ISO 80601-2-12:2023 This document applies to the basic safety and essential performance of a critical care ventilator in combination with its accessories, hereafter referred to as ME equipment:
- intended for use in an environment that provides specialized care for patients whose conditions can be life-threatening and who can require comprehensive care and constant monitoring in a professional healthcare facility;
NOTE 2 For the purposes of this document, such an environment is referred to as a critical care environment. Ventilators for this environment are considered life-sustaining.
NOTE 3 For the purposes of this document, such a critical care ventilator can provide ventilation during transport within a professional healthcare facility (i.e. be a transit-operable ventilator).
NOTE 4 A critical care ventilator intended for use in transport within a professional healthcare facility is not considered as an emergency medical services environment ventilator.
- intended to be operated by a healthcare professional operator; and
- intended for those patients who need differing levels of support from artificial ventilation including for ventilator-dependent patients.
A critical care ventilator is not considered to use a physiologic closed-loop-control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings.
This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system, or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator.
NOTE 5 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.
Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1.
NOTE 6 Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2.
This document is not applicable to ME equipment or an ME system operating in a ventilator-operational mode solely intended for patients who are not dependent on artificial ventilation.
NOTE 7 A critical care ventilator, when operating in such a ventilator-operational mode, is not considered life-sustaining.
This document is not applicable to ME equipment that is intended solely to augment the ventilation of spontaneously breathing patients within a professional healthcare facility.
This document does not specify the requirements for:
NOTE 8 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient.
- ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13;
- ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601-2-84;
- ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2-72;
- ventilators or accessories intended for home-care ventilatory support devices, which are given in ISO 80601-2-79 and ISO 80601-2-80;
- obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70;
- continuous positive airway pressure (CPAP) ME equipment.
- high-frequency ventilators, which are given in ISO 80601‑2‑87;
NOTE 9 A critical care ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilator-operational modes.
- respiratory high-flow therapy equipment, which are given in ISO 80601‑2‑90;
NOTE 10 A critical care ventilator can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathin

  • Standard
    171 pages
    French language
    sale 15% off

ISO 80601-2-84:2023 This document applies to the basic safety and essential performance of an EMS ventilator in combination with its accessories, hereafter also referred to as ME equipment:
- intended for patients who need differing levels of support from artificial ventilation including ventilator-dependent patients;
- intended to be operated by a healthcare professional operator;
- intended for use in the EMS environment; and
- intended for invasive or non-invasive ventilation.
NOTE 2 An EMS ventilator can also be used for transport within a professional healthcare facility.
An EMS ventilator is not considered to use a physiologic closed loop-control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings.
This document is also applicable to those accessories intended by their manufacturer to be connected to the ventilator breathing system, or to an EMS ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the EMS ventilator.
NOTE 3 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.
Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601‑1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1.
NOTE 4 Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2.
This document does not specify the requirements for the following:
NOTE 5 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient.
- ventilators or accessories intended for ventilator-dependent patients in critical care applications, which are given in ISO 80601-2-12.
- ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601-2-72.
- ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13.
- ventilators or accessories intended for ventilatory support equipment (intended only to augment the ventilation of spontaneously breathing patients), which are given in ISO 80601‑2-79 and ISO 80601-2-80.
- obstructive sleep apnoea therapy ME equipment, which are given in ISO 80601‐2‐70.
- user-powered resuscitators, which are given in ISO 10651‐4.
- gas-powered emergency resuscitators, which are given in ISO 10651‐5.
- continuous positive airway pressure (CPAP) ME equipment.
- high‐frequency jet ventilators (HFJVs), which are given in ISO 80601-2-87.
- high‐frequency oscillatory ventilators (HFOVs)[44], which are given in ISO 80601-2-87.
NOTE 6 An EMS ventilator can incorporate high-frequency jet or high-frequency oscillatory ventilation-modes.
- respiratory high-flow therapy equipment, which are given in ISO 80601-2-90.
NOTE 7 An EMS ventilator can incorporate high-flow therapy operational mode, but such a mode is only for spontaneously breathing patients.
- oxygen therapy constant flow ME equipment.
- cuirass or “iron‐lung” ventilators.

  • Standard
    148 pages
    French language
    sale 15% off

IEC 62220-2-1:2023 describes the performance metrics associated with DUAL-ENERGY IMAGING capable DIGITAL X-RAY IMAGING DEVICES meant for medical applications and specifies the methods for their determination. These metrics can be used to analyse TISSUE-SUBTRACTED IMAGES and to evaluate dose performance, noise characteristics, and tissue-subtraction efficacy of DIGITAL X-RAY IMAGING DEVICES. The described methods indicate the procedures to obtain MULTI-SPECTRAL PRIMARY DATA and to compute their derived TISSUE-SUBTRACTED IMAGES.
The intended users of this document are MANUFACTURERS and well-equipped test laboratories. This document is restricted to DIGITAL X-RAY IMAGING DEVICES that are used for single or multiple exposure dual-energy radiographic imaging based on, for example, CR systems, direct and indirect flat panel-detector based systems.
This document excludes and is not applicable to:
– DIGITAL X-RAY IMAGING DEVICES intended to be used in mammography or in dental RADIOGRAPHY;
– slot scanning DIGITAL X-RAY IMAGING DEVICES;
– COMPUTED TOMOGRAPHY or CONE-BEAM COMPUTED TOMOGRAPHY;
– photon-energy discriminating devices such as photon counting X-RAY IMAGING DEVICES;
– devices for dynamic imaging (where series of images are acquired, as in fluoroscopy or cardiac imaging).
– DIGITAL X-RAY IMAGING DEVICES intended to be used with RADIOTHERAPY beams.

  • Standard
    43 pages
    English and French language
    sale 15% off

ISO/IEC 80601-2-72:2023 This document applies to the basic safety and essential performance of a ventilator in combination with its accessories, hereafter referred to as ME equipment:
- intended for use in the home healthcare environment;
- intended for use by a lay operator; and
- intended for those patients who need differing levels of support from artificial ventilation including for ventilator-dependent patients.
This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system or to a ventilator where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator.

  • Standard
    142 pages
    English language
    sale 15% off
  • Standard
    153 pages
    French language
    sale 15% off

IEC 60601-2-46:2023 specifies safety requirements for operating tables, whether or not having electrical parts, including transporters, used for the transportation of the operating table top to or from the base or pedestal of an operating table with detachable operating table top. This particular standard does not apply to
- dental patient chairs (see ISO 7494-1),
- examination chairs and couches,
- patient-supporting systems of diagnostic, interventional and therapeutic equipment (see IEC 60601-2-54 or IEC 60601-2-43),
- operating table heating blankets (see IEC 60601-2-35),
- patient transfer equipment,
- delivery tables and delivery beds,
- medical beds (see IEC 60601-2-52 and EN 50637), and
- field tables.
IEC 60601-2-46:2023 cancels and replaces the third edition published in 2016. This edition constitutes a technical revision. This edition includes the following significant technical change with respect to the previous edition: structural alignment with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020.

  • Standard
    85 pages
    English language
    sale 15% off

IEC 61676:2023 is available as IEC 61676:2023 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 61676:2023 specifies the performance requirements of instruments as used in the non-invasive measurement of X-RAY tube voltage up to 150 kV and the relevant compliance tests. This document also describes the method for calibration and gives guidance for estimating the uncertainty in measurements performed under conditions different from those during calibration. Applications for such measurement are found in diagnostic radiology including mammography, computed tomography (CT), dental radiology and radioscopy. This document is not concerned with the safety aspect of such instruments. The requirements for electrical safety applying to them are contained in IEC 61010-1. IEC 61676:2023 cancels and replaces first edition published in 2002, Amendment 1:2008. This edition constitutes a technical revision. It includes an assessment of the combined standard uncertainty for the performance of a hypothetical instrument for the non-invasive measurement of the tube high voltage (in Annex A) which replaces Annex A of the edition 1.1 titled "Recommended performance criteria for the invasive divider".
The contents of the corrigendum of January 2024 have been included in this copy.

  • Standard
    67 pages
    English and French language
    sale 15% off

ISO 81060-3:2022 This document specifies the requirements and methods for the clinical investigation of continuous automated non-invasive sphygmomanometers used for the measurement of the blood pressure of a patient.
This document does not cover usability aspects such as the form and manner of the data display or output. This document does not specify a numerical threshold on the minimum output period. A continuous automated non-invasive sphygmomanometer providing blood pressure parameters (e.g., systolic blood pressure, diastolic blood pressure or mean arterial pressure) with an output period considerably larger than 30 s is not typically considered a continuous automated non-invasive sphygmomanometer.
This document covers both trending continuous automated non-invasive sphygmomanometers and absolute accuracy continuous automated non-invasive sphygmomanometers and focuses solely on requirements for the clinical investigation. Representation of output is not covered by this document.
NOTE 1 IEC 62366-1 provides requirements on the application of usability engineering to medical devices. The usability engineering process can be used to clarify for the intended user whether the displayed data concerns absolute accurate values or trending values.
The requirements and methods for the clinical investigation of continuous automated non-invasive sphygmomanometers provided in this document are applicable to any subject population, and any condition of use of the continuous automated non-invasive sphygmomanometers.
NOTE 2 Subject populations can, for example, be represented by age or weight ranges.
NOTE 3 This document does not provide a method to assess the effect of artefacts during the clinical investigation (e.g. motion artefacts induced by the movement of the subject or the movement of the platform supporting the subject).
This document specifies additional disclosure requirements for the accompanying documents of continuous automated non-invasive sphygmomanometers that have undergone clinical investigation according to this document.
This document is not applicable to:
- the clinical investigation of a non-automated sphygmomanometer as given in ISO 81060-1,
- the clinical investigation of an intermittent automated non-invasive sphygmomanometer as given in ISO 81060-2,
- an automated non-invasive sphygmomanometer as given in IEC 80601-2-30, or
- invasive blood pressure monitoring equipment as given in IEC 60601‑2‑34.

  • Standard
    36 pages
    English language
    sale 15% off
  • Standard
    36 pages
    French language
    sale 15% off

IEC 60601-2-43:2022 is available as IEC 60601-2-43:2022 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-43:2022 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of both FIXED and MOBILE X RAY EQUIPMENT declared by the MANUFACTURER to be suitable for RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, hereafter referred to as INTERVENTIONAL X-RAY EQUIPMENT. Its scope excludes, in particular:
- equipment for RADIOTHERAPY;
- equipment for COMPUTED TOMOGRAPHY;
- ACCESSORIES intended to be introduced into the PATIENT;
- mammographic X RAY EQUIPMENT;
- dental X RAY EQUIPMENT.
INTERVENTIONAL X-RAY EQUIPMENT declared by the MANUFACTURER to be suitable for RADIOSCOPICALLY GUIDED INTERVENTIONAL PROCEDURES, which does not include a PATIENT SUPPORT as part of the system, is exempt from the PATIENT SUPPORT provisions of this document.
If a clause or subclause is specifically intended to be applicable to INTERVENTIONAL X-RAY EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to INTERVENTIONAL X-RAY EQUIPMENT and to ME SYSTEMS, as relevant.
IEC 60601-2-54 applies only with regards to the cited subclauses; non-cited subclauses of IEC 60601-2-54 do not apply.
IEC 60601-2-43:2022 cancels and replaces the second edition published in 2010, Amendment 1:2017 and Amendment 2:2019. This edition constitutes a technical revision.
This edition includes editorial and technical changes to reflect the IEC 60601-1:2005/AMD2:2020 and IEC 60601-2-54:2022. It also contains corrections and technical improvements. Significant technical changes with respect to the previous edition are as follows:
a) a new specific term DOSIMETER is introduced to replace the general term DOSEMETER as in IEC 60601-2-54:2022;
b) several terms and definitions that are moved from IEC TR 60788:2004 to 201.3 of IEC 60601-2-54:2022 are also referenced from IEC 60601-2-54:2022.
c) the collateral standards IEC 60601-1-11:2015, IEC 60601-1-11:2015/AMD1:2020, IEC 60601-1-12:2014 and IEC 60601-1-12:2014/AMD1:2020 are applicable if MANUFACTURER so declares;
d) the former subclause 201.11.101 “Protection against excessive temperature of X-RAY TUBE ASSEMBLIES” is removed since covered by IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 and IEC 60601-2-28:2017, and the former subclause 201.11.102 is renumbered as 201.11.101, as in IEC 60601-2-54:2022;
e) to adopt changes in subclause 7.8.1 “Colours of indicator lights” in IEC 60601 1:2005/AMD2:2020, clarification of requirements is provided in 201.7.8.1 to avoid conflicts with requirements of indicator lights stipulated for X-RAY EQUIPMENT, as in IEC 60601-2-54:2022;
f) explanation of the term ESSENTIAL PERFORMANCE is provided in Annex AA to emphasize the performance of the clinical function under NORMAL CONDITIONS and SINGLE FAULT CONDITIONS.

  • Standard
    126 pages
    English and French language
    sale 15% off

IEC 60806:2022 is applicable to X-RAY SOURCE ASSEMBLIES and X-RAY TUBE ASSEMBLIES.
This document specifies a method for the determination of the greatest geometrically symmetrical RADIATION FIELD at a specified distance from the FOCAL SPOT for which the percentage AIR KERMA RATE along the major axes of the RADIATION FIELD does not fall below a permitted value.
In case multiple FOCAL SPOTS are not super-imposed, each FOCAL SPOT has its own REFERENCE AXIS. Then the maximum RADIATION FIELD can be given for each FOCAL SPOT separately.
IEC 60806:2022 cancels and replaces the first edition published in 1984. This edition constitutes a technical revision.
This edition includes the following significant technical change with respect to the previous edition:
a) addition of solid state detectors as they have become more common since the first edition of 1984.

  • Standard
    21 pages
    English and French language
    sale 15% off

IEC TR 60878:2022 provides a compilation, for easy reference, of graphical symbols (graphics, title, description) and safety signs for medical electrical equipment. The graphical symbols are grouped in sections according to their specific field of application (see Clause 4).
IEC TR 60878:2022 cancels and replaces the third edition published in 2015. This fourth edition constitutes a technical revision.
The main changes compared with IEC TR 60878:2015 are as follows:
• A total of 109 new symbols and safety signs that have been identified since the publication of the third edition have been added. For identification, the number of the new symbol or safety sign is printed in red followed by “New”.
• Of the symbols and safety signs in the third edition, 14 have changes in their title or description. For identification, the number of a modified symbol or safety sign is printed in red followed by “Mod”.
The contents of the corrigendum of February 2023 have been included in this copy.

  • Technical report
    226 pages
    English language
    sale 15% off

IEC 60601-2-54:2022 is available as IEC 60601-2-54:2022 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 60601-2-54:2022 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ME EQUIPMENT and ME SYSTEMS intended to be used for projection RADIOGRAPHY and INDIRECT RADIOSCOPY. IEC 60601-2-43 applies to ME EQUIPMENT and ME SYSTEMS intended to be used for interventional applications and refers to applicable requirements in this document. ME EQUIPMENT and ME SYSTEMS intended to be used for bone or tissue absorption densitometry, computed tomography, mammography or dental or radiotherapy applications are excluded from the scope of this document. The scope of this document also excludes radiotherapy simulators. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. IEC 60601-2-54:2022 edition cancels and replaces the first edition published in 2009, Amendment 1:2015 and Amendment 2:2018. This edition constitutes a technical revision. This edition includes editorial and technical changes to reflect the IEC 60601 1:2005/AMD2:2020. It also contains corrections and technical improvements. Significant technical changes with respect to the previous edition are as follows:
a) a new specific term DOSIMETER is introduced to replace the general term DOSEMETER;
b) terms and definitions taken exclusively from IEC TR 60788:2004 and which are specifically applicable in this document have been moved to 201.3;
c) the collateral standards IEC 60601-1-11:2015, IEC 60601-1-11:2015/AMD1:2020, IEC 60601-1-12:2014 and IEC 60601-1-12:2014/AMD1:2020 are applicable if MANUFACTURER so declares;
d) the subclause 201.11.101 “Protection against excessive temperatures of X-ray tube assemblies” has been removed from this document as its requirements are sufficiently and clearly covered by IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 and IEC 60601-2-28:2017;
e) to adopt changes which are introduced with respect to indicator lights in 7.8.1 of the IEC 60601-1:2005/AMD2:2020 clarification of requirements is provided to avoid conflicts with requirements of indicator lights stipulated for X-RAY EQUIPMENT;
f) explanation of the term ESSENTIAL PERFORMANCE is provided in Annex AA to emphasize the performance of the clinical function under NORMAL and SINGLE FAULT CONDITIONS. CONDITIONS.

  • Standard
    243 pages
    English language
    sale 15% off

IEC 60601-2-33:2022 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of MAGNETIC RESONANCE (MR) EQUIPMENT and MAGNETIC RESONANCE (MR) SYSTEMS.
This document does not cover the application of MR EQUIPMENT beyond the INTENDED USE.
If a clause or subclause is specifically intended to be applicable to MR EQUIPMENT only, or to MR SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to MR EQUIPMENT and to MR SYSTEMS, as relevant.
This document does not formulate additional specific requirements for MR EQUIPMENT or MR SYSTEMS used in INTERVENTIONAL MR EXAMINATIONS.
IEC 60601-2-33:2022 cancels and replaces the third edition published in 2010, Amendment 1:2013 and Amendment 2:2015. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) aligned with IEC 60601-1:2005 and its two amendments IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020;
b) addition of safety requirements for the EMERGENCY FIELD SHUT DOWN UNIT;
c) clarification of acoustic protection measures for the PATIENT and MR WORKER;
d) addition of noise emission declaration for exposure inside the MR EXAMINATION ROOM, to support occupational health assessment by the RESPONSIBLE ORGANIZATION;
e) addition of compliance methods for thermal safety of RF coils;
f) addition of RF transmit definitions to match MR CONDITIONAL labelling requirements for MEDICAL DEVICES;
g) clarification of requirements for MR CONDITIONAL labelling of ACCESSORIES;
h) alignment of static magnetic field limit for B0 HAZARD area to limits in other MEDICAL DEVICE standards (especially that for pacemakers, ISO 14117), the new limit value being 0,9 mT;
i) improved description of the magnetic field related plots in the Compatibility Technical Specification Sheet (CTSS);
j) provision of compatibility sequences (in the CTSS) to test auxiliary equipment by the MR manufacturer has become optional, and is expected to be eliminated in a future edition;
k) a separate section with requirements for a site-planning document containing safety information;
l) requirements for the alerting function (PATIENT to OPERATOR);
m) introduction of MROC as mandatory functionality for 1,5 T and 3 T systems to facilitate scanning of PATIENTs with MEDICAL DEVICES labelled as MR CONDITIONAL, unless such scanning is explicitly contra-indicated by the MR MANUFACTURER;
n) RF coil symbols in Table 201.A.102 have become mandatory, and the preferred and alternate signs have been swapped relative to the previous edition, with preferred now being the sign with color;
o) determination of the B1 stray field in 201.12.4.105.3.3 based on calculations only
The contents of the Interpretation Sheet 1 of May 2023 have been included in this copy.

  • Standard
    302 pages
    English and French language
    sale 15% off

ISO 80601-2-13:2022 This document is applicable to the basic safety and essential performance of an anaesthetic workstation for administering inhalational anaesthesia whilst continuously attended by a professional operator.
This document specifies particular requirements for a complete anaesthetic workstation and the following anaesthetic workstation components which, although considered as individual devices in their own right, may be utilized, in conjunction with other relevant anaesthetic workstation components, to form an anaesthetic workstation to a given specification:
- anaesthetic gas delivery system;
- anaesthetic breathing system;
- anaesthetic gas scavenging system (AGSS);
- anaesthetic vapour delivery system;
- anaesthetic ventilator;
- monitoring equipment;
- alarm system;
- protection device.
NOTE 1 Monitoring equipment, alarm systems and protection devices are summarized in Table AA.1.
An anaesthetic workstation supplied complete and its individual components are considered as ME equipment or ME systems with regard to the general standard.
NOTE 2 The applicability of this document is indicated in Table AA.2.
This document is also applicable to those accessories intended by their manufacturer to be connected to an anaesthetic workstation where the characteristics of those accessories can affect the basic safety and essential performance of the anaesthetic workstation.
If a clause or subclause is specifically intended to be applicable to anaesthetic workstation components or its accessories only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to an anaesthetic workstation and its individual components including accessories, as relevant.
Hazards inherent in the intended physiological function of an anaesthetic workstation and its individual components including accessories within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1.
NOTE 3 See also IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2.
This document is not applicable to any anaesthetic workstation intended for use with flammable anaesthetic agents, as determined by Annex BB.

  • Standard
    103 pages
    English language
    sale 15% off

IEC 61675-1:2022 is available as IEC 61675-1:2022 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 61675-1:2022 specifies terminology and test methods for declaring the characteristics of POSITRON EMISSION TOMOGRAPHS. POSITRON EMISSION TOMOGRAPHS detect the ANNIHILATION RADIATION of positron emitting RADIONUCLIDES by COINCIDENCE DETECTION. IEC 61675-1:2022 cancels and replaces the second edition published in 2013. This edition constitutes a technical revision. This edition includes the following significant technical change with respect to the previous edition: requirements have been changed or newly created regarding the technical aspects of SPATIAL RESOLUTION, sensitivity measurement, SCATTER FRACTION, COUNT RATE performance, image quality, PET/CT registration accuracy and time-of-flight resolution.

  • Standard
    87 pages
    English and French language
    sale 15% off

IEC 81001-5-1:2021 defines the LIFE CYCLE requirements for development and maintenance of HEALTH SOFTWARE needed to support conformance to IEC 62443-4-1 – taking the specific needs for HEALTH SOFTWARE into account. The set of PROCESSES, ACTIVITIES, and TASKS described in this document establishes a common framework for secure HEALTH SOFTWARE LIFE CYCLE PROCESSES.
The purpose is to increase the CYBERSECURITY of HEALTH SOFTWARE by establishing certain ACTIVITIES and TASKS in the HEALTH SOFTWARE LIFE CYCLE PROCESSES and also by increasing the SECURITY of SOFTWARE LIFE CYCLE PROCESSES themselves.
It is important to maintain an appropriate balance of the key properties SAFETY, effectiveness and SECURITY as discussed in ISO 81001-1.
This document excludes specification of ACCOMPANYING DOCUMENTATION contents.

  • Standard
    114 pages
    English and French language
    sale 15% off

IEC 61223-3-7:2021 applies to DENTAL CONE-BEAM COMPUTED TOMOGRAPHY X-RAY EQUIPMENT, hereafter also called DENTAL CBCT EQUIPMENT, that conforms to IEC 60601-2-63:2012 and IEC 60601-2-63:2012/AMD1:2017.
This document applies to ACCEPTANCE TESTS and CONSTANCY TESTS on DENTAL CONE-BEAM COMPUTED TOMOGRAPHY X-RAY EQUIPMENT.
The aim of ACCEPTANCE TESTS is to verify compliance of the installation or MAJOR SERVICE ACTION with specifications affecting the image quality, RADIATION OUTPUT and PATIENT positioning.
The requirements specified in this document are minimal requirements. The MANUFACTURER can establish criteria for the tests described here that exceed the levels contained in this document.
CONSTANCY TESTS are performed to ensure that the functional performance of ME EQUIPMENT meets established criteria and to enable the early recognition of changes in the properties of components of the ME EQUIPMENT, and to verify compliance with specifications affecting the image quality, RADIATION OUTPUT and PATIENT positioning.
This document also contains requirements for the ACCOMPANYING DOCUMENTS associated with ACCEPTANCE AND CONSTANCY TESTING of the DENTAL CBCT EQUIPMENT.
This document does not apply to:
- aspects of thermal, EMD (electromagnetic disturbances), mechanical and electrical safety;
- aspects of mechanical, electrical and software performance, unless they are essential for performing the ACCEPTANCE TESTS and CONSTANCY TESTS, and directly affect image quality, RADIATION OUTPUT and PATIENT positioning.
Equipment in the scope of IEC 61223-3-5 is excluded from the scope of this document.
DENTAL EXTRA-ORAL X-RAY EQUIPMENT can provide modalities which are in the scope of IEC 61223-3-4. In this case, the respective clauses of the IEC 61223-3-4 apply.
The object of this document is to establish:
- the essential parameters which describe the performance of DENTAL CBCT EQUIPMENT with regard to the image quality, RADIATION OUTPUT and PATIENT positioning;
- methods of testing and whether measured quantities related to those parameters comply with the specified requirements.
These methods rely on non-invasive measurements performed once the installation or a MAJOR SERVICE ACTION is completed.

  • Standard
    83 pages
    English and French language
    sale 15% off

IEC 62563-2:2021 establishes the performance CRITERIA and test frequencies for the ACCEPTANCE TESTS and CONSTANCY TESTS. The evaluation methods are defined in IEC 62563-1. The scope of this document is directed to practical tests that can be visually evaluated or measured using basic test equipment. This document applies to medical IMAGE DISPLAY SYSTEMS, which can display monochrome image information in the form of greyscale values on colour and greyscale IMAGE DISPLAY SYSTEMS. This document does not apply to information displays and to displays used solely for control of technical settings of all medical information.

  • Standard
    71 pages
    English and French language
    sale 15% off

IEC 80001-1:2021 specifies general requirements for ORGANIZATIONS in the application of RISK MANAGEMENT before, during and after the connection of a HEALTH IT SYSTEM within a HEALTH IT INFRASTRUCTURE, by addressing the KEY PROPERTIES of SAFETY, EFFECTIVENESS and SECURITY whilst engaging appropriate stakeholders.
IEC 80001-1:2021 cancels and replaces the first edition published in 2010. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) structure changed to better align with ISO 31000;
b) establishment of requirements for an ORGANIZATION in the application of RISK MANAGEMENT;
c) communication of the value, intention and purpose of RISK MANAGEMENT through principles that support preservation of the KEY PROPERTIES during the implementation and use of connected HEALTH SOFTWARE and/or HEALTH IT SYSTEMS.

  • Standard
    75 pages
    English and French language
    sale 15% off

This document applies to the basic safety and essential performance of respiratory high-flow therapy equipment, as defined in 201.3.220, hereafter also referred to as ME equipment or ME system, in combination with its accessories:
- intended for use with patients who can breathe spontaneously; and
- intended for patients who would benefit from improved alveolar gas exchange; and who would benefit from receiving high-flow humidified respiratory gases, which can include a patient whose upper airway is bypassed.
EXAMPLE 1 Patients with Type 1 Respiratory Failure who exhibit a reduction in arterial blood oxygenation.
EXAMPLE 2 Patients who would benefit from reduced work of breathing, as needed in Type 2 Respiratory Failure, where arterial carbon dioxide is high.
EXAMPLE 3 Patients requiring humidification to improve mucociliary clearance.
Respiratory high-flow therapy equipment can be intended for use in the home healthcare environment or intended for use in professional healthcare facilities.
NOTE 1 In the home healthcare environment, the supply mains is often not reliable.
Respiratory high-flow therapy equipment can be:
- fully integrated ME equipment; or
- a combination of separate items forming a ME system.
This standard also applies to other types of respiratory equipment when that equipment includes a respiratory high-flow therapy mode.
NOTE 2 This standard and ISO 80601-2-12[14] are applicable to a critical care ventilator with a high-flow therapy mode.
Respiratory high-flow therapy equipment can be transit-operable.
This document is also applicable to those accessories intended by their manufacturer to be connected to the respiratory high-flow therapy equipment, where the characteristics of those accessories can affect the basic safety or essential performance of the respiratory high-flow therapy equipment.
EXAMPLE 4 Breathing sets, connectors, humidifier, breathing system filter, external electrical power source, distributed alarm system, high-flow nasal cannula, tracheal tube, tracheostomy tube, face mask and supra-laryngeal airway.
If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.
Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in the general standard, 7.2.13 and 8.4.1.
NOTE 3 Additional information can be found in the general standard, 4.2.
This document does not specify the requirements for:
- ventilators or accessories for ventilator-dependent patients intended for critical care applications, which are given in ISO 80601‑2‑12[14];
- ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601‑2‑13[15];
- ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601‑2‑84[20];
- ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601‑2‑72[17];
- ventilatory support equipment or accessories intended for patients with ventilatory impairment, which are given in ISO 80601‑2‑79[18];
- ventilatory support equipment or accessories intended for patients with ventilatory insufficiency, which are given in ISO 80601‑2‑80[19];
- sleep apnoea therapy ME equipment, which are given in ISO 80601‑2‑70[16];
- continuous positive airway pressure (CPAP) ME equipment;
- high-frequency jet ventilators (HFJVs)[31], which are given in ISO 80601‑2‑87[21];
- gas mixers for medical use, which are given in ISO 11195[9];
- flowmeters, which are given in ISO 15002[11];
- high-frequency oscillatory ventilators (HFOVs), which are given in ISO 80601‑2‑87[21]; and
- cuirass or “iron-lung” venti

  • Standard
    89 pages
    English language
    sale 15% off

This document provides quality requirements for health apps and defines a health app quality label in order to visualize the quality and reliability of health apps.
This document is applicable to health apps, which are a special form of health software. It covers the entire life cycle of health apps.
This document is intended for use by app manufacturers as well as app assessment organizations in order to communicate the quality and reliability of a health app. Consumers, patients, carers, health care professionals and their organizations, health authorities, health insurers and the wider public can use the health app quality label and report when recommending or selecting a health app for use, or for adoption in care guidelines, care pathways and care contracts.
NOTE 1 Health apps can be subject to national legislation, such as for medical devices.
NOTE 2 See Annex C for additional details on the scope.
Outside the scope of this document are guidelines to comply to the medical device regulation.

  • Technical specification
    78 pages
    English language
    sale 15% off

This document applies to the basic safety and essential performance of a humidifier, also hereafter referred to as ME equipment, in combination with its accessories, the combination also hereafter referred to as ME system.
This document is also applicable to those accessories intended by their manufacturer to be connected to a humidifier where the characteristics of those accessories can affect the basic safety or essential performance of the humidifier.
EXAMPLE 1 Heated breathing tubes (heated-wire breathing tubes) or ME equipment intended to control these heated breathing tubes (heated breathing tube controllers).
NOTE 1 Heated breathing tubes and their controllers are ME equipment and are subject to the requirements of IEC 60601-1.
NOTE 2 ISO 5367 specifies other safety and performance requirements for breathing tubes.
This document includes requirements for the different medical uses of humidification, such as invasive ventilation, non-invasive ventilation, nasal high-flow therapy, and obstructive sleep apnoea therapy, as well as humidification therapy for tracheostomy patients.
NOTE 3 A humidifier can be integrated into other equipment. When this is the case, the requirements of the other equipment also apply to the humidifier.
EXAMPLE 2 Heated humidifier incorporated into a critical care ventilator where ISO 80601-2-12[10] also applies.
EXAMPLE 3 Heated humidifier incorporated into a homecare ventilator for dependent patients where ISO 80601-2-72[12] also applies.
EXAMPLE 4 Heated humidifier incorporated into sleep apnoea therapy equipment where ISO 80601-2-70[11] also applies.
EXAMPLE 5 Heated humidifier incorporated into ventilatory support equipment where either ISO 80601-2-79[13] or ISO 80601-2-80[14] also apply.
EXAMPLE 6 Heated humidifier incorporated into respiratory high-flow therapy equipment where ISO 80601-2-90[15] also applies.
This document also includes requirements for an active HME (heat and moisture exchanger), ME equipment which actively adds heat and moisture to increase the humidity level of the gas delivered from the HME to the patient. This document is not applicable to a passive HME, which returns a portion of the expired moisture and heat of the patient to the respiratory tract during inspiration without adding heat or moisture.
NOTE 4 ISO 9360-1 and ISO 9360-2[4] specify safety and performance requirements for a passive HME.
NOTE 5 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.
Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1.
NOTE 6 Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2.
This document does not specify the requirements for cold pass-over or cold bubble-through humidification devices, the requirements for which are given in ISO 20789[6].
This document is not applicable to equipment commonly referred to as “room humidifiers” or humidifiers used in heating, ventilation and air conditioning systems, or humidifiers incorporated into infant incubators.
This document is not applicable to nebulizers used for the delivery of a drug to patients.
NOTE 7 ISO 27427[7] specifies the safety and performance requirements for nebulizers.

  • Standard
    99 pages
    English language
    sale 15% off

This document specifies dimensions and requirements for the design and functional performance of small-bore connectors intended to be used for connections in intravascular applications or hypodermic connections in hypodermic applications of medical devices and accessories.
EXAMPLES Hypodermic syringes and needles or intravascular (IV) cannulae with male and female Luer slip connectors and Luer lock connectors.
NOTE 1 See Annex A.
NOTE 2 The Luer connector was originally designed for use at pressures up to 300 kPa.
This document does not specify requirements for the medical devices or accessories that use these connectors. Such requirements are given in particular documents for specific medical devices or accessories.
This document does not specify requirements for the following small-bore connectors, which are specified in other documents:
- haemodialyser, haemodiafilter and haemofilter blood compartment ports (ISO 8637 [5] and applicable portion of ISO 8638 [6] referencing blood compartment ports);
- haemodialysis, haemodiafiltration and haemofiltration equipment connectors (ISO 8637 [5]);
- infusion system closure piercing connectors (ISO 8536-€‘4 [4]).
NOTE 3 Manufacturers are encouraged to incorporate the small-bore connectors specified in this document into medical devices or accessories, even if currently not required by the relevant particular medical device documents. It is expected that when the relevant particular medical device documents are revised, requirements for small-bore connectors, as specified in ISO 80369, will be included.
NOTE 4 ISO 80369-€‘1:2018, Clause 7, specifies alternative methods of conformance with ISO 80369-€‘1:2018, for small-bore connectors intended for use with intravascular applications or hypodermic application medical devices or accessories, which do not conform with this document.

  • Standard
    47 pages
    English language
    sale 15% off

This document applies to the basic safety and essential performance of a high-frequency ventilator (HFV) in combination with its accessories, hereafter referred to as ME equipment:
- intended for use in an environment that provides specialized care for patients whose conditions can be life-threatening and who can require comprehensive care and constant monitoring in a professional healthcare facility;
NOTE 1 For the purposes of this document, such an environment is referred to as a critical care environment. High-frequency ventilators for this environment are considered life-sustaining.
NOTE 2 For the purposes of this document, such a high-frequency ventilator can provide transport within a professional healthcare facility (i.e., be a transit-operable ventilator).
NOTE 3 A high-frequency ventilator intended for use in transport within a professional healthcare facility is not considered as a ventilator intended for the emergency medical services environment.
- intended to be operated by a healthcare professional operator;
- intended for those patients who need differing levels of support from artificial ventilation including ventilator-dependent patients; and
- capable of providing more than 150 inflations/min.
There are three principal designations of HFV:
- high-frequency percussive ventilation [HFPV, with a typical HFV frequency of (60 to 1 000) HFV inflations/min];
- high-frequency jet ventilation [HFJV, with a typical HFV frequency of (100 to 1 500) HFV inflations/min]; and
- high-frequency oscillatory ventilation [HFOV, with a typical HFV frequency of (180 to 1200) HFV inflations/min and typically having an active expiratory phase].
Additionally, HFV designations can be combined together or with ventilation at rates less than 150 inflations/min.
*A high-frequency ventilator is not considered a physiologic closed loop-control system unless it uses a physiological patient variable to adjust the ventilation therapy settings.
This document is also applicable to those accessories intended by their manufacturer to be connected to an HFV breathing system, or to a high-frequency ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the high-frequency ventilator.
If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.
Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of IEC 60601-1:2005.
NOTE 4 Additional information can be found in 4.2 of IEC 60601-1:2005+AMD1:2012.
This document is not applicable to ME equipment that is intended solely to augment the ventilation of spontaneously breathing patients within a professional healthcare facility.
This document does not specify the requirements for:
- non-high-frequency ventilators or accessories which provide conventional ventilation for use in critical care environments, which are given in ISO 80601-2-12 [23];.
NOTE 5 An HFV can incorporate conventional critical care ventilator operational modes, in which case ISO 80601-2-12 is applicable to those modes.
- ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601-2-13 [24];
- ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601-2-84, the replacement for ISO 10651-3 [13];
NOTE 6 An HFV can incorporate EMS ventilator capability.
- ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601-€‘2-72 [26];
- ventilators or accessories intended for home-care ventilatory support devices, which are given in ISO

  • Standard
    136 pages
    English language
    sale 15% off

This document provides the principles, concepts, terms and definitions for health software and health IT systems, key properties of safety, effectiveness and security, across the full life cycle, from concept to decommissioning, as represented in Figure 1. It also identifies the transition points in the life cycle where transfers of responsibility occur, and the types of multi-lateral communication that are necessary at these transition points. This document also establishes a coherent concepts and terminology for other standards that address specific aspects of the safety, effectiveness, and security (including privacy) of health software and health IT systems.
This document is applicable to all parties involved in the health software and health IT systems life cycle including the following:
a) Organizations, health informatics professionals and clinical leaders designing, developing, integrating, implementing and operating health software and health IT systems – for example health software developers and medical device manufacturers, system integrators, system administrators (including cloud and other IT service providers);
b) Healthcare service delivery organizations, healthcare providers and others who use health software and health IT systems in providing health services;
c) Governments, health system funders, monitoring agencies, professional organizations and customers seeking confidence in an organization’s ability to consistently provide safe, effective and secure health software, health IT systems and services;
d) Organizations and interested parties seeking to improve communication in managing safety, effectiveness and security risks through a common understanding of the concepts and terminology used in safety, effectiveness and security management;
e) Providers of training, assessment or advice in safety, effectiveness and security risk management for health software and health IT systems;
f) Developers of related safety, effectiveness and security standards.

  • Standard
    61 pages
    English language
    sale 15% off

This document applies to basic safety and essential performance of cerebral tissue oximeter equipment, that employs light at multiple wavelengths to derive a quantitative measure of oxygen saturation of haemoglobin within the volume of tissue sampled under the probe attached to the head. The cerebral tissue oximeter equipment can be based on continuous light, frequency domain or time domain technologies. This document applies to ME equipment used in a hospital environment as well as when used outside the hospital environment, such as in ambulances and air transport. Additional standards may apply to ME equipment for those environments of use.
NOTE 1 Cerebral tissue oximeters are sometimes referred to as near infrared spectroscopy equipment in medical literature.
Not included within the scope of this document are:
- invasive tissue or vascular oximeters;
- oximeters that require a blood sample from the patient;
- equipment measuring dissolved oxygen;
- ME equipment, or part thereof, that measures path-length-dependent haemoglobin change. The requirements for functional near-infrared spectroscopy equipment are found in ISO 80601-2-71[4];
- ME equipment, or part thereof, that measures arterial saturation based on pulsatile changes in tissue optical properties (SpO2). The requirements for pulse oximeter equipment are found in ISO 80601‑2‑61[3];
- ME equipment, or any part thereof, that claims to monitor tissue in parts of the body other than the head.
This document also applies to cerebral tissue oximeter equipment, including cerebral tissue oximeter monitors, cerebral tissue oximeter probes and probe cable extenders, that have been remanufactured.
If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.
Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 201.11 and in 201.7.2.13 and 201.8.4.1 of the general standard.
NOTE 2 See also 4.2 of the general standard.
This document can also be applied to ME equipment and their accessories used for compensation or alleviation of disease, injury or disability.
This document is not applicable to remote or slave (secondary) equipment that displays StO2 values that are located outside of the patient environment.
NOTE 3 ME equipment that provides selection between diagnostic and monitoring functions is expected to meet the requirements of the appropriate document when configured for that function.

  • Standard
    107 pages
    English language
    sale 15% off

IEC TR 60601-4-5:2021 provides detailed technical specifications for SECURITY features of MEDICAL DEVICES used in MEDICAL IT-NETWORKS. MEDICAL DEVICES dealt with in this document include MEDICAL ELECTRICAL EQUIPMENT, MEDICAL ELECTRICAL SYSTEMS and MEDICAL DEVICE SOFTWARE. MEDICAL DEVICE SOFTWARE, although not in the scope of IEC 60601 (all parts), can also make use of this document. Based on the seven foundational requirements described in the state-of-the-art document IEC TS 62443 1 1:2009, this document provides specifications for different MEDICAL DEVICE capability SECURITY LEVELS (SL C). The specified SECURITY capabilities of a MEDICAL DEVICE can be used by various members of the medical community to integrate the device correctly into defined SECURITY ZONES and CONDUITS of a MEDICAL IT-NETWORK with an appropriate MEDICAL IT NETWORK's target SECURITY LEVEL (SL T).
This document is applicable to MEDICAL DEVICES with external data interface(s), for example when connected to a MEDICAL IT-NETWORK or when a human interface is used for processing – e.g. entering, capturing or viewing – CONFIDENTIAL DATA.
This document does not apply to other software used on a MEDICAL IT-NETWORK which does not meet the definition of MEDICAL DEVICE SOFTWARE.
This document does not apply to in-vitro diagnostic devices (IVD).

  • Technical report
    51 pages
    English language
    sale 15% off

IEC 60336:2020 applies to FOCAL SPOTS in medical diagnostic X-RAY TUBE ASSEMBLIES for medical use, operating at X-RAY TUBE VOLTAGES up to and including 150 kV.
This document describes the test methods employing digital detectors for determining:
a) FOCAL SPOT dimensions in terms of NOMINAL FOCAL SPOT VALUES, ranging from 0,1 to 3,0;
b) LINE SPREAD FUNCTIONS;
c) one-dimensional MODULATION TRANSFER FUNCTIONS;
d) FOCAL SPOT PINHOLE RADIOGRAMS,
and the means for indicating compliance.
In informative annexes, STAR PATTERN imaging and BLOOMING VALUE are described.
IEC 60336:2020 cancels and replaces the fourth edition published in 2005. This edition constitutes a technical revision.
The significant changes of this fifth edition with respect to the previous edition are detailed in Clause E.6. These changes are:
a) introduction of digital detectors and discretization errors;
b) fewer normative requirements;
c) support for both SLIT CAMERA and PINHOLE CAMERA;
d) reintroduction of distorted (skewed) FOCAL SPOT;
e) keeping of STAR PATTERNS and BLOOMING VALUE as informative.
The contents of the corrigendum of June 2022 have been included in this copy.

  • Standard
    90 pages
    English and French language
    sale 15% off

IEC 60522-1:2020 applies to X-RAY TUBE ASSEMBLIES and to FILTERING MATERIAL, in medical diagnostic applications up to a HIGH VOLTAGE of 150 kV. For HIGH VOLTAGES greater than 50 kV, this document applies to X-RAY TUBE ASSEMBLIES with tungsten or tungsten-alloy TARGETS only.
IEC 60522-1:2021 defines the concept of PERMANENT FILTRATION of X-RAY TUBE ASSEMBLIES, and it defines the term FILTERING MATERIAL.
Methods are given to determine the PERMANENT FILTRATION of an X-RAY TUBE ASSEMBLY and for determining the QUALITY EQUIVALENT FILTRATION of FILTERING MATERIALS.
It contains requirements for statements of compliance of X-RAY TUBE ASSEMBLIES in ACCOMPANYING DOCUMENTS and for markings on X-RAY TUBE ASSEMBLIES, and for indications and statements of compliance of FILTERING MATERIAL.
IEC 60522-1:2020 cancels and replaces the second edition of IEC 60522 published in 1999. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the IEC 60522:1999:
The scope of the IEC 60522-1 has been changed with respect to second edition of the IEC 60522 as follows:
a) As radiotherapy standards do not reference IEC 60522, radiotherapy is no longer in the scope. Consequently, the HIGH VOLTAGE is limited to 150 kV, and copper is no longer used as reference material.
b) While IEC 60522:1999 covers only PERMANENT FILTRATION, IEC 60522-1 also covers quite generally “material filtering the X-RAY BEAM incident on the PATIENT”. This concerns materials like ADDED FILTERS, table-tops, a breast COMPRESSION DEVICE, and materials in the BEAM LIMITING DEVICE. For these materials the defined term FILTERING MATERIAL has been introduced.
c) In order to provide technical and scientific background and rationale on the content of IEC 60522-1, IEC TR 60522-2 was introduced.

  • Standard
    37 pages
    English and French language
    sale 15% off

This document is applicable to the basic safety and essential performance of sleep apnoea breathing therapy equipment, hereafter referred to as ME equipment, intended to alleviate the symptoms of patients who suffer from obstructive sleep apnoea by delivering a therapeutic breathing pressure to the respiratory tract of the patient. Sleep apnoea breathing therapy equipment is intended for use in the home healthcare environment by lay operators as well as in professional healthcare institutions. * Sleep apnoea breathing therapy equipment is not considered to utilize a physiologic closed-loop-control system unless it uses a physiological patient variable to adjust the therapy settings. This document excludes sleep apnoea breathing therapy equipment intended for use with neonates. This document is applicable to ME equipment or an ME system intended for those patients who are not dependent on mechanical ventilation. This document is not applicable to ME equipment or an ME system intended for those patients who are dependent on mechanical ventilation such as patients with central sleep apnoea. This document is also applicable to those accessories intended by their manufacturer to be connected to sleep apnoea breathing therapy equipment, where the characteristics of those accessories can affect the basic safety or essential performance of the sleep apnoea breathing therapy equipment. Masks and application accessories intended for use during sleep apnoea breathing therapy are additionally addressed by ISO 17510. Refer to Figure AA.1 for items covered further under this document. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard. NOTE See also 4.2 of the general standard. This document is not applicable to high-frequency jet ventilators (HFJVs) or high-frequency oscillatory ventilators (HFOVs), which are given in ISO 80601-2-87[13]. This document does not specify the requirements for ventilators or accessories intended for critical care ventilators for ventilator-dependent patients, which are given in ISO 80601‑2‑12. This document does not specify the requirements for ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601-2-13[8]. This document does not specify the requirements for ventilators or accessories intended for home care ventilators for ventilator-dependent patients, which are given in ISO 80601-2-72[9]. This document does not specify the requirements for ventilators or accessories intended for emergency and transport, which are given in ISO 80601-2-84[12]. This document does not specify the requirements for ventilators or accessories intended for home-care ventilatory support, which are given in ISO 80601-2-79[10] and ISO 80601‑2‑80[11].

  • Standard
    68 pages
    English language
    sale 15% off

This document specifies requirements for the basic safety and essential performance of an oxygen concentrator in combination with its accessories, hereafter referred to as ME equipment, intended to increase the oxygen concentration of gas intended to be delivered to a single patient. Such oxygen concentrators are typically intended for use in the home healthcare environment by a single patient in various environments including any private and public transportation as well as in commercial aircraft.
NOTE 1 Such oxygen concentrators can also be used in professional healthcare facilities.
This document is applicable to a transit-operable and non-transit-operable oxygen concentrator. This document is applicable to an oxygen concentrator integrated into or used with other medical devices, ME equipment or ME systems.
EXAMPLE 1 An oxygen concentrator with integrated oxygen conserving equipment function or humidifier function.
EXAMPLE 2 An oxygen concentrator used with a flowmeter stand.
EXAMPLE 3 An oxygen concentrator as part of an anaesthetic system for use in areas with limited logistical supplies of electricity and anaesthetic gases[2].
EXAMPLE 4 An oxygen concentrator with an integrated liquid reservoir function or gas cylinder filling system function.
This document is also applicable to those accessories intended by their manufacturer to be connected to an oxygen concentrator, where the characteristics of those accessories can affect the basic safety or essential performance of the oxygen concentrator.
NOTE 2 Such accessories can include, but are not limited to, masks, cannulae, extension tubing, humidifiers, carts, carrying cases, external power sources and oxygen conserving equipment.
This document does not specify requirements for oxygen concentrators for use with a medical gas pipeline system.
If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant.
Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 7.2.13 and 8.4.1 of the general standard.
NOTE 3 See also 4.2 of the general standard.

  • Standard
    67 pages
    English language
    sale 15% off