CLC/TC 106X - Electromagnetic fields in the human environment
TC 106X deals with various aspects of the exposure of people to electromagnetic fields from 0 Hz to 300 GHz.
Electromagnetic fields in the human environment
TC 106X deals with various aspects of the exposure of people to electromagnetic fields from 0 Hz to 300 GHz.
General Information
- Amendment9 pagesEnglish languagesale 10% offe-Library read for1 day
This product standard applies to wireless communication devices used in close proximity to the human ear (e.g. mobile phones, wireless headsets). The applicable frequency range is from 300 MHz to 6 GHz. The objective of this standard is to demonstrate the compliance of such devices with the basic restrictions and exposure limit values related to human exposure to radio frequency electromagnetic fields. For devices used next to the body or in front of the face the applicable product standard is EN 50566:2017. For low power devices the applicable product standard is EN 50663:2017.
- Amendment8 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62311:2019 applies to electronic and electrical equipment for which no dedicated product standard or product family standard regarding human exposure to electromagnetic fields applies. It covers equipment with intentional or non-intentional radiators as well as a combination thereof. This document provides assessment methods and criteria to evaluate equipment against limits on exposure of people related to electric, magnetic and electromagnetic fields. The frequency range covered is from 0 Hz to 300 GHz. This second edition cancels and replaces the first edition published in 2007. This edition constitutes a technical revision. This document does not specify limits expressed by means of basic restrictions and/or reference levels. Such limits are subject to the applied assessment scheme, for example by means of regional limits. This edition includes the following significant technical changes with respect to the previous edition: a) a clear distinction between intentional and unintentional radiators has been introduced; b) the exposure to non-uniform fields is considered; c) the treatment of uncertainty for the assessment procedures has been improved; d) various summation regimes are described in Annex A; e) the information from meanwhile published basic standards has been used and hence all informative annexes of the previous edition have been removed. Key words: Human Exposure, Electromagnetic Fields (0 Hz to 300 GHz).
- Standard37 pagesEnglish languagesale 10% offe-Library read for1 day
Identical adoption of future IEC/IEEE 63195-2 into EN IEC/IEEE 63195-2
- Draft78 pagesEnglish languagesale 10% offe-Library read for1 day
Identical adoption of future IEC/IEEE 63195-1 into EN IEC/IEEE 63195-1.
- Draft146 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62232:2022 addresses the evaluation of RF field strength, power density and specific absorption rate (SAR) levels in the vicinity of base stations (BS), also called products or equipment under test (EUT), intentionally radiating in the radio frequency (RF) range 110 MHz to 300 GHz in accordance with the scope, see Clause 1. It does not address the evaluation of current density. RF exposure evaluation methods to be used for product compliance, product installation compliance and in-situ RF exposure assessments are specified in this document. Exposure limits are not specified in this document. The entity conducting RF exposure assessments refers to the set of exposure limits applicable where exposure takes place. Examples of applicable exposure limits considered in this document are provided in the Bibliography, for example ICNIRP-2020 [1], ICNIRP-1998 [2], IEEE Std C95.1™-2019 [3] and Safety Code 6 [4].
- Standard346 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62209-3: 2019 specifies measurement protocols and test procedures for the reproducible measurement of peak spatial-average specific absorption rate (psSAR) induced inside a simplified model of a human head or body by radio-frequency (RF) transmitting devices, with a specified measurement uncertainty. Requirements are provided for psSAR assessment using vector measurement-based systems. Such systems determine the psSAR by three-dimensional (3D) field reconstruction within the volume of interest in accordance with the requirements herein for the measurement system, calibration, uncertainty assessment and validation methods. The protocols and procedures apply for the psSAR assessments covering a significant majority of people including children during use of wireless communication devices operated in close proximity to the head or body. This document is applicable to wireless communication devices intended to be used at a position near the human head or body at distances up to and including 200 mm. This document may be employed to evaluate SAR compliance of different types of wireless communication devices used next to the ear, in front of the face, mounted on the body, combined with other RF-transmitting or non-transmitting devices or accessories (e.g. belt-clip), or embedded in garments. The overall applicable frequency range is from 600 MHz to 6 GHz. The system validation procedures provided within this document cover frequencies from 600 MHz to 6 GHz. With a vector measurement-based system this document can be employed to evaluate SAR compliance of different types of wireless communication devices. The wireless communication device categories covered include but are not limited to mobile telephones, cordless microphones, auxiliary broadcast devices and radio transmitters in personal computers, desktop and laptop devices, multi-band, multi-antenna, and push-to-talk devices. Key Words: Human Exposure, Hand-Held and Body Mounted Wireless Communication Devices.
- Standard140 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62764-1:2022 specifies a methodology for determining the exposure to multiple magnetic field sources for passenger cars and light commercial vehicles including standardized operating conditions and measurement volumes and/or surfaces. This part of IEC 62764 applies to the assessment of human exposure to low-frequency magnetic fields generated by automotive vehicles. For plug-in vehicles, this includes the electric vehicle supply equipment (EVSE) and associated cables provided by the car manufacturer. This excludes the charging station. This document specifies the measurement procedure for the evaluation of magnetic field exposures generated by electronic and electrical equipment (excluding intentionally transmitting radio frequency antennas) in selected automotive environments, for passenger cars and commercial vehicles of categories M1 and N1 as defined in ECE/TRANS/WP.29/78/Rev.3 [1], with respect to human exposure. It provides standardized operating conditions and defines recommended measurements to assess compliance with the applicable exposure requirements. This document covers the frequency range 1 Hz to 100 kHz and is applicable to any type of engine and/or internal energy source.
- Standard39 pagesEnglish languagesale 10% offe-Library read for1 day
This document provides general methods for measurement and calculation of quantities associated with human exposure to electromagnetic fields in the frequency range from 0 Hz to 300 GHz. It is intended specifically to be used for the assessment of emissions from products and comparison of these with the exposure limits for the general public given in Council Recommendation 1999/519/EC, and those given for workers in Directive 2013/35/EU, as appropriate. It also is intended to be used for assessment of human exposure to electromagnetic fields in the workplace to determine compliance with the requirements of Directive 2013/35/EU. This standard deals with quantities that can be measured or calculated external to the body, notably electric and magnetic field strength or power density, and includes the measurement and calculation of quantities inside the body that form the basis for protection guidelines. In particular the standard provides information on: — definitions and terminology, — characteristics of electromagnetic fields, — measurement of exposure quantities, — instrumentation requirements, — methods of calibration, — measurement techniques and procedures for evaluating exposure, — calculation methods for exposure assessment. Where an applicable electromagnetic field standard specific to a product or technology exists it is expected to be used rather than this document. EN 62311:—, Table 1 gives a list of relevant standards.
- Standard27 pagesEnglish languagesale 10% offe-Library read for1 day
The scope of this document is to provide a general procedure for the assessment of workers’ exposure to electric, magnetic and electromagnetic fields in a workplace in order to determine compliance with exposure limit values and/or action levels as stated in European Directive 2013/35/EU. The purpose of this document is to — specify how to perform an initial assessment of the levels of workers' exposure to electromagnetic fields (EMF), if necessary, including specific exposure assessment of such levels by measurements and/or calculations, — determine whether it is necessary to carry out a detailed risk assessment of EMF exposure. This document can be used by employers for the risk assessment and, where required, measurement and/or calculation of the exposure of workers. Based on specific workplace and other standards, it can be determined whether preventive measures/actions have to be taken to comply with the provisions of the Directive. The frequencies covered are from 0 Hz to 300 GHz. NOTE 1 This document relates to the exposure limits as specified in the Directive 2013/35/EU. It is intended to protect workers from risks to their health and safety arising or likely to arise from exposure to electromagnetic fields (0 Hz to 300 GHz) during their work. However, this and other Directives can include additional measures for the protection of specific groups of workers and/or specific workplaces for which the employer is required to investigate other protective measures as a part of the overall risk assessment. See Annex A. NOTE 2 Directive 2013/35/EU has been transposed into national legislation in all the EU member countries. It is intended that users of this standard consult the national legislation related to this transposition in order to identify the national regulations and requirements. These national regulations and requirements can have additional requirements that are not covered by this standard.
- Standard38 pagesEnglish languagesale 10% offe-Library read for1 day
This product standard applies to wireless communication devices used at distances up to and including 200 mm from the human body, i.e. when held in the hand or in front of the face, mounted on the body, combined with other transmitting or non-transmitting devices or accessories (e.g. belt-clip, camera or Bluetooth add-on), or integrated into garments. The applicable frequency range is from 30 MHz to 6 GHz. The objective of this standard is to demonstrate the compliance of such devices with the basic restrictions and exposure limit values related to human exposure to radio frequency electromagnetic fields. For devices used next to the ear the applicable product standard is EN 50360:2017 [1]. For low power devices the applicable product standard is EN 50663:2017 [2].
- Amendment7 pagesEnglish languagesale 10% offe-Library read for1 day
Identical adoption of future IEC/IEEE 62209-1528 into EN IEC/IEEE 62209-1528.
- Standard283 pagesEnglish languagesale 10% offe-Library read for1 day
This document provides the procedure for the specific assessment required in EN 50527 1:2016, Annex A, for workers with implanted neurostimulator systems (NS), specifically of the type used for spinal cord stimulation (SCS). It is recognized that implantable neurostimulators have been developed for a wide variety of clinical applications, however the SCS devices within the scope of this document represent the largest segment of the implantable neurostimulator applications thus far. NOTE 1 If the worker has other Active Implantable Medical Devices (AIMDs) implanted additionally, they are assessed separately according to EN 50527 1 or other particular standards within the EN 50527 series. The purpose of the specific assessment is to determine the risk for workers with implanted SCS devices arising from exposure to electromagnetic fields (EMF) at the workplace. The assessment includes the likelihood of clinically significant effects. NOTE 2 This document does not address risks from contact currents, or the effects upon any associated non-implantable devices (e.g. Patient Programmers). The techniques described in the different approaches can also be used for the assessment of publicly accessible areas. The frequency range to be observed is from 0 Hz to 3 GHz. Above 3 GHz no interference with the devices within the scope of this document is expected to occur. NOTE 3 The rationale for limiting the observation range to 3 GHz can be found in ISO 14708 3 [1]. NOTE 4 Further information concerning the functions of neurostimulator systems can be found at https://www.aans.org/Patients/Neurosurgical-Conditions-and-Treatments/Spinal-Cord-Stimulation.
- Standard106 pagesEnglish languagesale 10% offe-Library read for1 day
This Technical Report illustrates good practice and provides guidance with regard to the term "reasonably foreseeable use" as it relate to product compliance assessment standards concerning the exposure of humans to electric, magnetic and electromagnetic fields (EMF) as required in the Radio Equipment Directive (RED) and Low Voltage Directive (LVD). Other safety aspects and requirements from other directives are not covered by this Technical Report. This report uses the term "reasonably foreseeable use". In this context of this TR the terms "reasonably foreseeable conditions" and "conditions of use which can be reasonably foreseen" are interchangeable. The report provides guidance that is not specific to individual equipment. The report covers both occupational and general public use of equipment and also provides a rationale for the distinction between occupational use and use by the general public.
- Technical report24 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies the method for assessing overall exposure from all fixed radio frequency sources at a broadcast site. This assessment can be applied at any time but is carried out when the exposure situation changes in or around the aforementioned site. This document can play an essential role in the coordination of different stakeholders, with respect to ensuring EMF exposure compliance in the vicinity of a broadcast site especially for equipment installed within the site.
- Standard18 pagesEnglish languagesale 10% offe-Library read for1 day
This European Standard provides the procedure for the specific assessment required in Annex A of EN 50527 1:2015 for workers with implanted cardioverter defibrillators. It offers different approaches for doing the risk assessment. The most suitable one shall be used. If the worker has other Active Implantable Medical Devices (AIMDs) implanted additionally, they have to be assessed separately.
- Standard62 pagesEnglish languagesale 10% offe-Library read for1 day
This product standard applies to devices operating within the frequency range 0 Hz to 300 GHz, used in electronic article surveillance (EAS), radio frequency identification (RFID) and similar applications, in relation to exposure to electromagnetic fields The object of this generic standard is to provide a route for evaluation of such equipment against limits on human exposure to electric, magnetic and electromagnetic fields, and induced and contact current. NOTE Other standards can apply to products covered by this document. In particular this document is not designed to evaluate the electromagnetic compatibility with other equipment; nor does it reflect any product safety requirements other than those specifically related to human exposure to electromagnetic fields.
- Standard10 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62232:2017(E) provides methods for the determination of radio-frequency (RF) field strength and specific absorption rate (SAR) in the vicinity of radiocommunication base stations (RBS) for the purpose of evaluating human exposure. This document: - considers intentionally radiating RBS which transmit on one or more antennas using one or more frequencies in the range 110 MHz to 100 GHz; - considers the impact of ambient sources on RF exposure at least in the 100 kHz to 300 GHz frequency range; - specifies the methods to be used for RF exposure evaluation for compliance assessment applications, namely: - product compliance - determination of compliance boundary information for an RBS product before it is placed on the market; - product installation compliance - determination of the total RF exposure levels in accessible areas from an RBS product and other relevant sources before the product is put into service; - in-situ RF exposure assessment – measurement of in-situ RF exposure levels in the vicinity of an RBS installation after the product has been taken into operation; - describes several RF field strength and SAR measurement and computation methodologies with guidance on their applicability to address both the in-situ evaluation of installed RBS and laboratory-based evaluations; - describes how surveyors, with a sufficient level of expertise, establish their specific evaluation procedures appropriate for their evaluation purpose; - provides guidance on how to report, interpret and compare results from different evaluation methodologies and, where the evaluation purpose requires it, determine a justified decision against a limit value and - provides short descriptions of the informative example case studies given in the companion Technical Report IEC TR 62669 [1] This second edition cancels and replaces the first edition published in 2011 and constitutes a technical revision.
- Standard243 pagesEnglish languagesale 10% offe-Library read for1 day
The object of this generic standard is to provide a route for evaluation of equipment used by workers against limits on human exposure to electric, magnetic and electromagnetic fields, and induced and contact current when it is put into service in its operational environment, and also for in situ or post-market evaluation of such equipment. The frequency range covered is 0 Hz to 300 GHz. Other standards can apply to products covered by this document. In particular this document is not designed to evaluate the electromagnetic compatibility with other equipment; nor does it reflect any product safety requirements other than those specifically related to human exposure to electromagnetic fields. This standard applies to electronic and electrical equipment for which no dedicated put into service or in situ product or product family standard regarding worker exposure to electromagnetic fields exists. If such a standard does exist then it shall be used and this standard shall not.
- Standard10 pagesEnglish languagesale 10% offe-Library read for1 day
The object of this generic standard is to provide a route for evaluation of such equipment against limits on human exposure to electric, magnetic and electromagnetic fields, and induced and contact current. This standard applies to electronic and electrical equipment for which no dedicated product- or product family standard, or standard relating to low power equipment, regarding human exposure to electromagnetic fields exists. If such a standard does exist then it shall be used and this standard shall not. The frequency range covered is 0 Hz to 300 GHz. This standard is intended to cover both intentional and non-intentional radiators. It should be noted that the supplier of a specific piece of equipment might not know the overall exposure environment in which the equipment is being used. This product standard can only assess the human exposure from the specific equipment under evaluation. Other standards can apply to products covered by this document. In particular this document is not designed to evaluate the electromagnetic compatibility with other equipment; nor does it reflect any product safety requirements other than those specifically related to human exposure to electromagnetic fields.
- Standard10 pagesEnglish languagesale 10% offe-Library read for1 day
This European standard provides simple conformity assessment methods for low-power electronic and electrical equipment operating at frequencies between 10 MHz and 300 GHz to an electromagnetic field (EMF) exposure limit. If such equipment cannot be shown to comply with the applicable EMF exposure requirements using the exposure assessment methods in this standard, then other EMF product standards may be used for conformity assessment. Other standards can apply to products covered by this document. In particular this document is not designed to evaluate the electromagnetic compatibility with other equipment; nor does it reflect any product safety requirements other than those specifically related to human exposure to electromagnetic fields.
- Standard9 pagesEnglish languagesale 10% offe-Library read for1 day
This product standard is related to human exposure to radiofrequency electromagnetic fields transmitted by base station equipment in the frequency range 110 MHz to 100 GHz. The object is to assess the compliance of such equipment with the general public basic restrictions (directly or indirectly via compliance with reference levels) and the workers’ exposure limits values (directly or indirectly via compliance with action levels), when it is put into service in its operational environment.
- Standard9 pagesEnglish languagesale 10% offe-Library read for1 day
This product standard is related to human exposure to radiofrequency electromagnetic fields transmitted by base station equipment in the frequency range 110 MHz to 100 GHz. The object is to assess the compliance of such equipment with the general public basic restrictions (directly or indirectly via compliance with reference levels) and the workers’ exposure limit values (directly or indirectly via compliance with action levels), when it is placed on the market. For low power devices the applicable product standard is EN 50663:2017.
- Standard9 pagesEnglish languagesale 10% offe-Library read for1 day
This product standard applies to wireless communication devices used in close proximity to the human ear (e.g. mobile phones, wireless headsets). The applicable frequency range is from 300 MHz to 6 GHz. The objective of this standard is to demonstrate the compliance of such devices with the basic restrictions and exposure limit values related to human exposure to radio frequency electromagnetic fields. For devices used next to the body or in front of the face the applicable product standard is EN 50566:2017. For low power devices the applicable product standard is EN 50663:2017.
- Standard8 pagesEnglish languagesale 10% offe-Library read for1 day
This product standard applies to wireless communication devices used at distances up to and including 200 mm from the human body, i.e. when held in the hand or in front of the face, mounted on the body, combined with other transmitting or non-transmitting devices or accessories (e.g. belt-clip, camera or Bluetooth add-on), or integrated into garments. The applicable frequency range is from 30 MHz to 6 GHz. The objective of this standard is to demonstrate the compliance of such devices with the basic restrictions and exposure limit values related to human exposure to radio frequency electromagnetic fields. For devices used next to the ear the applicable product standard is EN 50360:2017 [1]. For low power devices the applicable product standard is EN 50663:2017 [2].
- Standard8 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment7 pagesEnglish languagesale 10% offe-Library read for1 day
This European Standard provides the procedure for the specific assessment required in EN 50527-1:2016, Annex A, for workers with implanted pacemakers. It offers different approaches for doing the risk assessment. The most suitable one will be used. If the worker has other Active Implantable Medical Devices (AIMDs) implanted additionally, they need to be assessed separately. The purpose of the specific assessment is to determine the risk for workers with implanted pacemakers arising from exposure to electromagnetic fields at the workplace. The assessment includes the likelihood of clinically significant effects and takes account of both transient and long-term exposure within specific areas of the workplace. NOTE 1 This standard does not address risks from contact currents. The techniques described in the different approaches may also be used for the assessment of publicly accessible areas. The frequency range to be observed is from 0 Hz to 3 GHz. Above 3 GHz no interference with the pacemaker occurs when the exposure limits are not exceeded. NOTE 2 The rationale for limiting the observation range to 3 GHz can be found in ISO 14117:2012, Clause 5.
- Standard70 pagesEnglish languagesale 10% offe-Library read for1 day
This European standard provides methods for assessing compliance with the requirements of the Directive 2013/35/EU] at a site operating one or more broadcast transmitters. This standard covers the frequency range up to 40 GHz. NOTE The Council and European Parliament Directive 2013/35/EU will be transposed into national legislation in all the EU member countries. Users of this standard shall consult the national legislation related to this transposition in order to identify the national regulations and requirements. These national regulations and requirements can have additional requirements that are not covered by this standard.
- Standard25 pagesEnglish languagesale 10% offe-Library read for1 day
The purpose of this Technical Report is to give advice on, and explanation of, the preparation of suitable EMF standards. It also aims to ensure that relevant deliverables from all CLC TCs will accurately reflect the current policy and legislative background on EMF exposure.
- Technical report12 pagesEnglish languagesale 10% offe-Library read for1 day
This European Standard provides a general procedure to assess workers’ exposure to electric and magnetic fields (EMF) in work places associated with the production, transmission and distribution of electric energy, and to demonstrate compliance with exposure limit values and action levels as stated in the Council and European Parliament "EMF" Directive 2013/35/EU [12]. NOTE 1 The Council and European Parliament Directive 2013/35/EU will be transposed into national legislation in all the EU member countries. It is important that users of this standard consult the national legislation related to this transposition in order to identify the national regulations and requirements. These national regulations and requirements may have additional requirements that are not covered by this standard It has the role of a specific workplace standard. It takes into account the non-binding application guide for implementing the EMF Directive [11] and it defines the assessment procedures and compliance criteria applicable to the electric industry. The frequency range of this standard covers from DC to 20 kHz, which is sufficient to include the power frequency used for electric power supply systems throughout Europe (50 Hz) and the various harmonics and inter-harmonics occurring in the supply system. In this extremely low frequency range, electric and magnetic fields are independent and, therefore, they both have to be addressed in the exposure assessment. NOTE 2 Electrical companies also use radio frequency transmissions to operate and maintain their networks and power plants. Similarly, other exposures to EMF may occur during maintenance operations, for instance, due to the use of hand-held electrical tools. All these EMF sources are outside the scope of this standard. NOTE 3 Regarding EMF in the low frequency range, the scientific basis of the EMF directive is the ICNIRP health guidelines published in 2010 [14]. Reference is made to this scientific basis when necessary for justifying or clarifying some of the technical statements of the present document.
- Standard54 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 61786-1:2013 provides guidance for measuring instruments used to measure the field strength of quasi-static magnetic and electric fields that have a frequency content in the range 1 Hz to 100 kHz and with DC magnetic fields to evaluate the exposure levels of the human body to these fields. Sources of fields include devices that operate at power frequencies and produce power frequency and power frequency harmonic fields, as well as devices that produce fields within the frequency range of this document, including devices that produce static fields, and the earth's static magnetic field. The magnitude ranges covered by this standard are 0,1 μT to 200 mT in AC (1 μT to 10 T in DC) and 1 V/m to 50 kV/m for magnetic fields and electric fields, respectively. When measurements outside this range are performed, most of the provisions of this standard will still apply, but special attention should be paid to specified uncertainty and calibration procedures. The first editions of IEC 61786-1 and IEC 61786-2 replace IEC 61786:1998. Part 1 deals with measuring instruments, and Part 2 deals with measurement procedures. The content of the standard was revised in order to give up-to-date and practical information to the user.
- Standard55 pagesEnglish languagesale 10% offe-Library read for1 day
This European Standard provides a procedure to assess the risk to workers bearing one or more active implantable medical devices from exposure to electric, magnetic and electromagnetic fields at a workplace. It describes how a general risk assessment should be performed and determines whether it is necessary to carry out a detailed risk assessment. NOTE 1 This European Standard does not cover indirect effects caused by non active implants. NOTE 2 The risk of human exposure to EMF considered is only due to malfunctioning of AIMD. Possibilities of AIMD contribution to the risk, e.g. local modification of the distribution of EMF produced by external source or production of own EMF, are covered by the respective product standards for the AIMD. Based on specific workplace standards it can be determined whether preventive measures/actions need to be taken to comply with the provisions of Directive 2013/35/EU. The work situation covered is considered to be under normal working conditions including normal operation, maintenance, cleaning and other situations being part of the normal work. The frequencies covered are from 0 Hz to 300 GHz. The European Parliament and Council Directive 2013/35/EU will be transposed into national legislation in all the EU member countries. It is recommended that users of this standard consult the national legislation related to this transposition in order to identify the national regulations and requirements. These national regulations and requirements may have additional requirements that are not covered by this standard and take precedence. NOTE 3 Performance requirements with respect to active implantable medical devices are excluded from the Scope of this standard. These are defined in the relevant particular standards for active implantable medical devices. The risk assessment described in this standard is only required if an AIMD-Employee is present. Active Implantable Medical Devices (AIMDs) are regulated by Directive 90/385/EEC and the amendments to it. NOTE 4 Product standards EN 45502-1 and of the EN 45502-2-X series describe the product requirements for different kinds of AIMDs. Different kinds of AIMDs are e.g. pacemaker (EN 45502-2-1), implantable cardioverter defibrillators (EN 45502-2-2), cochlear implants (EN 45502-2-3), implantable neurostimulators (ISO 14708-3), implantable infusion pumps (ISO 14708-4). In situations where the risk assessment following this standard does not lead to a conclusion, complementary provisions for the assessment of workers exposure for different kinds of AIMDs are given in particular standards for these specific AIMDs (see Figure 1). (...) Figure 1 - Structure of the EN 50527 family of standards
- Standard34 pagesEnglish languagesale 10% offe-Library read for1 day
- Corrigendum3 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62479:2010 provides simple conformity assessment methods for low-power electronic and electrical equipment to an exposure limit relevant to electromagnetic fields (EMF). If such equipment cannot be shown to comply with the applicable EMF exposure requirements using the methods included in this standard for EMF assessment, then other standards, including IEC 62311 or other (EMF) product standards, may be used for conformity assessment.
- Standard23 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62110:2009 establishes measurement procedures for electric and magnetic field levels generated by AC power systems to evaluate the exposure levels of the human body to these fields. This standard is not applicable to DC power transmission systems. IEC 62110:2009 is applicable to public exposure in the domestic environment and in areas accessible to the public. It specifies fundamental procedures for the measurement of fields, and, with regard to human exposure, for obtaining a field value that corresponds to a spatial average over the entire human body. IEC 62110:2009 is not applicable to occupational exposure associated with, for example, the operation and/or maintenance of the power systems. Such exposure may occur when working inside a distribution or transmission substation, a power plant, in a manhole or a tunnel for underground cables, or on an overhead line tower or pole.
- Standard53 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62369-1:2008 presents procedures for the evaluation of human exposure to electromagnetic fields (EMFs) from devices used in electronic article surveillance (EAS), radio frequency identification (RFID) and similar applications. It adopts a staged approach to facilitate compliance assessment. The first stage (Stage 1) is a simple measurement against the appropriate derived reference values. Stage 2 is a more complex series of measurements or calculations, coupled with analysis techniques. Stage 3 requires detailed modelling and analysis for comparison with the basic restrictions. When assessing any device, the most appropriate methodfor the exposure situation may be used. At the time of writing this International Standard, electronic article surveillance, radio frequency identification and similar systems do not normally operate at frequencies below 1 Hz or above 10 GHz. EMF exposure guidelines and standards can cover a wider range of frequencies, so clarification on the required range is included as part of the evaluation procedures. The devices covered by this document normally have non-uniform field patterns. Often these devices have a very rapid reduction of field strength with distance and operate under near-field conditions where the relationship between electric and magnetic fields is not constant. This, together with typical exposure conditions for different device types, is detailed in Annex A. Annex B contains comprehensive information to assist with numerical modelling of the exposure situation. It includes both homogeneous and anatomical models as well as the electrical properties of tissue. IEC 62369-1:2008 does not include limits. Limits can be obtained from separately published human exposure guidelines. Different guidelines and limit values may apply in different regions. Linked into the guidelines are usually methods for summation across wider frequency ranges and for multiple exposure sources. These shall be used. A simplified method for summation of multiple sources is contained in Annex C. This has to be used with care as it is simplistic and will overestimate the exposure; however it is useful as a guide, when the results of different evaluations are in different units of measure which are not compatible. Different countries and regions have different guidelines for handling the uncertainties from the evaluation. Annex D provides information on the two most common methods.
- Standard75 pagesEnglish languagesale 10% offe-Library read for1 day
This European Standard provides the procedure for the specific assessment required in Annex A of EN 50527 1:2010 for workers with implanted pacemakers. It offers different approaches for doing the risk assessment. The most suitable one shall be used. If the worker has other AIMDs implanted additionally, they have to be assessed separately. The purpose of the specific assessment is to determine the risk for workers with implanted pacemakers arising from exposure to electromagnetic fields at the workplace. The assessment includes the likelihood of clinically significant effects and takes account of both transient and long-term exposure within specific areas of the workplace. NOTE 1 This standard does not address risks to workers bearing a pacemaker from contact currents. The techniques described in the different approaches may also be used for the assessment of publicly accessible areas. The frequency range to be observed is from 0 Hz to 3 GHz. Above 3 GHz no interference with the pacemaker occurs when the exposure limits are not exceeded. NOTE 2 The rationale for limiting the observation range to 3 GHz can be found in Clause 5 of ANSI/AAMI PC69:2007.
- Standard65 pagesEnglish languagesale 10% offe-Library read for1 day
Applies to the frequency range for which exposure limits are based on the induction of voltages or currents in the human body when exposed to electric fields. Defines in detail the coupling factor K - introduced by the IEC 62226 series to enable exposure assessment for complex exposure situations, such as non-uniform magnetic field or perturbed electric field - for the case of simple models of the human body, exposed to uniform electric fields. The coupling factor K has different physical interpretations depending on whether it relates to electric or magnetic field exposure. It is the so called 'shape factor for electric field'. This part of IEC 62226 can be used when the electric field can be considered to be uniform, for frequencies up to at least 100 kHz. This situation of exposure to a 'uniform' electric field is mostly found in the vicinity of high voltage overhead power systems. For this reason, illustrations given in this part are given for power frequencies (50 Hz and 60 Hz).
- Standard57 pagesEnglish languagesale 10% offe-Library read for1 day
The object of this standard is to provide methods for assessing compliance with the requirements of the Directive 2004/40/EC [8] at a site operating one or more broadcast transmitters. This standard covers the frequency range up to 40 GHz. NOTE The Council and European Parliament Directive 2004/40/EC will be transposed into national legislation in all the EU member countries. It is recommended that users of this standard consult the national legislation related to this transposition in order to identify the national regulations and requirements. These national regulations and requirements may have additional requirements that are not covered by this standard.
- Standard20 pagesEnglish languagesale 10% offe-Library read for1 day
This standard applies to short wave broadcast transmitters and installations operating in the frequency range 3 MHz to 30 MHz. The objective of the standard is to specify, for such a frequency band, basic information allowing the definition of a method for assessment of compliance related to human exposure to radio frequency electromagnetic fields.
- Standard21 pagesEnglish languagesale 10% offe-Library read for1 day
Provides means for demonstrating compliance with the basic restrictions on human exposure to low and intermediate frequency electric and magnetic fields specified in exposure standards or guidelines such as those produced by IEEE and ICNIRP. The object of EN 62226 is - to propose a more realistic approach to the modelling of the human exposure to low frequency electric and magnetic fields, using a set of models of growing complexity for the field emission source, or the human body or both; - to propose standardised values for the electrical parameters of organs in human body: electrical conductivity and permittivity and their variation with the frequency. The present basic standard does not aim at replacing the definitions and procedures specified in exposure standards or guidelines, such as those produced by IEEE or ICNIRP, but aims at providing additional procedures with a view to allowing compliance assessment with these documents. The present basic standard provides means for demonstrating compliance with the basic restrictions without having to go to the sophisticated models. Nevertheless, when the exposure conditions are well characterized (such as in product standards, for example) and when results from such models are available, they can be used for demonstrating compliance with EMF standards or guidelines.
- Standard17 pagesEnglish languagesale 10% offe-Library read for1 day
This part of EN 62226 introduces the coupling factor K, to enable exposure assessment for complex exposure situations, such as non-uniform magnetic field or perturbed electric field. The coupling factor K has different physical interpretations depending on whether it relates to electric or magnetic field exposure. The aim of this part is to define in more detail this coupling factor K, for the case of simple models of the human body, exposed to non-uniform magnetic fields. It is thus called coupling factor for non-uniform magnetic field.
- Standard59 pagesEnglish languagesale 10% offe-Library read for1 day
This standard applies to a broadcast transmitter operating in the frequency range 30 MHz to 40 GHz when put on the market. The objective of the standard is to specify, for such equipment operating in typical conditions, the method for assessment of compliance distances according to the basic restrictions (directly or indirectly via compliance with reference levels) related to human exposure to radio frequency electromagnetic fields.
- Standard22 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft331 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft5 pagesEnglish languagesale 10% offe-Library read for1 day
Identical adoption of future IEC/IEEE 63184 into EN IEC/IEEE 63184
- Draft135 pagesEnglish languagesale 10% offe-Library read for1 day
This document provides general methods for measurement and calculation of quantities associated with human exposure to electromagnetic fields in the frequency range from 0 Hz to 300 GHz. It is intended specifically to be used for the assessment of emissions from products and comparison of these with the exposure limits for the general public given in Council Recommendation 1999/519/EC, and those given for workers in Directive 2013/35/EU, as appropriate. It also is intended to be used for assessment of human exposure to electromagnetic fields in the workplace to determine compliance with the requirements of Directive 2013/35/EU. This standard deals with quantities that can be measured or calculated external to the body, notably electric and magnetic field strength or power density, and includes the measurement and calculation of quantities inside the body that form the basis for protection guidelines. In particular the standard provides information on: - definitions and terminology, - characteristics of electromagnetic fields, - measurement of exposure quantities, - instrumentation requirements, - methods of calibration, - measurement techniques and procedures for evaluating exposure, - calculation methods for exposure assessment. Where an applicable electromagnetic field standard specific to a product or technology exists it is expected to be used rather than this document. EN 62311:-, Table 1 gives a list of relevant standards.
- Draft8 pagesEnglish languagesale 10% offe-Library read for1 day
This part of IEC 62209 series is applicable to any wireless communication device capable of transmitting electromagnetic fields (EMF) intended to be used at a position near the human body, in the manner described by the manufacturer, with the radiating part(s) of the device at distances up to and including 200 mm from a human body, i.e. when held in the hand or in front of the face, mounted on the body, combined with other transmitting or non-transmitting devices or accessories (e.g. belt-clip, camera or Bluetooth add-on), or embedded in garments. For transmitters used in close proximity to the human ear, the procedures of IEC 62209-1:2005 are applicable. This standard is applicable for radio frequency exposure in the frequency range of 30 MHz to 6 GHz, and may be used to measure simultaneous exposures from multiple radio sources used in close proximity to human body. Definitions and evaluation procedures are provided for the following general categories of device types: body-mounted, body-supported, desktop, front-of-face, hand-held, laptop, limb-mounted, multi-band, push-to-talk, clothing-integrated. The types of devices considered include but are not limited to mobile telephones, cordless microphones, auxiliary broadcast devices and radio transmitters in personal computers. This International Standard gives guidelines for a reproducible and conservative measurement methodology for determining the compliance of wireless devices with the SAR limits. Because studies suggest that exclusion of features to represent a hand in human models constitutes a conservative case scenario for SAR in the trunk and the head, a representation of a hand is not included if the device is intended to be used next to the head or supported on or near the torso [73], [80]. This standard does not apply for exposures from transmitting or non-transmitting implanted medical devices. This standard does not apply for exposure from devices at distances greater than 200 mm away from the human body. IEC 62209-2 makes cross-reference to IEC 62209-1:2005 where complete clauses or subclauses apply, along with any changes specified.
- Amendment6 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 62209-1:2016 specifies protocols and test procedures for measurement of the peak spatial-average SAR induced inside a simplified model of the head with defined reproducibility. It applies to certain electromagnetic field (EMF) transmitting devices that are positioned next to the ear, where the radiating structures of the device are in close proximity to the human head, such as mobile phones, cordless phones, certain headsets, etc. These protocols and test procedures provide a conservative estimate with limited uncertainty for the peak-spatial SAR that would occur in the head for a significant majority of people during normal use of these devices. The applicable frequency range is from 300 MHz to 6 GHz. This second edition cancels and replaces the first edition published in 2005. This edition constitutes a technical revision.
- Standard228 pagesEnglish languagesale 10% offe-Library read for1 day
2013-12-02 Publication editing allocated to svandriessche@cencenelec.eu * In hands of WG 1
- Amendment10 pagesEnglish languagesale 10% offe-Library read for1 day