Cable assemblies, cables, connectors and passive microwave components - Screening attenuation measurement by the reverberation chamber method

Describes the measurement of screening attenuation by the reverberation, or stirred, chamber, suitable for virtually any type of microwave component and having no theoretical upper frequency limit. It is only limited toward low frequencies due to the size of the test equipment, which is frequency dependent and is only one of several methods of measuring sreening attenuation. For the purpose of this standard, examples of microwave components are waveguides, phase shifters, diplexers/multiplexers, power dividers/combiners, etc.

Konfektionierte Kabel, Kabel, Steckverbinder und passive Mikrowellenbauteile - Messung der Schirmdämpfung mit dem Strahlungskammerverfahren

Câbles, cordons, connecteurs et composants hyperfréquence passifs - Mesure de l'atténuation d'écran par la méthode de la chambre réverbérante

Describes the measurement of screening attenuation by the reverberation, or stirred, chamber, suitable for virtually any type of microwave component and having no theoretical upper frequency limit. It is only limited toward low frequencies due to the size of the test equipment, which is frequency dependent and is only one of several methods of measuring sreening attenuation. For the purpose of this standard, examples of microwave components are waveguides, phase shifters, diplexers/multiplexers, power dividers/combiners, etc.

Cable assemblies, cables, connectors and passive microwave components - Screening attenuation measurement by the reverberation chamber method (IEC 61726:1999)

General Information

Status
Withdrawn
Publication Date
23-Mar-2000
Withdrawal Date
31-Dec-2002
Drafting Committee
Parallel Committee
Current Stage
9960 - Withdrawal effective - Withdrawal
Start Date
13-Oct-2018
Completion Date
13-Oct-2018

Relations

Buy Standard

Standard
EN 61726:2004
English language
24 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI SIST EN 61726:2004
STANDARD
april 2004
Cable assemblies, cables, connectors and passive microwave components -
Screening attenuation measurement by the reverberation chamber method (IEC
61726:1999)
ICS 33.120.10; 33.120.30 Referenčna številka
©  Standard je založil in izdal Slovenski inštitut za standardizacijo. Razmnoževanje ali kopiranje celote ali delov tega dokumenta ni dovoljeno

NORME CEI
INTERNATIONALE IEC
INTERNATIONAL
Deuxième édition
STANDARD
Second edition
1999-11
Câbles, cordons, connecteurs
et composants hyperfréquence passifs –
Mesure de l'atténuation d'écran
par la méthode de la chambre réverbérante
Cable assemblies, cables, connectors and
passive microwave components –
Screening attenuation measurement by the
reverberation chamber method
© IEC 2001 Droits de reproduction réservés ⎯ Copyright - all rights reserved
Aucune partie de cette publication ne peut être reproduite ni No part of this publication may be reproduced or utilized in
utilisée sous quelque forme que ce soit et par aucun procédé, any form or by any means, electronic or mechanical,
électronique ou mécanique, y compris la photocopie et les including photocopying and microfilm, without permission in
microfilms, sans l'accord écrit de l'éditeur. writing from the publisher.
International Electrotechnical Commission 3, rue de Varembé Geneva, Switzerland
Telefax: +41 22 919 0300 e-mail: inmail@iec.ch IEC web site http://www.iec.ch
CODE PRIX
Commission Electrotechnique Internationale
S
PRICE CODE
International Electrotechnical Commission
Pour prix, voir catalogue en vigueur
For price, see current catalogue

61726 © IEC:2001 – 3 –
CONTENTS
Page
FOREWORD . 5
INTRODUCTION .9
Clause
1 Scope . 11
2 Basic description of the reverberation chamber method . 11
3 Measurement of the screening attenuation of the device under test (DUT) . 13
4 Description of the test set-up. 13
4.1 Reverberation chamber . 13
4.2 Mode stirrer . 13
4.3 Antennas . 13
4.4 Test equipment . 15
4.5 Device under test (DUT). 17
4.6 Linking devices . 17
5 Measurement procedure. 19
5.1 General. 19
5.2 Measurement of the DUT . 19
5.3 Measurement of the insertion loss of the cavity. 19
5.4 Control of the test set-up. 21
5.5 Revolution speed of the mode stirrer . 23
5.6 Test frequencies . 23
5.7 Voltage standing wave ratio (VSWR) . 23
6 Evaluation of the test results . 23
6.1 Screening attenuation . 23
6.2 Relationship between transfer impedance parameters and screening attenuation. 25
6.3 Electromagnetic field surrounding the DUT . 27
Annex A (informative) Design of the reverberation chamber . 29
Annex B (informative) Design of the mode stirrer . 31
Annex C (informative) Example of a calibrator. 35
Annex D (informative) Relationship between transfer impedance and screening attenuation. 39
Bibliography . 43
Figure 1 – Example of a test set-up . 15
Figure B.1 – Construction diagram for mode-stirrer . 33
Figure B.2 – Details of collet for mounting tuner shaft to drive motor through the
wall of the test chamber.33
Figure C.1 – Basic construction details . 35

61726 © IEC:2001 – 5 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
CABLE ASSEMBLIES, CABLES, CONNECTORS AND
PASSIVE MICROWAVE COMPONENTS –
SCREENING ATTENUATION MEASUREMENT BY THE
REVERBERATION CHAMBER METHOD
FOREWORD
1) The IEC (International Electrotechnical Commission) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of the IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, the IEC publishes International Standards. Their preparation is
entrusted to technical committees; any IEC National Committee interested in the subject dealt with may
participate in this preparatory work. International, governmental and non-governmental organizations liaising
with the IEC also participate in this preparation. The IEC collaborates closely with the International Organization
for Standardization (ISO) in accordance with conditions determined by agreement between the two
organizations.
2) The formal decisions or agreements of the IEC on technical matters express, as nearly as possible, an
international consensus of opinion on the relevant subjects since each technical committee has representation
from all interested National Committees.
3) The documents produced have the form of recommendations for international use and are published in the form
of standards, technical specifications, technical reports or guides and they are accepted by the National
Committees in that sense.
4) In order to promote international unification, IEC National Committees undertake to apply IEC International
Standards transparently to the maximum extent possible in their national and regional standards. Any
divergence between the IEC Standard and the corresponding national or regional standard shall be clearly
indicated in the latter.
5) The IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with one of its standards.
6) Attention is drawn to the possibility that some of the elements of this International Standard may be the subject
of patent rights. The IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 61726 has been prepared by subcommittee 46A: Coaxial cables, of
IEC technical committee 46: Cables, wires, waveguides, r.f. connectors, and accessories for
communication and signalling.
This second edition cancels and replaces the first edition, which was issued as a type 3
technical report in 1995. It constitutes a technical revision and now has the status of an
International Standard.
This bilingual version (2001) replaces the English version.
The text of this standard is based on the following documents:
FDIS Report on voting
46A/356/FDIS 46A/359/RVD
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 3.

61726 © IEC:2001 – 7 –
Annexes A, B, C and D are for information only.
The committee has decided that the contents of this publication will remain unchanged
until 2005. At this date, the publication will be
 reconfirmed;
 withdrawn;
 replaced by a revised edition, or
 amended.
61726 © IEC:2001 – 9 –
INTRODUCTION
The requirements of modern electronic equipment have indicated a demand for a method of
testing screening attenuation of microwave components over their whole frequency range.
Convenient test methods exist for low frequencies and components of regular shape and these
test methods are described in the relevant IEC product specifications.
For higher frequencies and for components of irregular shape a new test method has become
necessary and such a test method is described in this International Standard.

61726 © IEC:2001 – 11 –
CABLE ASSEMBLIES, CABLES, CONNECTORS AND
PASSIVE MICROWAVE COMPONENTS –
SCREENING ATTENUATION MEASUREMENT BY THE
REVERBERATION CHAMBER METHOD
1 Scope
This International Standard describes the measurement of screening attenuation by the
reverberation chamber test method, sometimes named mode stirred chamber, suitable for
virtually any type of microwave component and having no theoretical upper frequency limit. It is
only limited toward low frequencies due to the size of the test equipment, which is frequency
dependent and is only one of several methods of measuring screening attenuation.
For the purpose of this standard, examples of microwave components are waveguides, phase
shifters, diplexers/multiplexers, power dividers/combiners etc.
2 Basic description of the reverberation chamber method
The reverberation chamber method for measurement of the screening attenuation of
microwave components consists of exposing the device under test (DUT) to an almost
homogeneous and isotropic electromagnetic field and then measuring the signal level induced
into the device.
These conditions are achieved by the use of a shielded enclosure, which acts as an oversized
cavity (in terms of wavelength), with a high quality factor. Its boundary conditions are
continuously agitated by a rotating reflective surface (mode stirrer), mounted within the
chamber, which enables the field to approach homogeneous and isotropic conditions during
one revolution.
Electromagnetic power is fed to the chamber by means of an input or transmitting antenna.
The strength of the field inside the chamber is measured through a reference antenna. The
ratio of the injected power (input antenna) to the received power (reference antenna) is the
insertion loss of the cavity. The insertion loss is strongly frequency dependent and is also
dependent on the quality factor of the cavity.
It has been shown that, due to the isotropic field, any antenna placed inside the cavity behaves
1)
as if its gain was unity [1] , therefore no directional effect is to be expected. If the device
under test is electrically short, its screening attenuation will be directly related to usual transfer
parameters (Z and Z). If the device under test is not electrically short the screening
t f
attenuation may still be related to Z and Z in some simple cases (evenly distributed leakage,
t f
periodically distributed leakage) using summing functions derived from antenna network theory.
___________
1)
Figures in square brackets refer to the Bibliography.

61726 © IEC:2001 – 13 –
3 Measurement of the screening attenuation of the device under test (DUT)
The measurement of screening attenuation is based on the comparison of the electromagnetic
field power outside the DUT to the electromagnetic field power induced into the DUT. The
screening attenuation is then defined as:
⎛ P ⎞
DUT
⎜ ⎟
a = −10 log (1)
s 10
⎜ ⎟
P
⎝ REF ⎠
or
⎛ ⎞
P
DUT
a = −10 log ⎜ ⎟ − ∆ (2)
s 10 ins
⎜ ⎟
P
⎝ INJ ⎠
where
P is the power coupled to the device under test (W);
DUT
P is the power coupled to the reference antenna (W);
REF
P is the power injected into the chamber (W);
INJ
∆ is the insertion loss of the chamber in decibels (dB).
ins
4 Description of the test set-up
4.1 Reverberation chamber
The reverberation chamber is a shielded enclosure having any shape, provided that its smallest
dimension exceeds three wavelengths at the lowest test frequency. A perfect cubic shape
should be avoided for optimum performance at lower frequencies. It shall be made of
conductive materials (copper, aluminium or steel) and shall not contain lossy materials.
The upper frequency limit depends only on the screening attenuation, which shall exceed 60 dB
for the whole frequency range. However, this value is not critical if the spectrum analyzer and
the connecting devices of the test set-up are sufficiently screened and if the quality factor of
the cavity remains sufficiently high. The quality factor shall be checked to verify that during one
revolution of the mode stirrer the ratio between the maximum and the minimum power at the
output of the reference antenna exceeds 20 dB. As a minimum, the test chamber and the test
instrumentation shall have a combined screening attenuation at each test frequency that is
10 dB greater than the screening requirement of the DUT.
The shielded enclosure is drilled with four coaxial feed-throughs: two for the output of the
antennas and two for the output of the measuring loop. For further details see annex A.
4.2 Mode stirrer
The mode stirrer shall be large with respect to wavelength and be bent at angles to the walls of
the chamber. The mode stirrer shall be at least two wavelengths from tip to tip at the lowest
test frequency. An example of a mode stirrer is described in annex B.
4.3 Antennas
The reverberation chamber is equipped with input an
...