Abstract

This document specifies general interface requirements for small-bore connectors that form part of a medical device or accessory that conveys liquids or gases to a patient.
This document also identifies the applications for which these small-bore connectors are intended to be used, which include, but are not limited to:
—     respiratory;
—     enteral;
—     limb cuff inflation;
—     neural;
—     intravascular or hypodermic;
—     other use cases utilizing an ISO 80369-7 small-bore connector.
This document provides the methodology to assess non-interconnectable characteristics of small-bore connectors based on their inherent design in order to reduce the risk of misconnections between medical devices or between accessories for different applications as specified in this document as well as those that can be developed under future parts of the ISO and IEC 80369 series.
NOTE            Clause A.2 contains guidance or rationale for this Clause.

Status
Published
Public Enquiry End Date
12-Sep-2023
Publication Date
29-Jun-2026
Technical Committee
VAZ - Healthcare
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
11-Jun-2026
Due Date
16-Aug-2026
Completion Date
30-Jun-2026

Buy Documents

Standard

SIST EN ISO 80369-1:2026 - BARVE

English language (55 pages)
Preview
Preview
e-Library read for
×1 day

SIST EN ISO 80369-1:2026 is the Slovenian adoption of EN ISO 80369-1:2026, which adopts ISO 80369-1:2025. It sets general interface requirements for small-bore connectors used in medical devices and accessories that convey liquids or gases to a patient, and it explains how to assess whether connector designs are non-interconnectable. The document is written for manufacturers, designers, risk managers, and evaluators who need to prevent misconnections between different healthcare applications.

What does SIST EN ISO 80369-1:2026 specify?

SIST EN ISO 80369-1:2026 specifies the general requirements for small-bore connectors used in healthcare applications and the method for checking whether different connector types cannot interconnect. Its role is to reduce the risk of wrong-route delivery of liquids and gases by keeping connector families separate by application.

The scope covers connectors that are part of a medical device or accessory and identifies the intended application areas, including respiratory, enteral, limb cuff inflation, neural, intravascular or hypodermic, and other use cases that use an ISO 80369-7 small-bore connector. Detailed connector designs for those application areas are handled in the relevant parts of the ISO and IEC 80369 series.

AnnexWhat it covers
Annex ARationale for the requirements
Annex BProcess for demonstrating non-interconnectable characteristics
Annex CAssessment procedures for new small-bore connector designs
Annex DApplications of small-bore connectors and the relevant standard
Annex EDesign assessment summary
Annex FReference to the IMDRF essential principles
Annex GAlphabetized index of defined terms
Annex ZARelationship to Regulation (EU) 2017/745

What are the key requirements of SIST EN ISO 80369-1:2026?

SIST EN ISO 80369-1:2026 requires connector families for different application categories to be non-interconnectable unless the series explicitly says otherwise. In practice, this means a connector intended for one clinical use must not be able to make an unintended connection with a connector for another use.

Non-interconnectable performance

Clause 4 sets the core rule: small-bore connectors for each application category shall be non-interconnectable with connectors from every other category. Conformity is checked by objective evidence showing that the Annex B acceptability criteria are met. The standard also states that using a connector specified by the ISO and IEC 80369 series is objective evidence, and it treats dimensional conformity together with a modulus of elasticity greater than 700 MPa as sufficient evidence of non-interconnectable characteristics.

Application-specific connector families

Clause 5 ties each clinical application to its corresponding connector standard. Respiratory applications use ISO 80369-2, enteral applications use ISO 80369-3, limb cuff inflation uses IEC 80369-5, neural applications use ISO 80369-6, and intravascular or hypodermic applications use ISO 80369-7. For a buyer or engineer, this is the first selection step: choose the connector family that matches the intended clinical use.

New clinical applications

Clause 5.1 allows new connector designs for applications not covered by the listed application clauses, but they must meet Annex C. Annex C requires a concept proposal, justification based on risk and evidence, design requirements, a user interface specification, engineering analysis, verification, and validation. This matters because a new connector family is not just a geometry change - it has to be justified, tested, and reviewed before inclusion in the series.

Other uses of ISO 80369-7 connectors

Clause 5.7 allows use of an ISO 80369-7 small-bore connector for other use cases outside the listed application categories, but only if the risk of unintended connection with reasonably foreseeable devices is assessed with ISO 14971 and found acceptable. The manufacturer must consider harm in both flow directions, apply marking in accordance with ISO 20417:2021, and, where safety information is used as the risk control, include a specific warning in the instructions for use. This clause gives a controlled route for legacy or special uses, but it also requires clear labeling and risk documentation.

Alternative connector designs

Clause 6 allows alternative small-bore connector designs, but only if they are evaluated against Annex B, do not create unacceptable risk, and are assessed using the engineering analysis of Clause C.3. Such devices or accessories must be marked in accordance with ISO 20417:2021 and, if necessary, carry a warning in the instructions for use. The clause also restricts claims: an alternative design may claim conformance with this document, but not generic conformity with the whole ISO and IEC 80369 series.

How non-interconnectability is tested

Annex B gives the test route. It starts with dimensional analysis of contactable surfaces and possible insertion paths, then uses misconnection testing when dimensions show a potential misconnection, and adds risk analysis if physical misconnection is possible. If a connector still misconnects, a leak test checks whether the misconnection can pass fluid. For practice, this annex is the evidence package behind design decisions, supplier qualification, and technical files.

What terms does SIST EN ISO 80369-1:2026 define?

  • Accessory - an additional part used with a medical device to achieve, adapt, facilitate, improve, or integrate its function.
  • Application - a specific healthcare use for the connector family.
  • Connector - one mating half of a medical device designed to join a conduit that carries liquids or gases.
  • Misconnection - a connection between two different connector types.
  • Unintended connection - a connection between the same connector type when it is used for different use cases.
  • Non-interconnectable - having design features that prevent different connectors from making a connection.
  • Small-bore - having an inner-fluid pathway with a diameter less than 8.5 mm.
  • Luer connector - a small-bore connector with a conical mating surface and a 6 % taper intended for intravascular or hypodermic use.

Who uses SIST EN ISO 80369-1:2026?

SIST EN ISO 80369-1:2026 is used by medical device and accessory manufacturers, connector designers, and suppliers who need to specify or verify small-bore connector systems. It is also used by risk management, usability, quality, and regulatory teams because it relies on ISO 14971 for risk management, IEC 62366-1 for usability engineering, and ISO 20417 for marking and user information.

National standards bodies and standards committees use it when reviewing new connector proposals in Annex C. Hospitals, buyers, and technical evaluators use it to check whether a device family matches the correct connector standard for its clinical application.

What changed in SIST EN ISO 80369-1:2026 from the previous edition?

SIST EN ISO 80369-1:2026 supersedes SIST EN ISO 80369-1:2019 and, as the European adoption, supersedes EN ISO 80369-1:2018. The third edition is technically revised.

The changes stated in the foreword are:

  • updated normative references
  • update according to ISO/IEC Directives, Part 2
  • addition of respiratory applications
  • addition of an interference test part and definitions for misconnection and unintended connection
  • deletion of the former materials requirements clause, with those requirements moved to the individual connector parts
  • revision of Annex B, including a clearer process description and a new dimensional-analysis case
  • replacement of the original Annex C with a normative reference to parts of ISO 20417
  • addition of Annex E summarizing the design assessments of the connector families
  • replacement of Annex F content with references to relevant IMDRF essential principles and labelling guidance
  • extension of ISO 80369-7 use to other medical devices and accessories where the risk is acceptable

Which standards are used with SIST EN ISO 80369-1:2026?

SIST EN ISO 80369-1:2026 relies on several normative references. ISO 14971 provides the risk management framework, IEC 62366-1 provides usability engineering methods, and ISO 20417 provides the rules for information supplied by the manufacturer and marking.

The application-specific connector parts are also essential companions:

  • ISO 80369-2 for respiratory applications
  • ISO 80369-3 for enteral applications
  • IEC 80369-5 for limb cuff inflation applications
  • ISO 80369-6 for neural applications
  • ISO 80369-7 for intravascular or hypodermic applications

What does the SIST EN ISO 80369-1:2026 document contain?

The document contains a short core body and several annexes that support design, testing, and regulatory use. Clause 4 states the non-interconnectable requirement, Clause 5 maps connector families to clinical applications, and Clause 6 covers alternative connector designs.

Annex B contains the detailed evaluation route, including dimensional analysis, misconnection testing, and leak testing. Annex C sets out the proposal, engineering analysis, verification, validation, and review process for a new connector design. Annex D lists example medical device uses for each application category, Annex E summarizes assessed connector interactions, and Annex F links the document to IMDRF essential principles. Annex ZA gives the European regulatory relationship to Regulation (EU) 2017/745.

Relations

Effective Date
01-Jul-2026

Buy Documents

Standard

SIST EN ISO 80369-1:2026 - BARVE

English language (55 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

SIST EN ISO 80369-1:2026 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements (ISO 80369-1:2025)". This standard covers: This document specifies general interface requirements for small-bore connectors that form part of a medical device or accessory that conveys liquids or gases to a patient. This document also identifies the applications for which these small-bore connectors are intended to be used, which include, but are not limited to: —     respiratory; —     enteral; —     limb cuff inflation; —     neural; —     intravascular or hypodermic; —     other use cases utilizing an ISO 80369-7 small-bore connector. This document provides the methodology to assess non-interconnectable characteristics of small-bore connectors based on their inherent design in order to reduce the risk of misconnections between medical devices or between accessories for different applications as specified in this document as well as those that can be developed under future parts of the ISO and IEC 80369 series. NOTE            Clause A.2 contains guidance or rationale for this Clause.

This document specifies general interface requirements for small-bore connectors that form part of a medical device or accessory that conveys liquids or gases to a patient. This document also identifies the applications for which these small-bore connectors are intended to be used, which include, but are not limited to: —     respiratory; —     enteral; —     limb cuff inflation; —     neural; —     intravascular or hypodermic; —     other use cases utilizing an ISO 80369-7 small-bore connector. This document provides the methodology to assess non-interconnectable characteristics of small-bore connectors based on their inherent design in order to reduce the risk of misconnections between medical devices or between accessories for different applications as specified in this document as well as those that can be developed under future parts of the ISO and IEC 80369 series. NOTE            Clause A.2 contains guidance or rationale for this Clause.

SIST EN ISO 80369-1:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.10 - Anaesthetic, respiratory and reanimation equipment; 11.040.20 - Transfusion, infusion and injection equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

SIST EN ISO 80369-1:2026 has the following relationships with other standards: It is inter standard links to SIST EN ISO 80369-1:2019. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

SIST EN ISO 80369-1:2026 is associated with the following European legislation: EU Directives/Regulations: 2017/745; Standardization Mandates: M/575, M/575 Amd 1, M/575 Amd 2. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.

SIST EN ISO 80369-1:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 accessory
  • 3.2 application
  • 3.3 cone
  • 3.4 connection
  • 3.5 connector
  • 3.6 contactable surface
  • 3.7 information for safety
  • 3.8 instructions for use
  • 3.9 interference test part
  • 3.10 least material condition
  • 3.11 Luer connector
  • 3.12 marking
  • 3.13 maximum material condition
  • 3.14 misconnection
  • 3.15 non-interconnectable
  • 3.16 patient
  • 3.17 small-bore
  • 3.18 socket
  • 3.19 test method
  • 3.20 unintended connection
  • 4 Small-bore connector non-interconnectable requirement
  • 5 Small-bore connectors for clinical applications
  • 5.1 Small-bore connectors for new clinical applications
  • 5.2 Small-bore connectors for respiratory applications
  • 5.3 Small-bore connectors for enteral applications
  • 5.4 Small-bore connectors for limb cuff inflation applications
  • 5.5 Small-bore connectors for neural applications
  • 5.6 Small-bore connectors for intravascular and hypodermic applications
  • 5.7 Other use cases utilizing an ISO 80369-7 small-bore connector
  • 6 Alternative small-bore connectors
  • Annex A
  • Annex B
  • Annex C
  • Annex D
  • Annex E
  • Annex F
  • Annex G
  • Bibliography

Introduction

In the 1990s, concern grew regarding the proliferation of medical devices fitted with Luer connectors [1] as specified in the ISO 594 series and the reports of patient death or injury arising from unintended connections that resulted in the inappropriate delivery of fluids and gases via incorrect routes. In addition to clinical and workplace protocols and warnings, attention was turned to engineering solutions to reduce the probability of wrong route administration of liquids and gases. Concerns regarding the use of Luer connectors with enteral feeding tubes and gas sampling and gas delivery systems were raised with CEN Bureau Technique (CEN/BT) and the European Commission. In November 1997, the newly created CEN Healthcare Forum (CHeF) steering group set up a Forum Task Group (FTG) to consider the problem.

[9] The FTG produced CEN Report, CEN/CR 13825, in which they concluded that there is a problem arising from the use of a single connector design to several different applications. In a coronary care unit, there could be as many as 40 Luer connectors on the medical devices used with a single patient until the use of connectors defined in the ISO and IEC 80369 documents started to be established. Therefore, it is not surprising that unintended connections were made.

Medical devices have, for many years, followed the established principle of “safety under single fault conditions.” Simply stated, this means that a single fault should not result in an unacceptable risk. This [6] principle is embodied in the requirements of numerous medical device standards. Extending this principle to the use of Luer connectors (i.e. that an unintended connection should not result in an unacceptable risk to a patient) the FTG recommended that the Luer connector should be restricted to medical devices intended to be connected to the vascular system or a hypodermic syringe. In addition, the FTG recommended that new designs of small-bore connectors should be developed for other applications, and these should be noninterconnectable with Luer connectors and each other. [10] International medical device regulators forum (IMDRF), GRRP N47:2024, Essential Principle 5.1.3, addresses this type of problem:

Risk control measures adopted by manufacturers for the design and manufacture of the medical device and IVD medical device should conform to safety principles, taking account of the generally acknowledged state of the art. When risk reduction is required, manufacturers should control risks so that the residual risk associated with each hazard as well as the overall residual risk is judged acceptable. In selecting the most appropriate solutions, manufacturers should, in the following order of priority:

a) eliminate or appropriately reduce risks through safe design and manufacture; b) where appropriate, take adequate protection measures, including alarms if necessary, in relation to risks that cannot be eliminated; c) provide information for safety (warnings/precautions/contra-indications) and, where appropriate, training to users.

It is understood that small-bore connector systems cannot be designed to overcome all chances of unintended connections with the potential for wrong route administration or to eliminate deliberate misuse. With these application-specific connectors now available, the risk of unintended connections and wrong route administration is greatly reduced thereby improving patient safety. Introduction of medical devices and accessories utilizing these small-bore connectors is progressing albeit slowly. The risks associated with unintended connections and subsequent wrong route administration of liquids and gases cannot be fully assessed until these small-bore connectors are part of a medical device or accessory. Therefore, the intended applications specified are recommendations. It is expected that particular medical device standards will reference the connectors from the relevant parts of the ISO and IEC 80369 series if considered appropriate.

This document contains the general requirements to reduce connections between small-bore connectors used in different applications as well as specifying those applications.

SIST EN ISO 80369-1:2026 ISO 80369-1:2025(en) It specifies the general requirements and test methods for assessing the non-interconnectable characteristics of small-bore connectors within the ISO and IEC 80369 series.

The Luer connector as originally defined in the withdrawn ISO 594 series has been widely used on many medical devices and accessories and in a wide range of clinical applications for many years. The clinical applications that present the highest risk to a patient from wrong route administration of liquids and gases have been identified and are those included in the application parts of the ISO and IEC 80369 series. ISO 80369-7, which replaces the ISO 594 series (i.e. the Luer connector), is intended for use with intravascular or hypodermic applications.

However, there are currently medical devices and accessories which incorporate a Luer connector, but do not fall into any of the applications specified by the ISO and IEC 80369 series. There are also some medical devices and accessories within the applications of the ISO and IEC 80369 series applications that incorporate a Luer connector. Those that present no unacceptable risk to the patient from an unintended connection to a medical device or accessory within intravascular or hypodermic application are suitable for consideration for the use of the Luer connector as specified in ISO 80369-7.

ISO 80369-20 specifies the common test methods for assessing the basic performance requirements specified in ISO 80369-2, ISO 80369-3, IEC 80369-5, ISO 80369-6 and ISO 80369-7 for small-bore connectors. ISO 80369-2, ISO 80369-3, IEC 80369-5, ISO 80369-6 and ISO 80369-7 specify the material requirements, dimensional requirements for the interfaces of the connectors and the specific performance requirements for assessing the interconnectability of the mating connectors. The designs and dimensions of small-bore connectors specified in ISO 80369-2, ISO 80369-3, IEC 80369-5, ISO 80369-6 and ISO 80369-7 have been successfully assessed according to the requirements of this document (i.e. have been proven to be acceptable with regard to the risk of misconnection with the other connectors of this series).

The risks of changing to these new small-bore connectors should be assessed before these small-bore connectors are incorporated into medical devices as they will when the relevant particular medical device standards are revised.

In this document, the conjunctive “or” is used as an “inclusive or” so a statement is true if any combination of the conditions is true.

In this document the following verbal forms are used.

  • “Shall” indicates requirements.
  • “Should” indicates recommendations.
  • “May” indicates permissions.
  • "Can" indicates possibility or capability.

This document uses italic type to distinguish defined terms from the rest of the text. It is important for the correct understanding of this document that those defined terms are identifiable throughout the text of this document. A list of the defined terms used in italics in this document is given in Annex G. Requirements in this document have been broken down so that each requirement is clearly delineated and listed individually. This has been done to support the common practice of automatic tracking of requirements and automatic verification of the requirements of this document. Annex A contains guidance and rationale on specific subclauses in this document.

SIST EN ISO 80369-1:2026

SIST EN ISO 80369-1:2026 International Standard ISO 80369-1:2025(en) Small-bore connectors for liquids and gases in healthcare applications — Part 1:

General requirements

1 Scope

NOTE Clause A.2 contains guidance or rationale for this Clause. This document specifies general interface requirements for small-bore connectors that form part of a medical device or accessory that conveys liquids or gases to a patient. This document also identifies the applications for which these small-bore connectors are intended to be used, which include, but are not limited to:

  • respiratory;
  • enteral;
  • limb cuff inflation;
  • neural;
  • intravascular or hypodermic;
  • other use cases utilizing an ISO 80369-7 small-bore connector. This document provides the methodology to assess non-interconnectable characteristics of small-bore connectors based on their inherent design in order to reduce the risk of misconnections between medical devices or between accessories for different applications as specified in this document as well as those that can be developed under future parts of the ISO and IEC 80369 series.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 14971, Medical devices — Application of risk management to medical devices ISO 20417:2021, Medical devices — Information to be supplied by the manufacturer ISO 80369-2, Small-bore connectors for liquids and gases in healthcare applications — Part 2: Connectors for respiratory applications ISO 80369-3, Small-bore connectors for liquids and gases in healthcare applications — Part 3: Connectors for enteral applications IEC 80369-5, Small-bore connectors for liquids and gases in healthcare applications — Part 5: Connectors for limb cuff inflation applications ISO 80369-6, Small bore connectors for liquids and gases in healthcare applications — Part 6: Connectors for neuraxial applications

SIST EN ISO 80369-1:2026 ISO 80369-1:2025(en) ISO 80369-7, Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for intravascular or hypodermic applications IEC 62366-1:2015+AMD1:2020, Medical devices — Part 1: Application of usability engineering to medical devices

3 Terms and definitions

For the purposes of this document, the terms and definitions given in ISO 14971, IEC 62366-1 and the following apply.

ISO and IEC maintain terminology databases for use in standardization at the following addresses:

- ISO Online browsing platform: available at https:// www .iso .org/ obp - IEC Electropedia: available at https:// www .electropedia .org/ NOTE For convenience, the sources of all defined terms that appear in italics in this document are given in Annex G.

3.1 accessory

additional part(s) for use with a medical device in order to:

  • achieve the intended use;
  • adapt it to some special use;
  • facilitate its use;
  • enhance its performance; or
  • enable its functions to be integrated with those of other medical devices [SOURCE: IEC 60601-1:2005+AMD1:2012+AMD2:2020, 3.3, modified — Replaced “equipment” with “a medical device”.]

3.2 application

specific healthcare use Note 1 to entry: Annex D lists examples of uses of the applications of small-bore (3.17) connectors (3.5).

3.3 cone

connector (3.5) with external sealing surface Note 1 to entry: The sealing surface need not be conical.

Note 2 to entry: This type of connector (3.5) was previously referred to as male.

3.4 connection

union or joining of two connectors (3.5)

3.5 connector

part of a medical device, consisting of one of two mating halves and designed to join a conduit to convey liquids or gases

SIST EN ISO 80369-1:2026 ISO 80369-1:2025(en)

3.6 contactable surface

any surface on a connector (3.5) that has an interaction potential in which physical contact occurs with any other surface on an opposing connector Note 1 to entry: Contactable surfaces may include, but are not limited to, sealing surfaces as intended by design, crest geometry of external or internal threads, faces, shrouds, grips. These are surfaces on a connector that can possibly interact with another connector.

3.7 information for safety

information provided to the user or responsible organization as a risk control measure EXAMPLE 1 Warnings, precautions or contraindications.

EXAMPLE 2 Instructions for the use of a medical device or accessory (3.1) to prevent use error or avoid a hazardous situation.

EXAMPLE 3 Explanation of a safety feature of a medical device. Note 1 to entry: Information for safety can be found in any or all types of information supplied by the manufacturer. Note 2 to entry: Information for safety can be located on the display of a medical device. [SOURCE: ISO 20417:2021, 3.9]

3.8 instructions for use

portion of the accompanying information that is essential for the safe and effective use of a medical device or accessory (3.1) directed to the user of the medical device Note 1 to entry: For the purposes of this document, a user can be either a lay user or professional user with relevant specialized training.

[SOURCE: ISO 20417:2021, 3.11, modified — Deleted admitted term “package insert”; deleted Notes 2 to 5 to entry.]

3.9 interference test part

component that physically represents a small-bore (3.17) connector (3.5) or connector feature Note 1 to entry: An interference test part is used to evaluate whether or not a contactable surface (3.6) can misconnect.

3.10 least material condition

LMC condition in which a feature contains the least amount of material within the stated tolerance EXAMPLE Maximum hole diameter, minimum shaft diameter.

3.11 Luer connector

small-bore (3.17) connector (3.5) that contains a conical mating surface with a 6 % (Luer) taper intended for use in intravascular or hypodermic applications (3.2) of medical devices and related accessories (3.1) Note 1 to entry: A Luer connector can be either a Luer slip connector or a Luer lock connector. [SOURCE: ISO 80369-7:2021, 3.2, modified — Deleted Note 2 to entry.]

SIST EN ISO 80369-1:2026 ISO 80369-1:2025(en)

3.12 marking

information, in text or graphical format, durably affixed, printed, etched (or equivalent) to a medical device or accessory (3.1) Note 1 to entry: For the purposes of this document, the term marked is used to designate the corresponding act. Note 2 to entry: For the purposes of this document, marking is different from ‘direct marking’ as commonly described in unique device identification (UDI) standards and regulations. A UDI ‘direct marking’ is a type of marking. [SOURCE: ISO 20417:2021, 3.16, modified — Deleted Note 3 to entry.]

3.13 maximum material condition

MMC condition in which a feature contains the maximum amount of material within the stated tolerance EXAMPLE Minimum hole diameter, maximum shaft diameter.

3.14 misconnection

connection (3.4) between two connectors (3.5) of a different type

3.15 non-interconnectable

having characteristics which incorporate geometries or other features that prevent different connectors (3.5) from making a connection (3.4)

3.16 patient

person undergoing a medical, surgical or dental procedure [SOURCE: IEC 60601-1:2005+A1:2012, 3.76, modified — Replaced “living being (person or animal)” with “person”; deleted the note.]

3.17 small-bore

inner-fluid pathway of a connection (3.4) with a diameter less than 8,5 mm Note 1 to entry: For the purposes of this document, the 8,5 mm cone (3.3) and socket (3.18) specified in ISO 5356-1 is not considered a small-bore (3.17) connector (3.5).

3.18 socket

connector (3.5) with internal sealing surface Note 1 to entry: This type of connector (3.5) was previously referred to as female.

3.19 test method

definitive procedure for evaluating connectors (3.5) that produces a test result

3.20 unintended connection

connection (3.4) between two connectors (3.5) of the same type for different use cases

4 Small-bore connector non-interconnectable requirement

NOTE 1 A.2.3 contains guidance and rationale for this Clause.

SIST EN ISO 80369-1:2026 ISO 80369-1:2025(en) Small-bore connectors of each application category specified in this document shall be non-interconnectable with any of the small-bore connectors of every other application category unless otherwise indicated within the ISO and IEC 80369 series.

Check conformity by confirming that objective evidence demonstrates that the acceptability criteria specified in Annex B are met.

The use of a connector specified by the ISO and IEC 80369 series is considered as objective evidence that the acceptability criteria specified in Annex B are met.

NOTE 2 For the purpose of this document, dimensional conformity and modulus of elasticity conformity with the requirements of the various application parts of the ISO and IEC 80369 series is considered sufficient objective evidence of non-interconnectable characteristics.

5 Small-bore connectors for clinical applications

NOTE 1 A.2.4 contains guidance and rationale for this Clause.

NOTE 2 Annex D lists examples of the sort of medical devices or accessories for which the small-bore connectors within each application are intended.

NOTE 3 Annex E summarizes the assessments of the designs of the small-bore connectors of this series of documents. NOTE 4 This document has been prepared to address the relevant essential principles and labelling guidances of the International Regulators Forum (IMDRF) as indicated in Annex F.

5.1 Small-bore connectors for new clinical applications

Designs of small-bore connectors for applications other than those specified in 5.2 to 5.6, for inclusion in the ISO and IEC 80369 series, and used in medical devices or accessories intended for use with a patient, shall meet the requirements of Annex C.

Check conformity by application of Annex C.

5.2 Small-bore connectors for respiratory applications

Small-bore connectors intended to be used for connections in the respiratory application shall conform to a) ISO 80369-2, unless the use of these connectors creates an unacceptable risk for a specific medical device or accessory, or b) Clause 6.

Check conformity by inspection of the documentation demonstrating that the small-bore connector conforms to the requirements of ISO 80369-2 or, in case of alternative connectors as per Clause 6, by applying the performance tests of ISO 80369-2 using appropriate reference connectors for the alternative design.

5.3 Small-bore connectors for enteral applications

Small-bore connectors intended to be used for connections in the enteral application shall conform to a) ISO 80369-3, unless the use of these connectors creates an unacceptable risk for a specific medical device or accessory, or b) Clause 6.

Check conformity by inspection of the documentation demonstrating that the small-bore connector conforms to the requirements of ISO 80369-3 or, in case of alternative connectors as per Clause 6, by applying the performance tests of ISO 80369-3 using appropriate reference connectors for the alternative design.

SIST EN ISO 80369-1:2026 ISO 80369-1:2025(en)

5.4 Small-bore connectors for limb cuff inflation applications

Small-bore connectors intended to be used for connections in limb cuff inflation application shall conform to a) IEC 80369-5, unless the use of these connectors creates an unacceptable risk for a specific medical device or accessory, or b) Clause 6.

Check conformity by inspection of the documentation demonstrating that the small-bore connector conforms to the requirements of IEC 80369-5 or, in case of alternative connectors as per Clause 6, by applying the performance tests of IEC 80369-5 using appropriate reference connectors for the alternative design.

5.5 Small-bore connectors for neural applications

Small-bore connectors intended to be used for connections in neural application shall conform to a) ISO 80369-6, unless the use of these connectors creates an unacceptable risk for a specific medical device or accessory, or b) Clause 6.

...