Expression of performance of gas analyzers - Part 1: General

IEC 61207-1:2010 specifies the terminology, definitions, requirements for statements by manufacturers and tests that are common to all gas analyzers. It is applicable to analyzers specified for permanent installation in any location (indoors or outdoors) and to such analyzers utilizing either a sample handling system or an in situ measurement technique. It is applicable to the complete analyzer when supplied by one manufacturer as an integral unit, comprised of all mechanical, electrical and electronic portions. It also applies to sensor units alone and electronic units alone when supplied separately or by different manufacturers. This second edition cancels and replaces the first edition published in 1994 and constitutes a technical revision. The significant technical changes with respect to the first edition are the following:
- all references (normative and informative) have been updated, deleted or added, as appropriate;
- all the terms and definitions relating to this International Standard have been updated;
- all references to 'errors' have been replaced by 'uncertainties' and appropriate updated definitions applied.
Where only one value is quoted for a performance specification, such as intrinsic uncertainty, linearity uncertainty or repeatability throughout a measurement range, this has now been defined as the maximum value, rather than an average or 'representative' value. This was previously undefined. Where zero and 100 % span calibration gases are used, there is now a defined requirement that the analyser must be able to respond within its standard performance specifications beyond its normal measurement range, to allow for any under or over response of the instrument to be recorded. A new Annex A has been added giving recommended standard values of influence.

Angabe zum Betriebsverhalten von Gasanalysatoren - Teil 1: Allgemeines

Expression des performances des analyseurs de gaz - Partie 1: Généralités

La CEI 61207-1:2010 comprend la terminologie, les définitions, les exigences de déclarations des fabricants ainsi que les essais communs à tous les analyseurs de gaz. Elle s'applique aux analyseurs spécifiés pour installation permanente en tout emplacement (à l'intérieur comme à l'extérieur) et aux analyseurs tels que ceux utilisant soit un système de manipulation des échantillons soit une technique de mesure sur site. Elle s'applique à l'analyseur complet lorsque celui-ci est fourni par un seul fabricant en une unité intégrée comprenant toutes les parties mécaniques, électriques et électroniques. Elle s'applique également aux unités de capteur seules et unités électroniques seules lorsqu'elles sont fournies séparément ou par des fabricants différents. Cette deuxième édition annule et remplace la première édition, parue en 1994, et constitue une révision technique. Cette édition inclut les modification techniques majeures suivantes:
- toutes les références (normatives et informatives) ont été mises à jour, retirées ou ajoutées comme il convient;
- tous les termes et définitions en rapport avec la présente Norme internationale ont été mis à jour;
- toutes les références aux erreurs ont été remplacées par 'incertitude' et les définitions ont été mises à jour comme il convient.
Lorsqu'une seule valeur est appelée pour une spécification de performance, telle que l'incertitude intrinsèque, l'incertitude de linéarité ou de répétabilité, sur une étendue de mesure, cette valeur est maintenant définie comme valeur maximale plutôt qu'une moyenne ou une valeur représentative. Auparavant, ceci était indéfini. Lorsque les gaz d'étalonnage pour le zéro et le 100 % sont utilisés, il y a maintenant une exigence définie à savoir il faut que l'analyseur soit capable de répondre dans ses performances normalisées au delà de son étendue normale de mesure, afin de permettre l'enregistrement des réponses en deçà ou au-delà. Une nouvelle Annexe A a été ajouté pour donner les valeurs d'influence normalisées recommandées.

Prikaz lastnosti analizatorjev plina - 1. del: Splošno (IEC 61207-1:2010)

Ta del IEC 61207 velja za analizatorje plina, ki se uporabljajo za ugotavljanje določenih sestavin plinskih mešanic. Ta del IEC 61207 določa terminologijo, opredelitve, zahteve za izjave proizvajalcev in preskuse, skupne za vse analizatorje plinov. Drugi mednarodni standardi v tej seriji, na primer IEC 61207-2, opisujejo vidike, ki so specifični za določene vrste (ki uporabljajo elektrokemične senzorje za visoko temperaturo). Ta del IEC 61207 je v skladu s splošnimi načeli, določenimi v IEC 60359 in IEC 60770.  Ta standard velja za analizatorje, namenjene stalni vgradnji na katerem koli mestu (v zaprtih prostorih ali zunaj), in analizatorje, ki uporabljajo sistem za rokovanje z vzorci ali tehniko za merjenje in situ. Ta standard velja za celoten analizator, če ga proizvajalec dobavi kot celoto, sestavljeno iz vseh mehanskih, električnih in elektronskih delov. Velja tudi za samostojne senzorske in elektronske enote, če so dobavljene posebej ali s strani različnih proizvajalcev. V tem standardu vsak regulator za omrežno napajanje ali neomrežno napajanje, ki je na voljo z analizatorjem ali ga določi proizvajalec, velja za del analizatorja, ne glede na to, ali je vgrajen v analizator ali pa se nahaja v ločenem ohišju. Varnostne zahteve obravnava IEC 61010-1. Če vzorec vsebuje eno ali več vnetljivih sestavin in je prisoten zrak ali druga plinska mešanica, ki vsebuje kisik ali drugo oksidacijsko sestavino, se koncentracijski razpon reaktivnih sestavin omeji na nevnetljive ravni. Standardni razpon analognega enosmernega toka in pnevmatski signali, ki se uporabljajo v sistemih za nadzor procesa, so obravnavani v IEC 60381-1 in IEC 60382. Specifikacije vrednosti za preskušanje vplivnih veličin so navedene v IEC 60654. Zahteve za dokumentacijo, ki jo je treba priložiti instrumentov, obravnava IEC 61187. Zahteve za splošna načela glede veličin, enot in simbolov obravnava ISO 1000. Glej tudi ISO 31-0. Ta del IEC 61207 ne velja za: - dodatke, kot so registratorji, pretvornike med analognim in digitalnim ali sisteme za pridobivanje podatkov, ki se uporabljajo v povezavi z analizatorjem, razen v primeru, ko se združita dva ali več analizatorjev in se prodajajo kot podsistem in je na voljo ena elektronska enota, ki zagotavlja neprekinjeno merjenje več lastnosti, in enota za odčitavanje velja za del analizatorja. Podobno so vključeni pretvorniki med električno poljsko jakostjo in tokom ter električno poljsko jakostjo in tlakom, ki so sestavni del analizatorja. Namen tega dela IEC 61207 je: - določiti splošne vidike pri terminologiji in opredelitvah, povezane z delovanjem analizatorjev plinov, ki se uporabljajo za neprekinjeno merjenje plinske sestave; - poenotiti metode, ki se uporabljajo pri izdelavi in preverjanju izjav o delovanju takšnih analizatorjev; - določiti preskuse in način njihove izvedbe za ugotavljanje delovanja; - zagotoviti osnovne dokumente za podporo uporabe standardov zagotavljanja kakovosti v okviru ISO 9001.

General Information

Status
Published
Publication Date
29-Jul-2010
Withdrawal Date
30-Jun-2013
Current Stage
6060 - Document made available - Publishing
Start Date
30-Jul-2010
Completion Date
30-Jul-2010

Relations

Standard
EN 61207-1:2010
English language
42 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-oktober-2010
1DGRPHãþD
SIST EN 61207-1:1998
Prikaz lastnosti analizatorjev plina - 1. del: Splošno (IEC 61207-1:2010)
Expression of performance of gas analyzers - Part 1: General (IEC 61207-1:2010)
Angabe zum Betriebsverhalten von Gasanalysatoren - Teil 1: Allgemeines (IEC 61207-
1:2010)
Expression des performances des analyseurs de gaz - Partie 1: Généralités (CEI 61207-
1:2010)
Ta slovenski standard je istoveten z: EN 61207-1:2010
ICS:
71.040.40 Kemijska analiza Chemical analysis
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD
EN 61207-1
NORME EUROPÉENNE
July 2010
EUROPÄISCHE NORM
ICS 19.080; 71.040.40 Supersedes EN 61207-1:1994

English version
Expression of performance of gas analyzers -
Part 1: General
(IEC 61207-1:2010)
Expression des performances  Angabe zum Betriebsverhalten
des analyseurs de gaz - von Gasanalysatoren -
Partie 1: Généralités Teil 1: Allgemeines
(CEI 61207-1:2010) (IEC 61207-1:2010)

This European Standard was approved by CENELEC on 2010-07-01. CENELEC members are bound to comply
with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard
the status of a national standard without any alteration.

Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the Central Secretariat or to any CENELEC member.

This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and notified
to the Central Secretariat has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus,
the Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy,
Latvia, Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia,
Spain, Sweden, Switzerland and the United Kingdom.

CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung

Management Centre: Avenue Marnix 17, B - 1000 Brussels

© 2010 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 61207-1:2010 E
Foreword
The text of document 65B/741/FDIS, future edition 2 of IEC 61207-1, prepared by SC 65B, Devices &
process analysis, of IEC TC 65, Industrial-process measurement, control and automation, was submitted
to the IEC-CENELEC parallel vote and was approved by CENELEC as EN 61207-1 on 2010-07-01.
This European Standard supersedes EN 61207-1:1994.
The significant technical changes with respect to EN 61207-1:1994 are the following:
– All references (normative and informative) have been updated, deleted or added, as appropriate.
– All the terms and definitions relating to this International Standard have been updated.
– All references to “errors” have been replaced by “uncertainties” and appropriate updated definitions
applied.
– Where only one value is quoted for a performance specification, such as intrinsic uncertainty, linearity
uncertainty or repeatability throughout a measurement range, this has now been defined as the
maximum value, rather than an average or “representative” value. This was previously undefined.
– Where zero and 100 % span calibration gases are used, there is now a defined requirement that the
analyser must be able to respond within its standard performance specifications beyond its normal
measurement range, to allow for any under or over response of the instrument to be recorded.
– A new Annex A has been added giving recommended standard values of influence.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN and CENELEC shall not be held responsible for identifying any or all such patent
rights.
The following dates were fixed:
– latest date by which the EN has to be implemented
at national level by publication of an identical
national standard or by endorsement (dop) 2011-04-01
– latest date by which the national standards conflicting
with the EN have to be withdrawn (dow) 2013-07-01
Annex ZA has been added by CENELEC.
__________
- 3 - EN 61207-1:2010
Endorsement notice
The text of the International Standard IEC 61207-1:2010 was approved by CENELEC as a European
Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards indicated:
IEC 61207-2 NOTE  Harmonized as EN 61207-2.
IEC 61298 series NOTE  Harmonized in EN 61298 series (not modified).
IEC 61326 series NOTE  Harmonized in EN 61326 series (not modified).
ISO 6141 NOTE  Harmonized as EN ISO 6141.
ISO 6142 NOTE  Harmonized as EN ISO 6142.
ISO 6143 NOTE  Harmonized as EN ISO 6143.
ISO 6144 NOTE  Harmonized as EN ISO 6144.
ISO 9001 NOTE  Harmonized as EN ISO 9001
ISO 16664 NOTE  Harmonized as EN ISO 16664.
__________
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications

The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.

NOTE  When an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD
applies.
Publication Year Title EN/HD Year

IEC 60068 Series Environmental testing EN 60068 Series

IEC 60359 2001 Electrical and electronic measurement EN 60359 2002
equipment - Expression of performance

IEC 60381-1 - Analogue signals for process control HD 452.1 -
systems -
Part 1: Direct current signals

IEC 60382 - Analogue pneumatic signal for process EN 60382 -
control systems
IEC 60654 Series Industrial-process measurement and control EN 60654 Series
equipment - Operating conditions -

IEC 60654-1 - Industrial-process measurement and control EN 60654-1 -
equipment - Operating conditions -
Part 1: Climatic conditions
IEC 60770 Series Transmitters for use in industrial-process EN 60770 Series
control systems
IEC 60770-1 - Transmitters for use in industrial-process EN 60770-1 -
control systems -
Part 1: Methods for performance evaluation

IEC 61010-1 - Safety requirements for electrical equipment EN 61010-1 -
for measurement, control and laboratory use -
Part 1: General requirements
IEC 61187 - Electrical and electronic measuring EN 61187 -
equipment - Documentation
ISO 31-0 - Quantities and units - - -
Part 0: General principles
ISO 1000 - SI units and recommendations for the use of - -
their multiples and of certain other units

IEC 61207-1 ®
Edition 2.0 2010-05
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Expression of performance of gas analyzers –
Part 1: General
Expression des performances des analyseurs de gaz –
Partie 1: Généralités
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
W
CODE PRIX
ICS 19.080; 71.040.40 ISBN 978-2-88910-947-0
– 2 – 61207-1 © IEC:2010
CONTENTS
FOREWORD.4
1 Scope and object.6
2 Normative references .7
3 Terms and definitions .7
3.1 General .7
3.2 Basic terms and definitions.8
3.3 General terms and definitions of devices and operations .11
3.4 Terms and definitions on manners of expression .15
3.5 Specific terms and definitions for gas analyzers .18
4 Procedure for specification .20
4.1 Specification of values and ranges .20
4.2 Operation, storage and transport conditions .21
4.3 Performance characteristics requiring statements of rated values.21
4.4 Uncertainty limits to be stated for each specified range .22
4.4.1 General .22
4.4.2 Limits of intrinsic uncertainty .22
4.4.3 Variations .22
4.5 Other performance characteristics.23
5 Procedure for compliance testing .23
5.1 General .23
5.1.1 Compliance tests .23
5.1.2 Test instruments.23
5.1.3 Test instrument uncertainties.23
5.1.4 Influence quantities .24
5.1.5 Operational conditions.24
5.2 Calibration gases .24
5.3 Adjustments made during tests.24
5.4 Reference conditions during measurement of intrinsic uncertainty.24
5.5 Reference conditions during measurement of influence quantity.24
5.6 Testing procedures.25
5.6.1 General .25
5.6.2 Intrinsic uncertainty .25
5.6.3 Linearity uncertainty .25
5.6.4 Repeatability .26
5.6.5 Output fluctuation .26
5.6.6 Drift .27
5.6.7 Delay time, rise time and fall time.27
5.6.8 Warm-up time.28
5.6.9 Interference uncertainty.28
5.6.10 Variations .29
Annex A (informative) Recommended standard values of influence – Quantities
affecting performance from IEC 60359.31
Annex B (informative) Performance characteristics calculable from drift tests .37
Bibliography.38

Figure 1 – Rise and fall times .20

61207-1 © IEC:2010 – 3 –
Figure 2 – Output fluctuations .26

Table A.1 – Mains supply voltage .35
Table A.2 – Mains supply frequency .35
Table A.3 – Ripple of d.c. supply .36
Table B.1 – Data: applied concentration 1 000 units .37

– 4 – 61207-1 © IEC:2010
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
EXPRESSION OF PERFORMANCE OF GAS ANALYZERS –

Part 1: General
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 61207-1 has been prepared by subcommittee 65B: Devices and
process analysis, of IEC technical committee 65: Industrial-process measurement, control and
automation.
This second edition cancels and replaces the first edition published in 1994 and constitutes a
technical revision.
The significant technical changes with respect to the first edition are the following:
a) All references (normative and informative) have been updated, deleted or added, as
appropriate.
b) All the terms and definitions relating to this International Standard have been updated.
c) All references to “errors” have been replaced by “uncertainties” and appropriate updated
definitions applied.
d) Where only one value is quoted for a performance specification, such as intrinsic
uncertainty, linearity uncertainty or repeatability throughout a measurement range, this

61207-1 © IEC:2010 – 5 –
has now been defined as the maximum value, rather than an average or “representative”
value. This was previously undefined.
e) Where zero and 100 % span calibration gases are used, there is now a defined
requirement that the analyser must be able to respond within its standard performance
specifications beyond its normal measurement range, to allow for any under or over
response of the instrument to be recorded.
f) A new Annex A has been added giving recommended standard values of influence.
The text of this standard is based on the following documents:
FDIS Report on voting
65B/741/FDIS 65B/752/RVD
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
A list of all parts of the IEC 61207 series, under the general title Expression of performance of
gas analyzers, can be found on the IEC website.
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
– 6 – 61207-1 © IEC:2010
EXPRESSION OF PERFORMANCE OF GAS ANALYZERS –

Part 1: General
1 Scope and object
This part of IEC 61207 is applicable to gas analyzers used for the determination of certain
constituents in gaseous mixtures.
This part of IEC 61207 specifies the terminology, definitions, requirements for statements by
manufacturers and tests that are common to all gas analyzers. Other international standards
in this series, for example IEC 61207-2, describe those aspects that are specific to certain
types (utilizing high-temperature electrochemical sensors).
This part IEC 61207 is in accordance with the general principles set out in IEC 60359 and
IEC 60770.
This standard is applicable to analyzers specified for permanent installation in any location
(indoors or outdoors) and to such analyzers utilizing either a sample handling system or an in
situ measurement technique.
This standard is applicable to the complete analyzer when supplied by one manufacturer as
an integral unit, comprised of all mechanical, electrical and electronic portions. It also applies
to sensor units alone and electronic units alone when supplied separately or by different
manufacturers.
For the purposes of this standard, any regulator for mains-supplied power or any non-mains
power supply, provided with the analyzer or specified by the manufacturer, is considered part
of the analyzer whether it is integral with the analyzer or housed separately.
Safety requirements are dealt with in IEC 61010-1.
If one or more components in the sample is flammable, and air or another gas mixture
containing oxygen or other oxidizing component is present, then the concentration range of
the reactive components are limited to levels which are not within flammability limits.
Standard range of analogue d.c. current and pneumatic signals used in process control
systems are dealt with in IEC 60381-1 and IEC 60382.
Specifications for values for the testing of influence quantities can be found in IEC 60654.
Requirements for documentation to be supplied with instruments are dealt with in IEC 61187.
Requirements for general principles concerning quantities, units and symbols are dealt with in
ISO 1000. See also ISO 31-0.
This part of IEC 61207 does not apply to:
– accessories such as recorders, analogue-to-digital converters or data acquisition systems
used in conjunction with the analyzer, except that when two or more such analyzers are
combined and sold as a subsystem and a single electronic unit is supplied to provide
continuous measurement of several properties, that read-out unit is considered to be part
of the analyzer. Similarly, e.m.f-to-current or e.m.f-to-pressure converters which are an
integral part of the analyzer are included.

61207-1 © IEC:2010 – 7 –
The object of this part of IEC 61207 is:
– to specify the general aspects in the terminology and definitions related to the
performance of gas analyzers used for the continuous measurement of gas composition;
– to unify methods used in making and verifying statements on the functional performance of
such analyzers;
– to specify which tests should be performed in order to determine the functional
performance and how such tests should be carried out;
– to provide basic documents to support the application of standards of quality assurance
within ISO 9001.
2 Normative references
The following referenced documents are indispensable for the application of this document.
For dated references, only the edition cited applies. For undated references, the latest edition
of the referenced document (including any amendments) applies.
IEC 60068 (all parts), Environmental testing
IEC 60359:2001, Electrical and electronic measurement equipment – Expression of
performance
IEC 60381-1, Analogue signals for process control systems – Part 1: Direct current signals
IEC 60382, Analogue pneumatic signal for process control systems
IEC 60654 (all parts), Industrial-process measurement and control equipment – Operating
conditions
IEC 60654-1, Industrial-process measurement and control equipment – Operating conditions –
Part 1: Climatic conditions
IEC 60770 (all parts), Transmitters for use in industrial-process control systems
IEC 60770-1, Transmitters for use in industrial-process control systems – Part 1: Methods for
performance evaluation
IEC 61010-1, Safety requirements for electrical equipment for measurement, control and
laboratory use – Part 1: General requirements
IEC 61187, Electrical and electronic measurement equipment – Documentation
ISO 31-0, Quantities and units – General principles
ISO 1000, SI units and recommendations for the use of their multiples and of certain other
units
3 Terms and definitions
3.1 General
For the purposes of this document, the following terms and definitions apply. The definitions
in 3.2 (excepting 3.2.17), 3.3 and 3.4 are taken from IEC 60359.

– 8 – 61207-1 © IEC:2010
3.2 Basic terms and definitions
3.2.1
measurand
quantity subjected to measurement, evaluated in the state assumed by the measured system
during the measurement itself
NOTE 1 The value assumed by a quantity subjected to measurement when it is not interacting with the measuring
instrument may be called unperturbed value of the quantity.
NOTE 2 The unperturbed value and its associated uncertainty can only be computed through a model of the
measured system and of the measurement interaction with the knowledge of the appropriate metrological
characteristics of the instrument that may be called instrumental load.
3.2.2
(result of a) measurement
set of values attributed to a measurand, including a value, the corresponding uncertainty and
the unit of measurement
[IEC 60050-311, 311-01-01, modified]
NOTE 1 The mid-value of the interval is called the value (see 3.2.3) of the measurand and its half-width the
uncertainty (see 3.2.4).
NOTE 2 The measurement is related to the indication (see 3.2.5) given by the instrument and to the values of
correction obtained by calibration.
NOTE 3 The interval can be considered as representing the measurand provided that it is compatible with all
other measurements of the same measurand.
NOTE 4 The width of the interval, and hence the uncertainty, can only be given with a stated level of confidence
(see 3.2.4, NOTE 1).
3.2.3
(measure-) value
mid element of the set assigned to represent the measurand
NOTE The measure-value is no more representative of the measurand than any other element of the set. It is
singled out merely for the convenience of expressing the set in the format V ± U, where V is the mid element and U
the half-width of the set, rather than by its extremes. The qualifier "measure-" is used when deemed necessary to
avoid confusion with the reading-value or the indicated value.
3.2.4
uncertainty (of measurement)
parameter, associated with the result of a measurement, that characterizes the dispersion of
the values that could reasonably be attributed to the measurand
NOTE 1 The parameter can be, for example, a standard deviation (or a given multiple of it), or a half-width of an
interval having a stated level of confidence.
NOTE 2 Uncertainty of measurement comprises, in general, many components. Some of these components can
be evaluated from the statistical distribution of the results of a series of measurements and can be characterized
by experimental standard deviations. The other components, which can also be characterized by standard
deviations, are evaluated from the assumed probability distributions based on experience or other information.
[IEC 60050-311, 311-01-02, ISO/IEC Guide 99, 2.26 modified]
NOTE 3 It is understood that the result of the measurement is the best estimate of the value of the measurand,
and that all components of uncertainty, including those arising from systematic effects, such as components
associated with corrections and reference standards, contribute to the dispersion.
NOTE 4 The definition and notes 1 and 2 are from GUM, Clause B.2.18. The option used in this standard is to
express the uncertainty as the half-width of an interval with the GUM procedures with a coverage factor of 2. This
choice corresponds to the practice now adopted by many national standards laboratories. With the normal
distribution a coverage factor of 2 corresponds to a level of confidence of 95 %. Otherwise statistical elaborations
are necessary to establish the correspondence between the coverage factor and the level of confidence. As the
data for such elaborations are not always available, it is deemed preferable to state the coverage factor. This
interval can be "reasonably" assigned to describe the measurand, in the sense of the GUM definition, as in most

61207-1 © IEC:2010 – 9 –
usual cases it ensures compatibility with all other results of measurements of the same measurand assigned in the
same way at a sufficiently high confidence level.
NOTE 5 Following CIPM document INC-1 and ISO/IEC Guide 98-3, the components of uncertainty that are
evaluated by statistical methods are referred to as components of category A, and those evaluated with the help of
other methods as components of category B.
3.2.5
indication or reading-value
output signal of the instrument
[IEC 60050-311, 311-01-01, modified]
NOTE 1 The indicated value can be derived from the indication by means of the calibration curve.
NOTE 2 For a material measure, the indication is its nominal or stated value.
NOTE 3 The indication depends on the output format of the instrument:
– for analogue outputs it is a number tied to the appropriate unit of the display;
– for digital outputs it is the displayed digitized number;
– for code outputs it is the identification of the code pattern.
NOTE 4 For analogue outputs meant to be read by a human observer (as in the index-on-scale instruments) the
unit of output is the unit of scale numbering; for analogue outputs meant to be read by another instrument (as in
calibrated transducers) the unit of output is the unit of measurement of the quantity supporting the output signal.
3.2.6
calibration
set of operations which establishes the relationship which exists, under specified conditions,
between the indication and the result of a measurement
[IEC 60050-311, 311-01-09]
NOTE 1 The relationship between the indications and the results of measurement can be expressed, in principle,
by a calibration diagram.
NOTE 2 The calibration must be performed under well-defined operating conditions for the instrument. The
calibration diagram representing its result is not valid if the instrument is operated under conditions outside the
range used for the calibration.
NOTE 3 Quite often,e specially for instruments whose metrological characteristics are sufficiently known from
past experience, it is convenient to predefine a simplified calibration diagram and perform only a verification of
calibration (see 3.3.12) to check whether the response of the instrument stays within its limits. The simplified
diagram is, of course, wider than the diagram that would be defined by the full calibration of the instrument, and
the uncertainty assigned to the results of measurements is consequently larger.
3.2.7
calibration diagram
portion of the co-ordinate plane, defined by the axis of indication and the axis of results of
measurement, which represents the response of the instrument to differing values of the
measurand
[IEC 60050-311, 311-01-10]
3.2.8
calibration curve
curve which gives the relationship between the indication and the value of the measurand
NOTE 1 When the calibration curve is a straight line passing through zero, it is convenient to refer to the slope
which is known as the instrument constant.
[IEC 60050-311, 311-01-11]
NOTE 2 The calibration curve is the curve bisecting the width of the calibration diagram parallel to the axis of
results of measurement, thus joining the points representing the values of the measurand.

– 10 – 61207-1 © IEC:2010
3.2.9
indicated value
value given by an indicating instrument on the basis of its calibration curve
[IEC 60050-311, 311-01-08]
NOTE The indicated value is the measure-value of the measurand when the instrument is used in a direct
measurement (see 3.3.7) under all the operating conditions for which the calibration diagram is valid.
3.2.10
(measurement) compatibility
property satisfied by all the results of measurement of the same measurand, characterized by
an adequate overlap of their intervals
[IEC 60050-311, 311-01-14]
NOTE 1 The compatibility of any result of a measurement with all the other ones that represent the same
measurand can be asserted only at some level of confidence, as it depends on statistical inference, a level that
should be indicated, at least by implicit convention or through a coverage factor.
NOTE 2 The compatibility of the results of measurements obtained with different instruments and methods is
ensured by the traceability (see 3.2.16) to a common primary standard (see 3.3.6) of the standards used for the
calibration of the several instruments (and of course by the correctness of the calibration and operation
procedures).
NOTE 3 When two results of a measurement are not compatible it must be decided by independent means
whether one or both results are wrong (perhaps because the uncertainty is too narrow), or whether the measurand
is not the same.
NOTE 4 Measurements carried out with wider uncertainty yield results which are compatible on a wider range,
because they discriminate less among different measurands allowing to classify them with simpler models; with
narrower uncertainties the compatibility calls for more detailed models of the measured systems.
3.2.11
intrinsic uncertainty of the measurand
minimum uncertainty that can be assigned in the description of a measured quantity
NOTE 1 No quantity can be measured with narrower and narrower uncertainty, in as much as any given quantity
is defined or identified at a given level of detail. If one tries to measure a given quantity with uncertainty lower than
its own intrinsic uncertainty one is compelled to redefine it with higher detail, so that one is actually measuring
another quantity. See also GUM D.1.1.
NOTE 2 The result of a measurement carried out with the intrinsic uncertainty of the measurand may be called the
best measurement of the quantity in question.
3.2.12
(absolute) instrumental uncertainty
uncertainty of the result of a direct measurement of a measurand having negligible intrinsic
uncertainty
NOTE 1 Unless explicitly stated otherwise, the instrumental uncertainty is expressed as an interval with coverage
factor 2.
NOTE 2 In single-reading direct measurements of measurands having intrinsic uncertainty small with respect to
the instrumental uncertainty, the uncertainty of the measurement coincides, by definition, with the instrumental
uncertainty. Otherwise the instrumental uncertainty is to be treated as a component of category B in evaluating the
uncertainty of the measurement on the basis of the model connecting the several direct measurements involved.
NOTE 3 The instrumental uncertainty automatically includes, by definition, the effects due to the quantization of
the reading-values (minimum evaluable fraction of the scale interval in analogic outputs, unit of the last stable digit
in digital outputs).
NOTE 4 For material measures the instrumental uncertainty is the uncertainty that should be associated to the
value of the quantity reproduced by the material measure in order to ensure the compatibility of the results of its
measurements.
NOTE 5 When possible and convenient the uncertainty may be expressed in the relative form (see 3.4.3) or in the
fiducial form (see 3.4.4). The relative uncertainty is the ratio U/V of the absolute uncertainty U to the measure

61207-1 © IEC:2010 – 11 –
value V, and the fiducial uncertainty the ratio U/V of the absolute uncertainty U to a conventionally chosen value
f
V .
f
3.2.13
conventional value measure
value of a standard used in a calibration operation and known with uncertainty negligible with
respect to the uncertainty of the instrument to be calibrated
NOTE This definition is adapted to the object of this standard from the definition of "conventional true value (of a
quantity)": value attributed to a particular quantity and accepted, sometimes by convention, as having an
uncertainty appropriate for a given purpose (see IEC 60050-311, 311-01-06, ISO/IEC Guide 99, 2.13 modified).
3.2.14
influence quantity
quantity which is not the subject of the measurement and whose change affects the
relationship between the indication and the result of the measurement
NOTE 1 Influence quantities can originate from the measured system, the measuring equipment or the
environment.
NOTE 2 As the calibration diagram depends on the influence quantities, in order to assign the result of a
measurement it is necessary to know whether the relevant influence quantities lie within the specified range.
[IEC 60050-311, 311-06-01]
NOTE 3 An influence quantity is said to lie within a range C’ to C" when the results of its measurement satisfy the
relationship: C' ≤ V – U < V + U ≤ C".
3.2.15
steady-state conditions
operating conditions of a measuring device in which the variation of the measurand with the
time is such that the relation between the input and output signals of the instruments does not
suffer a significant change with respect to the relation obtaining when the measurand is
constant in time
3.2.16
traceability
property of the result of a measurement or of the value of a standard such that it can be
related to stated references, usually national or international standards, through an unbroken
chain of comparisons all having stated uncertainties
[IEC 60050-311, 311-01-15, ISO/IEC Guide 99, 2.41 modified]
NOTE 1 The concept is often expressed by the adjective traceable.
NOTE 2 The unbroken chain of comparisons is called a traceability chain.
NOTE 3 The traceability implies that a metrological organization be established with a hierarchy of standards
(instruments and material measures) of increasing intrinsic uncertainty. The chain of comparisons from the primary
standard to the calibrated device adds indeed new uncertainty at each step.
NOTE 4 Traceability is ensured only within a given uncertainty that should be specified.
3.2.17
mean
summation of the individual values divided by the total number of values for a set of values
3.3 General terms and definitions of devices and operations
3.3.1
(measuring) instrument
device intended to be used to make measurements, alone or in conjunction with
supplementary devices
[IEC 60050-311, 311-03-01, ISO/IEC Guide 99, 3.1 modified]

– 12 – 61207-1 © IEC:2010
NOTE The term "(measuring) instruments" includes both the indicating instruments and the material measures.
3.3.2
indicating (measuring) instrument
measuring instrument which displays an indication
NOTE 1 The display can be analogue (continuous or discontinuous), digital or coded [IEV].
NOTE 2 Values of more than one quantity can be displayed simultaneously [IEV].
NOTE 3 A displaying measuring instrument can also provide a record [IEV].
NOTE 4 The display can consist of an output signal not directly readable by a human observer, but able to be
interpreted by suitable devices [IEV].
[IEC 60050-311, 311-03-02, ISO/IEC Guide 99, 3.3 modified]
NOTE 5 An indicating instrument may consist of a chain of transducers with the possible addition of other process
devices, or it may consist of one transducer.
NOTE 6 The interaction between the indicating instrument, the measured system and the environment generates
a signal in the first stage of the instrument (called sensor). This signal is elaborated inside the instrument into an
output signal which carries the information on the measurand. The description of the output signal in a suitable
output format is the indication supplied by the instrument.
NOTE 7 A chain of instruments is treated as a single indicating instrument when a single calibration diagram is
available that connects the measurand to the output of the last element of the chain. In this case the influence
quantities must be defined for the whole chain.
3.3.3
material measure
device intended to reproduce or supply, in a permanent manner during its use, one or more
known values of a given quantity
NOTE 1 The quantity concerned may be called the supplied quantity [IEV].
[IEC 60050-311, 311-03-03, ISO/IEC Guide 99, 3.6 modified]
NOTE 2 The definition covers also the devices, such as signal generators and standard voltage or current
generators, often referred to as supply instruments.
NOTE 3 The identification of the value and uncertainty of the supplied quantity is given by a number tied to a unit
of measurement or a code term, called nominal value or marked value of the material measure.
3.3.4
electrical measuring instrument
measuring instrument intended to measure an electrical or non-electrical quantity using
electrical or electronic means
[IEC 60050-311, 311-03-04]
3.3.5
transducer
technical device which performs a given elaboration on an input signal, transforming it into an
output signal
NOTE All indicating instruments contain transducers and they may consist of one transducer. When the signals
are elaborated by a chain of transducers, the input and output signals of each transducer are not always directly
and univocally accessible.
3.3.6
primary standard
standard that is designated or widely acknowledged as having the highest metrological
qualities and whose value is accepted without reference to other standards of the same
quantity
NOTE 1 The concept of a primary standard is equally valid for base quantities and derived quantities.

61207-1 © IEC:2010 – 13 –
NOTE 2 A primary standard is never used directly for measurement other than for comparison with other primary
standards or reference standards.
[IEC 60050-311, 311-04-02, ISO/IEC Guide 99, 5.4 modified]
3.3.7
direct (method of) measurement
method of measurement in which the value of a measurand is obtained directly, without the
necessity for supplementary calculations based on a functional relationship between the
measurand and other quantities actually measured
NOTE 1 The value of the measurand is considered to be obtained directly even when the scale of a measuring
instrument has values which are linked to corresponding values of the measurand by means of a table or a graph
[IEV].
NOTE 2 The method of measurement remains direct even if it is necessary to make supplementary measurements
to determine the values of influence quantities in order to make corrections [IEV].
[IEC 60050-311, 311-02-01]
NOTE 3 The definitions of the metrological characteristics of the instruments refer implicitly to their use in direct
measurements.
3.3.8
indirect (method of) measurement
method of measurement in which the value of a quantity is obtained from measurements
made by direct methods of measurement of other quantities linked to the measurand by a
known relationship
[IEC 60050-311, 311-02-02]
NOTE 1 In order to apply an indirect method of measurement a model is needed which is able to supply the
relationship, and which is fully explicit, between the measurand and the parameters that are measured by direct
measurement.
NOTE 2 The computations must be carried out on both values and uncertainties, and therefore require accepted
rules for the propagation of the uncertainty as provided by GUM.
3.3.9
(method of) measurement by repeated observations
method of measurement by which the result of the measurement is assigned on the basis of a
statistical analysis on the distribution of the data obtained by several observations repeated
under nominally equal conditions
NOTE 1 One should resort to a statistical analysis when the instrumental uncertainty is too small to ensure the
measurement compatibility. This may happen in two quite different sets of circumstances:
a) when the measurand is a quantity s
...