(Main)Cryogenic vessels - Pressure-relief accessories for cryogenic service - Part 3: Sizing and capacity determination (ISO 21013-3:2026)
- Abstract
This document specifies calculation methods for determining the required mass flow to be relieved for each specified conditions described in Table 1.
Recommendations for pressure relief devices for cryostats are given in Annex A.
- Status
- Published
- Public Enquiry End Date
- 02-May-2025
- Publication Date
- 08-Nov-2026
- Technical Committee
- TLP - Pressure vessels
- Current Stage
- 6060 - National Implementation/Publication (Adopted Project)
- Start Date
- 14-Oct-2026
- Due Date
- 19-Dec-2026
- Completion Date
- 09-Nov-2026
SIST EN ISO 21013-3:2026 is the SIST-published European adoption of ISO 21013-3 for cryogenic vessels. It specifies calculation methods for sizing pressure-relief accessories by determining the mass flow that must be relieved under the vessel, insulation, pressure build-up, piping and fire conditions covered by the standard. Engineers use it when they need to size or verify pressure relief valves and bursting discs for cryogenic service, including cryostats.
What does SIST EN ISO 21013-3:2026 specify?
SIST EN ISO 21013-3:2026 covers the sizing and capacity determination of pressure-relief accessories for cryogenic service. It gives separate calculation routes for the specified vessel and service conditions, so the user can work from heat input to required mass flow and then to device sizing.
The document is organized into Clauses 1 to 7, followed by Annex A and Annex ZA. Clause 3 defines symbols, Clause 4 calculates heat transfer into the vessel, Clause 5 converts that heat input into required mass flow, Clause 6 addresses piping effects, and Clause 7 sizes pressure relief devices.
| Annex | What it covers |
|---|---|
| Annex A | Cryostats and additional recommendations for their pressure-relief design |
| Annex ZA | Relationship with the essential requirements of Directive 2014/68/UE |
What are the key requirements of SIST EN ISO 21013-3:2026?
Heat input must be calculated for the actual service condition
Clause 4 sets out the heat-transfer cases that feed the sizing calculations. SIST EN ISO 21013-3:2026 treats normal vacuum insulation, pressure build-up devices, loss of vacuum, supports and piping, fire exposure, and air or nitrogen condensation as separate contributors.
That matters in practice because the required relief capacity depends on the worst credible heat load, not just on the steady operating case. The document also fixes the fire-case external temperature at 922 K and makes the user check both fire and pressure build-up conditions, because fire is not always the governing case.
Loss of vacuum and fire conditions are treated differently
Clause 4 distinguishes between vacuum-insulated vessels that retain insulation, vessels where insulation is lost or degraded, and non-vacuum-insulated vessels. For low-temperature fluids with a saturation temperature below 75 K at 1 bar, air or nitrogen condensation is included through the method and figure given in Clause 4.
In practice, this means the designer must identify which insulation state and which ambient condition apply before selecting the heat-transfer equation. The standard also points users to thermal-conductivity values and heat-transfer coefficients for the relevant fluids and conditions.
The required mass flow is derived from heat input and fluid properties
Clause 5 converts the total heat transfer rate, W_T, into the mass flow to be relieved, Q_m. For relieving pressure below the critical pressure, the document uses vaporization properties at relieving conditions or, alternatively, the normal evaporation rate (NER) method.
For relieving pressure at or above the critical pressure, Clause 5 uses the enthalpy-to-volume expansion ratio, L’, and a property-based search for the relieving temperature. In practice, the user must evaluate the cryogen at the actual relieving condition, because the selection of T and L’ comes from thermophysical property data rather than a fixed rule of thumb.
Piping between the vessel and the device affects the sizing result
Clause 6 requires the effect of piping to be included when the line between the outer jacket and the safety relief device is longer than 0.6 m. It also requires consideration of heat transfer into the released flow, pressure drop in the inlet and outlet pipework, and back pressure on the relief device.
For conventional direct acting relief valves, the document recommends keeping upstream pressure drop at or below 3% of set pressure and built-up back pressure at or below 10% of set pressure. That matters because excessive pressure drop can reduce discharge capacity and cause valve instability or chatter.
Combined device capacity must meet the required discharge
Clause 7 says that all pressure relief devices that must discharge together at the same relieving pressure shall have a combined relieving capacity equal to or greater than the required mass flow. The clause then distinguishes gas flow through relief valves as critical or subcritical by comparing the pressure ratio with the isentropic exponent, κ.
The sizing sequence is practical and stepwise: determine inlet and outlet conditions, check whether flow is critical or subcritical, calculate the minimum orifice area, then select the next larger available device and verify the installation again with piping losses included. This is the part that turns the heat-load calculation into an actual valve or bursting disc selection.
Bursting discs are sized with the whole pipe network in mind
Clause 7.3 treats bursting discs separately from relief valves. The complete pipe network resistance, including the disc, is used in the calculation, and the document recommends using manufacturer-certified post-rupture flow area.
That matters because a bursting disc may not provide its nominal full opening after rupture. The user therefore has to verify the network resistance and the net flow area after rupture, not just the rated disc size.
Cryostats get extra guidance in Annex A
Annex A adds recommendations for cryostats, including the need to include any heat generated by upset process conditions in the relief calculation. It also gives design recommendations for gas space, lowest set pressure, outlet pressure, and bursting-disc flow area data.
What terms does SIST EN ISO 21013-3:2026 define?
- Normal evaporation rate (NER) - the percentage of vessel mass that evaporates per day, used as an alternative basis for sizing in normal operation.
- Relieving pressure, P - the actual absolute pressure inside the vessel that is used in the mass-flow and device-sizing equations.
- Total heat transfer rate, W_T - the combined heat input for the selected condition, and the starting point for calculating required relieved mass flow.
- Coefficient of discharge, K_d - the device flow performance factor used in valve sizing, with the document also referring to a derated coefficient for sizing calculations.
- Isentropic exponent, κ - the ratio of specific heat at constant pressure to specific heat at constant volume, used to decide whether gas flow is critical or subcritical.
- Entalpy-to-volume expansion ratio, L’ - the quantity used for critical or gas-full-vessel flow calculations when relieving pressure is at or above the critical pressure.
Who uses SIST EN ISO 21013-3:2026?
SIST EN ISO 21013-3:2026 is used by engineers who design cryogenic vessels and their pressure-relief accessories. It is also used by suppliers selecting pressure relief valves or bursting discs, and by project teams checking piping losses, back pressure, and discharge capacity for cryogenic installations.
Quality, compliance and certification teams use it to review whether the relief arrangement matches the calculated duty for the vessel or cryostat. It is particularly relevant in cryogenic vessel design, pressure equipment work, and systems that handle liquid hydrogen, liquid helium, nitrogen and similar fluids.
What changed in SIST EN ISO 21013-3:2026 from the previous edition?
SIST EN ISO 21013-3:2026 replaces the second edition, ISO 21013-3:2016, and the foreword says the document has been technically revised. The stated changes are:
- precision added on the total outer surface area of the pipe network containing flow of fluid
- inclusion of the coefficient of discharge symbol
- precision added on the overall convective heat transfer coefficient of the ambient air vaporizer
- additional information added on the fire case and PBU conditions
- correction of the molar mass in kg/kmol and modification of formulas 32 and 43
- supplementary information added on cryostats
What does the SIST EN ISO 21013-3:2026 document contain?
The document contains the symbol list, the calculation methods for heat transfer and relieved mass flow, and the piping and device-sizing procedures that support relief-system design. It includes tables of thermal conductivity and specific heat for refrigerated fluids, plus tables of pipe friction and flow-resistance coefficients for elbows, tees, size changes and valves.
It also includes a figure for air or nitrogen condensation as a function of insulation layers, worked examples for liquid hydrogen valve sizing and bursting-disc network sizing, and an informative annex for cryostats. Annex ZA gives the relationship to Directive 2014/68/UE for pressure equipment.
Relations
- Effective Date
- 01-Dec-2026
- Effective Date
- 28-Oct-2026
- Effective Date
- 28-Oct-2026
- Effective Date
- 28-Oct-2026
- Effective Date
- 28-Oct-2026
Get Certified
Connect with accredited certification bodies for this standard
CIS Institut d.o.o.
Personal Protective Equipment (PPE) certification body. Notified Body NB-2890 for EU Regulation 2016/425 PPE.
Institut za varilstvo d.o.o. (Welding Institute)
Slovenia's leading welding institute since 1952. ISO 3834, EN 1090, pressure equipment certification, NDT personnel, welder qualification. Only IIW Au

Kiwa BDA Testing
Building and construction product certification.
Sponsored listings
Frequently Asked Questions
SIST EN ISO 21013-3:2026 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Cryogenic vessels - Pressure-relief accessories for cryogenic service - Part 3: Sizing and capacity determination (ISO 21013-3:2026)". This standard covers: This document specifies calculation methods for determining the required mass flow to be relieved for each specified conditions described in Table 1. Recommendations for pressure relief devices for cryostats are given in Annex A.
This document specifies calculation methods for determining the required mass flow to be relieved for each specified conditions described in Table 1. Recommendations for pressure relief devices for cryostats are given in Annex A.
SIST EN ISO 21013-3:2026 is classified under the following ICS (International Classification for Standards) categories: 13.240 - Protection against excessive pressure; 23.020.40 - Cryogenic vessels. The ICS classification helps identify the subject area and facilitates finding related standards.
SIST EN ISO 21013-3:2026 has the following relationships with other standards: It is inter standard links to SIST EN ISO 21013-3:2016, SIST EN 14398-2:2004+A2:2008, SIST EN 14398-1:2004, SIST EN 13458-2:2003, SIST EN 13648-2:2002. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
SIST EN ISO 21013-3:2026 is associated with the following European legislation: EU Directives/Regulations: 2008/68/EC, 2014/68/EU; Standardization Mandates: M/071, M/601, M/TDG. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
SIST EN ISO 21013-3:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 1 Scope
- 2 Normative references
- 3 Terms, definitions and symbols
- 3.1 Symbols
- 4 Calculation of the total quantity of heat transferred per unit time from the hot wall
- 4.1 General
- 4.2 Under conditions other than fire
- 4.3 Under fire conditions
- 4.4 Air or nitrogen condensation
- 4.5 Heat transfer per unit time (watts)
- 5 Calculation of the mass flow to be relieved by pressure relief devices
- 5.1 Relieving pressure, P, less than the critical pressure
- 5.2 Relieving pressure, P, equal to or greater than the critical pressure
- 5.3 Example
- 6 Piping for pressure relief devices
- 6.1 Pressure drop
- 6.2 Back pressure consideration
- 6.3 Heat transfer
- 7 Sizing of pressure relief devices
- 7.1 General
- 7.2 Sizing of pressure relief valves
- 7.3 Sizing of bursting discs
- Annex A
- Bibliography
1 Scope
This document specifies calculation methods for determining the required mass flow to be relieved for each specified conditions described in Table 1.
a Table 1 — Subclause required for calculating mass flow rate based on specified condition Insulation Type of system (outer Pressure Use clauses insu- Vacuum Inner vessel jacket + Ambient buildup Condition including any lation status temperature insulating conditions system subclauses system material) condition condition normal (ambient 4.2.3,4.2.4, 4.5.1, 1 intact tempera- 4.5.2, 4.5.7, 5, 6, 7 nonture) N/A vacuum normal fully or condition 4.3.1, 4.5.1, 4.5.7, 5, 2 partially in fire (922 K) 6, 7 tempera- place ture of the 4.2.1, 4.2.4, 4.5.1,
contents at 4.5.2, 4.5.7, 5, 6, 7 the specified normal normal 4.2.1, 4.2.2, 4.2.4, relieving (ambient 4 intact full open 4.5.1, 4.5.3, 4.5.7, 5, pressure tempera- 6, 7 ture) 4.2.3, 4.2.4, 4.5.1,
4.5.4, 4.5.7 ,5, 6, 7 loss of fully or vacuum vacuum 4.3.1, 4.4.3, 4.5.1, 6 partially in fire (922 K) 4.5.5, 4.5.7, 5, 6, 7 place normal (ambient 4.4.2, 4.5.1, 4.5.4, 7 normal loss of tempera- 4.5.7, 5, 6 ,7 condition intact vacuum ture) contents with with air or relieving 4.4.3, 4.5.1, 4.5.5, 8 nitrogen temperature 4.5.7, 5, 6, 7 in the below 75 K Vacuum vacuum fire (922 K) and 4.5.1, 4.5.6, 4.5.7, 5, space 9 totally lost non- 6, 7 vacuum When no appropriate calculation method is provided for a condition, the required mass flow is determined using good engineering practice based on well-established theoretical physical science. Recommendations for pressure relief devices for cryostats are given in Annex A.
SIST EN ISO 21013-3:2026 ISO 21013-3:2026(en)
2 Normative references
There are no normative references in this document.
3 Terms, definitions and symbols
No terms and definitions are listed in this document.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https:// www .iso .org/ obp
- IEC Electropedia: available at https:// www .electropedia .org/
3.1 Symbols
A arithmetic mean of inner and outer surface areas of vessel insulating material m
A actual flow area of a pipe element m B
A total outer surface area of the pipe network containing flow of fluid, from the outer m e jacket up to the flow temperature location x under consideration; applicable only when l > 0,6 m (See 6.1.1)
A minimum flow area (reference area) in a pipe network m F
A minimum flow area (reference area) in the pipe network, downstream of relief valve m Fd
A minimum flow area (reference area) in the pipe network, upstream of relief valve m Fu
A total outside surface area of inner vessel m
A total outer surface area of pipe network between the inner and outer jackets m j (interspace)
A larger flow area of a pipe element containing two different flow area sizes m L
A cross-sectional area, support, or pipe material m n A area ratio, A /A — R S L
A smaller flow area of a pipe element containing two different flow area sizes m S
A actual flow (orifice) area of a pressure relief valve mm V
A actual flow (orifice) area of pressure relief valve selected for final analysis mm Va
A minimum required relief valve flow (orifice) area mm V1
A external heat transfer surface area of ambient air vaporizer m
C experimentally determined flow rate through a pipe element or device m /h V K flow resistance coefficient of a pipe element in terms of A — A F K flow resistance coefficient of a pipe element in terms of A — B B K subcritical flow coefficient — b K coefficient of discharge — d
SIST EN ISO 21013-3:2026 ISO 21013-3:2026(en) K derated coefficient of discharge — dr K derated coefficient of discharge of next largest available valve orifice area greater — dr,a than A V1 K derated coefficient of discharge of initially analysed valve — dr,1 K flow resistance coefficient of complete pipe network in terms of reference area, A — R F K flow resistance coefficient at the transition between critical and subcritical flow — RC K overall flow resistance coefficient of pipe network, downstream of pressure relief — Rd valve K overall flow resistance coefficient of pipe network, upstream of pressure relief valve — Ru K total flow resistance coefficient of a series or parallel pipe network — SUM 3 1) K experimentally determined flow rate through a pipe element or device m /h/bar V L latent heat of vaporization of cryogenic liquid at relieving conditions kJ/kg 1) L latent heat of vaporization of cryogenic liquid at a pressure of 1,013 bar kJ/kg a L’ enthalpy-to-volume expansion ratio for critical or all-gas fluid flow conditions kJ/kg M molar mass kg/kmol N normal evaporation rate (NER) %/day P relieving pressure, inner vessel in absolute pressure bar P pressure, safety relief valve outlet in absolute pressure bar b P pressure at relief valve outlet for a downstream built-up backpressure of 10 % bar b10 P pressure at pipe network exit in absolute pressure bar exit P pressure, safety relief valve inlet in absolute pressure bar
P pressure relief valve set pressure in absolute pressure bar S Q mass flow rate kg/h m Q mass flow rate of a relief valve within a given pipe network kg/h ma Q mass flow rate due to the normal evaporation rate kg/h mNER R universal gas constant J/(mol·K) T relieving temperature, inner vessel K T maximum external ambient temperature, non-fire condition K a T temperature at relief valve outlet K b,Pb T temperature at relief valve outlet for a downstream built-up backpressure of 10 % K b10 T external temperature for a given condition K e 5 2 1) bar = 0,1 MPa = 10 Pa; 1 MPa = 1 N/mm .
SIST EN ISO 21013-3:2026 ISO 21013-3:2026(en) T temperature at pipe network exit K exit,Pb T external temperature, fire condition K f T temperature, safety relief valve inlet K
T temperature of fluid at a given flow start location along the pipe network K n T saturation temperature of fluid at a pressure of 1 bar K sat T temperature of fluid at a given location x along the pipe network K
U heat transfer coefficient of insulating material, normal vacuum, non-fire condition W/ (m ·K)
U overall convective heat transfer coefficient of ambient air vaporizer W/ (m ·K)
U heat transfer coefficient of insulating material with air or gaseous lading, non-fire W/ (m ·K)
condition
U heat transfer coefficient, air or nitrogen condensation, loss of vacuum, non-fire W/m 3a condition
U heat transfer coefficient of insulating material with air or gaseous lading, fire W/ (m ·K)
condition
U heat transfer coefficient, air or nitrogen condensation, loss of vacuum, fire condition W/m 5a
U overall heat transfer coefficient of a pipe network for given temperature conditions W/ (m ·K) p W total heat transfer rate for specified conditions Watt [W] T W total heat transfer rate under normal operation Watt [W] T1 W total NER heat transfer rate under normal operation Watt [W] T1NER W total heat transfer rate under normal operation, including pressure build-up device Watt [W] T2 W total NER heat transfer rate under normal operation, including pressure build-up Watt [W] T2NER device W total heat transfer rate, loss of vacuum, insulation in place, non-fire condition, Watt [W] T3 T ˃ 75 K sat W total heat transfer rate, loss of vacuum, insulation in place, non-fire condition, Watt [W] T3a T ≤ 75 K sat W total heat transfer rate, loss of vacuum, insulation in place, fire condition, T ˃ 75 K Watt [W] T5 sat W total heat transfer rate, loss of vacuum, insulation in place, fire condition, T ≤ 75 K Watt [W] T5a sat W total heat transfer rate, loss of vacuum, insulation not in place, fire condition Watt [W] T6 W heat transfer rate through insulation system, normal vacuum, non-fire condition Watt [W]
W heat transfer rate through pressure build-up device, fully open regulator Watt [W]
W heat transfer rate through insulation system, loss of vacuum, non-fire condition Watt [W]
W heat transfer rate through air or nitrogen condensation, loss of vacuum, Watt [W] 3a non-fire condition W heat transfer rate through interspace supports and piping Watt [W]
SIST EN ISO 21013-3:2026 ISO 21013-3:2026(en) W heat transfer rate through vessel walls, insulation in place, fire condition Watt [W]
W heat transfer rate through air or nitrogen condensation, loss of vacuum, Watt [W] 5a fire condition W heat transfer rate through vessel walls, insulation not in place, fire condition Watt [W]
X number of insulation layers — Y heat transfer rate U or U — 3a 5a Z compressibility factor at pressure P and temperature T — i i i a c constant pressure specific heat capacity at the average of T and T kJ/(kg·K) p n e e nominal insulating material thickness, in normal vacuum, non-fire condition m
e minimum insulating material thickness, considering loss of vacuum, non-fire m
condition e insulating material thickness remaining in place during fire conditions m
f pipe flow friction coefficient — T h enthalpy of fluid at conditions of ν kJ/kg h specific enthalpy at relief valve inlet and outlet kJ/kg r k mean thermal conductivity of insulating material, normal vacuum, W/(m·K)
non-fire condition k mean thermal conductivity of insulating material with air or gaseous lading, W/(m·K)
non-fire condition k mean thermal conductivity of an individual support or pipe, between T and T W/(m·K) n a l length, pipe element m l length of support or pipe in vacuum interspace m n m maximum mass capacity of vessel kg max r pipe elbow transition radius m x lengthwise location along a pipe network m φ — PP P exit exit pressure ratio 1 P P κ isentropic exponent — λ subcritical flow coefficient —
ν specific volume of critical or all-gas fluid at a given temperature at pressure P m /kg
ν specific volume at relief valve outlet for a downstream built-up backpressure of 10 % m /kg b10
ν specific volume at pressure relief valve outlet, evaluated at h and a trial value of P m /kg b,Pb r b
SIST EN ISO 21013-3:2026 ISO 21013-3:2026(en)
ν average downstream specific volume, for a downstream built-up backpressure m /kg d10 of 10 %
ν maximum average downstream specific volume, as per desired backpressure limit m /kg dmax
ν specific volume at pipe network exit for a downstream built-up backpressure of 10 % m /kg exit10
ν specific volume at pipe network exit, evaluated at P and T m /kg exit,Pb exit exit,Pb
ν specific volume of saturated gas at relieving pressure, P m /kg g
ν specific volume of saturated gas at a pressure of 1,013 bar m /kg ga
ν specific volume, safety relief valve inlet m /kg
ν specific volume of saturated liquid at relieving pressure, P m /kg
ν specific volume of saturated liquid at a pressure of 1,013 bar m /kg la
ν average specific volume of flowing fluid upstream of pressure relief valve inlet m /kg u 3/2 1/2 ψ expression for determining Q and T for critical or gas-full-vessel fluid flow m ·kg / m conditions kJ
4 Calculation of the total quantity of heat transferred per unit time from the hot wall
(outer jacket) to the cold wall (inner vessel)
4.1 General
P (in bar absolute) is the actual relieving pressure inside the vessel which is used for calculating the required mass flow through pressure relief devices.
T (in K) is the maximum ambient temperature for conditions other than fire (as specified, for example, by a a regulation or standard).
T (in K) is the external environment temperature under fire conditions which is taken to be 922 K in this f document.
T (in K) is the relieving temperature in the vessel to be taken into account. a) For subcritical fluids, T is the saturation temperature of the liquid at pressure P. b) For critical or supercritical fluids, T is calculated from 5.2.
SIST EN ISO 21013-3:2026 ISO 21013-3:2026(en)
4.2 Under conditions other than fire
4.2.1 Vacuum-insulated vessels under normal vacuum
W is the quantity of heat transferred per unit time (in watts) by heat leak through the insulation system, as
per Formula (1).
WUAT T (1)
11 a where U is the overall heat transfer coefficient of the insulating material under normal vacuum, in W/
k
(m ⋅K): U = ;
e
k is the mean thermal conductivity of the insulating material under normal vacuum, between T
and T , in W/(m⋅K); a e is the nominal insulating material thickness, in m;
A is the arithmetic mean of the inner and outer surface areas of the vessel insulating material, in
m ; T is the relieving temperature, inner vessel; T is the maximum external ambient temperature, non-fire condition. a
4.2.2 Pressure build-up device
W is the quantity of heat transferred per unit time (in watts) by the pressure build-up device circuit with
the regulator fully open. W is determined from the type (ambient air, water or steam, electrical, etc.) and
design of the pressure build-up device circuit. For example, in the case of an ambient air vaporizer, W is
calculated using Formula (2):
WUAT T (2)
22 2 a where
U is the overall convective heat transfer coefficient of the ambient air vaporizer, in W/(m ⋅K);
A is the external heat transfer surface area of the vaporizer, in m ;
T is the relieving temperature, inner vessel; T is the maximum external ambient temperature, non-fire condition. a As a first approximation, Formula (3) or Formula (4) may be used:
UT TT19 000 W/mfor 75K (3)
2 a
UT TT2 850 W/mfor 75K (4)
2 a U shall not be lower than the values proposed but manufacturers can use greater values if their design
requires.
SIST EN ISO 21013-3:2026 ISO 21013-3:2026(en)
4.2.3 Vacuum-insulated vessels in the case of loss of vacuum and non-vacuum insulated vessels
W is the quantity of heat transferred per unit time (in watts) by heat leak through the insulating material. It
is calculated using Formula (5):
WUAT T (5)
33 a where U is the overall heat transfer coefficient of the insulating material when saturated with
gaseous lading or air at atmospheric pressure, whichever is greater, in W/(m ·K); T is the relieving temperature, inner vessel; T is the maximum external ambient temperature, non-fire condition; a A is the arithmetic mean of inner and outer surface areas of vessel insulating material. If the insulation is fully effective for conduction, convection, and radiation heat transfer at 328 K, U may be
calculated using Formula (6):
k
U = (6)
e
where k is the mean thermal conductivity of the insulating material saturated with gaseous lading
or air at atmospheric pressure, whichever provides the greater coefficient, between T and T , in W/(m·K) (values of k for gases are listed in Table 2); a 3 e is the minimum insulating material thickness taking into account the manufacturing
tolerances or effects of sudden loss of vacuum, in m.
NOTE This formula might not be applicable at temperatures below 75 K with a small thickness of insulating material as the maximum heat transfer coefficient would be given by air condensation. Vacuum space, gas space, or space occupied by the deteriorated insulation shall not be included in the thickness of the insulation. The effectiveness of these spaces or deteriorated insulation in reducing conduction, convection, or radiation heat transfer may be evaluated separately and included in the overall heat transfer coefficient, U , using methods found in published heat transfer literature. Deterioration of the
insulation can be caused by the following:
- moisture condensation;
- air condensation;
- increase in the density of the insulation due to a sudden loss of vacuum.
SIST EN ISO 21013-3:2026 ISO 21013-3:2026(en) Table 2 — Thermal conductivity for refrigerated (cryogenic) fluids at the mean temperature between saturation and 328 K (k) and 922 K (k) at 1 bar 3 5 k k 3 5 Fluid [W/(m·K)] [W/(m·K)] Air 0,019 0,043 Argon 0,013 0,027 Carbon dioxide 0,017 0,039 Carbon monoxide 0,020 0,039 Helium 0,104 0,211 Hydrogen 0,116 0,217 Methane 0,024 0,074 Neon 0,034 0,067 Nitrogen 0,019 0,040 Oxygen 0,019 0,043 Krypton 0,007 0,015 Xenon 0,005 0,009 Ethane 0,016 0,064 Trifluoromethane 0,012 0,027 Ethylene (ethene) 0,015 0,056 Nitrous oxide 0,014 0,038 Values are for reference only. Use fluid properties given in current databases, such as from NIST.
4.2.4 Supports and piping
W is the quantity of heat transferred per unit time (in watts) by supports and piping located in the
interspace, given by Formula (7):
WwwwwT T (7)
41 23 na where w is the heat leak per degree K contributed by one of the supports or the pipes, in W/K, given by n Formula (8):
A n wk (8)
nn
n where k is the mean thermal conductivity of the support or pipe material between T and T , in n a W/(m⋅K);
A is the support or pipe section area, in m ; n l is the support or pipe length in the vacuum interspace, in m. n
SIST EN ISO 21013-3:2026 ISO 21013-3:2026(en)
4.3 Under fire conditions
4.3.1 Insulation system remains fully or partially in place during fire conditions
W is the quantity of heat transferred per unit time (in watts) by heat leak through the vessel walls, given by
Formula (9):
08, 2 WT26, 922 UA (9)
where if the insulation is fully effective for conduction, convection, and radiation heat transfer for an external temperature of 922 K, U may be calculated using Formula (10):
k
U = (10)
e
where U is the overall heat transfer coefficient of the container-insulating material when saturated
with gaseous lading or air at atmospheric pressure, whichever is greater, in W/(m ·K); k is the mean thermal conductivity of the insulating material saturated with gaseous lading
or air at atmospheric pressure, whichever provides the greater coefficient, between T and 922 K, in W/(m·K) (values of k for gases are listed in Table 2);
e is the thickness of the insulating material remaining in place during fire conditions, in m;
A is the arithmetic mean of the inner and outer surface areas of the insulating material
remaining in place during fire conditions, in m ; T is the relieving temperature, inner vessel.
Vacuum space, gas space, or space occupied by the deteriorated insulation shall not be included in the thickness of the insulation. The effectiveness of these spaces or deteriorated insulation in reducing conduction, convection, or radiation heat transfer may be evaluated separately and included in the overall heat transfer coefficient, U , using methods found in published heat transfer literature. Deterioration of the
insulation can be caused by the following:
- moisture condensation;
- air condensation;
- increase in the density of the insulation due to a sudden loss of vacuum;
- degradation due to heat.
If the outer jacket remains in place during fire conditions, but the insulating material is entirely destroyed, U is equal to the overall heat transfer coefficient with gaseous lading or air at atmospheric pressure in the
space between the outer jacket and the inner vessel, whichever provides the greater coefficient, between T and 922 K. A is equal to the mean surface area of the interspace.
4.3.2 Insulation system does not remain in place during fire conditions
W is the quantity of heat transferred per unit time (in watts) by heat leak through the vessel walls,
calculated using Formula (11):
40,82 WA71, 10 (11) 6 i
where A is the total outside surface area of the inner vessel, in m .
SIST EN ISO 21013-3:2026 ISO 21013-3:2026(en) The heat transferred by supports and piping located in the interspace can be neglected in this case.
4.4 Air or nitrogen condensation
4.4.1 General
The air or nitrogen condensation case for the loss of vacuum condition shall be considered for fluids with a saturation temperature below 75 K at 1 bar.
...