Radiation protection instrumentation - Installed personnel surface contamination monitoring assemblies

Defines mechanical and operational characteristics, minimum performance characteristics and general test procedures for personnel monitoring equipment. Applies to contamination warning assemblies, meters and monitors used for the monitoring of radioactive contamination on the surface of personnel whether they be clothed or not.

Strahlenschutz-Messgeräte - Festinstallierte Personenkontaminationsmonitore

Instrumentation pour la radioprotection - Ensembles fixes pour la surveillance de la contamination de surface du personnel

Définit les caractéristiques mécaniques et d'utilisation, les performances minimales et les procédures générales d'essai pour les équipements de surveillance du personnel. S'applique aux ensembles de prévention, de mesure et de surveillance utilisés pour la surveillance de la contamination radioactive en surface des personnels, que ceux-ci soient habillés ou non.

Oprema za varstvo pred sevanjem - Vgrajeni sestavi za nadzor kontaminacije delovnih površin (IEC 61098:2003)

General Information

Status
Published
Publication Date
26-Jul-2007
Withdrawal Date
31-Mar-2010
Current Stage
6060 - Document made available - Publishing
Start Date
27-Jul-2007
Completion Date
27-Jul-2007

Relations

Buy Standard

Standard
EN 61098:2007
English language
56 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-november-2007
Oprema za varstvo pred sevanjem - Vgrajeni sestavi za nadzor kontaminacije
delovnih površin (IEC 61098:2003)
Radiation protection instrumentation - Installed personnel surface contamination
monitoring assemblies
Strahlenschutz-Messgerte - Festinstallierte Personenkontaminationsmonitore
Instrumentation pour la radioprotection - Ensembles fixes pour la surveillance de la
contamination de surface du personnel
Ta slovenski standard je istoveten z: EN 61098:2007
ICS:
13.280 Varstvo pred sevanjem Radiation protection
17.240 Merjenje sevanja Radiation measurements
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD
EN 61098
NORME EUROPÉENNE
July 2007
EUROPÄISCHE NORM
ICS 13.280
English version
Radiation protection instrumentation -
Installed personnel surface contamination monitoring assemblies
(IEC 61098:2003, modified)
Instrumentation pour la radioprotection -  Strahlenschutz-Messgeräte -
Ensembles fixes pour la surveillance Festinstallierte
de la contamination de surface Personenkontaminationsmonitore
du personnel (IEC 61098:2003, modifiziert)
(CEI 61098:2003, modifiée)
This European Standard was approved by CENELEC on 2007-04-01. CENELEC members are bound to
comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration.

Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the Central Secretariat or to any CENELEC member.

This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and
notified to the Central Secretariat has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Cyprus, the
Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy,
Latvia, Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia,
Slovenia, Spain, Sweden, Switzerland and the United Kingdom.

CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung

Central Secretariat: rue de Stassart 35, B - 1050 Brussels

© 2007 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 61098:2007 E
Foreword
The text of the International Standard IEC 61098:2003, prepared by SC 45B, Radiation protection
instrumentation, of IEC TC 45, Nuclear instrumentation, together with the common modifications
prepared by the CENELEC BTTF 111-3, Nuclear instrumentation and radiation protection
instrumentation, was submitted to the formal vote and was approved by CENELEC as EN 61098 on
2007-04-01.
The following dates were fixed:
– latest date by which the EN has to be implemented
at national level by publication of an identical
national standard or by endorsement (dop) 2008-04-01
– latest date by which the national standards conflicting
with the EN have to be withdrawn (dow) 2010-04-01
Clauses, subclauses, notes, tables and figures which are additional to those in IEC 61098:2003 are
prefixed “Z”.
Annexes ZA, ZB and ZC have been added by CENELEC.
__________
– 3 – EN 61098:2007
Endorsement notice
The text of the International Standard IEC 61098:2003 was approved by CENELEC as a European
Standard with agreed common modifications as given below.
COMMON MODIFICATIONS
1 Scope and object
First paragraph, first line: delete “, meters”.
Second paragraph: replace “face” with “head”.

2 Normative references
Replace “IEC 60050(151):2001” with “IEC 60050-151:2001”.
Replace “IEC 60050(393):1996” with “IEC 60050-393:2003, International Electrotechnical
Vocabulary (IEV) – Part 393: Nuclear instrumentation – Physical phenomena and basic
concepts”.
Replace “IEC 60050(394):1995” with “IEC 60050-394:1995 + A1:1996 + A2:2000“.
Replace “IEC 61000-4-2:1995“ with “IEC 61000-4-2:1995 + A1:1998 + A2:2000“.
Replace “IEC 61000-4-3:2002“ with “IEC 61000-4-3:2002 + A1:2002“.
Replace “IEC 61000-4-5:1995“ with “IEC 61000-4-5:1995 + Corr. 1995 + A1:2000“.
Replace “IEC 61000-4-6:2003“ with “IEC 61000-4-6:2003 + A1:2004“.
Replace “IEC 61000-4-8:1993“ with “IEC 61000-4-8:1993 + A1:2000“.
Replace “IEC 61000-4-12:1995” with “IEC 61000-4-12:1995 + A1:2000“.
Add the following references:
IEC 60038:1983 + A1:1994 + A2:1997, IEC standard voltages
IEC 60068 series, Environmental testing
IEC 60359:2001, Electrical and electronic measurement equipment – Expression of
performance
ISO 7503-1:1988, Evaluation of surface contamination – Part 1: Beta-emitters (maximum beta
energy greater than 0,15 MeV) and alpha-emitters
ISO 8769-2:1996, Reference sources for the calibration of surface contamination monitors –
Part 2: Electrons of energy less than 0,15 MeV and photons of energy less than 1,5 MeV
ISO 11929 series, Determination of the detection limit and decision threshold for ionizing
radiation measurements
3 Terms and definitions
3.2
surface emission rate of a source
Add below the definition: “(according to ISO 8769 definition)”.

3.5
high efficiency source
Replace “0,5 keV” with “0,59 keV”.

3.6
small source
Add “surface” between “active” and “dimension”.

3.14
decision threshold
The whole definition to be replaced with a note, reading
NOTE Not used in EN 61098, see 3.Z1.

3.15
decay probability
The whole definition to be replaced with a note, reading
NOTE Not used in this document.

Add the following definitions:
3.Z1
minimum detectable surface emission rate
specific performance criteria taking account both the statistical fluctuation in the background
count rates and nominal factors of the background to take account of short term spatial and
temporal changes in background as well as changes due to the operator mass
NOTE This definition is specific for this standard only.
3.Z2
lowest level of detection
minimum count rate above background that will not trigger a false alarm due to statistical
fluctuations
3.Z3
lowest limit of detection
limit of detection associated to the lowest level of detection

4 Classification of assemblies
4.1 According to type of radiation to be measured
Replace “contamination warning assemblies and monitors” with “contamination monitors and
warning assemblies” all over the clause.
Fourth dash: add at the end: “(where the alpha and beta contaminations are indicated
separately)”.
Fifth dash: replace the text in brackets with “(where the beta and gamma contaminations are
indicated separately)”.
Sixth dash: replace the text in brackets with “(where the alpha, beta and gamma
contaminations are indicated separately)”.

4.2 According to type of surface
First dash: replace “face” with “head”.
Second dash: delete “hand warning”.
Third dash: delete “foot warning”.
Fourth dash: delete “hand and foot warning”.

– 5 – EN 61098:2007
Add the following clause after 4.3:
4.Z1 Compressed overall classification
The equipment may also be supported with a compressed overall classification based on the
intended field of use, as shown in Table Z1 below.

Table Z1 – Classification according to the intended field of use
Common  Options
Category Abbreviation Energy range Energy Ambient background compensation Control
extension type
Alpha A 3 MeV to 6 MeV – O: Without any compensation Hands,

CP: Post compensation Feet,
CPA: Alpha natural compensation Whole body
CPB: Beta high energy compensation
CS: Simultaneous compensation
CSA: Alpha natural dynamic
compensation
CSB: Beta high energy dynamic
compensation
Beta B E of  – O: Without any compensation Hands,
max
150 keV to 3 MeV
CP: Post compensation Feet,
CPA: Alpha natural compensation Whole body
CPB: Beta ambient compensation
CPGX: Gamma and X-ray ambient
compensation
CS: Simultaneous compensation
CSA: Alpha natural dynamic
compensation
CSB: Beta ambient dynamic
compensation
CSGX: Gamma and X-ray ambient
dynamic compensation
Gamma G 500 keV to 1,5 MeV L1: Low limit 1 O: Without any compensation Hands,
at 50 keV
CP: Post compensation Feet,
L2: Low limit 2
CPB: Beta ambient compensation Whole body
at 150 keV
CPGX: Gamma and X-ray ambient
compensation
CS: Simultaneous compensation
CSB: Beta ambient dynamic
compensation
CSGX: Gamma and X-ray ambient
dynamic compensation
X-rays X 5 keV to 20 keV H: High limit O: Without any compensation Hands,
at 50 keV
CP: Post compensation Feet,
CPB: Beta ambient compensation Whole body
CPGX: Gamma and X-ray ambient
compensation
CS: Simultaneous compensation
CSB: Beta ambient dynamic
compensation
CSGX: Gamma and X-ray ambient
dynamic compensation
5 Design characteristics
5.3 Hand monitoring facilities
Add at the end of the third paragraph: “if not otherwise agreed between manufacturer and user
for existing equipment.”
5.4 Foot monitoring facilities
In the second paragraph, replace “30 cm” with “35 cm”.

6 Performance requirements and test procedures
6.1 General test procedure
6.1.1 Nature of tests
Add a note below the first paragraph, reading:
NOTE Other test methods may be agreed between manufacturer and purchaser to be regarded as additional
acceptance tests. Examples may be published as informative annexes to this standard in future.

6.3 Reference sources
Start the second and third paragraphs as follows: “For assemblies for the monitoring of …”

7 Radiation characteristics
7.1 Variation of response with source position
7.1.1 For clothing or the body
7.1.1.1 Requirement
Add an additional paragraph below the current text, reading:
In order to improve the performance and meet the requirements of this standard, the
manufacturer may in the case of beta and/or gamma monitoring additionally sum the response
of any number of adjacent detectors.

7.1.1.2 Method of test
7.1.1.2.1 Alpha monitoring systems
Add a note below the current text, reading:
NOTE For a possible further test (for further characterization of the sensitive volume) see ZA.2.
7.1.1.2.2 Beta monitoring systems
Change the existing note to become NOTE 1.
Add a second note reading:
NOTE 2 For a possible further test (for further characterization of the sensitive volume) see ZA.2.
7.1.1.2.3 Gamma monitoring systems
Add a note below the current text, reading:
NOTE For a possible further test (for further characterization of the sensitive volume) see ZA.1.

– 7 – EN 61098:2007
7.3 Minimum detectable surface emission rate
Delete “Decision threshold” in the headline.
Replace “decision threshold” with “minimum detectable surface emission rate” wherever it
appears (also in any subclause).
Add new paragraph between the first and the second paragraphs, reading:
Since the minimum detectable surface emission rate depends on the response of the equipment
and this is going to depend on where on the body the activity is to be measured, the minimum
detectable surface emission rate is dependent on where the activity is. In order to provide some
indication of performance an average minimum detectable activity shall be determined from the
average efficiency. Since also the signal from the background can be large by comparison to
the signal from the activity being measured, and this background will be far from constant both
in terms of time and position, allowance in determining the minimum detectable surface
emission rate shall take account of this. The user also has a significant effect on the
background. In this document the minimum detectable surface emission rate has a special
meaning and includes arbitrary factors not used in the more precise decision threshold defined
in ISO 11929.
7.3.1 For clothing or body
Replace the first paragraph with:
For the purposes of this document, the minimum detectable surface emission rate shall relate to
the 4 π body average efficiency to clothing (body). This shall be determined as below, or the
manufacturer shall undertake similar methods of determining an average efficiency. In the latter
case the manufacturer shall provide the purchaser with a full description of the phantom used,
all measurements taken, the assumptions made in calculating the average efficiency and the
calculations made to determine this “average”. The 4 π efficiency shall be determined from both
the vertical response characteristic shown in Figure 1 and the polar response given in Figure 2.
Delete the third sentence of the third paragraph.
In the last paragraph, replace the second sentence with “Where monitoring is undertaken by
two or more steps, the sum of the times taken for each monitoring step shall not exceed the
time specified above or a time agreed between the manufacturer and purchaser.”

7.4 Variation of response with energy
7.4.1 Beta
7.4.1.1 Requirements
90 90 90
Replace “ Sr” with “ Sr/ Y”.
7.4.1.2 Method of test
Add in the second line of item b) “or by default 10 cm x 10 cm (performing two measurements
without overlap in order to cover the whole detector area and using the mean response for
further reference)” behind “15 cm x 10 cm”.
Add in the second line of item c) “or by default 10 cm x 10 cm (performing three measurements
without overlap in order to cover the whole detector area and using the mean response for
further reference)” behind “30 cm x 10 cm”.

7.4.3 Gamma
7.4.3.3 Method of test
Item a), first paragraph, second sentence: add “of the nuclide” between “A small source” and “of
interest”.
Item b), first paragraph, last word: replace “counter” with “detector”.

Add in the second line of item b) “or by default 10 cm x 10 cm (performing two measurements
without overlap in order to cover the whole detector area and using the mean response for
further reference)” behind “15 cm x 10 cm”.
Add in the second line of item c) “or by default 10 cm x 10 cm (performing three measurements
without overlap in order to cover the who
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.