fact that definitions in IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 are numbered 3.1 through 3.154, additional definitions in this document are numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and additional items aa), bb), etc.

SIST EN IEC 80601-2-52:2026

Subclauses, figures or tables which are additional to those of a collateral standard are numbered starting from 20x, where "x" is the number of the collateral standard, e.g. 202 for IEC 60601-1-2, 203 for IEC 60601-1-3, etc.

Where there is no corresponding clause or subclause in this particular standard, the clause or subclause of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or applicable collateral standard, although possibly not relevant, applies without modification; where it is intended that any part of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or applicable collateral standard, although possibly relevant, is not to be applied, a statement to that effect is given in this document.

201.2 Normative references The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

NOTE Informative references are listed in the Bibliography.

IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, Clause 2, applies except as follows:

Replacement:

IEC 60529:1989, Degrees of protection provided by enclosures (IP Code) IEC 60529:1989/AMD1:1999 IEC 60529:1989/AMD2:2013 Addition:

IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance IEC 60601-1:2005/AMD1:2012 IEC 60601-1:2005/AMD2:2020 ISO 3746, Acoustics - Determination of sound power levels and sound energy levels of noise sources using sound pressure - Survey method using an enveloping measurement surface over a reflecting plane EN 597-1, Furniture - Assessment of the ignitability of mattresses and upholstered bed bases - Part 1: Ignition source smouldering cigarette EN 597-2, Furniture - Assessment of the ignitability of mattresses and upholstered bed bases - Part 2: Ignition source match-flame equivalent 201.3 Terms and definitions For the purposes of this document, the terms and definitions given in IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, and the following apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • IEC Electropedia: available at https://www.electropedia.org/
  • ISO Online browsing platform: available at https://www.iso.org/obp

SIST EN IEC 80601-2-52:2026

201.3.8 APPLIED PART Addition:

Note 4 to entry: The APPLIED PART includes all parts of the MEDICAL BED that are within the reach of the PATIENT even if they are underneath the MATTRESS SUPPORT PLATFORM.

201.3.76 PATIENT Replacement:

living person undergoing a medical PROCEDURE or PERSON WITH DISABILITY 201.3.109 SAFE WORKING LOAD SWL Replacement:

sum of:

1) the PATIENT; 2) the mattress; 3) the ACCESSORIES of the MEDICAL BED, only if they are supported by the support system of the MEDICAL BED; and 4) the SAFE WORKING LOADS supported by those ACCESSORIES, excluding PATIENT weight 201.3.131 * TRAPPING ZONE Replacement:

location where the PATIENT or other persons can become entrapped, entangled, wedged, or stuck in or between parts of the MEDICAL BED, such as the SIDE RAILS, HEAD BOARD, FOOT BOARD, MATTRESS SUPPORT PLATFORM or mattress Additional terms and definitions:

201.3.201 * APPLICATION ENVIRONMENT 1 intensive care or critical care provided in a hospital where 24 h medical supervision and constant monitoring is required and provision of life support equipment used in medical PROCEDURES is essential to maintain or improve the vital functions of the PATIENT 201.3.202 * APPLICATION ENVIRONMENT 2 acute care provided in a hospital or other medical facility where medical supervision and monitoring is required and MEDICAL BED used in medical PROCEDURES is often provided to help maintain or improve the condition of the PATIENT 201.3.203 * APPLICATION ENVIRONMENT 3 long-term care in a medical area where medical supervision is required and monitoring is provided if necessary and MEDICAL BED used in medical PROCEDURES may be provided to help maintain or improve the condition of the PATIENT Note 1 to entry: This includes use in nursing homes and in rehabilitation and geriatric facilities.

SIST EN IEC 80601-2-52:2026

201.3.204 * APPLICATION ENVIRONMENT 4 care provided in a domestic area where MEDICAL BED is used to alleviate or compensate for an injury, disability or disease Note 1 to entry: This excludes use in all other APPLICATION ENVIRONMENTs (e.g. nursing homes, rehabilitation and geriatric facilities) when a MEDICAL BED is purely designed for APPLICATION ENVIRONMENT 4. 201.3.205 * APPLICATION ENVIRONMENT 5 outpatient care or ambulatory care, which is provided in a hospital or other medical facility, under medical supervision where MEDICAL BED, is provided for the need of persons with illness, injury or disability for treatment, diagnosis or monitoring 201.3.206 * APPLICATION ENVIRONMENT 6 psychiatric care or mental health care environment where medical supervision is required, and monitoring is provided. MEDICAL BED used in medical PROCEDURES may be provided to help maintain, improve condition and protect the PATIENT Note 1 to entry: Environment where a patient may be a harm to themselves or others. Note 2 to entry: Includes prisons, jails, correctional facilities. 201.3.207 BED-LIFT height adjustable mechanism on which a MATTRESS SUPPORT PLATFORM can be mounted Note 1 to entry: The combination of a BED LIFT and a compatible non-MEDICAL BED as specified by the MANUFACTURER is considered to be a MEDICAL BED.

201.3.208 CHILD PATIENT having a physical size equal to or less than 155 cm and a mass equal to or less than 70 kg and may display cognitive immaturity, exploratory behaviours, RISK-taking tendencies or any combination Note 1 to entry: Physical size is measured from crown to sole. [SOURCE: IEC 80601-2-89:2025[1], 201.3.207, modified – “Body length” replaced with “physical size” in the definition and in Note 1 to entry.] 201.3.209 PERSON WITH DISABILITY person with one or more impairments, one or more activity limitations, one or more participation restrictions or a combination thereof [SOURCE: ISO 9999:2022[3], 3.10] 201.3.210 HEAD OR FOOT BOARD assembly mounted to a MEDICAL BED, which identifies for the PATIENT the edge of the head or foot end of either the MEDICAL BED or MATTRESS SUPPORT PLATFORM or both Note 1 to entry: It can be used as handles to push a MEDICAL BED intended to transport PATIENTS. 201.3.211 LIFTING POLE ACCESSORY attached to a MEDICAL BED and intended to assist support of a PATIENT when changing position by providing a gripping support above the PATIENT

SIST EN IEC 80601-2-52:2026

201.3.212 MATTRESS OVERLAY supplementary mattress, which is intended to be placed on an existing mattress, and generally used for prophylactic or therapeutic effect 201.3.213 MATTRESS SUPPORT PLATFORM structure which supports a PATIENT surface Note 1 to entry: The platform can articulate or change positions to facilitate various therapeutic, diagnostic and convenience positions (See Figure 201.101).

Key

1 HEAD BOARD

2 Back section (part of MATTRESS SUPPORT PLATFORM)

3 Seat section (part of MATTRESS SUPPORT PLATFORM)

4 Upper leg section (part of MATTRESS SUPPORT PLATFORM)

5 Lower leg section (part of MATTRESS SUPPORT PLATFORM)

6 FOOT BOARD

Figure 201.101 – MEDICAL BED, general arrangement (example, schematic presentation only) 201.3.214 * MEDICAL BED device for which the INTENDED USE is sleeping, resting or both that contains a MATTRESS SUPPORT PLATFORM and intended to assist in diagnosis, monitoring, prevention, treatment, alleviation of disease or compensation for an injury or disability 201.3.215 MOTION LOCKOUT CONTROL auxiliary subsystem that deactivates motion controls

SIST EN IEC 80601-2-52:2026

201.3.216 PENDANT CONTROL hand-held device, which has a FUNCTIONAL CONNECTION to the MEDICAL BED, controlling at least MEDICAL BED articulations or movements or both Note 1 to entry: PENDANT CONTROLS can be wired, or wireless, and can integrate other functions such as communications, radio, tv, etc.

201.3.217 SIDE RAIL physical barrier, which may be a detachable ACCESSORY or integral to the overall construction of a MEDICAL BED and is mounted to the side(s) of the MEDICAL BED Note 1 to entry: When a SIDE RAIL is located in the "up" position it provides a physical barrier, which is intended to reduce the RISK of the PATIENT accidentally slipping or rolling off the mattress. 201.3.218 SPECIALTY MATTRESS mattress intended for prophylactic or therapeutic effect 201.3.219 TEST BED BOARD flat, rigid loading board of dimensions as specified by the MANUFACTURER representing the dimensions of the MEDICAL BED 201.3.220 TRENDELENBURG specific therapeutic position used in some medical PROCEDURES requiring a minimum 12° head down angle of the entire MATTRESS SUPPORT PLATFORM with respect to horizontal 201.3.221 UNDERCARRIAGE all components of the MEDICAL BED or BED-LIFT below the MATTRESS SUPPORT PLATFORM 201.3.222 * ADULT PATIENT having a physical size equal to or more than 146 cm, a mass equal to or more than 40 kg and a body mass index (BMI) equal to or more than 17 201.3.223 HEAD DOWN TILT entire MATTRESS SUPPORT PLATFORM positioned in a head down angle less than 12° with respect to horizontal

SIST EN IEC 80601-2-52:2026

Key

1 HEAD BOARD

2 FOOT BOARD

3 MATTRESS SUPPORT PLATFORM

Figure 201.102 – HEAD DOWN TILT, Example 201.3.224 FOOT DOWN TILT entire MATTRESS SUPPORT PLATFORM positioned in any foot down angle with respect to horizontal

Key

1 HEAD BOARD

2 FOOT BOARD

3 MATTRESS SUPPORT PLATFORM

Figure 201.103 – FOOT DOWN TILT, Example 201.4 General requirements IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, Clause 4 applies, except as follows:

201.4.2.2 * General requirement for RISK MANAGEMENT Addition:

The MANUFACTURER shall include, in the RISK MANAGEMENT PROCESS, HAZARDS (e.g. entrapment, PATIENT falls) related to PATIENTS taller than 185 cm, or having a physical size, mass and body mass index (BMI) outside the definition of ADULT.

Bariatric PATIENT population shall also be considered in the RISK MANAGEMENT PROCESS. Alternative means not described in this document to ensure an equivalent level of BASIC SAFETY and ESSENTIAL PERFORMANCE shall be justified by the MANUFACTURER in the RISK MANAGEMENT FILE.

SIST EN IEC 80601-2-52:2026

The ACCOMPANYING DOCUMENT shall document any HAZARDS associated with this use case For MEDICAL BEDS intended for APPLICATION ENVIRONMENT 6, the RISK MANAGEMENT shall include:

  • strangulation HAZARDS;
  • snagging HAZARDS;
  • disassembly resulting in choking, ingestion or both;
  • material flammability.

Compliance is checked by inspection of the RISK MANAGEMENT FILE. 201.4.6 ME EQUIPMENT or ME SYSTEM parts that contact the PATIENT Addition:

All parts within the reach of the PATIENT, even if they are underneath the MATTRESS SUPPORT PLATFORM (see Figure 201.118) shall be included in the RISK MANAGEMENT PROCESS for determination if they should be treated as APPLIED PARTS.

For APPLICATION ENVIRONMENT 4 and APPLICATION ENVIRONMENT 6, all ACCESSIBLE PARTS shall be treated as APPLIED PARTS (the entire MEDICAL BED).

201.5 General requirements for testing of ME EQUIPMENT IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, Clause 5 applies, except as follows:

201.5.9.2 ACCESSIBLE PARTS Addition:

NOTE For APPLICATION ENVIRONMENT 6, accessibility of parts considers the use of specialty or safety fasteners and robustness of ENCLOSURE material to prevent access or removal of parts (see 201.15.4.104). Additional subclauses:

201.5.101 * Entrapment test tools Figure 201.104 a) and b) illustrate entrapment test TOOLS (a cone TOOL and a cylinder TOOL respectively).

SIST EN IEC 80601-2-52:2026

Dimensions in millimeters

Key

1 Marked centre line

2 Drilling hole for weight adjustment, if necessary

Total weight 5,13 kg ±0,05 kg

...