SIST EN IEC 61326-1:2021/A11:2026
(Amendment)Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
- Abstract
This part of EN 61326 is a product family standard specifying requirements for immunity and emissions regarding electromagnetic compatibility (EMC) for electrical equipment, operating from a supply or battery of less than 1 000 V AC or 1 500 V DC or from the circuit being measured. Equipment intended for professional, industrial-process, industrial-manufacturing and educational use is covered by this part. It includes equipment and computing devices for
- measurement and test;
- control;
- LABORATORY use;
- accessories intended for use with the above (such as sample handling equipment), intended to be used in industrial and non-industrial locations.
- Status
- Not Published
- Public Enquiry End Date
- 30-Dec-2023
- Technical Committee
- MOV - Measuring equipment for electromagnetic quantities
- Current Stage
- 6060 - National Implementation/Publication (Adopted Project)
- Start Date
- 30-Apr-2026
- Due Date
- 05-Jul-2026
SIST EN IEC 61326-1:2021/A11:2026 is a Slovenian amendment to EN IEC 61326-1:2021 for electromagnetic compatibility (EMC) of electrical equipment for measurement, control and laboratory use. It updates immunity and emission requirements, performance criteria, and user-documentation rules for equipment used in professional, industrial-process, industrial-manufacturing and educational settings.
What does SIST EN IEC 61326-1:2021/A11:2026 specify?
SIST EN IEC 61326-1:2021/A11:2026 specifies changes to the general EMC requirements for this equipment family. It keeps the focus on immunity and emissions for equipment operating from a supply or battery below 1 000 V AC or 1 500 V DC, or from the circuit being measured.
The amendment is organized as modifications to the Introduction and Clauses 1, 4, 5, 6, 7 and 9, plus Annex A, replacement of Annex ZA, and a new informative Annex ZZ.
| Annex | What it covers |
|---|---|
| Annex A | Immunity test requirements for portable test and measurement equipment powered by battery or from the circuit being measured |
| Annex ZA | Normative references to international publications with corresponding European publications |
| Annex ZZ | Relationship between this European Standard and Directive 2014/30/EU |
What are the key requirements of SIST EN IEC 61326-1:2021/A11:2026?
SIST EN IEC 61326-1:2021/A11:2026 makes the EMC test plan and the user documentation part of the compliance decision. In practice, manufacturers need to define the intended EMC environment, choose the right test set-up, and state the permitted performance level or loss of performance before testing starts.
Test planning and test items
Clause 4 allows testing on one EUT or split across multiple EUTs, but every result must be traceable to the item tested. Clause 5 says the equipment shall be tested in the worst-case performance-relevant arrangement and as type tests. That matters for modular systems, because the test setup has to reflect the most demanding intended configuration, not just a convenient one.
Immunity requirements
Clause 6 uses separate immunity tables for the BASIC, INDUSTRIAL and CONTROLLED electromagnetic environments, and Annex A adds a table for portable battery-powered or circuit-powered test and measurement equipment. The tables pair each port and phenomenon with a test method and performance criterion, so the reader can match the equipment, the environment and the expected behaviour.
Clause 6 also defines what the performance criteria mean in practice:
- Criterion A - the EUT keeps operating as intended during and after the test, with no loss of function or data.
- Criterion B - temporary loss of function is allowed during the test if the EUT recovers by itself afterward.
- Criterion C - loss of function is allowed during the test if recovery is self-recoverable or restored by user action, and the recovery procedure must be in the user documentation.
These criteria matter because they tell the manufacturer what level of disturbance is acceptable for operation, measurement, control, outputs, communication, data retention and HMI.
Emission requirements
Clause 7 points radio-frequency emission limits to EN 55011 for Group 1 and Group 2 CLASS A and CLASS B EQUIPMENT. For low-frequency emissions, it directs CLASS B EQUIPMENT with input current up to 16 A to EN 61000-3-2 and EN 61000-3-3, and higher currents up to 75 A per phase to EN 61000-3-12 and EN 61000-3-11. In practice, this tells the manufacturer which emission standard to use for the product’s current rating and equipment class.
User information
Clause 9 requires the manufacturer to state the intended electromagnetic environment, the applicable immunity test requirements, and the emission standards used. If a PERFORMANCE LEVEL or permissible LOSS OF PERFORMANCE applies under immunity conditions, that must also be described in the product specification or user documentation.
What terms does SIST EN IEC 61326-1:2021/A11:2026 define?
- EMC (electromagnetic compatibility) - the subject of the standard, covering both immunity to disturbance and emissions of disturbance.
- EUT (equipment under test) - the item being tested against the EMC requirements.
- Performance criterion A - the EUT must keep operating as intended, with no loss of function or data.
- Performance criterion B - temporary loss of function is allowed during testing if the EUT recovers automatically afterward.
- Performance criterion C - loss of function is allowed during testing if recovery is automatic or user-restored, and the recovery procedure is documented.
- HMI (human-machine interface) - buttons, tactile panels, displays and audible information used by the operator.
- CONTROLLED ELECTROMAGNETIC ENVIRONMENT - the environment category used for more sheltered places such as analysis, test and service laboratories.
Who uses SIST EN IEC 61326-1:2021/A11:2026?
Manufacturers of measurement, control and laboratory equipment use this amendment to choose EMC tests, set performance criteria and write product documentation. EMC test laboratories use it to plan type tests and select the correct disturbance levels for the intended environment.
Quality managers, compliance engineers and technical buyers use it to check whether equipment is intended for BASIC, INDUSTRIAL or CONTROLLED electromagnetic environments, and whether the declared performance under disturbance matches the intended use. It is also relevant for accessories such as sample handling equipment that are used with the covered equipment.
What changed in SIST EN IEC 61326-1:2021/A11:2026 from the previous edition?
SIST EN IEC 61326-1:2021/A11:2026 modifies EN IEC 61326-1:2021. The stated technical changes are updates to the immunity requirements, the performance criterion text, and the emission requirements.
It also updates the Introduction, Scope, General requirements, EMC test plan, instructions for use, Annex A, Annex ZA, and adds Annex ZZ for the link to Directive 2014/30/EU.
Which standards are used with SIST EN IEC 61326-1:2021/A11:2026?
| Standard | What it contributes |
|---|---|
| IEC 60050-161 | EMC vocabulary |
| IEC 61000-3-2 | Harmonic current emission limits for equipment up to 16 A per phase |
| IEC 61000-3-3 | Voltage changes, voltage fluctuations and flicker for equipment up to 16 A per phase |
| IEC 61000-3-11 | Voltage changes, voltage fluctuations and flicker for equipment up to 75 A and subject to conditional connection |
| IEC 61000-3-12 | Harmonic current limits for equipment above 16 A and up to 75 A per phase |
| IEC 61000-4-2 | Electrostatic discharge immunity test |
| IEC 61000-4-3 | Radiated RF electromagnetic field immunity test |
| IEC 61000-4-4 | Electrical fast transient/burst immunity test |
| IEC 61000-4-5 | Surge immunity test |
| IEC 61000-4-6 | Conducted RF immunity test |
| IEC 61000-4-8 | Power frequency magnetic field immunity test |
| IEC 61000-4-11 | Voltage dips, short interruptions and voltage variations immunity tests |
| CISPR 11 / EN 55011 | Radio-frequency disturbance limits and measurement methods |
What does the SIST EN IEC 61326-1:2021/A11:2026 document contain?
The amendment contains replacement text for the scope, general rules, test planning, immunity criteria, emission clauses and user instructions. It also includes immunity test tables for BASIC, INDUSTRIAL and CONTROLLED environments, plus a separate table for portable equipment powered by battery or from the measured circuit.
Annex ZA gives the European publication mapping for the referenced standards, which helps users identify the correct harmonized versions. Annex ZZ connects the document to Directive 2014/30/EU, which is important for conformity assessment and for understanding how the standard is used in regulatory contexts.
Relations
- Effective Date
- 28-Apr-2026
- Effective Date
- 28-Apr-2026
- Effective Date
- 28-Apr-2026
- Effective Date
- 28-Apr-2026
- Effective Date
- 28-Apr-2026
- Effective Date
- 01-Feb-2022
Get Certified
Connect with accredited certification bodies for this standard
DVS-ZERT GmbH
German welding certification society.
IMP NDT d.o.o.
Non-destructive testing services. Radiography, ultrasonic, magnetic particle, penetrant, visual inspection.
Inštitut za kovinske materiale in tehnologije
Institute of Metals and Technology. Materials testing, metallurgical analysis, NDT.
Sponsored listings
Frequently Asked Questions
SIST EN IEC 61326-1:2021/A11:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements". This standard covers: This part of EN 61326 is a product family standard specifying requirements for immunity and emissions regarding electromagnetic compatibility (EMC) for electrical equipment, operating from a supply or battery of less than 1 000 V AC or 1 500 V DC or from the circuit being measured. Equipment intended for professional, industrial-process, industrial-manufacturing and educational use is covered by this part. It includes equipment and computing devices for - measurement and test; - control; - LABORATORY use; - accessories intended for use with the above (such as sample handling equipment), intended to be used in industrial and non-industrial locations.
This part of EN 61326 is a product family standard specifying requirements for immunity and emissions regarding electromagnetic compatibility (EMC) for electrical equipment, operating from a supply or battery of less than 1 000 V AC or 1 500 V DC or from the circuit being measured. Equipment intended for professional, industrial-process, industrial-manufacturing and educational use is covered by this part. It includes equipment and computing devices for - measurement and test; - control; - LABORATORY use; - accessories intended for use with the above (such as sample handling equipment), intended to be used in industrial and non-industrial locations.
SIST EN IEC 61326-1:2021/A11:2026 is classified under the following ICS (International Classification for Standards) categories: 17.220.20 - Measurement of electrical and magnetic quantities; 19.080 - Electrical and electronic testing; 25.040.40 - Industrial process measurement and control; 33.100.01 - Electromagnetic compatibility in general; 33.100.20 - Immunity. The ICS classification helps identify the subject area and facilitates finding related standards.
SIST EN IEC 61326-1:2021/A11:2026 has the following relationships with other standards: It is inter standard links to SIST EN IEC 55011:2025, SIST EN 62952-1:2017, SIST EN IEC 61987-1:2025, SIST EN 50549-10:2023, SIST EN 61069-4:2017, SIST EN IEC 61326-1:2021. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
SIST EN IEC 61326-1:2021/A11:2026 is associated with the following European legislation: EU Directives/Regulations: 2014/30/EU; Standardization Mandates: M/552. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
SIST EN IEC 61326-1:2021/A11:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI STANDARD
SIST EN IEC 61326-1:2021/oprAA:2023
01-december-2023
Električna oprema za merjenje, nadzor in laboratorijsko uporabo - Zahteve za
elektromagnetno združljivost (EMC) - 1. del: Splošne zahteve
Electrical equipment for measurement, control and laboratory use - EMC requirements -
Part 1: General requirements
Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV-Anforderungen - Teil 1:
Allgemeine Anforderungen
Matériel électrique de mesure, de commande et de laboratoire - Exigences relatives à la
CEM - Partie 1: Exigences générales
Ta slovenski standard je istoveten z: EN IEC 61326-1:2021/prAA:2023
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
33.100.01 Elektromagnetna združljivost Electromagnetic compatibility
na splošno in general
SIST EN IEC 61326-1:2021/oprAA:2023 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
SIST EN IEC 61326-1:2021/oprAA:2023
SIST EN IEC 61326-1:2021/oprAA:2023
EUROPEAN STANDARD DRAFT
EN IEC 61326-1:2021
NORME EUROPÉENNE
EUROPÄISCHE NORM
prAA
October 2023
ICS 25.040.40; 17.220.20; 33.100.20 -
English Version
Electrical equipment for measurement, control and laboratory
use - EMC requirements - Part 1: General requirements
Matériel électrique de mesure, de commande et de Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV-
laboratoire - Exigences relatives à la CEM - Partie 1: Anforderungen - Teil 1: Allgemeine Anforderungen
Exigences générales
This draft amendment prAA, if approved, will modify the European Standard EN IEC 61326-1:2021; it is submitted to CENELEC members
for enquiry.
Deadline for CENELEC: 2024-01-12.
It has been drawn up by CLC/TC 65X.
If this draft becomes an amendment, CENELEC members are bound to comply with the CEN/CENELEC Internal Regulations which
stipulate the conditions for giving this amendment the status of a national standard without any alteration.
This draft amendment was established by CENELEC in three official versions (English, French, German).
A version in any other language made by translation under the responsibility of a CENELEC member into its own language and notified to
the CEN-CENELEC Management Centre has the same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Türkiye and the United Kingdom.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are aware and to
provide supporting documentation.
Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without notice and
shall not be referred to as a European Standard.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2023 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Project: 75662 Ref. No. EN IEC 61326-1:2021/prAA:2023 E
SIST EN IEC 61326-1:2021/oprAA:2023
EN IEC 61326-1:2021/prAA:2023 (E)
9 Contents Page
10 European foreword . 3
11 1 Modifications to Introduction . 5
12 2 Modifications to Clause 1, “Scope” . 5
13 3 Modifications to Clause 4, “General” . 5
14 4 Modifications to Clause 5, “EMC test plan” . 5
15 5 Modifications to Clause 6, “Immunity requirements” . 5
16 6 Modifications to Clause 7, “Emission requirements” .11
17 7 Modifications to Clause 9, “Instruction for use” .12
18 8 Modifications to Annex A ” Immunity test requirements for PORTABLE TEST AND MEASUREMENT
19 EQUIPMENT powered by battery or from the circuit being measured “ .12
20 9 Replacement of Annex ZA, “Normative references to international publications with their
21 corresponding European publications” .12
22 Annex ZA (normative) Normative references to international publications with their
23 corresponding European publications .14
24 Annex ZZ (informative) Relationship between this European Standard and the essential
25 requirements of Directive 2014/30/EU [2014 OJ L96] aimed to be covered .16
SIST EN IEC 61326-1:2021/oprAA:2023
EN IEC 61326-1:2021/prAA:2023 (E)
26 European foreword
27 This document (EN IEC 61326-1:2021/prAA:2023) has been prepared by CLC/TC65X “Industrial-process
28 measurement, control and automation”
29 This document is currently submitted to the Enquiry.
30 The following dates are proposed:
• latest date by which the existence of this (doa) dor + 6 months
document has to be announced at national
level
• latest date by which this document has to be (dop) dor + 12 months
implemented at national level by publication of
an identical national standard or by
endorsement
• latest date by which the national standards (dow) dor + 36 months
conflicting with this document have to be (to be confirmed or
withdrawn modified when voting)
31 This document will modify EN IEC 61326-1:2021.
32 EN IEC 61326:2021/prAA:2023 includes the following significant technical changes with respect to
33 EN IEC 61326-1:2021:
34 — Update of immunity requirements
35 — Update of performance criterion
36 — Update of emission requirements
37 This document has been prepared under a standardization request addressed to CENELEC by the European
38 Commission. The Standing Committee of the EFTA States subsequently approves these requests for its
39 Member States.
40 For the relationship with EU Legislation, see informative Annex ZZ, which is an integral part of this document.
SIST EN IEC 61326-1:2021/oprAA:2023
EN IEC 61326-1:2021/prAA:2023 (E)
41 Introduction
42 The purpose of this amendment is to identify and modify the requirements that would not fit the EMCD
43 requirements, so that the amendment could be harmonized under this Directive.
SIST EN IEC 61326-1:2021/oprAA:2023
EN IEC 61326-1:2021/prAA:2023 (E)
45 1 Modification to Introduction
46 Replace the last paragraph with:
47 “This part of EN 61326 specifies the EMC requirements that are generally applicable to all equipment within its
48 scope. For certain types of equipment, these requirements will be supplemented or modified by the special
49 requirements of one, or more than one, particular part EN 61326-2 (all parts). EN 61326-2 (all parts) should be
50 read in conjunction with the EN 61326-1 requirements.
51 NOTE This informative statement applies to any edition of these standards.”
52 2 Modification to Clause 1, “Scope”
53 Replace the first paragraph of Clause 1 with:
54 “This part of EN 61326 is a product family standard specifying requirements for immunity and emissions
55 regarding electromagnetic compatibility (EMC) for electrical equipment, operating from a supply or battery of
56 less than 1 000 V AC or 1 500 V DC or from the circuit being measured. Equipment intended for professional,
57 industrial-process, industrial-manufacturing and educational use is covered by this part. It includes equipment
58 and computing devices for
59 — measurement and test;
60 — control;
61 — LABORATORY use;
62 — accessories intended for use with the above (such as sample handling equipment),
63 intended to be used in industrial and non-industrial locations.”
64 3 Modification to Clause 4, “General”
65 Replace the last paragraph of Clause 4 with:
66 “The manufacturer may select to perform all tests on either a single EUT or divide the tests among multiple
67 EUTs. If the latter, each test result shall be traceable to the EUT tested.”
68 4 Modification to Clause 5, “EMC test plan”
69 Replace 5.2.1 “General” with:
70 “Measurement, control and LABORATORY equipment often consists of systems with no fixed configuration. The
71 kind, number and installation of different subassemblies within the equipment may vary from system to system.
72 Thus, the tests shall be performed on the worst-case performance-relevant arrangement(s) covering all intended
73 uses and carried out as TYPE TESTS.”
74 5 Modification to Clause 6, “Immunity requirements”
75 Replace 6.2” Immunity test requirements” with:
76 “Table 1 gives the immunity requirements for equipment intended to be used in a BASIC ELECTROMAGNETIC
77 ENVIRONMENT.
78 Table 2 gives the immunity requirements for equipment intended to be used in an INDUSTRIAL ELECTROMAGNETIC
79 ENVIRONMENT.
80 Table 3 gives the immunity requirements for equipment intended to be used in a CONTROLLED ELECTROMAGNETIC
81 ENVIRONMENT.
SIST EN IEC 61326-1:2021/oprAA:2023
EN IEC 61326-1:2021/prAA:2023 (E)
82 Table A.1 gives the immunity requirements for portable test and measurement equipment that is powered by
83 battery or from the circuit being measured.
84 The performance criteria A, B, and C that are mentioned in the following tables are described in Table 4, Table 5
85 and Table 6.
86 Table 1 — Immunity test requirements for equipment
87 intended to be used in a BASIC ELECTROMAGNETIC ENVIRONMENT
Performance
Port Phenomenon Basic standard Test value
criterion
Enclosure ESD EN 61000-4-2:2009 ± 4 KV contact discharge B
± 8 KV air discharge B
Electromagnetic field EN 61000-4-3:2006 3 V/m (80 MHz to 1 GHz; A
+ EN 61000-4-3:2006/A1:2008 1,4 GHz to 6 GHz),
+ EN 61000-4-3:2006/A2:2010
Power frequency EN 61000-4-8:2010 3 A/m (50 Hz) A
b
magnetic field
AC power Burst EN 61000-4-4:2012 ±1 kV (5 kHz) B
(including protective
Surge EN 61000-4-5:2014 + ±0,5 kV line-to-line B
earth)
EN 61000-4-5:2014/A1:2017
±1 kV line-to-ground B
Conducted RF EN 61000-4-6:2014 3 V (150 kHz to 80 MHz) A
Voltage dip EN IEC 61000-4-11:2020 0 % during half cycle B
0 % during 1 cycle B
c
70 % during 25/30 cycles C
c
Short interruptions EN IEC 61000-4-11:2020 0 % during 250/300 cycles C
d, e
DC power Burst EN 61000-4-4:2012 ±1 kV (5 kHz) B
(including protective
Surge EN 61000-4-5:2014 + ±0,5 kV line-to-line B
earth)
EN 61000-4-5:2014/A1:2017
±1 kV line-to-ground B
Conducted RF EN 61000-4-6:2014 3 V (150 kHz to 80 MHz) A
e d
I/O signal/control Burst EN 61000-4-4:2012 ±0,5 kV (5 kHz) B
(including functional
a
Surge EN 61000-4-5:2014 + ±1 kV line-to-ground B
earth)
EN 61000-4-5:2014/A1:2017
d
Conducted RF EN 61000-4-6:2014 3 V (150 kHz to 80 MHz) A
e d
I/O signal/control Burst EN 61000-4-4:2012 ±1 kV (5 kHz) B
connected directly to a
Surge EN 61000-4-5:2014 + ±0,5 kV line-to-line B
mains supply
EN 61000-4-5:2014/A1:2017
±1 kV line-to-ground B
d
Conducted RF EN 61000-4-6:2014 3 V (150 kHz to 80 MHz) A
a
Only in the case of lines > 30 m (called LONG-DISTANCE LINES).
b
Only to equipment containing Reed relays or Hall effect circuits (called magnetically sensitive equipment).
c
For example, “25/30 cycles” means “25 cycles for 50 Hz test” or “30 cycles for 60 Hz test”.
d
Only in the case of lines > 3 m.
e
DC POWER PORTS intended to be connected to a low voltage DC supply (≤60 V), where secondary circuits (isolated from the
AC mains supply) are not subject to transient overvoltages (i.e. reliably-grounded, capacitively-filtered DC secondary
circuits) shall be regarded as I/O signal/control PORTS.
SIST EN IEC 61326-1:2021/oprAA:2023
EN IEC 61326-1:2021/prAA:2023 (E)
88 Table 2 — Immunity test requirements for equipment intended
89 to be used in an INDUSTRIAL ELECTROMAGNETIC ENVIRONMENT
Performance
Port Phenomenon Basic standard Test value
criterion
Enclosure ESD EN 61000-4-2:2008 ±4 kV contact discharge B
±8 kV air discharge B
Electromagnetic field EN 61000-4-3:2006 + 10 V/m (80 MHz to 1 GHz) A
EN 61000-4-3:2006/A1:2008
3 V/m (1,4 GHz to 6 GHz) A
+
EN 61000-4-3:2006/A2:2010
Power frequency EN 61000-4-8:2010 30 A/m (50 Hz) A
b
magnetic field
AC power Burst EN 61000-4-4:2012 ±2 kV (5 kHz) B
(including protective
Surge EN 61000-4-5:2014 + ±1 kV line-to-line B
earth)
EN 61000-4-5:2014/A1:2017
±2 kV line-to-ground B
Conducted RF EN 61000-4-6:2014 10 V (150 kHz to 80 MHz) A
Voltage dip EN IEC 61000-4-11:2020 0 % during 1 cycle B
c
40 % during 10/12 cycles C
c
70 % during 25/30 cycles C
c
Short interruptions EN IEC 61000-4-11:2020 0 % during 250/300 cycles C
d, e
DC power Burst EN 61000-4-4:2012 ±2 kV (5 kHz) B
(including protective
Surge EN 61000-4-5:2014 + ±1 kV line-to-line B
earth)
EN 61000-4-5:2014/A1:2017
±2 kV line-to-ground B
Conducted RF EN 61000-4-6:2014 10 V (150 kHz to 80 MHz) A
e d
I/O signal/control Burst EN 61000-4-4:2012 ±1 kV (5 kHz) B
(including functional
a
Surge EN 61000-4-5:2014 + ±1 kV line-to-ground B
earth)
EN 61000-4-5:2014/A1:2017
d
Conducted RF EN 61000-4-6:2014 10 V (150 kHz to 80 MHz) A
e d
I/O signal/control Burst EN 61000-4-4:2012 ±2 kV (5 kHz or 100 kHz) B
(connected directly to a
Surge EN 61000-4-5:2014 + ±1 kV line-to-line B
mains supply)
EN 61000-4-5:2014/A1:2017
±2 kV line-to-ground B
d
Conducted RF EN 61000-4-6:2014 10 V (150 kHz to 80 MHz) A
a
Only in the case of lines > 30 m (called LONG-DISTANCE LINES).
b
Only to equipment containing Reed relays or Hall effect circuits (called magnetically sensitive equipment).
c
For example, “25/30 cycles” means “25 cycles for 50 Hz test” or “30 cycles for 60 Hz test”.
d
Only in the case of lines > 3 m.
e
DC power PORTS intended to be connected to a low voltage DC supply (≤60 V), where secondary circuits (isolated from the
AC mains supply) are not subject to transient overvoltages (i.e. reliably-grounded, capacitively-filtered DC secondary
circuits) shall be regarded as I/O signal/control PORTS.
SIST EN IEC 61326-1:2021/oprAA:2023
EN IEC 61326-1:2021/prAA:2023 (E)
90 Table 3 — Immunity test requirements for equipment intended
91 to be used in a CONTROLLED ELECTROMAGNETIC ENVIRONMENT
Performance
Port Phenomenon Basic standard Test value
criterion
Enclosure ESD EN 61000-4-2:2008 ±4 kV contact discharge B
B
±8 kV air discharge
Electromagnetic field EN 61000-4-3:2006 + 1 V/m (80 MHz to 1 GHz; A
EN 61000-4-3:2006/A1:2008
1,4 GHz to 6 GHz)
+
EN 61000-4-3:2006/A2:2010
AC power Burst EN 61000-4-4:2012 ±1 kV (5 kHz) B
(including protective
Surge EN 61000-4-5:2014 + ±0,5 kV line-to-line B
earth)
EN 61000-4-5:2014/A1:2017 B
±1 kV line-to-ground
Conducted RF EN 61000-4-6:2014 1 V (150 kHz to 80 MHz) A
a, b
DC power Burst EN 61000-4-4:2012 ±1 kV (5 kHz) B
(including protective
earth)
Conducted RF EN 61000-4-6:2014 1 V (150 kHz to 80 MHz) A
a b
I/O signal/control Burst EN 61000-4-4:2012 ±0,5 kV (5 kHz) B
(including functional
b
Conducted RF EN 61000-4-6:2014 1 V (150 kHz to 80 MHz) A
earth)
a
DC power PORTS intended to be connected to a low voltage DC supply (≤60 V), where secondary circuits
(isolated from the AC mains supply) are not subject to transient overvoltages (i.e. reliably-grounded, capacitively-filtered DC
secondary circuits) shall be regarded as I/O signal/control PORTS.
b
Only in the case of lines > 3 m.
92 NOTE In general, analysis, test and service laboratories have controlled electromagnetic environments, and personnel in
93 these areas are usually trained to interpret results. Such environments typically contain equipment which requires protection
94 by devices such a
...