SIST IEC 61010-3-043:2002
(Main)Safety requirements for electrical equipment for measurement, control and laboratory use - Part 3-043: Conformity verification report for IEC 61010-2-043:1997 - Particular requirements for dry heat sterilizers using either hot air or hot inert gas for treatment of medical materials and for laboratory processes
Safety requirements for electrical equipment for measurement, control and laboratory use - Part 3-043: Conformity verification report for IEC 61010-2-043:1997 - Particular requirements for dry heat sterilizers using either hot air or hot inert gas for treatment of medical materials and for laboratory processes
A technical report for use by test houses and other users. It assists them with determining and recording verification of conformity of the equipment under test with the requirements of IEC61010-2-043 (1997). If the paper version of the report is ordered, a diskette is provided with fill-in tables for recording the test results. If the report is downloaded directly from the web, it consists of two files: the technical report as a ZIP file and the fill-in tables as a revisable file.
Safety requirements for electrical equipment for measurement, control and laboratory use - Part 3-043: Conformity verification report for IEC 61010-2-043:1997 - Particular requirements for dry heat sterilizers using either hot air or hot inert gas for treatment of medical materials and for laboratory processes
General Information
Standards Content (Sample)
SLOVENSKI STANDARD
SIST IEC 61010-3-043:2002
01-maj-2002
Safety requirements for electrical equipment for measurement, control and
laboratory use - Part 3-043: Conformity verification report for IEC 61010-2-043:1997
- Particular requirements for dry heat sterilizers using either hot air or hot inert gas
for treatment of medical materials and for laboratory processes
Safety requirements for electrical equipment for measurement, control and laboratory
use - Part 3-043: Conformity verification report for IEC 61010-2-043:1997 - Particular
requirements for dry heat sterilizers using either hot air or hot inert gas for treatment of
medical materials and for laboratory processes
Ta slovenski standard je istoveten z: IEC/TR 61010-3-043
ICS:
11.080.10 Sterilizacijska oprema Sterilizing equipment
19.080 (OHNWULþQRLQHOHNWURQVNR Electrical and electronic
SUHVNXãDQMH testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
SIST IEC 61010-3-043:2002 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
---------------------- Page: 1 ----------------------
SIST IEC 61010-3-043:2002
---------------------- Page: 2 ----------------------
SIST IEC 61010-3-043:2002
TECHNICAL IEC
REPORT
TR 61010-3-043
First edition
2001-07
Safety requirements for electrical equipment
for measurement, control, and laboratory use –
Part 3-043:
Conformity verification report for
IEC 61010-2-043:1997 –
Particular requirements for dry heat sterilizers
using either hot air or hot inert gas for treatment
of medical materials and for
laboratory processes
Règles de sécurité pour appareils électriques de mesurage,
de régulation et de laboratoire –
Partie 3-043:
Rapport de vérificatoin de la conformité de
la CEI 61010-2-043:1997 –
Prescriptions particulières pour les stérilisateurs à chaleur
utilisant de l'air chaud ou un gaz inerte chaud pour
le traitement des matériels à usage médical et durant
les procédés de traitement de laboratoire
IEC 2001 Copyright - all rights reserved
No part of this publication may be reproduced or utilized in any form or by any means, electronic or
mechanical, including photocopying and microfilm, without permission in writing from the publisher.
International Electrotechnical Commission 3, rue de Varembé Geneva, Switzerland
Telefax: +41 22 919 0300 e-mail: inmail@iec.ch IEC web site http://www.iec.ch
Commission Electrotechnique Internationale
PRICE CODE
XA
International Electrotechnical Commission
For price, see current catalogue
---------------------- Page: 3 ----------------------
SIST IEC 61010-3-043:2002
– 2 – TR 61010-3-043 IEC:2001(E)
CONTENTS
FOREWORD . 3
Conformity verification report IEC 61010-2-043. 5
5 Marking and documentation .10
6 Protection against electric shock .14
7 Protection against mechanical hazards.21
8 Mechanical resistance to shock and impact .23
9 Equipment temperature limits and protection against the spread of fire .23
10 Resistance to heat .24
11 Protection against hazards from fluids .25
12 Protection against radiation, including laser sources, and against sonic
and ultrasonic pressure.26
13 Protection against liberated gases, explosion and implosion .26
14 Components .27
15 Protection by interlocks .29
16 Measuring circuits .29
Annex K…………. .30
Table 1 – Documents attached to this report . 7
Table 2 – Test equipment list. 8
Table 3 – List of components relied on for safety . 9
Form A.1 to Form A.29 .30 to 59
---------------------- Page: 4 ----------------------
SIST IEC 61010-3-043:2002
TR 61010-3-043 IEC:2001(E) – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
___________
SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT
FOR MEASUREMENT, CONTROL, AND LABORATORY USE –
Part 3-043: Conformity verification report for IEC 61010-2-043:1997–
Particular requirements for dry heat sterilizers using either hot air
or hot inert gas for treatment of medical materials
and for laboratory processes
FOREWORD
1) The IEC (International Electrotechnical Commission) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of the IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, the IEC publishes International Standards. Their preparation is
entrusted to technical committees; any IEC National Committee interested in the subject dealt with may
participate in this preparatory work. International, governmental and non-governmental organizations liaising
with the IEC also participate in this preparation. The IEC collaborates closely with the International Organization
for Standardization (ISO) in accordance with conditions determined by agreement between the two
organizations.
2) The formal decisions or agreements of the IEC on technical matters express, as nearly as possible, an
international consensus of opinion on the relevant subjects since each technical committee has representation
from all interested National Committees.
3) The documents produced have the form of recommendations for international use and are published in the form
of standards, technical specifications, technical reports or guides and they are accepted by the National
Committees in that sense.
4) In order to promote international unification, IEC National Committees undertake to apply IEC International
Standards transparently to the maximum extent possible in their national and regional standards. Any
divergence between the IEC Standard and the corresponding national or regional standard shall be clearly
indicated in the latter.
5) The IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with one of its standards.
6) Attention is drawn to the possibility that some of the elements of this International Standard may be the subject
of patent rights. The IEC shall not be held responsible for identifying any or all such patent rights.
The main task of IEC technical committees is to prepare International Standards. However, a
technical committee may propose the publication of a technical report when it has collected
data of a different kind from that which is normally published as an International Standard, for
example “state of the art”.
Technical reports do not necessarily have to be reviewed until the data they provide are
considered to be no longer valid or useful.
IEC 61010-3-043, which is a technical report, has been prepared by IEC technical com-
mittee 66: Safety of measuring, control, and laboratory equipment.
The text of this technical report is based on the following documents:
Committee draft Report on voting
66/223/CDV 66/230/RVC
Full information on the voting for the approval of this technical report can be found in the report
on voting indicated in the above table.
---------------------- Page: 5 ----------------------
SIST IEC 61010-3-043:2002
– 4 – TR 61010-3-043 IEC:2001(E)
This publication has been drafted in accordance with the ISO/IEC Directives, Part 3.
This report is a Technical Report and is of a purely informative nature and is therefore by itself
not to be regarded as an International Standard. It is for use by test houses and other users to
assist them with determining and recording verification of conformity of the equipment under
test with the requirements of:
IEC 61010-2-043: 1997
and
IEC 61010-1:1990 + amendment 1:1992 + amendment 2:1995
The protocol for completion of this report is contained in publication IEC 61010-3:1997.
The IEC sells read-only PDF files as a general rule. In the present instance, and quite
exceptionally, to enable the user to fill in the forms, a revisable file is included in a pocket
affixed to the back cover of this publication.
This file can also be downloaded from the Web as a PDF file. There is, however, at the end of
the document, a revisable file containing the forms. Please use the zip/unzip function.
A bilingual version will not be issued.
A French version may be issued.
The committee has decided that the contents of this publication will remain unchanged
until 2004. At this date, the publication will be
• reconfirmed;
• withdrawn;
• replaced by a revised edition, or
• amended.
---------------------- Page: 6 ----------------------
SIST IEC 61010-3-043:2002
TR 61010-3-043 IEC:2001(E) – 5 –
Conformity Verification Report
IEC 61010-2-043
Safety requirements for electrical equipment for measurement, control, and laboratory use –
Part 2-043: Particular requirements for dry heat sterilizers
using either hot air or hot inert gas for treatment of
medical materials, and for laboratory processes
Report reference No.:
Compiled by (+ signature).:
Approved by (+ signature).:
Date of issue.:
Testing organization.:
Address.:
Testing location .:
Applicant.:
Address.:
Standard.: IEC 61010-2-043:1997
and
IEC 61010-1:1990 + amendment 1:1992 + amendment 2:1995
Copyright blank test report.: This report has been prepared by IEC/TC 66, which retains
responsibility for any changes or corrections required.
Test procedure.:
Procedure deviation.:
Non-standard test method.:
Type of item tested.: Laboratory
Trademark. .:
Model/type reference.:
Manufacturer.:
Rating.:
Copy of rating plate:
---------------------- Page: 7 ----------------------
SIST IEC 61010-3-043:2002
– 6 – TR 61010-3-043 IEC:2001(E)
Description of equipment function.:
INSTALLATION/OVERVOLTAGE CATEGORY:
OLLUTION DEGREE
P :
Environmental rating: Standard Other (specify):
Equipment mobility: Portable Hand-held Floorstanding Fixed
Built in Benchmounted Other (specify):
Connection to mains supply: Permanent Detachable Non-detachable None
Operating conditions: Continuous Short-time Intermittent
Overall size of the equipment (Length x Width x Height):
Mass of the equipment (kg):
Marked degree of protection to IEC 60529: IP___ ___
Accessories and detachable parts included in the evaluation:
Options:
NOTE "(see form A.X)" refers to a form appended to the report.
---------------------- Page: 8 ----------------------
SIST IEC 61010-3-043:2002
TR 61010-3-043 IEC:2001(E) – 7 –
Table 1 – Documents attached to this report
Document No Document description Number of pages
---------------------- Page: 9 ----------------------
SIST IEC 61010-3-043:2002
– 8 – TR 61010-3-043 IEC:2001(E)
Table 2 – Test equipment list
Item Type Equipment Calibration date Comments
1)
No. Last Due
1)
or interval between calibrations.
---------------------- Page: 10 ----------------------
SIST IEC 61010-3-043:2002
TR 61010-3-043 IEC:2001(E) – 9 –
Table 3 – List of components relied on for safety
Unique component Application/Function Manufacturer and part number RATING Licence number,
reference or location file number or other
(note 1) (note 2)
(including drawing documentary evidence
reference if required) of acceptance
NOTE 1 List all manufacturers concerned.
NOTE 2 Electrical, mechanical, flammability, etc.
---------------------- Page: 11 ----------------------
SIST IEC 61010-3-043:2002
– 10 – TR 61010-3-043 IEC:2001(E)
Clause Requirement Result Comments
Subclause
5 Marking and documentation
5.1.1 General
Required equipment markings are:
a) visible:
– from the exterior
or
– after removing a cover
or
– opening a door
or
– after removal from a rack or panel
b) not put on parts which can be removed
by an OPERATOR
c) Letter symbols (IEC 60027) used
d) Graphic symbols
(IEC 61010-1, Table 1) used
5.1.2 Identification
Equipment is identified by:
– manufacturer's name or registered
trade mark
– model number, name or other means
5.1.3 Mains supply
Equipment is marked as follows:
a) nature of supply:
–a.c. RATED mains frequency or range of
frequencies
– d.c. with symbol 1
b) RATED supply voltage(s) or range
c) – maximum RATED power (W or VA)
or input current
If more than one voltage range:
– separate values marked
or
– values differ by less than
20 % (see Form A.3)
F
d) OPERATOR – set for different RATED
supply voltages:
– indicates the equipment set voltage
– PORTABLE EQUIPMENT indication is
visible from the exterior
– changing the setting changes the
indication
Result: P = Pass F = Fail NA = Not Applicable
---------------------- Page: 12 ----------------------
SIST IEC 61010-3-043:2002
TR 61010-3-043 IEC:2001(E) – 11 –
Clause Requirement Result Comments
Subclause
e) Accessory mains socket-outlets
accepting standard mains plugs are
marked:
– with the voltage, if it is different from
the mains supply voltage
– for use only with specific equipment
If not marked for specific equipment, it is
marked with:
– the maximum RATED current or power,
and maximum permitted leakage current
or
– symbol 14 with full details in the
documentation
F The measured value not more than
110 % (see Form A.3)
5.1.4 Fuses
OPERATOR replaceable fuse marking (see
also 5.4.5)
5.1.5 Measuring circuit TERMINALS Not applicable
5.1.6 TERMINALS and operating devices
Where necessary for safety, indication of
purpose of TERMINALS, connectors, controls
and indicators
Mains supply TERMINALS identified
Power supply switch on or off position
marked if used as disconnecting device
TERMINAL marking:
a) FUNCTIONAL EARTH TERMINALS
b) PROTECTIVE CONDUCTOR TERMINALS:
– symbol 6 is placed close to or on the
TERMINAL
or
– part of appliance inlet
c) TERMINALS of measuring and control
circuits
d) TERMINALS supplied from the interior
e) ACCESSIBLE FUNCTIONAL EARTH
TERMINALS
NORMAL USE
aa) setting a control could
cause a hazard, an indicating device is
provided
5.1.7 Equipment protected by DOUBLE
INSULATION or REINFORCED INSULATION
Protected throughout (symbol 11 used)
Only partially protected (symbol 11 not
used)
Result: P = Pass F = Fail NA = Not Applicable
---------------------- Page: 13 ----------------------
SIST IEC 61010-3-043:2002
– 12 – TR 61010-3-043 IEC:2001(E)
Clause Requirement Result Comments
Subclause
5.1.8 Battery charging
Equipment with means to charge re-
chargeable batteries is marked:
– to warn against the charging of non-
rechargeable batteries
– to indicate the type of rechargeable
battery used
5.2 Warning markings
– visible when ready for NORMAL USE
– if necessary, marked with symbol 14
– are near or on applicable parts
– statement to isolate or disconnect
– advice how to avoid contact with
ACCESSIBLE HAZARDOUS LIVE
parts
– easily touched high temperature parts
(symbol 13)
– if lockable door, to lock and retain key
– load type
5.2.101 Equipment with high ACCESSIBLE current
If equipment must be permanently
connected to protect against high
ACCESSIBLE current, a warning marking is
provided
5.2.102 Hot items falling out of equipment
If hot items can fall onto the equipment
support surface, a warning marking is
provided
5.3 Durability of markings
The required markings remain clear and
F legible in NORMAL USE (see Form A.4)
5.4 Documentation
5.4.1 General
Equipment is accompanied by
documentation which includes:
– technical specification
– instructions for use
– name and address of manufacturer or
supplier
Definition of INSTALLATION CATEGORY
Warning statements and a clear
explanation of warning symbols:
– provided in the documentation
or
– information is marked on the equipment
Result: P = Pass F = Fail NA = Not Applicable
---------------------- Page: 14 ----------------------
SIST IEC 61010-3-043:2002
TR 61010-3-043 IEC:2001(E) – 13 –
Clause Requirement Result Comments
Subclause
– instructions on warning signs for
country of use
5.4.2 Equipment RATINGS
Documentation includes:
– supply voltage or voltage range
– frequency or frequency range
RATING
– power or current
– a description of all input and output
connections
– RATING of insulation of external circuits,
when such circuits are nowhere ACCESSIBLE
– statement of the range of
environmental conditions
5.4.3 Equipment Installation
Instructions include:
– location and mounting instructions
including space for maintenance
– individual weights of principle heavy
subassemblies
– assembly instructions
– main supply requirements and
connections
– ventilation
– protective earthing
– sound power data requirements
– requirements for special services
5.4.3.101 Drying-out
After transport or storage in humid
conditions, equipment could fail to meet
the requirements in the standard;
instructions shall include a warning
5.4.4 Equipment operation
Instructions include:
a) operating controls:
i) interconnection
ii) use
b) positioned so that disconnection is
difficult
c) accessories:
i) interconnection
ii) suitability
iii) detachable parts and special materials
Result: P = Pass F = Fail NA = Not Applicable
---------------------- Page: 15 ----------------------
SIST IEC 61010-3-043:2002
– 14 – TR 61010-3-043 IEC:2001(E)
Clause Requirement Result Comments
Subclause
d) limits for intermittent operation
e) explanation of symbols used
f) cleaning
g) lockable door closure stop:
– correct use
– retain key
h) safe use of override key
i) action in case of malfunction
5.4.5 Equipment maintenance
Instructions include:
i) maintenance on threaded parts
ii) safety devices fitted
– replacement of hoses, etc.
– specific battery type
– any manufacturer specified parts
– RATING and characteristics of fuses
6 Protection against electric shock
F (see Form A.5)
6.1 General
Conformity is checked by the determination
of 6.2 and 6.3 followed by the tests of 6.4
to 6.12
Abestos is not used
6.1.1 Exceptions
F Capacitance test (see Forms A.6
and A.7)
6.2 Determination of ACCESSIBLE parts
F (see Form A.6)
6.3 Permissible limits for ACCESSIBLE parts
6.3.1 Values in NORMAL CONDITION
(see Form A.7)
F
6.3.1.2 Levels of PERMANENTLY CONNECTED
equipment are 1.5 times the values (see
Form A.7)
6.3.2 Values in SINGLE FAULT CONDITION
(see Form A.8)
F
6.3.2.2 Levels of PERMANENTLY CONNECTED
equipment are 1,5 times the values (see
Form A.8)
6.4 Protection in NORMAL CONDITION
(see 6.8 and 8.1)
Result: P = Pass F = Fail NA = Not Applicable
---------------------- Page: 16 ----------------------
SIST IEC 61010-3-043:2002
TR 61010-3-043 IEC:2001(E) – 15 –
Clause Requirement Result Comments
Subclause
6.5 Protection in SINGLE FAULT CONDITION
Additional protection is provided by:
– one or more of 6.5.1 to 6.5.3
or
– automatic disconnection of the supply
6.5.1 Protective earthing
ACCESSIBLE conductive parts:
– bonded to the protective conductor
terminal
or
– separated by screen or BARRIER from
parts which are HAZARDOUS LIVE
(For indirect bonding of measurement and
test equipment, see 6.5.1.4)
6.5.1.1 PROTECTIVE BONDING
PROTECTIVE BONDING consists of directly
connected structural parts or discrete
conductors or both; and withstands thermal
and dynamic stresses
6.5.1.2 Bonding impedance of plug-connected
F equipment (see Form A.9)
6.5.1.3 Bonding impedance of PERMANENTLY
F CONNECTED EQUIPMENT (see Form A.9)
6.5.1.4 Indirect bonding for measuring and test
F equipment (see Form A.9)
6.5.2 DOUBLE INSULATION and REINFORCED
INSULATION
(see 6.7, 6.8 and 6.9.2)
6.5.3 PROTECTIVE IMPEDANCE
F (see Form A.10)
Components wires and connections are
RATED as required
6.5.4 Built-in panel meters
If, after building-in, the requirements of
6.5.1 to 6.5.3 are not met, the panel meter:
– has no ACCESSIBLE conductive parts
– has basic insulation of accessible
surfaces
– has DOUBLE/REINFORCED INSULATION of
ACCESSIBLE
surface of parts intended to be
grasped
6.6 External circuits
6.6.1 Separation of internal circuits
F (see list in Forms A.1 and A.5)
If the other internal circuit exceeds the
values of 6.3.2 in NORMAL CONDITION and
only BASIC INSULATION:
Result: P = Pass F = Fail NA = Not Applicable
---------------------- Page: 17 ----------------------
SIST IEC 61010-3-043:2002
– 16 – TR 61010-3-043 IEC:2001(E)
Clause Requirement Result Comments
Subclause
a) short-circuit could not make external
circuit HAZARDOUS LIVE
b) manufacturer's instructions include:
– a statement that the TERMINAL for
external circuits is for use only with
equipment which has no live parts which
are ACCESSIBLE
– the RATING of the insulation required for
external circuits
– the connection to be used at the remote
end of external circuits
– the type of equipment which may be
connected to the TERMINAL
6.6.2 TERMINALS for external circuits
ACCESSIBLE TERMINALS are not HAZARDOUS
LIVE, except as permitted by 6.1.1
The following terminals are not hazardous
live:
– PROTECTIVE CONDUCTOR TERMINALS
– FUNCTIONAL EARTH TERMINALS
– headphone TERMINALS
TERMINALS which receive a charge from an
F internal capacitor (see Form A.7)
High-voltage TERMINALS energized from the
interior are:
– not accessible
or
–marked
6.6.3 Circuits with TERMINALS which are
HAZARDOUS LIVE
These circuits:
ACCESSIBLE
– are not connected to
conductive parts
or
– are connected to ACCESSIBLE
conductive parts, but are not mains circuits
and have one TERMINAL contact at earth
potential
No ACCESSIBLE conductive parts are
HAZARDOUS LIVE
6.7 CLEARANCES and CREEPAGE DISTANCES
(see annex D of IEC 61010-1 and
F Form A.11)
6.8 Dielectric strength tests
(see annex E of IEC 61010-1 and
F Form A.12)
Protection against the spread of fire
(see 9.1)
Result: P = Pass F = Fail NA = Not Applicable
---------------------- Page: 18 ----------------------
SIST IEC 61010-3-043:2002
TR 61010-3-043 IEC:2001(E) – 17 –
Clause Requirement Result Comments
Subclause
6.9 Constructional requirements for
protection against electric shock
6.9.1 General
In circuits exceeding the values of 6.3.2:
– security of wiring connections
– screws securing removable covers
– accidental loosening
6.9.2 ENCLOSURES of equipment with DOUBLE
INSULATION or REINFORCED INSULATION
NCLOSURE
E surrounds all metal parts
except for small metal parts which are
separated
NCLOSURES
E or parts made of insulating
material
Protection for metal ENCLOSURES or parts
by:
– protective impedance
or
– an insulating coating or BARRIER on the
inside
or
– CLEARANCES and CREEPAGE DISTANCES
cannot be reduced by loosening of parts or
wires
6.9.3 Equipment using PROTECTIVE BONDING
a) OPERATOR removable parts
b) Movable conductive connections
c) Exterior metal braids of cables
d) Mains passed through the equipment
e) Protective earthing conductors
green/yellow
Exceptions:
– earthing braids
– internal protective conductors
f) Equipment using PROTECTIVE BONDING
6.9.4 Over-range indication
6.10 Connection to mains supply source and
connections between parts of equipment
6.10.1 Mains supply cords
RATED for maximum equipment current
(see 5.1.3c)
Cable complies with IEC 60227 or
IEC 60245 or is a certified cord
Heat resistant if likely to contact hot parts
Result: P = Pass F = Fail NA = Not Applicable
---------------------- Page: 19 ----------------------
SIST IEC 61010-3-043:2002
– 18 – TR 61010-3-043 IEC:2001(E)
Clause Requirement Result Comments
Subclause
RATING
Temperature (cord and inlet)
Green/yellow used only for connection to
PROTECTIVE CONDUCTOR TERMINALS
Detachable cords with IEC 60320 mains
connectors:
– comply with IEC 60799
or
– have the current RATING of the mains
connector
6.10.2 Fitting of non-detachable mains supply
cords
6.10.2.1 Cord entry
Non-detachable cord protection:
– inlet smoothly rounded with radius
≥1,5 D
or
– insulated cord guard protruding ≥5 D
6.10.2.2 Cord anchorage
The protective earth conductor is the last
to take the strain
Cord anchorages:
– the cord is not clamped by direct
pressure from a screw
– knots are not used
– cannot push the cord into the
equipment to cause a hazard
– no failure of cord insulation in
anchorage with metal parts
– compression bushing:
a) clamps all types and sizes of mains
cords
and
b) is suitable:
i) for connection to TERMINALS provided
or
ii) it is designed for screened mains cord
– cord replacement does not cause a
hazard and method of strain relief is clear
F Push-pull test (see Form A.13)
– not used to attach any other component
6.10.3 Plugs and connectors
a) Mains supply plugs, connectors etc.,
conform with relevant specifications
b) If equipment supplied at voltages below
6.3.2.1:
– plugs of mains supply cords do not fit
mains sockets above RATED supply voltage
Result: P = Pass F = Fail NA = Not Applicable
---------------------- Page: 20 ----------------------
SIST IEC 61010-3-043:2002
TR 61010-3-043 IEC:2001(E) – 19 –
Clause Requirement Result Comments
Subclause
– mains type plugs used only for
connection to mains supply
c) Plug pins which receive a charge from
F an internal capacitor (see Form A.7)
d) Accessory mains socket outlets:
– if a standard mains plug is accepted,
there is a marking (see 5.1.3e)
– input has a protective earth conductor if
TERMINAL
outlet has earth contact
6.10.101 Connection of non-detachable mains
cords to TERMINALS
Flexible cables or cords attached to
TERMINAL blocks:
– no special preparation of conductor
required
– designed and positioned so that
conductor does not slip out
6.11 TERMINALS
6.11.1 ACCESSIBLE TERMINALS
a) No risk of accidental contact
(see also 5.1.6c)
b) Will not work loose
6.11.2 PROTECTIVE CONDUCTOR TERMINAL
a) Appliance inlet (no requirement)
b) For rewirable cords and PERMANENTLY
CONNECTED EQUIPMENT, PROTECTIVE
CONDUCTOR TERMINAL is close to mains
TERMINALS
supply
c) If no mains supply is required, any
PROTECTIVE CONDUCTOR TERMINAL:
– is near TERMINALS of circuit for which
protective earthing is necessary
– external if other TERMINALS external
d) Equivalent current-carrying capacity to
mains supply TERMINALS
e) Soldered connections:
i) independently secured
ii) not used for other purposes
– Screw connections are secured
f) Contact surfaces are metal
g) If plug-in, makes first and breaks last
h) Protective conductor of measuring
circuit:
– current RATING
Result: P = Pass F = Fail NA = Not Applicable
---------------------- Page: 21 ----------------------
SIST IEC 61010-3-043:2002
– 20 – TR 61010-3-043 IEC:2001(E)
Clause Requirement Result Comments
Subclause
– protective bonding:
i) not interrupted
or
ii) indirect bonding
6.11.3 FUNCTIONAL EARTH TERMINALS
Independent connection
6.12 Disconnection from supply source
6.12.1 General
Disconnection device provided
6.12.1.1 Exception to 6.12.1
Short-circuit or overload cannot cause a
hazard
6.12.2 Requirements according to type of
equipment
6.12.2.1 PERMANENTLY CONNECTED EQUIPMENT
– switch or circuit breaker is part of the
equipment
or
– documentation specifies switch location
and marking
6.12.2.2 Single-phase cord-connected equipment
– switch or circuit breakers
or
– appliance coupler (disconnectable
without TOOL)
or
– separable plug (without locking device)
6.12.2.3 Hazards arising from function
Emergency switch
Emergency switch <1 m from the moving
part
6.12.3 Disconnecting devices
Electrically close to the supply
6.12.3.1 Switches and circuit-breakers
When used as disconnection device:
– meet IEC 60947-1 and IEC 60947-3
– contact separation
– contact position evident in off position
– marked to indication function
– not incorporated in mains cord
– does not interrupt protection earth
conductor
– if has other contacts, meets separation
requirements of 6.6 and 6.7
Result: P = Pass F = Fail NA = Not Applicable
---------------------- Page: 22 ----------------------
SIST IEC 61010-3-043:2002
TR 61010-3-043 IEC:2001(E) – 21 –
Clause Requirement Re
...